We thank you for purchasing a Thermcraft, Inc. heat treating furnace.
Since our establishment in 1971, Thermcraft, Inc., has manufactured reliable,
high quality, electric resistance and gas equipment. Our products are used in a
variety of applications both domestically and around the world.
For over 40 years, Thermcraft, Inc. has manufactured furnaces and ovens
to comply with predetermined custom and industrial specifications. This furnace
has been completely checked for mechanical and electrical compliance prior to
shipment.
We trust you have received your furnace in acceptable condition and that
you will find it meets or exceeds your expectations and requirements.
Prior to installation and operation of your new Thermcraft, Inc. furnace, we
strongly urge you to read this manual it its entirety and comply with all
instructions herein.
If you have any questions, feel free to contact us.
Upon receipt of your new furnace, inspect for visible exterior damage.
Note and describe any damage found on the freight bill and file a claim with the
carrier.
Once unpacked, carefully inspect for concealed loss or damage on the
unit itself, both the interior and exterior. If necessary, the carrier will arrange for
official inspection to substantiate your claim.
Verify that all of the equipment shown on the packing slip is included with
the unit. Carefully check all packing materials.
RETURN SHIPMENT
Save the shipping box/crate until you are sure all is well. If for any reason
you must return the unit, first contact our Sales Department at 336-784-4800.
Please have the unit’s serial number and model number available when you call
(located on the rating plate). Note: all returns MUST have a Return Authorization
Number.
3.) ASSEMBLY
The furnace will arrive pre-assembled; however, some assembly may be
required prior to start-up.
NOTE: These furnaces are heavy and care should be taken to use lifting
devices which are sufficiently rated for these loads. Doors, handles and knobs
are not adequate for lifting or stabilization. The unit should be completely
restrained to prevent tipping during lifting and transport.
For larger furnaces some parts such as doors, door tracks, car bottom
tracks, elevator lifts, blower motors, and conveyors may have been disassembled
from the furnace to accommodate shipping. Before proceeding with installation,
review all of drawings and diagrams included in this manual.
1. Refer to designs, making sure all furnace stands, control boxes,
terminal boxes, doors, etc., are in place and secure. If not, locate
the appropriate parts that are included in the shipment and follow
the furnace design drawings in order to complete the assembly.
2. Locate the thermocouple hole and carefully pull out the alignment
pin. Carefully slide thermocouple into the hole, do not force.
If any questions arise concerning assembly of the furnace, please contact
Thermcraft, Inc. at 336-784-4800.
4.) CONTROL ENVIRONMENTAL CONDITIONS
Normal Operating Conditions
1. Ambient temperature: 0 – 50°C
2. Rate of change in temperature: 10°C/h or less
3. Ambient humidity: 20-90% RH (no condensation allowed)
4. Altitude: 2000m or less above sea level
Maximum Effects from Operating Conditions
1) Temperature effects
1.1. Thermocouple, DC mV and DC V input: +/- 2u V/°C or +/-0.02% of
F.S/°C, whichever is the larger
1.2. Resistance temperature detector: +/-0.05°C/°C
1.3. Analog output: +/-0.05% of F.S./°C
2) Effect from fluctuation of power supply voltage (within rated voltage range)
2.1. Analog input: +/- 0.2u V/V or +/-0.002% of F.S/V, whichever is the
larger
2.2. Analog output: +/-0.05% of F.S./V
Transportation and Storage Conditions
1) Temperature: -25 to 70°C
2) Humidity: 5 to 95% RH (no condensation allowed)
5.) INSTALLATION AND POWER CONNECTION
After the above assembly procedures have been completed, the furnace is
ready for connection to incoming power as follows:
1. Open the furnace and remove the protective packing covering the
heating elements.
2. Consult an electrical professional and the NEC (National Electric
Code) specifications to select wire size to adequately carry the line
amperage shown in section 1.) Furnace Specifications. Connect
electrical service to a fused power source disconnect. Such
disconnect device shall be marked as a disconnect device for the
furnace.
3. For overcurrent protection device selection, please refer to
amperage shown on the furnace rating plate.
NOTE: Be sure to check all external strain relief fittings for tightness. See
Torque Recommendation for Strain Relief Fittings chart below:
TORQUE RECOMMENDATIONS FOR STRAIN RELIEF FITTINGS
Torque in Inch Pounds - in lb (Newton Meters - Nm)
METAL STRAIN RELIEF FITTINGS PLASTIC STRAIN RELIEF FITTINGS
Note: The values shown above are for reference only.
Note: Nylon strain relief fittings installed in a threaded connection and exposed to elevated temperature
should be retained with a sealant such as Locktite® to avoid a loose connection when temperature fluctuates.
DOME NUTS
36.9 (4.17) 55.3 (6.25) 22.1 (2.50) 33.2 (3.75)
59.0 (6.67) 88.5 (10.00) 44.2 (5.00) 66.4 (7.50)
THREAD AND LOCK
NUTS
DOME NUTS
THREAD AND LOCK
NUTS
6.) OPERATIONAL SAFETY
.
