Testo Testo 160 IAQ Manual

Page 1
testo 160
Instruction manual
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Page 2
Contents
Contents
1 Safety and waste disposal ................................................................... 5
1.1 About this document .............................................................................. 5
1.2 Symbols and writing standards .............................................................. 5
1.3 Safety ..................................................................................................... 5
1.4 Warning notices ..................................................................................... 6
1.5 Disposal ................................................................................................. 6
2 Description of the instrument ............................................................... 7
2.1 Using the testo 160 ................................................................................ 7
2.2 WiFi data loggers ................................................................................... 7
2.2.1 testo 160 TH ............................................................................ 7
2.2.2 testo 160 E .............................................................................. 8
2.2.3 testo 160 THE .......................................................................... 8
2.2.4 testo 160 THL .......................................................................... 9
2.2.5 testo 160 IAQ ......................................................................... 10
2.3 External probes .................................................................................... 10
2.3.1 S-TH ...................................................................................... 11
2.3.2 S-LuxUV ................................................................................ 12
2.3.3 S-Lux ..................................................................................... 12
2.3.4 Extension cable ..................................................................... 13
2.4 Deco-covers ......................................................................................... 13
3 Technical data ................................................................................... 14
3.1 WiFi data loggers ................................................................................. 14
3.2 External probes .................................................................................... 18
3.3 Deco-covers ......................................................................................... 21
4 Operation ........................................................................................... 22
4.1 Commissioning .................................................................................... 22
4.2 Logging into the Testo Cloud ............................................................... 24
4.2.1 Configuration via the Setup assistant .................................... 24
4.2.2 Configuration via the web interface (WPA2 Personal) ........... 25
4.2.3 Configuration via the testo Saveris 2 App ............................. 26
4.2.4 Offline configuration via PDF form ......................................... 28
4.3 Logging WiFi data loggers out of the Testo Cloud ............................... 29
4.4 Status LED signals ............................................................................... 30
Page 3
Contents
4.5 Inserting into/removing from the wall bracket ..................................... 31
4.5.1 Installing the probe on the data logger ................................. 31
4.5.2 Changing batteries ................................................................ 32
4.5.3 Deco-cover installation ......................................................... 32
4.5.4 Wall bracket .......................................................................... 33
4.6 testo Saveris 2 App ............................................................................. 34
4.6.1 Receiving push alerts ............................................................ 34
4.6.2 Network analysis ................................................................... 34
4.7 Analysis and reports (web) .................................................................. 34
4.8 Alarms ................................................................................................. 36
4.8.1 Alarm list ............................................................................... 36
4.8.2 Alarm settings ....................................................................... 37
4.8.2.1 Creating and displaying alarm settings ................ 37
4.8.2.2 Configuring and editing a displayed alarm setting
............................................................................. 37
4.9 System warnings ................................................................................. 39
4.9.1 Creating and displaying system warnings ............................ 39
4.9.2 Configuring and editing a displayed system warning ........... 39
4.10 Configuration ....................................................................................... 40
4.10.1 Standard users ...................................................................... 40
4.10.2 Creating and editing new users ............................................ 40
4.10.3 User roles .............................................................................. 41
4.10.4 User management ................................................................. 43
4.10.4.1 User settings ........................................................ 43
4.10.4.2 Account information ............................................. 43
4.10.4.3 Change password ................................................ 43
4.10.4.4 Logoff ................................................................... 43
4.10.5 Account ID ............................................................................ 43
4.10.6 Creating and editing a measuring point group...................... 43
4.10.7 Areas ..................................................................................... 44
4.10.7.1 Creating and editing areas ................................... 44
4.10.7.2 Deleting an area ................................................... 45
4.10.8 WiFi data loggers .................................................................. 45
4.10.9 Firmware updates ................................................................. 46
4.11 Command bar ...................................................................................... 46
Page 4
Contents
4.11.1 Opening the Setup assistant ................................................. 46
4.11.2 Opening the Online Help ....................................................... 47
4.11.3 Opening system messages ................................................... 47
4.12 System and status information ............................................................ 47
5 FAQ ................................................................................................... 48
6 Cloud licences ................................................................................... 52
7 Authorizations .................................................................................... 52
Page 5

1 Safety and waste disposal

Result of an action
Requirement
1 Safety and waste disposal

1.1 About this document

Use
• The instruction manual is an integral part of the instrument.
Pay particular attention to the safety instructions and warning notices in
order to prevent injuries or damage to the product.
Keep this documentation to hand so that you can refer to it when
necessary.
• Always use the complete original instruction manual.
Hand this documentation on to any subsequent users of the product.

1.2 Symbols and writing standards

Display Explanation
Note: basic or further information
1.
2.

1.3 Safety

General safety instructions
Only operate the product properly, for its intended purpose, and within the
parameters specified in the technical data. Do not apply any force.
• Do not operate the instrument if there are signs of damage on the housing.
Dangers may also arise from the systems to be measured or from the
measuring environment: Always comply with the locally valid safety regulations when carrying out measurements.
Action: several steps, the sequence must be followed.
5
Page 6
1 Safety and waste disposal
CAUTION
Indicates possible damage to equipment
Temperature information given on probes/sensors relates only to the
measuring range of the sensor technology. Do not expose handles and feed lines to temperatures in excess of 70 °C (158°F), unless they are expressly authorised for use at higher temperatures.
Do not carry out any contact measurements on uninsulated, live parts.
Do not store the product together with solvents. Do not use any desiccants.
Only perform that maintenance and repair work on this instrument which is
described in the documentation. Follow the prescribed steps exactly when doing the work. Use only original spare parts from Testo.
Batteries
Improper use of batteries may cause destruction of the batteries, injuries
due to current surges, fire or the escape of chemicals.
Only use the batteries supplied in accordance with the instructions in the
instruction manual.
Do not short-circuit the batteries.
• Do not take the batteries apart and do not modify them.
Do not expose the batteries to heavy impacts, water, fire or temperatures in
excess of 55 °C.
Do not store the batteries near any metal objects.
In the event of contact with battery acid: rinse affected areas thoroughly
with water, and if necessary consult a doctor.
Do not use any leaky or damaged batteries.

1.4 Warning notices

Always pay attention to any information marked with the following warning notices along with warning pictograms. Implement the specified precautionary measures!

1.5 Disposal

Dispose of spent batteries in accordance with the relevant legal
specifications.
6
Page 7

2 Description of the instrument

160 TH data logger to carry out temperature and
At the end of its useful life, deliver the product to the separate collection
point for electric and electronic devices (observe local regulations) or return the product to Testo for disposal.
2 Description of the instrument

2.1 Using the testo 160

The testo 160 WiFi data logger system is a modern solution for monitoring the climate and lighting conditions, e.g. in museums, archives, galleries and libraries.
The system consists of WiFi data loggers, external probes and a Cloud data storage system. The testo 160 WiFi data loggers reliably record temperature and humidity, CO2 levels, illuminance and UV radiation at adjustable intervals and transmit the readings directly to the Testo Cloud via WLAN. Via the web interface of this Cloud, the data can be analyzed at any time and anywhere, using an internet-ready smartphone, tablet or PC. The WiFi data loggers are programmed and reports are generated via this interface. Limit value violations are immediately reported via e-mail or optionally via SMS, and alarm notifications are possible via Push (Saveris 2 App).

