Pos: 1 /TD/Überschriften/MUF/Sicherheit und Umwel t @ 3\mod_1234793958627_79.doc @ 26223 @ 1
1 Safety and the environment
Pos: 2 /TD/Sicherheit und Umwelt/Sicherheit gewähr lei s ten/MUF 63xx/Elektrische Gefahren vermeiden @ 3\ mod_ 1234 794609299_79.doc @ 26280 @ 5
Avoiding electrical hazards
> Never use the instrument and connected probes to measure on
or near live parts!
> Damaged mains cables must only be replaced by authorized
personnel.
> Only have the transmitter wired and connected by authorized
personnel with the voltage disconnected.
> You must always comply with the regulations applicable in your
Pos: 3 /TD/Sicherheit und Umwelt/Sicherheit gewähr lei s ten/MUF 63xx/Personen- und Sachschäden vermeiden @ 3\ mod_1234794744768_79.doc @ 26299 @ 5
country for opening and repairing electrical equipment.
Avoiding personal injury and damage to equipment
> Installation, setting and calibration work must only be carried
out by qualified and authorized personnel!
> Only open the instrument when this is expressly described in
the instruction manual for installation, maintenance or repair
purposes.
> Observe the permissible storage, transport and operating
Pos: 4 /TD/Sicherheit und Umwelt/Sicherheit gewähr lei s ten/Nicht mit Lösungsmitteln lagern @ 0\mod_11756923 75179_79.doc @ 583 @
temperature.
> Do not store the product together with solvents. Do not use any
Pos: 5 /TD/Sicherheit und Umwelt/Sicherheit gewähr lei s ten/MUF 63xx/Bei Wartung MUF nicht zur Regelung verwende n @ 3\ mod_1234794852377_79.doc @ 26318 @
desiccants.
> Do not use the instrument for control purposes at the same time
Pos: 6 /TD/Sicherheit und Umwelt/Sicherheit gewähr lei s ten/Produkt bestimmungsgemäß verwenden @ 0\mod_11737 81261848_79.doc @ 386 @
as operating or servicing the transmitter.
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not
Pos: 7 /TD/Sicherheit und Umwelt/Sicherheit gewähr lei s ten/Nur beschriebene Wartungsarbeiten durchf ühren @ 0\mod_1175692705195_79.doc @ 601 @
use any force.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts from
Pos: 8 /TD/Sicherheit und Umwelt/Sicherheit gewähr lei s ten/MUF 63xx/Fachpersonal @ 3\mod_1234794940409_7 9.doc @ 26337 @
Testo.
Any additional work must only be carried out by authorized
personnel. Otherwise testo will not accept any responsibility for the
proper functioning of the instrument after repair and for the validity
of certifications.
Pos: 12 /TD/Sicherheit und Umwelt/Zu diesem Dokument/Ver wendung (Standard) @ 0\mod_1173775068554_79.doc @ 337 @ 5
regulations) or return the product to Testo for disposal.
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay
particular attention to the safety instructions and warning advice
in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
Pos: 13 /TD/Sicherheit und Umwelt/Zu diesem Dokument/Sy mbole und Schreibkonventionen/Warnhinweis W ARNUNG @ 2\mod_1207646966234_79.doc @ 14398 @
Pos: 14 /TD/Sicherheit und Umwelt/Zu diesem Dokument/Sy mbole und Schreibkonventionen/Warnhinweis VORSIC HT @ 2\mod_1207651416515_79.doc @ 14416 @
Pos: 15 /TD/Sicherheit und Umwelt/Zu diesem Dokument/Sy m bo le und Schreibkonv. Software [Standard] @ 0\mod_11902 0333 2543_79.doc @ 4883 @ 5
product.
WARNING
CAUTION
Indicates potential serious injuries
indicates potential minor injuries
Symbols and writing standards
Representa-
Explanation
tion
Note: Basic or further information.
1. ...
2. ...
Action: more steps, the sequence must be
followed.
> ... Action: a step or an optional step.
- ... Result of an action.
Menu
[OK]
Elements of the program interface.
Buttons of the program interface.
... | ... Functions/paths within a menu.
“...” Example entries
The measuring uncertainty increases with high flow values.
1
Above 0 °Ctd the dewpoint temperature is shown and below 0 °Ctd the frost
point temperature is shown. In rare cases at frost point temperatures between
-35 °Ctd and 0 °Ctd differences between the 6781 and a dew mirror hygrometer
may occur.
2
Expanded basic uncertainty (K = 2) at a process temperature of 25 °C.
Measuring uncertainty according to GUM (Guide to the E xpression of Uncertainty in Measurement):
ISO guideline for determining the measuring uncertainty in order to render global measurement results
comparable.
The following uncertainties are used during the inquiry:
• Hysteresis
• Linearity
• Reproducibility
• Adjustment area/factory calibration
• Test location
An additional uncertainty factor independent of the dewpoint temperature and
process of ±0.03 K x reading (in °C
) + 0.2 K x (25 °C - process temperature in
td
°C) results from this overall view
Outside the stated range, a measurement inaccuracy of +/-5K (typically) applies.
3
At process temperatures > 25 °C the lower limit of the measuring range
increases by 0.5 K/K. The uncertainty values are valid for the respective
measuring range.
