2串行外设接口(SPI) is a trademark of Motorola.
3All other trademarks are the property of their respective owners.
PRODUCTION DATA information is current as of publication date.
Products conform to specifications per the terms of the Texas
Instruments standard warranty. Production processing does notEnglish Data Sheet: SBAS459
necessarily include testing of all parameters.
Please be aware that an important notice concerning availability, standard warranty, and use in critical applications of Texas
Instruments semiconductor products and disclaimers thereto appears at the end of this data sheet.
This integrated circuit can be damaged by ESD. Texas Instruments recommends that all integrated circuits be handled with
appropriate precautions. Failure to observe proper handling and installation procedures can cause damage.
ESD damage can range from subtle performance degradation to complete device failure. Precision integrated circuits may be more
susceptible to damage because very small parametric changes could cause the device not to meet its published specifications.
Over operating free-air temperature range, unless otherwise noted.
ADS1294, ADS1296, ADS1298
ADS1294R, ADS1296R, ADS1298R
AVDD to AVSS–0.3 to +5.5V
DVDD to DGND–0.3 to +3.9V
AVSS to DGND–3 to +0.2V
V
input to AVSSAVSS – 0.3 to AVDD + 0.3V
REF
Analog input to AVSSAVSS – 0.3 to AVDD + 0.3V
Digital input voltage to DGND–0.3 to DVDD + 0.3V
Digital output voltage to DGND–0.3 to DVDD + 0.3V
Input current (momentary)100mA
Input current (continuous)10mA
Operating
temperature
range
ESD ratings
Storage temperature range–60 to +150°C
Maximum junction temperature (TJ)+150°C
(1) Stresses above these ratings may cause permanent damage. Exposure to absolute maximum conditions for extended periods may
degrade device reliability. These are stress ratings only, and functional operation of the device at these or any other conditions beyond
those specified is not implied.
Minimum/maximum specifications apply for all commercial grade (0°C to +70°C) devices and from –40°C to +85°C for
industrial grades devices. Typical specifications are at +25°C. All specifications at DVDD = 1.8V, AVDD – AVSS = 3V
V
= 2.4V, external f
REF
PARAMETERTEST CONDITIONSMINTYPMAXUNIT
ANALOG INPUTS
Full-scale differential input voltage (AINP – AINN)±V
Input common-mode rangesubsection of the PGA Settings and Input
Input capacitance20pF
Input bias currentTA= 0°C to +70°C, input = 1.5V±1nA
DC input impedanceCurrent source lead-off detection500MΩ
Offset error±500µV
Offset error drift2µV/°C
Gain errorExcluding voltage reference error±0.2±0.5% of FS
Gain driftExcluding voltage reference drift5ppm/°C
Gain match between channels0.3% of FS
(1) Performance is applicable for 5V operation as well. Production testing for limits is performed at 3V.
(2) Noise data measured in a 10-second interval. Test not performed in production. Input-referred noise is calculated with input shorted
(without electrode resistance) over a 10-second interval.
= 2.048MHz, data rate = 500SPS, High-Resolution mode, and gain = 6, unless otherwise noted.
CLK
TA= +25°C, input = 1.5V±200pA
TA= –40°C to +85°C, input = 1.5V±1.2nA
No lead-off1000MΩ
Pull-up resistor lead-off detection10MΩ
Data rates up to 8kSPS, no missing codes24Bits
32kSPS data rate17Bits
f
= 2.048MHz, High-Resolution mode50032000SPS
CLK
f
= 2.048MHz, Low-Power mode25016000SPS
CLK
(2)
Gain = 6
Gain = 6, 256 points, 0.5 seconds of data47μV
Gain settings other than 6, data rates other
than 500SPS
Full-scale with gain = 6, best fit8ppm
Full-scale with gain = 6, best fit,
–20dBFS with gain = 6, best fit,
ADS1294R/6R/8R channel 1
Minimum/maximum specifications apply for all commercial grade (0°C to +70°C) devices and from –40°C to +85°C for
industrial grades devices. Typical specifications are at +25°C. All specifications at DVDD = 1.8V, AVDD – AVSS = 3V
V
= 2.4V, external f
REF
PARAMETERTEST CONDITIONSMINTYPMAXUNIT
CHANNEL PERFORMANCE (continued)
AC Performance
Common-mode rejection ratio (CMRR)fCM= 50Hz, 60Hz
Power-supply rejection ratio (PSRR)fPS= 50Hz, 60Hz90dB
CrosstalkfIN= 50Hz, 60Hz–126dB
Signal-to-noise ratio (SNR)fIN= 10Hz input, gain = 6112dB
Total harmonic distortion (THD)
DIGITAL FILTER
–3dB bandwidth0.262f
Digital filter settlingFull setting4Conversions
RIGHT LEG DRIVE (RLD) AMPLIFIER AND PACE AMPLIFIERS
RLD integrated noiseBW = 150Hz7μV
PACE integrated noiseBW = 8kHz20µV
PACE amplifier crosstalkCrosstalk between PACE amplifiers60dB
Gain bandwidth product50kΩ || 10pF load, gain = 1100kHz
Slew rate50kΩ || 10pF load, gain = 10.25V/μs
PACE and RLD amplifier drive strength
PACE and RLD current
PACE amplifier output resistance100Ω
Total harmonic distortionfIN= 100Hz, gain = 1–70dB
Common-mode input rangeAVSS + 0.7AVDD – 0.3V
Common-mode resistor matchingInternal 200kΩ resistor matching0.1%
Short-circuit current±0.25mA
Quiescent power consumptionEither RLD or PACE amplifier20μA
Minimum/maximum specifications apply for all commercial grade (0°C to +70°C) devices and from –40°C to +85°C for
industrial grades devices. Typical specifications are at +25°C. All specifications at DVDD = 1.8V, AVDD – AVSS = 3V
V
= 2.4V, external f
REF
PARAMETERTEST CONDITIONSMINTYPMAXUNIT
LEAD-OFF DETECT
FrequencySee the Register Map section for settings0, fDR/4kHz
CurrentSee the Register Map section for settings6, 12, 18, 24nA
Current accuracy±20%
Comparator threshold accuracy±30mV
RESPIRATION (ADS1294R/6R/8R Only)
Frequency
Phase shiftSee the Register Map section for settings22.590157.5Degrees
Impedance rangeI
Minimum/maximum specifications apply for all commercial grade (0°C to +70°C) devices and from –40°C to +85°C for
industrial grades devices. Typical specifications are at +25°C. All specifications at DVDD = 1.8V, AVDD – AVSS = 3V
V
Minimum/maximum specifications apply for all commercial grade (0°C to +70°C) devices and from –40°C to +85°C for
industrial grades devices. Typical specifications are at +25°C. All specifications at DVDD = 1.8V, AVDD – AVSS = 3V
V
= 2.4V, external f
REF
PARAMETERTEST CONDITIONSMINTYPMAXUNIT
POWER DISSIPATION (Analog Supply = 5V, RLD, WCT, and PACE Amplifiers Turned Off)
The ADS1294R/6R/8R channel performance differs from the ADS1294/6/8 in regards to
respiration circuitry found on channel one. Unless otherwise noted, ADS129x refers to all
specifications and functional descriptions of the ADS1294, ADS1296, ADS1298,
ADS1294R, ADS1296R, and ADS1298R.
The ADS129x noise performance can be optimized by adjusting the data rate and PGA setting. As the averaging
is increased by reducing the data rate, the noise drops correspondingly. Increasing the PGA value reduces the
input-referred noise, which is particularly useful when measuring low-level biopotential signals. Table 1 and
Table 2 summarize the noise performance of the ADS129x in the High-Resolution (HR) mode and Low-Power
(LP) mode, respectively, with a 3V analog power supply. Table 3 and Table 4 summarize the noise performance
of the ADS129x in the HR mode and LP mode, respectively, with a 5V analog power supply. The data are
representative of typical noise performance at TA= +25°C. The data shown are the result of averaging the
readings from multiple devices and are measured with the inputs shorted together. A minimum of 1000
consecutive readings are used to calculate the RMS and peak-to-peak noise for each reading. For the two
highest data rates, the noise is limited by quantization noise of the ADC and does not have a gaussian
distribution. Thus, the ratio between rms noise and peak-to-peak noise is approximately 10. For the lower data
rates, the ratio is approximately 6.6.
Table 1 to Table 4 show measurements taken with an internal reference. The data are also representative of the
ADS129x noise performance when using a low-noise external reference such as the REF5025.
DOUT43Digital outputSPI data out
GPIO244Digital input/outputGeneral-purpose input/output pin
GPIO345Digital input/outputGeneral-purpose input/output pin
GPIO446Digital input/outputGeneral-purpose input/output pin
DRDY47Digital outputData ready; active low
DVDD48SupplyDigital power supply
DGND49SupplyDigital ground
DVDD50SupplyDigital power supply
1Analog inputDifferential analog negative input 8 (ADS1298)
2Analog inputDifferential analog positive input 8 (ADS1298)
3Analog inputDifferential analog negative input 7 (ADS1298)
4Analog inputDifferential analog positive input 7 (ADS1298)
5Analog inputDifferential analog negative input 6 (ADS1296/8)
6Analog inputDifferential analog positive input 6 (ADS1296/8)
7Analog inputDifferential analog negative input 5 (ADS1296/8)
8Analog inputDifferential analog positive input 5 (ADS1296/8)
9Analog inputDifferential analog negative input 4
10Analog inputDifferential analog positive input 4
11Analog inputDifferential analog negative input 3
12Analog inputDifferential analog positive input 3
13Analog inputDifferential analog negative input 2
14Analog inputDifferential analog positive input 2
15Analog inputDifferential analog negative input 1
16Analog inputDifferential analog positive input 1
17Analog input/buffer output Internal test signal/single-ended buffer output based on register settings
18Analog input/outputInternal test signal/single-ended buffer output based on register settings
Master clock period414514414514ns
CS low to first SCLK, setup time617ns
SCLK period5066.6ns
SCLK pulse width, high and low1525ns
DIN valid to SCLK falling edge: setup time1010ns
Valid DIN after SCLK falling edge: hold time1011ns
SCLK falling edge to invalid DOUT: hold time1010ns
SCLK rising edge to DOUT valid: setup time1732ns
CS high pulse22t
CS low to DOUT driven1020ns
Eighth SCLK falling edge to CS high44t
Command decode time44t
CS high to DOUT Hi-Z1020ns
DAISY_IN valid to SCLK rising edge: setup time1010ns
DAISY_IN valid after SCLK rising edge: hold time1010ns
The ADS1294R/6R/8R channel performance differs from the ADS1294/6/8 in regards to
respiration circuitry found on channel one. Unless otherwise noted, ADS129x refers to all
specifications and functional descriptions of the ADS1294, ADS1296, ADS1298,
ADS1294R, ADS1296R, and ADS1298R.
The ADS129x are low-power, multichannel, simultaneously-sampling, 24-bit delta-sigma (ΔΣ) analog-to-digital
converters (ADCs) with integrated programmable gain amplifiers (PGAs). These devices integrate various
ECG-specificfunctionsthatmakethemwell-suitedforscalableelectrocardiogram(ECG),
electroencephalography (EEG), and electromyography (EMG) applications. The devices can also be used in
high-performance, multichannel data acquisition systems by powering down the ECG-specific circuitry.
The ADS129x have a highly programmable multiplexer that allows for temperature, supply, input short, and RLD
measurements. Additionally, the multiplexer allows any of the input electrodes to be programmed as the patient
reference drive. The PGA gain can be chosen from one of seven settings (1, 2, 3, 4, 6, 8, and 12). The ADCs in
the device offer data rates from 250SPS to 32kSPS. Communication to the device is accomplished using an
SPI-compatible interface. The device provides four GPIO pins for general use. Multiple devices can be
synchronized using the START pin.