2.2 WiFi data loggers

2.2.1 testo 160 TH

You can use the testo humidity measurements.
Element Element
7
Page 8
2 Description of the instrument
1
and relative humidity
2
1
sensor
2
3
Internal sensors for temperature
Micro-USB port

2.2.2 testo 160 E

The external probes S-TH, S-LuxUV and S-Lux can be connected to the testo 160 E WiFi data logger.
Element Element
Connector socket for external
Connector socket for external sensor
USB connection

2.2.3 testo 160 THE

You can use the testo 160 THE WiFi data logger to carry out temperature and humidity measurements. The external probes S-TH, S-LuxUV or S-Lux can also be connected.
Element Element
8
Page 9
2 Description of the instrument
1
and relative humidity
2
sensor
3
4
sensor
1
UV sensor
2
Lux sensor
3
4
Internal sensor for temperature
USB connection
Connector socket for external
Connector socket for external

2.2.4 testo 160 THL

You can use the testo 160 THL WiFi data logger to measure temperature, humidity, illuminance and UV radiation.
Element Element
Internal sensor for temperature and relative humidity
USB connection
9
Page 10
2 Description of the instrument
1
2
3
4
5
QR code
6
Button
7
8
If the WiFi data logger is in Single Mode (without external power supply

2.2.5 testo 160 IAQ

You can use the testo 160 IAQ WiFi data logger to carry out temperature, humidity, carbon dioxide concentration and atmospheric
Element Element
pressure measurements.
Status LED Air quality light
Display CO2 sensor
USB connection
Internal sensor for temperature and relative humidity
If the WiFi data logger is in Continuous Mode (external power supply via USB mains unit), the air quality light stays on permanently. Temperature and humidity readings are displayed alternately.
via USB mains unit), the air quality light only comes on briefly during the measurement. Only the temperature is displayed. When switching to Single Mode, the WiFi data logger does not supply any readings for at least 10 min. “CAL” is shown on the display until the next measurement.

2.3 External probes

The external probes S-TH, S-LuxUV and S-Lux extend the range of functions of the 160 THE WiFi data logger and, in conjunction with the 160 E WiFi data logger, form an extremely versatile measurement system.
10
Page 11
The external probes are only approved in conjunction with the
To make it easy to install, the probe can be pushed out of the wall
1
Sensor
2
Screw thread
3
4
testo 160 THE and testo 160 E WiFi data loggers.

2.3.1 S-TH

The external probe S-TH can be connected to the following WiFi data loggers: testo 160 THE and testo 160 E. You can use the S-TH probe to carry out temperature and humidity measurements.
bushing. The probe can also be used without this wall bushing.
2 Description of the instrument
Element Element
Securing nut
Jack plug
11
Page 12
2 Description of the instrument
1
Lux sensor
2
UV sensor
3
Jack plug

2.3.2 S-LuxUV

The external probe S-LuxUV can be connected to the following WiFi data loggers: testo 160 THE and testo 160 E. You can use the S-
Element Element
LuxUV probe to carry out illuminance and UV measurements.

2.3.3 S-Lux

The external probe S-Lux can be connected to the following WiFi data loggers: testo 160 THE and testo 160 E. You can use the S-Lux probe
Element Element
12
to carry out illuminance measurements.
Page 13
2 Description of the instrument
1
2
The sensors are supplied with the 60 cm cable (0554 2004) as
combined. The maximum total length is approx. 10 m.
Lux sensor
Jack plug

2.3.4 Extension cable

standard. A 2.5 m long cable is optionally available (0554 2005), to be able to adapt the measurement system to all measuring situations. Since these are digital probes, multiple extension cables can also be

2.4 Deco-covers

3 different deco-covers are optionally available. Cover 0554 2006 is intended for the testo 160 TH, testo 160 THE and testo 160 E WiFi data loggers. Cover 0554 2009 is intended for the testo 160 THL WiFi data logger and cover 0554 2012 for the testo 160 IAQ WiFi data logger.
13
Page 14

3 Technical data

CAUTION
Damage to the humidity probe
Temperature measurement
Measuring range
-10 °C to 50 °C
see ext. probes
Accuracy
± 0.5 °C
Resolution
0.1 °C
Humidity measurement
± 1% RH/year drift
Lux measurement
Measuring range
see ext. probes
see ext. probes
Accuracy
Resolution
UV measurement
Measuring range
see ext. probes
see ext. probes
3 Technical data