Safe operation of your new furnace is mandatory. As manufacturer of your furnace, we at
Thermcraft, Inc. have adhered to all known safety standards and strongly suggest you install your
furnace in accordance with national electric and fire protection codes. Do not exceed the
electrical and temperature rating printed on the furnace rating plate.
It is important to pay particular attention to those regulations that are applicable to the
specific governing operations entities including, but not limited to legislated and accredited
national and local industry standards and the type of facility in which the furnace will be
employed, including zoning requirements, local building codes, and local electrical codes.
SYMBOLS USED ON THE EQUIPMENT
Caution: risk of electric shock.
Caution: refer to accompanying documents.
The potential for electric shock is always present when electrically operated equipment is
in use. The following suggestions are recommended for your protection:
1. Before beginning service / maintenance procedures on your furnace, the power source
must be locked out and tagged out per Occupational Safety and Health Administration
(OSHA) regulations.
2. Insure the furnace is properly grounded and electrically protected. Grounding lugs and
ground wires have been installed prior to shipment. The ground wire will terminate within
the external terminal box and is visible to the installer / operator. If unsure of the
grounding status of the equipment, consult the NEC (National Electric Code) or a
licensed professional.
3. If electrically conductive material is to be heat treated, the operator must be protected
from becoming a conductor to the ground. To avoid electric shock observe the following
operating practices:
a. Wear insulated gloves specific to the task that guard against electric shock if the
operator will be in contact with an electric current.
b. When using any metallic instrument to introduce items into the furnace, insure
the handle is properly insulated and the instrument is adequately grounded.
c. Install rubber mats in front of and behind the equipment to protect the operator.
Caution: hot surface.
C A U TI O N : If this equipment is used in a manner other than described in this
manual, protection provided by the equipment may be compromised or impaired.
To preclude hazard and minimize risk, follow all instructions and operate within
the design limits noted on the rating plate.
Do not use components or materials not specifically designed for this equipment.
Failure to comply with this precaution could result in damage to the equipment
used or the furnace and may create an overheat situation. Also, do not use
anything other than OEM exact replacement parts. Not using OEM replacement
parts could cause faulty instrumentation readings, inoperative equipment, or a
temperature overshoot.
Avoid combustible product which generate toxic or hazardous vapor or fumes.
Work should be done in a properly vented environment.
Observe the following precautions:
* Never stand in front of an open hot furnace.
* Wear protective eyewear.
* Wear protective gloves.
* Use tongs to insert and remove furnace load.
* Do not allow the load to touch the furnace walls.
WARNING: Do not use combustible gases directly in this furnace. Process
gasses must always be contained in a separate tube.
Modification of equipment for use other than that for which it is explicitly designed
for could cause severe injury or death. Any customer after market retrofit violates
the warranty.
Do not modify or disconnect any safety features provided. Disconnection of the
units safety features could allow the unit to become overheated and start on fire,
causing personal injury or death, product and property damage.
Periodically disconnect power from the equipment and allow it to cool completely. Once
cool, inspect for loose or broken heaters and for worn wire coils on the inside of the heated
chamber. Heaters may have to be replaced if damage is severe. Avoid contact with any
exposed heater coils / elements. Do not touch elements with bare hands or oily gloves. Contact
Thermcraft, Inc. if unsure of the safety of the heating elements in the conditions you have
observed.
Only qualified electrical professionals should remove the upper and lower terminal covers
of the terminal boxes. Keep all guards (guide-tabs, backstops, upper and lower terminal covers)
provided with the equipment in place while the furnace is in operation. Observe all safety labels.
If questions arise concerning the operation of your furnace, contact Thermcraft, Inc. at (336) 7844800, fax at (336) 784-0634 or email at sales@thermcraftinc.com.
7.) FURNACE START UP PROCEDURE
Follow your company’s lock-out/tag-out procedures before beginning any
work on the heaters or heated chamber of your furnace.
The following procedures should be performed by a trained electrical
professional due to potential hazard.
Check to ensure that the thermocouple is securely mounted and
undamaged. Check the thermocouple wiring connections.
CAUTION! Failure to check thermocouple wiring connection before initial
start up could result in damage to the furnace.
DRYING OUT PROCEDURE
1. It is imperative that the furnace is allowed to vent the moisture out of the
furnace. For split tube furnaces make sure that either the tube is slightly
open or that the vestibules are open for venting.
2. Heat furnace up to 300°F. This will dry the moisture out of the refractory
lining.
3. Maintain the furnace at 300°F for four to six hours or till all smoke/steam
has subsided.
4. Ramp the furnace up to 600°F and hold this temperature until all
steam/smoke has dissipated.
5. If the furnace is still steaming/smoking after 600F, repeat this step 4 at
900°F.
6. After all steam/smoke is gone, run furnace up to operating temperature.
7. If steam/smoke appears at any time during run up, do not increase
temperature until steaming/smoking stops.
Be careful when moving the furnace once the furnace has reached normal
operating temperatures. The insulation becomes brittle and could possibly
crack or break apart. With prolonged use, hairline cracks can develop in
the insulation material. These minor cracks will not affect the furnace’s
performance.
