value or the minimum/maximum value) can be printed out
on site using the Testo report printer (optional). In addiition
to this, the measurement values can be cyclically printed –
the intervals can be individually set.
www.testo.com
Page 2
testo 922
Technical data
Sensor typeType K (NiCr-Ni)
testo 922
testo 922, 2 channel temperature
measuring instrument T/C Type K,
connection of an optional radio probe, with
battery and calibration protocol
Part no. 0560 9221
Differential temperature set
testo 922
testo 922, 2 channel temperature measuring instrument T/C Type K,
2 x Velcro probes for temperature measurement on pipes with
max. Ø 120 mm, Handy bag for measuring instrument and probes,
with battery and calibration protocol
Part no. 0563 9222
Measuring range
Accuracy
±1 digit
Resolution
General technical data
Operating temperature-20 to +50 °C
Storage temperature-40 to +70 °C
Housing materialABS
Battery type9V block battery, 6F22
Battery life200 h (connected probe, backlight off)
Dimensions182 x 64 x 40 mm
Weight171 g
-50 to +1000 °C
±(0.5 °C +0.3% of m.v.) (-40 to +900 °C)
±(0.7 °C +0.5% of m.v.) (remaining range)
0.1 °C (-50 to +199.9 °C)
1 °C (remaining range)
45 h (radio mode, backlight off)
68 h (connected probe, backlight always on)
33 h (radio mode, backlight always on)
Optional protective case TopSafeWireless measurement with radio
probes
Simultaneous recording of
temperature by two connected
probes and display of differential
temperature
2 probe conneections
Page 3
testo 922
Accessories
Radio module for upgrading measuring instrument with radio option
Radio module for measuring instrument, 869.85 MHz, approval for the countries: DE, FR, UK, BE, NL, ES, IT, SE, AT,
DK, FI, HU, CZ, PL, GR, CH, PT, SI, MT, CY, SK, LU, EE, LT, IE, LV, NO
Printer and Accessories
Part no.Accessories for measuring instrument
0515 00259V rech. battery for instrument, instead of battery
0554 0188
0554 0190Radio module for measuring instrument, 915.00 MHz FSK, approval for USA, CA, CL
0554 0549Testo fast printer IrDA with wireless infrared interface; 1 roll thermal paper; 4 AA batteries
measurement data documentation legible for up to 10 years
Transport and Protection
TopSafe, protects from impact and dirt (incl. 2 attachment magnets)
Service case for measuring instrument, probe and accessories, dimensions 454 x 316 x 111 mm
Service case for measuring instrument and probe, dimensions 454 x 316 x 111 mm
Case for measuring instrument and probes
Other features
Calibration Certificates
for air/immersion probes, calibration points -18°C; 0°C; +60°C
meas. instr. with air/immersion probe; cal. points 0°C; +150°C; +300°C
meas. instr. with air/immersion probe; calibration points 0°C; +300°C; +600°C
meas. instr. with surface probe; calibration points +60°C; +120°C; +180°C
meas. instr. with air/immersion probe; calibration points -20 °C; 0 °C; +60 °C
contact surface temperature probes; calibration points +100°C; +200°C; +300°C
0554 0568Spare thermal paper for printer (6 rolls), permanent ink
0516 0222
0516 1200
0516 1201
0516 0191
0409 1092Handle for attachable measurement tips, applicable for all Testo probes with miniature thermocouple plugs
0554 0592Extension cable, 5m, for thermocouple probe Type K
0554 0004Silicone heat paste (14g), Tmax = +260°C, improves heat transfer in surface probes
0520 0001ISO calibration certificate/temperature
0520 0021ISO calibration certificate/temperature (Applies only to immersion/penetration probe 0602 2693)
Radio handles and probe head for air-/ immersion-penetration-meas.
