TESTO 6610 Instruction Manual

Page 1
testo 6383 · differential pressure transmitter
testo 6610 · Probes P2A software · Parameterizing, adjusting and analyzing
software
Instruction manual Volume 1
Page 2
Page 3
1 Safety and the environment
3
Pos: 1 /TD/Überschriften/MUF/Sicher heit und Umwelt @ 3\mod_1234793958627_79. doc @ 26223 @ 1
1 Safety and the environment
Pos: 2 /TD/Sicherheit und Umwelt/Sicher heit gewährleisten/MUF 63xx/Elektr i sche Gefahren vermeiden @ 3\mod_12347946092 99_79.doc @ 26280 @ 5
Avoiding electrical hazards
> Never use the instrument and connected probes to measure on
or near live parts!
> Damaged mains cables must only be replaced by authorize d
personnel.
> Only have the transmitter wired and connected by authorized
personnel with the voltage disconnected.
> You must always comply with the regulations applicable in your
country for opening and repairing electrical equipm ent.
Pos: 3 /TD/Sicherheit und Umwelt/Sicher heit gewährleisten/MUF 63xx/Personen- und Sachschäden vermeiden @ 3\mod_1234794744 768_79.doc @ 26299 @ 5
Avoiding personal injury and damage to equipment
> Installation, setting and calibration work must only be carried
out by qualified and authorized personnel!
> Only open the instrument when this is expressly described in
the instruction manual for installation, maintenance or repair purposes.
> Observe the permissible storage, transport and operating
temperature.
Pos: 4 /TD/Sicherheit und Umwelt/Sicher heit gewährleisten/Nicht mit Lösungsmi tteln lagern @ 0\mod_1175692375179_79.d oc @ 583 @
> Do not store the product together with solvents. Do not use any
desiccants.
Pos: 5 /TD/Sicherheit und Umwelt/Sicher heit gewährleisten/MUF 63xx/Bei W ar tung MUF nicht zur Regelung verwenden @ 3\mod_123479 48523 77_79.doc @ 26318 @
> Do not use the instrument for control purposes at the same time
as operating or servicing the transmitter.
Pos: 6 /TD/Sicherheit und Umwelt/Sicher heit gewährleisten/Produkt bestim mun gsg emäß verwenden @ 0\mod_1173781261848_79. doc @ 386 @
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not use any force.
Pos: 7 /TD/Sicherheit und Umwelt/Sicher heit gewährleisten/Nur beschriebene W artungsarbeiten durchführen @ 0\mod_11756 92705195_79.doc @ 601 @
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the prescribed steps exactly. Use only original spare parts from Testo.
Pos: 8 /TD/Sicherheit und Umwelt/Sicher heit gewährleisten/MUF 63xx/Fachperson al @ 3\mod_1234794940409_79.doc @ 26337 @
Any additional work must only be carried out by authorized personnel. Otherwise testo will not accept any responsibility for the proper functioning of the instrument after repair and for the validity of certifications.
Page 4
2 About this document
Pos: 9 /TD/Überschriften/MUF/Umwelt sc hü tzen @ 3\mod_1234858757571_79.doc @ 26363 @ 5
Protecting the environment
Pos: 10 /TD/Sicherheit und Umwelt/Umwel t schützen/Produkt entsorgen @ 0\mod_1173780 307072_79.doc @ 357 @
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local regulations) or return the product to Testo for disposal.
Pos: 11 /TD/Überschriften/MUF/Zu diese m Dokument @ 3\mod_1234793991331_79.doc @ 26242 @ 1
2 About this document
Pos: 12 /TD/Sicherheit und Umwelt/Zu diese m Dokument/Verwendung (Standard) @ 0\ mod_1173775068554_79.doc @ 337 @ 5
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay particular attention to the safety instructions and warning advice in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.
Pos: 13 /TD/Sicherheit und Umwelt/Zu diese m Dokument/Symbole und Schreibkonventi one n/Warnhinweis WARNUNG @ 2\mod_1207646966234_7 9.doc @ 14398 @
WARNING
Indicates potential serious injuries
Pos: 14 /TD/Sicherheit und Umwelt/Zu diese m Dokument/Symbole und Schreibkonventi one n/Warnhinweis VORSICHT @ 2\mod_1207651 416515_79.doc @ 14416 @
CAUTION
indicates potential minor injuries
Pos: 15 /TD/Sicherheit und Umwelt/Zu diese m Dokument/Symbole und Schreibkonv. Sof t ware [Standard] @ 0\mod_1190203332543_79.d oc @ 4883 @ 5
Symbols and writing standards Representa-
tion
Explanation
Note: Basic or further information.
