> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay
particular attention to the safety instructions and warning advice
in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.
Warnings
Always pay attention to information that is marked by the following
warnings with warning pictograms. Implement the specified
precautionary measures.
Representation Explanation
WARNING
NOTICE
Symbols and writing standards
Represen-
Indicates potential serious injuries
indicates circumstances that may lead to
damage to the products
Explanation
tation
Note: Basic or further information.
1. ...
Action: more steps, the sequence must be followed.
2. ...
> ... Action: a step or an optional step.
- ... Result of an action.
Menu Elements of the instrument, the instrument displays
[OK] Control keys of the instrument or buttons of the
program interface.
... | ... Functions/paths within a menu.
2.2. Ensure safety
“...” Example entries
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not
use any force.
> Do not operate the instrument if there are signs of damage at
the housing, mains unit or feed lines.
The thermal imager must not be directed at the sun or other
sources of intense radiation (e.g. objects with temperatures greater
than 500 °C, 800 °C while using the high-temperature filter) during
operation. This may lead to serious damage to the detector. The
manufacturer accepts no warranty for this type of damage to the
microbolometer detector.
> The objects to be measured or the measurement environment
may also pose risks: Note the safety regulations valid in your
area when performing the measurements.
> Do not store the product together with solvents. Do not use any
desiccants.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts from
Testo.
> Improper use of rechargeable batteries can lead to destruction
or injuries by means of current surges, fire or escaping
chemicals. The following instructions must be observed to avoid
such hazards:
• Only use in accordance with the directions in the instruction
manual.
• Do not short, take apart or modify.
• Do not expose to heavy impacts, water, fire or temperatures
above 60 °C.
• Do not store in the proximity of metal objects.
• Do not use leaky or damaged rechargeable batteries. In the
event of contact with battery acid: Thoroughly wash affected
area with water and consult a doctor, if necessary.
The testo 875i is a handy and robust thermal imager. It opens the
door to contactless determination and illustration of the temperature
distribution on surfaces for you.
Typical areas of application are:
• Building inspection (heating, ventilation and air conditioning
trade, company technicians, engineering offices, experts):
Energy evaluation of buildings
• Preventative maintenance (servicing): Mechanical and electrical
inspection of systems and machines
• Production monitoring (quality assurance): Monitoring
production processes
The testo 875i is available in several versions, which are tailored to
various usage requirements:
• testo 875-1i: High-quality wide-angle lens 32° x 23, detector
160 x 120, NETD < 50 mK, 2 GB SD card for approx. 1000
images, minimum focus distance 10 cm, integrated digital
camera
• testo 875-2i: High-quality wide-angle lens 32° x 23°, detector
160 x 120, NETD < 50 mK, 2 GB SD card for approx. 1000
images, minimum focus distance 10 cm, integrated digital
camera with power LEDs for illumination, isotherm, min./max.
on area, audio comments, display of surface moisture
distribution (real-time display with optional radio humidity
probe), high-temperature filter (optional)
3 Specifications
Export control
Thermal imagers may be subject to the export restrictions of the
European Union.
The testo 875i with 33 Hz in particular can only be exported to
countries outside of the European Union with an export permit from
the authorities.
Please observe the national regulations for export control when
exporting.
0 - 350 °C (32 - 662 °F)
Measuring range 3 (optional, testo 875-2i
only):
350 - 550 °C (662 - 1022 °F)
Accuracy
With measuring range 1 activated, for
readings within the range of -20 - 100 °C
(-4 - 212 °F):
±2 °C (±3.6 °F)
With measuring range 2 activated, for
readings within the range of 0 - 350 °C
(32 - 662 °F):
±2 °C (±3.6 °F) or ±2 % of reading
(whichever is the greater)
With measuring range 3 activated (optional,
testo 875-2i only), for readings within the
range of 350 - 550 °C (662 - 1022 °F):
±3 % of reading
Reproducibility
±1 °C (±1.8 °F) or ±1 % (whichever is the
greater)
Measuring point
minimum diameter
On-time (time to
Standard lens: 10 mm at 1 m
Telephoto lens: 3 mm at 1 m
30 s
image)
Humidity and
ambient temperature
0…100%rF,
-20…70°C (-4…158°F)
measurement with
radio probe (optional,
only testo 875-2i)
Standard measurement (1-point), 2-point
measurement, Cold-/Hotspot, solar (manual
entry of the radiation intensity)
Also with testo 875-2i: isotherm, min./max.
on area, Display of humidity image via
manual entry of the ambient humidity /
temperature (real-time display with optional
radio humidity probe)
4 Metric thread: For fastening the included tripod adapter.
Do not use desktop tripods, danger of tilting!
