HaloguardTM monitors are carefully packed, inspected and delivered to the carrier in good condition. If
damage occurs in transit it is the responsibility of the carrier. Carefully inspect the unit upon receipt.
Any damage should be reported to the carrier and an inspection requested. After inspection by the
carrier and your receipt of his acknowledgment as to the damage, contact us for return authorization
(RMA). We cannot file claims for damaged goods on your behalf, but we will assist you in any way we
can. In order to expedite our work, please provide serial number and model number when you call.
WE CANNOT ACCEPT RETURNS WITHOUT PRIOR APPROVAL.
BEFORE INSTALLATION
1. Remove Haloguard IR from carton. Check contents against packing list. If shipment is incomplete,
contact us immediately.
2. Select a site for the installation:
a. Power requirements: Provide a clean source of power 115-230 VAC, 50/60 Hz, single phase 15A
minimum.
b. Sample tube pick-up should be located near potential leaks in a quiet area or downstream from
leak source in area with air movement.
c. Sample tube pick-up should be located 18” - 24” above the floor. If using optional scanner, locate
sample tubes at points near all chiller locations. Make sure each sample tube is equipped with an
end of line filter.
d. Make sure the area selected is not subject to flooding, potential impact or severe ambient
temperature and humidity changes (i.e. boiler blow-down or near roll-up doors).
e. Narrow band infrared detector is highly selective but may respond to the presence of other
refrigerants if interfering gas reaches a very high concentration. For accurate leak detection
select a location free from interfering vapors. Contact us for specific gas interference.
3. The Haloguard IR should be installed indoors, about five feet above the floor and at a location easily
visible to operators, in an area with minimal vibration, and with temperature and humidity changes
like sample pick-up location. DO NOT MOUNT DIRECTLY TO CHILLER
Remote Alar
Haloguard
Panel and
Local Alarm
Fig. 1-Typical Area Monitoring Installation
Rev.1/26/01
Sample poin
Chillers
Sample poin
Remote Alarm
Sample poin
Refrigerant
Storage Area
3
Fan
IMPORTANT
)
p
CONSULT LOCAL AND NATIONAL ELECTRICAL CODES FOR ANY SPECIAL
REQUIREMENTS OR RESTRICTIONS BEFORE INSTALLING HALOGUARD
FUNCTION OF LED’S, PUSHBUTTONS, and LCD DISPLAY INFORMATION
1. READY LED-Steady LED means Haloguard IR is measuring ppm; blinking indicates otherwise
2. POWER LED- Green LED indicates Haloguard IR is receiving power.
3. FAULT LED – LED indicates malfunction. LCD indicates specific type of failure (See Figure 3)
4. ALARM 1, ALARM 2, ALARM 3 LED’s-Adjustable Alarm levels & relays - Factory set for
particular refrigerant (See Figure 12 ). User adjustable as follows:
RANGE
1 PPM
INCREMENTS
10 PPM
INCREMENTS
0 - 1000 ppm10 - 100100 - 1000
5. MANUAL SCAN push-button
6. RESET/SILENCE push-button - Silences alarms, disables relays, and resets instrument for 20
minutes. LCD displays RESET/SILENCE with 20 minute countdown. Alarms are automatically
reactivated if alarm condition is not corrected. Reset is automatically initiated on start-up and
after power failure.
7. Optional Display Features: TWA Integration – Haloguard IR calculates an 8 hr. Time Weighted
Average (TLV-TWA), displays this value on LCD, activates Alarm 1 LED and relay if factory
set PPM value is exceeded. Used for gases with TLV-TWA less than 1000ppm. TLV-TWA
Alarm is factory set and not user adjustable.
-Zero Calibrate, connect scrubber,
follow on screen directions
-Span Calibrate, insert key from gas cal kit
Figure 3 - LCD Display Features
4
(4)
p
y
Reset/Silence
n
n
Pushbutto
Test
Pushbutto
Local Strobe Alarm-Opt.
LCD
Enclosure is 12” x 14” x 4.5”; NEMA 250 rated
Alarm 2
Mounting Holes
0.25”dia.
