5.3. Reading out data ........................................................................ 21
6 Maintaining the product ..................................................................... 23
6.1. Changing the batteries ................................................................ 23
6.2. Cleaning the instrument .............................................................. 24
7 Tips and assistance ........................................................................... 25
3
Page 4
2 Safety and the environment
2 Safety and the environment
2.1. About this document
Use
> This instruction manual is an essential component of the
product.
> Please read this documentation through carefully and
familiarize yourself with the product before using it. Pay
particular attention to the safety instructions and warning
notices in order to prevent injuries and damage to the product.
> Keep this document to hand so that you can refer to it when
necessary.
> Make sure that the instruction manual is read by subsequent
users of the product.
2.2. Ensure safety
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not
use any force.
> Do not use the product if there are signs of damage to the
housing.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts
from Testo.
2.3. Protecting the environment
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
regulations) or return the product to Testo for disposal.
WEEE Reg. Nr. DE 75334352
4
Page 5
3 Specifications
3.1. Use
The testo 184 USB transport data loggers are used to save and
read out individual readings and measurement series. They have
been specifically designed for monitoring the transportation of
products subject to cold chain requirements.
Temperature and humidity readings are saved throughout the
entire duration of the measurement program.
Acceleration readings are monitored throughout the duration of the
measurement program and saved when the set limit value is
exceeded.
Data logger programming and measurement report output are
implemented via PDF files, no software needs to be installed.
Product versions T1 and T2 are single-use data loggers with a
time-limited service life.
Configuration and readout of the testo 184 G1 USB
transport data logger are not possible with the testo
Saveris CFR Software and its transport add-ons.
3 Specifications
3.2.Technical data
testo 184 T1
Feature Values
Display No
Probe type internal NTC temperature sensor
Measuremen
1 internal
t channels
Measuremen
Temperature [°C, °F]
t parameters
[unit]
Measuring
-35 to 70 °C
range
Accuracy ±0.5 K
Resolution 0.1 °C
Operating
-35 to 70 °C
temperature
5
Page 6
3 Specifications
logger)
Feature Values
Storage
-35 to 70 °C
temperature
Battery type internal, non-replaceable
Operating
time (single-
90 days from first program start (5-minute
measuring cycle, -35 °C)
use data
Protection
IP67 (when USB protective cap closed)
class
Measuring
1 minute to 24 hours
interval
Memory 16,000 readings
Dimensions 33 x 9 x 74 mm
Weight 25 g
Directives,
standards,
certificates
2014/30/EU, EN 12830, HACCP-certified,
temperature calibration certificate traceable
according to ISO 17025
testo 184 T2
Feature Values
Display Yes
Probe type internal NTC temperature sensor
Measuremen
1 internal
t channels
Measuremen
Temperature [°C, °F]
t parameters
[unit]
Measuring
-35 to 70 °C
range
Accuracy ±0.5 K
Resolution 0.1 °C
Operating
-35 to 70 °C
temperature
Storage
-35 to 70 °C
temperature
Battery type internal, non-replaceable
6
Page 7
3 Specifications
Feature Values
Operating
time (single-
150 days from first program start (5-minute
measuring cycle, -35 °C)
use data
logger)
Protection