The internal reference can be programmed to either 2.4V or 4V. The internal oscillator generates a 2.048MHz
clock. The versatile right leg drive (RLD) block allows the user to choose the average of any combination of
electrodes to generate the patient drive signal. Lead-off detection can be accomplished either by using a
pull-up/pull-down resistor or a current source/sink. An internal ac lead-off detection feature is also available. The
device supports both hardware PACE detection and software PACE detection. The Wilson Central Terminal
(WCT) block can be used to generate the WCT point of the standard 12-lead ECG.
Additionally, the ADS1294R, ADS1296R, and ADS1298R provide options for an internal respiration modulator
and a demodulator circuit in the signal path of channel 1.
This section discusses the details of the ADS129x internal functional elements. The analog blocks are reviewed
first, followed by the digital interface. Blocks implementing ECG-specific functions are covered in the end.
Throughout this document, f
denotes the frequency of the signal at the CLK pin, t
CLK
denotes the period of the
CLK
signal at the CLK pin, fDRdenotes the output data rate, tDRdenotes the time period of the output data, and f
denotes the frequency at which the modulator samples the input.
EMI FILTER
An RC filter at the input acts as an EMI filter on all of the channels. The –3dB filter bandwidth is approximately
3MHz.
INPUT MULTIPLEXER
The ADS129x input multiplexers are very flexible and provide many configurable signal switching options.
Figure 24 shows the multiplexer on a single channel of the device. Note that the device has eight such blocks,
one for each channel. TEST_PACE_OUT1, TEST_PACE_OUT2, and RLD_IN are common to all eight blocks.
VINP and VINN are separate for each of the eight blocks. This flexibility allows for significant device and
sub-system diagnostics, calibration, and configuration. Selection of switch settings for each channel is made by
writing the appropriate values to the CHnSET[2:0] register (see the CHnSET: Individual Channel Settings section
for details) and by writing the RLD_MEAS bit in the CONFIG3 register (see the CONFIG3: Configuration Register
3 subsection of the Register Map section for details). More details of the ECG-specific features of the multiplexer
are presented in the Input Multiplexer subsection of the ECG-Specifc Functions section.
Setting CHnSET[2:0] = 001 sets the common-mode voltage of (AVDD – AVSS)/2 to both inputs of the channel.
This setting can be used to test the inherent noise of the device in the user system.
Test Signals (TestP and TestN)
Setting CHnSET[2:0] = 101 provides internally-generated test signals for use in sub-system verification at
power-up. This functionality allows the entire signal chain to be tested out. Although the test signals are similar to
the CAL signals described in the IEC60601-2-51 specification, this feature is not intended for use in compliance
testing.
Control of the test signals is accomplished through register settings (see the CONFIG2: Configuration Register 2
subsection in the Register Map section for details). TEST_AMP controls the signal amplitude and TEST_FREQ
controls switching at the required frequency.
The test signals are multiplexed and transmitted out of the device at the TESTP_PACE_OUT1 and
TESTN_PACE_OUT2 pins. A bit register (CONFIG2.INT_TEST = 0) deactivates the internal test signals so that
the test signal can be driven externally. This feature allows the calibration of multiple devices with the same
signal. The test signal feature cannot be used in conjunction with the external hardware PACE feature (see the
External Hardware Approach subsection of the ECG-Specific Functions section for details).
When hardware PACE detect is not used, the TESTP_PACE_OUT1 and TESPN_PACE_OUT2 signals can be
used as a multiplexed differential input channel. These inputs can be multiplexed to any of the eight channels.
The performance of the differential input signal fed through these pins is identical to the normal channel
performance.
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
Temperature Sensor (TempP, TempN)
The ADS129x contain an on-chip temperature sensor. This sensor uses two internal diodes with one diode
having a current density 16x that of the other, as shown in Figure 25. The difference in current densities of the
diodes yields a difference in voltage that is proportional to absolute temperature.
As a result of the low thermal resistance of the package to the printed circuit board (PCB), the internal device
temperature tracks the PCB temperature closely. Note that self-heating of the ADS129x causes a higher reading
than the temperature of the surrounding PCB.
The scale factor of Equation 1 converts the temperature reading to °C. Before using this equation, the
temperature reading code must first be scaled to μV.
Setting CHnSET[2:0] = 011 sets the channel inputs to different supply voltages of the device. For channels 1, 2,
5, 6, 7, and 8, (MVDDP – MVDDN) is [0.5 × (AVDD – AVSS)]; for channel 3 and for channel 4, (MVDDP –
MVDDN) is DVDD/4. Note that to avoid saturating the PGA while measuring power supplies, the gain must be
set to '1'. For example, if AVDD = 2.5V and AVSS = –2.5V, then the measurement result would be 2.5V.
Lead-Off Excitation Signals (LoffP, LoffN)
The lead-off excitation signals are fed into the multiplexer before the switches. The comparators that detect the
lead-off condition are also connected to the multiplexer block before the switches. For a detailed description of
the lead-off block, refer to the Lead-Off Detection subsection in the ECG-Specific Functions section.
Auxiliary Single-Ended Input
The RLD_IN pin is primarily used for routing the right leg drive signal to any of the electrodes in case the right leg
drive electrode falls off. However, the RLD_IN pin can be used as a multiple single-ended input channel. The
signal at the RLD_IN pin can be measured with respect to the voltage at the RLD_REF pin using any of the eight
channels. This measurement is done by setting the channel multiplexer setting to '010' and the RLD_MEAS bit of
the CONFIG3 register to '1'.
ANALOG INPUT
The analog input to the ADS1298 is fully differential. Assuming PGA = 1, the differential input (INP – INN) can
span between –V
and AVDD + 0.3V. Refer to Table 8 for an explanation of the correlation between the analog input and the digital
codes. There are two general methods of driving the analog input of the ADS1298: single-ended or differential,
as shown in Figure 26 and Figure 27. Note that INP and INN are 180°C out-of-phase in the differential input
method. When the input is single-ended, the INN input is held at the common-mode voltage, preferably at
mid-supply. The INP input swings around the same common voltage and the peak-to-peak amplitude is the
(common-mode + 1/2V
is given by (INP + INN)/2. Both the INP and INN inputs swing from (common-mode + 1/2V
common-mode – 1/2V
differential configuration.
REF
to +V
REF
REF
. Note that the the absolute range for INP and INN must be between AVSS – 0.3V
REF
) and the (common-mode – 1/2V
). When the input is differential, the common-mode
REF
). For optimal performance, it is recommended that the ADS1298 devices be used in a
www.ti.com.cn
REF
to
Figure 26. Methods of Driving the ADS1298: Single-Ended or Differential
Figure 27. Using the ADS1298 in the Single-Ended and Differential Input Modes
PGA SETTINGS AND INPUT RANGE
The PGA is a differential input/differential output amplifier, as shown in Figure 28. It has seven gain settings (1,
2, 3, 4, 6, 8, and 12) that can be set by writing to the CHnSET register (see the CHnSET: Individual Channel
Settings subsection of the Register Map section for details). The ADS129x have CMOS inputs and hence have
negligible current noise. Table 6 shows the typical values of bandwidths for various gain settings. Note that
Table 6 shows the small-signal bandwidth. For large signals, the performance is limited by the slew rate of the
The resistor string of the PGA that implements the gain has 120kΩ of resistance for a gain of 6. This resistance
provides a current path across the outputs of the PGA in the presence of a differential input signal. This current
is in addition to the quiescent current specified for the device in the presence of a differential signal at the input.
Input Common-Mode Range
The usable input common-mode range of the front end depends on various parameters, including the maximum
differential input signal, supply voltage, PGA gain, etc. This range is described in Equation 2:
where:
V
MAX_DIFF
= maximum differential signal at the input of the PGA
CM = common-mode range(2)
For example:
If VDD= 3V, gain = 6, and V
MAX_DIFF
= 350mV
Then 1.25V < CM < 1.75V
www.ti.com.cn
Input Differential Dynamic Range
The differential (INP – INN) signal range depends on the analog supply and reference used in the system. This
range is shown in Equation 3.
(3)
The 3V supply, with a reference of 2.4V and a gain of 6 for ECGs, is optimized for power with a differential input
signal of approximately 300mV. For higher dynamic range, a 5V supply with a reference of 4V (set by the
VREF_4V bit of the CONFIG3 register) can be used to increase the differential dynamic range.
ADC ΔΣ Modulator
Each channel of the ADS129x has a 24-bit ΔΣ ADC. This converter uses a second-order modulator optimized for
low-power applications. The modulator samples the input signal at the rate of f
mode and f
shaped until f
= f
MOD
MOD
/8 for Low-Power mode. As in the case of any ΔΣ modulator, the noise of the ADS129x is
CLK
/2, as shown in Figure 29. The on-chip digital decimation filters explained in the next section
MOD
= f
/4 for High-Resolution
CLK
can be used to filter out the noise at higher frequencies. These on-chip decimation filters also provide antialias
filtering. This feature of the ΔΣ converters drastically reduces the complexity of the analog antialiasing filters that
are typically needed with Nyquist ADCs.
The digital filter receives the modulator output and decimates the data stream. By adjusting the amount of
filtering, tradeoffs can be made between resolution and data rate: filter more for higher resolution, filter less for
higher data rates. Higher data rates are typically used in ECG applications for implement software PACE
detection and ac lead-off detection.
The digital filter on each channel consists of a third-order sinc filter. The decimation ratio on the sinc filters can
be adjusted by the DR bits in the CONFIG1 register (see the Register Map section for details). This setting is a
global setting that affects all channels and, therefore, in a device all channels operate at the same data rate.
Sinc Filter Stage (sinx/x)
The sinc filter is a variable decimation rate, third-order, low-pass filter. Data are supplied to this section of the
filter from the modulator at the rate of f
then decimates the data stream into parallel data. The decimation rate affects the overall data rate of the
converter.
Equation 4 shows the scaled Z-domain transfer function of the sinc filter.
The frequency domain transfer function of the sinc filter is shown in Equation 5.
. The sinc filter attenuates the high-frequency noise of the modulator,
The sinc filter has notches (or zeroes) that occur at the output data rate and multiples thereof. At these
frequencies, the filter has infinite attenuation. Figure 30 shows the frequency response of the sinc filter and
Figure 31 shows the roll-off of the sinc filter. With a step change at input, the filter takes 3 × tDRto settle. After a
rising edge of the START signal, the filter takes t
time to give the first data output. The settling time of the
SETTLE
filters at various data rates are discussed in the START subsection of the SPI Interface section. Figure 32 and
Figure 33 show the filter transfer function until f
shows the transfer function extended until 4 × f
itself at every f
in frequencies around multiples of f
. The input R-C anti-aliasing filters in the system should be chosen such that any interference
MOD
are attenuated sufficiently.
MOD
/2 and f
MOD
. It can be seen that the passband of the ADS129x repeats
MOD
/16, respectively, at different data rates. Figure 34
MOD
www.ti.com.cn
Figure 30. Sinc Filter Frequency ResponseFigure 31. Sinc Filter Roll-Off
Figure 32. Transfer Function of On-ChipFigure 33. Transfer Function of On-Chip
Decimation Filters Until f
/2Decimation Filters Until f
MOD
MOD
/16
Figure 34. Transfer Function of On-Chip Decimation Filters
Figure 35 shows a simplified block diagram of the internal reference of the ADS129x. The reference voltage is
generated with respect to AVSS. When using the internal voltage reference, connect VREFN to AVSS.