9
Page 10
4 Transmitter
Response time
≤ 3 s when changing from -75 °Ctd to -30 °Ctd
t
63
t
≤ 9 s when changing from -75 °Ctd to -30 °Ctd
90
t
≤ 300 s when changing from -30 °Ctd to -75 °C
63
td
t90 ≤ 1080 s when changing from -30 °Ctd to -75 °Ctd
Resolution
Measuring range Resolution
0 to 100 % RH 0.001
0.001 to 28 g/kg 0.001
0.01 to 194 gr/lb 0.01
0 to 31 g/m3 0.001
0.001 to 14 gr/ft3 0.001
1 to 42,500 ppm (vol) 1
-90 to 30 °Ctd 0.1
-130 to 86 °Ftd 0.1
-110 to 30 °C
-165 to 86 °F
0.1
tdA
0.1
tdA
-40 to 70 °C 0.01
-40 to 158 °F 0.01
10
Meas. cycle
• 1/sec
Interface
• Mini-DIN for P2A software (adjustment and parameterization
software)
Voltage supply
• 4-wire (separate signal and supply lines): 20 to 30 V AC/DC,
300 mA power consumption
Page 11
4 Transmitter
Maximum load
• 4-wire: 500 Ω (power output)
Maximal load
• 4-wire: 10 kΩ (voltage output)
Analog output
• 0 to 1 V ± 1.5 mV (4-wire) or
• 0 to 5 V ± 7.5 mV (4-wire) or
• 0 to 10 V ± 15 mV (4-wire) or
• 0 to 20 mA ± 0.03 mA (4-wire) or
• 4 to 20 mA ± 0.03 mA (4-wire)
Resolution of analog output
• 12 bit
Display
• 2-line LCD with plain text line (optional)
Operating temperature without display
• -40 to 70 °C
Operating temperature with display
• 0 to 50 °C
11
Page 12
4 Transmitter
Storage temperature without display
• -40 to 80 °C
Storage temperature with display
• -20 to 80 °C
Process temperature
• -40 to 70 °C
Application humidity
• 0 to 100 % RH
Pressure application range
• 50,000 hPa
Housing, weight
• Metal, 0.5 kg
Process connection
• G 1/2 or
• NPT 1/2"
12
Protection class
• IP 65 only if the transmitter is wired and/or sealing plugs are
inserted
Directives, standards and tests
• EC Directive: 2004/108/EC
• 6781 was manufactured in accordance with EC pressure
equipment Directive 97/23/EC, Article 3 Paragraph 3 as per
"sound engineering practice", taking the regulations to be
observed into account.
Warranty
• Duration: 2 years
• Warranty conditions: see website www.testo.com/warranty
corresponding to the presetting of the
measurement chamber (1), a measurement
chamber (1) is required for the rapid
• 1: Measurement chamber
• testo 6781 with
G1/2 thread
assembly (without de-pressurizing the
installation).
With clean air and a process pressure
B
greater than or less than 7000 hPa (1000 to
35,000 hPa) that therefore does not
correspond to the presetting of the inflow
valve, a measurement chamber (1) with a
• 1: Measurement chamber
• 2: Flow meter
• testo 6781 with
G1/2 thread
flow meter (2) is required.
14
Page 15
4 Transmitter
Application System components
For rapid assembly (without de-pressurizing
C
the installation), a measurement chamber (1)
is required. Connect a filter (4) upstream for
oily media.
• 4: Filter
• 1: Measurement chamber
• testo 6781 with
G1/2 thread
With dry air (max. 140 °C), a PTFE tube (3)
D
is used and the valve of the measurement
chamber (1) is opened to maximum (suitable
up to -60 °C
For process temperatures of 50 °C to 200 °C
E
).
td
and a process pressure corresponding to the
presetting, it is necessary to use a cooling
coil (5) and a measurement chamber (1).
Connect a filter upstream for oily media.
• 3: PTFE tube
• 1: Measurement chamber
• testo 6781 with
G1/2 thread
• Filter, if necessary
• 5: Cooling coil
• 1: Measurement chamber
• testo 6781 with
G1/2 thread
If A to E are not required. The thread is
F
screwed directly into the process.
De-pressurized tube required during
installation.
Pos: 36 /TD/Erste Schritte/MUF 63xx/Mechanische Mon tag e 6781 @ 3\mod_1239807163104_79.doc @ 31204 @ 3444
range limits are decisive.
4 Transmitter
4.3.1. Mechanical assembly
• Carefully perform assembly work.
• Take special care with leaktightness. We recommend
inserting a metal sealing ring (internal diameter 21 mm).
• Do not use any hygroscopic materials for the sealing.
> Only use force on the union nut.
Depending on the application, there are three different assembly
options:
Assembly Application
Without stainless steel
measurement chamber
and cooling coil
• Process temperatures 0 to 50 °C
• Attachment of the sensor directly in
the process possible
• No fast assembly/dismantling of the
testo 6781 required and sufficient
inflow of the sensor (1 l/min.) exists
17
Page 18
4 Transmitter
Assembly Application
With stainless steel
measurement chamber
• Process temperatures 0 to 50 °C
• Fast assembly/dismantling of the
testo 6781 required and/or
• No sufficient inflow of the sensor
(1 l/min.) exists
With stainless steel
• Process temperatures 50 to 200 °C
measurement chamber
and cooling coil
4.3.1.1. Without stainless steel measurement chamber and cooling
coil
CAUTION
Compressed air! Danger of injury!
> De-pressurize pipe sections (e.g. compressor off or use of
bypass) and vent them before opening.
Clamp ring screw connection is already pre-assembled.
Screw connection body (1) and union nut (2) have a pointshaped marking (4).
18
1. Screw thread in process.
2. Turn housing (3) to desired position without much exertion
of energy.
3. Tighten union nut (2) finger-tight.
4. Hold screw connection body (1) with a suitable spanner.
5. Tighten union nut (2) slightly past the marking (4).
An increased resistance can be felt when tightening.
Page 19
4 Transmitter
6. Close ventilation of the pipe section and pressurize the pipe
section again.
