Testo Testo 510-KIT Manual

Page 1
Instruction manual
testo 510
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Page 2
Short manual testo 510 11
Protection cap: Park positionDifferential pressure sensor
nipple connection
DisplayControl keysBattery compartment, holding
magnets (on rear)
Basic settings
Instrument off >press and hold 2s > select with ( ), confirm with ( ):
Unit of pressure/velocity: hPa, mbar, Pa, mmH2O, mmHg, psi, inH2O, inHG, m/s, fpm
> Density of the medium (Density) > Auto off function: OFF, ON
Switching the instrument on
Press .
Switching the display light on (for 10s)
Instrument on > press
Select display mode
Instrument on > select with :
Current reading > Hold: Readings are held > Max: Maximum values > Min: Minimum values > Hold Avg: Timed mean value calculation, see below
Timed mean value calculation for flow
Press several times, until until
---- appears > Releasing starts mean value calculation,
the current readings are displayed > End measurement: press the mean value is displayed.
Switching the instrument off:
Instrument on >press and hold 2s.
Hold and Avg appear > Hold down
de
enfresitptsvnl????
,
Page 3
Safety and the enviroment12
Safety and the environment
About this document
i Please read this documentation through carefully and
familiarise yourself with the product before putting it to use. Keep this document to hand so that you can refer to it when necessary. Hand this documentation on to any subsequent users of the product.
i Pay particular attention to information emphasised by the
following symbols:
Important.
Avoid personal injury/damage to equipment
i Only operate the measuring instrument properly, for its
intended purpose and within the parameters specified in the technical data. Do not use force.
i Never store the product together with solvents, acids or other
aggressive substances.
i Only carry out the maintenance and repair work that is
described in the documentation. Follow the prescribed steps when doing so. Use only OEM spare parts from Testo.
Strong magnets. Damage to other instruments! i Keep a safe distance from products which could be damaged
by magnetism (e.g. monitors, computers, pacemakers, credit cards).
Protecting the environment
i Take faulty rechargeable batteries as well as spent batteries to
the collection points provided for them.
i Send the product back to Testo at the end of its useful life. We
will ensure that it is disposed of in an environmentally friendly manner.
Page 4
Specifications 13
Specifications
Functions and use
The testo 510 is a differential pressure measuring instrument. It is normally used to measure small differences in pressure (e.g. to check filter permeability), gas flow pressure measurement, fine draught measurement as well as flow velocity measurement with a Pitot tube.
Technical data
Measurement data
· Sensor: Differential pressure sensor
· Parameters: Pa, hPa, mbar, mmH inH2O, psi, m/s, fpm
· Measuring ranges:
0...100 hPa, 0...40.15 inH
· Resolutions:
0.01 hPa, 0.01 inH
· Accuracies (Nominal temperatue 22 °C, ±1 Digit): ±0.03 hPa (0...0.30 hPa), ±0.05 hPa (0.31...1.00 hPa), ±(0.1 hPa+1.5 % of reading) (rest of range), ±0.01 inH
2
±0.02 inH2O (0.13...0.40 inH2O), ±(0.04 inH2O +1.5 % of reading) (rest of range)
· Measuring rate:
0.5 s
O, mmHg, inHG,
2
O
2
O
2
O (0...0.12 inH2O),
Further instrument data
· Protection class: IP40
· Pitot-factor: 1
· Ambient conditions:
0...50 °C, 32...122 °F
· Storage/transport conditions:
-40...70 °C, -40...158 °F
· Voltage supply: 2x 1.5 V type AAA
· Battery life: 50 h (without display light)
· Dimensions: 119x46x25mm / 4.7x1.8x1.0 in (inc. protection cap)
· Weight: 90 g / 3.2 oz (inc. batteries and protection cap)
Directives, standards and tests
· EC Directive: 2004/108/EEC
Warranty
· Duration: 2 years
· Warranty conditions: see guarantee card
de
enfresitptsvnl????