SHUT DOWN INSTRUCTIONS
Open main circuit breaker or disconnect switch (Note: the incoming side of the
main circuit breaker still has voltage, even after main device is open.)
8.) PREVENTIVE MAINTENANCE
CAUTION: Maintenance should only be performed by trained personnel.
WARNING: Prior to performing maintenance to the furnace or the
controls, the main power must be disconnected.
Prior to maintaining this equipment, read the applicable MSDS at the back
of the manual.
General Furnace Maintenance
1. Inspect and verify all electrical connections for tightness. Due to thermal
expansion during the heat- up/cool-down cycles electrical connection may
become loosened.
2. Inspect the heating elements. Do not allow build up of foreign material on
the element. Oil should be cleaned off of any part of the furnace prior to
heating. Use caution not to displace the heating elements as the wire may
be brittle and break easily.
3. If furnace has a door, verify that the door safety switch is operating and
cutting the power to the heaters once the door is opened.
4. If your furnace is equipped with a blower motor, or has moving parts,
lubricate grease fittings as required.
5. Check all hardware fasteners to ensure that they are tight.
6. Inspect the tip of the thermocouple for excessive heat deterioration.
General Control Cabinet Maintenance
1. Verify that all electrical connections are tight.
2. If applicable, clean the cabinet cooling fan filter.
3. Check all cabinet to furnace interconnection wiring.
4. If applicable, check to ensure that all signal lamps light.
5. If applicable, check to ensure that all alarms sound.
9.) THERMOCOUPLE
A thermocouple is used to measure the voltage that is generated as the
temperature increases. It provides an input signal for the control system that
monitors the temperature of the heated chamber.
At high temperatures, the elements within the thermocouple may undergo
some oxidation or corrosion causing the signals to be read incorrectly by the
control system.
To monitor temperature, If possible, the tip of the thermocouple must be
inside the heated chamber at least 1/8 deep. Also, the thermocouple tip must not
be blocked or shielded by items being heat treated.
If the control system indicates that a thermocouple is no longer
functioning, it is important to replace the malfunctioning thermocouple with the
same type model thermocouple that was originally provided with the furnace.
Replacement extension wires and connectors must also be compatible with the
thermocouple and the controller. If your system has a Eurotherm controller a
failed thermocouple will be indicated by SBR on the display and for a Yokogawa
controller it will indicate B.OUT. Refer to the replacement parts list to determine
the type of thermocouple required.
For optimal performance the thermocouple should be replaced one a year.
In some cases a more frequent replacement schedule may be required.
NOTE: Polarity of the thermocouple extension wires is the
OPPOSITE of the polarity of typical electrical wiring. The red leg is always
negative for thermocouples while the black leg is always positive.
10.) CERAMIC FIBER INSULATION SAFETY
Dear Valued Customer:
Please keep in mind that your Thermcraft, Inc. furnace is completely safe
when used for its intended purpose and with the precautions listed herein. The
following information is to make you aware of the potential health effects of the
ceramic and insulation components incorporated into your furnace.
Thermcraft, Inc. manufactures a broad range of high temperature furnace
systems. These systems incorporate various ceramic materials. Airborne
particulates produced in the handling of these materials should be considered
nuisance dust. Always wear a dust mask and safety glasses to avoid such
irritation due to inhalation of, or eye contact with, this or any other ceramic dust.
Chronic Effect – there has been no increased incidence of respiratory
disease in the Refractory Ceramic Fiber Coalition’s studies examining
occupationally exposed workers. In their animal testing, long-term laboratory
exposure to doses hundreds of times higher than normal occupational exposures
has produced fibrosis, lung cancer, and mesothelioma in rats or hamsters. The
fibers used in those studies were specially sized to maximize rodent reparability.
Please go to the Refractory Ceramic Fiber Coalition’s website at www.rcfc.net for
up-to-date information.
The recommended handling procedures for these ceramic materials are
outlined on the next page (Section 9) as well as in the Fibercraft Material Safety
Data Sheets (MSDS) at the end of this manual. Please make sure this
information is available to all personnel who may be operating, handling or
repairing this furnace.
As always, if you have any questions or concerns, please feel free to
contact Thermcraft, Inc. at (336) 784-4800, fax at (336) 784-0634 or email at
sales@thermcraftinc.com.
Recommended Safe Handling Procedures For Ceramic Fiber Products
1. Minimize presence of airborne fiber at all times by avoiding applications of
pressurized air from air canisters or pneumatic systems to any fibers that
have already been heated.
2. Wear an appropriate NIOSH – or MSHA-approved high efficiency air
purifying respirator mask when handling any ceramic fiber products.
3. Wear long-sleeved, loose-fitting clothing when handling ceramic fiber
products. Use protective coveralls over clothing. Do not wear contact
lenses and always wear safety glasses. Do not allow employees to take
soiled clothing, or any clothing in which fibers have become embedded
from the facility where the furnace is employed to his/her dwelling. Have
employees store, maintain and wash work clothing on site separately from
other clothing. Rinse washing machine thoroughly after washing clothing
worn when handling ceramic fibers.