4.3.1.2. With stainless steel measurement chamber (max. 35,000 hPa)
1. Connect measurement chamber without testo 6781 transmitter
to quick-release compressed-air fastener for approx. 10 - 30
sec. to blow out any dirt deposits.
2. Remove measurement chamber from quick-release
compressed-air fastener.
3. Screw process connection (G1/2) of the testo 6781 transmitter
into the thread of the measurement chamber.
4. Mount 6781 as described under Without stainless steel
measurement chamber and cooling coil page 18.
5. Engage quick-release fastening of the measurement chamber
in the standardized socket of the compressed air line.
4.3.1.3. With stainless steel measurement chamber and cooling coil
(max. 35,000 hPa)
1. Connect measurement chamber without testo 6781 transmitter
to quick-release compressed-air fastener for approx. 10 - 30
sec. to blow out any dirt deposits.
2. Remove measurement chamber from quick-release
compressed-air fastener.
3. Screw process connection (G1/2) of the testo 6781 transmitter
into the thread of the measurement chamber.
4. Mount 6781 as described under Without stainless steel
measurement chamber and cooling coil page 18.
5. Engage quick-release fastening of the measurement chamber
in quick-release fastening of the cooling coil.
6. Engage second quick-release fastening of the cooling coil in the
Conventional trace humidity sensors show a steep rise in
measuring uncertainty at low humidities. With the trace humidity
sensor of the testo 6781, these measuring uncertainties are
corrected by means of an automatic self-adjustment process. This
means that extremely accurate measuring results are also attained
in the range up to -90 °C
To this end, a temperature sensor which is used as a heater is
fitted on the back of the sensor element. A humidity and
temperature value pair is taken in both the unheated and heated
state. The deviation of the probe obtained from these pairs of
values is automatically corrected.
.
td
4 Transmitter
21
Page 22
4 Transmitter
• Deactivating the self-adjustment function of the testo
6781 will reduce measuring accuracy and should
therefore be restricted to the shortest possible length of
time.
• During the self-adjustment phase, the analog outputs,
the display value and the output value are "frozen".
Self-adjustment active is shown in the display until it
has finished. The self-adjustment time (incl. heating
time, calculation time, cooling time) is 30 minutes. The
cycle time can be edited in the P2A software, see Selfadjustment page 54.
• At temperature wn > 0 °C, the self-adjustment time
becomes longer.
4.3.2.2. 1-point adjustment by entering a reference value
Using the 1-point adjustment, you can enter a reference value for
one of the working points you have selected (e.g. -40 °C
this, you achieve a minimum target/actual deviation around this
working point.
But the further away the measurement is from the working point,
the greater the deviation can become. The 1-point adjustment
should therefore only be used for a relatively narrow measuring
range (working range)
✓ A reference measuring instrument (e.g. dewpoint mirror
hygrometer) is ready.
1. Expose the reference measuring instrument and testo 6781 to
the same constant conditions and wait for the equalization
period to expire.
The ideal accuracy is reached when at least two adjustment
cycles are run through at dewpoint temperatures above -60
°C
, whereby four adjustment cycles are recommended at
td
dewpoint temperatures below -60 °C
.
td
2. Measure the reference value and compare with the reading of
the testo 6781.
3. In the event of deviations of the values: Enter reference value
• via the user menu (see Enter reference value for 1-point
adjustment page 35) or
• via the P2A software (see 1-point adjustment page 61).
The purpose of adjusting the analog outputs is to adjust the signal
chain from the digital signal (within the transmitter) to the analog
outputs. The signal type that was appointed for the transmitter is
adjusted respectively for each channel (e.g. 4 to 20 mA or 0 to 1 V,
✓ Load of max. 500 Ω is connected to channel 1 (see Plug-in
connection for power supply and channel page 21)
✓ A precise multimeter (minimum requirement: resolution 6.5
digits, at least 5-times more accurate than the 6781) is
available.
If only a simple multimeter is available, the analog outputs
may not be adjusted.
1. Connect the inputs of the multimeter with the contacts (1) and
(2) for channel 1.
2. Transfer the reference analog value measured with the
multimeter into the P2A software (see Adjusting the analog
output page 62) or enter it via the user menu (see Performing
analog adjustment page 36).
3. Disconnect connections between the multimeter and the
contacts of the testo 6781.
4 Transmitter
1 Contact ch. 1+
2. Contact ch. 1-
3. Service interface
23
Page 24
4 Transmitter
4.4. Operation
Pos: 43 /TD/Produkt verwenden/MUF 63xx/Zusammenhang Bedi enmenü – Mini DIN Buchse aktiv @ 3\mod_1234454016014_79. doc @ 25213 @ 3
4.4.1. Relationship between user menu and mini-DIN
socket is active
The testo 6781 can be parameterized using either the user menu or
the P2A software (see Parameterizing, adjusting and analyzing
software (P2A software) page 42).
The testo 6781 transmitter can only be operated via the
display and keypad if the display option is available.
If the testo 6781 is connected to the P2A software, the user
menu is blocked for the duration of the communication. The
message Service plug is shown in the display of the testo
6781. As soon as the P2A software is disconnected, the
The user menu can be protected with a four-digit numerical code
(see Calling up Main Menu page 35) so that access to the user
menu is denied to unauthorized persons not familiar with this
numerical code.
If the password protection is not to be used, the numerical code
The alarm statuses are shown on the display. You can choose
whether the alarm is to be used to monitor limit values or as a
collective alarm. If an alarm is to be used to monitor limit values,
you can choose between monitoring the minimum or maximum
value and set a limit value and hysteresis for each alarm.
In addition, every alarm can be linked to a clearly visible visual
alarm (display background lighting flashes).