IP67 (when USB protective cap closed)
class
Measuring
1 minute to 24 hours
interval
Memory 40,000 readings
Dimensions 44 x 12 x 97 mm
Weight 45 g
Directives,
standards,
certificates
2014/30/EU, EN 12830, HACCP-certified,
temperature calibration certificate traceable
according to ISO 17025
testo 184 T3
Feature Values
Display Yes
Probe type internal NTC temperature sensor
Measuremen
1 internal
t channels
Measuremen
Temperature [°C, °F]
t parameters
[unit]
Measuring
-35 to 70 °C
range
Accuracy ±0.5 K
Resolution 0.1 °C
Operating
-35 to 70 °C
temperature
Storage
-35 to 70 °C
temperature
Battery type CR2450, replaceable
Battery life
500 days (15-minute measuring cycle, 25 °C)
(reusable
data logger)
7
Page 8
3 Specifications
Weight
45 g
Feature
Values
Feature Values
Protection
class
Measuring
interval
Memory 40,000 readings
Dimensions 44 x 12 x 97 mm
IP67 (when USB protective cap closed)
1 minute to 24 hours
Directives,
standards,
certificates
testo 184 T4
Display No
Probe type Internal PT1000 temperature sensor
Measuremen
t channels
Measuremen
t parameters
[unit]
Measuring
range
Accuracy ±0.8 K (-80 to -35.1 °C), ±0.5 K (-35.0 to 70 °C)
Resolution 0.1 °C
Operating
temperature
Storage
temperature
Battery type TLH-2450, replaceable
Battery life
(reusable
data logger)
Protection
class
Measuring
interval
2014/30/EU, EN 12830, HACCP-certified,
temperature calibration certificate traceable
according to ISO 17025
1 internal
Temperature [°C, °F]
-80 to 70 °C
-80 to 70 °C
-80 to 70 °C
100 days (15-minute measuring cycle, -80 °C)
IP67 (when USB protective cap closed)
1 minute to 24 hours
8
Page 9
3 Specifications
Feature
Values
±1 %RH/year drift
Feature Values
Memory 40,000 readings
Dimensions 44 x 12 x 97 mm
Weight 45 g
Directives,
standards,
certificates
2014/30/EU, EN 12830, HACCP-certified,
temperature calibration certificate traceable
according to ISO 17025
testo 184 H1
Display Yes
Probe type internal digital humidity/temperature sensor
Measuremen
2 internal
t channels
Measuremen
temperature [°C, °F], relative humidity [%]
t parameters
[unit]
Measuring
range
-20 to 70 °C
0 to 100% (not for condensing atmospheres)
1
Accuracy ±0.5 K (0.0 to 70 °C), ±0.8 K (-20 to -0.1 °C)
±1,8 % RH, +3 % of the reading (at 25 °C, 5 %
to 80 % RH)
1
±0,03 % RH (at 0 to 60 °C)
±1 %RH hysteresis
1
The humidity sensor has the highest accuracy between 5 °C and 60 °C, and
between 20 %RH and 80 %RH. If the instrument is exposed to high humidity for
a longer period of time, the measuring accuracy decreases. The humidity sensor
is regenerated within 48 hours via storage at 50 %RH ±10 % and 20 °C ±5 °C.
9
Resolution 0.1 °C
0.1 %RH
Operating
-20 to 70 °C
temperature
Storage
conditions
-55 to 70 °C
30 to 60 %RH
Page 10
3 Specifications
Feature
Values
Feature
Values
Feature Values
Battery type CR2450, replaceable
Battery life
500 days (15-minute measuring cycle, 25 °C)
(reusable
data logger)
Protection
IP 30
class
Measuring
1 minute to 24 hours
interval
Memory 64,000 readings (temperature and relative humidity)
Probe type internal digital humidity/temperature sensor &
internal 3-axis accelerometer
Measuremen
5 internal
t channels
Measuremen
t parameters
temperature [°C, °F], relative humidity [%],
acceleration [g, m/s²]
[unit]
Measuring
range
-20 to 70 °C
0 to 100 % (not for condensing atmospheres)
2
0 to 27 g
2
The humidity sensor has the highest accuracy between 5 °C and 60 °C, and
between 20 % and 80 %RH. If the instrument is exposed to high humidity for a
longer period of time, the measuring accuracy decreases. The humidity sensor
is regenerated within 48 hours via storage at 50 %RH ±10 % and 20 °C ±5 °C.