1. ...
2. ...
Action: more steps, the sequence must be followed.
> ... Action: a step or an optional step.
- ... Result of an action.
Menu
Elements of the program interface.
[OK]
Buttons of the program interface. ... | ... Functions/paths within a menu. “...” Example entries
Pos: 16 /TD/--- Seitenwechsel --- @ 0\mod_1173774430601_0.doc @ 283 @
Page 5
3 Contents
5
Pos: 17 /TD/Überschriften/MUF/Inhal t @ 3\mod_1234794019831_79.doc @ 26261 @ 1
3 Contents
1 Safety and the environment....................................................................3
2 About this document...............................................................................4
3 Contents...................................................................................................5
4 Transmitter...............................................................................................7
4.1. Specifications ..................................................................................7
4.1.1. Functions and use ...........................................................................................7
4.1.2. Scope of delivery .............................................................................................7
4.1.3. Accessories .....................................................................................................7
4.1.4. Technical data .................................................................................................8
4.1.5. Dimensions....................................................................................................12
4.2. Product description........................................................................13
4.2.1. At a glance.....................................................................................................13
4.2.2. Usable probes................................................................................................14
4.2.3. Display and keypad........................................................................................15
4.2.4. Service interface............................................................................................15
4.2.5. Relay board (option) ......................................................................................15
4.2.6. Analog outputs...............................................................................................16
4.2.7. Parameters....................................................................................................16
4.2.8. Scaling .........................................................................................................17
4.2.9. Alarm handling...............................................................................................19
4.3. Commissioning..............................................................................20
4.3.1. Mounting preparations ...................................................................................20
4.3.2. Connecting the instrument.............................................................................21
4.3.2.1. Overview of terminals........................................................................22
4.3.2.2. Connecting voltage supply and analog outputs.................................23
4.3.2.3. Connecting the relay outputs ............................................................24
4.3.2.4. Closing the instrument......................................................................27
4.3.2.5. Adjusting the instrument....................................................................29
4.3.2.6. Overview: Adjustment keys and test contacts...................................30
4.3.2.7. 1-point adjustment (offset - humidity/temperature)............................30
4.3.2.8. Analog output adjustment .................................................................32
4.3.2.9. n-point adjustment (pressure)...........................................................33
4.3.2.10. High-humidity adjustment for testo 6614....................................34
4.3.2.11. Self adjustment of testo 6615 trace humidity probe....................35
4.4. Operation.......................................................................................36
4.4.1. Relationship between user menu and mini-DIN socket is active....................36
4.4.2. Password protection ......................................................................................36
4.4.3. Structure of user menu ..................................................................................37
4.4.4. Overview of the testo 6383 user menu...........................................................38
4.4.5. The individual main menus ............................................................................41
4.4.5.1. Editing main menu of channel 1........................................................41
Page 6
3 Contents
4.4.5.2. Editing Main Menu Channel 2 (if this option is available)..................41
4.4.5.3. Editing Main Menu Channel 3 (if this option is available)..................42
4.4.5.4. Editing Main Menu Alarm..................................................................42