An alarm delay between 0 and 240 seconds can still be assigned to
every alarm used for limit value monitoring so that the visual alarm
is delayed. If the alarm status goes out within the set alarm delay
time, the visual alarm is not triggered. .
With an alarm status present, the visual alarm can be reset by
means of acknowledgement. The triggering of a new alarm cannot
be enabled until after the alarm status goes out.
1. In the Measuring Mode press SET, select Main Menu Alarm
with X or S and confirm selection with SET.
- Four alarms can be parameterized.
2. Select Alarm x with X or S and confirm selection with SET.
4 Transmitter
Using alarm to monitor limit values
Monitoring minimum Monitoring maximum
Hysteresis
Limit value Limit value
Hysteresis
3. Select Channel x (e.g. "Channel 1") with X or S and confirm
selection with SET.
4. Select Max control or Min control with X or S (see graphic).
5. Press SET and edit Limit value as well as Hysteresis: Scroll
one digit to the right using X and increase value of digit by 1
using S. Confirm with SET or abort entry with ESC.
6. Select Visual alarm with X or S. Select YES or NO with X or
S. Confirm with SET or abort entry with ESC.
7. Press SET and edit Alarm delay: Scroll one digit to the right
using X and increase value of digit by 1 using S. Confirm with
SET or abort entry with ESC.
8. Return to Channel x with ESC.
29
Page 30
4 Transmitter
9. Return to Alarm x with ESC.
10. Change to the other alarms using X or S and perform settings
Using alarm as collective alarm or not using it at all
If an alarm is assigned to the collective alarm, the display flashes
as soon as (at least) one of the warning or error messages of the
testo 6781 transmitter becomes active.
✓ Alarm is selected (see previous steps 1 and 2).
1. Use X or S to determine whether Alarm x should be used as a
2. If collective alarm is selected: Select Visual alarm with X or S.
3. Change to another alarm using X or S and perform settings in
4. Return to Main Menu Alarm with ESC.
5. Continue to Main Menu Settings with X or S or return to
The messages affecting the collective alarm can only be
selected in the P2A software, see Using the software page
46
Collective alarm or not used. Confirm selection with SET.
Select YES or NO with X or S. Confirm with SET and return to
Alarm x.
the same way.
Measuring Mode with ESC.
4.4.5.3. Editing Main Menu Settings
You can edit instrument settings and other settings.
> In Measuring Mode, press SET, select Main Menu Settings
using X or S and confirm selection with SET.
You can edit settings for:
• Display
• Language
• Code
• Unit
◦ Absolute pressure
◦ Temperature
• Humidity process data
30
Page 31
4 Transmitter
Editing display settings
You can set the brightness and contrast of the display.
1. Select Display Settings with X or S and confirm selection with
SET.
2. Select Backlight or Contrast with X or S and confirm
selection with SET.
One of the following parameters can now be selected using X or
S, after which the selection must be confirmed with SET:
• Backlight
The display illumination is changed.
Edit/select parameter with X or S, confirm with SET or cancel
entry with ESC (the effect of the change in parameter can be
seen during input).
• Contrast
The brightness difference between the display background and
the displayed values is changed.
Edit/select parameter with X or S, confirm with SET or cancel
entry with ESC (the effect of the change in parameter can be
seen during input).
• Backlight 24h on
Select On or Off using X or S and confirm with SET.
Off: The display light switches off automatically if no button was
pressed for 10 seconds.
On: The display light is activated
3. Return to Display Settings with ESC and useX or S to
continue to Language.
Selecting language
You can se
lect the language for the plain text line in the display.
> Press SET, select required language with X or S, confirm
selection with SET and return to Language.
Only choose a language that you can understand well.
31
Page 32
4 Transmitter
Editing code settings
You can set the access code (password).
1. Select Code with X or S and confirm selection with SET.
2. Scroll one digit to the right using X and increase value of digit
3. Return to Code with ESC and useX or S to continue to
Select unit
This setting affects the unit of the process data.
1. Press SET, select Change parameters with X or S, confirm
2. Select Change unit with X or S, confirm selection with SET or
3. Select the required variable (absolute pressure/temperature)
4. Select the required unit with X or S, confirm selection with SET
5 Return to Change unit with ESC.
6. Select Hu
7. Select Humidity process pressure with X or S, confirm
8. Scroll one digit to the right using X and increase value of digit
Pos: 50 /TD/Produkt verwenden/MUF 63xx/Hauptmenü Ana ly se bearbeiten 6781 @ 3\mod_1241425854399_79.doc @ 32344 @ 455
9. Return to Main Menu Settings by pressing ESC three times.
If a code other than "0000" (factory setting) is set, the
transmitter can only be operated once this code has been
entered via the menu.
by 1 using S. Confirm with SET or abort entry with ESC.
Change parameters.
selection with SET or cancel with ESC.
cancel with ESC.
with X or S, confirm selection with SET or cancel with ESC.
or cancel with ESC.
midity process data with X or S, confirm selection
with SET or cancel with ESC.
selection with SET or cancel with ESC.
by 1 using S. Confirm with SET or abort entry with ESC.
4.4.5.4. Editing Main Menu Analysis
You can check the function of analog outputs. In addition, you can
read off the minimum and maximum values (since the last voltage
supply or reset of the min./max. values).
32
Testing functionality of analog outputs
This function affects the analog outputs directly, not only
the test contacts.
Page 33
1. In the Measuring Mode press SET, select Main Menu Analysis
- Test Analog Output is shown.
