Testo 922 Specifications

Page 1
Data sheet
testo 922
Temperature measuring instrument (2-channel)
testo 922 – For fast (differential) temperature measurement
Ideally suited for applications in the HVAC field
2-channel temperature measuring instrument with optional
wireless probes
Display of differential temperature
TopSafe, the indestructible protective cover, protects from
dirt and impact
Continuous display of min./max. values
Hold-button for freezing measurement values
Cyclic printing of measurement values, e.g. once per minute
testo 922 is a temperature measuring instrument which
is especially suited to applications in the HVAC field. The
measuring instrument records the temperature values of
two connected thermocouple probes, and shows these
simultaneously together with the differential temperature in
allows the measurement value from a further temperature
probe to be displayed wirelessly.
With the TopSafe, and the probe plugged in, the measuring
instrument has protection class IP65.
°C
The instrument continuously displays the minimum and
maximum values. The measurement values shown in the
display (current measurement value, fixed measurement
value or the minimum/maximum value) can be printed out
on site using the Testo report printer (optional). In addiition
to this, the measurement values can be cyclically printed –
the intervals can be individually set.
www.testo.com
Page 2
testo 922
Technical data
Sensor type Type K (NiCr-Ni)
testo 922
testo 922, 2 channel temperature measuring instrument T/C Type K, connection of an optional radio probe, with battery and calibration protocol
Part no. 0560 9221
Differential temperature set testo 922
testo 922, 2 channel temperature measuring instrument T/C Type K, 2 x Velcro probes for temperature measurement on pipes with max. Ø 120 mm, Handy bag for measuring instrument and probes, with battery and calibration protocol
Part no. 0563 9222
Measuring range
Accuracy ±1 digit
Resolution
General technical data
Operating temperature -20 to +50 °C
Storage temperature -40 to +70 °C
Housing material ABS
Battery type 9V block battery, 6F22
Battery life 200 h (connected probe, backlight off)
Dimensions 182 x 64 x 40 mm
Weight 171 g
-50 to +1000 °C
±(0.5 °C +0.3% of m.v.) (-40 to +900 °C) ±(0.7 °C +0.5% of m.v.) (remaining range)
0.1 °C (-50 to +199.9 °C) 1 °C (remaining range)
45 h (radio mode, backlight off) 68 h (connected probe, backlight always on) 33 h (radio mode, backlight always on)
Optional protective case TopSafe Wireless measurement with radio
probes
Simultaneous recording of temperature by two connected probes and display of differential temperature
2 probe conneections
Page 3
testo 922
Accessories
Radio module for upgrading measuring instrument with radio option
Radio module for measuring instrument, 869.85 MHz, approval for the countries: DE, FR, UK, BE, NL, ES, IT, SE, AT, DK, FI, HU, CZ, PL, GR, CH, PT, SI, MT, CY, SK, LU, EE, LT, IE, LV, NO
Printer and Accessories
Part no.Accessories for measuring instrument
0515 00259V rech. battery for instrument, instead of battery
0554 0188
0554 0190Radio module for measuring instrument, 915.00 MHz FSK, approval for USA, CA, CL
0554 0549Testo fast printer IrDA with wireless infrared interface; 1 roll thermal paper; 4 AA batteries
measurement data documentation legible for up to 10 years
Transport and Protection
TopSafe, protects from impact and dirt (incl. 2 attachment magnets)
Service case for measuring instrument, probe and accessories, dimensions 454 x 316 x 111 mm
Service case for measuring instrument and probe, dimensions 454 x 316 x 111 mm
Case for measuring instrument and probes
Other features
Calibration Certificates
for air/immersion probes, calibration points -18°C; 0°C; +60°C
meas. instr. with air/immersion probe; cal. points 0°C; +150°C; +300°C
meas. instr. with air/immersion probe; calibration points 0°C; +300°C; +600°C
meas. instr. with surface probe; calibration points +60°C; +120°C; +180°C
meas. instr. with air/immersion probe; calibration points -20 °C; 0 °C; +60 °C
contact surface temperature probes; calibration points +100°C; +200°C; +300°C
0554 0568Spare thermal paper for printer (6 rolls), permanent ink
0516 0222
0516 1200
0516 1201
0516 0191
0409 1092Handle for attachable measurement tips, applicable for all Testo probes with miniature thermocouple plugs
0554 0592Extension cable, 5m, for thermocouple probe Type K
0554 0004Silicone heat paste (14g), Tmax = +260°C, improves heat transfer in surface probes
0520 0001ISO calibration certificate/temperature
0520 0021ISO calibration certificate/temperature (Applies only to immersion/penetration probe 0602 2693)
0520 0031ISO calibration certificate/temperature
0520 0071ISO calibration certificate/temperature
0520 0211DAkkS calibration certificate/temperature
0520 0271DAkkS calibration certificate/temperature
Page 4
testo 922
Radio probes
Radio handles and probe head for air-/ immersion-penetration-meas.