5 Right interface terminal: Radio module compartment.
6 2 LEDs: For illumination for the visual image (testo 875-2i only).
7 Digital camera lens: For recording visual images
8 Laser (not available in all countries): For aiming at the
measuring field.
Laser radiation. Do not look into the beam.
CLASS 2 LASER
9 Infrared camera lens: For recording thermography images.
Exchangeable lens (testo 875-2i only).
10 Rotating ring Manual focusing: For manual focusing.
11 Lens lock ring
12 [Trigger]: For recording (freezing/saving) images.
13 Left interface terminal: Memory card slot. Headset slot. USB
interface. Mains socket, for connecting the included mains unit.
Buffer battery slot.
14 Status LED: off (no mains unit connected), flashes (mains unit
connected and rechargeable battery charging), lights up (mains
4.2. Basic properties
unit connected and battery charging process completed).
14
Voltage supply
The power is supplied to the instrument via an exchangeable
rechargeable battery or via the included mains unit.
With an attached mains unit, the power is supplied automatically
via the mains unit and the instrument's rechargeable battery is
charged (only at ambient temperatures from 0 to 45 °C).
Charging the battery is also possible using a desktop charging
station (accessory).
The instrument is equipped with a buffer battery (type CR 1632) to
maintain the system data during an interruption in the power
supply (e.g. when the battery is changed).
2. Connect the mains unit to the mains socket ( ) (2).
3. Connect the mains plug to a mains socket.
- The thermal imager starts automatically.
When charging the rechargeable battery, the imager can
remain switched on or be switched off. This has no effect
on the length of the charging process.
- The charging of the rechargeable battery begins.
- The charge status is indicated on the status LED:
• LED flashes: Charging in process.
• LED lights up: Battery charged, charging process finished.
4. Charge the rechargeable battery completely, then disconnect
the instrument from the mains unit.
- After the initial charging of the battery, the thermal imager is
ready for operation.
Charging the battery is also possible using a desktop charging
station (accessory).
Rechargeable battery care:
• Do not fully exhaust rechargeable batteries.
• Store rechargeable batteries only in charged condition
and at low temperatures, but not below 0°C (best
storage conditions with a charge level of 50-80%, at an
ambient temperature of 10-20°C, recharge completely
before use).
• During longer breaks, you should discharge and
recharge the batteries every 3-4 months. Trickle
charging should not exceed 2 days
1. Open the cover on the left side of the thermal imager.
2. Insert memory card (SD card) into the card slot (SD) (1).
> To remove the memory card: Press on the memory card to
release the lock.
Inserting a radio module (accessory, testo 875-2i only)
Using a radio module (accessory) the instrument can be connected
with a radio humidity probe. The radio module must be connected
before switching on the instrument in order for the instrument to be
identified.
Radio probes may only be used in countries in which
they are permitted (see separate instructions for use of
radio probe).
> Switch off instrument before inserting the radio module.
1. Open the cover on the right side of the thermal imager.
2. Release the lock (1) and open the retaining clip (2).
3. Insert radio module into the module slot (3).
> To remove the radio module: push radio module down slightly
1. Open the cover on the left side of the thermal imager.
2. Insert stereo jack of the headset into the headset slot (1).
Mount/remove infrared protection glass
Installation:
1. Attach the protection glass (with black fitting) fastened to the
red mounting ring to the lens and turn the mounting ring
clockwise to the stop.
2. Remove the red mounting ring from the protection glass.
Removal:
1. Attach the red mounting ring to the protection glass.
2. Turn the mounting ring anticlockwise and remove the protection
glass.
After mounting/removing the protection glass, activate or deactivate
the Protection glass option, see Optics..., page 34. If this option is
set incorrectly, the specified measuring accuracy is not ensured.
1. Attach the high-temperature filter (with red fitting) fastened to
the red mounting ring to the lens and turn the mounting ring
clockwise to the stop.