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
(1) For V
= 2.4V: R1 = 12.5kΩ, R2 = 25kΩ, and R3 = 25kΩ. For V
REF
= 4V: R1 = 10.5kΩ, R2 = 15kΩ, and R3 = 35kΩ.
REF
Figure 35. Internal Reference
The external band-limiting capacitors determine the amount of reference noise contribution. For high-end ECG
systems, the capacitor values should be chosen such that the bandwidth is limited to less than 10Hz, so that the
reference noise does not dominate the system noise. When using a 3V analog supply, the internal reference
must be set to 2.4V. In case of a 5V analog supply, the internal reference can be set to 4V by setting the
VREF_4V bit in the CONFIG2 register.
Alternatively, the internal reference buffer can be powered down and VREFP can be applied externally. Figure 36
shows a typical external reference drive circuitry. Power-down is controlled by the PD_REFBUF bit in the
CONFIG3 register. By default the device wakes up in external reference mode.
The ADS129x provide two different methods for device clocking: internal and external. Internal clocking is ideally
suited for low-power, battery-powered systems. The internal oscillator is trimmed for accuracy at room
temperature. Over the specified temperature range the accuracy varies; see the Electrical Characteristics. Clock
selection is controlled by the CLKSEL pin and the CLK_EN register bit.
The CLKSEL pin selects either the internal or external clock. The CLK_EN bit in the CONFIG1 register enables
and disables the oscillator clock to be output in the CLK pin. A truth table for these two pins is shown in Table 7.
The CLK_EN bit is useful when multiple devices are used in a daisy-chain configuration. It is recommended that
during power-down the external clock be shut down to save power.
The ADS129x output 24 bits of data per channel in binary twos complement format, MSB first. The LSB has a
weight of V
full-scale input produces an output code of 800000h. The output clips at these codes for signals exceeding
full-scale. Table 8 summarizes the ideal output codes for different input signals. Note that for DR[2:0] = 000 and
001, the device has only 17 and 19 bits of resolution, respectively. The last seven (in 17-bit mode) or five (in
19-bit mode) bits can be ignored.
/(223– 1). A positive full-scale input produces an output code of 7FFFFFh and the negative
REF
www.ti.com.cn
Table 8. Ideal Output Code versus Input Signal
INPUT SIGNAL, V
(AINP – AINN)IDEAL OUTPUT CODE
≥ V
REF
+V
/(223– 1)000001h
REF
0000000h
–V
/(223– 1)FFFFFFh
REF
≤ –V
(1) Only valid for 24-bit resolution data rates.
(2) Assumes gain = 1.
(3) Excludes effects of noise, linearity, offset, and gain error.
The SPI-compatible serial interface consists of four signals: CS, SCLK, DIN, and DOUT. The interface reads
conversion data, reads and writes registers, and controls the ADS129x operation. The DRDY output is used as a
status signal to indicate when data are ready. DRDY goes low when new data are available.
Chip Select (CS)
Chip select (CS) selects the ADS129x devices for SPI communication. While CS is low the serial interface is
active CS must remain low for the entire duration of the serial communication. After the serial communication is
finished, always wait four or more t
is reset, SCLK and DIN are ignored, and DOUT enters a high-impedance state. DRDY asserts when data
conversion is complete, regardless of whether CS is high or low.
While ADS129x is selected the device will attempt to decode and execute commands every eight serial clocks. If
the devices ceases to execute serial commands, it is possible extra clock pulses were presented and placed the
serial interface in an unknown state. Take CS high and back low to reset the serial interface to a known state.
Serial Clock (SCLK)
SCLK is the serial peripheral interface (SPI) serial clock. It is used to shift in commands and shift out data from
the device. The serial clock (SCLK) features a Schmitt-triggered input and clocks data on the DIN and DOUT
pins into and out of the ADS129x. Even though the input has hysteresis, it is recommended to keep SCLK as
clean as possible to prevent glitches from accidentally forcing a clock event. The absolute maximum limit for
SCLK is specified in the Serial Interface Timing table.
While ADS129x is selected(CS = LOW), the device attempts to decode and execute commands every eight
serial clocks. It is therefore recommended that multiples of 8 SCLKs be presented every serial transfer to keep
the interface in a normal operating mode. If the interface ceases to function because of extra serial clocks, it can
be reset by toggling CS high and back to low.
cycles before taking CS high. When CS is taken high, the serial interface
CLK
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
For a single device, the minimum speed needed for the SCLK depends on the number of channels, number of
bits of resolution, and output data rate. (For multiple cascaded devices, see the Cascade Mode subsection of the
Multiple Device Configuration section.)
t
SCLK
< (tDR– 4t
CLK
)/(N
BITS
× N
CHANNELS
+ 24)(6)
For example, if the ADS1298 is used in a 500SPS mode (eight channels, 24-bit resolution), the minimum SCLK
speed is 110kHz.
Data retrieval can be done either by putting the device in RDATAC mode or by issuing a RDATA command for
data on demand. The above SCLK rate limitation applies to RDATAC. For the RDATA command, the limitation
applies if data must be read between two consecutive DRDY signals. The above calculation assumes that there
are no other commands issued between data captures.
Data Input (DIN)
The data input pin (DIN) is used along with SCLK to communicate with the ADS129x (opcode commands and
register data). The device latches data on DIN on the falling edge of SCLK.
The data output pin (DOUT) is used with SCLK to read conversion and register data from the ADS129x. Data on
DOUT are shifted out on the rising edge of SCLK. DOUT goes to a high-impedance state when CS is high. In
read data continuous mode (see the SPI Command Definitions section for more details), the DOUT output line
also indicates when new data are available. This feature can be used to minimize the number of connections
between the device and the system controller.
Figure 37 shows the data output protocol for ADS1298.
Figure 37. SPI Bus Data Output for the ADS1298 (Eight Channels)
www.ti.com.cn
Data Retrieval
Data retrieval can be accomplished in one of two methods. The read data continuous command (see the
RDATAC: Read Data Continuous section) can be used to set the device in a mode to read the data continuously
without sending opcodes. The read data command (see the RDATA: Read Data section) can be used to read
just one data output from the device (see the SPI Command Definitions section for more details). The conversion
data are read by shifting the data out on DOUT. The MSB of the data on DOUT is clocked out on the first SCLK
rising edge. DRDY returns to high on the first SCLK falling edge. DIN should remain low for the entire read
operation.
The number of bits in the data output depends on the number of channels and the number of bits per channel.
For the ADS1298/8R, the number of data outputs is (24 status bits + 24 bits × 8 channels) = 216 bits. The format
of the 24 status bits is: (1100 + LOFF_STATP + LOFF_STATN + bits[4:7] of the GPIO register). The data format
for each channel data are twos complement and MSB first. When channels are powered down using the user
register setting, the corresponding channel output is set to '0'. However, the sequence of channel outputs
remains the same. For the ADS1294/4R and ADS1296/6R, the last four and two channel outputs shown in
Figure 37 are zeros. The four and six channels parts require only 120 and 168 SCLKs to shift data out,
respectively. Status and GPIO register bits are loaded into the 24-bit status word 2t
s before DRDY goes low.
CLK
The ADS129x also provide a multiple readback feature. The data can be read out multiple times by simply giving
more SCLKs, in which case the MSB data byte repeats after reading the last byte. The DAISY_EN bit in
CONFIG1 register must be set to '1' for multiple readbacks.
Data Ready (DRDY)
DRDY is an output. When it transitions low, new conversion data are ready. The CS signal has no effect on the
data ready signal. Regardless of the status of the CS signal, a rising edge on SCLK pulls DRDY high. Hence,
when using multiple devices in the SPI bus, it is recommended that SCLK be gated with CS. The behavior of
DRDY is determined by whether the device is in RDATAC mode or the RDATA command is being used to read
data on demand. (See the RDATAC: Read Data Continuous and RDATA: Read Data subsections of the SPI
Command Definitions section for further details).
When reading data with the RDATA command, the read operation can overlap the occurrence of the next DRDY
without data corruption.
The START pin or the START command is used to place the device either in normal data capture mode or pulse
data capture mode.
Figure 38 shows the relationship between DRDY, DOUT, and SCLK during data retrieval (in case of an ADS1298
with a selected data rate that gives 24-bit resolution). DOUT is latched out at the rising edge of SCLK. DRDY is
pulled high at the falling edge of SCLK. Note that DRDY goes high on the first falling edge SCLK regardless of
whether data are being retrieved from the device or a command is being sent through the DIN pin.
Figure 38. DRDY with Data Retrieval (CS = 0 in RDATA mode)
GPIO
The ADS129x have a total of four general-purpose digital I/O (GPIO) pins available in the normal mode of
operation. The digital I/O pins are individually configurable as either inputs or as outputs through the GPIOC bits
register. The GPIOD bits in the GPIO register control the level of the pins. When reading the GPIOD bits, the
data returned are the logic level of the pins, whether they are programmed as inputs or outputs. When the GPIO
pin is configured as an input, a write to the corresponding GPIOD bit has no effect. When configured as an
output, a write to the GPIOD bit sets the output value.
If configured as inputs, these pins must be driven (do not float). The GPIO pins are set as inputs after power-on
or after a reset. Figure 39 shows the GPIO port structure. The pins should be shorted to DGND if not used.
GPIO1 can be used as the PACEIN signal; GPIO2 is multiplexed with RESP_BLK signal; GPIO3 is multiplexed
with the RESP signal; and GPIO4 is multiplexed with the RESP_PH signal.
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
Figure 39. GPIO Port Pin
Power-Down (PWDN)
When PWDN is pulled low, all on-chip circuitry is powered down. To exit power-down mode, take the PWDN pin
high. Upon exiting from power-down mode, the internal oscillator and the reference require time to wake up. It is
recommended that during power-down the external clock is shut down to save power.
There are two methods to reset the ADS129x: pull the RESET pin low, or send the RESET opcode command.
When using the RESET pin, take it low to force a reset. Make sure to follow the minimum pulse width timing
specifications before taking the RESET pin back high. The RESET command takes effect on the eighth SCLK
falling edge of the opcode command. On reset it takes 18 t
cycles to complete initialization of the configuration
CLK
registers to the default states and start the conversion cycle. Note that an internal RESET is automatically issued
to the digital filter whenever registers CONFIG1 and RESP are set to new values with a WREG command.
START
The START pin must be set high for at least 2 t
s or the START command sent to begin conversions. When
CLK
START is low or if the START command has not been sent, the device does not issue a DRDY signal
(conversions are halted).
When using the START opcode to control conversion, hold the START pin low. The ADS129x feature two modes
to control conversion: continuous mode and single-shot mode. The mode is selected by SINGLE_SHOT (bit 3 of
the CONFIG4 register). In multiple device configurations the START pin is used to synchronize devices (see the
Multiple Device Configuration subsection of the SPI Interface section for more details).
Settling Time
The settling time (t
) is the time it takes for the converter to output fully settled data when START signal is
SETTLE
pulled high. Once START is pulled high, DRDY is also pulled high. The next falling edge of DRDY indicates that
data are ready. Figure 40 shows the timing diagram and Table 9 shows the settling time for different data rates.
The settling time depends on f
register). Table 8 shows the settling time as a function of t
and the decimation ratio (controlled by the DR[2:0] bits in the CONFIG1
CLK
. Note that when START is held high and there is a
CLK
step change in the input signal, it takes 3 × tDRfor the filter to settle to the new value. Settled data are available
on the fourth DRDY pulse. This time must be considered when trying to measure narrow PACE pulses for PACE
detection.