Radio handle for plug-in probe heads, incl. T/C adapter, approval for the countries: DE, FR, UK, BE, NL, ES, IT, SE, AT, DK, FI, HU, CZ, PL, GR, CH, PT, SI, MT, CY, SK, LU, EE, LT, IE, LV, NO; Radio freq. 869.85 MHz FSK
T/C probe head for air/immersion/penetration measurement (T/C Type K)
Radio handle for plug-in probe heads, incl. T/C adapter, approval for USA, CA, CL; Radio freq. 915.00 MHz FSK
T/C probe head for air/immersion/penetration measurement (T/C Type K)
Dimensions Probe shaft/probe shaft tip
100 mm
Ø 5 mm
30 mm
Ø 3,4 mm
Measuring range
-50 to +350 °C Short-term to +500 °C
Accuracy
Radio handle: ±(0.5 °C +0.3% of m.v.) (-40 to +500 °C) ±(0.7 °C +0.5% of m.v.) (remaining range) T/C probe head: Class 2
Resolution
0.1 °C (-50 to +199.9 °C)
1.0 °C (remaining range)
t99
(in
t
99
water) 10 s
Radio handles and probe head for surface measurement
Radio handle for plug-in probe heads, incl. T/C adapter, approval for the countries: DE, FR, UK, BE, NL, ES, IT, SE, AT, DK, FI, HU, CZ, PL, GR, CH, PT, SI, MT, CY, SK, LU, EE, LT, IE, LV, NO; Radio freq. 869.85 MHz FSK
T/C probe head for surface measurement (T/C Type K)
Radio handle for plug-in probe heads, incl. T/C adapter, approval for USA, CA, CL; Radio freq. 915.00 MHz FSK
T/C probe head for surface measurement (T/C Type K)
Dimensions Probe shaft/probe shaft tip
120 mm
Ø 5 mm
40 mm
Ø 12 mm
Measuring range
-50 to +350 °C Short-term to +500 °C
Accuracy
Radio handle: ±(0.5 °C +0.3% of m.v.) (-40 to +500 °C) ±(0.7 °C +0.5% of m.v.) (remaining range) T/C probe head: Class 2
Resolution
0.1 °C (-50 to +199.9 °C)
1.0 °C (remaining range)
t
99
5 s
Part no.
0554 0189
0602 0293
0554 0191
0602 0293
Part no.
0554 0189
0602 0394
0554 0191
0602 0394
Radio handles for attachable T/C probes
Radio handle for plug-in probe heads, incl. T/C adapter, approval for the countries: DE, FR, UK, BE, NL, ES, IT, SE, AT, DK, FI, HU, CZ, PL, GR, CH, PT, SI, MT, CY, SK, LU, EE, LT, IE, LV, NO ; Radio freq. 869.85 MHz FSK
Radio handle for plug-in probe heads, incl. T/C adapter, approval for USA, CA, CL; Radio freq. 915.00 MHz FSK
Illustration
Measuring
Accuracy
Resolution
range
-50 to +1000 °C
±(0.7 °C +0.3% of m.v.) (-40 to +900 °C) ±(0.9 °C +0.5% of m.v.) (remaining range)
0.1 °C (-50 to +199.9 °C)
1.0 °C (remaining range)
Technical data Radio probes
Radio immersion/penetration probe, NTC
Battery type 2 x 3V button cell (CR 2032)
Common Technical Data
Measuring rate
0.5 s or 10 s, adjustable on handle
Battery life 150 h (meas. rate 0.5 s)
2 months (meas. rate 10 s)
Radio handle
Battery type 2 x 3V button cell (CR 2032)
Battery life 215 h (meas. rate 0.5 s)
Radio coverage
Radio transmission
Operating temperature
Up to 20 m (without obstructions)
Unidirectional
-20 to +50 °C
Storage temperature -40 to +70 °C
6 months (meas. rate 10 s)
Part no.