2. Press SET, scroll one digit to the right using X and increase
3. Accept setting by pressing SET and test with multimeter
3. Return to Test Analog Output with ESC and continue to
Reading off min./max. values of channels
To reset the max./min. values, see Editing Reset main menu page
37
1. Read off the min./max. values of the channel in succession with
2. Continue to Main Menu Message with X or S or return to
Pos: 51 /TD/Produkt verwenden/MUF 63xx/Hauptmenü Meld unge n bearbeiten 6781 @ 3\mod_1240905240318_79.doc @ 32203 @ 4
4 Transmitter
with X or S and confirm selection with SET.
value of digit by 1 using S. Any analog output value can be
predefined, e.g. for an analog output of 4 to 20 mA, the value
"6.0 mA". Confirm with SET or abort entry with ESC.
(minimum requirement: resolution 6.5 digits, at least 2-times
more accurate than the 6781) as follows:
Analog output 1: Via test contacts under service flap, see
diagram.
1 Channe l 1
test contacts
2 Multimeter
Max./min. values with X or S.
X or S and return to the Main Menu Analysis with ESC.
Measuring Mode with ESC.
33
Page 34
4 Transmitter
4.4.5.5. Editing Message main menu
Messages can be confirmed/acknowledged, the last messages can
be called up and the display of the messages can be switched on
or off.
Using the P2A software (see Parameterizing, adjusting and
analyzing software (P2A software) page 42) you can
predefine whether messages are to be shown in the
display.
1. In the Measuring Mode press SET, select Main Menu Message
with X or S and confirm selection with SET.
2. Confirm Confirm message using SET.
3. Select Last messages with X or S and confirm with SET.
4. Scroll between the warning and error messages recorded so far
using X or S and return to Last messages using ESC.
5. Select Information with X or S and confirm with SET.
6. Scroll between the status messages recorded so far using X or
S and return to Information using ESC.
7. Continue to Display of message with X or S.
ON: Measurements are shown on the display in Measuring
Mode.
OFF: No messages shown on display.
8. Select ON or OFF with X or S and confirm selection with SET.
9. Return to Main Menu Message with ESC.
1 Message text
2 Additional information on
the message
3 Message num ber
(example: "4/7" refers to
the fourth of seven
messages)
4 Number of messages
present (example: "4/7"
refers to the fourth of
seven messages)
5 Message code (see
Status, warning and
error messages page
37).
34
Page 35
10. Continue to Main Menu Ident with X or S or return to
Measuring Mode with ESC.
An overview of the messages can be found in Status,
1. In the Measuring Mode press SET, select Main Menu Ident
with X or S and confirm selection with SET.
- The type, firmware version, build number and serial number of
the transmitter are displayed.
This information is required when servicing.
2. Return to Main Menu Ident with ESC.
3. Continue to Main MenuAdjust with X or S or return to
Pos: 53 /TD/Produkt verwenden/MUF 63xx/Hauptmenü Abg lei c h bearbeiten 6781 @ 3\mod_1239867127747_79.doc @ 31254 @ 455
Measuring Mode with ESC.
4.4.5.7. Editing Main Menu Adjust
A reference value can be entered for the temperature (°C/°F) and a
reference value can be entered for the dewpoint (°C
1-point adjustment. Please refer to the description in Adjusting the
analog output page 62.
In addition, the analog outputs can be adjusted. Also see Analog
output adjustment page 23 for instructions on how to do this.
Enter reference value for 1-point adjustment
Please also refer to Analog output adjustment page 23.
1. In the Measuring Mode press SET, select Main Menu Adjust
with X or S and confirm selection with SET.
2. Select Reference value temp or Reference value Td with X
or S.
3. Press SET, edit value: Scroll one digit to the right using X and
increase value of digit by 1 using S. Confirm with SET or abort
entry with ESC.
4. Continue to Analog Adj. Ch. 1 with X or S.
5. Continue with the adjustment of the analog outputs (see below,
step 2) or press ESC to return to Main Menu Adjust.
Performing analog adjustment
Please refer to Analog output adjustment page 23.
1. In the Measuring Mode press SET, select Main Menu Adjust
with X or S and confirm selection with SET.
Each channel is adjusted at three points in the analog
range (at 10 %; 50 %; 90°% of the analog scale).
2. Select Analog Adj. Ch. 1 with X or S and confirm with SET.
3. Use X or S to select Adj. Point 1.
4. Press SET. Read off multimeter display (e.g. 5601 mA) and
enter this value in the user menu. Do this by scrolling one digit
to the right using X and increasing the value of digit by 1 using
S. Confirm with SET or abort entry with ESC.
5. Use X or S to select Adj. Point 2.
6. Press SET. Read off multimeter display (e.g. 12,001 mA) and
enter this value in the user menu. Do this by scrolling one digit
to the right using X and increasing the value of digit by 1 using
S. Confirm with SET or abort entry with ESC.
7. Use X or S to select Adj. Point 3.
8. Press SET. Read off multimeter display (e.g. 18,401 mA) and
enter this value in the user menu. Do this by scrolling one digit
to the right using S and increasing the value of digit by 1 using
X. Confirm with SET or abort entry with ESC.
9. Return to Main Menu Adjust with ESC.
10. Continue to Main Menu Reset with X or S or return to
Measuring Mode with ESC.
36
Page 37
4.4.5.8. Editing Reset main menu
You can reset the factory settings for the following:
• Instrument
Pos: 55 /TD/Überschriften/MUF/1.5 Status-, W arn- und Fehlermeldungen @ 3\mod_1234442925442_79.doc @ 24963 @ 2
4.5. Status, warning and error messages
Pos: 56 /TD/Produkt verwenden/MUF 63xx/Status-W arnmeldungen/Status-, Warn- und Fehlermeldungen _6781 @ 3\ m od_1 241441355544_79.doc @ 32477 @
Due to a processor
error, the transmitter
performs an automatic
If the problem occurs
frequently, contact
Testo Service.
restart.