4. Wear eye protection (safety glasses or goggles) and protective gloves at
all times.
5. Wash exposed skin areas gently with soap and cold water immediately
after handling ceramic fiber product.
6. Particular care should be taken when working with “used” material which
has been in service at elevated temperatures (greater than 1600° F).
Such products may undergo partial conversion to Cristobalite, a form of
crystalline silica that can cause respiratory disease.
11.) REPLACEMENT PARTS LIST
Description Quantity Required
FH-1212393-L 1
(Lower heating element, includes thermocouple)
CHEMTREC Assist: CHEMTREC will provide assistance for chemical emergencies.
Call 1-800-424-9300
2. COMPOSITION / INFORMATION ON INGREDIENTS
COMPONENTS CAS NUMBER % BY WEIGHT
Refractories, Fibers, Aluminosilicate 142844-00-6 100
(See Section 8 "Exposure Controls / Personal Protection" for exposure guidelines)
3. HAZARDS IDENTIFICATION
EMERGENCY OVERVIEW
WARNING!
POSSIBLE CANCER HAZARD BY INHALATION.
(See Section 11 for more information)
CHRONIC EFFECT
There has been no increased incidence of respiratory disease in studies examining
occupationally exposed workers. In animal studies, long-term laboratory exposure to doses
hundreds of times higher than normal occupational exposures has produced fibrosis, lung cancer,
and mesothelioma in rats or hamsters. The fibers used in those studies were specially sized to
maximize rodent respirability.
OTHER POTENTIAL EFFECTS
TARGET ORGANS:
Respiratory Tract (nose & throat), Eyes, Skin
RESPIRATORY TRACT (nose & throat) IRRITATION:
If inhaled in sufficient quantity, may cause temporary, mild mechanical irritation to respiratory
tract.Symptoms may include scratchiness of the nose or throat, cough or chest discomfort.
EYE IRRITATION:
May cause temporary, mild mechanical irritation. Fibers may be abrasive; prolonged contact may
cause damage to the outer surface of the eye.
SKIN IRRITATION:
May cause temporary, mild mechanical irritation. Exposure may also result in inflammation, rash
or itching.
GASTROINTESTINAL IRRITATION:
Unlikely route of exposure.
MEDICAL CONDITIONS AGGRAVATED BY EXPOSURE:
Pre-existing medical conditions, including dermatitis, asthma or chronic lung disease may be
aggravated by exposure; individuals who have a history of allergies may experience greater
amounts of skin and respiratory irritation.
HAZARD CLASSIFICATION
Although studies, involving occupationally exposed workers, have not identified any increased
incidence of respiratory disease, results from animal testing have been used as the basis for
hazard classification. In each of the following cases, the conclusions are qualitative only and do
not rest upon any quantitative analysis suggesting that the hazard actually may occur at current
occupational exposure levels.
In October 2001, the International Agency for Research on Cancer (IARC) confirmed that
Group 2b (possible human carcinogen) remains the appropriate IARC classification for RCF.
The Seventh Annual Report on Carcinogens (1994), prepared by the National Toxicology Program (NTP), classified respirable RCF as "reasonably anticipated" to be a carcinogen.
The American Conference of Governmental Industrial Hygienists (ACGIH) has classified
RCF as “A2-Suspected Human Carcinogen.”
The Commission of The European Communities (DG XI) has classified RCF as a substance
that should be regarded as if it is carcinogenic to man.
The State of California, pursuant to Proposition 65, The Safe Drinking Water and Toxic
Enforcement Act of 1986, has listed "ceramic fibers (airborne fibers of respirable size)" as a
chemical known to the State of California to cause cancer.
The Canadian Environmental Protection Agency (CEPA) has classified RCF as "probably
carcinogenic" (Group 2).
The Canadian Workplace Hazardous Materials Information System (WHMIS) – RCF is
classified as Class D2A – Materials Causing Other Toxic Effects
The Hazardous Materials Identification System (HMIS) –
Health 1* Flammability 0 Reactivity 0 Personal Protection Index: X (Employer Determined)
(* denotes potential for chronic effects)
4. FIRST AID MEASURES
FIRST AID PROCEDURES
RESPIRATORY TRACT (nose & throat) IRRITATION:
If respiratory tract irritation develops, move the person to a dust free location. Get medical
attention if the irritation continues. See Section 8 for additional measures to reduce or eliminate
exposure.
EYE IRRITATION:
If eyes become irritated, flush immediately with large amounts of lukewarm water for at least 15
minutes.
Eyelids should be held away from the eyeball to ensure thorough rinsing. Do not rub eyes. Get
medical attention if irritation persists.
SKIN IRRITATION:
If skin becomes irritated, remove soiled clothing. Do not rub or scratch exposed skin. Wash area
of contact thoroughly with soap and water. Using a skin cream or lotion after washing may be
helpful.
GASTROINTESTINAL IRRITATION:
If gastrointestinal tract irritation develops, move the person to a dust free environment.
