> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay
particular attention to the safety instructions and warning advice
in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.
Symbols and writing standards
Representation Explanation
Warning advice, risk level according to the
signal word:
Warning! Serious physical injury may occur.
Caution! Slight physical injury or damage to
the equipment may occur.
> Implement the specified precautionary
measures.
Note: Basic or further information.
1. ...
2. ...
Action: more steps, the sequence must be
followed.
> Do not operate the instrument if there are signs of damage at
the housing, mains unit or feed lines.
> Do not perform contact measurements on non-insulated, live
parts.
> Do not store the product together with solvents. Do not use any
desiccants.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts
from Testo.
> Dangers may also arise from the systems being measured or
the measuring environment: Note the safety regulations valid in
your area when performing the measurements.
Testo guarantees the functionality of its products when used in
accordance with their intended purpose. This guarantee does not apply
to features of Testo products in combination with unauthorised thirdparty products. Competitor products are not authorised by Testo.
As is common practice, Testo generally excludes support, warranty or
guarantee claims relating to functionality that has not been guaranteed
by Testo as part of the product offered. Claims shall also be excluded
in the event of improper use or handling of the products, e.g. in
combination with unauthorised third-party products.
2 Safety and the environment
For products with Bluetooth®
Changes or modifications that have been made without the explicit
consent of the responsible approval authority, may cause the
retraction of the type approval.
Data transfer may be disturbed by equipment that uses the same
ISM-band, e.g. WLAN, microwave ovens, ZigBee.
The use of radio communication links is not permitted in aeroplanes
and hospitals, among others. For this reason the following points
must be ensured before entering:
> Switch off the instrument.
> Disconnect control unit from all external power sources (mains
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
3 Specifications
regulations)
The testo 420 is used for volume flow measurements (main
application), Pitot tube measurements and pressure measurements
for air conditioning and ventilation systems. Thanks to its
interchangeable flow hoods, the testo 420 can be used for air inlets
and outlets of various sizes.
With the additional App (Android/iOS), the readings can be
conveniently displayed on a tablet or smartphone, and in addition a
4 Technical data
measurement can be started, stopped and saved via the App.
The use of the wireless module is subject to the regulations and
stipulations of the respective country of use, and the module may only
be used in countries for which a country certification has been granted.
The user and every owner has the obligation to adhere to these
regulations and prerequisites for use, and acknowledges that the resale, export, import etc. in particular in countries without wireless
permits, is his responsibility.
1 Battery compartment, on rear of the instrument
2 Display
3 Control keys
4 Probe socket Mini-DIN (only for use on the measurement base)
5 Micro-USB port
6 Connection for pressure measurement
Actual volume flow:
The current ambient conditions are used to
calculate the volume flow. The actual
barometric pressure is measured with the
internal sensor. In applications with volume
flow hood, the temperature is measured by
the integrated temperature/humidity sensor,
with Pitot tube measurement the actual
temperature needs to be input manually.
Standard volumetric flow:
The standard settings for temperature and
barometric pressure (21 °C / 1013 hPa in
accordance with the standard conditions of
the NIST, National Institute of Standards and
Technology) are used to calculate the volume
flow.
Factor by which the current reading is
multiplied. Depends on which outlet
measurements are being taken at.
The Pitot tube factor for Pitot tubes is
generally the same and must be entered:
• Pitot tubes from Testo: 1.00
• Pitot tubes from other manufacturers: the
Pitot tube factor can be found in the
instruction manual or you should ask your
supplier.
10
Control keys
Key Function
[►, ■] Holds / starts / stops a measurement
[ESC]
[►, ▲, ▼,◄] Navigation within the menu
[]
Menu
Switches to the previous view / to the
measurement view
In order to be able to establish a connection via Bluetooth,
you need a tablet or smartphone with the Testo App testo 420 already installed on it.
You can get the App for iOS instruments in the App Store
or for Android instruments in the Play Store.
Information about compatibility can be found in the relevant
App Store.
Measurements can be performed and saved with App into
the instruments memory. Measuring mode and save
function on the instrument itself is not available during
Bluetooth communication is established with a mobile
device.
Switching on Bluetooth
1. Hold ▲ down for 3 seconds.
- Once the Bluetooth icon is shown on the display, Bluetooth is
switched on.
- If no connection is established, Bluetooth switches off after 10
mins.
or
1. Press -> Device Settings -> Bluetooth, ► and using ▲/▼
-> select Off. Confirm with .
- Once the Bluetooth icon is shown on the display,
Bluetooth is switched on.
- If no connection is established, Bluetooth switches off after 10
mins.
Display Explanation
flashes
is permanently
There is no Bluetooth connection, or a
potential connection is being searched for.
If the readings fluctuate widely, it is advisable to damp the
readings. The time range for the damping can be set manually
between 5-20 seconds.
1. Press , then select Device settings and Gliding average.
- The damping can be set between 5-20 seconds.