10
Page 11
3 Specifications
±1,1,1 m/s² + 5 % of the reading
Feature Values
Accuracy ±0.5 K (0.0 to 70 °C), ±0.8 K (-20 to -0.1 °C)
±1,8 % RH, +3 % of the reading (at 25 °C, 5 %
to 80 % RH)
Close the USB cover while using the data logger in order to prevent
moisture or liquids from getting into the instrument and damaging
the instrument.
CAUTION
Humidity sensor may be damaged!
After using the data logger for up to 60 hours in high-humidity areas
above 80%, it must be brought to relaxation. Here, the conditions
should be +25 °C +/- 5 °C and a relative humidity of 50 % +/- 10 %.
EU conformity
12
The EU Declaration of Conformity can be found on the
testo homepage www.testo.com under the product
specific downloads.
EU countries:
Belgium (BE), Bulgaria (BG), Denmark (DK), Germany (DE), Estonia
(EE), Finland (FI), France (FR), Greece (GR), Ireland (IE), Italy (IT),
Latvia (LV), Lithuania (LT), Luxembourg (LU), Malta (MT),
Netherlands (NL), Austria (AT), Poland (PL), Portugal (PT), Romania
(RO), Sweden (SE), Slovakia (SK), Slovenia (SI), Spain (ES), Czech
Republic (CZ), Hungary (HU), United Kingdom (GB), Republic of
Cyprus (CY).
EFTA countries:
Iceland, Liechtenstein, Norway, Switzerland
Page 13
4 Product description
4.1. Status LEDs
To increase the battery life, the status LEDs are not permanently
illuminated. They flash once every 5 seconds.
In hibernation mode, the status LEDs are disabled.
Alarm
Feature LED colour
No alarm green
Alarm red
Battery
Feature LED colour
Battery life > 10 days green
Battery life < 10 days red
Mode
Feature LED colour
WAIT mode (wait for program start) green and red
Rec mode (measurement program running) green
End mode (measurement program finished) red
4 Product description
13
Page 14
4 Product description
20 °C).
measuring accuracy. We recommend the use of fully
4.2. Display (LCD)
Not available for all product versions.
1 Measurement program running
2 Measurement program finished
3 Wait for start of measurement program
4 Highest saved reading
5 Lowest saved reading
6 Reading
7 Status information: start criterion Date/time programmed/time
mark, xyz measurement axes for acceleration measurement,
Alarm set limit value(s) exceeded, set limit value(s) not
exceeded
8 Units
9 Battery capacity: sufficient, part empty, low,
(flashing) empty
10 Lower limit value exceeded
11 Upper limit value exceeded
For technical reasons, the display speed of liquid crystal
displays becomes slower at temperatures below 0 °C
(approx. 2 seconds at -10 °C, approx. 6 seconds at This has no influence on the measuring accuracy.
For technical reasons, battery performance decreases at
lower temperatures. This has no influence on the
14
Page 15
4 Product description
charged batteries in order to avoid a reset of the instrument
at low temperatures.
4.3.Key functions
Commissioning
The data loggers are delivered in hibernation mode to prolong the
battery life. In this mode, the status LEDs and the display are
disabled.
> Press the START key or the STOP key.
- Wait mode is activated.
START key
✓Wait mode and start criterion Button Start programmed.
> Press START key for approximately 3 seconds to start the
measurement program.
- The measurement program starts: Mode status LED flashes
green, Rec appears on the display.
✓Product versions with display:
> Press START key to toggle between the displays.
Display sequence (max. display range per version, certain data is
not displayed depending on the operating mode):
Display T2 T3 H1 G1
Current temperature reading (°C / °F) X X X X
Current average value MKT (Mean Kinetic
Temperature)
Current relative humidity reading (%) - - X X
Current acceleration reading, X axis (x, g) - - - X
Current acceleration reading, Y axis (y, g) - - - X
Current acceleration reading, Z axis (z, g) - - - X
Maximum temperature reading (Max, °C / °F)
Minimum temperature reading (Min, °C / °F)
Maximum relative humidity reading (Max, %)
Minimum relative humidity reading (Min, %)
X X X X
X X X X
X X X X
- - X X
- - X X
15
Page 16
4 Product description
Display T2 T3 H1 G1
Maximum acceleration reading , X axis (Max,
x, g)
Maximum acceleration reading, Y axis (Max,
y, g)
Maximum acceleration reading, Z axis (Max,
z, g)
Time mark ( )
Battery life in days ()
STOP key
✓Rec mode and stop criterion Button Stop programmed.