Sample Inlet
1 inlet Std.
2,4,8 inlet Opt
Inlet can be
extended 300’
using 0.125”
tubing. All lines
must have end
of line filter.
Power In
120/230Vac
50/60Hz
Rev.1/26/01
Relay/Binary Output
4x5A. dry contact Std
7x10A. dr
Local Audible alarm-Opt.
Analog/ Digital Output
4-20mA, 0-5VDC
contact Opt.
RS 232/485 serial O
Aux. Inputs and Remote Expansion
Module In
Fig. 4-Haloguard IR
5
Sample
Outlet
Remote
Reset In
t.
INSTALLATION
1. Check power supply and determine best location. (See Section “Before Installation”)
2. Disconnect power before beginning.
Refer to figures 2, 3, 4, 5 and 6 before proceeding.
3. Installation
a. Mounting Haloguard
IR
1. Securely mount the unit to a wall or support using the (4) mounting holes.
2. Remove cover.
3. Install sample tube to 18-24” above floor, install end of line filter. Sample inlet may be
extended up to 300FT with 0.125” OD x 0.093” ID nylon, copper or stainless steel tubing.
When extending sample tubing:
a. Keep sample lines as short as possible and free of kinks.
b. Exhaust to atmosphere. DO NOT install any device which might restrict the flow.
4. Route sample inlet tubing through entry indicated ( Figure 4 ) and insert into one-touch
fitting provided. To remove tubing pull plastic ring on end of one-touch fitting to release.
b. Wiring Connections:
1. Relay Terminals – Route wire through entry indicated. (Figure 4 ). Remove safety cover.
Connect alarm relays NO or NC (Figure 6) required.
2. Power Supply Wiring-Route wire through entry indicated. (Figure 4 ). Remove safety cover.
All connections are made to the terminal block labeled IN 120/250VAC (Figure 6). Connect
HOT AC wire to Terminal L. Connect NEUTRAL wire to terminal N. Connect a GROUND
wire to terminal G. Auxiliary 120/250VAC output is provided to activate remote audible and
strobe light alarms. This output is not fused. Limit load to 0.5A.max total.
**WARNING** analog output is a driven signal, do not connect any wire with voltage or
current to analog output terminals
3. Analog Output (Optional) – Two analog outputs are provided for connection to remote
recording devices (Figure 6); PPM analog output is for gas concentration; Channel analog
output is for channel location. Connect 18 ga. twisted shielded pair as follows, maximum
length 300 ft.:
Analog Output RangeTerminal Position
0-10VDCV and GND
4 - 20mAI and GND
Figure 5 - Analog Jumper Settings
4. Auxiliary Input(VDC) – Five analog inputs (EXT INPUT 3-8) are provided for connection
of remote input devices such as oxygen sensors, CO sensors, and flammables sensors
(Figure 6)
5. Auxiliary Input(mA) – Two analog inputs (Remote IR 1&2) are provided for connection of
remote input devices such as Haloguard Remote Expansion Module (Figure 6)
Rev.1/26/01
6
6. Uninterruptable Power Supply (Optional) – Make sure UPS is fully charged. Plug UPS into
power supply, turn switch ON. Connect Haloguard IR to UPS, with power cord provided
c. On power-up, RESET and 20 minute count down will appear on LCD. For 20 minutes after
power up all alarms and relays are automatically disabled. Whenever power is disrupted,
RESET mode will be automatic (unless optional UPS is supplied) to protect against false alarms.
Allow 1 hour for warm-up from cold start.