7.2.2. Hood Calibration
This input is provided for the filing of calibration data by the relevant
calibration laboratory. The flow hood-specific adjustment data can
be input manually for supply air and exhaust air and has a direct
effect on the measurement results. Input option of 0.001-9.999.
1. Press , then select Device settings and Hood adjustment.
- Calibration of the hood can be set for supply air and exhaust air.
7.2.3. Zeroing Interval (Automatic Zeroing)
The pressure sensor carries out automatic zeroing at regular
intervals. These intervals can be set via the automatic zeroing.
1. Press , then select Device settings and Zeroing int.
- The zeroing interval can be set between 1-20 seconds.
7.3. Measuring
7.3.1. Pitot tube measurement
1. Connect the hoses to the testo 420 and to the Pitot tube.
2. Press -> Application -> Pitot tube and there set the duct
geometry, the Pitot tube factor and the temperature, and
choose between actual and standard.
3. Select the required measurement program.
4. Carry out the measurement.
With a timed/multi-point measurement program, the
required number of measuring points can be recorded
using [►, ■]. To end the measurement, press and hold
down [►, ■] for at least 3 s. At least one measuring point
must be recorded before the measurement can be
terminated.
5. Press to save the measurement data. Unsaved
measurement data is lost once you carry out the next
measurement.
- The target folder and file name are displayed, confirm with
to save the measurement data under this name and in the
selected folder.
The Pitot tube factor for Pitot tubes is generally the same
and must be entered:
Pitot tubes from Testo, Pitot tube factor: 1.00
Straight Pitot tubes from Testo, Pitot tube factor: 0.67
Air flow velocity matrix 0699.7077, Pitot tube factor: 0.82
For Pitot tubes from other manufacturers, refer to the
instruction manual for the Pitot tube factor or ask your
supplier.
7.3.2. Differential pressure measurement
1. Connect the hoses to the testo 420 at + and -.
2. Press -> Application -> Pressure only.
3. Carry out the measurement.
4. Press to save the measurement data. Unsaved
measurement data is lost once you carry out the next
measurement.
- The target folder and file name are displayed, confirm with
to save the measurement data under this name and in the
selected folder.
7 Using the product
7.4. Saving
A maximum of 99 measurements can be saved in one
folder.
> -> Memory ->
- The folder overview is shown on the display. Create a new
folder via New Folder.
A maximum 100 folders can be created.
Open folder
> Using the arrow keys, navigate to the required folder and press
.
- The selected folder is opened and the individual files are
displayed.
1. Using the arrow keys, navigate to the required folder and press
.
2. Select the menu item Delete Folder and confirm with .
- The folder that you want to delete is shown on the display.
3. Confirm once again with in order to delete the folder,
or cancel by pressing Esc.
Set as Logging Folder
This setting establishes which folder is to be specified as the
standard storage location for saving the measurements.
The folder which is set as the standard storage location is
marked in black.
1. Using the arrow keys, navigate to the required folder and press
.
2. Select the menu item Set as Logging Folder and confirm with
.
- The selected folder is set as the standard storage location.
- The stored location can be changed during the saving process.
7.5. Transferring measurement data to the PC
No Bluetooth connection is possible while there is a
connection to the PC. The existing Bluetooth connection is
aborted.
The PC identifies the testo 420 as a removable medium. In
the event of any formatting, please make sure that the FAT
format is always selected under the file system.
1. Connect the testo 420 to your PC using the USB cable.
- The testo 420 switches on automatically, a window appears on
the PC, select Open folder here. The folders and files saved on
your testo 420 are displayed. The files are available in the file
format *.txt.
L Series BLE Module (08 Mai 2013) based on TI
CC254X chip
Qualified
Design ID
B016552
Bluetooth® radio class Class 3
Bluetooth® company ID 10274
IC Warnings
This instrument complies with Part 15C of the FCC Rules and Industry Canada RSS210 (revision 8). Commissioning is subject to the following two conditions:
(1) This instrument must not cause any harmful interference and
(2) this instrument must be able to cope with interference, even if
this has undesirable effects on operation.
Cet appareil satisfait à la partie 15C des directives FCC et au standard Industrie
Canada RSS-210 (révision 8). Sa mise en service est soumise aux deux conditions
suivantes :
(1) cet appareil ne doit causer aucune interférence dangereuse et
(2) cet appareil doit supporter toute interférence, y compris des interférences qui
provoquerait des opérations indésirables.
FCC Warnings
Information from the FCC (Federal Communications Commission)
For your own safety
Shielded cables should be used for a composite interface. This is to ensure continued
protection against radio frequency interference.
FCC warning statement
This equipment has been tested and found to comply with the limits for a Class C
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference
to radio communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful interference to
radio or television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more of the
following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user's authority to operate the equipment. Shielded interface
cable must be used in order to comply with the emission limits.
Warning
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received,
including interference that may cause undesired operation.