Testo 875-1 Operating Manual

testo 875 · Thermal imager
Instruction manual
2
1 Contents
1 Contents
1 Contents ...................................................................................................3
2 Safety and the environment....................................................................4
2.1. About this document........................................................................4
2.2. Ensure safety...................................................................................5
2.3. Protecting the environment..............................................................6
3 Specifications ..........................................................................................7
3.1. Use ..................................................................................................7
3.2. Technical data .................................................................................8
4 Product description...............................................................................12
4.1. Overview........................................................................................12
4.2. Basic properties.............................................................................13
5 First steps ..............................................................................................14
5.1. Commissioning ..............................................................................14
5.2. Getting to know the product...........................................................16
6 Using the product..................................................................................21
6.1. Menu functions ..............................................................................21
6.1.1. Measuring functions.......................................................................................21
6.1.2. Image gallery .................................................................................................22
6.1.3. Scale..............................................................................................................24
6.1.4. Display........................................................................................................... 24
6.1.5. Emissivity.......................................................................................................25
6.1.6. Palette ...........................................................................................................28
6.1.7. Configuration .................................................................................................28
6.2. Measuring......................................................................................30
7 Maintaining the product........................................................................32
8 Tips and assistance...............................................................................34
8.1. Questions and answers .................................................................34
8.2. Accessories and spare parts .........................................................35
2 Safety and the environment
2 Safety and the environment
2.1. About this document
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay particular attention to the safety instructions and warning advice in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.
Warnings
Always pay attention to information that is marked by the following warnings with warning pictograms. Implement the specified precautionary measures.
Representation Explanation
CAUTION
indicates circumstances that may lead to damage to the products
Symbols and writing standards Represen-
Explanation
tation
Note: Basic or further information.
1. ...
Action: more steps, the sequence must be followed.
2. ...
> ... Action: a step or an optional step.
- ... Result of an action.
Menu Elements of the instrument, the instrument displays
or the program interface.
[OK] Control keys of the instrument or buttons of the
program interface.
... | ... Functions/paths within a menu.
“...” Example entries
4
2.2. Ensure safety
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not use any force.
> Do not operate the instrument if there are signs of damage at
the housing, power supply or feed lines.
The thermal imager must not be directed at the sun or other sources of intense radiation (e.g. objects with temperatures greater than 500 °C) during operation. This may lead to serious damage to the detector. The manufacturer accepts no warranty for this type of damage to the detector.
> The objects to be measured or the measurement environment
may also pose risks: Note the safety regulations valid in your area when performing the measurements.
> Do not store the product together with solvents. Do not use any
desiccants.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the prescribed steps exactly. Use only original spare parts from Testo.
> Improper use of rechargeable batteries can lead to destruction
or injuries by means of current surges, fire or escaping chemicals. The following instructions must be observed to avoid such hazards:
• Only use in accordance with the directions in the instruction manual.
• Do not short, take apart or modify.
• Do not expose to heavy impacts, water, fire or temperatures above 60 °C.
• Do not store in the proximity of metal objects.
• Do not use leaky or damaged rechargeable batteries. In the event of contact with battery acid: Thoroughly wash affected area with water and consult a doctor, if necessary.
• Only charge in the instrument or the recommended charging station.
• Immediately stop the charging process if this is not completed in the given time.
2 Safety and the environment
2 Safety and the environment
• In the event of improper function or signs of overheating,
immediately remove the rechargeable battery from the measuring instrument/charging station. Caution:
2.3. Protecting the environment
Rechargeable battery may be hot!
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local regulations) or return the product to Testo for disposal.
6
3 Specifications
3.1. Use
The testo 875 is a handy and robust thermal imager. It opens the door to contactless determination and illustration of the temperature distribution on surfaces for you.
Typical areas of application are:
• Building inspection (heating, ventilation and air conditioning trade, company technicians, engineering offices, experts): Energy evaluation of buildings
• Preventative maintenance (servicing): Mechanical and electrical inspection of systems and machines
• Production monitoring (quality assurance): Monitoring production processes
The testo 875 is available in several versions, which are tailored to various usage requirements:
• testo 875-1: High-quality wide-angle lens 32° x 23, detector 160 x 120, manual focus, 2 GB SD card for approx. 1000 images, minimum focus distance 10 cm
• testo 875-2: High-quality wide-angle lens 32° x 23, detector 160 x 120, manual focus, 2 GB SD card for approx. 1000 images, minimum focus distance 10 cm, integrated digital camera, exchangeable lens, display of surface moisture distribution
3 Specifications
Export control
Thermal imagers may be subject to the export restrictions of the European Union.
