6.3 Reading out data ...........................................................................30
7 Maintaining the product........................................................................30
7.1 Changing the batteries ..................................................................30
7.2 Cleaning the instrument.................................................................31
8 Tips and assistance...............................................................................32
8.1 Questions and answers .................................................................32
8.2 Accessories and spare parts .........................................................32
19
Page 3
2 Safety and the environment
2 Safety and the environment
2.1 About this document
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay
particular attention to the safety instructions and warning advice
in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.
Symbols and writing standards
Representation Explanation
Warning advice, risk level according to the
signal word:
Warning! Serious physical injury may occur.
Caution! Slight physical injury or damage to
the equipment may occur.
> Implement the specified precautionary
measures.
Note: Basic or further information.
1. ...
2. ...
Action: more steps, the sequence must be
followed.
> ... Action: a step or an optional step.
- ... Result of an action.
Menu Elements of the instrument, the instrument
display or the program interface.
[OK] Control keys of the instrument or buttons of
the program interface.
... | ... Functions/paths within a menu.
“...” Example entries
20
Page 4
2.2 Ensure safety
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not
use any force.
> Never use the instrument to measure on or near live parts.
> Keep the read-out contacts on the rear of the data logger in
particular away from live parts!
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts from
2.3 Protecting the environment
Testo.
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
3 Specifications
regulations) or return the product to Testo for disposal.
3 Specifications
3.1 Use
The testo 174 data loggers are used for storing and reading out
individual readings and series of measurements.
Readings are measured and stored with the testo 174 and
transmitted to the PC via the interface, where they can be read out
and evaluated using the testo ComSoft software. The data logger
can also be programmed individually via the software.
Examples of applications
testo 174T is ideally suited for temperature measurement in in
refrigerators, freezers, cold rooms and cold shelves.
testo 174H monitors the climatic conditions, e.g. in warehouses,
offices and in the manufacturing sector.
21
Page 5
3 Specifications
3.2Technical data
testo 174T
Characteristic Values
Probe type Internal NTC temperature sensor
Measuring range -30 to +70 °C
Accuracy ± 0.5 °C (-30 to +70 °C)
Resolution 0.1 °C
Operating
temperature
Storage
temperature
Battery type 2 x 3 V button cell (2 x CR 2032 lithium)
Life 500 days (15 min measuring cycle, +25 °C)
Protection class IP65
Meas. cycle 1 min - 24 h (can be selected)
Memory 16,000 readings
Software can be used with Windows XP, Vista, Win7
Standards
Warranty
-30 to +70 °C
-40 to +70 °C
2004/108/EC, complies with the directives in
accordance with the standard EN 12830
2
24 months, warranty conditions: see website
www.testo.com/warranty
2
Please note that with this instrument in accordance with EN 12830 a regular
inspection and calibration as per EN 13486 must be performed
(recommendation: annually). Contact us for more information.
22
Page 6
3 Specifications
testo 174H
Characteristic Values
Probe type
NTC temperature sensor and internal
capacitive humidity sensor
Measuring range
0 to 100 % RH (non-dewing atmosphere),
-20 to +70 °C
Humidity accuracy ± 3 % RH (2 % RH to 98 % RH)
± 1 digit +0.03 % RH/K
Temperature
± 0.5 °C (-20 to +70 °C)
accuracy
Resolution 0.1 % RH, 0.1 °C
Operating
-20 to +70 °C
temperature
Storage
-40 to +70 °C
temperature
Battery type 2 x 3 V button cell (2 x CR 2032 lithium)
Life 1 year (15 min measuring cycle, +25 °C)
Protection class IP20
Meas. cycle 1 min - 24 h (can be selected)
Memory 2 x 8000 readings
Software can be used with Windows XP, Vista, Win7
Warranty 24 months, warranty conditions: see website
www.testo.com/warranty
EC Directive 2004/108/EC
23
Page 7
4 First steps
4 First steps
4.1 Releasing/securing data logger
The data logger is delivered secured.
Releasing data logger
1. Push both locks on the bottom of the wall bracket outwards.
2. Slide data logger out of the wall bracket.
Securing data logger
1. Slide data logger into the wall bracket.
2. Push both locks on the bottom of the wall bracket inwards.
24
Page 8
4.2 Inserting batteries
1. Place the data logger on its front.
4 First steps
2 Open battery compartment cover on the rear of the data logger
by turning to the left and remove it.
Use a coin for this.
3. Insert supplied batteries (3 V button cell, CR 2032 lithium) so
that the + pole is always visible.
4. Place battery compartment cover on the data logger and close
by turning to the right.
Use a coin for this.
- The current temperature is shown on the display screen.
25
Page 9
4 First steps
4.3 Connecting data logger to PC
1. Install testo ComSoft software.
The software is available in the Internet as a free download
The instructions for the installation and operation of the
software can be found in the testo ComSoft instruction
manual that is downloaded together with the software.
The software can be ordered on CD (article no. 0572 0580)
if the download from the Internet is not desired.
2. Connect the connecting cable of the interface to a free USB
interface on the PC.
3. Slide the data logger into the bracket of the interface.
4. Configure data logger, see separate testo ComSoft instruction
manual.
26
Page 10
5 Display and control elements
5 Display and control elements
5.1 Display
Depending on the operating status, various information may
be shown in the display. A detailed depiction of the
information that can be called up can be found in the quick
instruction manual that is included with every data logger.