3.1 WiFi data loggers

Measurement-specific data
The humidity sensor attains the highest degree of accuracy in temperatures between + 5 °C and + 60 °C and 20% to 80% RH. If the instrument is exposed to higher humidity for a long period of time, this can falsify the readings by up to 3% RH. After 48 hours at 50% RH ±
WiFi data loggers testo 160 TH testo 160 THE testo 160 E Order number 0572 2021 0572 2023 0572 2022
10 % and +20 °C ± 5 °C, the sensor regenerates by itself.
- The probe must never be exposed to a humidity level of 100 % RH for
longer than 3 days.
Measuring range 0 to 100% RH (non-condensing) see ext. probes Accuracy ± 2% RH @ 25 °C & 20 to 80% RH
Resolution 0.1% RH
Accuracy Resolution
14
± 3% RH @ 25 °C & <20% RH &
>80% RH
± 1% RH hysteresis
Page 15
3 Technical data
Order number
0572 2014
0572 2024
Temperature measurement
Measuring range
0 °C to 50 °C
-10 °C to 50 °C
Accuracy
± 0.5 °C
± 0.5 °C
Resolution
0.1 °C
0.1 °C
Humidity measurement
condensing)
condensing)
Resolution
0.1% RH
0.1% RH
Lux measurement
Measuring range
0 to 20,000 lux
UV measurement
Measuring range
0 to 10,000 mW/m2
22 °C)
Resolution
0.1 mW/m2
CO2 measurement
reading) (@ 25 °C)
WiFi data loggers testo 160 IAQ testo 160 THL
Measuring range 0 to 100% RH (non-
Accuracy ± 2% RH @ 25 °C & 20 to
80% RH ± 3% RH @ 25 °C & <20% RH & >80% RH ± 1% RH hysteresis ± 1% RH / year drift
Accuracy DIN 5032-7 Class C-
Resolution 0.1 lux
Accuracy ± 5 mW/m2 or 5% of the
0 to 100% RH (non-
± 2% RH @ 25 °C & 20 to 80% RH ± 3% RH @ 25 °C & <20% RH & >80% RH ± 1% RH hysteresis ± 1% RH/year drift
compliant or: ± 3 lux or ± 3% of the
reading (based on the external reference DIN 5032-7 Class L)
reading (based on the external reference at
Measuring range 0 to 5,000 ppm Accuracy ± (50 ppm + 3% of the
Resolution 1 ppm
reading) (@ 25 °C) Battery-operated: ± (100 ppm + 3% of the
15
Page 16
3 Technical data
Pressure
Measuring range
600 to 1100 mbar
Accuracy
± 3 mbar @ 22 °C
Resolution
1 mbar
Order number
0572 2021
0572 2023
0572 2024
temperature
temperature
Protection class
IP20
WiFi data loggers testo 160 IAQ testo 160 THL
The time between the system warning "Battery almost discharged” and "Measurement data stop" is at the most one day during standard operation and a measuring cycle & communication cycle of 1 min (day
testo 160 WiFi data loggers come with a factory calibration protocol as standard. For museums, we recommend having an annual test carried out by Testo Customer Service. Moreover, there is an option of having ISO certificates created for the WiFi data loggers. These can be implemented by Testo Industrial Services (TIS) .
General data
WiFi data loggers testo 160 TH testo 160 THE testo 160 THL
& night) (battery type: Varta Industrial).
Operating
Storage
-10 °C to 50 °C
-20 °C to 50 °C
Measuring cycle Depends on the Cloud licence
Basic: 15 min to 24 h / Advanced 1 min to 24 h flexible
Communication cycle
Basic: 15 min to 24 h / Advanced 1 min to 24 h flexible
Depends on the Cloud licence
Memory 32,000 readings (sum of all channels) Voltage supply 4 x AAA batteries 1.5 V
Alternatively mains unit via USB connection
Battery life 18 months
At +25 °C, 15-minute measuring cycle and 6-hour
communication cycle
(depending on the WLAN structure) Dimensions 64 x 76 x 22 mm 64 x 76 x 22 mm 64 x 92 x 24 mm Weight including
batteries
94 g 94 g 113 g
16
Page 17
3 Technical data
Order number
0572 2014
0572 2022
temperature
Protection class
IP20
flexible (battery operation)
Advanced 1 min to 24 h flexible
USB connection
USB connection
reception quality)
reception quality)
batteries
Order number
0572 2021
0572 2023
0572 2024
WLAN
Standard
802.11 b/g/n
WiFi data
testo 160 IAQ testo 160 E
loggers
Operating
Storage
0 °C to 50 °C -10 °C to 50 °C
0 °C to 50 °C -20 °C to 50 °C
temperature
Measuring cycle Depends on the Cloud
licence Basic: 15 min to 24 h / Advanced 1 min to 24 h flexible (mains operation)
Depends on the Cloud licence Basic: 15 min to 24 h / Advanced 1 min to 24 h flexible
Advanced 5 min to 24 h
Communication cycle
Depends on the Cloud licence
Basic: 15 min to 24 h /
Memory 32,000 readings (sum of all channels) Voltage supply 4 x AA batteries
Alternatively mains unit via
Battery life 12 months at +25 °C,
15-minute measuring cycle and 8-hour communication cycle
(depending on the WLAN
4 x AAA batteries 1.5V Alternatively mains unit via
18 months at +25 °C, 15-minute measuring cycle and 6-hour communication cycle
(depending on the WLAN
Dimensions 82 x 117 x 32 mm 64 x 76 x 22 mm Weight including
WiFi-specific data
WiFi data loggers testo 160 TH testo 160 THE testo 160 THL
269 g 96 g
17
Page 18
3 Technical data
Order number
0572 2014
0572 2022
WLAN
Standard
802.11 b/g/n
WPA Personal, WPA2 (AES), WPA (TKIP), WEP
Ports
No bi-directional port approvals are necessary.
During the initial configuration, it is possible to select whether DHCP or
testo 160 application
WiFi data loggers testo 160 TH testo 160 THE testo 160 THL Security WPA2 Enterprise: EAP-TLS, EAP-TTLS-TLS, EAP-TTLS-
MSCHAPv2, EAP-TTLS-PSK, EAP-PEAP0-TLS, EAP­PEAP0-MSCHAPv2, EAP-PEAP0-PSK, EAP-PEAP1-TLS, EAP-PEAP1-MSCHAPv2, EAP-PEAP1-PSK;
WPA Personal, WPA2 (AES), WPA (TKIP), WEP
WiFi data
testo 160 IAQ testo 160 E
loggers
Security
WPA2 Enterprise: EAP-TLS, EAP-TTLS-TLS, EAP-TTLS­MSCHAPv2, EAP-TTLS-PSK, EAP-PEAP0-TLS, EAP­PEAP0-MSCHAPv2, EAP-PEAP0-PSK, EAP-PEAP1-TLS, EAP-PEAP1-MSCHAPv2, EAP-PEAP1-PSK;
Technical data for a secure wireless LAN
The testo 160 WiFi data loggers use the MQTT protocol, which communicates via port TCP 1883 and 8883. These UDP port approvals are also required:
Port 53 (DNS name resolution)
Port 123 (NTP time synchronisation)
All ports only have to be able to communicate externally to the Cloud.
Static IP is used (select Expert mode for the corresponding information). (Not possible in the Setup assistant.)
The testo 160 application is accessible via a normal, up-to-date browser (www). The standard TCP ports http (80) and https (443) are used.

3.2 External probes

Measurement-specific data
18
Page 19
3 Technical data
Order number
0572 2156
0572 2157
0572 2158
Temperature measurement
Measuring range
-10 °C to 50 °C
Accuracy
± 0.5 °C
Resolution
0.1 °C
Humidity measurement
(non-condensing)
Resolution
0.1 % RH
Lux measurement
Measuring range
0 to 20,000 Lux
Class L)
Resolution 0.1 lux
UV measurement
mW/m²
22 °C)
Probes S-TH S-LuxUV S-Lux
Measuring range 0 to 100 % RH
Accuracy ± 2% RH @ 25 °C
& 20 to 80% RH ± 3% RH @ 25 °C
& <20 % RH & >80% RH ± 1% RH hysteresis ± 1 % RH / year drift
Accuracy DIN 5032-7 Class C-compliant.
or: ± 3 lux or ± 3% of the reading (based
on the external reference DIN 5032-7
Measuring range 0 to 10,000
Accuracy ± 5 mW / m² or ±
5 % of the reading (based on the external reference at
Resolution 0.1 mW/m²
General data
19
Page 20
3 Technical data
Order number
0572 2156
0572 2157
0572 2158
temperature
Dimensions
38 x 16 mm
28 x 56 x 15 mm
28 x 56 x 15 mm
Weight
13 g
15 g
13 g
Probes
S-TH S-Lux UV S-Lux
Operating
Storage temperature
-10 °C to 50 °C
-20 °C to 50 °C
20
Page 21

3.3 Deco-covers

Cover
Order number
0554 2006
0554 2009
0554 2012
THE / E
Dimensions
82 x 69 x 23 mm
97 x 69 x 23 mm
121 x 88 x 32 mm
Weight
22 g
18 g
41 g
General data
3 Technical data
Use testo 160 TH /
testo 160 THL testo 160 IAQ
21
Page 22

4 Operation

The external probes must be connected to the WiFi data logger before
WiFi data logger logged in again.
CAUTION
Damage to WiFi data loggers!
CAUTION
Potential damage to the optical surfaces (THL, S-Lux and S-LuxUV)
CAUTION
Potential damage to the optical components (IAQ)
2
2
4 Operation

4.1 Commissioning

logging into the Cloud for the first time. If an additional probe is to be connected at a later stage, the WiFi data logger must first be logged out of the Cloud. The external probe can then be connected and the
- Do not place near any solvents.
- Do not clean using solvents.
- Do not use sharp objects.
- Only use soft cleaning cloths.
- Do not use aggressive cleaning agents.
- Avoid any vibrations, the factory calibration may be altered. Check the
readings in fresh air 350 to 450 ppm CO
(urban air up to 700 ppm CO2).
- Prevent condensation. This can result in elevated CO
- Do not use aggressive cleaning agents.
The data loggers must only be mounted vertically. Here, the connections must point downwards. In the case of data loggers with a display, you need to pay attention to the reading direction. Otherwise, the measuring accuracy might be diminished.
22
readings.
Page 23
4 Operation
1
- Mount the wall bracket at the
2
- Open battery compartment cover.
3
- Remove battery safety strips.
4
- Close the battery compartment.
5
- Insert the data logger into the wall
The IAQ data logger has a higher energy requirement. This reduces the
designated location using suitable mounting materials (screws, cable ties or the supplied 3M adhesive strips).
bracket.
minimum measuring cycle to 5 minutes when battery-operated. Operation via mains unit is therefore recommended. An appropriate USB cable can also be purchased as an accessory.
23
Page 24
4 Operation
before
1
Only for testo 160 E and testo 160 THE: The external probes must be connected
for the first time. If an additional probe is to be connected at a later stage, the data logger must first be logged out of the Cloud. The external probe can then be connected and the data logger logged in again.
The testo 160 WiFi data loggers can also be powered via the USB port instead of being run on batteries. However, the WiFi data loggers do not have a charging function, i.e. no rechargeable batteries in the WiFi data logger can be charged up via the USB port. If you connect the WiFi data logger to the USB port on your PC, the WiFi data logger automatically switches to mass storage and configuration mode. A PC is therefore not suitable as a voltage source for logger operation.
logging into the Cloud