Conversions begin when the START pin is taken high for at least 2 t
is sent. As seen in Figure 41, the DRDY output goes high when conversions are started and goes low when data
are ready. Conversions continue indefinitely until the START pin is taken low or the STOP opcode command is
transmitted. When the START pin is pulled low or the stop command is issued, the conversion in progress is
allowed to complete. Figure 42 and Table 10 show the required timing of DRDY to the START pin and the
START/STOP opcode commands when controlling conversions in this mode. To keep the converter running
continuously, the START pin can be permanently tied high. Note that when switching from pulse mode to
continuous mode, the START signal is pulsed or a STOP command must be issued followed by a START
command. This conversion mode is ideal for applications that require a fixed-continuous stream of conversions
results.
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
s or when the START opcode command
CLK
(1) START and STOP opcode commands take effect on the seventh SCLK falling edge.
Figure 41. Continuous Conversion Mode
(1) START and STOP commands take effect on the seventh SCLK falling edge at the end of the opcode transmission.
Figure 42. START to DRDY Timing
Table 10. Timing Characteristics for Figure 42
SYMBOLDESCRIPTIONMINUNIT
t
SDSU
t
DSHD
(1) START and STOP commands take effect on the seventh SCLK falling edge at the end of the opcode transmission.
START pin low or STOP opcode to DRDY setup time
to halt further conversions
START pin low or STOP opcode to complete current
conversion
The single-shot mode is enabled by setting the SINGLE_SHOT bit in CONFIG4 register to '1'. In single-shot
mode, the ADS129x perform a single conversion when the START pin is taken high or when the START opcode
command is sent. As seen in Figure 42, when a conversion is complete, DRDY goes low and further conversions
are stopped. Regardless of whether the conversion data are read or not, DRDY remains low. To begin a new
conversion, take the START pin low and then back high for at least 2 t
s, or transmit the START opcode again.
CLK
Note that when switching from continuous mode to pulse mode, make sure the START signal is pulsed or issue
a STOP command followed by a START command.
Figure 43. DRDY with No Data Retrieval in Single-Shot Mode
This conversion mode is provided for applications that require non-standard or non-continuous data rates.
Issuing a START command or toggling the START pin high resets the digital filter, effectively dropping the data
rate by a factor of four. This mode leaves the system more susceptible to aliasing effects, thus requiring more
complex analog or digital filtering. Loading on the host processor increases because it must toggle the START
pin or send a START command to initiate a new conversion cycle.
www.ti.com.cn
MULTIPLE DEVICE CONFIGURATION
The ADS129x are designed to provide configuration flexibility when multiple devices are used in a system. The
serial interface typically requires four signals: DIN, DOUT, SCLK, and CS. With one additional chip select signal
per device, multiple devices can be connected together. The number of signals needed to interface n devices is
3 + n.
The right leg drive amplifiers can be daisy-chained as explained in the RLD Configuration with Multiple Devices
subsection of the ECG-Specific Functions section. To use the internal oscillator in a daisy-chain configuration,
one of the devices must be set as the master for the clock source with the internal oscillator enabled (CLKSEL
pin = 1) and the internal oscillator clock brought out of the device by setting the CLK_EN register bit to '1'. This
master device clock is used as the external clock source for the other devices.
When using multiple devices, the devices can be synchronized with the START signal. The delay from START to
the DRDY signal is fixed for a fixed data rate (see the START subsection of the SPI Interface section for more
details on the settling times). Figure 44 shows the behavior of two devices when synchronized with the START
signal as an example.
There are two ways to connect multiple devices with a optimal number of interface pins: cascade mode and
daisy-chain mode.
Figure 45a shows a configuration with two devices cascaded together. One of the devices is an ADS1298 (eight
channels) and the other is an ADS1294 (four channels). Together, they create a system with 12 channels.
DOUT, SCLK, and DIN are shared. Each device has its own chip select. When a device is not selected by the
corresponding CS being driven to logic 1, the DOUT of this device is high-impedance. This structure allows the
other device to take control of the DOUT bus. This configuration method is suitable for the majority of
applications.
Daisy-Chain Mode
Daisy-chain mode is enabled by setting the DAISY_EN bit in the CONFIG1 register. Figure 45b shows the
daisy-chain configuration. In this mode SCLK, DIN, and CS are shared across multiple devices. The DOUT of
one device is hooked up to the DAISY_IN of the other device, thereby creating a chain. One extra SCLK must be
issued between each data set. Also, when using daisy-chain mode, the multiple readback feature is not
available. Short the DAISY_IN pin to digital ground if not used. Figure 2 describes the required timing for the
ADS1298 shown in Figure 46. Data from the ADS1298 appear first on DOUT, followed by a don’t care bit, and
finally by the status and data words from the ADS1294.
www.ti.com.cn
(1) To reduce pin count, set the START pin low and use the START serial command to synchronize and start conversions.
In a case where all devices in the chain operate in the same register setting, DIN can be shared as well and
thereby reduce the SPI communication signals to four, regardless of the number of devices. However, because
the individual devices cannot be programmed, the RLD driver cannot be shared among the multiple devices.
Furthermore, an external clock must be used.
Note that from Figure 2, the SCLK rising edge shifts data out of the ADS129x on DOUT. The SCLK rising edge is
also used to latch data into the device DAISY_IN pin down the chain. This architecture allows for a faster SCLK
rate speed, but it also makes the interface sensitive to board level signal delays. The more devices in the chain,
the more challenging it could become to adhere to setup and hold times. A star pattern connection of SCLK to all
devices, minimizing length of DOUT, and other PCB layout techniques help. Placing delay circuits such as
buffers between DOUT and DAISY_IN are ways to mitigate this challenge. One other option is to insert a D
flip-flop between DOUT and DAISY_IN clocked on an inverted SCLK. Note also that daisy-chain mode requires
some software overhead to recombine data bits spread across byte boundaries.
The maximum number of devices that can be daisy-chained depends on the data rate at which the device is
being operated. The maximum number of devices can be estimated with Equation 7.
where:
N
= device resolution (depends on data rate), and
BITS
N
CHANNELS
= number of channels in the device (4, 6, or 8).(7)
For example, when the ADS1298 (eight-channel, 24-bit version) is operated at a 2kSPS data rate with a 4MHz
f
The ADS129x provide flexible configuration control. The opcode commands, summarized in Table 11, control
and configure the operation of the ADS129x. The opcode commands are stand-alone, except for the register
read and register write operations that require a second command byte plus data. CS can be taken high or held
low between opcode commands but must stay low for the entire command operation (especially for multi-byte
commands). System opcode commands and the RDATA command are decoded by the ADS129x on the seventh
falling edge of SCLK. The register read/write opcodes are decoded on the eighth SCLK falling edge. Be sure to
follow SPI timing requirements when pulling CS high after issuing a command.
Table 11. Opcode Command Definitions
COMMANDDESCRIPTIONFIRST BYTESECOND BYTE
System Commands
RDATAC0001 0000 (10h)
SDATACStop Read Data Continuously mode0001 0001 (11h)
RDATARead data by command; supports multiple read back.0001 0010 (12h)
Register Read Commands
RREGRead n nnnn registers starting at address r rrrr001r rrrr (2xh)
WREGWrite n nnnn registers starting at address r rrrr010r rrrr (4xh)
(1) When in RDATAC mode, the RREG command is ignored.
(2) n nnnn = number of registers to be read/written – 1. For example, to read/write three registers, set n nnnn = 0 (0010). r rrrr = starting
register address for read/write opcodes.
Enable Read Data Continuous mode.
This mode is the default mode at power-up.
(1)
(2)
(2)
www.ti.com.cn
000n nnnn
000n nnnn
(2)
(2)
WAKEUP: Exit STANDBY Mode
This opcode exits the low-power standby mode; see the STANDBY: Enter STANDBY Mode subsection of the
SPI Command Definitions section. Time is required when exiting standby mode (see the Electrical
Characteristics for details). There are no restrictions on the SCLK rate for this command and it can be
issued any time. Any subsequent command must be sent after 4 t
CLK
cycles.
STANDBY: Enter STANDBY Mode
This opcode command enters the low-power standby mode. All parts of the circuit are shut down except for the
reference section. The standby mode power consumption is specified in the Electrical Characteristics. There areno restrictions on the SCLK rate for this command and it can be issued any time. Send a WAKEUP
command to return device to normal operation. Serial interface is active, thus register read/write commands are
permitted while in this mode.
RESET: Reset Registers to Default Values
This command resets the digital filter cycle and returns all register settings to the respective default values. See
the Reset (RESET) subsection of the SPI Interface section for more details. There are no restrictions on theSCLK rate for this command and it can be issued any time. 18 t
cycles are required to execute the
CLK
RESET command. Avoid sending any commands during this time.
This opcode starts data conversions. Tie the START pin low to control conversions by command. If conversions
are in progress this command has no effect. The STOP opcode command is used to stop conversions. If the
START command is immediately followed by a STOP command, there must be a gap of 4 t
the two commands. When the START opcode is sent to the device, keep the START pin low until the STOP
command is issued. (See the START subsection of the SPI Interface section for more details.) There are no
restrictions on the SCLK rate for this command and it can be issued any time.
STOP: Stop Conversions
This opcode stops conversions. Tie the START pin low to control conversions by command. When the STOP
command is sent, the conversion in progress completes and further conversions are stopped. If conversions are
already stopped, this command has no effect. There are no restrictions on the SCLK rate for this command
and it can be issued any time.
RDATAC: Read Data Continuous
This opcode enables the output of conversion data on each DRDY without the need to issue subsequent read
data opcodes. This mode places the conversion data in the output register and may be shifted out directly. The
read data continuous mode is the default mode of the device and the device defaults to this mode on power-up
and reset.
RDATAC mode is cancelled by the Stop Read Data Continuous command. If the device is in RDATAC mode, a
SDATAC command must be issued before any other commands can be sent to the device. There is no
restriction on the SCLK rate for this command. However, the subsequent data retrieval SCLKs or the SDATAC
opcode command should wait at least 4 t
shows, there is a keep out zone of 4 t
cycles. The timing for RDATAC is shown in Figure 47. As Figure 47
CLK
cycles around the DRDY pulse when this command cannot be issued.
CLK
If no data are retrieved from the device, DOUT and DRDY behave similarly in this mode. To retrieve data from
the device after RDATAC command is issued, make sure either the START pin is high or the START command
is issued. Figure 47 shows the recommended way to use the RDATAC command. RDATAC is ideally suited for
applications such as data loggers or recorders where registers are set once and do not need to be re-configured.
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
cycles between
CLK
(1) t
UPDATE
= 4/f
. Do not read data during this time.
CLK
Figure 47. RDATAC Usage
SDATAC: Stop Read Data Continuous
This opcode cancels the Read Data Continuous mode. There is no restriction on the SCLK rate for this
command, but the subsequent command must wait for 4 t
Issue this command after DRDY goes low to read the conversion result (in Stop Read Data Continuous mode).
There is no restriction on the SCLK rate for this command, and there is no wait time needed for the subsequent
commands or data retrieval SCLKs. To retrieve data from the device after RDATA command is issued, make
sure either the START pin is high or the START command is issued. When reading data with the RDATA
command, the read operation can overlap the occurrence of the next DRDY without data corruption. Figure 48
shows the recommended way to use the RDATA command. RDATA is best suited for ECG- and EEG-type
systems where register settings must be read or changed often between conversion cycles.
www.ti.com.cn
Figure 48. RDATA Usage
Sending Multi-Byte Commands
The ADS129x serial interface decodes commands in bytes and requires 4 t
Therefore, when sending multi-byte commands, a 4 t
period must separate the end of one byte (or opcode)
CLK
cycles to decode and execute.