2. Remove the red mounting ring from the high-temperature filter.
Removal:
1. Attach the red mounting ring to the high-temperature filter.
2. Turn the mounting ring anticlockwise and remove the hightemperature filter.
Activate/deactivate the high-temperature measuring range (350.0 -
550.0 °C or 662.0 - 1022 °F), see Measuring range, page 28, after
mounting or removing the high-temperature filter, otherwise the
specified measuring accuracy is not ensured
Mount tripod adapter
Using the included tripod adapter, the imager can be fastened to
the testo tripod (accessory) or to a commercially available standard
tripod. With the tripod adapter mounted, the battery cannot be
changed!
1. Place the tripod adapter on the lower end of the handle and
screw on using the included Allen key (ISO 2936, size 4).
2. Slide thermal imager into the fastening plate of the testo tripod
and lock or fasten it to the commercially available standard
tripod (screw-type fastening).
Mounting Softcase
The Softcase combines the functions of instrument protection, Sun
Shield and carrying function (strap).
1. Set the Softcase on the imager from above (1) and pull both
sides of the Softcase down around the display (2).
2. Guide Velcro strip through the two plates (3) and close (4).
Changing lens
Only lenses that were adjusted to the respective thermal imager
can be used. The serial number on the lens must match the serial
number of the instrument, see Optics..., page 34.
> Switch instrument off before changing the lens.
> To keep the lens from falling out: Hold the instrument with the
lens pointing up.
1. Turn lens lock ring anticlockwise to the stop (approx. 2 cm).
2. Remove lens.
3. Insert new lens, taking care to align the white markings on the
Always store lenses not in use in the container designed for this in
the case.
> Turn the lens ring clockwise to the stop, place the lens in the
container and close it.
The following steps are only relevant for USA:
5. Note or remember the serial number (Serial no.) which is
displayed on the lens.
6. Press [].
If you have changed from a wide-angle to a telephoto lens or vice
versa:
- The information No lens/wrong lens appears.
> Press [OK] to confirm the information.
If you have changed between two wide-angle or two telephoto
lenses:
> [OK] | Configuration… | [OK] | Optics… | [OK].
- The dialog Optics opens.
7. Move joystick up/down to select the lens used.
- The lens selected is outlined in orange.
8. Press function key [Apply].
- The information Restart instrument appears.
9. Press [
].
- The instrument switches off.
> To switch on the instrument: Press [].
Switch the imager on/off
1. Remove protection cap from the lens.
2. Press [].
- The start screen appears on the display. The type designation
of the imager and the firmware version are shown.
- The measurement view is opened after the warm-up period.
- The imager performs an automatic zeroing approx. every 60
sec. This can be recognized by means of a "click". The image is
frozen briefly when this occurs.
> To switch off: Press [
- The display goes out and the imager is switched off.
].
Manually focusing the image
> Turn Rotating ring of the lens until the image is in focus.
1. Move Joystick left or right to open the Configure key function
list for the left or right quick select button.
2. Move Joystick up/down to select the desired function.
Function Description
Image type Change display mode: Infrared image, real
image or infrared/real image.
Laser Switch on laser for homing in on the
measuring surface (hold button down).
Light (testo 875-2i
only)
Switch on/off LEDs for lighting up objects for
the digital camera.
2
3
Palette Change colour palette for the infrared
image.
Emissivity...
Scale...
Adjustment
3. Press [OK] to activate the selected function.
Set emissivity and reflected temperature.
Set scale limits.
Perform manual zero point calibration.
Navigating in the menu
1. Press [OK] button.
- The Menu is opened.
2. Select navigation/function:
• Move Joystick up or down to select the menu item.
• Press [OK] to confirm the selection. Or for menu items that
are marked with an arrow (►): Move Joystick to the right.
• Move Joystick to the left to return to the previous menu
level.
• Press [OK] to select the function.
• Press [ESC] to cancel the process and change to
Measuring Mode.
2
Real image and infrared lens have a different angle of view due to the spacial
separation. The image areas therefore do not correspond completely (parallax
error).
3
Laser and lens have a different angle of view due to the spacial separation.
The laser sighting and the centre of the image thus do not correspond
completely (parallax error).
2. Select left or right Temperature 1 or Temperature 2 quick
select button to activate one of the two crosshairs.
- The activated crosshairs are shown red.