4.2 Logging into the Testo Cloud

You need an account for the Testo Cloud.
For your new testo 160 WiFi data logger to be able to connect to your account in the Testo Cloud, it requires the three following pieces of information at minimum:
1. The ID of your account in the Cloud. You will find this in your account under the menu item Configuration - Account ID.
2. The network name of your WLAN (SSID), which the WiFi data logger will use to connect to the internet.
3. The password for this network.
Storage of this information on the WiFi data logger is called “Configuring the WiFi data logger". Four different options are available for this process.

4.2.1 Configuration via the Setup assistant

The Setup assistant in the web interface of the Testo Cloud is provided to assist you when you take your first steps with commissioning the testo 160. It can help you with logging in WiFi data loggers.
To be able to carry out the configuration, you need to be logged into the web interface
- Click on the symbol above the menu bar.
24
Page 25
4 Operation
- Data logger has not been configured
1
- Briefly press the button on the side of the data logger. (On the
- Data logger has already been configured (logger is in sleep mode)
1
- Press and hold down the button on
When configuring for WPA2 Enterprise, pay attention to the correct spelling and suffixes of certificate names. Depending on the encryption
In addition, they must be available individually and not in a bundle.
- The WiFi data logger is already in configuration mode and flashes at
The Setup assistant launches and assists you with the configuration. Follow the instructions there.
4.2.2 Configuration via the web interface (WPA2
Personal)
yet, the LED on the side of the data logger flashes once after the batteries are inserted.
testo 160 IAQ, the button is on the front.)
Data logger switches into configuration mode (LED flashes at one-
or
The WiFi data loggers can also be set up for the WPA2 Enterprise security standard via the web-based configuration. In this mode, the WiFi data logger functions as a web server on which you can log in via WLAN with the IP address 192.168.1.1 via smartphone, tablet or PC.
second intervals).
the side of the data logger for more than 3 s.
Data logger switches into configuration mode (LED flashes at one­second intervals)
method, the following 3 certificates must be available: ca.pem, client.pem, private.key. The certificates must be available either in the PEM or BASE64 format.
one-second intervals.
25
Page 26
4 Operation
1
- Select the network name of the WiFi data logger you wish to
2
- Open web browser on the PC, tablet, smartphone, etc.
3
- Enter IP address 192.168.1.1 in the web browser.
4
- Enter the testo Account ID (shown in the web interface of the Testo
5
- Enter network name (SSID).
6
- Enter configuration slot.
The testo 160 WiFi data loggers can be configured for up to three
7
- The security standard can be selected under “Security”. (Depending
8
- Enter password for the network.
9
- Confirm the configuration via “Configure”.
configure under network settings on the PC/tablet (e.g. testo 160 Sn:
12345678).
PC/tablet is connected to the WLAN hotspot of the WiFi data logger.
Website of the WLAN configuration opens.
Cloud under account information).
WLAN networks. Network name (SSID), password and security settings can be stored for each profile.
on the selection, further input options appear.)
WiFi data logger is fully configured and connected to the Cloud. The LED flashes green twice. The WiFi data logger then switches to measuring mode.

4.2.3 Configuration via the testo Saveris 2 App

Once you have downloaded the testo Saveris 2 App, you need to first log in with your access data.
Any number of WiFi data loggers can be commissioned at the same time.
26
Page 27
Saveris 2 App has already been configured via the Setup assistant, see
4.6.
- Data logger has not been configured
1
- Briefly press the button on the side of the data logger. (On the
- Data logger has already been configured (logger is in sleep mode)
1
- Press and hold down the button on
- The WiFi data logger is in configuration mode and flashes at one-
1
- Launch the testo Saveris 2 App once you have activated the WLAN
2
- Select Logger commissioning in the testo app.
3
- Enter the access data for your WLAN, so that this can be transmitted
Preparation
yet, the LED on the side of the data logger flashes once after the batteries are inserted.
testo 160 IAQ, the button is on the front.)
Data logger switches into configuration mode (LED flashes at one­second intervals).
or
the side of the data logger for more than 3 s.
Data logger switches into configuration mode (LED flashes at one­second intervals)
Configuration
4 Operation
second intervals.
- Press the button again to switch the WiFi data logger to App-config
mode. (The LED flashes every 200 ms for 10 seconds)
on your smartphone or tablet.
to all loggers.
27
Page 28
4 Operation
PDF form correctly. If you have not installed Adobe Reader, you can go
- Make sure that the batteries are inserted.
1
- Connect the data logger to the PC via USB connection.
2
- Open the file WiFiConf.pdf on the external drive testo 160.
3
- Copy your Account ID and paste it into the relevant field on the PDF
Configuration -> Account ID
4
- Enter configuration slot.
The testo 160 WiFi data loggers can be configured for up to three
settings can be stored for each profile.
5
- Enter the Network name (SSID) and, if necessary, your WLAN
password
6
- Click on the Save configuration button.
7
- Select the external drive testo 160 as the storage location and save
WiFiConf_Daten.xml
WiFi data loggers to be configured are displayed in the testo app and set up automatically.

4.2.4 Offline configuration via PDF form

As an alternative to creating the configuration file in the Quick Start Guide with subsequent download of the XML configuration file, the WiFi data logger can also be configured via a PDF form.
You need the Adobe Reader program (version 10 or later) to use the
to the following address to download it free of charge: http://get.adobe.com/reader/.
form. You will find the Account ID in the web interface of the Testo Cloud under
.
WLAN networks. Network name (SSID), password and security
in the relevant fields on the PDF form.
A dialogue opens for exporting the form data.
the form data (configuration file
The green and red LEDs light up simultaneously until the PDF document is completely generated.
28
) on it.
Page 29
4 Operation
8
- Disconnect the USB connection to the PC to complete the
- The WiFi data logger is logged into the Testo Cloud.
1
- Select Configuration ->WiFi data logger in the web interface.
2
- Select the WiFi data logger you require.
3
- Press Details.
4
- At the bottom of the menu, select the button Remove data logger.
communicates with the Cloud. Depending on the communication cycle
configuration of the data logger.
You can also save the configuration file locally on your computer. Other WiFi data loggers can be configured even faster by simply copying the XML configuration file onto the external drive testo 160.
4.3 Logging WiFi data loggers out of the
Testo Cloud
It may be necessary to log the WiFi data logger out of the Cloud again. A logger cannot be operated in two different accounts simultaneously, therefore it must be logged out before switching accounts. Similarly, any technical changes to the WiFi data logger, e.g. due to adding or removing external sensors, can only be registered by logging back into the Cloud.
All WiFi data loggers logged in are displayed.
The WiFi data logger is removed.
The log-out still needs to also be transmitted to the WiFi data logger. This happens automatically the next time the WiFi data logger
selected, this may take some time. You can instruct the WiFi data logger to establish a connection with the Cloud right away by briefly pressing the button. This process is indicated by the green LED flashing briefly. The WiFi data logger is logged out. After logging out of the Cloud, press the buttons once briefly so that the WiFi data logger receives the log-out.
29
Page 30
4 Operation
LED does not flash (TH, E, THE, THL)
Sleep mode
(for 5 min, then 1 long red flash)
10 seconds)
mode press button < 3 sec
any external probes connected.)
flashes
successful
violation
(measurement data transmitted)
successfully
LED gives 4 red flashes
Batteries spent