NOTES TO PHYSICIANS:
Skin and respiratory effects are the result of temporary, mild mechanical irritation; fiber exposure
does not result in allergic manifestations.
NFPA Unusual Hazards: None
Flammable Properties: None
Flash Point: None
Hazardous Decomposition Products: None
Unusual Fire and Explosion Hazard: None
Extinguishing Media: Use extinguishing media suitable for type of surrounding fire.
6. ACCIDENTAL RELEASE MEASURES
SPILL PROCEDURES
Minimize creating airborne dust. Dust suppressing cleaning methods such as wet sweeping or
vacuuming should be used to clean the work area. If vacuuming, the vacuum must be equipped
with a HEPA filter. Compressed air or dry sweeping should not be used for cleaning.
7. HANDLING AND STORAGE
STORAGE
Store in original container in a dry area. Keep container closed when not in use.
HANDLING
Handle ceramic fiber carefully. Limit use of power tools unless in conjunction with local exhaust.
Use hand tools whenever possible. Frequently clean the work area with HEPA filtered vacuum or
wet sweeping to minimize the accumulation of debris. Do not use compressed air for clean-up.
EMPTY CONTAINERS
Product packaging may contain residue. Do not reuse.
* There is no specific regulatory standard for RCF in the U.S. OSHA’s “Particulate Not Otherwise
Regulated (PNOR)” standard [29 CFR 1910.1000, Subpart Z, Air Contaminants] applies
generally; Total Dust 15 mg/m³;Respirable Fraction 5 mg/m³.
** The Refractory Ceramic Fibers Coalition (RCFC) has sponsored comprehensive toxicology and
epidemiology studies to identify potential RCF-related health effects [see Section 11 for more
details],
consulted experts familiar with fiber and particle science, conducted a thorough review of the
RCF-related scientific literature, and further evaluated the data in a state-of-the-art quantitative
risk assessment. Based on these efforts and in the absence of an OSHA PEL, RCFC has
adopted a recommended exposure guideline, as measured under NIOSH Method 7400 B. The
manufacturers’ REG is intended to promote occupational health and safety through prudent
exposure control and reduction and it reflects relative technical and economic feasibility as
determined by extensive industrial hygiene monitoring efforts undertaken pursuant to an
agreement with the U.S. Occupational Safety and Health Administration (OSHA).
0.5 f/cc. The objectives and criteria underlying each of these OEL decisions also vary. The
evaluation of occupational exposure limits and determining their relative applicability to the
workplace is best performed, on a case-by case basis, by a qualified Industrial Hygienist.
ENVIRONMENTAL CONTROLS
Use environmental controls such as local exhaust ventilation, point of generation dust collection,
down draft work stations, emission controlling tool designs, and materials handling equipment
designed to minimize airborne fiber emissions.
PERSONAL PROTECTION EQUIPMENT
Respiratory Protection – RCF:
When engineering and/or administrative controls are insufficient to maintain workplace
concentrations within the 0.5 f/cc REG, the use of appropriate respiratory protection, pursuant to
the requirements of OSHA Standards 29 CFR 1910.134 and 29 CFR 1926.103, is recommended.
The following information is provided as an example of appropriate respiratory protection for
aluminosilicate fibers. The evaluation of workplace hazards and the identification of appropriate
respiratory protection is best performed, on a case by case basis, by a qualified Industrial
Hygienist.
MANUFACTURER’S RESPIRATORY PROTECTION RECOMMENDATIONS
WHEN HANDLING RCF PRODUCTS
Not yet determined but expected to be below A respirator with a filter efficiency of at least 95%
5.0 f/cc based on operation
"Reliably" less than 0.5 f/cc Optional
0.5 f/cc to 5.0 f/cc A single use respirator or half-face, air purifying
respirator with a filter efficiency of at least 95%
5.0 f/cc to 25 f/cc Full-face piece, air purifying respirator equipped
with a NIOSH certified particulate filter cartridge
with a filter efficiency of at least 95% or PAPR
Greater than 25 f/cc PAPR with tight-fitting full face piece or a
supplied air respirator in continuous flow mode
When individual workers request respiratory A NIOSH certified respirator, such as a single
use Protection as a matter of personal comfort particulate respirator with a filter efficiency of a
least or choice where exposures are “reliably” 95%
below .0.5 f/cc
The 95% filter efficiency recommendation is based on NIOSH respirator selection logic
†
sequence for exposure to particulates. Selection of filter efficiency (i.e. 95%, 99% or 99.9 7%)
depends on how much filter leakage can be accepted. Higher filter efficiency means lower filter
leakage. Other factors to consider are the NIOSH filter series N, R or P. (N) Not resistant to oil,
(R) Resistant to oil and (P) oil Proof. These recommendations are not designed to limit informed
choices, provided that respiratory protection decisions comply with 29 CFR 1910.134.
Other Information
-Concentrations based upon an eight hour time weighted average (TWA) as determined by air
samples collected and analyzed pursuant to NIOSH method 7400 (B) for airborne fibers.