Please observe the national regulations for export control when exporting.
3 Specifications
3.2. Technical data
Infrared image output Characteristic Values
Field of vision/min. focusing distance
Standard lens: 32° x 23°/0.1 m (0.33 ft)
testo 875-2 only (optional): Telephoto lens: 9° x 7°/0.5 m (1.64 ft)
Thermal sensitivity (NETD)
< 100 mK typically, < 110 mK at 30 °C (86 °F)
Geometric resolution Standard lens: 3.3 mrad
Telephoto lens: 1 mrad
Refresh rate 9 Hz
Focus manual
Detector type FPA 160 x 120 pixels, a.Si
Spectral range 8 - 14 μm
Visual image output (testo 875-2 only) Characteristic Values
Field of vision/min.
33° x 25°/0.4 m (1.31 ft)
focusing distance
Image size 640 x 480 pixels
Refresh rate 8 to 15 Hz
Image presentation Characteristic Values
Image display 3.5" LCD with 320 x 240 pixels
Display options testo 875-1: Infrared image
testo 880-2: Infrared image/real image/infrared and real image
Video output USB 2.0
Video stream 9 Hz
Color palettes 4 options
8
3 Specifications
Measuring Characteristic Values
Temperature range (can be changed)
Accuracy
-20 - 100 °C/0 - 280°C (-4 - 212 °F/32 ­536°F)
±2 °C (±3.6 °F) or ±2 % of reading (whichever is the greater)
Reproducibility
±1 °C (±1.8 °F) or ±1 % (whichever is the greater)
Measuring point minimum diameter
On-time (time to
Standard lens: 10 mm at 1 m
Telephoto lens: 3 mm at 1 m
30 s
image)
Measuring functions
Standard measurement (1-point), Cold-/ Hotspot
testo 875-2: Display of the surface moisture distribution via manual entry of the dewpoint
Compensation for
manual
reflected temperature
Setting emissivity 0.01 - 1.00
Image storage Characteristic Values
File format .bmt
Export options in .bmp, .jpg, .csv
Removable memory SD card
Memory capacity
Scope of delivery: 2 GB (approx. 1000 images)
Lens Characteristic Values
Standard lens 32° x 23°
Telephoto lens (testo
9° x 7°
872-2 optional)
Aperture 1.0
3 Specifications
Power supply Characteristic Values
Battery type
Fast-charging, Li-ion battery can be changed on-site
Operating time approx. 4 h at 20 - 30 °C (68 - 86 °F)
Charging option In instrument/in charging station (optional)
Power supply
Yes, with power supply 0554 8808
operation
Power supply output
5 V/4 A
voltage
Ambient conditions Characteristic Values
Operating
-15 - 40 °C (5 - 113 °F)
temperature
Storage
-30 - 60 °C (-22 - 140 °F)
temperature
Air humidity 20 - 80 %RH, not condensing
Enclosure rating of the housing
Physical features
IP54 (Interface terminals closed, battery connected, lens mounted)
Characteristic Values
Weight 900 g (including battery)
Dimensions 152 x 108 x 262 mm (5.98 x 4.17 x 10.31")
Tripod mounting Yes, with adapter (included in delivery)
Housing ABS
Protection class IP 54
Vibration max. 2 G
10
3 Specifications
PC software Characteristic Values
System requirements
Operating system Windows XP Service Pack 2 or Windows Vista
USB 2.0 interface
Standards, tests, warranty Characteristic Values
EU Directive 2004/108/EC
Vibration
Warranty
IEC 60068-2-6
2 years, warranty conditions: see website www.testo.com/warranty
11
4 Product description
4 Product description
4.1. Overview
Product components
1 Display.
2 Control keys:
Key Functions
[ ] [OK] and Joystick
Switch the imager on/off.
• Press [OK]:
Open menu, confirm selection/setting.
• Move [OK] up/down/right/left = Joystick
function: Select functions, navigate
[Esc]
Left/right ["xy"] quick select button
Cancel action.
Call up a function. The function assigned to the quick select button at any one time is shown in the display.
3 Instrument rechargeable battery lock release button.
12
4 Metric thread: For fastening the included tripod adapter.
Do not use desktop tripods, danger of tilting!