Low adjustment
temperature
High adjustment
temperature
Shown on the
display
4
The ambient
temperature is too low
during the pressure
adjustment.
The ambient
temperature is too high
during the pressure
adjustment.
Can be used for
collective alarm5
Take necessary
measures to raise
ambient temperature,
e.g. through heating.
Take necessary
measures to lower
ambient temperature,
e.g. through venting.
Additional
message end6
New limit value x
Scaling changed x
Alarm 1 x
Alarm 2 x
Alarm 3 x
Alarm 4 x
Transmitter reset x
Analog out adjust x
4
If multiple messages/alarms are activated at the same time, only the last
message/alarm is shown. If this is cancelled, the other messages that are still
active are no longer displayed.
5
The message can be assigned the collective alarm function, which means that
the collective alarm is activated as soon as at least one of the messages
assigned to it is activated.
6
The message is shown upon the occurrence of the event causing the message
as well as when closing. Two entries appear in the history in the P2A software:
Message text_start and Message text_end.
40
Page 41
4 Transmitter
Shown on the
display4
Can be used for
collective alarm5
T ambient high x x
T ambient low x x
Supply voltage low x x
Watchdog error x
Perform the Confirm message function (acknowledgement of the
alarm via the control keys on the transmitter):
• The message/alarm is no longer shown on the display and the
optical alarm goes out, where applicable. If multiple
messages/alarms are active at the same time, all are reset
simultaneously.
• If at least one message is assigned to the collective alarm, the
Pos: 61 /TD/Überschriften/MUF/1.6 Wartung und Reinigu ng @ 3\mod_1234443039129_79.doc @ 24982 @ 2
If used in process conditions involving oil or dust, the protection cap
and, if used, the measurement chamber and cooling coil must be
cleaned regularly.
• Unscrew/remove protection cap, measurement chamber and
cooling coil and blow out with compressed air or place in an
ultrasonic bath.
Page 42
5 Parameterizing, adjusting and analyzing software (P2A software)
• Replace a heavily soiled protection cap with the same type of
protection cap.
The accuracy is only maintained when the ordered
protection cap is replaced by the same type of protection
cap.
Sensor cleaning
During sensor cleaning, avoid any contact with the sensor.
Do not mechanically clean the sensor, as otherwise the
cover electrode may be damaged.
1. Carefully unscrew protection cap.
2. Carefully rinse with isopropanol and/or distilled water.
The P2A software is used for the parameterizing, adjustment and
analysis of testo transmitters. The following applies:
• Generally, all newer testo transmitters (as of 2007) are
supported.
• Included with every testo transmitter that is bought new is a CD
that contains a free upgrade of the software, which includes the
device drivers for all transmitters that can be attached at this
time.
• This upgrade can be downloaded at any time via the testo
homepage "www.testo.com/Download/P2A".
The software must only be bought one time, even for owners of
several testo transmitters.
42
Page 43
5 Parameterizing, adjusting and analyzing software (P2A software)
5.1.1. Functions and use
In the P2A software, two different file types are used: The
instrument and the parameter file.
Instrument file
The parameters of a particular transmitter are stored in its so-called
instrument file. Using this file, the parameters can be edited and the
instrument can be tested and adjusted.
Instrument files also contain the respective histories in addition to
the parameter data, i.e. "log books" are kept for the previous
parameterizations, adjustments and messages (see Transmitter
Parameter files are not tied to a specific individual transmitter and
contain only parameter data/no history data.
If you use various instruments of the same type, you can create
parameter files once (e.g. by saving the appropriate instrument file
as the parameter file) and transmit these onto the other
instruments.
Parameter files are ".cfp" format files.
5.1.2. System requirements
Operating system
• Windows® 2000 SP4
• Windows® XP Home/Professional
• Windows® Vista
Computer
• Pentium processor of at least 400 MHz or equivalent
• 128 MB RAM
• Graphics resolution of at least 1024 x 768
• Unused hard drive capacity of at least 15 MB
• CD-ROM drive
• USB interface
43
Page 44
5 Parameterizing, adjusting and analyzing software (P2A software)
• At least Internet Explorer 5.0.
Software
The P2A software must be purchased and installed separately from
the transmitter. If it is a new software version, the transmitter is
already supported completely. Older P2A software versions can be
updated via the P2A software upgrade (cf. product CD included
Shows the Windows dialogue for
searching and opening files.
Save as
Saves the parameters of an
instrument or parameter file
under a new name.
Edit Copy
Copies the parameters of the
marked instrument or parameter
file in the cache.
Paste
Pastes the parameters from the
cache in the marked instrument
or parameter file.
View Toolbar
Status bar
Activates/deactivates the toolbar
or status bar.
Page 47
5 Parameterizing, adjusting and analyzing software (P2A software)
Menu Command Explanation
?
Check instrument
connections
Checks the connections to a
connected instrument without the
instrument having to be
activated.
Service
A text file with the most important
information on the computer and
the software is opened via
Display service data.
Information
Shows the version number of the
P2A software.
2 Toolbar: Shows the Windows-compliant icons for editing.
3 File:
Icon File Explanation
Symbol shows a
transmitter
Instrument
file
Instrument file
Connection to the instrument has
been established.
<Type> <Serial number>.cfm
File name should not be
changed.
Symbol shows a
transmitter with a red
minus sign in the upper
left corner
Symbol shows a
transmitter with a white
P for parameter file in
the upper left corner
Instrument
file
Parameter
file
Instrument file
Connection to the instrument has
not been established.
<Type> <Serial number>
<Date> <Time>.cfp
File name can be changed.
The name can be selected freely,
but it is recommended that you
retain the reference to the
instrument.
Parameter files are always
marked red; the parameter
values they contain are only
forwarded to the instrument after
being transmitted to the
instrument file.