> Press STOP key for approximately 3 seconds to end the
- The measurement program ends: Mode status LED flashes red,
START + STOP key
The data loggers may be placed in hibernation mode to prolong the
battery life. In this mode, the status LEDs and the display are
disabled.
✓WAIT or End operating mode.
> Press the START key and the STOP key simultaneously for
- Hibernation mode is activated.
- - - X
- - - X
- - - X
X X X X
X X X X
measurement program.
End appears on the display.
approximately 3 seconds.
4.4. Important information and glossary of terms
• Single-use data logger (versions T1 and T2): the data logger
has a time-limited service life, which begins from the first time
the program is started.
• Start and stop setting: the criteria for starting and stopping the
program are defined in the configuration file.
One of the criteria must be selected for starting the program.
When selecting the criterion, a time delay can be entered
(program starts x minutes after pressing the key).
Both criteria can also be selected for stopping the program. The
criterion that occurs first stops the program.
• Measuring interval: the measurement interval defines the
intervals at which readings are saved.
16
Page 17
4 Product description
• Time mark: time marks can be set for documentation by
pressing the START key for 3 seconds during the
measurement, e.g. when responsibility is switched to another
institution. A maximum of 10 time marks can be set. Setting a
time mark resets the statistical values Min, Max and MKT.
• Acceleration (shock): the (positive and negative) acceleration
is measured in 3 measurement axes. Only those readings that
exceed the set limit value (maximum value per 1 second) are
stored and displayed.
The acceleration readings of the 3 measurement axes are
displayed separately on the display of the data logger.
The maximum cumulative value (peak) of the 3 measurement
axes is displayed in the PDF report.
• Report time zone: defines the time zone to which all time
specifications in the measurement report refer. Any time zone
changes during the measurement are not taken into account.
If the data logger was in rSt mode and could not be
reconfigured, configuration via a copy of an XML file
may result in the time & time zone not being correct.
• Reset mode (rSt): is triggered by an interruption to the power
supply, e.g. during battery replacement. A reconfiguration of the
data logger is required to resume operation. Data that has
already been recorded will not be affected.
• MKT (mean kinetic temperature): the MKT is a single calculated
temperature. MKT may be considered an isothermal storage
temperature. It simulates the non-isothermal effects of storage
temperature variations.
Calculation:
=
/
+
/
/
+
/
Tmkt = mean kinetic temperature in degrees Kelvin
∆E = activation energy (standard value: 83.144 kJ/mol)
R = universal gas constant (0.0083144 kJ/mol)
T1 = average temperature in degrees Kelvin during the first time
period
Tn = average temperature in degrees Kelvin during the nth time
period
• MKT activation energy: the default activation energy is set at
83.144 kJ/mol, as recommended in USP <1160>. If other
estimates are available as a result of studies that have been
carried out, the activation energy can be customised.
• Individual alarm: an alarm is triggered when the set limit value
is exceeded.
17
Page 18
4 Product description
• Cumulative alarm (only for temperature and humidity
measurement): an alarm is not triggered when the set limit
value is first exceeded, but only once the overall duration,
during which limit values are exceeded, exceeds the set waiting
period (allowed time).
• Wall bracket (delivery includes version G1): for the
acceleration measurement, the data logger must be
permanently connected to the object to be monitored.
Mount the wall bracket using 2 screws or 2 cable ties and then
push the data logger into the wall bracket.
18
Page 19
5 Using the product
5.1. Configuring the data logger
Showing/changing the configuration
Adobe Reader software (version X or later) is required.