5 Right interface terminal: Not assigned
6 Digital camera lens: For recording visual images (testo 875-2
only)
7 Infrared camera lens: For recording thermography images.
Exchangeable lens (testo 875-2 only).
8 Rotating ring Manual focusing: For focusing manually. 9 Lens lock ring 10 [Trigger]: For recording (freezing/saving) images.
11 Left interface terminal: Memory card slot. USB interface. Power
supply socket, for connecting the included power supply. Backup battery slot.
12 Status LED: off (no power supply connected), flashes (power
supply connected and rechargeable battery charging), lights up (power supply connected and battery charging process
4.2. Basic properties
completed).
Power supply
The power is supplied to the instrument via an exchangeable rechargeable battery or via the included power supply.
With the power supply connected, the power is supplied automatically and the instrument's rechargeable battery is charged (only at ambient temperatures from 0 to 45 °C).
Charging the battery is also possible using a desktop charging station (accessory).
The instrument is equipped with a backup battery (type CR 1632) to maintain the system data during an interruption in the power supply (e.g. when the battery is changed).
4 Product description
File formats and file names
The images are saved according to the following pattern: XX_YYYYY.ZZZ
XX: IR for infrared image without attached real image, IV for infrared image with attached real image, VI for real image.
YYYYY: 5-digit consecutive number ZZZ (file extension): BMT for infrared image without/with attached
real image, BMP for real image.
13
5 First steps
5 First steps
5.1. Commissioning
Connect rechargeable battery
The imager is delivered with a rechargeable battery that is inserted in the rechargeable battery slot but not connected.
> Push the battery all the way into the rechargeable battery slot
until this is flush with the bottom of the handle.
-
T h e thermal imager starts automatically.
Perform basic settings
> Remove the protective film from the display.
- The start screen appears on the display.
- The Landeseinstellungen (Country settings) dialogue is
opened.
Instrument language (Language) and temperature unit (Unit) can be set.
1. Move Joystick up/down to select the desired option.
- The selected option has an orange border.
2. Confirm the selection with [OK].
14
- The selected setting is highlighted orange ( ). The arrows ( )
show that the setting can be changed.
3. Move Joystick up/down to change the setting.
4. Confirm the entry with [OK].
5. Confirm the settings with Übernehmen (Apply).
6. Press [ ] to switch off the thermal imager.
5 First steps
Initial charging of rechargeable battery
The thermal imager is delivered with a partially charged battery. Completely charge the battery before use.
> Connect the country-specific adapter required for the existing
mains to the power supply.
1. Open the cover on the left side of the thermal imager (1).
2. Connect the power supply cord to the camera (
) (2).
3. Connect the power supply unit to a wall outlet.
- The thermal imager starts automatically.
When charging the rechargeable battery, the imager can remain switched on or be switched off. This has no effect on the length of the charging process.
- The charging of the rechargeable battery begins.
- The charge status is indicated on the status LED:
• LED flashes: Charging in process.
• LED lights up: Battery charged, charging process finished.
4. Charge the battery fully, then disconnect the instrument from the power supply.
- After the initial charging of the battery, the thermal imager is ready for operation.
Charging the battery is also possible using a desktop charging station (accessory).
15
5 First steps
5.2. Getting to know the product
Insert the memory card
1. Open the cover on the left side of the thermal imager.
2. Insert memory card (SD card) into the card slot (SD) (1).
> To remove the memory card: Press on the memory card to
release the lock.
Mount/remove infrared protection glass
Installation:
1. Attach the protection glass fastened to the red mounting ring to
the lens and turn the mounting ring clockwise to the stop.
2. Remove the red mounting ring from the protection glass.
Removal:
1. Attach the red mounting ring to the protection glass.
2. Turn the mounting ring anticlockwise and remove the protection
glass.
After mounting/removing the protection glass, activate or deactivate the Protection glass option, Optics... page 29. If this option is set incorrectly, the stated measuring accuracy is not guaranteed.
Mount tripod adapter
Using the included tripod adapter, the imager can be fastened to the testo tripod (accessory) or to a commercially available standard tripod. With the tripod adapter mounted, the battery cannot be changed!
1. Place the tripod adapter on the lower end of the handle and
screw on using the included Allen key (ISO 2936, size 4).
2. Slide thermal imager into the fastening plate of the testo tripod
and lock or fasten it to the commercially available standard tripod (screw-type fastening).
16
5 First steps
Mount Sun Shield
The Sun Shield increases the readability of the display with a high exposure to light.