4.4.5.5. Editing Main Menu Settings..............................................................44
4.4.5.6. Editing Main Menu Analysis..............................................................47
4.4.5.7. Editing Message main menu.............................................................49
4.4.5.8. Calling up Main Menu Ident..............................................................50
4.4.5.9. Editing Main Menu Adjust.................................................................50
4.4.5.10. Editing Reset main menu...........................................................52
4.5. Status, warning and error messages ............................................53
4.5.1. Status messages...........................................................................................54
4.5.2. Warning messages........................................................................................55
4.5.3. Transmitter error messages ...........................................................................56
4.5.4. Handling alarm messages .............................................................................58
4.5.5. Namur fault conditions...................................................................................59
4.6. Maintenance and cleaning............................................................ 61
4.6.1. Maintaining the instrument.............................................................................61
4.6.2. Cleaning the instrument.................................................................................61
Pos: 18 /TD/--- Seitenwechsel --- @ 0\mod_1173774430601_0.doc @ 283 @
Page 7
4 Transmitter
7
Pos: 19 /TD/Überschriften/MUF/1 Messu mfor mer @ 3\mod_1234258401060_79.doc @ 23894 @ 1
4 Transmitter
Pos: 20 /TD/Überschriften/MUF/1.1/2. 1/ 3.1 Leistungsbeschreibung @ 3\mod_123 4258595211_79.doc @ 23951 @ 2
4.1. Specifications
Pos: 21 /TD/Leistungsbeschreibung/Ver wendung/MUF63xx/MUF 6383,84 @ 4\mod_125 1793 491331_79.doc @ 47763 @ 3
4.1.1. Functions and use
The testo 6383 transmitter is suitable for the following applications, amongst others:
• Clean rooms
Complex room climate applications
Pos: 22 /TD/Leistungsbeschreibung/Li eferumfang/MUF 63xx/MUF 6383, 6385 @ 4\mod_1 254749193798_79.doc @ 51251 @ 3
4.1.2. Scope of delivery
The scope of delivery of the testo 6383 transmitter includes the following:
With KMAT version D04: Integrated humidity probe
• Sealing frame
Instruction manual
• Calibration report
CD-ROM with operating instructions (PDF), configuration files for Ethernet module and P2A update (this can only be used in conjunction with the P2A software, which has to be ordered separately).
Pos: 23 /TD/Leistungsbeschreibung/Li eferumfang/MUF 63xx/Zubehör Übersic ht 638x @ 3\mod_1234448071530_79.doc @ 25136 @ 3
4.1.3. Accessories
The following accessories are available for the testo 6383 transmitter, amongst others:
Protection caps for probes
• Mains unit
P2A software (parameterizing, adjusting and analyzing software)
• Assembly accessories
Information about accessories and their order numbers can be found in volume 2, Accessories and spare parts or on the website at www.testo.com.
Page 8
4 Transmitter
Pos: 24 /TD/Leistungsbeschreibung/ Technische Daten/MUF 63xx/MUF 6383 @ 4\mod_12 5249 9023732_79.doc @ 48869 @ 355555555555555 5555 55555
4.1.4. Technical data
Parameters
• Differential pressure
• Temperature
• Humidity
Differential pressure accuracy
The specifications are only valid if the positive pressure is applied at the positive pressure connection.
0.3 % of measuring range, additional 0.3 Pa intrinsic error 1
• T
K slope drift
= 0.02 % of measuring range per degree Kelvin of
deviation from nominal temperature 22 °C
• T
K zero point drift
= 0 %, as zeroing with solenoid valve2
Humidity and temperature accuracy
Depends on probe
Humidity and temperature measuring range
Depends on probe
1
Measuring uncertainty in accordance with GUM: ±0.5 % of measuring range
final value ±0.3 Pa.
GUM (Guide to the Expression of Uncertainty in Measurement): ISO guideline for determining the measuring uncertainty in order to render global measurement results comparable.
The following uncertainties are used during the inquiry:
• Hysteresis
• Linearity
• Reproducibility
Adjustment area/factory calibration
• Test location
2
Minor mixtures of the media may occur at the positive and negative pressure
side due to the automatic zeroing cycle.
Page 9
4 Transmitter
9
Pressure measuring range, resolution and overload of differential pressure
Pressure measuring range depending on version ordered
Resolution Overload
0 to 50 Pa 0.1 Pa 20,000 Pa 0 to 50 Pa 0.1 Pa 20,000 Pa 0 to 100 Pa 0.1 Pa 20,000 Pa 0 to 500 Pa 0.1 Pa 20,000 Pa 0 to 10 hPa 0.01 hPa 200 hPa
-10 to 10 Pa 0.1 Pa 20,000 Pa
-50 to 50 Pa 0.1 Pa 20,000 Pa
-100 to 100 Pa 0.1 Pa 20,000 Pa
-500 to 500 Pa 0.1 Pa 20,000 Pa
-10 to 10 hPa 0.01 hPa 200 hPa
Upon delivery and following a factory reset the readings are shown in the display in the unit that was ordered via the KMAT option Fxx, see Ordering options for testo 6383 transmitter (0555 6383), page 133.