3. Use the Joystick to move the activated crosshairs in order to
read off the individual temperatures.
6 Using the product
Cold-/Hotspot
Cold-/Hotspot shows the minimum and maximum temperature
point on the display. If Cold-/Hotspot is activated (), the quick
select buttons with the Coldspot and Hotspot functions can be
permanently assigned and can no longer be changed.
With frozen images, the crosshairs can be moved and the
min./max. point is visible. The min./max. point is not saved.
With saved images, Cold-/Hotspot must be activated for the
respective opened image.
> [OK] | Measurement | Cold-/Hotspot | [OK].
> Select left or right Coldspot or Hotspot quick select button to
Min./Max. on area (testo 875-2i only)
Min./Max. on area shows the minimum and maximum temperature
value within the area selection shown on the display. The area
selection is neither scalable nor adjustable. If Min./Max. on area is
activated (), the quick select buttons with the Scale... and
Emissivity... functions can be permanently assigned and can no
longer be changed.
With frozen images, the crosshairs can be moved. The Min./Max.
on area values are saved together with every image.
With saved images, Min./Max on area must be activated for the
respective opened image.
> [OK] | Measurement | Min./Max. on area | [OK].
Isotherm (testo 875-2i only)
Isotherm shows temperatures in a selectable temperature range in
a uniform colour. If Isotherm is activated (), the quick select
buttons with the Isotherm and Emissivity... functions can be
permanently assigned and can no longer be changed.
With frozen images, the crosshairs can be moved.
> [OK] | Measurement | Isotherm | [OK].
To set the isotherm colour and range:
1. [Isotherm]
- The Isotherm dialogue is opened.
2. Move Joystick left/right to select the desired option: Isotherm
- The selected option has an orange border ( ).
3. Move Joystick up/down to change the colour or the value(s).
- Changes are applied immediately so that the effects on the
4. Close dialogue with [OK] or [Esc].
- The changes are saved.
activate/deactivate the respective function.
colour (), lower limit value ( ), temperature range ( ) or
upper limit value ( ).
display of the infrared image can be checked directly.
Moisture (testo 875-2i only)
The ambient temperature and humidity
measured by the (optional) radio humidity probe are used to
calculate the relative surface moisture for each pixel. The values
are shown as a humidity image. The special colour palette
indicates which areas are at risk of mould:
Colour Surface moisture Assessment
green 0…64%RH non-critical
yellow/or
65…80%RH potentially critical
ange
red >80%RH critical
If this function is enabled (), the shortcut buttons are
permanently assigned the functions Humidity (only if no radio
humidity probe is logged on) and Emission level….
The values for relative humidity and ambient temperature are
shown at the top of the display.
1. [OK] | Measurement | [OK] | Humidity | [OK].
- If a radio humidity probe is logged on: No further information is
required.
- If no radio humidity probe is logged on: The Humidity dialogue
box is opened. The values for the ambient temperature and
humidity can be entered.
2. Move Joystick up/down to select the desired option.
- The selected option has an orange border.
3. Confirm the selection with [OK].
- The selected number is highlighted orange.
4. Move Joystick up/down to set the desired value. Move
Joystick right/left to change between the numbers.
5. Confirm the entry with [OK].
6. Confirm the settings with Apply.
Solar
To detect and document failures in photovoltaic systems, the solar
function can be selected. The sun's radiation intensity measured
(with an external instrument) can be entered for documentation
purposes. This value is saved with the IR- image.
If this function is enabled (), the shortcut buttons are
permanently assigned the functions Intensity and Hotspot.
The radiation intensity value is shown at the top of the display.
3. Move Joystick up/down to set the required value. Move
Joystick right/left to switch between the numbers.
4. Confirm entry with [OK].
5. Confirm settings with [Apply].
Measuring range
The measuring range can be set to adapt to the respective area of
application.
1. [OK] | Measurement | [OK] | Measurement range.
2. Select the desired temperature range and confirm with [OK].
For measuring range 350 - 550 °C (662 - 1022 °F):
• The high-temperature filter must be installed,
dg_ref_source_inline>HT-Filter
montieren</dg_ref_source_inline>.
• With temperature differences larger than 350 K the use
of the Ironbow HT palette is recommended as it
enables lower temperatures to be shown with higher
contrast, see Palette, page 33see Mounting/removing
high-temperature filter (testo 875-2i only), page 19.