> Set the Sun Shield on the thermal imager from above (1) and
pull both sides of the Sun Shield down around the display (2).
Mounting Softcase
The Softcase combines the functions of instrument protection, Sun Shield and carrying function (strap).
1. Set the Softcase on the imager from above (1) and pull both sides of the Softcase down around the display (2).
2. Guide Velcro strip through the two plates (3) and close (4).
Change/activate lens (testo 875-2 only)
Only lenses that were adjusted to the respective thermal imager can be used. The serial number on the lens must match the serial number of the instrument. After Changing the lens, the new lens must be activated in the imager.
Attention! Always carry out the activation of the lens, otherwise the measuring accuracy of the instrument is not guaranteed.
17
5 First steps
> Switch instrument off before changing the lens.
> To keep the lens from falling out: Hold the instrument with the
lens pointing up.
1. Turn lens lock ring anticlockwise to the stop (approx. 2 cm).
2. Remove lens.
3. Insert new lens, taking care to align the white markings on the
lens and the instrument with one another.
4. Turn the lens lock ring clockwise to the stop.
Always store lenses not in use in the container designed for this in the case.
> Turn the lens ring clockwise to the stop, place the lens in the
container and close it.
5. Note or remember the serial number (Serial no.) which is
displayed on the lens.
6. Press [
].
If you have changed from a wide-angle to a telephoto lens or vice versa:
- The information No lens/wrong lens appears. > Press [OK] to confirm the information.
If you have changed between two wide-angle or two telephoto lenses:
> [OK] | Configuration… | [OK] | Optics… | [OK].
- The dialog Optics opens.
7. Move joystick up/down to select the lens used.
- The lens selected is outlined in orange.
8. Press function key [Apply].
- The information Restart instrument appears.
9. Press [
].
- The instrument switches off.
> To switch on the instrument: Press [
].
18
Switch the imager on/off
1. Remove protection cap from the lens.
2. Press [
].
- The start screen appears on the display. The type designation
of the imager and the firmware version are shown.
5 First steps
- The measurement view is opened after the warm-up period.
- The imager performs an automatic zeroing approx. every 60 sec. This can be recognized by means of a "click". The image is frozen briefly when this occurs.
> To switch off: Press [
- The display goes out and the imager is switched off.
].
Manually focusing the image
> Turn Rotating ring of the lens until the image is in focus.
Record (freeze/save) image
1. Press [Trigger].
- The image is frozen (fixed image).
If the image is to be saved, the desired storage location can be set using the left [Folder] quick select button, see Select the storage location (folder): page 31.
testo 875-1:
• Infrared image is shown: Infrared image is saved.
testo 875-2:
• Infrared image or infrared/real image is shown: Infrared image is saved, real image is saved in the same file as an attachment to the infrared image.
• Real image is shown: Real image is saved.
2. Save image: Press [Trigger] again, or: Press [OK]. or Discard image: Press [ESC].
Getting to know the quick select buttons
The quick select buttons can be assigned functions that are required often so that you can call them up directly.
Factory settings:
• Left quick select button: [Scale].
• Right quick select button: [Palette] (testo 875-1) or [Image
type] (testo 875-2).
19
5 First steps
Changing assignment of the quick select buttons
1. Move Joystick left or right to open the Configure key function
list for the left or right quick select button.
2. Move Joystick up/down to select the desired function.
Function Description
Image type
(testo 875-2 only)
Palette Emissivity... Scale... Adjustment
Change display mode: Infrared image, real image or infrared/real image. 1
Change color palette for the infrared image.
Set emissivity and reflected temperature.
Set scale limits.
Perform manual zero point calibration.
3. Press [OK] to activate the selected function.
Navigating in the menu
1. Press [OK] button.
- The Menu is opened.
2. Select navigation/function:
• Move Joystick up or down to select the menu item.
• Press [OK] to confirm the selection. Or for menu items that are marked with an arrow (): Move Joystick to the right.
• Move Joystick to the left to return to the previous menu level.
• Press [OK] to select the function.
• Press [ESC] to cancel the process and change to Measuring Mode.
- Depending on the selected menu item, press [OK] to perform a
setting or open a submenu/dialogue, see Menu functions page
21.
1
Real image and infrared lens have a different angle of view due to the spacial separation. The image areas therefore do not correspond completely (parallax error).
