TESTO 6681 Profibus Instruction Manual

testo 6681 Profibus · Humidity transmitte
r
with Profibus module
testo 6610 · Probes
P2A software · Parameterizing, adjusting and analyzing
Instruction manual Volume 1 en
Safety and the environment
3
Safety and the environment
Avoiding electrical hazards
Never use the instrument and connected probes to measure on or near live
parts!
Damaged mains cables must only be replaced by authorized personnel.
Only have the transmitter wired and connected by authorized personnel with
the voltage disconnected.
You must always comply with the regulations applicable in your country to the
opening and repair of electrical equipment.
Avoiding personal injury/damage to equipment
Installation, setting and calibration work must only be carried out by qualified
and authorized personnel!
Only open the instrument when this is expressly described in the instruction
manual for installation, maintenance or repair purposes.
Observe the permissible storage, transport and operating temperatures.
Never store or operate the product with solvents and do not use any
desiccants.
Do not use the instrument for control purposes at the same time as operating
or servicing the transmitter.
Only operate the product properly, for its intended purpose and within the
parameters specified in the technical data. Do not use force.
Carry out only the maintenance and repair work that is described in the
documentation. Follow the prescribed steps when doing so. Use only OEM spare parts from Testo.
Any additional work must only be carried out by authorized personnel. Otherwise Testo will not accept any responsibility for the proper functioning of the instrument after repair and for the validity of certifications.
Protecting the environment
Send the product back to Testo at the end of its useful life. We will ensure
that it is disposed of in an environmentally friendly manner.
4
About this document
About this document
Please read this documentation through carefully and familiarize yourself with
the product before putting it to use. Keep this document to hand so that you can refer to it when necessary. Hand this documentation on to any subsequent users of the product.
The following conventions are followed in this document:
Character/ display:
Explanation/example
With the signal word Warning!:
Warns against hazards which could result in serious physical injury if the precautionary measures indicated are not taken, e.g.:
De-energize the mains connection before connecting the transmitter!
With the signal word Caution!:
Warns against hazards which could result in minor physical injury or damage to equipment if the precautionary measures indicated are not taken, e.g.:
Observe the permissible operating temperature!
Important information, e.g.:
Administrator rights are required to install the program under Windows
®
2000, XP and Vista.
Aim of action, e.g.:
Assembling the instrument at the process
connection:
Requirement that must be met, e.g.:
USB drivers are installed.
About this document
5
Character/ display:
Explanation/example
1
Steps are numbered if a certain sequence of actions must be followed, e.g.:
1 Loosen and remove housing screws. 2 Remove the upper part of housing.
A step is not numbered if there are no further steps or if the step is optional, e.g.:
Insert probe connector into socket of testo 6681 until it
engages.
" to "
Example entries are in inverted commas, e.g.:
The value "0" results in ....
Bold type
Elements of the program interface or instrument display, e.g.:
The instrument designation appears in the instrument/parameter file list.
Select Main Menu Channel 1 and confirm with SET.
…>…
Functions/paths within a menu, e.g.:
Start > All Programmes > Testo > P2A Software.
[ ]
Buttons which start an action, e.g.:
Confirm the software key with [OK].
CAPITAL LETTERS
Keys on the instrument or keypad, e.g.:
Press ESC.