0554 0189
0554 0191
Page 5
testo 922
Probes
Probe type
Air probes
Robust air probe, T/C Type K, Fixed cable 1.2 m
Immers./penetr. probes
Efficient and fast-action immersion probe, waterproof, TC Type K, Fixed cable 1.2 m
Fast-action, waterproof immersion/ penetration probe, TC Type K, Fixed cable 1.2 m
Immersion tip, flexible, TC Type K
Immersion measurement tip, flexible, for measurements in air/ exhaust gases (not suitable for measurements in smelters), TC Type K
Immersion tip, flexible, TC Type K
Dimensions Probe shaft/probe shaft tip
115 mm
Ø 4 mm
Ø 1.5 mm
Ø 1.5 mm
Ø 3 mm
Ø 1.5 mm
300 mm
60 mm 14 mm
Ø 5 mm Ø 1.5 mm
500 mm
1000 mm
500 mm
Measuring range
-60 to +400 °C
-60 to +1000 °C
-60 to +800 °C
-200 to +1000 °C
-200 to +1300 °C
-200 to +40 °C
Accuracy
1)
Class 2
1)
Class 1
1)
Class 1
1)
Class 1
1)
Class 1
1)
Class 3
t
99
200 s
2 s
3 s
5 s
4 s
5 s
Part no.
0602 1793
0602 0593
0602 2693
0602 5792
0602 5693
0602 5793
1)
Waterproof immersion/penetration probe, TC Type K, Fixed cable 1.2 m
114 mm 50 mm
Ø 5 mm Ø 3.7 mm
-60 to +400 °C
Class 2
7 s
Surface probes
Fast-reaction paddle surface probe, for measurements in inaccessible places, e.g. narrow apertures and slots, TC Type K,
145 mm
Ø 8 mm
40 mm
Fixed cable
Fast-action surface probe with sprung thermocouple strip, also for uneven surfaces, measurement range short-term to +500°C, TC Type K, Fixed cable 1.2 m
Waterproof surface probe with widened measurement tip for flat surfaces, T/C Type K, Fixed cable
1.2 m
The measuring instrument inside TopSafe is waterproof with this probe.
1) According to standard EN 60584-2, the accuracy of Class 1 refers to -40 to +1000 °C (Type K), Class 2 to -40 to +1200 °C (Type K), Class 3 to -200 to +40 °C (Type K). A probe always corresponds to only one accuracy class.
115 mm
Ø 5 mm
115 mm
Ø 5 mm Ø 6 mm
Ø 12 mm
0 to +300 °C
Ø 7 mm
-60 to +300 °C
-60 to +400 °C
Class 2
Class 2
Class 2
1)
1)
1)
5 s
3 s
30 s
0602 1293
0602 0193
0602 0393
0602 1993
Page 6
testo 922
Probes
Probe type
Surface probes
Fast-action surface probe with sprung thermocouple strip, bent, also for uneven surfaces, measurement range short-term to +500°C, TC Type K, Fixed cable
1.2 m
Efficient, waterproof surface probe with small measurement head for flat surfaces, TC Type K, Fixed cable 1.2 m
Flat head surface probe with telescopic handle max. 680 mm for measurements at hard-to-access points, TC Type K, Fixed cable 1.6 m (correspondingly shorter when telescope extended)
Magnetic probe, adhesive force approx. 20 N, with magnets, for measurements on metal surfaces, TC Type K, Fixed cable 1.6 m
Dimensions Probe shaft/probe shaft tip
80 mm 50 mm
Ø 5 mm
150 mm
Ø 2.5 mm Ø 4 mm
985 ±5 mm 12 mm
35 mm
Ø 20 mm
Ø 12 mm
Ø 25 mm
Measuring range
-60 to +300 °C
-60 to +1000 °C
-50 to +250 °C
-50 to +170 °C
Accuracy
1)
Class 2
1)
Class 1
1)
Class 2
1)
Class 2
t
99
3 s
20 s
3 s
150 s
Part no.