Radio handle for plug-in probe heads, incl. T/C adapter, approval for the countries: DE, FR, UK, BE, NL, ES, IT, SE, AT,
DK, FI, HU, CZ, PL, GR, CH, PT, SI, MT, CY, SK, LU, EE, LT, IE, LV, NO; Radio freq. 869.85 MHz FSK
T/C probe head for air/immersion/penetration measurement (T/C Type K)
Radio handle for plug-in probe heads, incl. T/C adapter, approval for USA, CA, CL; Radio freq. 915.00 MHz FSK
T/C probe head for air/immersion/penetration measurement (T/C Type K)
Dimensions
Probe shaft/probe shaft tip
100 mm
Ø 5 mm
30 mm
Ø 3,4 mm
Measuring
range
-50 to +350 °C
Short-term to
+500 °C
Accuracy
Radio handle:
±(0.5 °C +0.3% of m.v.) (-40 to +500 °C)
±(0.7 °C +0.5% of m.v.) (remaining range)
T/C probe head: Class 2
Resolution
0.1 °C (-50 to
+199.9 °C)
1.0 °C (remaining
range)
t99
(in
t
99
water)
10 s
Radio handles and probe head for surface measurement
Radio handle for plug-in probe heads, incl. T/C adapter, approval for the countries: DE, FR, UK, BE, NL, ES, IT, SE, AT,
DK, FI, HU, CZ, PL, GR, CH, PT, SI, MT, CY, SK, LU, EE, LT, IE, LV, NO; Radio freq. 869.85 MHz FSK
T/C probe head for surface measurement (T/C Type K)
Radio handle for plug-in probe heads, incl. T/C adapter, approval for USA, CA, CL; Radio freq. 915.00 MHz FSK
T/C probe head for surface measurement (T/C Type K)
Dimensions
Probe shaft/probe shaft tip
120 mm
Ø 5 mm
40 mm
Ø 12 mm
Measuring
range
-50 to +350 °C
Short-term to
+500 °C
Accuracy
Radio handle:
±(0.5 °C +0.3% of m.v.) (-40 to +500 °C)
±(0.7 °C +0.5% of m.v.) (remaining range)
T/C probe head: Class 2
Resolution
0.1 °C (-50 to
+199.9 °C)
1.0 °C (remaining
range)
t
99
5 s
Part no.
0554 0189
0602 0293
0554 0191
0602 0293
Part no.
0554 0189
0602 0394
0554 0191
0602 0394
Radio handles for attachable T/C probes
Radio handle for plug-in probe heads, incl. T/C adapter, approval for the countries: DE, FR, UK, BE, NL, ES, IT, SE, AT,
DK, FI, HU, CZ, PL, GR, CH, PT, SI, MT, CY, SK, LU, EE, LT, IE, LV, NO ; Radio freq. 869.85 MHz FSK
Radio handle for plug-in probe heads, incl. T/C adapter, approval for USA, CA, CL; Radio freq. 915.00 MHz FSK
Illustration
Measuring
Accuracy
Resolution
range
-50 to +1000 °C
±(0.7 °C +0.3% of m.v.) (-40 to +900 °C)
±(0.9 °C +0.5% of m.v.) (remaining range)
0.1 °C (-50 to +199.9 °C)
1.0 °C (remaining range)
Technical data Radio probes
Radio immersion/penetration probe, NTC
Battery type2 x 3V button cell (CR 2032)
Common Technical Data
Measuring rate
0.5 s or 10 s, adjustable on handle
Battery life150 h (meas. rate 0.5 s)
2 months (meas. rate 10 s)
Radio handle
Battery type2 x 3V button cell (CR 2032)
Battery life215 h (meas. rate 0.5 s)
Radio coverage
Radio transmission
Operating temperature
Up to 20 m (without obstructions)
Unidirectional
-20 to +50 °C
Storage temperature-40 to +70 °C
6 months (meas. rate 10 s)
Part no.