-The manufacturer recommends the use of a full face piece air purifying respirator equipped with
an appropriate particulate filter cartridge during furnace tear out events and the removal of used
RCF to control exposures to airborne fiber and the potential presence of crystalline silica. If
exposure levels are known, the respiratory protection charge provided above may be applied. Potential exposure to other airborne contaminants should be evaluated by a qualified industrial
hygienist for the selection of appropriate respiratory protection and air monitoring.
Skin Protection:
Wear gloves, head coverings and full body clothing as necessary to prevent skin irritation.
Washable or disposable clothing may be used. If possible, do not take unwashed clothing home.
If soiled work clothing must be taken home, employers should ensure employees are thoroughly
trained on the best practices to minimize non-work dust exposure (e.g., vacuum clothes before
leaving the work area, wash work clothing separately, rinse washer before washing other
household clothes, etc.).
Eye Protection:
Wear safety glasses with side shields or other forms of eye protection in compliance with
appropriate OSHA standards to prevent eye irritation. The use of contact lenses is not
recommended, unless used in conjunction with appropriate eye protection. Do not touch eyes
with soiled body parts or materials. If possible, have eye-washing facilities readily available where
eye irritation can occur.
9. PHYSICAL AND CHEMICAL PROPERTIES
ODOR AND APPEARANCE: White, odorless, fibrous material
CHEMICAL FAMILY: Vitreous Aluminosilicate Fibers
BOILING POINT: Not Applicable
WATER SOLUBILITY (%): Not Soluble in Water
MELTING POINT: 1760° C (3200° F)
SPECIFIC GRAVITY: 2.50 – 2.75
VAPOR PRESSURE: Not Applicable
pH: Not Applicable
VAPOR DENSITY (Air = 1): Not Applicable
% VOLATILE: Not Applicable
MOLECULAR FORMULA: Not Applicable
10. STABILITY AND REACTIVITY
CHEMICAL STABILITY: Stable under conditions of normal use.
INCOMPATIBILITY: Soluble in hydrofluoric acid, phosphoric acid, and
concentrated alkali.
CONDITIONS TO AVOID: None.
HAZARDOUS DECOMPOSITION
PRODUCTS: None.
HAZARDOUS POLYMERIZATION: Not Applicable.
11. TOXICOLOGICAL INFORMATION
HEALTH DATA SUMMARY
Epidemiological studies of RCF production workers have indicated no increased incidence
of respiratory disease nor other significant health effects. In animal studies, long-term,
high-dose inhalation exposure resulted in the development of respiratory disease in rats
and hamsters.
EPIDEMIOLOGY
In order to determine possible human health effects following RCF exposure, the University of
Cincinnati in the United States and the Institute of Occupational Medicine (IOM) in Europe have
conducted medical surveillance studies on RCF workers in U.S. and European manufacturing
facilities. The University of Cincinnati study has been in progress for over 20-years, collecting
data from respiratory questionnaires, lung function tests, and chest X-rays, exposure monitoring,
and worker mortality.
The results of this study of RCF plant workers exposed from 1953 to the present have shown
(LeMasters et al, 2003): No excess mortality related to all deaths, all cancers, or lung cancer No
statistically significant increase in interstitial findings (fibrosis), and
No mesotheliomas or increase in lung cancer The initial cross-sectional spirometry studies in the
U.S. (LeMasters et al.1998) and Europe (Cowie et al.2001) revealed lung function decrements in
the RCF-exposed cohort that were associated with heavier historical exposures. Subsequently,
longitudinal studies have revealed no RCF exposure related decrements in lung function
associated with current exposure levels.
Through 1996, pleural plaques seen on chest X-rays in 2.7% of the workers. Pleural plaques are
considered a marker of exposure and not disease. The prevalence of pleural plaques has
remained relatively constant over time, perhaps as a result of lower current exposure levels.
Thus, this long term epidemiology study has demonstrated an absence of interstitial fibrosis, no
increased mortality risk and no decrement in lung function associated with current exposures.
TOXICOLOGY
Early animal studies of RCF effects by intraperitoneal and intrapleural injections, as well as by
inhalation, resulted in mostly negative results. In an effort to eliminate any questions posed by the
results of these early studies, a definitive Maximum Tolerated Dose Study (MTD) by nose only,
lifetime inhalation in rats and hamsters, was designed in the 1980s. The MTD study appeared to
confirm that RCF was an animal carcinogen under certain test conditions, e.g., extremely high
concentrations of approximately 200 f/cc inhaled directly into the lungs.
A later review of the MTD pathology indicated that the animals’ lungs were likely “overloaded”
because of large quantities of non-fibrous particles, and that this overload condition was likely
responsible for the disease observed. In fact, evaluation of the aerosol samples used confirmed
the presence of significant quantities of particulate matter.
In a subsequent multi-dose animal inhalation study at 25 f/cc, 75 f/cc, and 115 f/cc; a no observed effect level (NOEL) was found at 25 f/cc. This level is 50 times the RCFC recommended REG of
0.5 f/cc for
humans.
12. ECOLOGICAL INFORMATION
No ecological concerns have been identified.