6
Contents
Contents
1 TRANSMITTER ............................................................................................ 9
1.1 Specifications ........................................................................................... 9
1.1.1 Functions and use ...................................................................... 9
1.1.2 Scope of delivery ........................................................................ 9
1.1.3 Accessories............................................................................... 10
1.1.4 Technical Data .......................................................................... 10
1.1.5 Dimensions ............................................................................... 12
1.2 Product description ................................................................................ 13
1.2.1 At a glance ................................................................................ 13
1.2.2 Usable probes ........................................................................... 14
1.2.3 Display and keypad .................................................................. 15
1.2.4 Service interface ....................................................................... 15
1.2.5 Relay board (option) ................................................................. 15
1.2.6 Analog outputs .......................................................................... 16
1.2.7 Parameters ............................................................................... 16
1.2.8 Scaling ...................................................................................... 17
1.2.9 Alarm handling .......................................................................... 19
1.3 Commissioning ...................................................................................... 20
1.3.1 Insert Profibus module (Order no. 0554 6686) ......................... 20
1.3.2 Assembling the instrument ....................................................... 22
1.3.3 Connecting the instrument ........................................................ 25
1.3.4 Connect instrument to the bus system ..................................... 42
1.3.5 Configure Profibus module using EDD ..................................... 49
1.3.6 Adjusting the instrument ........................................................... 55
1.4 Operation ............................................................................................... 66
1.4.1 Relationship between user menu and mini-DIN socket is
active ......................................................................................... 66
1.4.2 Key cover .................................................................................. 67
1.4.3 Password protection ................................................................. 68
1.4.4 Structure of user menu ............................................................. 69
1.4.5 Overview of the testo 6681 user menu ........................................... 70
1.4.6 The individual main menus ....................................................... 72
Contents
7
1.5 Status, warning and error messages ..................................................... 84
1.5.1 Status messages ...................................................................... 84
1.5.2 Warning messages ................................................................... 85
1.5.3 Transmitter error messages ...................................................... 87
1.5.4 Status codes in cyclical service ................................................ 88
1.5.5 Handling alarm messages ........................................................ 88
1.5.6 Namur fault conditions .............................................................. 90
1.6 Maintenance and cleaning ..................................................................... 91
1.6.1 Maintaining the instrument ........................................................ 91
1.6.2 Cleaning the instrument ............................................................ 91
8
Certificate
testo 6681 Profibus - 1.1 Specifications
9
1 Transmitter
1.1 Specifications
1.1.1 Functions and use
The testo 6681 humidity transmitter with Profibus module is used in conjunction with plug-in, adjusted probes from the testo 6610 range.
Please refer to volume 2, chapter 2, page 97 for information about commissioning, operating and maintaining the testo 6610 probe.
The testo 6681 humidity transmitter is suitable for the following applications with Profibus networking, for example:
Process instrumentation Clean rooms Test benches Drying processes Production and storage air quality Complex room climate applications.
1.1.2 Scope of delivery
The scope of delivery of the testo 6681 humidity transmitter includes the following:
Key cover Rear panel bracket
10
testo 6681 Profibus - 1.1 Specifications
1.1.3 Accessories
The following accessories are available for the testo 6681 humidity transmitter:
Protection caps for probes Mains unit P2A software (parameterizing, adjusting and analyzing software) Assembly accessories.
Information about accessories and their order numbers can be found in volume 2, chapter 4.2, page 167 or on the website at www.testo.com.
1.1.4 Technical Data
Parameters
Humidity (various variables and units) Temperature (°C/°F)
Measuring range
Depends on probe
Accuracy
Depends on probe
Resolution
0.1 % RH or 0.1 °C/0.1 °F
Meas. cycle
1/s
Interface
Mini-DIN for P2A software
(parameterizing and adjusting software) and handheld testo 400/650
Optional Profibus-DP module
Voltage supply
4-wire (separate signal and supply