- A confirmation request appears as to whether all files on the
memory card should be deleted. Folders are not deleted.
2. Confirm erase with [OK] or cancel process with [Esc].
6.1.3.Scale...
Set scale limits
You can choose between auto-scaling (continuous, automatic
adjustment to the current min./max. values) and manual scaling.
The scale limits can be set within the activated measuring range
(see Measuring range, page 28). All temperatures that undershoot
or exceed the limit value are shown in the colour of the limit value
(depending on the selected colour palette, see Palette, page 33).
Temperature ranges that are not relevant can thus be hidden.
2. Move Joystick left/right to select the desired option: Auto-
scaling ( ), min. value ( ), temperature range ( ) or max.
value ( )
- The selected option has an orange border ( ).
3. With the selection of min. value, temperature range or max.
value: Move Joystick up/down to change the value(s).
- Changes are applied immediately so that the effects on the
display of the infrared image can be checked directly.
4. Close dialogue with [OK] or [Esc].
- The changes are saved.
6.1.4.Display...
Select display options
The presentation can be adapted by means of showing/hiding the
following elements: Crosshairs, temperature scale and quick select
buttons.
Hidden function keys can be shown again by pressing a quick
select button: The first press of the button shows the function keys
again. To trigger a function, the button must be pressed again.
1. [OK] | Display… | [OK].
- The Display options dialogue is opened.
2. Move Joystick up/down to select the desired option.
- The selected option has an orange border ( ).
3. Activate () or deactivate () option with [OK].
4. Confirm settings with the [Apply] quick select button.
or
Discard settings with [Esc].
6.1.5.Emissivity...
Emissivity
The emissivity describes the capability of a body to emit
electromagnetic radiation. This is material-specific and must be
adapted for correct measurement results.
Non-metals (paper, ceramic, gypsum, wood, paints and coatings),
plastics and food have high emissivity, which means that the
surface temperature can be easily measured using infrared.
Because of their low or non-uniform emissivity, bright metals and
metal oxides are only somewhat suited for infrared measurement.
You must anticipate highly inaccurate measurements. A remedy for
this is coatings that increase the emissivity, e.g. paint or emission
Using this offset factor, the reflection is calculated out due to the
low emissivity and the accuracy of the temperature measurement
with infrared instruments is improved. In most cases, the reflected
temperature is identical to the ambient air temperature. Only when
objects with strong emissions with much higher temperatures (such
as ovens or machines) are in the proximity of the object being
measured should the radiation temperature of theses heat sources
be determined and used (e.g. using a globe thermometer). The
reflected temperature has only little effect on objects with high
emissivity.
Set emissivity/reflected temperature
You can choose between a user-defined emissivity and 8 materials
with recorded emissivities.
The reflected temperature can be set individually.
1. [OK] | Emissivity… | [OK].
- The Emissivity dialogue is opened.
2. Move Joystick up/down to select the desired material.
- The selected material is highlighted orange ( ). The
3. Move Joystick to the right to switch to the setting of the
- The selected option has an orange border ( ).
4. Confirm the selection with [OK].
- The selected number is highlighted orange ( ).
5. Move Joystick up/down to set the desired value. Move
6. Confirm the entry with [OK].
7. Confirm the settings with Apply.
- The changes are saved.
The designation of the materials in the selection list is
configured during commissioning of the thermal imager in
the language selected by the user. A translation does not
occur in the event of subsequent changeover of the
language in the instrument.
Using the PC software, other materials can be imported into
the instrument from an existing list. The language for the
materials here depends on the operating system language
of the PC.
corresponding emissivity is shown to the right next to the
material.
emissivity (only available when selecting User defined) or to
Reflected temp..
Joystick right/left to change between the numbers.
- The selected emissivity ( ) is shown at the bottom right in the
display in measurement view.
6.1.6.Palette
Change colour palette for the infrared image
You can choose between 10 existing palettes. The Ironbow HT
palette is intended specifically for measurements in the hightemperature measuring range (low temperatures are shown with
higher contrast). The currently activated option is marked with a tick
().
1. [OK] | Palette| [OK].
2. Move Joystick up/down to select the desired option.
3. Confirm the selection with [OK].
6.1.7.Configuration
Time/date...