20
6 Using the product
6.1. Menu functions
6.1.1. Measuring functions
1-point measurement
The 1-point measurement is the standard measuring function. If
this is activated ( ), all available options can be selected via the quick select buttons.
> [OK] | Measurement | [OK] | 1-point measurement | [OK].
Cold-/Hotspot
Cold-/Hotspot shows the minimum and maximum temperature
point on the display. If Cold-/Hotspot is activated ( ), the quick select buttons with the Coldspot and Hotspot functions can be permanently assigned and can no longer be changed.
With frozen images, the crosshairs can be moved and the min./max. point is visible. The min./max. point is not saved.
With saved images, Cold-/Hotspot must be activated for the respective opened image.
> [OK] | Measurement | Cold-/Hotspot | [OK]. > Select left or right Coldspot or Hotspot quick select button to
activate/deactivate the respective function.
6 Using the product
Humidity (testo 875-2 only)
The surface humidity is calculated using the manually entered dewpoint temperature and the measured surface temperatures. In the display, the areas that are at risk of developing mould are shown using a special color palette:
Color Surface moisture Assessment
green 0…64% noncritical
yellow-
65…80% potentially critical
orange
red >80% critical
If this function is activated ( ), the quick select buttons are permanently assigned the Dewpoint and Emissivity... functions.
21
6 Using the product
The dewpoint temperature is shown at the upper edge of the display.
1. [OK] | Measurement | [OK] | Humidity | [OK].
- The Dewpoint dialogue is opened. The dewpoint temperature
2. Move Joystick up/down to select the desired option.
- The selected option has an orange border.
3. Confirm the selection with [OK].
- The selected number is highlighted orange.
4. Move Joystick up/down to set the desired value. Move
5. Confirm the entry with [OK].
6. Confirm the settings with Apply.
Measuring range
The measuring range can be set to adapt to the respective area of application.
1. [OK] | Measurement | [OK] | Measurement range.
2. Select the desired temperature range and confirm with [OK].
can be entered. In addition, the values for ambient temperature and ambient humidity can be entered for informational purposes (evaluation by means of IRSoft software).
Joystick right/left to change between the numbers.
Depending on the scene temperature, the measuring range may be undershot or exceeded. In this case, --- or +++ is shown instead of the readings.
> If this occurs, change the measuring range.
6.1.2. Image gallery
Show images…
1. [OK] | Image gallery | [OK] | Show images… | [OK].
- The Folder dialogue is opened.
- The folder name of the opened folder is displayed in the header.
ROOT designates the base folder (top level).
- The saved data is shown as preview images (image overview). The image saved last is shown directly behind the folders.
22
6 Using the product
Optional: > Press the right [Page scroll] quick select button to change to
page scroll.
In page scroll, you can navigate in groups of 3 x 3
images/folders (page by page). To select/open an image/folder, you must change back into Single Image.
2. Navigate:
• In page scroll: Move Joystick up/down to change between the pages.
• In single image: Move Joystick up/down/left/right to select a preview image/a folder (orange border).
• Open selected folder with [OK].
3. In single image: Press [OK] to open selected preview image. Press Gallery to return to the image overview.
Delete image
1. In single image: Select image (orange border) or select and open image.
2. Press Delete to delete the selected or open image.
- A confirmation request appears as to whether the image should really be deleted.
3. Confirm with [OK] or cancel process with [Esc].
Create new folder
1. In the Folder dialogue: Select New Folder with the joystick and confirm with[OK].
- The Folder name dialogue is opened.
2. Name folders: Use the Joystick to select the desired letters and confirm with[OK].
The folder name can consist of up to 8 characters. Incorrect entries can be deleted using the right [C] quick select button.
3. Press left [Create Folder] quick select button to create the folder.
23
6 Using the product
Erase all
1. [OK] | Image gallery | [OK] | Erase all… | [OK].
- A confirmation request appears as to whether all files on the
memory card should be deleted. Folders are not deleted.
2. Confirm erase with [OK] or cancel process with [Esc].
6.1.3. Scale...
Set scale limits
You can choose between auto-scaling (continuous, automatic adjustment to the current min./max. values) and manual scaling. The scale limits can be set within the activated measuring range (see Measuring range page 22). All temperatures that undershoot or exceed the limit value are shown in the color of the limit value (depending on the selected color palette, see Palette page 28). Temperature ranges that are not relevant can thus be hidden.
1. [OK] | Scale… | [OK].
- The Set scale dialogue is opened.