4.4 Status LED signals

The following table provides an overview of the meaning of the various status LED signals of the testo 160 WiFi data logger.
Signal Description
LED flashes green every 30 seconds
Normal state
(IAQ) LED flashes green at one-second
intervals
LED flashes green every 200 ms (for
Configuration mode (hotspot) - press button > 3 sec
Configuration app: During hotspot
LED gives 2 red flashes Connection to WLAN failed (incorrect
SSID, incorrect SSID password, incorrect account ID or incorrect account password, attempt to log the testo 160 E into the Cloud without
If XML is correct, LED gives 1 long
Configuration via USB/PDF
green flash If XML is incorrect, LED gives 3 red
LED gives 2 green flashes Connection to WLAN and Cloud
LED gives 1 long red flash Alarm activated due to limit value
LED gives 5 green flashes Reset WiFi data logger to factory
settings Press key > 20 sec
LED gives 1 green flash (measurement data collected)
LED gives 2 short green flashes
Send measurement data to the Testo Cloud (website): press key < 3 sec
Measurement data transmitted
LED flashes alternately green and red Firmware update via USB or wireless
30
Page 31
4 Operation
1
- Insert the unlocking tool into the
2
- Push back the locking pin using the unlocking tool.
3
- Pull the data logger up and out of
The external probes must be connected to the WiFi data logger before
logger logged in again.
1
- Connect the probe plug to the
4.5 Inserting into/removing from the wall
bracket
unlocking opening.
the wall bracket.

4.5.1 Installing the probe on the data logger

logging into the Cloud for the first time. If an additional probe is to be connected at a later stage, the data logger must first be logged out of the Cloud. The external probe can then be connected and the data
designated jack on the data logger.
The external probe is ready for use.
31
Page 32
4 Operation
CAUTION
Incorrectly inserted batteries! The instrument may be damaged!
1
- Open battery compartment cover.
2
- Change batteries. Pay attention to
3
- Close the battery compartment.
1
- Break out the required, pre-punched

4.5.2 Changing batteries

A battery change stops a measurement that is currently running. However, stored data is preserved.
- Pay attention to the polarity when inserting the batteries.
Only use new branded batteries. If a partially exhausted battery is inserted, the battery capacity will not be calculated correctly.
polarity.

4.5.3 Deco-cover installation

knock-out points on the deco-cover.
32
Page 33
4 Operation
2
- Place the deco-cover onto the data
3
- Always make sure that the deco-
4
- Then connect external probes or the
CAUTION
Incorrect readings!
CAUTION
Damage to the sensor!
logger from the side and press it into place.
cover is positioned correctly so as not to obscure sensors.
external voltage supply once again.
- Make sure that the deco-cover is positioned correctly.
- Let painted or varnished deco-covers dry out and out-gas sufficiently
before fitting.

4.5.4 Wall bracket

The wall bracket supplied, which comes with an adhesive pad, is only intended for testo 160 loggers and ensures that the loggers stay securely in place. Any other use is not deemed appropriate and may result in the wall bracket being damaged.
33
Page 34
4 Operation
Apart from the adhesive pad, no other mounting materials are included
1
- Click on the checkbox in front of the required recipient.
settings are changed. After finishing the configuration, these may need
reports are regularly generated automatically by the system (Generated
in the delivery. Please select suitable mounting materials (cable ties or screws) that are appropriate for the required mounting location.

4.6 testo Saveris 2 App

The free testo Saveris 2 App offers you practical functions for commissioning, network analysis and for receiving testo 160 push messages. It must be installed on your smartphone or tablet.
Once you have downloaded the testo Saveris 2 App, you need to first

4.6.1 Receiving push alerts

Push alerts can be set in the web interface via the Alarm settings menu. (See Section 4.8.2)
log in with your access data.
Due to an iOS system limitation, logger commissioning and network analysis can only be carried out on Android devices.
The selected recipient will receive push messages in future.
When configuring the WiFi data loggers using the testo Saveris 2 App, the smartphone functions as a hot-spot and any default hot-spot
ast

4.6.2 Network analysis

to be adjusted again manually.
Use the network analysis to analyze your WLAN network in advance in order to check the requirements for operating the testo 160 loggers correctly.
You can also optionally create and send status reports.

4.7 Analysis and reports (web)

According to the settings specified by the user (Report settings),
34
reports).
Page 35
4 Operation
1
- Click on the “Automatic reports” button.
2
- Enter the data required to create an automatic report.
3
- Click on the “Create an automatic report” button.
- A summary of the reports already generated is displayed.
1
- Click on the arrow symbol to open the tab.
2
- Click on the "Download" button.
3
- “Edit this report series” button.
The following settings can be defined and edited:
• Name of the report: Designation of the automatic report.
Measuring points for the report: Measuring points that are to be covered in
the report. Click on the checkbox in front of the channel designation.
How often is the report to be created?: Interval at which the reports are to
be generated. Select a report cycle from the drop-down menu.
File format: File format in which the reports are to be generated. Select a file
format from the drop-down menu.
Data views: Data views in which the data in the report is to be displayed.
Click on the checkbox in front of the data view designation.
Also send report via e-mail: as well as saving reports under Generated
reports, these can also be sent as e-mails. Click on the checkbox to open the input screen for e-mail addresses.
Only users created with a supplied e-mail address are listed as possible e-mail recipients. It is not possible to directly input an e-mail address.
The first report will be created on the following day.
Generated reports
More information is displayed.
The report is downloaded.
Settings are displayed and can be edited.
35
Page 36
4 Operation
1
- Click on the "Actions" button.
2
- Click “Edit”
1
- Click on the "Actions" button.
2
- Click “Delete”
1
- Click on the checkbox in front of the measuring point
Report settings
Automatic reports which have already been created are displayed in a table.
Settings are displayed and can be edited.
Automatic report is deleted.

4.8 Alarms

With the testo Saveris 2 App, you can receive push alerts and you also get an additional overview.
The Testo Saveris 2 App must be installed on your Android or iOS device. You can get the App from the App Store for iOS devices or from the Play Store for Android devices.
Push alerts will not be forwarded if your provider service is unavailable or if the smartphone is switched off.

4.8.1 Alarm list

Display of alarms
A summary of all triggered alarms and system warnings is displayed. Unread alarms and system warnings are shown in bold.
The display can be filtered according to the following characteristics:
group/measuring point.
The alarms are sorted and displayed by measuring point
36
group/measuring point.
Page 37
1.1
- Click on the start date/end date.
1.2
- Select the start date/start time or end date/end time.
1
- Click on the arrow to open the tab and display more information.
1
- Click on the “Mark all as read” button.
1
- Click on the "+ New alarm setting” button.
1
- Click on the title of an alarm setting.
(required field)
The alarms are sorted and displayed by start or end date.
Detailed information on alarms
On displaying the detailed information, the alarm message/system warning is marked as "read" and the alarm counter is reduced.
All alarm messages are marked as “read”.