The data logger must not be in Rec mode.
1. Connect the data logger to a PC via the USB port.
- The status LEDs are disabled, uSb is displayed (instruments
with display). The device drivers are installed automatically.
- The window Automatic playback is displayed.
2. Click on Open folder to view files.
- The file explorer opens.
3. Open the file testo 184 configuration.pdf.
4. Make changes to the configuration. Please note:
• The type of instrument used must be set correctly.
• Existing configuration data can be imported by clicking on
the Import button. The configuration data to be imported
must be available in XML data format.
• When using the Configuration Assistant, some functions are
predefined or filled in automatically. Expert Mode must be
enabled in order to use and manually set all instrument
functions.
5. Export changes to the configuration by clicking on the
button on the data logger.
- A window opens for exporting form data.
6. Select the data logger as the storage location
(Drive TESTO 184) and export the configuration data by
clicking on the Save button.
- The configuration is stored on the data logger as an XML file.
The XML file can be used as a template for other data loggers (via
the import function in the configuration PDF)
5 Using the product
19
Page 20
5 Using the product
ATTENTION
Incorrect configuration of the time settings!
> Configuration of the data logger via copy/paste of the XML
file directly to the data logger memory is not recommended,
since the local time and time zone settings are not accepted
in this way. Use the configuration PDF to apply the time
settings of the PC used.
7. Close file.
A message Do you want to save the changes to “testo 184
configuration.pdf” before closing? may appear.Answer this
question with No.
8. Disconnect the data logger from the PC.
- The data logger switches to Wait mode, the Mode status LED
flashes green/red.
Configuring several data loggers with the same settings
> Configure your measurement protocol using the configuration
PDF or import an existing XML file.
> Connect the testo 184 USB transport data logger to a USB
connection.
> Click to save the configuration on the connected
testo 184 USB transport data logger.
> Leave the configuration PDF open. Connect the next testo 184
USB transport data logger. Repeat the last step in order to
export the identical configuration
20
Changing the logo for the measurement data report
A logo is inserted in the measurement data report. This can be
customised.
The logo must be available in JPEG data format, the file size
should not exceed 5 kB and the file name must be Logo.jpg.
> Create a logo that meets the above criteria and copy it to the
data logger.
Configuring the data logger using Testo PC software
Alternatively, the data logger can also be configured using the testo
Comfort Software Professional (version 4.3 service pack 2 or later),
testo Comfort Software CFR (version 4.3 service pack 2 or later)
and testo Saveris CFR Software including transport add-ons.
Please consult the relevant software instruction manual for this.
Page 21
In order to ensure conformity of the testo Comfort Software 21
CFR Part 11 and the testo Saveris CFR Software, configuration
via the PDF file is no longer possible once the testo 184 USB
transport data logger has been configured using the abovementioned software.
5.2.Measuring
Starting measurement
Depending on the configuration of the data logger, the
measurement program is started via one of the following criteria:
• Button Start: hold down the START key for > 3 seconds.
• Time Start: the measurement starts automatically once the
configured time has been reached.
- The data logger switches to Rec mode, the Mode status LED
flashes green.
Setting time marks
While the measurement program is running (Rec mode), up to
10 time marks can be set. These are used to document the transfer
of responsibility, for example.
> Hold down the START key for > 3 seconds.
- The number of set time marks is displayed for 3 seconds and
flashes three times (instruments with display), Mode status LED
flashes green/red three times.
5 Using the product
Ending measurement
Depending on the configuration of the data logger, the
measurement program is ended via one of the following criteria:
• Button Stop: hold down the STOP key for > 3 seconds.
• Time Stop: the measurement stops automatically once the
configured time has been reached.
- The data logger switches to End mode, the Mode status LED
flashes red.
5.3.Reading out data
Displaying a measurement data report
Adobe Reader software (version 5 or later) or compatible software
is required to display PDF/A files.