Humidity and temperature resolution
0.1 % RH or 0.01 °C/0.01 °F
Meas. cycle
• 1/sec
Interface
Mini-DIN for P2A software (parameterizing and adjusting software) and handheld testo 400/650
optional: Ethernet interface
Voltage supply
4-wire (separate signal and supply lines): 20 to 30 V AC/DC, 300 mA power consumption
Maximum load
• 4-wire: 500 Ω (power output)
Page 10
4 Transmitter
10
Maximal load
4-wire: 10 k (voltage output)
Analog output
0 to 1 V ± 1.5 mV (4-wire) or
0 to 5 V ± 7.5 mV (4-wire) or
0 to 10 V ± 15 mV (4-wire) or
0 to 20 mA ± 0.03 mA (4-wire) or
4 to 20 mA ± 0.03 mA (4-wire)
Resolution of analog output
• 12 bit
Relay
4 relays, 250 V AC/DC, 3 A (optional)
Display
2-line LCD with plain text line (optional)
Operating temperature
-5 to 50 °C/23 to 122 °F
Page 11
4 Transmitter
11
Storage temperature
-20 to 60 °C/-4 to 140 °F
Process temperature
-20 to 65 °C/-4 to 149 °F
Oper. humidity
0 to 90 % RH
Housing, weight
• Metal/plastic
Version with integrated humidity probe: approx. 1.35 kg
Version with preparation for external humidity probe: approx.
1.26 kg
Protection class, frontal
IP 65 only if the transmitter is wired and/or sealing plugs are inserted
Directives, standards and tests
EC Directive: 2004/108/EC
Warranty
• Duration: 2 years
Warranty conditions: see website www.testo.com/warranty
Page 12
4 Transmitter
12
Pos: 25 /TD/Leistungsbeschreibung/ Technische Daten/MUF 63xx/MUF Panel lang Ab messungen @ 4\mod_1255092618418_79.doc @ 51743 @ 3
4.1.5. Dimensions
Page 13
4 Transmitter
13
Pos: 26 /TD/Überschriften/MUF/1.2/2. 2 P r oduktbeschreibung @ 3\mod_123425872 3551 _79.doc @ 24008 @ 2
4.2. Product description
Pos: 27 /TD/Produktbeschreibung/Übersi cht/MUF 63xx/Auf einen Blick 6383 @ 4\mod_1 252499456455_79.doc @ 48901 @ 3
4.2.1. At a glance
1 Sealing plugs on the positive
pressure test connection
2 Sealing plugs on the
negative pressure test
connection 3 Socket for service plug 4 Display (optional) 5 Keys (only with optional
display) 6 Test rods for the analog
outputs 7 Service cover screw
connection (self-locking,
2 pcs.) 8 Openings for screws for
fastening to the wall 9 Only with integrated humidity
probe (KMAT option D04):
Opening for the integrated
humidity probe
Page 14
4 Transmitter
14
10 9-pin connector for the
current/voltage supply
11 Covered Ethernet interface
(non-functional)
12 Relay cover with 12-pin
connector underneath for relay connection (option)
13 Negative pressure
connection
14 Positive pressure connection
15 Probe socket
Pos: 28 /TD/Produktbeschreibung/Übersi cht/MUF 63xx/Verwendbare Fühler Pane l @ 4\mod_1252670326817_79.doc @ 50119 @ 3
4.2.2. Usable probes
The testo 6383 transmitter can be used with the following probes:
Probes Article no. Characteristic
Integrated humidity probe
KMAT option D04
Wall probe version; accuracy to ±1 % RH; temperature range -20 to +70 °C/-4 to +158 °F, plug-on sensor
testo 6612 0555 6610-L12
Duct probe version; accuracy to ±1 % RH; temperature range -30 to +150 °C/
-22 to +302 °F, sensor soldered
testo 6613 0555 6610-L13
Cable probe version; accuracy to ±1 % RH; temperature range -40 to +180 °C/
-40 to +356 °F, sensor soldered
testo 6614 0555 6610-L14
Heated cable probe version; accuracy to ±1.0 % RH; temperature range -40 to +180 °C/
-40 to +356 °F, sensor soldered
Page 15
4 Transmitter
15
Probes Article no. Characteristic
testo 6615 0555 6610-L15
Trace humidity cable probe version; accuracy ±1 K at 0 °Ctd/+32 °Ftd; temperature range
-40 to 120 °C/-40 to +248 °F, sensor soldered
testo 6617 0555 6610-L17
Cable with cover electrode monitoring probe version; accuracy to ± 1.2 % RH; temperature range -40 to 180 °C/
-40 to +356 °F, sensor soldered
Pos: 29 /TD/Produktbeschreibung/Übersi cht/MUF 63xx/Display und Tastatur @ 3\ mod_1234773965059_79.doc @ 25650 @ 3
4.2.3. Display and keypad
The display option allows operation of the testo 6383 transmitter via the display and four keys.