Time and date can be set. Time and date format are automatically
set based on the selected instrument language.
The lenses adjusted to the instrument are shown. Only the lenses
that are shown may be used.
testo 875-2i: The high-temperature filter registered in the
instrument is shown. With use of another high-temperature filter,
the specified accuracy is not ensured.
> [OK] | Configuration… | [OK] | Optics… | [OK].
Using the Protection glass option, you can set whether an infrared
protection glass is used or not. Ensure the correct setting to
prevent corruption of the measurement results.
> Press [OK] to activate ( ) or deactivate ( ) the Protection
- With the option activated, the protection glass symbol ( ) is
SuperResolution (optional)
SuperResolution is a technology to improve image quality. Each
time an image is recorded, a sequence of images is saved on the
thermal imager and used to calculate a higher-resolution image
with the aid of the PC software (no interpolation). The pixel count
increases by a factor of 4, while the geometric resolution (IFOV) is
improved by a factor of 1.6.
To use the function, the imager must be hand-held (no tripod).
This function is available as an additional option (article
no. 0554 7806) and must be activated before it can be
used, if the option was not ordered at the same time as
the imager.
Activate function (only with subsequent order):
You will receive an envelope with an access code
(identification code)
When all necessary data has been entered, an
activation code is generated and can be used to
activate the function via the IRSoft PC software. Please
observe the installation requirements and steps on the
website or in the email sent to you.
> Move Joystick up/down to change the setting. Confirm the
- Switch off LCD or Power off imager: The selected option can
> Press [OK] to activate ( ) or deactivate ( ) the function.
4. Confirm the settings with [Apply].
Radio…( testo 875-2i only)
A radio humidity probe can be logged onto the instrument.
The function Radio… can only be selected if a radio module is
connected (accessory), see Inserting a radio module (accessory,
testo 875-2i only), page 18.
Every radio probe has a probe identification number (RFID). This is
made up of the three-digit RFID number on the probe and the
position of the switch in the battery compartment of the radio probe
(H or L). To be logged on at the instrument, the radio probe must
be switched on and the transfer rate must be set to 0.5 sec. This
occurs by briefly pressing the on/off button when switching on (also
see instructions for use of radio probe).
1. [OK] | Configuration… | [OK] | Radio… | [OK].
- The Radio... dialogue box is opened.
- The probe-IDs of the radio probes found are displayed.
If no radio probe was found, this may be because of the following:
• The radio probe is not switched on or the battery of the radio
• The radio probe is outside the range of the instrument.
• Sources of interference are influencing radio transmission (e.g.
> If necessary: Rectify the possible causes of the fault in radio
2. Press [OK] to enable () or disable () the radio
3. Move Joystick right to switch to a list of the radio probes found.
4. Move Joystick up/down to select a probe.
-
5. Confirm selection with Apply.
entry with [OK].
be activated or deactivated.
- With the option activated, the length of time until shut-down
can be set.
> Move Joystick to the right and press [OK]. Move Joystick
up/down to change the setting. Confirm the entry with [OK].
probe is spent.
reinforced concrete, metal objects, walls or other barriers
between transmitter and receiver, other transmitters of the
same frequency, strong electromagnetic fields).
2. Move Joystick up/down to select the desired option.
- The selected option has an orange border ( ).
3. Move Joystick left/right to change the setting.
4. Confirm the settings with [Apply].
Format…
The memory card can be formatted. During the formatting, all
saved data on the SD card is lost.
1. [OK] | Configuration… | [OK] | Format… | [OK].
- A confirmation request appears as to whether the memory card
should really be formatted.
2. Confirm with [OK] or cancel process with [Esc].
Factory reset
The instrument settings can be reset to the factory settings.
Time/date and country settings are not reset.
Please note: After a factory reset, the numbering of images begins
anew. When saving images, images that have already been saved
and have the same number are overwritten!
> Transfer all saved images onto a PC before performing a reset
Damage to the infrared detector by measuring high scene
temperatures.
> testo 875i without high-temperature filter: do not direct the
thermal imager at areas > 500°C / 932°F.
> testo 875-2i with high-temperature filter: do not direct the
thermal imager at areas > 800°C / 1472°F.
testo 875-2i: When measuring temperatures > 350°C or
> 662°F, always install the high-temperature filter, see
Mounting/removing high-temperature filter (testo 875-2i
only), page 19, and adjust the high-temperature
measurement range, see Measuring range, page 28.