1. Connect the data logger to a Windows PC via the USB port.
21
Page 22
5 Using the product
The readings displayed in the diagram in the PDF report are limited
The default setting for the NFC function is "Off". If the NFC function
- The status LEDs are disabled, uSb is displayed (instruments
- The window Automatic playback is displayed.
2. Click on Open folder to view files.
- The file explorer opens.
3. Open the file testo 184 measurement report.pdf.
- The measurement data report is displayed.
> Save or print the report as required.
to a measurement series of 324 readings. The internal algorithm
automatically selects the readings for the diagram and displays
these. Thus important readings may not be displayed in the
diagram. We recommend using testo ComSoft to read and display
all readings for measurement series of more than 324 readings.
Detailed measurement data analysis
To carry out detailed analysis and to further process readings, the
testo Comfort Software Professional (version 4.3 service pack 2 or
later), testo Comfort Software CFR (version 4.3 service pack 2 or
later) or testo Saveris CFR Software including transport add-ons is
required (accessory). Please consult the relevant software
instruction manual for this. The shock values of the testo 184 G1
data logger are not available as separately readable data. The
shock values are only shown in the graphic.
with display). The device drivers are installed automatically.
Measurement data output via NFC
The data loggers are equipped with an NFC (Near Field
Communication) transmitter. This allows instrument data to be read
out via short-range radio using compatible devices (for example
a report printer with NFC). Unfortunately, it is not possible to read
the NFC interface via android or Apple devices.
• The NFC function on your data logger can be enabled/disabled
in the configuration file.
• You do not need any additional software to transfer data to
a compatible Testo report printer (e.g. mobile printer for data
loggers 0572 0576).
• To transfer data, the data logger must be placed on the NFC
transmitter () of the target device.
• Please also consult the instruction manual for the target device.
is to be used, it must be set to "On" prior to use.
22
Page 23
6 Maintaining the product
6.1. Changing the batteries
It is not possible to change the batteries for instrument types T1
and T2 (single-use data loggers).
A battery replacement stops any measurement that is
currently running. The saved measurement data and
generated PDF reports, however, remain permanently
stored.
An interruption of the power supply leads to the testo 184
USB transport data logger's time settings being reset. In
order to reinstate the correct time setting, configuration
must be carried out via the PDF file, the Comfort Software
or the testo Saveris 184 config tool.
1. Read out stored data.
2. Place the data logger on its front.
3. Open the battery compartment on the back of the data logger
by turning anti-clockwise. Use a coin for this purpose.
4. Take the spent battery out of the battery compartment.
After removing the old battery, please wait 10 seconds or
press the start key approx. 5~10 times until the indicator on
the LCD display goes completely blank and no LED is
flashing.
5. Insert new batteries (see technical data for type required) into
the instrument so that the positive terminal is visible.
When changing batteries, only use new, full batteries. If a
partially exhausted battery is inserted, the battery capacity
will not be calculated correctly.
Only change the battery when the battery icon on the LCD
display is empty.
Before and during use, check that the battery LED flashes
red and the battery icon on the LCD display is "full".
6. Place the battery compartment cover on the battery
compartment and close it by turning clockwise. Use a coin for
this purpose.
- The data logger is in reset mode, rSt is illuminated
(instruments with display), status LEDs are disabled.
6 Maintaining the product
23
Page 24
6 Maintaining the product
7. Reconfigure the data logger, see the section Configuring the
data logger.
For verification of battery performance, we only used
batteries from EVE and Panasonic. This is why we
recommend using batteries from these manufacturers in
our testo 184 product.
6.2.Cleaning the instrument
CAUTION
Damage to the sensor!
> Ensure that no liquid enters the inside of the housing.
> If the housing of the instrument is dirty, clean it with a damp
cloth.
Do not use any aggressive cleaning agents or solvents! Weak
household cleaning agents or soap suds can be used.