CLK
and the next.
Assume CLK is 2.048MHz, then t
SDECODE
in 500ns. This byte transfer time does not meet the t
(4 t
) is 1.96µs. When SCLK is 16MHz, one byte can be transferred
CLK
SDECODE
specification; therefore, a delay must be inserted so
the end of the second byte arrives 1.46µs later. If SCLK is 4MHz, one byte is transferred in 2µs. Because this
transfer time exceeds the t
SDECODE
specification, the processor can send subsequent bytes without delay. In this
later scenario, the serial port can be programmed to cease single-byte transfer per cycle to multiple bytes.
This opcode reads register data. The Register Read command is a two-byte opcode followed by the output of the
register data. The first byte contains the command opcode and the register address. The second byte of the
opcode specifies the number of registers to read – 1.
First opcode byte: 001r rrrr, where r rrrr is the starting register address.
Second opcode byte: 000n nnnn, where n nnnn is the number of registers to read – 1.
The 17th SCLK rising edge of the operation clocks out the MSB of the first register, as shown in Figure 49. When
the device is in read data continuous mode, it is necessary to issue a SDATAC command before a RREG
command can be issued. An RREG command can be issued any time. However, because this command is a
multi-byte command, there are restrictions on the SCLK rate depending on the way the SCLKs are issued. See
the Serial Clock (SCLK) subsection of the SPI Interface section for more details. Note that CS must be low for
the entire command.
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
Figure 49. RREG Command Example: Read Two Registers Starting from Register 00h (ID Register)
(OPCODE 1 = 0010 0000, OPCODE 2 = 0000 0001)
WREG: Write to Register
This opcode writes register data. The Register Write command is a two-byte opcode followed by the input of the
register data. The first byte contains the command opcode and the register address.
The second byte of the opcode specifies the number of registers to write – 1.
First opcode byte: 010r rrrr, where r rrrr is the starting register address.
Second opcode byte: 000n nnnn, where n nnnn is the number of registers to write – 1.
After the opcode bytes, the register data follows (in MSB-first format), as shown in Figure 50. WREG command
can be issued any time. However, because this command is a multi-byte command, there are restrictions on the
SCLK rate depending on the way the SCLKs are issued. See the Serial Clock (SCLK) subsection of the SPI
Interface section for more details. Note that CS must be low for the entire command.
Figure 50. WREG Command Example: Write Two Registers Starting from 00h (ID Register)
ID: ID Control Register (Factory-Programmed, Read-Only)
Address = 00h
BIT 7BIT 6BIT 5BIT 4BIT 3BIT 2BIT 1BIT 0
DEV_ID7DEV_ID6DEV_ID510DEV_ID2DEV_ID1DEV_ID0
The ID Control Register is programmed during device manufacture to indicate device characteristics.
Bits[7:5]DEV_ID[7:5]: Device family identification
These bits indicate the device family.
000 = Reserved
011 = Reserved
100 = ADS129x device family
101 = Reserved
110 = ADS129xR device family
111 = Reserved
Bit 4This bit reads high.
Bit 3This bit reads low.
Bits[2:0]DEV_ID[2:0]: Channel number identification
These bits indicates number of channels.
000 = 4-channel ADS1294 or ADS1294R
001 = 6-channel ADS1296 or ADS1296R
010 = 8-channel ADS1298 or ADS1298R
011 = Reserved
111 = Reserved
This bit determines whether the device runs in Low-Power or High-Resolution mode.
0 = Low-Power mode (default)
1 = High-Resolution mode
Bit 6DAISY_EN: Daisy-chain/multiple readback mode
This bit determines which mode is enabled.
0 = Daisy-chain mode (default)
1 = Multiple readback mode
Bit 5CLK_EN: CLK connection
This bit determines if the internal oscillator signal is connected to the CLK pin when the CLKSEL pin = 1.
0 = Oscillator clock output disabled (default)
1 = Oscillator clock output enabled
Bits[4:3]Must always be set to '0'
Bits[2:0]DR[2:0]: Output data rate
For High-Resolution mode, f
These bits determine the output data rate of the device.
(1) Additional power is consumed when driving external devices.
(1)
MOD
= f
/4. For low power mode, f
CLK
MOD
= f
CLK
/8.
www.ti.com.cn
BITDATA RATEHIGH-RESOLUTION MODE
000f
001f
010f
011f
100f
101f
110 (default)f
/1632kSPS16kSPS
MOD
/3216kSPS8kSPS
MOD
/648kSPS4kSPS
MOD
/1284kSPS2kSPS
MOD
/2562kSPS1kSPS
MOD
/5121kSPS500SPS
MOD
/1024500SPS250SPS
MOD
111Do not usen/an/a
(1) Additional power is consumed when driving external devices.
(2) f
Configuration Register 2 configures the test signal generation. See the Input Multiplexer section for more details.
Bits[7:6]Must always be set to '0'
Bit 5WCT_CHOP: WCT chopping scheme
This bit determines whether the chopping frequency of WCT amplifiers is variable or fixed.
0 = Chopping frequency varies, see Table 13
1 = Chopping frequency constant at f
Bit 4INT_TEST: TEST source
This bit determines the source for the Test signal.
0 = Test signals are driven externally (default)
1 = Test signals are generated internally
Bit 3Must always be set to '0'
Bit 2TEST_AMP: Test signal amplitude
These bits determine the calibration signal amplitude.
0 = 1 × –(VREFP – VREFN)/2.4mV (default)
1 = 2 × –(VREFP – VREFN)/2.4mV
Bits[1:0]TEST_FREQ[1:0]: Test signal frequency
These bits determine the calibration signal frequency.
00 = Pulsed at f
01 = Pulsed at f
10 = Not used
11 = At dc
Configuration Register 3 configures multi-reference and RLD operation.
Bit 7PD_REFBUF: Power-down reference buffer
This bit determines the power-down reference buffer state.
0 = Power-down internal reference buffer (default)
1 = Enable internal reference buffer
Bit 6Must always be set to '1'
Default is '1' at power-up.
Bit 5VREF_4V: Reference voltage
This bit determines the reference voltage, VREFP.
0 = VREFP is set to 2.4V (default)
1 = VREFP is set to 4V (use only with a 5V analog supply)
Bit 4RLD_MEAS: RLD measurement
This bit enables RLD measurement. The RLD signal may be measured with any channel.
0 = Open (default)
1 = RLD_IN signal is routed to the channel that has the MUX_Setting 010 (V
Bit 3RLDREF_INT: RLDREF signal
This bit determines the RLDREF signal source.
0 = RLDREF signal fed externally (default)
1 = RLDREF signal (AVDD – AVSS)/2 generated internally
Bit 2PD_RLD: RLD buffer power
This bit determines the RLD buffer power state.
0 = RLD buffer is powered down (default)
1 = RLD buffer is enabled
Bit 1RLD_LOFF_SENS: RLD sense function
This bit enables the RLD sense function.
0 = RLD sense is disabled (default)
1 = RLD sense is enabled
Bit 0RLD_STAT: RLD lead-off status
This bit determines the RLD status.
0 = RLD is connected (default)
1 = RLD is not connected
This bit determines the lead-off detection mode.
0 = Current source mode lead-off (default)
1 = Pull-up/pull-down resistor mode lead-off
Bits[3:2]ILEAD_OFF[1:0]: Lead-off current magnitude
These bits determine the magnitude of current for the current lead-off mode.
00 = 6nA (default)
01 = 12nA
10 = 18nA
11 = 24nA
Bits[1:0]FLEAD_OFF[1:0]: Lead-off frequency
These bits determine the frequency of lead-off detect for each channel.
00 = When any bits of the LOFF_SENSP or LOFF_SENSN registers are turned on, make sure that FLEAD[1:0] are either
set to 01 or 11 (default)
01 = AC lead-off detection at fDR/4
10 = Do not use
11 = DC lead-off detection turned on
The CH[1:8]SET Control Register configures the power mode, PGA gain, and multiplexer settings channels. See
the Input Multiplexer section for details. CH[2:8]SET are similar to CH1SET, corresponding to the respective
channels.
Bit 7PD: Power-down
This bit determines the channel power mode for the corresponding channel.
0 = Normal operation (default)
1 = Channel power-down.
When powering down a channel, it is recommended that the channel be set to input short by setting the appropriate
MUXn[2:0] = 001 of the CHnSET register.
Bits[6:4]GAIN[2:0]: PGA gain
These bits determine the PGA gain setting.
000 = 6 (default)
001 = 1
010 = 2
011 = 3
100 = 4
101 = 8
110 = 12
Bit 3Always write '0'
Bits[2:0]MUXn[2:0]: Channel input
These bits determine the channel input selection.
000 = Normal electrode input (default)
001 = Input shorted (for offset or noise measurements)
010 = Used in conjunction with RLD_MEAS bit for RLD measurements. See the Right Leg Drive (RLD DC Bias Circuit)
subsection of the ECG-Specific Functions section for more details.
011 = MVDD for supply measurement
100 = Temperature sensor
101 = Test signal
110 = RLD_DRP (positive electrode is the driver)
111 = RLD_DRN (negative electrode is the driver)
www.ti.com.cn
RLD_SENSP
Address = 0Dh
BIT 7BIT 6BIT 5BIT 4BIT 3BIT 2BIT 1BIT 0
RLD8PRLD7PRLD6PRLD5PRLD4PRLD3PRLD2PRLD1P
This register controls the selection of the positive signals from each channel for right leg drive derivation. See the
Right Leg Drive (RLD DC Bias Circuit) subsection of the ECG-Specific Functions section for details.
Note that registers bits[5:4] are not available for the ADS1294/4R. Bits[7:6] are not available for the
ADS1294/6/4R/6R.
RLD_SENSN
Address = 0Eh
BIT 7BIT 6BIT 5BIT 4BIT 3BIT 2BIT 1BIT 0
RLD8NRLD7NRLD6NRLD5NRLD4NRLD3NRLD2NRLD1N
This register controls the selection of the negative signals from each channel for right leg drive derivation. See
the Right Leg Drive (RLD DC Bias Circuit) subsection of the ECG-Specific Functions section for details.
Note that registers bits[5:4] are not available for the ADS1294/4R. Bits[7:6] are not available for the
ADS1294/6/4R/6R.
This register selects the positive side from each channel for lead-off detection. See the Lead-Off Detection
subsection of the ECG-Specific Functions section for details. Note that the LOFF_STATP register bits are only
valid if the corresponding LOFF_SENSP bits are set to '1'.
Note that registers bits[5:4] are not available for the ADS1294/4R. Bits[7:6] are not available for the
ADS1294/6/4R/6R.
LOFF_SENSN
Address = 10h
BIT 7BIT 6BIT 5BIT 4BIT 3BIT 2BIT 1BIT 0
LOFF8NLOFF7NLOFF6NLOFF5NLOFF4NLOFF3NLOFF2NLOFF1N
This register selects the negative side from each channel for lead-off detection. See the Lead-Off Detection
subsection of the ECG-Specific Functions section for details. Note that the LOFF_STATN register bits are only
valid if the corresponding LOFF_SENSN bits are set to '1'.
Note that registers bits[5:4] are not available for the ADS1294/4R. Bits[7:6] are not available for the
ADS1294/6/4R/6R.
This register controls the direction of the current used for lead-off derivation. See the Lead-Off Detection
subsection of the ECG-Specific Functions section for details.