Test Equipment Depot - 800.517.8431
99 Washington Street, Melrose, MA 02176
TestEquipmentDepot.com
Page 5
Product description14
Product description
At a glance
Protection cap: Park positionDifferential pressure sensor
nipple connection
DisplayControl keysBattery compartment, holding
magnets (on rear)
First steps
Inserting batteries:
²
1 To open the battery compartment, push the battery cover
down.
2 Insert batteries (2x 1.5 V type AAA). Observe the polarity! 3 To close the battery compartment, push the battery cover
back on.
Basic settings (configuration mode):
²
Adjustable functions
· Unit of pressure/velocity: hPa, mbar, Pa, mmH2O, mmHg, psi, inH2O, inHG, m/s, fpm
· Only if a unit of velocity is selected: Density of the medium (Density): Adjust flashing numbers with , change to the next number with .
· Auto off function: OFF, ON (instrument switches off automatically if no key is pressed for 10 minutes)
1 When switching the instrument on, press and hold until
and appear on the display (configuration mode).
- The adjustable function is displayed. The current setting flashes.
2 Press ( ) several times until the desired setting flashes. 3 Press ( ) to confirm the input. 4 Repeat steps 2 and 3 for all functions.
- The instrument changes to measuring mode.
Page 6
Using the product 15
Using the product
For velocity to be measured, the unit of density and density value must be set correctly, see chapter 'First steps', section 'Basic settings (configuration mode)'.
Switching the instrument on:
²
i Press .
- Measuring mode is opened.
S
witching the display light on:
²
The instrument is switched on.
i Press .
- The display light goes out automatically if no key is pressed for 10 seconds.
Zeroing the measuring instrument:
²
The measur position of the measuring instrument. After zeroing, the position of the measuring instrument must not be changed. Carry out zeroing before every measurement in order to compensate faulty positioning or long-term zero-point drift. Zeroing is only possible in a range of 0...25% of the measuring range.
The instrument can only be zeroed in the range up to 10 hPa.
i Press .
- Zeroing takes place.
hanging the display view:
C
²
Adjustable views
· Current reading
· Hold: Readings are held.
· Max: Maximum values since the instrument was last switched on or last reset.
· Min: Minimum values since the instrument was last switched on or last reset.
· Hold Avg: Timed mean value calculation, see below.
i Press several times until the desired view appears.
ement values can be falsified by a change in the
de
enfresitptsvnl????
Page 7
Using the product16
Timed mean value calculation for flow:
²
Function only available when unit m/s or fpm is set.
1 Press several times, until
display. The last result of mean value calculation is displayed.
- If mean value calculation was carried out since the last time the instrument was switched on, the last result is displayed.
2 Hold down until
value calculation automatically, the current readings are displayed.
3 End measurement: press . The mean value is displayed.
i For further mean value calculation: hold down . 4 End mean value calculation: press briefly.
Resetting Max/Min values:
²
1 Press several times until the desired view appears. 2 Press and hold until 3 Repeat steps 1 and 2 for all values that are to be reset.
Switching the instrument off:
²
i Press and hold until the display goes out.
---- flashes. Releasing starts mean
Hold and Avg appear in the
---- appears.
Maintaining the product
Changing batteries:
±
1 To open the battery compartment, push the battery cover
down.
2 Remove used batteries and insert new batteries
(2x 1.5 V type AAA). Observe the polarity!
3 To close the battery compartment, push the battery cover
back on.
Cleaning the housing:
±
i Clean the housing with a moist cloth (soap suds) if it is dirty.
Do not use aggressive cleaning agents or solvents!
Page 8
Tips and assistance 17
Tips and assistance
Questions and answers
Question Possible causes/solutions
Hi or Lo · Readings outside the measuring range (too high, too
low): Keep to the permitted measuring range.
· Residual capacity <10 min: Change batteries.
de
enfresitptsvnl????
Page 9
Notes18
Page 10
Test Equipment Depot - 800.517.8431
99 Washington Street, Melrose, MA 02176
TestEquipmentDepot.com
0973.5102/04/Sh/dr/10.04.2008