0554 0189
0554 0191
Page 5
testo 922
Probes
Probe type
Air probes
Robust air probe, T/C Type K,
Fixed cable 1.2 m
Immers./penetr. probes
Efficient and fast-action immersion
probe, waterproof, TC Type K,
Fixed cable 1.2 m
Fast-action, waterproof immersion/
penetration probe, TC Type K,
Fixed cable 1.2 m
Immersion tip, flexible, TC Type K
Immersion measurement tip,
flexible, for measurements in air/
exhaust gases (not suitable for
measurements in smelters), TC
Type K
Immersion tip, flexible, TC Type K
Dimensions
Probe shaft/probe shaft tip
115 mm
Ø 4 mm
Ø 1.5 mm
Ø 1.5 mm
Ø 3 mm
Ø 1.5 mm
300 mm
60 mm14 mm
Ø 5 mmØ 1.5 mm
500 mm
1000 mm
500 mm
Measuring
range
-60 to +400 °C
-60 to +1000 °C
-60 to +800 °C
-200 to
+1000 °C
-200 to
+1300 °C
-200 to +40 °C
Accuracy
1)
Class 2
1)
Class 1
1)
Class 1
1)
Class 1
1)
Class 1
1)
Class 3
t
99
200 s
2 s
3 s
5 s
4 s
5 s
Part no.
0602 1793
0602 0593
0602 2693
0602 5792
0602 5693
0602 5793
1)
Waterproof immersion/penetration
probe, TC Type K, Fixed cable 1.2
m
114 mm50 mm
Ø 5 mmØ 3.7 mm
-60 to +400 °C
Class 2
7 s
Surface probes
Fast-reaction paddle surface
probe, for measurements in
inaccessible places, e.g. narrow
apertures and slots, TC Type K,
145 mm
Ø 8 mm
40 mm
Fixed cable
Fast-action surface probe with
sprung thermocouple strip, also
for uneven surfaces, measurement
range short-term to +500°C, TC
Type K, Fixed cable 1.2 m
Waterproof surface probe with
widened measurement tip for flat
surfaces, T/C Type K, Fixed cable
1.2 m
The measuring instrument inside TopSafe is waterproof with this probe.
1) According to standard EN 60584-2, the accuracy of Class 1 refers to -40 to +1000 °C (Type K), Class 2 to -40 to +1200 °C (Type K), Class 3 to -200 to +40 °C (Type K).
A probe always corresponds to only one accuracy class.
115 mm
Ø 5 mm
115 mm
Ø 5 mmØ 6 mm
Ø 12 mm
0 to +300 °C
Ø 7 mm
-60 to +300 °C
-60 to +400 °C
Class 2
Class 2
Class 2
1)
1)
1)
5 s
3 s
30 s
0602 1293
0602 0193
0602 0393
0602 1993
Page 6
testo 922
Probes
Probe type
Surface probes
Fast-action surface probe with
sprung thermocouple strip,
bent, also for uneven surfaces,
measurement range short-term to
+500°C, TC Type K, Fixed cable
1.2 m
Efficient, waterproof surface probe
with small measurement head for
flat surfaces,
TC Type K, Fixed cable 1.2 m
Flat head surface probe with
telescopic handle max. 680 mm for
measurements at hard-to-access
points, TC Type K, Fixed cable 1.6
m (correspondingly shorter when
telescope extended)
Magnetic probe, adhesive force
approx. 20 N, with magnets, for
measurements on metal surfaces,
TC Type K, Fixed cable 1.6 m
Dimensions
Probe shaft/probe shaft tip
80 mm50 mm
Ø 5 mm
150 mm
Ø 2.5 mmØ 4 mm
985 ±5 mm12 mm
35 mm
Ø 20 mm
Ø 12 mm
Ø 25 mm
Measuring
range
-60 to +300 °C
-60 to +1000 °C
-50 to +250 °C
-50 to +170 °C
Accuracy
1)
Class 2
1)
Class 1
1)
Class 2
1)
Class 2
t
99
3 s
20 s
3 s
150 s
Part no.