This register stores the status of whether the positive electrode on each channel is on or off. See the Lead-Off
Detection subsection of the ECG-Specific Functions section for details. Ignore the LOFF_STATP values if the
corresponding LOFF_SENSP bits are not set to '1'.
'0' is lead-on (default) and '1' is lead-off. When the LOFF_SENSEP bits are '0', the LOFF_STATP bits should be
This register stores the status of whether the negative electrode on each channel is on or off. See the Lead-Off
Detection subsection of the ECG-Specific Functions section for details. Ignore the LOFF_STATN values if the
corresponding LOFF_SENSN bits are not set to '1'.
'0' is lead-on (default) and '1' is lead-off. When the LOFF_SENSEN bits are '0', the LOFF_STATP bits should be
ignored.
GPIO: General-Purpose I/O Register
Address = 14h
BIT 7BIT 6BIT 5BIT 4BIT 3BIT 2BIT 1BIT 0
GPIOD4GPIOD3GPIOD2GPIOD1GPIOC4GPIOC3GPIOC2GPIOC1
The General-Purpose I/O Register controls the action of the three GPIO pins. Note that when RESP_CTRL[1:0]
is in mode 01 and 11, the GPIO2, GPIO3, and GPIO4 pins are not available for use.
www.ti.com.cn
Bits[7:4]GPIOD[4:1]: GPIO data
These bits are used to read and write data to the GPIO ports.
When reading the register, the data returned correspond to the state of the GPIO external pins, whether they are
programmed as inputs or as outputs. As outputs, a write to the GPIOD sets the output value. As inputs, a write to the
GPIOD has no effect. GPIO is not available in certain respiration modes.
Bits[3:0]GPIOC[4:1]: GPIO control (corresponding GPIOD)
These bits determine if the corresponding GPIOD pin is an input or output.
0 = Output
1 = Input (default)
PACE: PACE Detect Register
Address = 15h
BIT 7BIT 6BIT 5BIT 4BIT 3BIT 2BIT 1BIT 0
000PACEE1PACEE0PACEO1PACEO0PD_PACE
This register provides the PACE controls that configure the channel signal used to feed the external PACE detect
circuitry. See the Pace Detect subsection of the ECG-Specific Functions section for details.
Bits[7:5]Must always be set to '0'
Bits[4:3]PACEE[1:0]: PACE even channels
These bits control the selection of the even number channels available on TEST_PACE_OUT1. Note that only one channel
may be selected at any time.
00 = Channel 2 (default)
01 = Channel 4
10 = Channel 6, ADS1296/8/6R/8R only
11 = Channel 8, ADS1298/8R only
Bits[2:1]PACEO[1:0]: PACE odd channels
These bits control the selection of the odd number channels available on TEST_PACE_OUT2. Note that only one channel
may be selected at any time.
00 = Channel 1 (default)
01 = Channel 3
10 = Channel 5, ADS1296/8/6R/8R only
11 = Channel 7, ADS1298/8R only
Bit 0PD_PACE: PACE detect buffer
This bit is used to enable/disable the PACE detect buffer.
0 = PACE detect buffer turned off (default)
1 = PACE detect buffer turned on
This register provides the controls for the respiration circuitry; see the Respiration section for details.
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
Bit 7RESP_DEMOD_EN1: Enables respiration demodulation circuitry (ADS1294R/6R/8R only, for ADS1294/6/8 always
Bit 6RESP_MOD_EN1: Enables respiration modulation circuitry (ADS1294R/6R/8R only, for ADS1294/6/8 always write '0')
Bit 5Reserved
Bits[4:2]RESP_PH[2:0]: Respiration phase
Bits[1:0]RESP_CTRL[1:0]: Respiration control
(1) RESP_PH[2:0] phase control bits only for internal respiration (RESP_CTRL = 10) and external respiration (RESP_CTRL = 01) modes
when the CONFIG4.RESP_FREQ[2:0] register bits are 000b or 001b.
write '0')
This bit enables/disables the demodulation circuitry on channel 1.
0 = RESP demodulation circuitry turned off (default)
1 = RESP demodulation circuitry turned on
This bit enables/disables the modulation circuitry on channel 1.
0 = RESP modulation circuitry turned off (default)
1 = RESP modulation circuitry turned on
Must always be set to '1' for ADS1294R/6R/8R. This bit does not have any effect for the ADS1294/6/8.
(1)
These bits control the phase of the respiration demodulation control signal. (GPIO4 is out-of-phase with GPIO3 by the
phase determined by the RESP_PH bits.)
These bits set the mode of the respiration circuitry.
00 = No respiration (default)
01= External respiration
10 = Internal respiration with internal signals
11 = Internal respiration with user-generated signals
Bits[7:5]RESP_FREQ[2:0]: Respiration modulation frequency
These bits control the respiration control frequency when RESP_CTRL[1:0] = 10 or RESP_CTRL[1:0] = 10
000 = 64kHz modulation clock
001 = 32kHz modulation clock
010 = 16kHz square wave on GPIO3 and GPIO04. Output on GPIO4 is 180 degree out of phase with GPIO3.
011 = 8kHz square wave on GPIO3 and GPIO04. Output on GPIO4 is 180 degree out of phase with GPIO3.
100 = 4kHz square wave on GPIO3 and GPIO04. Output on GPIO4 is 180 degree out of phase with GPIO3.
101 = 2kHz square wave on GPIO3 and GPIO04. Output on GPIO4 is 180 degree out of phase with GPIO3.
110 = 1kHz square wave on GPIO3 and GPIO04. Output on GPIO4 is 180 degree out of phase with GPIO3.
111 = 500Hz square wave on GPIO3 and GPIO04. Output on GPIO4 is 180 degree out of phase with GPIO3.
Modes 000 and 001 are modulation frequencies in internal and external respiration modes. In internal respiration mode, the
control signals appear at the RESP_MODP and RESP_MODN terminals. All other bit settings generate square waves as
described above on GPIO4 and GPIO3.
(1) These frequencies assume f
= 2.048MHz.
CLK
www.ti.com.cn
(1)
.
Bit 4Must always be set to '0'
Bit 3SINGLE_SHOT: Single-shot conversion
This bit sets the conversion mode.
0 = Continuous conversion mode (default)
1 = Single-shot mode
Bit 2WCT_TO_RLD: Connects the WCT to the RLD
This bit connects WCT to RLD.
0 = WCT to RLD connection off (default)
1 = WCT to RLD connection on
Bit 1PD_LOFF_COMP: Lead-off comparator power-down
This bit powers down the lead-off comparators.
0 = Lead-off comparators disabled (default)
1 = Lead-off comparators enabled
INPUT MULTIPLEXER (REROUTING THE RIGHT LEG DRIVE SIGNAL)
The input multiplexer has ECG-specific functions for the right leg drive signal. The RLD signal is available at the
RLDOUT pin once the appropriate channels are selected for the RLD derivation, feedback elements are installed
external to the chip, and the loop is closed. This signal can be fed after filtering or fed directly into the RLDIN pin
as shown in Figure 51. This RLDIN signal can be multiplexed into any one of the input electrodes by setting the
MUX bits of the appropriate channel set registers to 110 for P-side or 111 for N-side. Figure 51 shows the RLD
signal generated from channels 1, 2, and 3 and routed to the N-side of channel 8. This feature can be used to
dynamically change the electrode that is used as the reference signal to drive the patient body. Note that the
corresponding channel cannot be used and can be powered down.
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
(1) Typical values for example only.
Figure 51. Example of RLDOUT Signal Configured to be Routed to IN8N
INPUT MULTIPLEXER (MEASURING THE RIGHT LEG DRIVE SIGNAL)
Also, the RLDOUT signal can be routed to a channel (that is not used for the calculation of RLD) for
measurement. Figure 52 shows the register settings to route the RLDIN signal to channel 8. The measurement is
done with respect to the voltage on the RLDREF pin. If RLDREF is chosen to be internal, it would be at (AVDD +
AVSS)/2. This feature is useful for debugging purposes during product development.
In the standard 12-lead ECG, WCT voltage is defined as the average of Right Arm (RA), Left Arm (LA), and Left
Leg (LL) electrodes. This voltage is used as the reference voltage for the measurement of the chest leads. The
ADS129x has three integrated low-noise amplifiers that generate the WCT voltage. Figure 53 shows the block
diagram of the implementation.
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
Figure 53. WCT Voltage
The devices provide flexibility to choose any one of the eight signals (IN1P to IN4N) to be routed to each of the
amplifiers to generate the average. Having this flexibility allows the RA, LA, and LL electrodes to be connected to
any input of the first four channels depending on the lead configuration.
Each of the three amplifiers in the WCT circuitry can be powered down individually with register settings. By
powering up two amplifiers, the average of any two electrodes can be generated at the WCT pin. Powering up
one amplifier provides the buffered electrode voltage at the WCT pin. Note that the WCT amplifiers have limited
drive strength and thus should be buffered if used to drive a low-impedance load.
See Table 5 for performance when using any 1, 2, or 3 of the WCT buffers.
As can be seen in Table 5, the overall noise reduces when more than one WCT amplifier is powered up. This
noise reduction is a result of the fact that noise is averaged by the passive summing network at the output of the
amplifiers. Powering down individual buffers gives negligible power savings because a significant portion of the
circuitry is shared between the three amplifiers. The bandwidth of the WCT node is limited by the RC network.
The internal summing network consists of three 30kΩ resistors and a 80pF capacitor. It is recommended that an
external 100pF capacitor be added for optimal performance. The effective bandwidth depends on the number of
amplifiers that are powered up, as shown in Table 5.
The WCT node should be only be used to drive very high input impedances (typically greater than 500MΩ).
Typical application would be to connect this WCT signal to the negative inputs of a ADS129x to be used as a
reference signal for the chest leads.
As mentioned previously in this section, all three WCT amplifiers can be connected to one of eight analog input
pins. The inputs of the amplifiers are chopped and the chopping frequency varies with the data rates of the
ADS129x. The chop frequency for the three highest data rates scale 1:1. For example, at 32kSPS data rate, the
chop frequency is 32kHz in HR mode with WCT_CHOP = 0. The chopping frequency of the four lower data rates
is fixed to 4kHz. When WCT_CHOP = 1, the chop frequency is fixed to highest data rate (that is, f
frequency, as shown in Table 13. The chop frequency shows itself at the output of the WCT amplifiers as a small
square wave riding on dc. The amplitude of the square wave is the offset of the amplifier and is typically 5mVPP.
This artifact as a result of chopping is out-of-band and thus does not interfere with ECG-related measurements.
As a result of the chopping function, the input current leakage on the pins with WCT amplifiers connected sees
increased leakage currents at higher data rates and as the input common voltage swings closer to 0V (AVSS),
as described in Figure 54.
Note that if the output of a channel connected to the WCT amplifier (for example, the V lead channels) is
connected to one of the PACE amplifiers for external PACE detection, the artifact of chopping appears at the
PACE amplifier output.
www.ti.com.cn
/16)
MOD
Figure 54. WCT Input Leakage Current versusFigure 55. WCT Input Leakage Current versus
In the typical implementation of the 12-lead ECG with eight channels, the augmented leads are calculated
digitally. In certain applications, it may be required that all leads be derived in analog rather than digital. The
ADS1298/8R provides the option to generate the augmented leads by routing appropriate averages to channels
5 to 7. The same three amplifiers that are used to generate the WCT signal are used to generate the Goldberger
Central Terminal signals as well. Figure 56 shows an example of generating the augmented leads in analog
domain. Note that in this implementation it takes more than eight channels to generate the standard 12 leads.