2. Move Joystick left/right to select the desired option: Auto-
scaling (
value (
)
), min. value ( ), temperature range ( ) or max.
- The selected option has an orange border ( ).
3. With the selection of min. value, temperature range or max.
value: Move Joystick up/down to change the value(s).
- Changes are applied immediately so that the effects on the
display of the infrared image can be checked directly.
4. Close dialogue with [OK] or [Esc].
- The changes are saved.
6.1.4. Display...
Select display options
The presentation can be adapted by means of showing/hiding the following elements: Crosshairs, temperature scale and quick select buttons.
Hidden function keys can be shown again by pressing a quick select button: The first press of the button shows the function keys again. To trigger a function, the button must be pressed again.
24
6 Using the product
1. [OK] | Display… | [OK].
- The Display options dialogue is opened.
2. Move Joystick up/down to select the desired option.
- The selected option has an orange border (
3. Activate (
) or deactivate ( ) option with [OK].
4. Confirm settings with the [Apply] quick select button. or Ignore changed settings with [Esc].
6.1.5. Emissivity...
Emissivity
The emissivity describes the capability of a body to emit electromagnetic radiation. This is material-specific and must be adapted for correct measurement results.
Non-metals (paper, ceramic, gypsum, wood, paints and coatings), plastics and food have high emissivity, which means that the surface temperature can be easily measured using infrared.
Because of their low or non-uniform emissivity, bright metals and metal oxides are only somewhat suited for infrared measurement. You must anticipate highly inaccurate measurements. A remedy for this is coatings that increase the emissivity, e.g. paint or emissivity tape (art. no. 0554 0051) which must be applied to the object being measured.
The following table gives typical emissivities of important materials. These values can be used as orientation with the user-defined settings.
Material (material temperature) Emissivity
Aluminum, bright rolled (170 °C) 0.04
Cotton (20 °C) 0.77
Concrete (25 °C) 0.93
Ice, smooth (0 °C) 0.97
Iron, emery ground (20 °C) 0.24
Iron with casting skin (100 °C) 0.80
Iron with rolling skin (20 °C) 0.77
Gypsum (20 °C) 0.90
Glass (90 °C) 0.94
).
25
6 Using the product
Material (material temperature) Emissivity
Rubber, hard (23 °C) 0.94
Rubber, soft grey (23 °C) 0.89
Wood (70 °C) 0.94
Cork (20 °C) 0.70
Radiator, black anodized (50 °C) 0.98
Copper, slightly tarnished (20 °C) 0.04
Copper, oxidized (130°C) 0.76
Plastics: PE, PP, PVC (20 °C) 0.94
Brass, oxidized (200°C) 0.61
Paper (20 °C) 0.97
Porcelain (20 °C) 0.92
Black paint, matt (80 °C) 0.97
Steel, heat-treated surface (200 °C) 0.52
Steel, oxidized (200 °C) 0.79
Clay, burnt (70 °C) 0.91
Transformer paint (70 °C) 0.94
Brick, mortar, plaster (20 °C) 0.93
26
Reflected temperature
Using this offset factor, the reflection is calculated out due to the low emissivity and the accuracy of the temperature measurement with infrared instruments is improved. In most cases, the reflected temperature is identical to the ambient air temperature. Only when objects with strong emissions with much higher temperatures (such as ovens or machines) are in the proximity of the object being measured should the radiation temperature of theses heat sources be determined and used (e.g. using a globe thermometer). The reflected temperature has only little effect on objects with high emissivity.
6 Using the product
Set emissivity/reflected temperature
You can choose between a user-defined emissivity and 8 materials with recorded emissivities. The reflected temperature can be set individually.
The designation of the materials in the selection list is configured during commissioning of the thermal imager in the language selected by the user. A translation does not occur in the event of subsequent changeover of the language in the instrument.
Using the PC software, other materials can be imported into the instrument from an existing list. The language for the materials here depends on the operating system language of the PC.
1. [OK] | Emissivity… | [OK].
- The Emissivity dialogue is opened.
2. Move Joystick up/down to select the desired material.
- The selected material is highlighted orange (
). The corresponding emissivity is shown to the right next to the material.
3. Move Joystick to the right to switch to the setting of the emissivity (only available when selecting User defined) or to
Reflected temp..
- The selected option has an orange border ( ).
4. Confirm the selection with [OK].
- The selected number is highlighted orange ( ).
5. Move Joystick up/down to set the desired value. Move
Joystick right/left to change between the numbers.