4.8.2 Alarm settings

4.8.2.1 Creating and displaying alarm settings
New alarm setting can be set.
Existing alarm settings are displayed below the button.
4 Operation
An existing setting is displayed.
4.8.2.2 Configuring and editing a displayed alarm setting
The following settings can be defined and edited.
Setting Description Title Designation of the alarm setting
37
Page 38
4 Operation
in the marked period.
sensor.
+ Add button.
Setting Description Measuring points Measuring point group/measuring
point which is to be monitored. Click on the checkbox in front of the measuring point group/measuring point.
Alarm thresholds 1 and 2 Different limit value ranges which can
be defined for different periods. Lower limit, upper limit Values which are to be monitored Alarm delay Minimum duration of a limit value
violation before an alarm is triggered.
The time intervals between
measurements (measuring cycle)
should be lower than the alarm delay
(e.g. measuring cycle = 5 minutes,
alarm delay = 15 minutes). Time control Define individual alarm periods for
which the alarm limit values 1 and 2,
or no alarm limit value at all, apply. To
define the alarm value 1 and 2,
double-click on a time point in the
table or pull open the desired time
period with the mouse. During
periods in which the table is left
empty, you receive no alarm. If you
have not defined alarm periods, the
alarm limits will be active 24 hours a
day.
If alarm periods have been defined,
the limit value alarms are active only
Channel alarms Alarms in the event of a defective
E-mail recipient Addressees who are informed when
38
an alarm occurs. Click on the
checkbox in front of the recipient or
enter the name and e-mail address of
other recipients and click on the
Page 39
4 Operation
on the + Add button.
1
- Click on the "+ New system warning” button.
1
- Click on the title of a system warning.
the WiFi data logger for interruptions.
Setting Description SMS recipient Addressees who are informed when
an alarm occurs. Click on the checkbox in front of the recipient or enter the name and mobile phone number of other recipients and click
Save The settings are saved. Delete The alarm settings are deleted.

4.9 System warnings

4.9.1 Creating and displaying system warnings

A new system warning is created.
Existing system warnings are displayed below the button.
An existing setting is displayed.
4.9.2 Configuring and editing a displayed system
warning
The following settings can be defined and edited.
Settings Description Title Designation of the system warning
Battery almost discharged Monitor the WiFi data logger for
Power supply interrupted Monitor the external power supply of
(required field)
discharged battery.
39
Page 40
4 Operation
communication cycle.
click on the + Add button.
on the + Add button.
Settings Description WiFi data logger is not responding Monitor the WiFi data logger for data
transmission failure. Click on the
“Activate” button and configure the
monitoring cycle using the slide
control. The set time should be
greater than the WiFi data logger's
WiFi data logger WiFi data logger that needs to be
monitored. Click on the checkbox in
front of the WiFi data logger. E-mail recipient Addressees who are informed when
an alarm occurs. Click on the
checkbox in front of the recipient or
enter the names and e-mail
addresses of other recipients and
SMS recipient Addressees who are informed when
an alarm occurs. Click on the
checkbox in front of the recipient or
enter the name and mobile phone
number of other recipients and click
Save The settings are saved. Delete The alarm settings are deleted.

4.10 Configuration

4.10.1 Standard users

By default, two users are created in the system:
Account Owner (name can be changed), with Administrator user role (role
cannot be changed)
Support Testo (name can be changed), with Testo User Support user role
(role cannot be changed)

4.10.2 Creating and editing new users

Other users with different roles can be created and edited.
40
Page 41
4 Operation
1
- Click on the “Add a new user” button to create a new user.
2
- Click on the name of a user to display the settings.
3
- Click on the “Edit” button to change the settings.
Title
Title of the user.
First name
First name of the user (required field).
Second name
Second name of the user.
stage.
the system.
address also changes the login name.
and system warnings).
Active from
Date from which the user is active.
Active to
Date up to which the user is active.
specific information.
Save
The settings can be saved.
The following settings can be defined and edited:
Settings Description
Surname Surname of the user (required field). Password and Repeat password User password. The user password
User role Defines the user permissions within
E-mail address & login E-mail address of the user. The e-mail
Change e-mail address & login The field is only available when
Existing users are displayed in a list.
can be changed by the user at a later
address is also the login name. The e­mail address is also used for system notifications (alarms, system warnings).
editing the user account of the account holder. Enter a new e-mail address. Entering a new e-mail
Mobile number Phone number of the user. This is
used for system notifications (alarms
Details Text field for entering other user-

4.10.3 User roles

A description of the available user roles can be displayed.
41
Page 42
4 Operation
1
- Click on the title of a user role to display a description of it
Display created users
X X X
X
users
Display Account ID
X - - X Login WiFi data loggers
X - -
X
WiFi data loggers
areas
system warnings
warnings (= mark as read)
Create automatic areas
X - -
X
Users have different permissions depending on their allocated user role.
Permissions Admin Analyst Auditor Operator
Create, edit and delete
Configure and deactivate
Create, edit and delete
Display, create, edit and
X - - -
X - - X
X - - X
X - - X
delete alarm settings and
Read and analyze readings X X X X Display details about
X X X X
alarms and system
X = available, - = not available
42
Page 43
4 Operation
1
- Click on User to open the user menu.
Language
Language of the user interface.
Unit
Measurement parameters
Save
Settings can be saved.
1
- Enter the new password in both text fields ("New password" and
2
- Click on the “Save” button to save the new password.
1
- Click on the “Logoff” button to log off.

4.10.4 User management

The user management provides information and settings options for the user account.
4.10.4.1 User settings
The following user-specific settings can be made:
Setting Description
Time zone Time zone for the date and time
display.
4.10.4.2 Account information
Information about your testo 160 account is displayed.
4.10.4.3 Change password
"New password (repeat)".
4.10.4.4 Logoff

4.10.5 Account ID

The Account ID is the unique address of your user account in the Testo Cloud. This is needed to configure the WiFi data loggers in order to ensure that they send your data to the correct user account.

4.10.6 Creating and editing a measuring point group

Measuring points can be organized into measuring point groups. Allocating measuring points to a measuring point group (e.g. Room 1, Room 2, etc.) makes the administration of multiple measuring points easier.
43
Page 44
4 Operation
1
- Click on the “New measuring point group” button to create a new
is to be allocated to.
arrow
cross
group.
1
- Click on the “New area” button to create a new measuring point
2
- Click on the "Actions” button and then “Edit”.
For higher-level grouping, measuring point groups can be allocated to an area (e.g. ground floor, first floor, etc.).
measuring point group.
The following settings can be defined and edited:
Settings Description Title Designation of the measuring point
Description Description of the measuring point
Area Area that the measuring point group
Measuring points Measuring points that are available
Measuring point groups that already exist are displayed in a list.
group (required field).
group.
and those allocated to the measuring
point group are displayed. Click on
the
to allocate a measuring point to the group. Click on the to delete a measuring point from the
Save The settings can be saved. Delete The settings can be deleted.