Also, this feature is not available in the ADS1296/6R and ADS1294/4R.
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
Figure 56. Analog Domain Augmented Leads
Right Leg Drive with the WCT Point
In certain applications, the out-of-phase version of the WCT is used as the right leg drive reference. The
ADS1298 provides the option to have a buffered version of the WCT terminal at the RLD_OUT pin. This signal
can be inverted in phase using an external amplifier and used as the right leg drive. Refer to the Right Leg Drive
Patient electrode impedances are known to decay over time. It is necessary to continuously monitor these
electrode connections to verify a suitable connection is present. The ADS129x lead-off detection functional block
provides significant flexibility to the user to choose from various lead-off detection strategies. Though called
lead-off detection, this is in fact an electrode-off detection.
The basic principle is to inject an excitation signal and measure the response to determine if the electrode is off.
As shown in the lead-off detection functional block diagram in Figure 59, this circuit provides two different
methods of determining the state of the patient electrode. The methods differ in the frequency content of the
excitation signal. Lead-off can be selectively done on a per channel basis using the LOFF_SENSP and
LOFF_SENSN registers. Also, the internal excitation circuitry can be disabled and just the sensing circuitry can
be enabled.
DC Lead-Off
In this approach, the lead-off excitation is with a dc signal. The dc excitation signal can be chosen from either a
pull-up/pull-down resistor or a current source/sink, shown in Figure 57. The selection is done by setting the
VLEAD_OFF_EN bit in the LOFF register. One side of the channel is pulled to supply and the other side is pulled
to ground. The pull-up resistor and pull-down resistor can be swapped (as shown in Figure 58) by setting the bits
in the LOFF_FLIP register. In case of current source/sink, the magnitude of the current can be set by using the
ILEAD_OFF[1:0] bits in the LOFF register. The current source/sink gives larger input impedance compared to the
10MΩ pull-up/pull-down resistor.
www.ti.com.cn
Figure 57. DC Lead-Off Excitation OptionsFigure 58. LOFF_FLIP Usage
Sensing of the response can be done either by looking at the digital output code from the device or by monitoring
the input voltages with an on-chip comparator. If either of the electrodes is off, the pull-up resistors and/or the
pull-down resistors saturate the channel. By looking at the output code it can be determined that either the P-side
or the N-side is off. To pinpoint which one is off, the comparators must be used. The input voltage is also
monitored using a comparator and a 4-bit DAC whose levels are set by the COMP_TH[2:0] bits in the LOFF
register. The output of the comparators are stored in the LOFF_STATP and LOFF_STATN registers. These two
registers are available as a part of the output data stream. (See the Data Output Protocol (DOUT) subsection of
the SPI Interface section.) If dc lead-off is not used, the lead-off comparators can be powered down by setting
the PD_LOFF_COMP bit in the CONFIG4 register.
An example procedure to turn on dc lead-off is given in the Lead-Off subsection of the Guide to Get Up and
This method uses an out-of-band ac signal for excitation. The ac signal is generated by alternatively providing
pull-up resistors and pull-down resistors at the input with a fixed frequency. The ac signal is passed through an
anti-aliasing filter to avoid aliasing. The frequency can be chosen by the FLEAD_OFF[1:0] bits in the LOFF
register. The excitation frequency is a function of the output data rate and is fDR/4. This out-of-band excitation
signal is passed through the channel and measured at the output.
Sensing of the ac signal is done by passing the signal through the channel to digitize it and measure at the
output. The ac excitation signals are introduced at a frequency that is above the band of interest, generating an
out-of-band differential signal that can be filtered out separately and processed. By measuring the magnitude of
the excitation signal at the output spectrum, the lead-off status can be calculated. Therefore, the ac lead-off
detection can be accomplished simultaneously with the ECG signal acquisition.
The ADS129x provide two modes for determining whether the RLD is correctly connected:
•RLD lead-off detection during normal operation
•RLD lead-off detection during power-up
The following sections provide details of the two modes of operation.
RLD Lead-Off Detection During Normal Operation
During normal operation, the ADS129x RLD lead-off at power-up function cannot be used because it is
necessary to power off the RLD amplifier.
RLD Lead-Off Detection At Power-Up
This feature is included in the ADS129x for use in determining whether the right leg electrode is suitably
connected. At power-up, the ADS129x provide two measurement procedures to determine the RLD electrode
connection status using either a current or a voltage pull-down resistor, as shown in Figure 60. The reference
level of the comparator is set to determine the acceptable RLD impedance threshold.
www.ti.com.cn
Figure 60. RLD Lead-Off Detection at Power-Up
When the RLD amplifier is powered on, the current source has no function. Only the comparator can be used to
sense the voltage at the output of the RLD amplifier. The comparator thresholds are set by the same LOFF[7:5]
bits used to set the thresholds for other negative inputs.
The right leg drive (RLD) circuitry is used as a means to counter the common-mode interference in a ECG
system as a result of power lines and other sources, including fluorescent lights. The RLD circuit senses the
common-mode of a selected set of electrodes and creates a negative feedback loop by driving the body with an
inverted common-mode signal. The negative feedback loop restricts the common-mode movement to a narrow
range, depending on the loop gain. Stabilizing the entire loop is specific to the individual user system based on
the various poles in the loop. The ADS129x integrates the muxes to select the channel and an operational
amplifier. All the amplifier terminals are available at the pins, allowing the user to choose the components for the
feedback loop. The circuit shown in Figure 61 shows the overall functional connectivity for the RLD bias circuit.
The reference voltage for the right leg drive can be chosen to be internally generated (AVDD + AVSS)/2 or it can
be provided externally with a resistive divider. The selection of an internal versus external reference voltage for
the RLD loop is defined by writing the appropriate value to the RLDREF_INT bit in the CONFIG3 register.
If the RLD function is not used, the amplifier can be powered down using the PD_RLD bit (see the CONFIG3:
Configuration Register 3 subsection of the Register Map section for details). This bit is also used in daisy-chain
mode to power-down all but one of the RLD amplifiers.
The functionality of the RLDIN pin is explained in the Input Multiplexer section. An example procedure to use the
RLD amplifier is shown in the Right Leg Drive subsection of the Guide to Get Up and Running section.
In certain applications, the right leg drive is derived as the average of RA, LA, and LL. This level is the same as
the WCT voltage. The WCT amplifier has limited drive strength and thus should be used only to drive very high
impedances directly. The ADS129x provide an option to internally buffer the WCT signal by setting the
WCT_TO_RLD bit in the CONFIG4 register. The RLD_OUT and RLD_INV pins should be shorted external to the
device. Note that before the RLD_OUT signal is connected to the RLD electrode, an external amplifier should be
used to invert the phase of the signal for negative feedback.
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
Figure 62. Using the WCT as the Right Leg Drive
RLD Configuration with Multiple Devices
Figure 63 shows multiple devices connected to an RLD.
The ADS129x provide flexibility for PACE detection either in software or by external hardware. The software
approach is made possible by providing sampling rates up to 32kSPS. The external hardware approach is made
possible by bringing out the output of the PGA at two pins: TESTP_PACE_OUT1 and TESTN_PACE_OUT2.
Note that if the WCT amplifier is connected to the signal path, the user sees switching noise as a result of
chopping; see the Wilson Central Terminal (WCT) section for details.
Software Approach
To use the software approach, the device must be operated at 8kSPS or more to be able to capture the fastest
pulse. Afterwards, digital signal processing can be used to identify the presence of the pacemaker pulse. The
software approach gives the utmost flexibility to the user to be able to program the PACE detect threshold on the
fly using software. This becomes increasingly important as pacemakers evolve over time. Two parameters must
be considered while measuring fast PACE pulses:
1. The PGA bandwidth shown in Table 6.
2. For a step change in input, the digital decimation filter takes 3 × tDRto settle. The PGA bandwidth determines
the gain setting that can be used and the settling time determines the data rate that the device must be
operated at.
External Hardware Approach
One of the drawbacks of using the software approach is that all channels on a single device must operate at
higher data rates. For systems where it is of concern, the ADS129x provide the option of bringing out the output
of the PGA. External hardware circuitry can be used to detect the presence of the pulse. The output of the PACE
detection logic can then be fed into the device through one of the GPIO pins. The GPIO data are transmitted
through the SPI port and loaded 2t
register bits in the PACE register, one from the odd-numbered channels and the other from the even-numbered
channels. During the differential to single-ended conversion, there is an attenuation of 0.4. Therefore, the total
gain in the PACE path is equal to (0.4 × PGA_GAIN). The PACE out signals are multiplexed with the TESTP and
TESTN signals through the TESTP_PACE_OUT1 and TESTN_PACE_OUT2 pins respectively. The channel
selection is done by setting bits[4:1] of the PACE register. If the PACE circuitry is not used, the PACE amplifiers
can be turned off using the PD_PACE bit in the PACE register.
Note that if the output of a channel connected to the WCT amplifier (for example, the V lead channels) is
connected to one of the PACE amplifiers for external PACE detection, the artifact of chopping appears at the
PACE amplifier output. Refer to the Wilson Central Terminal (WCT) section for more details.
s before DRDY goes low. Two of the eight channels can be selected using
The ADS1294R/6R/8R provide three modes for respiration impedance measurement: external respiration,
internal respiration using on-chip modulation signals, and internal respiration using user-generated modulation
signals, as shown in Table 14.
In this mode, GPIO2, GPIO3, and GPIO4 are automatically configured as outputs. The phase relationship
between the signals is shown in Figure 65. GPIO2 is the exclusive-OR of GPIO3 and GPIO4, as shown in
Figure 66. GPIO3 is the modulation signal, and GPIO4 is the de-modulation signal. While in this mode, the
general-purpose pin function of GPIO2, GPIO3, and GPIO4 is not available. The modulation frequency can be
64kHz or 32kHz using RESP_FREQ[2:0] bits in the CONFIG4 register. The remaining bit options of
RESP_FREQ[2:0] generate square waves on GPIO3 and GPIO4. The exclusive-OR out on GPIO2 is only
available on 64kHz and 32kHz modes. The phase of GPIO4, relative to GPIO3, is set by RESP_PH[2:0] bits in
the RESP register.
The mode can be used when the user implements custom respiration impedance circuitry external to the
ADS129x.
Figure 67 shows a block diagram of the internal respiration circuitry. The internal modulation and demodulator
circuitry can be selectively used. The modulation block is controlled by the RESP_MOD_EN bit and the
demodulation block is controlled by the RESP_DEMOD_EN bit. The modulation signal is a square wave of
magnitude VREFP – AVSS. In this mode, the output of the modulation circuitry is available at the RESP_MODP
and RESP_MODN terminals of the device. This availability allows custom filtering to be added to the square
wave modulation signal. In this mode, GPIO2, GPIO3, and GPIO4 can be used for other purposes. The
modulation frequency can be 64kHz or 32kHz, as set by the RESP_FREQ[2:0] bits in the CONFIG4 register. The
phase of the internal demodulation signal is set by the RESP_PH[2:0] bits in the RESP register.
The ADS1294R/6R/8R channel 1 with respiration enabled mode cannot be used to acquire ECG signals. If the
RA and LA leads are intended to measure respiration and ECG signals, the two leads can be wired into channel
1 for respiration and channel 2 for ECG signals.
In this mode GPIO2, GPIO3, and GPIO4 are automatically configured as inputs. GPIO2, GPIO3, and GPIO4
cannot be used for other purposes. The signals must be provided as described in Figure 68. The internal master
clock is not recommended in this mode.