lines): 20 - 30 V AC/DC, 300 mA power consumption
Maximum load
- 4-wire: 500  (power output)
Analogue output*
0 to 1 V ± 1.5 mV (4-wire) or 0 to 5 V ± 7.5 mV (4-wire) or 0 to 10 V ± 15 mV (4-wire) or 0 to 20 mA ± 0.03 mA (4-wire) or 4 to 20 mA ± 0.03 mA (4-wire)
Resolution of analog output
12 bit
Relay
4 relays, 250 V AC/DC, 3 A (optional)
testo 6681 Profibus - 1.1 Specifications
11
Display
2-line LCD with plain text line
(optional)
Housing operating temperature
-40 to +70 °C/-40 to +158 °F, with display from 0 to 50 °C/+32 to +122 °F
With integrated relay: -40 - +60 °C
Storage temperature
- 40 to 80 °C/-40 to +176 °F
Housing, weight
Metal: 1.960 kg Profibus: 0.610 kg
Protection class
IP 65 only if the transmitter is wired
and/or seal plugs are inserted
Directives, standards and tests
EC Directive: 2004/108/EC
Warranty
Duration: 2 years Warranty conditions: See web page
www.testo.com/warranty
12
testo 6681 Profibus - 1.1 Specifications
1.1.5 Dimensions
Dimensions in mm a b
With M 20 cable couplings 144 147
With NPT cable couplings 144 144
With M plug-in connections 143
testo 6681 Profibus - 1.2 Product description
13
1.2 Product description
1.2.1 At a glance
1 Keys (with optional display) 2 Service flap screw
connection (self-locking,
2x) 3 Display (optional) 4 Service flap 5 M 16 x 1.5 screw
connection*, e.g. analog
outputs 6 M 16 x 1.5 screw
connection*, e.g. voltage
supply 7 Eyelet for measuring point
panel 8 M 20 x 1.5 screw
connection*, e.g. relay R 3
and R 4 9 Earthing/PE connection 10 M 20 x 1.5 screw
connection*, e.g. relay R 1
and R 2 11 Probe connector
(testo 6610) 12 Upper part of housing
* Alternatively, NPT cable
couplings or M plug-in connections are available
A
Profibus module
13 Rotary encoder switch for
addressing 14 Profibus M12 socket 15 Profibus M12 connector 16 LED: Bus error 17 LED: System error 18 LED: Supply
14
testo 6681 Profibus - 1.2 Product description
19 Housing screws (4x) 20 Socket for probe connector 21 Hole for fastening to rear
panel bracket (M3 x 6 screw)
22 Plastic bracket for assembly
on rear panel
22
21
1.2.2 Usable probes
The testo 6681 humidity transmitter can be used with the following probes:
Probe Article no. Characteristic
testo 6611 0555 6610-L11 Wall probe version; accuracy to ± 1 % RH;
temperature range -20 °C to +70 °C/-4 to +158 °F, sensor plugged
testo 6612 0555 6610-L12 Duct probe version; accuracy to ± 1 % RH;
temperature range -30 °C to 150 °C/-22 to +302 °F, sensor soldered/optionally plugged
testo 6613 0555 6610-L13 Cable probe version; accuracy to ± 1 % RH;
temperature range -70 °C to +180 °C/-94 to +356 °F, sensor soldered/optionally plugged
testo 6614 0555 6610-L14 Heated cable probe version; accuracy to ± 1.0 %
RH; temperature range -40 °C to +180 °C/
-40 to +356 °F, sensor soldered
testo 6681 Profibus - 1.2 Product description
15
Probe Article no. Characteristic
testo 6615 0555 6610-L15 Trace humidity cable probe version;
accuracy ± 6 K at -60 °Ctd; temperature range -40 °C to +120 °C/-40 to +248 °F, sensor soldered
testo 6617 0555 6610-L17 Cable with cover electrode monitoring probe
version; accuracy to ± 1.2 % RH; temperature range
-40 °C to +180 °C/-40 to +356 °F, sensor soldered
1.2.3 Display and keypad
The display option allows the testo 6681 humidity transmitter to be operated via the display and four keys.
The LCD display consists of two 7-segment lines for displaying readings and units and of an information line (for status messages, for example).
The brightness and contrast of the display and the background lighting (permanent or off) can be changed via the user menu or the P2A software.
1.2.4 Service interface
Behind the service flap is the parameterizing socket (mini-DIN) as an interface to the P2A software or Testo handheld instrument (testo 400/testo 650).
1.2.5 Relay board (option)
This has a floating switch capacity of 250 V AC/3 A. The switching limits and hysteresis can be set via the display or the P2A software or via the Profibus. Further features include:
Function of changeover contacts (NC/NO contacts) freely selectable 12 terminals for a total of 4 relays.
If no relays are available, settings for monitoring limit values or alarms can still be controlled via the display. The alarm status will be shown on the display.
Only have the transmitter wired and connected by authorized personnel with the voltage disconnected.
16
testo 6681 Profibus - 1.2 Product description
1.2.6 Analog outputs
For analog outputs, the testo 6681 has either
2 or optionally 3 current outputs of 4 to 20 mA/0 to 20 mA (4-wire) or 2 or optionally 3 voltage outputs of 0 to 1 V/0 to 5 V/0 to 10 V (4-wire).
The transmitter can be ordered with a third analog output as an option.
If the Profibus module is integrated directly into the testo 6681 humidity transmitter (order code B77), it has two current outputs 4 ...20 mA.
1.2.7 Parameters
The following parameters are displayed:
Relative humidity in %RH (technical) Relative humidity in % WMO* (calculation according to the WMO
standard)
Temperature °C and °F Dewpoint in °Ctd and °Ftd Absolute humidity in g/m³ and gr/ft³ Degree of humidity in g/kg and gr/lb Enthalpy in kJ/kg and BTU/lb Psychrometer temperature in °Ctw and °Ftw Water vapour partial pressure in hPa and "H
2
O (inch H2O)
Water content in ppm
vol
and % vol
Dewpoint of H
2O2
mixture in °Ctm and °Ftm
Mixture humidity H
2O2
in %RHm
*It is possible that condensation appears as of a displayed humidity starting from 70 % and is shown on the display. This unit is used in meteorology, for example. According to the WMO, the Magnus coefficient is used for sub-cooled water in the calculation of relative humidity.