47
Page 48
5 Parameterizing, adjusting and analyzing software (P2A software)
4 Function buttons: Dialogues on editing and testing the
instrument are opened by means of the buttons.
[Change parameterization] see Changing instrument/parameter
file page 48
[Test/analyze transmitter] see Chapter Analyzing/testing the
transmitter page 57
[Adjusting the transmitter] see Chapter Adjusting the transmitter
page 60
[Transmitter history] see Transmitter history page 63
5 File information:
Status Shown in the window
An instrument file is
selected
A parameter file is
selected
Type, serial number, firmware version of the
instrument.
Type, serial number and firmware version of
instrument for which the parameter file was
created.
Connection status Green = connection is active
Red = connection is inactive
6 Status bar: Shows the current status when editing via the menu
✓ The desired instrument/parameter file is marked.
1. Click on [Change parameterization].
- The Properties of <Instrument type> <Serial number>
dialogue is opened with the Change parameterization register.
If the parameters were transmitted from other parameter files into
the instrument file, a message is shown with which you can
transmit the new parameters to the connected instrument using
[Yes].
> If the parameters should not be transmitted, click on [No].
Pos: 75 /TD/Produkt verwenden/MUF 63xx/P2A/Einhei t / Analogausgang ohne Relais @ 4\mod_1244205188941_79.doc @ 44503 @
48
2. Change or enter parameters in the corresponding fields.
Page 49
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
Unit/
Analog output
All analog outputs are parameterized in this
mask.
Unit/analog output
(graphic)
Unit: 0 to 1 V/5 V/10 V or 0...20 mA, 4 to 20
mA.
Vertical: Current version of the analog output
(cannot be changed).
Horizontal: Min./max. scale end points of
selected unit.
The curve changes in accordance with the
entered value of scale minimum and
maximum.
Scale minimum/
maximum
The endpoints of the scaling can be selected
up to the stored scale minimum and
maximum. In the process, scaling can take
place beyond the measuring range in order
to adjust the analog output to the customer
system.
Unit Selection of the physical unit.
When changing the unit, standard values are
set for scale minimum and maximum.
Signal delay
(graphic)
49
Curve changes according to the set signal
delay.
Page 50
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
Signal delay Time interval in stages 1 – 15:
1 = no delay
15 = longest delay.
The signal delay is added to the reaction time
of the sensor. The signal delay shows
averaging over the time interval of the
selected stage in seconds:
Example
Stage 10 = average of the readings from the
last 10 seconds.
The delay of the signal in relation to the change in the
process is also significantly influenced by the selection of
3. Select storage location and enter the file name.
4. Click on [Save].
- The new parameter file is shown in the file list.
Only the parameters are saved from an instrument file; the history
data are not adopted.
55
Page 56
5 Parameterizing, adjusting and analyzing software (P2A software)
The original name (Instrument type, Serial number) is
suggested with the current date/time as standard, e.g.
"testo 6781 01234578 061120 1403.cfp".
For a standard installation, the files are saved under
"C:\Documents and Settings\All Users\Shared
Documents\P2A Software". The path can differ depending
Pos: 82 /TD/Produkt verwenden/MUF 63xx/P2A/Parame ter da tei öffnen @ 3\mod_1234358634164_79.doc @ 24341 @ 4
on the version of the operating system.
5.3.2.4. Opening the parameter file
All parameter files stored in the standard directory path are
automatically displayed in the file list when the software is started.
You can also open parameter files that are stored in other
directories.
1. Click on File > Open in the menu bar.
2. Select the storage location and click on the requisite file.
3. Click on [Open].
- The selected file is opened. This can be changed and saved
5 Parameterizing, adjusting and analyzing software (P2A software)
5.3.4.1. 1-point adjustment
A dew mirror hygrometer is recommended as the reference
measuring instrument for the 1-point adjustment (offset).
1. Expose the reference measuring instrument and the instrument
to be adjusted to the same constant conditions and wait for
equalization period to lapse.
2. Mark the instrument file of the connected instrument.
3. Click on [Adjusting the transmitter].
- The Properties of <Instrument type> <Serial number>
dialogue is opened with the Adjusting the transmitter register.
4. Enter reference value and click on [Carry out 1-point
adjustment].
5. Confirm confirmation request.
- The adjustment is carried out.
Field Explanation
°Ctd/°Ftd
Selection of the unit; only for dewpoint
adjustment.
°C/°F
Selection of the unit; only for temperature
adjustment.
Current reading Reading in the selected unit °Ctd/°Ftd or °C/°F.
Readings are updated every second.
61
Page 62
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
Reference value
Entry of the read-off value from the reference
measuring instrument.
Permissible entries:
• max. 10 °C
deviation (sum of all 1-point
td
adjustments)
• max. 2 K (°C) deviation (sum of all 1-point
adjustments)
> Click on [Set offset to zero] to reset a transmitted reference
Required field: Entry of the value read off at
the multimeter.
Parameterizations, adjustment processes and messages that have
occurred are registered in the transmitter with an operating hours
stamp.
In the history overviews (explained later in more detail), past
processes and events can be made visible.
63
Page 64
5 Parameterizing, adjusting and analyzing software (P2A software)
For parameter changes or adjustments that are performed
directly at the instrument (via the user menu), "Transmitter"
is entered in the User field and only the operating hour is
entered in the Date/time field instead of operating
hour/date/time.
For entries that are performed using the P2A software, the
name of the user logged into Windows appears in the User
field, while the date/time and operating hour are shown in
the Date/time field.
1. Mark the instrument file of the connected instrument.
2. Click on the [Transmitter history] button.
- The Properties of <Instrument type> <Serial number>
dialogue is opened with the Transmitter history register.