With the high-temperature measuring range (350.0 -
550.0 °C or 662.0 - 1022 °F) activated, the stated accuracy
specifications are not valid for readings below 350 °C or
662 °F as this range is not calibrated.
Important framework conditions
Please observe the following framework conditions in order to
obtain significant measurement results.
Humidity measurement (only testo 875-2i):
• Pivot radio humidity probe or separate humidity measuring
instrument slightly to shorten the adjustment time. (e.g.
breathed air).
Building thermography, investigating the building shell:
• Considerable temperature difference between inside and
outside required (ideal: > 15 °C/> 27 °F).
• Consistent weather conditions, no intensive sunlight, no
precipitation, no strong wind.
To ensure the highest level of accuracy, the imager requires an
equalization period of 5 minutes after being switched on.
Important settings
Before recording an image, check whether the protection glass
option is set correctly to prevent corruption of the measurement
results, see Optics..., page 34.
Before saving an image, ensure that it is focused correctly (in
focus), see Manually focusing the image, page 21. Images that are
not in focus cannot be corrected retroactively!
To obtain precise readings, the emissivity and reflected
temperature must be set correctly, see Set emissivity/reflected temperature, Seite 32. A subsequent adjustment is possible via
the PC software as required.
With an activated auto scaling, the colour scale is continuously
adapted to the min./max. values of the current measurement
image. The colour that is to be assigned to a specific temperature
thereby changes constantly. To be able to compare multiple
images by means of the colouring, the scaling must be manually
set to fixed values, see Scale... page 29 or must be subsequently
adjusted to uniform values using the PC software.
Depending on the scene temperature, the measuring range
may be undershot or exceeded. In this case, --- or +++ is
shown instead of the readings.
> If this occurs, change the measuring range, see
Measuring range, page 28.
Select the storage location (folder)
1. While an image is frozen (fixed image): Press the left [Folder]
quick select button.
- The Folder dialogue is opened.
To create a new folder, see Create new folder, Seite 29.
2. Select the desired folder with the joystick and open with [OK].
- The folder name of the opened folder is displayed in the
header.
3. Press left [Apply] quick select button to select the opened
folder.
1. Selection position from which the recording should be
overwritten: Begin playback with [] and stop at the
desired point [].
2. Use [] to overwrite the recording as of the desired point.
• Delete recording:
> Press [].
- The complete recording is deleted.
• Exit Audio dialogue:
> Press [ESC].
Aiming at the measuring surface
Using the laser, the measuring surface can be aimed at.
WARNING
Eye damage due to laser beam!
> Do not look into the laser beam.
> Do not point the laser beam at people.
To change the quick select button functions, see Changing
assignment of the quick select buttons, page 23.
> Press and hold [Laser].
- When the laser is active, the laser symbol () flashes.
Caution: Laser and lens have a different angle of view due
to the spacial separation. The laser sighting and the centre
of the image thus do not correspond completely (parallax
error). The laser dot is therefore not suitable for marking
locations that were aimed at using the crosshairs.
To prevent losing the instrument settings: Only change the
rechargeable battery when a buffer battery is inserted in the
instrument or the mains unit is connected.
1. Press lock release button.
- The rechargeable battery is released and sticks out a little from
the rechargeable battery slot. The locking function keeps the
battery from falling out.
2. Completely remove the battery from the rechargeable battery
slot.
3. Insert a new battery into the rechargeable battery slot until this
is flush with the bottom of the handle.
- The thermal imager starts automatically.
Clean thermal imager
> If the housing of the thermal imager is dirty, clean it with a damp
cloth. Do not use any aggressive cleaning agents or solvents!
Weak household cleaning agents and soap suds may be used.
> Large dust particles can be removed with a clean optical
cleaning brush (available in specialized photography shops).
> Use a lens cleaning cloth for slight contamination. Do not use
rubbing alcohol!
7 Maintaining the product
Change buffer battery
To prevent losing the instrument settings: Only change the buffer
battery when a charged rechargeable battery is inserted or the
mains unit is connected.
1. Open the cover on the left side of the thermal imager.
2. Remove battery holder, including buffer battery, from the battery
slot.
3. Remove empty battery from the holder and insert new battery
(type CR 1632).