Calculated humidity variables correspond to the medium of air. With other gases/gas compositions, deviations may occur, e.g. with the enthalpy
testo 6681 Profibus - 1.2 Product description
17
1.2.8 Scaling
There are three types of min./max. values:
1 The measuring range
The maximum sensor performance is in this range. Values outside of the measuring range are displayed via messages, for example. Refer to table (see below) for the measuring range.
2 Standard scaling
The output signals are assigned to this measuring range as standard:
- during delivery if no entries are made in the order code
- after exchanging the unit, the measuring range recorded in the instrument is applied as standard.
The transmitter even retains its scaling with the voltage disconnected.
Refer to table (see below) for the measuring range.
3 The maximum settings for the manual scaling
- the values are not expressly given in the table. The maximum limits can be calculated as follows:
X = difference between MIN. and MAX. value of the standard scaling
(Max. value of standard) + (50 % of X)
(Min. value of standard) - (50 % of X)
- It is thus possible to scale beyond the measuring range, e.g. for the adjustment of the scaling limits to standard values of a PLC. However, the physical measurement range limits are the deciding factor when defining alarms.
18
testo 6681 Profibus - 1.2 Product description
Parameter Unit Probe
Measuring
range
Physical
at 1013 hPa
Standard
scaling
MUF
measuring
range
MIN MAX MIN MAX
Temperature °C 6611 -20
+70
-20 +70
°F 6611 -4 +158
-4
+158
°C 6612 -30 +150
-30 +150
°F 6612 -22 +302
-22 +302
°C 6613 -70 +180
-70 +180
°F 6613 -94 +356
-94 +356
°C 6614, 6617 -40 +180
-40 +180
°F 6614, 6617 -40 +356
-40 +356
°C 6615 -40 +120
-40 +120
°F 6615 -40 +248
-40 +248
relative humidity % RH 0 +100
0
+100
WMO relative humidity % RH 0 +100
0
+100
Mixture humidity H2O2 % RHm 0 100
0
100
Dewpoint °Ctd 6611 -20
+70
-80 +100
°Ftd 6611 -4 +158 -112 +212
°Ctd
6612, 6613, 6614, 6617
-20
+100
-80 +100
°Ftd
6612, 6613, 6614, 6617
-112 +212 -112 +212
°Ctd 6615 -60
+30
-80 +100
°Ftd 6615 -148 +212 -112 +212
Mixture dewpoint (H2O2) °Ctm -20 +100
-20 +100
°Ftm -4 +212
-4
+212
Absolute humidity g/m3 all probes 0
600
0
2000
gr/ft3 0
250
0
800
Degree of humidity g/kg all probes 0 13300
0
9500
gr/lb 0 93000
0
66500
Enthalpy kJ/kg -40 99999
-40 8000
BTU/lb -18 43000
-18 3500
Psychrometer temperature °Ctw -40
100
-40 180
°Ftw -58
210
-40 356
testo 6681 Profibus - 1.2 Product description
19
Parameter Unit Probe
Measuring
range
Physical
at 1013 hPa
Standard
scaling
MUF
measuring
range
MIN MAX MIN MAX
Water content
ppm vol H2O 0 99999
0
99999
% vol 0
100
0
100
Water vapour partial pressure hPa 0 1000
0
7000
inchH2O 0
400
0
2800
1.2.9 Alarm handling
For upper and lower alarm limits, individual alarms as well as collective alarms can be specified. If the collective alarm function is activated, an alarm is triggered as soon as the alarm limit of an alarm is exceeded, if this alarm is assigned to the collective alarm.
The testo 6681 monitors limit values using relays. If a reading is outside the limit values, a relay to be specified by the user is switched.
If the reading reverts to more than a specified hysteresis below or above the limit value, the alarm is cancelled.
In addition, information about the occurrence of error/status messages can be provided by means of a collective alarm relay, see chapter 1.5, Status, warning and error messages, page 84.
If multiple alarm messages are activated at the same time, the last alarm is shown. If the alarm is cancelled again, the previous messages are no longer shown.
Example:
If the condensation of the probe begins, the "Condensation" message appears on the display and the "Start" status display. If the condensation is over, the status display changes from "Start" to "End".