1 Measurement program running
2 Measurement program over
3 Wait for start of measurement program
4 Highest saved reading
5 Lowest saved reading
6 Reading
7 Units
8 Start criterion Date/time programmed
9 Battery capacity
27
Page 11
5 Display and control elements
Icon Capacity
10 Lower alarm value:
• Flashes: programmed alarm value is shown
• Lights up: programmed alarm value was undershot
11 Upper alarm value
• Flashes: programmed alarm value is shown
• Lights up: programmed alarm value was exceeded
The display speed of liquid crystal displays slows down at
temperatures below 0 °C (approx. 2 seconds at -10 °C,
approx. 6 seconds at -20 °C) for technical reasons. This
does not influence the measuring accuracy.
75-100%
50-75%
25-50%
10-25%
<10%
Battery empty (measurement program was
stopped).
> Read out data and change battery (see
Changing the batteries page 30 ).
5.2 Key functions
✓ Operating status Wait and start criterion Button start
programmed.
> Press and hold GO button approx. 3 seconds to start the
measurement program.
- The measurement program starts and Rec appears in the
display.
✓ Operating status Wait:
28
Page 12
> Press GO button in order to change between displays of the
upper alarm value, lower alarm value, battery life and last
reading.
- These are shown in the specified sequence in the display.
✓ Operating status Rec or End:
> Press GO button in order to change between displays of the
highest saved reading, lowest saved reading, upper alarm
value, lower alarm value, battery life and last reading.
6 Using the product
- These are shown in the specified sequence in the display.
6.1 Programming data logger
In order to adapt the programming of your data logger to your
individual requirements, you require the testo ComSoft software. It
is available in the Internet as a free download requiring registration
The instructions for the installation and operation of the
software can be found in the testo ComSoft instruction
manual that is downloaded together with the software.
6 Using the product
6.2 Mounting wall bracket
Mounting materials (e.g. screws, anchor plugs) are not
included in the delivery.
✓ Data logger is removed from the wall bracket (see Releasing
data logger page 24).
1. Position wall bracket at the desired location.
2. Using a pen or similar, mark the location for the fastening
screw.
3. Prepare the fastening location for the attachment in accordance
with the material (e.g. drill hole, insert anchor plug).
4. Fasten wall bracket using an appropriate screw.
29
Page 13
7 Maintaining the product
6.3 Reading out data
The data logger is read out and the read-out data are processed
further by means of the testo ComSoft software, see separate
7 Maintaining the product
7.1 Changing the batteries
instruction manual.
The running measurement program is stopped when the
battery is changed. The stored data are preserved,
however.
1. Read out stored data, see testo ComSoft software instruction
manual.
✓ If it is no longer possible to read out the stored data because
the battery capacity is too low:
> Change batteries and then read out the stored data.
2. Place the data logger on its front.
3. Open battery cover on the rear of the data logger by turning to
the left. Use a coin for this.
30
4. Remove empty batteries from the battery compartment.
5. Insert two new batteries (3 V button cells, CR 2032 lithium) into
the instrument so that the + pole is always visible.
Page 14
7 Maintaining the product
Only use new branded batteries. If a partially used battery
is inserted, the calculation of the battery capacity is not
performed correctly.
6. Place battery compartment cover on the battery compartment
and close by turning to the right. Use a coin for this.
- The current reading is shown on the display.
The data logger must be reconfigured. To do so, the testo
ComSoft software must be installed on the computer and a
connection to the data logger must be set up (see
Connecting data logger to PC page 26).
7. Slide the data logger into the bracket of the interface.
8. Start testo ComSoft software and create a connection to the
data logger.
9. Reconfigure data logger or install the old stored configuration,
see separate testo ComSoft instruction manual.
7.2 Cleaning the instrument
- The data logger is once again ready for use.
CAUTION
Damage to the sensor!
> Ensure that no liquid enters the inside of the housing.
> If the housing of the instrument is dirty, clean it with a damp
cloth.
Do not use any aggressive cleaning agents or solvents! Weak
household cleaning agents or soap suds can be used.
31
Page 15
8 Tips and assistance
8 Tips and assistance
8.1 Questions and answers
If you have any questions, please contact your dealer or Testo
Customer Service. The contact details can be found on the back of
this document or on the Internet at www.testo.com/service-
8.2Accessories and spare parts
contact.
Description Article no.
testo 174H mini data logger, 2-channel, incl.
wall bracket, battery (2 x CR 2032 lithium)
and calibration protocol
testo 174T mini data logger, 1-channel, incl.
wall bracket, battery (2 x CR 2032 lithium)
and calibration protocol
USB interface for programming and reading
out testo 174T and testo 174H loggers
testo ComSoft CD (if free download requiring
registration from website is not desired)
Battery, 3 V button cell (CR 2032 lithium),
please order 2 batteries per logger
ISO calibration certificate humidity, calibration
points 11.3 % RH; 50.0 % RH; 75.3 % RH at
+25 °C/+77 °F; per channel/instrument
ISO calibration certificate temperature,
calibration points -18 °C; 0 °C; +60 °C; per
channel/instrument
0572 6560
0572 1560
0572 0500
0572 0580
0515 0028
0520 0176
0520 0151
32
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.