3. Click on the required entry in the list to change the display.
Field Explanation
64
Operating hours /
date/time
User
Comments
Operating hour/time stamp at which the
change at the instrument was performed.
Name with which the user is logged into the
operating system.
"Transmitter" entry if the change was
performed at the instrument.
Type of parameter change, e.g. "Unit of
channel 1 from % RH to Ctpd".
Page 65
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
Selection of adjustment histories: 1-point adjustments / Analog
adjustments.
Operating hours /
date/time
User
Operating hour/time stamp at which the
change at the instrument was performed.
Name with which the user is logged into the
operating system.
"Transmitter" entry if the change was
performed at the instrument.
Reference value Value read off at reference instrument.
Unit Unit during the adjustment.
Actual value before
adjustment
1-point adjustment: Reading of the
transmitter. If no changes were performed,
no value is displayed.
Offset from
1-point adjustment: Value before the
adjustment.
Offset to
1-point adjustment: Value after the
adjustment.
Channel Analog adjustment: Channel 1.
Specification Analog adjustment: Actual value.
Reading Analog adjustment: Entered reference value.
65
Page 66
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
Offset
Analog adjustment: Deviation at time of
adjustment.
The table is shown only for error and status messages that were
generated in the transmitter and were transferred and saved there
via the connection to the P2A software.
Operating hours
Operating hour at which the message
appeared in the instrument.
Serial number
Serial number of the instrument to which the
message is assigned
Message
e.g. "Scaling changed". The scaling was
changed.
Type of message e.g. early warning, status message.
> To print out the history data, click on [Print].
The printing job is automatically sent to the default printer
for the operating system.
With [Set up printer …] a different printer can be selected
or the printout can be edited.
4. Click on [OK] or [Cancel] to close the dialogue.
66
Page 67
Pos: 96 /TD/Überschriften/8. Tipps und Hilfe @ 0\mod_11 7378 9887985_79.doc @ 406 @ 1
6 Tips and assistance
Pos: 97 /TD/Überschriften/8.1 Fragen und Antworten @ 0\ mo d_11 77402017078_79.doc @ 1093 @ 2
6.1. Questions and answers
Pos: 98 /TD/Tipps und Hilfe/Fragen und Antworten/MUF 6781 @ 3\ mod_1239863529560_79.doc @ 31233 @
Question Possible causes/solution
Connection to instrument
Check connection cable/plug contacts
cannot be established
Reading is too high or too
low
• There is still no equilibrium
condition in the system.
• Clean sensor or sintered cap, see
Maintenance and cleaning page 41
A message appears on the
display
Malfunction (with and
without display)
See Status, warning and error
messages page 37
Analysis using the P2A software, see
Analyzing/testing the transmitter page
57
Undo adjustment
A 1-point temperature/humidity
adjustment can be reset to the current
readings using [Set offset to zero]
The actual values before the
adjustment can be read out from the
corresponding history table.
If we could not answer your question, please contact your dealer or
Testo Customer Service. For contact details see the rear side of
Pos: 99 /TD/Überschriften/8.2 Zubehör und Ersatztei le @ 0\mod_1177402058734_79.doc @ 1102 @ 2
this document or the web page www.testo.com/service-contact
6 Tips and assistance
6.2. Accessories and spare parts
Pos: 100 /TD/Tipps und Hilfe/Zubehör und Ersatzteile/ MUF63xx/Zubehör Ersatzteile MUF 678x @ 3\mod_123850429 9533_79.doc @ 30505 @
Description Article no.
Measurement chamber and cooling coil
Precision measurement chamber up to
0554 3312
35,000 hPa with adjustable inflow
Cooling coil for process temperatures above
0554 3304
50 °C (up to 200 °C), only with measurement
chamber
P2A software (parameterizing, adjusting,
analyzing) incl. USB adapter
Plug-in connections
M12 plug-in connection (socket) for power
and signal lines
Supply
Mains unit (desktop, wall-mounted) 0554 1748
Mains unit (top-hat rail mounting) 0554 1749
Hose
PTFE tube with compressed air connections,
2 m, up to 140 °C and max. 9000 hPa
External display
Process display for testo 54-2 AC.... 5400 7553
Process display for testo 54-7 AC.... 5400 7555
Calibration
Standard ISO calibration certificate for
dewpoint
(-40 °C
selectable points
Standard DKD calibration certificate for
dewpoint at -10 °C
to 0 °Ctp at 6000 hPa), freely
tp
and -40 °Ctp
tp
0554 3312
0554 3304
0554 6020
0554 6689
0699 2824/4
0520 0116
0520 0136
68
For a complete list of all accessories and spare parts, please refer
to the product catalogues and brochures or look up our website at:
www.testo.com
Page 69
6 Tips and assistance
Pos: 101 /TD/Tipps und Hilfe/Zubehör und Ersatzteile/ MUF63xx/Bestelloptionen MUF 6781 @ 3\mod_12392700482 66_79.doc @ 31074 @ 3
6.2.1. Ordering options for testo 6781 transmitter (0555
6781)
Order code Characteristic
Axx Process
connection
A01 Process connection G1/2
A02 Process connection NPT 1/2"
Bxx Analog
output/supply
B02 0 to 1 V (4-wire, 24 V AC/DC)
B03 0 to 5 V (4-wire, 24 V AC/DC)
B04 0 to 10 V (4-wire, 24 V AC/DC)
B05 0 to 20 mA (4-wire, 24 V AC/DC)
B06 4 to 20 mA (4-wire, 24 V AC/DC)
Cxx Display
C00 Without display
C02 With display/English
C03 With display/German
C04 With display/French
C05 With display/Spanish
C06 With display/Italian
C07 With display/Japanese
C08 With display/Swedish