4.10.7 Areas

Measuring point groups can be organized into areas. Allocating measuring point groups to an area (e.g. ground floor, first floor, etc.) makes the administration of multiple measuring point groups easier.
4.10.7.1 Creating and editing areas
group.
44
Areas already created are displayed in a list.
The settings are displayed and can be edited.
Page 45
4 Operation
field).
is to be allocated to.
1
- Click on the "Actions” button.
2
- Click on the “Delete” button to delete the area.
1
- Click on the “Details” button to display more information.
2
- Click on the “Deactivate” or “Activate” button to deactivate or
1
- Click on the “Configure” button to change the configuration.
must be selected.
The following settings can be defined and edited:
Setting Description Display name Designation of the area (required
Description Description of the area. Area Area that the measuring point group
Save The settings can be saved. Delete The settings can be deleted.
4.10.7.2 Deleting an area

4.10.8 WiFi data loggers

A summary of all logged on WiFi data loggers is displayed.
activate WiFi data loggers.
If a logger is deactivated, the measurement and alarm system are switched off. The logger remains logged into the Cloud.
Configuring a WiFi data logger
The following settings can be defined and edited:
Setting Description Name of the WiFi data logger Designation of the WiFi data logger
(required field). Condition at delivery:
"Model_Serial number". Description Description of the WiFi data logger. Select battery type Set the battery type used. For the
battery capacity to be displayed
correctly, the correct battery type
45
Page 46
4 Operation
using the slide control.
displayed.
Save
The settings can be saved.
activated.
1
- Click on the “Activate” button to install a firmware update, if this
1
Setting Description Display Switch the display of the WiFi data
logger on or off. (if present)
Name of the measuring point Specify the designation of the
measuring points.
Measuring cycle Interval at which readings are
obtained. Set the measuring cycle
Day communication cycle and energy-saving mode
Interval at which readings are transmitted to the Testo Cloud. Select the start time for the day communication cycle and the energy­saving mode. Set the communication cycle using the slide control.
Select unit Unit in which the readings are
Deactivate or Activate Measurement channels or WiFi data
loggers can be deactivated or
Remove The WiFi data loggers can be logged
off the system.

4.10.9 Firmware updates

A list of available firmware updates for the WiFi data loggers is displayed. Firmware updates can be installed on the data loggers via WiFi.
update is available as an optional update. Otherwise, the button is activated automatically.

4.11 Command bar

4.11.1 Opening the Setup assistant

The Setup assistant can help you log in WiFi data loggers.
- Click on the icon to open the Setup assistant.
46
Page 47
4 Operation
1
- Click on the question mark icon to open the Online Help.
1
- Click on the envelope icon to open the system messages.
2
- Click on the title of a system message to display more information.
1
- Click on the green checkmark or the alarm bell to open the Alarm

4.11.2 Opening the Online Help

The Setup assistant takes you through the menu step by step.
The Online Help (this document) provides you with support for issues related to the product components.

4.11.3 Opening system messages

The system messages contain important information relating to the product.
The number of unread system messages is displayed above the icon.
A summary of all system messages is displayed.
Unread system messages are shown in bold.
On displaying the detailed information, the system message is marked as "read" and the message counter is reduced.

4.12 System and status information

Unacknowledged alarms (green checkmark): no alarms active. Unacknowledged alarms (alarm bell): alarms active, number of unread alarms is
displayed.
list.
47
Page 48

5 FAQ

5 FAQ
Can the WiFi data logger be connected to the PC using any USB cable?
We recommend that you use the USB cable supplied with the WiFi data logger to guarantee stable data transmission. Longer USB cables are suitable for the power supply only.
Can the WiFi data logger also be used in networks with WPA2
Enterprise encryption? testo 160 data loggers can be used in networks with the following WPA2 Enterprise encryption methods. WPA2 Enterprise: EAP-TLS, EAP-TTLS-TLS, EAP-TTLS-MSCHAPv2, EAP­TTLS-PSK, EAP-PEAP0-TLS, EAP-PEAP0-MSCHAPv2, EAP-PEAP0-PSK, EAP-PEAP1-TLS, EAP-PEAP1-MSCHAPv2, EAP-PEAP1-PSK, WPA Personal, WPA2 (AES), WPA (TKIP), WEP
To integrate the loggers into the WPA2 Enterprise network, proceed as follows:
1. Open the PDF file stored on the logger and generate a corresponding XML file by selecting the programming options step by step.
2. Copy your company-specific WPA2 Enterprise certificates and the generated .XML file to the logger's mass storage via USB using drag & drop.
3. Please note that the configuration of the WiFi data logger will only be fully transferred once the USB connector has been removed.
The XML configuration file is not being applied by the WiFi data logger,
what can I do? Depending on the operating system, there may be difficulties with the data transfer if the configuration file name has been changed. Leave the default file name.
The humidity sensor has been stored at a high temperature (> 30 °C)
and in very high humidity (> 80% RH) for a long period of time, what can I do? The sensor requires a long period of time to regenerate itself again. This process can be accelerated by storing the sensor in a well-ventilated location at a high temperature (> 30 °C) and in low humidity (< 20% RH) for at least 12 hours.
The WiFi data logger's wireless connection to the access point was
interrupted, what can I do?
1. Press the control key on the WiFi data logger to start searching for a WLAN connection manually.
2. Change the alignment or position of the WiFi data logger or the access point (WLAN router).
48
Page 49
5 FAQ
The error codes can be read out using a web browser via a smartphone/tablet or PC. Press the probe button for 3 seconds. Then enter the following IP address 192.168.1.1 in the web browser.
The error codes below are only displayed on the testo 160 IAQ.
The WiFi data logger (160 IAQ) is displaying error code E03, E04, E05
or E09, what can I do? An error has occurred in the WiFi data logger. The error will automatically be corrected by the firmware of the WiFi data logger. After a few seconds the error code should no longer be displayed, you do not need to do anything.
The WiFi data logger (160 IAQ) is displaying error code E12, what can I
do? The configuration file WifiConfig.xml indicates an error. Use the Quick Start Guide to create a new configuration file and save this on the WiFi data logger.
The WiFi data logger (160 IAQ) is displaying error code E23, what can I
do? The most common reason for this error is low battery. Insert new batteries into the WiFi data logger. If this does not solve the problem: Reset the WiFi data logger to its factory settings. To do this, press and hold down the control key for > 20 s until the display goes blank. If the error code continues to be displayed, then there is a hardware problem.
The WiFi data logger (160 IAQ) is displaying error code E26, what can I
do?
1. The access point (WLAN router) has no connection to the internet. Check the access point's internet connection.
2. The routing within the network infrastructure is not working, check whether too many terminal devices are logged into the access point.
The WiFi data logger (160 IAQ) is displaying error code E32, what can I
do? The WiFi data logger has not obtained an IP address. There are 2 possible reasons for this error:
1. The network password is incorrect. Check the password of the WLAN network. Use the Quick Start Guide to create a new configuration file with the correct password and save this on the WiFi data logger.
2. The access point (WLAN router) has a MAC filter or does not permit the integration of new devices.
Please contact our Customer Service.
Check the settings for the access point.
49
Page 50
5 FAQ
The WiFi data logger (160 IAQ) is displaying error code E35, what can I
do? The WiFi data logger has not received any reply to its test ping from the access point (WLAN router). Make sure that a ping to the gateway is allowed within the access point configuration.
The WiFi data logger is displaying error code E36, what can I do?
No DNS available or accessible. Contact the operator of the WLAN network.
The WiFi data logger is displaying error code E41, what can I do?
The WiFi data logger cannot obtain any current time from a time server (pool.ntp.org).
1. The access point (WLAN router) has no connection to the internet. Check the access point's internet connection.
2. The NTP port (123/UDP) of the access point (WLAN router) is not open. Check whether the NTP port (123/UDP) is opened.
The WiFi data logger (160 IAQ) is displaying error code E51, what can I
do? The WiFi data logger was not able to connect to the Testo Cloud.
1. If the WiFi data logger has already been connected to the Testo Cloud and this connection is suddenly no longer possible: The Testo Cloud servers are not currently accessible. The servers will be monitored and should be accessible again within a few hours.
2. If the WiFi data logger has not yet been connected to the Testo Cloud: The TCP ports (1883 or 8883) of the access point (WLAN router) are not open. Check whether the TCP ports (1883 or 8883) are open in both directions.
The WiFi data logger is displaying error code E52, what can I do?
The WiFi data logger could not log into the Cloud because it is already logged into another account. Please log the WiFi data logger out of the existing account first.
The WiFi data logger (160 IAQ) is displaying error code E63, what can I
do? The WiFi data logger could not send any data to the Testo Cloud.
1. The internet connection was interrupted during the transmission. Check whether there is a stable connection from the WiFi data logger to the access point (WLAN router). Check the access point's internet connection. The data will be transferred during the next communication cycle. Alternatively: Initiate data transmission manually by pressing the control key on the WiFi data logger.
2. The Testo Cloud server was not able to process the request for data storage. The servers will be monitored and should be accessible again within a few hours.
50
Page 51
5 FAQ
The WiFi data logger is displaying error code E69, what can I do?
1. The Account ID contained in the configuration file is missing or is not valid. Create a new configuration file and save this on the WiFi data logger.
2. An attempt was made to log the testo 160 E WiFi data logger into the Cloud without any external probes connected. external probes before logging in.
The WiFi data logger (160 IAQ) is displaying error code E75, what can I
do? A firmware update for the WiFi data logger failed. The internet connection was interrupted during the transmission or the data was not received intact by the WiFi data logger for other reasons. Check whether there is a stable connection from the WiFi data logger to the access point (WLAN router). Check the access point's internet connection. The data will be transferred during the next communication cycle. Alternatively: Initiate data transmission manually by pressing the control key on the WiFi data logger.
The WiFi data logger (160 IAQ) is displaying the warning message
Err AccountID, what can I do? The AccountID contained in the configuration file is not valid. Use the Quick Start Guide to create a new configuration file and save this on the WiFi data logger.
The WiFi data logger (160 IAQ) is displaying the warning message
no AccountID, what can I do? There is no AccountID in the configuration file. Use the Quick Start Guide to create a new configuration file and save this on the WiFi data logger.
The WiFi data logger is displaying the warning message no License,
what can I do? The WiFi data logger cannot be logged in because the number of WiFi data loggers permitted to log in has been exceeded or your testo 160 licence has expired. Log off another WiFi data logger, extend or renew your testo 160 licence.
The WiFi data logger (160 IAQ) is displaying the warning message
not Active, what can I do? The WiFi data logger has been deactivated. It is not storing, and therefore not sending, any measurement data to the Testo Cloud. Activate the WiFi data logger (under Configuration --> WiFi data logger) when the WiFi data logger needs to store and send measurement data again.
Connect the required
51
Page 52