Rev.1/26/01
Figure 6 Circuit Board
7
SETUP and SYSTEM CONFIGURATION
Locate configuration jumpers on Figure 6. Select configuration settings from the following:
JPFUNCTION
RUN
SETUP
TEST
Select run for normal operation
Set gas type and number of active channels
Sequentially activates all alarm & Fault LED’s &
relays
ALARM
CAL
SPAN
JP5
Set Point Adjustment for Alarm 1, 2, and 3
Zero Calibration
For use only with span gas and key
Factory only
Figure 7 – Setup & Configuration Jumper Settings
1. RUN - Remove jumper to initiate other modes. Replace jumper when finished.
2. SETUP - Gas type will be pre-set at the factory. Number of active channels will be the
maximum possible for the instrument as supplied (Fig.12). To change gas type or number of
active channels use jumper to initiate the set-up mode. Select channel with cursor by using
the UP/DOWN/ENTER pushbuttons. Select gas type for each channel by using UP/DOWN
pushbuttons. “NONE” selected in the gas type menu will skip channel
3. TEST-Select test mode. Select alarm with cursor by using UP/DOWN/ENTER pushbuttons
4. ALARM - Alarm levels will be pre-set at the factory (Fig.12). To change select ALARM
with cursor using UP/DOWN/ENTER pushbuttons. Change PPM using UP/DOWN/ENTER
pushbuttons. OFF selected turns alarm off.
5. CAL – Install optional activated charcoal filter in channel 1, follow on screen directions to
zero calibrate using ENTER pushbutton.
6. SPAN –Key and instructions for use are included with Span Gas Calibration Kit
7. JP5- Factory only.
System Configuration:
1. Relay Setup - Factory recommended setting is unlatched mode; e.g., the alarm LED’s and
relays will return to a normal status when gas concentration goes below alarm setpoint. To
select latched mode place jumper on ALARM setting. Select LATCH mode with the cursor
using UP/DOWN/ENTER pushbuttons. LATCH ON appears on the LCD Display. Latched
mode requires manual RESET of all alarms
2. Analog Output (optional) is factory set at 0 - 5 VDC or 4 - 20mA as ordered. Use Analog
trim pot (Figure 6) to adjust for up to 0-10VDC range
4. Two, Four or Eight Point Scanner - (Optional) Scanner is factory installed. Scanner
automatically sequences through each port (1-2, 1-4 or 1-8).
a. Manual Scan - While in RUN Mode,
1. Press MAN SCAN button on side of controller. Each time the button is pressed,
display will advance to next sample point. Controller will return to automatic
sequencing within 2 minutes.
2. Adjust Scan Rate.-Rate is factory set at 4 seconds per point. Rate is adjustable from
3-10 seconds. To change scan display rate press ENTER from RUN mode; adjust
UP/DOWN; press ENTER to return to RUN mode
Rev.1/26/01
8
5. Installation of Optional Remote IR Expansion Module
(Op
)
un
Setup
Refer to FIGURE 4, 6, 8, and 9 before proceeding.
DO NOT CUT SUPPLIED CABLE
Controller is provided with 18” cable with DIN connector and expansion module is provided
with 6-1/2’ cable. If cable extension is required we recommend 3 conductor 18 ga. cable with
foil shield, Carol C2535 or equal. Use 3/4” conduit if required. Maximum 1000 ft. length
Remove 18” cable with DIN connector from controller circuit board. Splice or solder extension
cable to end of 18” cable. Reconnect extended sensor cable to terminal block on circuit board of
controller.
Provided
Terminal
CHARedRedScanner
SIGWhiteWhiteSignal
GNDBlk/BlkBlackGround
Note - CHA Terminal is used for expansion modules with scanner only.
To avoid nuisance alarms remove alarm jumper “AUD” before beginning.
To compensate for line losses adjust signal voltage with a digital VOM as follows:
1. Open cover on Remote IR expansion module. Move jumper from RUN to SETUP. Wait
approximately 1 minute before proceeding .
2. Open cover on Haloguard Controller .
3. With VOM on 5 VDC scale, place VOM probe on SIG (+) and GND (-) Test points on
Controller (see Figure 6).
Adjust potentiometer on Controller Circuit Board, with screwdriver until it reads 4.975
+/-0.025 VDC.
Display will show FAULT if out of range and SETUP, if trimmed to 4.975+/-0.025 VDC.
4. If Remote expansion module is provided with scanner it will be necessary to lock the scanner
on channel 1 using the jumper provided (Figure 9).
