Your benefits with the testo 6682 humidity transmitter
The new testo 6682 humidity transmitter for Ex applications meets the most
stringent requirements for process monitoring in explosive applications for
pharmaceutical, chemical and process engineering.
Process safety and system availability are supported by the excellent features of
teso 6682 which enthuse those working in the field:
• Highest accuracy and long-term stability of ±1% RH
• Preventive maintenance through self-monitoring and early warning
• Historical traceability thanks to internal log
• Flexible calibration concept
• Replaceable digital testo 6616 probe
• Display and operating menu for easy operation with parameterization,
adjustment and analysis
Testo’s humidity sensor: The core piece of high quality humidity transmitters
now also for Ex areas
Testo is the first choice when it comes to superior humidity transmitters for a
critical climate and drying processes. Using our many years of experience, the
sensor concept and signal processing have now been completely overhauled.
Accuracy and long-term stability were tested as part of 5 years of interlaboratory
tests by different national calibration laboratories (PTB, CETIAT, NIST etc.). Even
without a new adjustment, the ± 1% RH limit was not exceeded.
You are on the safe side thanks to the accuracy, stability and efficiency of the
Testo humidity transmitter.
Top electrode
Bottom electrode
Substrate
Connections
Dielectric layer
Polymer, changes
permittivity continuously as
a function of relative
humidity
Allows humidity to
penetrate to the
dielectric layer and
protects against dirt
Ceramic substrate for
mechanical protection
Special anti-
corrosion design
II Instrument group: Use in non-firedamp endangered areas
1/2 Instrument category: Application range Electronics Category 2,
Application range Probe shaft Category 1
2 (1) Instrument category: Instrument can be used in Category 2 and facilitates
the connection of instruments which can be used in Category 1
G Media: For explosive mixtures of air and combustible gases, vapours or
mist
Ex Explosion-proof electrical equipment in accordance with European
standards
ia Ignition protection type: Intrinsic safety
[ia] [ia] The instrument provides an intrinsic electric circuit for an additional
instrument
IIC The instrument can be used for gases in the IIC gas group
T4/T3 Temperature class T4 or T3
Ex pprotection cclasses
The testo 6682 humidity transmitter is intended
for protection class
ATEX III 22 ((1) GG EEx iia [[ia] IIIC
T4
, the digital testo 6616 humidity probe for protection
class
ATEX III 11/2 GG EEx iia IIIC TT4/T3
.
The sensor tip can be used in a Zone 0 explosive area,
the transmitter in Zone 1.
Basic safety and health requirements are fulfilled by
compliance with EN 60079-0:2006, EN 60079-11:2007
as well as EN606079-26:2007 for testo 6616.
A
A
Explosive area, Zone 0
B
Explosive area, Zone 1
C
Non-explosive area
B
C