20
testo 6681 Profibus - 1.3 Commissioning
1.3 Commissioning
1.3.1 Insert Profibus module (Order no. 0554 6686)
The Profibus module can be ordered retroactively as an accessory. It can easily be installed in the testo 6681 humidity transmitter.
The probe connector must be disconnected.
4 Loosen screw connection (1) of service flap and open the flap.
5 Loosen and remove housing screws (2).
6 Remove upper part of housing (3) and place on a clean surface.
testo 6681 Profibus - 1.3 Commissioning
21
7 Place Profibus module (A) on lower part of instrument (4).
If necessary, manually set the address using the rotary encoder switches (see chapter 1.3.3.6, page 37 before fixing the instrument in place.
8 Set on upper part of instrument (5) and fix in place using housing screws
(2).
22
testo 6681 Profibus - 1.3 Commissioning
1.3.2 Assembling the instrument
1.3.2.1 Wall mounting
(for testo 6611/6613/6614/6615/6617 probes)
Attaching rear panel bracket
1 Remove locking screw (item (4) of drawing on page 21) and detach rear
panel bracket from plastic bracket (item (2) of drawing on page 21).
2 Hold rear panel bracket in assembly position and mark the three drill
holes.
3 Drill three holes ( 5 mm) and insert dowels where necessary.
4 Screw on rear panel bracket.
Remember that the clamping brackets (1) must face the wall.
testo 6681 Profibus - 1.3 Commissioning
23
Fastening instrument to rear panel bracket
2
3
4
5
1 Slide plastic bracket (2) on the back of instrument onto rear panel bracket
until it engages (see arrows).
2 Insert screw (4) through hole (3) and screw into rear panel bracket.
3 Insert probe connector (5) into socket until it engages.
24
testo 6681 Profibus - 1.3 Commissioning
1.3.2.2 Duct mounting (for testo 6612 probes)
3
10
5
1
4
11
2
6
9
12
7
8
1 Hold wall/duct bracket (order no. 0554 6651) (6) against duct wall (8) and
mark drill holes for wall/duct bracket and probe shaft.
2 Drill a hole ( 12,5 mm) in the duct wall to feed through the probe shaft.
3 Fasten wall/duct bracket (6) to duct wall with screws (5).
4 Push probe shaft (9) with filter (10) through the middle hole of the
mounting bracket.
The wall/duct bracket (6) has an O-ring (7) to seal it against the duct. Feed the probe shaft (9) carefully through the wall/duct bracket so that the O-ring is not damaged.
5 Fix the correct position of the probe shaft (9) with screw (11) and mark
(insert probe shaft as far as possible).
testo 6681 Profibus - 1.3 Commissioning
25
6 Slide plastic bracket (2) on the back of the transmitter onto bracket (3, 4)
until it engages.
Take the weight of the transmitter into account. Ensure that the brackets (4, 6) are fastened securely.
7 Insert screw (1) through hole on the top of the instrument and screw into
bracket (3).
8 Insert probe connector (12) into socket until it engages.
1.3.3 Connecting the instrument
Opening the instrument
1 Loosen screw connection (1) of service flap and open the flap.
26
testo 6681 Profibus - 1.3 Commissioning
2 Loosen and remove housing screws (2).
Important.
The Profibus module (A) is already separated from the upper and lower parts of the instrument when the housing screws
(2) are removed.
3 Remove upper part of housing (3) and place on a clean surface.
4 Remove Profibus module (A) from lower part of housing (4) and also
place on a clean surface.
testo 6681 Profibus - 1.3 Commissioning
27
Warning!
Electrical voltage.
Danger of short-circuit! De-energize the mains connection before connecting the transmitter!
Only have the transmitter wired and connected by authorized personnel with the voltage disconnected.
28
testo 6681 Profibus - 1.3 Commissioning
1.3.3.1 Overview of terminals
1 Lower part of housing 2 Relay board (option) 3 Relay terminals 4 Insulating trough for relay
board
5 Terminal strip for voltage
supply and analog outputs*
* The transmitter testo 6681 with
integrated Profibus module (order code B 77) has two current outputs 4 to 20 mA.
6 Terminal board
7 Earthing terminal (internal) 8 M 16 x 1.5 screw
connection** 9 Earthing terminal (external) 10 M 20 x 1.5 screw
connection* 11 Eyelet for measuring point
panel
** Alternatively, NPT cable coupling
or M plug-in connection.
The following description of the terminals refers to this overview and its numbering.
Loading...
+ 64 hidden pages