0602 0993
0602 0693
0602 2394
0602 4792
1)
Magnetic probe, adhesive force
approx. 10 N, with magnets, for
higher temp., for measurements on
75 mm
-50 to +400 °C
Ø 21 mm
Class 2
metal surfaces, TC Type K, Fixed
cable
1)
Pipe wrap probe with velcro strip;
for temperature measurement on
pipes with diameter up to max. 120
395 mm
-50 to +120 °C
20 mm
Class 1
90 s
mm; Tmax. +120 °C; TC Type K ,
Fixed cable
1)
Pipe wrap probe for pipe diameter
-60 to +130 °C
Class 2
5 s
5 to 65 mm, with exchangeable
measuring head. Meas. range
short-term up to +280 °C, TC Type
K, Fixed cable
1)
Spare meas. head for pipe wrap
probe, TC Type K
Clamp probe for measurements on
35 mm
-60 to +130 °C
15 mm
-50 to +100 °C
Class 2
Class 2
5 s
1)
5 s
pipes, pipe diameter 15 to 25 mm
(max. 1“), meas. range short-term
up to +130°C, TC Type K, Fixed
cable
The measuring instrument inside TopSafe is waterproof with this probe.
1) According to standard EN 60584-2, the accuracy of Class 1 refers to -40 to +1000 °C (Type K), Class 2 to -40 to +1200 °C (Type K), Class 3 to -200 to +40 °C (Type K).
A probe always corresponds to only one accuracy class.
0602 4892
0628 0020
0602 4592
0602 0092
0602 4692
Page 7
testo 922
Probes
Probe type
Food probes
Waterproof food probe made of
stainless steel (IP65), TC Type K,
Fixed cable
Waterproof robust immersion/
penetration probe with metal
protection hose Tmax +230°C, e.g.
for monitoring temp. in cooking oil,
T/C Type K, Fixed cable
Thermocouples
Thermocouple with TC adapter,
flexible, 800 mm long, fibre glass,
TC Type K
Thermocouple with TC adapter,
flexible, length 1500 mm ,
fibreglass, TC Type K
Thermocouple with TC adapter,
flexible, 1500 mm long, PTFE, TC
Type K
Dimensions
Probe shaft/probe shaft tip
125 mm30 mm
Ø 4 mmØ 3.2 mm
240 mm
Ø 4 mm
800 mm
Ø 1.5 mm
1500 mm
Ø 1.5 mm
1500 mm
Ø 1.5 mm
Measuring
range
-60 to +400 °C
-50 to +230 °C
-50 to +400 °C
-50 to +400 °C
-50 to +250 °C
Accuracy
1)
Class 2
1)
Class 1
1)
Class 2
1)
Class 2
1)
Class 2
t
99
7 s
15 s
5 s
5 s
5 s
Part no.
0602 2292
0628 1292
0602 0644
0602 0645
0602 0646
The measuring instrument inside TopSafe is waterproof with this probe.
1) According to standard EN 60584-2, the accuracy of Class 1 refers to -40 to +1000 °C (Type K), Class 2 to -40 to +1200 °C (Type K), Class 3 to -200 to +40 °C (Type K).
A probe always corresponds to only one accuracy class.
Information on surface measurement:
• The response times t
stated are measured on ground steel or aluminium plates at +60 °C.
99
• The stated accuracies are sensor accuracies.
• The accuracy in your application is dependent on the surface structure (roughness), material of the measurement
object (heat capacity and heat transfer), as wel as sensor accuracy. Testo creates a corresponding calibration
certificate for the deviations of your measurement system in your application. For this purpose, Testo uses a surface
test bench developed in cooperation with the PTB (Physikalisch Technische Bundesanstalt).
Page 8
testo 922
0981 9294/msp/I/05.2018
www.testo.com
Subject to change without notice.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.