5. When done, replace expansion module jumper in RUN mode. Replace AUD jumper to
activate alarms circuit.
Scanner Circuit Board
Connector for Digital Calibrator
tional
Cable to Controller
(DO NOT REMOVE)
DO NOT ADJUST POTS
Rev.1/26/01
Red
White
Black
Figure 9 - Expansion Module Circuit Board
Lower Circuit
Board
R
9
IR Module Run/
Setup Jumper
CALIBRATION
Haloguard IR monitors are factory calibrated to the specific gas of interest. Field calibration is not
required for operation in typical machine room environments during the first twelve months of operation
unless a change in gas type is required.
For those who wish to check the factory calibration or change the gas type an optional Span Gas Kit is
available. Instructions for use are included with kit
TESTING
The Haloguard IRTM should be tested at start-up, after any changes to alarm protocol, and after each
alarm level condition has been corrected. During testing, display will indicate TEST MODE
a. Alarm Test –Enter TEST mode with jumper, select alarm with cursor by using
UP/DOWN/ENTER pushbuttons. Check all alarms sequentially. Alarm relays, lights, audible
and remote alarms (if equipped) will activate if selected.
b. Reset Test - Push RESET/SILENCE button; alarm relays, audible and remote alarms (if
equipped) will become inactive for 20 minutes. Display will read RESET and indicate minutes
to completion of reset period.
TROUBLESHOOTING
The most frequent cause of nuisance alarms and fault is improper set-up, primarily due to:
a. Failure to install in accordance with instructions.
b. Not allowing sufficient warm-up time.
c. Poor sensor location.
To check Zero Calibration, fill a clean plastic trash bag with clean outside air, tie off with twist-tie, let
sample stabilize inside for 15 - 20 minutes. Insert sample tube into bag . For models equipped with
scanner allow up to 2.5 minutes per channel for clean air sample analysis
.
Observe LCD response, if displayed PPM value drops lower then Haloguard IR is detecting the presence
of the gas of interest or another gas which absorbs infrared light at the same frequency.
Remove sample tube from bag and observe LCD response. If displayed PPM value moves higher, look
for refrigerant leaks or possible sources of interfering gases (See “Before Installation). and relocate
sample point.. Keep in mind that the Haloguard IR is much more sensitive than hand held leak detectors,
it may detect a leak when hand held units show no response
We wish to aid with any problems you might experience with the Haloguard IR
. When contacting
Thermal Gas Systems, Inc. Service Department, please have the following information ready
1. Model Number and Serial Number.
2. Description of the problem.
3. Remove cover. Place jumper on Alarm setting function and record settings. Place jumper on
Setup function and record settings
4. Advise the condition of all LED’s and the information displayed on the LCD.
Rev.1/26/01
10
MAINTENANCE
There is no scheduled maintenance for the first 12 months operation; after 12 months the internal and
external filters should be changed and the instrument should be gas tested in accordance with the
instructions. Self-diagnostics will indicate the cause of any unusual malfunction. To obtain spare or
replacement parts please contact Thermal Gas Systems, Inc.
Fault Diagnostics: Indicator Light and LCD Readout
Enclosure: NEMA 250, NEMA4/12 optional
Weight: 20 lbs.
Dimensions: 12” W x 14” H x 4.5” D
Sample Rate 1.5Liter/min.
Response Time:< 1 Minute single point
o
+40
F to +120oF Non-condensing
(See Figure 5)
< 2.5 Minutes per point with optional
scanner.
Part No.
Initial factory settings are:
FunctionSetting
Channel
1 2 3 4 5 6 7 8
Gas
Alarm 1
Alarm 2
Alarm 3
Figure 12 - Initial Factory Settings
Rev.1/26/01
GAS
R-11
R-12 1000
R-22 1000
R-123 50
R-134a 1000
Figure 13- Refrigerant Exposure limits
11
TLV-
TWA*CEILPEL
•
1000
•
•
•
1000
1000
• •
•
1000
•
• •
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.