The ADS1294R, ADS1296R, and ADS1298R channel 1 with respiration enabled mode cannot be used to
acquire ECG signals. If the RA and LA leads are intended to measure respiration and ECG signals, the two leads
can be wired into channel 1 for respiration and channel 2 for ECG signals, as shown in Figure 69.
NOTE: Patient and input protection circuitry not shown.
Figure 70 shows a respiration test circuit. Figure 71 and Figure 72 plot noise on channel 1 for the
ADS1294R/6R/8R as baseline impedance, gain, and phase are swept. The x-axis is the baseline impedance,
normalized to a 29µA modulation current (as shown in Equation 8).
Figure 70. Respiration Noise Test CircuitFigure 71. Channel 1 Noise versus Impedance for
32kHz Modulation Clock and Phase
(BW = 150Hz, Respiration Modulation Clock =
32kHz)
www.ti.com.cn
Figure 72. Channel 1 Noise versus Impedance for 64kHz Modulation Clock and Phase
is the modulation current, as calculated by (VREFP – AVSS) divided by the impedance of the
modulation circuit.(8)
For example, if modulation frequency = 32kHz, R
ACTUAL
= 3kΩ, I
ACTUAL
= 50µA, and R
NORMALIZED
= (3kΩ ×
50µA)/29µA = 5.1kΩ.
Referring to Figure 71 and Figure 72, it can be noted that gain = 4 and phase = 112.5° yield the best
performance at 6.4µVPP. Low-pass filtering this signal with a high-order 2Hz cutoff can reduce the noise to less
than 600nVPP. The impedance resolution is 600nVPP/29µA = 20mΩ.
When the modulation frequency is 32kHz, gains of 3 and 4 and phase of 112.5° and 135° are recommended.
When the modulation frequency is 64kHz, gains of 2 and 3 and phase of 135° and 157° are recommended for
The ADS129x have three supplies: AVDD, AVDD1, and DVDD. Both AVDD and AVDD1 should be as quiet as
possible. AVDD1 provides the supply to the charge pump block and has transients at f
recommended that AVDD1 and AVSS1 be star-connected to AVDD and AVSS. It is important to eliminate noise
from AVDD and AVDD1 that is non-synchronous with the ADS129x operation. Each supply of the ADS129x
should be bypassed with 1μF and 0.1μF solid ceramic capacitors. It is recommended that placement of the digital
circuits (DSP, microcontrollers, FPGAs, etc.) in the system is done such that the return currents on those devices
do not cross the analog return path of the ADS129x. The ADS129x can be powered from unipolar or bipolar
supplies.
Capacitors used for decoupling can be of surface-mount, low-cost, low-profile, multi-layer ceramic type. In most
cases, the VCAP1 capacitor can also be a multi-layer ceramic, but in systems where the board is subjected to
high- or low-frequency vibration, it is recommend to install a non-ferroelectric capacitor such as a tantalum or
class 1 capacitor (C0G or NPO). EIA class 2 and class 3 dielectrics such as (X7R, X5R, X8R, etc.) are
ferroelectric. The piezoelectric property of these capacitors can appear as electrical noise coming from the
capacitor. When using internal reference, noise on the VCAP1 node results in performance degradation.
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
. Therefore, it is
CLK
Connecting the Device to Unipolar (+3V/+1.8V) Supplies
Figure 73 illustrates the ADS129x connected to a unipolar supply. In this example, analog supply (AVDD) is
referenced to analog ground (AVSS) and digital supplies (DVDD) are referenced to digital ground (DGND).
NOTE: Place the capacitors for supply, reference, WCT, and VCAP1 to VCAP4 as close to the package as possible.
Connecting the Device to Bipolar (±1.5V/1.8V) Supplies
Figure 74 illustrates the ADS129x connected to a bipolar supply. In this example, the analog supplies connect to
the device analog supply (AVDD). This supply is referenced to the device analog return (AVSS), and the digital
supply (DVDD) is referenced to the device digital ground return (DGND).
www.ti.com.cn
NOTE: Place the capacitors for supply, reference, WCT, and VCAP1 to VCAP4 as close to the package as possible.
Figure 74. Bipolar Supply Operation
Shielding Analog Signal Paths
As with any precision circuit, careful PCB layout ensures the best performance. It is essential to make short,
direct interconnections and avoid stray wiring capacitance—particularly at the analog input pins and AVSS.
These analog input pins are high-impedance and extremely sensitive to extraneous noise. The AVSS pin should
be treated as a sensitive analog signal and connected directly to the supply ground with proper shielding.
Leakage currents between the PCB traces can exceed the input bias current of the ADS129x if shielding is not
implemented. Digital signals should be kept as far as possible from the analog input signals on the PCB.
Analog Input Structure
The analog input of the ADS129x is as shown in Figure 75.
Before device power-up, all digital and analog inputs must be low. At the time of power-up, all of these signals
should remain low until the power supplies have stabilized, as shown in Figure 76. At this time, begin supplying
the master clock signal to the CLK pin. Wait for time t
the configuration register must be programmed, see the CONFIG1: Configuration Register 1 subsection of the
Register Map section for details. The power-up sequence timing is shown in Table 17.
, then transmit a RESET pulse. After releasing RESET,
POR
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
Figure 76. Power-Up Timing Diagram
Table 17. Power-Up Sequence Timing
SYMBOLDESCRIPTIONMINTYPMAXUNIT
t
POR
t
RST
Wait after power-up until reset2
Reset low width2t
16
t
CLK
CLK
SETTING THE DEVICE FOR BASIC DATA CAPTURE
The following section outlines the procedure to configure the device in a basic state and capture data. This
procedure is intended to put the device in a data sheet condition to check if the device is working properly in the
user's system. It is recommended that this procedure be followed initially to get familiar with the device settings.
Once this procedure has been verified, the device can be configured as needed. For details on the timings for
commands refer to the appropriate sections in the data sheet. Also, some sample programming codes are added
for the ECG-specific functions.
Sample code to set dc lead-off with pull-up/pull-down resistors on all channels
WREG LOFF 0x13 // Comparator threshold at 95% and 5%, pull-up/pull-down resistor // DC lead-off
WREG CONFIG4 0x02 // Turn-on dc lead-off comparators
WREG LOFF_SENSP 0xFF // Turn on the P-side of all channels for lead-off sensing
WREG LOFF_SENSN 0xFF // Turn on the N-side of all channels for lead-off sensing
Observe the status bits of the output data stream to monitor lead-off status.
Right Leg Drive
Sample code to choose RLD as an average of the first three channels.
WREG RLD_SENSP 0x07 // Select channel 1—3 P-side for RLD sensing
WREG RLD_SENSN 0x07 // Select channel 1—3 N-side for RLD sensing
WREG CONFIG3 b’x1xx 1100 // Turn on RLD amplifier, set internal RLDREF voltage
Sample code to route the RLD_OUT signal through channel 4 N-side and measure RLD with channel 5. Make
sure the external side to the chip RLDOUT is connected to RLDIN.
WREG CONFIG3 b’xxx1 1100 // Turn on RLD amplifier, set internal RLDREF voltage, set RLD measurement bit
WREG CH4SET b’1xxx 0111 // Route RLDIN to channel 4 N-side
WREG CH5SET b’1xxx 0010 // Route RLDIN to be measured at channel 5 w.r.t RLDREF
ZHCS313I –JANUARY 2010– REVISED JANUARY 2012
PACE Detection
Sample code to select channel 5 and 6 outputs for PACE
WREG PACE b’0001 0101 // Power-up PACE amplifier and select channel 5 and 6 for PACE out
•Updated description of Analog Input section ...................................................................................................................... 24
•Changed description of Data Ready (DRDY) section ......................................................................................................... 32
•Changed description of START pin in START section ....................................................................................................... 34
•Changed conversion description in Continuous Mode section ........................................................................................... 35
•Changed Unit column in Table 10 ...................................................................................................................................... 35
•Changed conversion description in Single-Shot Mode section .......................................................................................... 36
•Added power-down recommendation to bit 7 description of CHnSET: Individual Channel Settings section ..................... 50
•Changed description of bit 5 in RESP: Respiration Control Register section .................................................................... 53
•Corrected name of bit 6 in WCT2: Wilson Central Terminal Control Register section ....................................................... 56
•Changed description of AVSS and AVDD in PAG Pin Assignments table ......................................................................... 14
•Changed title of Figure 20 .................................................................................................................................................. 18
•Changed title of Table 8 ...................................................................................................................................................... 30
•Changed description of STANDBY: Enter STANDBY Mode section .................................................................................. 40
•Changed bit name for bits 5, 6, and 7 in ID register of Table 12 ....................................................................................... 44
•Changed bit name for bits 5, 6, and 7 in ID: ID Control Register section .......................................................................... 45
•Added new paragraph to Respiration section ..................................................................................................................... 70
•Added footnote to Figure 69 ............................................................................................................................................... 73
•Changed description of solid ceramic capacitor in Power Supplies and Grounding section .............................................. 75
•Changed description of Connecting the Device to Bipolar (±1.5V/1.8V) Supplies section ................................................. 76
The marketing status values are defined as follows:
ACTIVE: Product device recommended for new designs.
LIFEBUY: TI has announced that the device will be discontinued, and a lifetime-buy period is in effect.
NRND: Not recommended for new designs. Device is in production to support existing customers, but TI does not recommend using this part in a new design.
PREVIEW: Device has been announced but is not in production. Samples may or may not be available.
OBSOLETE: TI has discontinued the production of the device.
Status
(1)
Package Type Package
Drawing
PinsPackage Qty
Eco Plan
& no Sb/Br)
(2)
Lead/
Ball Finish
SNAGCULevel-3-260C-168 HR
MSL Peak Temp
(3)
(Requires Login)
24-Jan-2012
Samples
(2)
Eco Plan - The planned eco-friendly classification: Pb-Free (RoHS), Pb-Free (RoHS Exempt), or Green (RoHS & no Sb/Br) - please check http://www.ti.com/productcontent for the latest availability
information and additional product content details.
TBD: The Pb-Free/Green conversion plan has not been defined.
Pb-Free (RoHS): TI's terms "Lead-Free" or "Pb-Free" mean semiconductor products that are compatible with the current RoHS requirements for all 6 substances, including the requirement that
lead not exceed 0.1% by weight in homogeneous materials. Where designed to be soldered at high temperatures, TI Pb-Free products are suitable for use in specified lead-free processes.
Pb-Free (RoHS Exempt): This component has a RoHS exemption for either 1) lead-based flip-chip solder bumps used between the die and package, or 2) lead-based die adhesive used between
the die and leadframe. The component is otherwise considered Pb-Free (RoHS compatible) as defined above.
Green (RoHS & no Sb/Br): TI defines "Green" to mean Pb-Free (RoHS compatible), and free of Bromine (Br) and Antimony (Sb) based flame retardants (Br or Sb do not exceed 0.1% by weight
in homogeneous material)
(3)
MSL, Peak Temp. -- The Moisture Sensitivity Level rating according to the JEDEC industry standard classifications, and peak solder temperature.
Important Information and Disclaimer:The information provided on this page represents TI's knowledge and belief as of the date that it is provided. TI bases its knowledge and belief on information
provided by third parties, and makes no representation or warranty as to the accuracy of such information. Efforts are underway to better integrate information from third parties. TI has taken and
continues to take reasonable steps to provide representative and accurate information but may not have conducted destructive testing or chemical analysis on incoming materials and chemicals.
TI and TI suppliers consider certain information to be proprietary, and thus CAS numbers and other limited information may not be available for release.
In no event shall TI's liability arising out of such information exceed the total purchase price of the TI part(s) at issue in this document sold by TI to Customer on an annual basis.