6. Confirm the entry with [OK].
7. Confirm the settings with Apply.
- The changes are saved.
- The selected emissivity ( ) is shown at the bottom right in the display in measurement view.
27
6 Using the product
6.1.6. Palette
Change color palette for the infrared image
You can choose between 4 existing palettes. The currently
activated option is marked with a tick ( ).
1. [OK] | Palette| [OK].
2. Move Joystick up/down to select the desired option.
3. Confirm the selection with [OK].
6.1.7. Configuration
Time/date...
Time and date can be set. Time and date format are automatically set based on the selected instrument language.
1. [OK] | Configuration… | [OK] | Time/date… | [OK].
- The Set time/date dialogue is opened.
2. Move Joystick up/down to select the desired option.
- The selected option has an orange border ( ).
3. Confirm the selection with [OK].
28
- The selected number is highlighted orange ( ).
4. Move Joystick up/down to set the value. Move Joystick
right/left to change between the numbers.
5. Confirm the entry with [OK].
6. Confirm the settings with Apply.
Country settings…
Instrument language and temperature unit can be set.
1. [OK] | Configuration… | [OK].| Country settings… | [OK].
- The Country settings dialogue is opened.
2. Move Joystick up/down to select the desired option.
- The selected option has an orange border (
).
3. Confirm the selection with [OK].
- The selected setting is highlighted orange ( ).
4. Move Joystick up/down to change the setting.
6 Using the product
5. Confirm the entry with [OK].
6. Confirm the settings with Apply.
Optics...
The lenses adjusted to the instrument are shown. Only the lenses that are shown may be used.
> [OK] | Configuration… | [OK] | Optics… | [OK]. Using the Protection glass option, you can set whether an infrared
protection glass is used or not. Ensure the correct setting to prevent corruption of the measurement results.
> Press [OK] to activate (
) or deactivate ( ) the Protection
glass option.
- With the option activated, the protection glass symbol (
) is
shown in the bottom right of the display.
Power-save options…
Options can be set to extend the rechargeable battery running time.
1. [OK] | Configuration… | [OK] | Power-save options… | [OK].
- The Power-save options dialogue is opened.
2. Move Joystick up/down to select the desired option.
- The selected option has an orange border (
).
3. Confirm the selection with [OK].
- LCD backlight: The selected setting is highlighted orange and
can be changed (
).
> Move Joystick up/down to change the setting. Confirm the
entry with [OK].
- Switch off LCD or Power off imager: The selected option can be activated or deactivated.
> Press [OK] to activate (
) or deactivate ( ) the function.
- With the option activated, the length of time until shut-down
can be set.
> Move Joystick to the right and press [OK]. Move Joystick
up/down to change the setting. Confirm the entry with [OK].
4. Confirm the settings with Apply.
29
6 Using the product
Format…
The memory card can be formatted. During the formatting, all saved data on the SD card is lost.
1. [OK] | Configuration… | [OK] | Format… | [OK].
- A confirmation request appears as to whether the memory card
2. Confirm with [OK] or cancel process with [Esc].
Factory reset
The instrument settings can be reset to the factory settings. Time/date and country settings are not reset.
Please note: After a factory reset, the numbering of images begins anew. When saving images, images that have already been saved and have the same number are overwritten!
> Transfer all saved images onto a PC before performing a reset
1. [OK] | Configuration… | [OK] | Factory settings… | [OK].
- A confirmation request appears as to whether the instrument
2. Confirm reset with [OK] or cancel process with [Esc].
should really be formatted.
to prevent possible overwriting.
should really be reset to the factory settings.
6.2. Measuring
Important framework conditions
Please observe the following framework conditions in order to obtain significant measurement results.
Humidity measurement (testo 875-2 only):
• Gently wave the separate dewpoint measuring instrument to
shorten the equalization period. Avoid sources of interference (e.g. breathed air).
Building thermography, investigating the building shell:
• Considerable temperature difference between inside and
outside required (ideal: > 15 °C/> 27 °F).
• Consistent weather conditions, no intensive sunlight, no
precipitation, no strong wind.
To ensure the highest level of accuracy, the imager requires an equalization period of 5 minutes after being switched on.
30
6 Using the product
Important settings
Before recording an image, check whether the protection glass option is set correctly to prevent corruption of the measurement results, see Optics... page 29.