0602 0993
0602 0693
0602 2394
0602 4792
1)
Magnetic probe, adhesive force approx. 10 N, with magnets, for higher temp., for measurements on
75 mm
-50 to +400 °C
Ø 21 mm
Class 2
metal surfaces, TC Type K, Fixed cable
1)
Pipe wrap probe with velcro strip; for temperature measurement on pipes with diameter up to max. 120
395 mm
-50 to +120 °C
20 mm
Class 1
90 s
mm; Tmax. +120 °C; TC Type K , Fixed cable
1)
Pipe wrap probe for pipe diameter
-60 to +130 °C
Class 2
5 s
5 to 65 mm, with exchangeable measuring head. Meas. range short-term up to +280 °C, TC Type K, Fixed cable
1)
Spare meas. head for pipe wrap probe, TC Type K
Clamp probe for measurements on
35 mm
-60 to +130 °C
15 mm
-50 to +100 °C
Class 2
Class 2
5 s
1)
5 s
pipes, pipe diameter 15 to 25 mm (max. 1“), meas. range short-term up to +130°C, TC Type K, Fixed cable
The measuring instrument inside TopSafe is waterproof with this probe.
1) According to standard EN 60584-2, the accuracy of Class 1 refers to -40 to +1000 °C (Type K), Class 2 to -40 to +1200 °C (Type K), Class 3 to -200 to +40 °C (Type K). A probe always corresponds to only one accuracy class.
0602 4892
0628 0020
0602 4592
0602 0092
0602 4692
Page 7
testo 922
Probes
Probe type
Food probes
Waterproof food probe made of stainless steel (IP65), TC Type K, Fixed cable
Waterproof robust immersion/ penetration probe with metal protection hose Tmax +230°C, e.g. for monitoring temp. in cooking oil, T/C Type K, Fixed cable
Thermocouples
Thermocouple with TC adapter, flexible, 800 mm long, fibre glass, TC Type K
Thermocouple with TC adapter, flexible, length 1500 mm , fibreglass, TC Type K
Thermocouple with TC adapter, flexible, 1500 mm long, PTFE, TC Type K
Dimensions Probe shaft/probe shaft tip
125 mm 30 mm
Ø 4 mm Ø 3.2 mm
240 mm
Ø 4 mm
800 mm
Ø 1.5 mm
1500 mm
Ø 1.5 mm
1500 mm
Ø 1.5 mm
Measuring range
-60 to +400 °C
-50 to +230 °C
-50 to +400 °C
-50 to +400 °C
-50 to +250 °C
Accuracy
1)
Class 2
1)
Class 1
1)
Class 2
1)
Class 2
1)
Class 2
t
99
7 s
15 s
5 s
5 s
5 s
Part no.
0602 2292
0628 1292
0602 0644
0602 0645
0602 0646
The measuring instrument inside TopSafe is waterproof with this probe.
1) According to standard EN 60584-2, the accuracy of Class 1 refers to -40 to +1000 °C (Type K), Class 2 to -40 to +1200 °C (Type K), Class 3 to -200 to +40 °C (Type K). A probe always corresponds to only one accuracy class.
Information on surface measurement:
• The response times t
stated are measured on ground steel or aluminium plates at +60 °C.
99
• The stated accuracies are sensor accuracies.
• The accuracy in your application is dependent on the surface structure (roughness), material of the measurement
object (heat capacity and heat transfer), as wel as sensor accuracy. Testo creates a corresponding calibration
certificate for the deviations of your measurement system in your application. For this purpose, Testo uses a surface
test bench developed in cooperation with the PTB (Physikalisch Technische Bundesanstalt).
Page 8
testo 922
0981 9294/msp/I/05.2018
www.testo.com
Subject to change without notice.
Loading...