24
Page 25
7Tips and assistance
Question
Possible causes/solution
all status LEDs are flashing
If you have used the Comfort Software
184
Repeat the configuration to restore
Questions and answers
7 Tips and assistance
E0x is displayed
(instruments with display),
red
---- is displayed
(instruments with display)
Hi is displayed
(instruments with display)
Lo is displayed
(instruments with display)
En is displayed
(instruments with display)
Err is displayed
(instruments with display)
Configuration via the PDF
file is not possible
The size of the PDF
configuration file is 0 kB or
the file is damaged.
Different time or time zone
in the report
An error has occurred.
• E01: Configuration unsuccessful /
PDF file faulty.
• E02, E03, E04 or E05: sensor
faulty.
• E06: maximum number of time
marks set, new time mark cannot
be set.
• No reading available (after setting
a time mark)
• Reading invalid.
Reading is above the measuring
range.
Reading is below the measuring
range.
Set time marks function is disabled.
Configuration not possible,
e.g. because Rec mode is active.
21 CFR Part 11 for the configuration,
configuration via the PDF file is
disabled.
Copy the PDF file of another testo
USB transport data logger or
download the configuration file from
the Testo website:
http://www.testo.com/.
> The testo 184 USB transport data
logger has not been configured
after a battery replacement.
25
the correct time settings.
> Check whether the PC which was
used for the configuration has the
correct time settings.
Page 26
7 Tips and assistance
PDF file.
Please check whether the LCD display
Please check whether the LED display
Question Possible causes/solution
No measurement protocol
was created.
The PDF configuration is
not ready for use.
There is no LCD display.
There is no LED display.
There is no time mark
display.
There is no NFC. Please check whether the NFC
> Please check whether the data
logger is in record/end mode.
> Please reconnect the data logger
to the PC.
> Please check whether there is
enough free memory capacity
available on the data logger.
> Please check whether the data
logger is in record mode.
> Please check whether the correct
data logger mode is selected.
> Please check whether the
testo 184 USB transport data
logger was configured via the
Comfort Software 21 CFR Part 11.
This prevents configuration via the
is disabled in the configuration.
is disabled in the configuration.
Please check whether the time mark
display is disabled in the configuration.
display is disabled in the configuration.
26
Page 27
Question Possible causes/solution
relative humidity of more than 80%
with a different PDF reader
The humidity value
measured is out of
tolerance.
Measurement protocol
cannot be opened or
shows 0 KB.
At what point is a shock
experience critical for an
object?
Can a logger be set up
(e.g. Foxit Reader)?
7 Tips and assistance
a Was the t99 response time
achieved?
b Was the H1/G1 data logger stored
for longer than 60 hours at a
without an airtight bag?
c Was the H1/G1 data logger used
for longer than 60 hours at a
relative humidity of more than
80%?
Solution for points b and c:
> Store the instrument in a well-
ventilated location for 12 hours at
>30°C and a relative humidity of
less than 20%.
or
> Store the device for 12 hours at
20°C to 30°C and a relative
humidity of around 75%.
- Connect the testo 184 data logger to
the PC.
- Open the Explorer
- Right-click on the logger and then on
"Format".
- After successful formatting, remove
the data logger from the PC and
connect it again.
- The measurement protocol can be
opened again.
This depends on the object and
cannot be answered in general terms.
G-forces of 10 g and more may be
considered problematic (damage,
cracks, etc.).
No, unfortunately this is not possible.
For compatibility reasons, Acrobat
Reader must be used.
27
Page 28
7 Tips and assistance
If you have any questions please contact your local dealer or the
Testo Customer Service. Contact details can be found on the
internet: www.testo.com/service-contact
28
Page 29
Page 30
Testo SE & Co. KGaA
Internet: www.testo.com
Celsiusstraße 2
79822 Titisee-Neustadt
Germany
Telefon: +49 7653 681-0
E-Mail: info@testo.de
0970 1842 en 10
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.