Before saving an image, ensure that it is focused correctly (in focus), see Manually focusing the image page 19. Images that are not in focus cannot be corrected retroactively!
To obtain precise readings, the emissivity and reflected temperature must be set correctly, see Set emissivity/reflected temperature page 27. A subsequent adjustment is possible via the PC software as required.
With an activated auto scaling, the color scale is continuously adapted to the min./max. values of the current measurement image. The color that is to be assigned to a specific temperature thereby changes constantly. To be able to compare multiple images by means of the coloring, the scaling must be manually set to fixed values, see Scale... page 24, or must be subsequently adjusted to uniform values using the PC software.
Select the storage location (folder):
1. While an image is frozen (fixed image): Press the left [Folder] quick select button.
- The Folder dialogue is opened.
To create a new folder, see Create new folder page 23.
2. Select the desired folder with the joystick and open with [OK].
- The folder name of the opened folder is displayed in the header.
3. Press left [Apply] quick select button to select the opened folder.
31
7 Maintaining the product
7 Maintaining the product
Changing the battery
To prevent losing the instrument settings: Only change the rechargeable battery when a backup battery is inserted in the instrument or the power supply is connected.
1. Press lock release button.
- The rechargeable battery is released and sticks out a little from
the rechargeable battery slot. The locking function keeps the battery from falling out.
2. Completely remove the battery from the rechargeable battery
slot.
3. Insert a new battery into the rechargeable battery slot until this
is flush with the bottom of the handle.
- The thermal imager starts automatically.
Clean thermal imager
> If the housing of the thermal imager is dirty, clean it with a damp
cloth. Do not use any aggressive cleaning agents or solvents! Weak household cleaning agents and soap suds may be used.
32
Cleaning lens/infrared protection glass
> Large dust particles can be removed with a clean optical
cleaning brush (available in specialized photography shops).
> Use a lens cleaning cloth for slight contamination. Do not use
rubbing alcohol!
Change backup battery
To prevent losing the instrument settings: Only change the backup battery when a charged rechargeable battery is inserted or the power supply is connected.
1. Open the cover on the left side of the thermal imager.
2. Remove battery holder, including backup battery, from the
battery slot.
7 Maintaining the product
3. Remove empty battery from the holder and insert new battery (type CR 1632).
NOTICE
Incorrectly inserted batteries can lead to damage of the product!
> When inserting the batteries, observe the polarity (label on the
battery holder).
4. Insert battery holder into the battery slot.
33
8 Tips and assistance
8 Tips and assistance
8.1. Questions and answers
Question Possible causes/solution
Message Backup battery
empty or not present is
shown.
Error No memory card
inserted! is shown.
1. Confirm the message with [OK].
2. Change backup battery, see Change backup battery page 32.
Memory card defective or not inserted.
1. Confirm the message with [OK].
2. Check or insert memory card.
Error Memory card full! is shown.
Insufficient memory present.
1. Confirm the message with [OK].
2. Insert new memory card.
Error No lens/wrong lens! is shown.
1. Switch off thermal imager.
2. Check lens
> Change lens if necessary.
Message Charging not
possible! is shown.
Ambient temperature is outside the permissible range for charging the battery.
1. Confirm the message with [OK].
2. Keep to the permissible ambient temperature.
Error Permissible
instrument temperature exceeded! is shown.
1. Switch off thermal imager.
2. Let instrument cool off and keep to the permissible ambient temperature.
--- or +++ is shown instead
of the reading.
Reading is outside of the measuring range.
> Change measuring range.
xxx is shown instead of the
reading.
Reading cannot be calculated.
> Check parameter settings for
plausibility.
If we could not answer your question, please contact your dealer or Testo Customer Service. For contact details see the rear side of this document or the web page www.testo.com/service-contact
34
8.2. Accessories and spare parts
Description Article no.
Lens protection glass 0554 8805
Sun Shield 0554 8806
SoftCase 0554 8814
Rech. batt. 0554 8802
Desktop charging station 0554 8801
Tripod 0554 8804
Power supply 0554 8808
SD card 0554 8803
USB cable 0449 0047
ISO calibration certificate for thermal imager:
• Calibration points 0 °C, 25 °C, 50 °C in measuring range -20 °C - 100 °C
• Calibration points at 0 °C, 100 °C, 200 °C in measuring range 0 °C to 280°C
8 Tips and assistance
0520 0489
0520 0490
For further accessories and spare parts, please refer to the product catalogues and brochures or look up our website: www.testo.com
35