6 Cloud licences

be used in countries for which a country certification has been granted.
testo 160 TH
0572 2021
07.06.2018
testo 160 E
0572 2022
07.06.2018
testo 160 THE
0572 2023
07.06.2018
07.06.2018
07.06.2018
E 1561
6 Cloud licences
When you purchase the testo 160 system, you receive a free basic licence. Some measuring functions displayed in the testo 160 system are inactive.
Upgrade your licence in order to activate and use these functions. In the header you can see which licence you are currently using.
Purchasing a licence
Click on the licence display in the header to be redirected to the licence shop. Here you can choose and purchase your licence upgrade. Or Click on the “Upgrade” symbol next to the inactive measuring function. This also opens the licence shop for you to upgrade your licence.

7 Authorizations

The use of the wireless module is subject to the regulations and stipulations of the respective country of use, and the module may only
The user and every owner has the obligation to adhere to these regulations and prerequisites for use, and acknowledges that the re­sale, export, import etc. in particular in countries without wireless permits, is his responsibility.
Product Mat.-No. Date
testo 160 THL 0572 2024 testo 160 IAQ 0572 2014
Country Comments Australia
52
Page 53
Country Comments
em sistemas devidamente autorizados.
Testo 160 IAQ: CMIIT ID: 2017DJ3243
Brazil
testo 160 TH
testo 160 E
testo 160 THE
testo 160 THL
testo 160 IAQ
Este equipamento não tem direito à proteção contra interferência prejudicial e não pode causar interferência
7 Authorizations
Canada Contains IC : 21461-LSD4WF0459
TH/E/THE/THL: IC: 6127B-0572202X IAQ: IC: 6127B-05722014
IC Warnings
China Testo 160 TH: CMIIT ID: 2017DJ4557
Testo 160 E: CMIIT ID: 2017DJ4559 Testo 160 THE: CMIIT ID: 2017DJ4564 Testo 160 THL: CMIIT ID: 2017DJ4547
53
Page 54
7 Authorizations
Japan Information
KCC Warning
Emirates
FCC Warnings
Country Comments Europa + EFTA
EU countries: Belgium (BE), Bulgaria (BG), Denmark (DK), Germany (DE),
Estonia (EE), Finland (FI), France (FR), Greece (GR), Ireland (IE), Italy (IT), Latvia (LV), Lithuania (LT), Luxembourg (LU), Malta (MT), Netherlands (NL), Austria (AT), Poland (PL), Portugal (PT), Romania (RO), Sweden (SE), Slovakia (SK), Slovenia (SI), Spain (ES), Czech Republic (CZ), Hungary (HU), United Kingdom (GB), Republic of Cyprus (CY).
EFTA countries: Iceland, Liechtenstein, Norway, Switzerland
Japan
South Africa ICASA
The EU Declaration of Conformity can be found on the testo homepage under the product specific downloads.
South Korea
United Arab
USA Contains FCC ID: N8NLSD4WF0459
54
Radio Equipment Type Approval Number: testo 160 IAQ: TA-2018/075
testo 160 TH: R-CRM-te2-05722021 testo 160 THL: R-CRM-te2-05722024 testo 160 IAQ: R-CRM-te2-05722014
Authorization Number: ER57487/17
TH/E/THE/THL: FCC ID: WAF-0572202X
IAQ: FCC ID: WAF-05722014
Page 55
7 Authorizations
Feature
Values
WLAN Range
100 m
WLAN type
LSD4WF0459-01D0
802.11b/g/n
Company
Lierda Technology Group co., LTD
RF Band
2412-2472MHz
Transmitter Power
13.42dBm
Country Comments Wi-Fi-Module
WLAN radio class Accord with the standard of IEEE
IC Warnings:
This instrument complies with Part 15C of the FCC Rules and Industry Canada RSS-210 (revision 8). Commissioning is subject to the following two conditions: (1) This instrument must not cause any harmful interference and (2) this instrument must be able to cope with interference, even if this has undesirable effects on operation.
Cet appareil satisfait à la partie 15C des directives FCC et au standard Industrie Canada RSS-210 (révision 8). Sa mise en service est soumise aux deux conditions suivantes :
(1) cet appareil ne doit causer aucune interférence dangereuse et (2) cet appareil doit supporter toute interférence, y compris des interférences
qui provoquerait des opérations indésirables.
FCC Warnings:
Information from the FCC (Federal Communications Commission) For your own safety Shielded cables should be used for a composite interface. This is to ensure
continued protection against radio frequency interference.
FCC warning statement
This equipment has been tested and found to comply with the limits for a Class C digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio
55
Page 56
7 Authorizations
frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
Caution
Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. Shielded interface cable must be used in order to comply with the emission limits.
Warning
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Japan Information:
当該機器には電波法に基づく、技術基準適合証明等を受けた特定無線設備を装 着している。
KCC Warning
해당 무선 설비는 운용 전파혼신 가능성이 있음.
56
Page 57
0970 1601 en 05
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Loading...