13. DISPOSAL CONSIDERATIONS
WASTE MANAGEMENT
To prevent waste materials from becoming airborne during waste storage, transportation and
disposal, a covered container or plastic bagging is recommended.
DISPOSAL
RCF, as manufactured, is not classified as a hazardous waste according to Federal regulations
(40 CFR 261). Any processing, use, alteration or chemical additions to the product, as purchased,
may alter the disposal requirements. Under Federal regulations, it is the waste generator's
responsibility to properly characterize a waste material, to determine if it is a "hazardous" waste.
Check local, regional, state or provincial regulations to identify all applicable disposal
requirements.
14. TRANSPORT INFORMATION
U.S. DEPARTMENT OF TRANSPORTATION (DOT)
Hazard Class: Not Regulated United Nations (UN) Number: Not Applicable
Labels: Not Applicable North America (NA) Number: Not Applicable
Placards: Not Applicable Bill of Lading: Product Name
INTERNATIONAL
Canadian TDG Hazard Class & PIN: Not regulated
Not classified as dangerous goods under ADR (road), RID (train) or IMDG (ship).
15. REGULATORY INFORMATION
UNITED STATES REGULATIONS
EPA: Superfund Amendments and Reauthorization Act (SARA) Title III – This
product does not contain any substances reportable under Sections 302, 304,
313, (40 CFR 372). Sections 311 and 312 (40 CFR 370) apply (delayed hazard).
Toxic Substances Control Act (TSCA) - RCF has been assigned a CAS
number; however, it is an "article" under TSCA and therefore exempt from listing
on the TSCA inventory.
Comprehensive Environmental Response, Compensation and Liability Act
(CERCLA) and the Clean Air Act (CAA) - RCF contains fibers with an average
diameter greater than one micron and thus is not considered a hazardous air
pollutant.
OSHA: Comply with Hazard Communication Standards 29 CFR 1910.1200 and 29
CFR 1926.59 and the Respiratory Protection Standards 29 CFR 1910.134 and
29 CFR 1926.103.
Ceramic fibers (airborne particles of respirable size)” is listed in Proposition 65,
The California: Safe Drinking Water and Toxic Enforcement Act of
1986 as a chemical known to the State of California to cause cancer.
Other RCF products are not known to be regulated by states other than California
States: however, state and local OSHA and EPA regulations may apply to these
products. If in doubt, contact your local regulatory agency.
INTERNATIONAL REGULATIONS
Canada: Canadian Workplace Hazardous Materials Information System (WHMIS) –
RCF is classified as Class D2A – Materials Causing Other Toxic Effects
Canadian Environmental Protection Act (CEPA) - All substances in this
product are listed, as required, on the Domestic Substance List (DSL)
European European Directive 97/69/EC classified RCF as a Category 2 carcinogen; that
is it “should be regarded as if it is carcinogenic to man.”
16. OTHER INFORMATION
RCF DEVITRIFICATION
As produced, all RCF fibers are vitreous (glassy) materials which do not contain crystalline silica.
Continued exposure to elevated temperatures may cause these fibers to devitrify (become
crystalline). The first crystalline formation (mullite) begins to occur at approximately 985° C
(1805° F). Crystalline phase silica may begin to form at temperatures of approximately 1200° C
(2192° F). When the glass RCF fibers devitrify, they form a mixed mineral crystalline silica
containing dust. The crystalline silica is trapped in grain boundaries within a matrix predominately
consisting of mullite. The occurrence and extent of crystalline phase formation is dependent on
the duration and temperature of exposure, fiber chemistry and/or the presence of fluxing agents.
The presence of crystalline phases can be confirmed only through laboratory analysis of the "hot
face" fiber.
IARC’s evaluation of crystalline silica states “Crystalline silica inhaled in the form of quartz or
cristobalite from occupational sources is carcinogenic to humans (Group 1)” and additionally
notes “carcinogenicity in humans was not detected in all industrial circumstances studied.” IARC
also studied mixed mineral crystalline silica containing dusts such as coal dusts (containing 5 –
15 % crystalline silica) and diatomaceous earth without seeing any evidence of disease. (IARC
Monograph Vol. 68, 1997). NTP lists all polymorphs of crystalline silica amongst substances
which may "reasonably be anticipated to be carcinogens".
IARC and NTP did not evaluate after-service RCF, which may contain various crystalline phases.
However, an analysis of after-service RCF samples obtained pursuant to an exposure monitoring
agreement with the USEPA, found that in the furnace conditions sampled, most did not contain
detectable levels of crystalline silica. Other relevant RCF studies found that (1) simulated afterservice RCF showed little, or no, activity where exposure was by inhalation or by intraperitoneal
injection; and (2) after-service RCF was not cytotoxic to macrophage-like cells at concentrations
up to 320 g/cm² - by comparison, pure quartz or cristobalite were significantly active at much
lower levels (circa 20 g/cm²).
RCF AFTER-SERVICE REMOVAL
Respiratory protection should be provided in compliance with OSHA standards. During removal
operations, a full face respirator is recommended to reduce inhalation exposure along with eye
and respiratory tract irritation. A specific evaluation of workplace hazards and the identification of
appropriate respiratory protection is best performed, on a case by case basis, by a qualified
industrial hygiene professional.
PRODUCT STEWARDSHIP PROGRAM
The Refractory Ceramic Fibers Coalition (RCFC) and the U.S. Occupational Safety and Health
Administration (OSHA) introduced a voluntary worker protection program entitled PSP HTW (High
Temperature Wools), a comprehensive, multi-faceted risk management program designed to
control and reduce workplace exposures to refractory ceramic fiber (RCF). For more information
regarding PSP HTW, please refer to the RCFC web site: http://www.rcfc.net.
MSDS Prepared By: RISK MANAGEMENT DEPARTMENT
DISCLAIMER
The information presented herein is presented in good faith and believed to be accurate as of the
effective date of this Material Safety Data Sheet. Employers may use this MSDS to supplement
other information gathered by them in their efforts to assure the health and safety of their
employees and the proper use of the product. This summary of the relevant data reflects
professional judgment; employers should note that information perceived to be less relevant has
not been included in this MSDS. Therefore, given the summary nature of this document,
Thermcraft, Inc. does not extend any warranty (expressed or implied), assume any responsibility,
or make any representation regarding the completeness of this information or its suitability for the
purposes envisioned by the user.
DEFINITIONS
ACGIH: American Conference of Governmental Industrial Hygienists
ADR: Carriage of Dangerous Goods by Road (International Regulation)
CAA: Clean Air Act
CAS: Chemical Abstracts Service
CERCLA: Comprehensive Environmental Response, Compensation and Liability Act
DSL: Domestic Substances List
EPA: Environmental Protection Agency
EU: European Union
f/cc: Fibers per cubic centimeter
HEPA: High Efficiency Particulate Air
HMIS: Hazardous Materials Identification System
HTW: High Temperature Wools
IARC: International Agency for Research on Cancer
IATA: International Air Transport Association
IMDG: International Maritime Dangerous Goods Code
mg/m³: Milligrams per cubic meter of air
mmpcf: Million particles per cubic meter
NFPA: National Fire Protection Association
NIOSH: National Institute for Occupational Safety and Health
OSHA: Occupational Safety and Health Administration
29 CFR 1910.134 & 1926.103: OSHA Respiratory Protection Standards
29 CFR 1910.1200 & 1926.59: OSHA Hazard Communication Standards
PEL: Permissible Exposure Limit (OSHA)
PIN: Product Identification Number
PNOC: Particulates Not Otherwise Classified
PNOR: Particulates Not Otherwise Regulated
PSP: Product Stewardship Program
RCFC: Refractory Ceramic Fibers Coalition
RCRA: Resource Conservation and Recovery Act
REG: Recommended Exposure Guideline (RCFC)
REL: Recommended Exposure Limit (NIOSH)
RID: Carriage of Dangerous Goods by Rail (International Regulations)
SARA: Superfund Amendments and Reauthorization Act
SARA Title III: Emergency Planning and Community Right to Know Act
SARA Section 302: Extremely Hazardous Substances
SARA Section 304: Emergency Release
SARA Section 311: MSDS/List of Chemicals and Hazardous Inventory
SARA Section 312: Emergency and Hazardous Inventory
SARA Section 313: Toxic Chemicals and Release Reporting
STEL: Short Term Exposure Limit`
SVF: Synthetic Vitreous Fiber
TDG: Transportation of Dangerous Goods
TLV: Threshold Limit Value (ACGIH)
TSCA: Toxic Substances Control Act
TWA: Time Weighted Average
WHMIS: Workplace Hazardous Materials Information System (Canada)
14.) CONTROL SYSTEM OPERATION
The temperature control system controls the amount of power to the resistive
heating elements in order to achieve and maintain the desired temperature as
measured through the control thermocouple.
The temperature control system consists of the following:
1. Setpoint temperature controller
2. Solid state power controller or mechanical contactor
3. Matching control thermocouple
Heat/cool control systems provide heating control to the unit in addition to cooling
control.
Additional optional features include:
1. Over-temperature controller
2. Current limit SCR power controller
3. Multiple zone independent control
INSTALLATION
1. Position control cabinet in the desired location making sure it is
permanently secured. NOTE: Do not position controller in such a manner
as to be difficult to operate disconnect.
2. Study all instrument manuals before energizing the control system and
furnace.
3. Review all electrical wiring diagrams.
4. Make all external wire connections specified in the electrical wiring
diagrams including connection of the cables between the furnace and the
control console. All conduit supplied shall be permanently secured /
mounted during the installation process. CAUTION: Control system must
be grounded according to local wiring requirements.
5. Connect the thermocouple to the control system.
6. Close main circuit breaker or disconnect switch.
7. Turn on main power.
8. Check red and green indicating lights on control cabinet and temperature
controller for control power indication.
9. Set temperature controller and over temperature controller to desired
temperature setting. (Check instrument control manuals for fine-tuning).
10. Turn on control cabinet power. The furnace will continue to heat until the
controller setpoint is reached. The controller will start cycling to maintain
selected setpoint.
ɸ
ɸ