The LCD display consists of two 7-segment lines for displaying readings and units and of an information line (for status messages, for example).
The brightness and contrast of the display and the backgrou nd lighting (permanent or off) can be changed via the user menu or the P2A software.
Pos: 30 /TD/Überschriften/MUF/1.2. x Serviceschnittstelle @ 3\mod_12373068 91654_79.doc @ 29795 @ 3
4.2.4. Service interface
Pos: 31 /TD/Produktbeschreibung/Übersi cht/MUF 63xx/Serviceschnittste lle 638x @ 3\mod_1234774092911_79.doc @ 25669 @
The parameterizing socket (mini-DIN) is located behind the service flap as an interface to the P2A software or Testo handheld instrument (testo 400/testo 650).
Pos: 32 /TD/Produktbeschreibung/Übersi cht/MUF 63xx/Relaisplatine (Op ti on) @ 3\mod_1234774184843_79.doc @ 25688 @ 3
4.2.5. Relay board (option)
This has a floating switch capacity of 250 V AC/3 A. The switching limits and hysteresis as well as the function as relay for the collective alarm can be set via the display or the P2A software.
Further features include:
Function of changeover contacts (NC/NO contacts) freely selectable
12 terminals for a total of 4 relays.
If no relays are available, settings for monitoring limit values or alarms can still be controlled via the display. The alarm status will be shown on the display.
Only have the transmitter wired and connected by authorized personnel with the voltage disconnected.
Page 16
4 Transmitter
16
Pos: 33 /TD/Produktbeschreibung/Übersi cht/MUF 63xx/Analogausgänge 638 x @ 3\mo d_1 234774341528_79.doc @ 25707 @ 3
4.2.6. Analog outputs
As analog outputs, the testo 6383 has either
1 or optionally 3 current outputs of 0 to 20 mA (4-wire)/4 to
20 mA (4-wire) or
1 or optionally 3 voltage outputs of 0 to 1 V/0 to 5 V/0 to 10 V
(4-wire).
The transmitter can be ordered with three analog outputs as an option.
The optional three channels are galvanically isolated.
Pos: 34 /TD/Produktbeschreibung/Übersi cht/MUF 63xx/Messgrößen 6383/6385 @ 3\ m od_1 234774693900_79.doc @ 25745 @ 3
4.2.7. Parameters
The following parameters are displayed
Differential pressure in Pa, hPa, kPa, mbar, bar, mmH
2
O, inch
H
2
O, inch HG, kg/cm2, PSI
Relative humidity in % RH (technical)
Relative humidity in % WMO* (calculation according to the
WMO standard)
Degree of humidity in g/kg and gr/lb
Absolute humidity in g/m³ and gr/ft³
Water content in ppm (vol) and % vol
Psychrometer temperature in °C
tw
and °Ftw
Enthalpy in kJ/kg and BTU/lb
Water vapour partial pressure in hPa and inch H
2
O
Dewpoint temperature in °C
td
and °Ftd
Standardized dewpoint in °C
tdA ,
standardized at atmospheric
pressure (1013 hPa); precondition: Absolute process pressure.
Dewpoint of H
2O2
mixture in °Ctm and °Ftm.
* It is possible that condensation appears as of a displayed
humidity starting from 70 % and is shown on the display. This unit is used in meteorology, amongst others. When calculating the relative humidity the Magnus coefficient with undercooled water is used in accordance with WMO.
Calculated humidity variables correspond to the medium of air. With other gases/gas compositions, deviations may occur, e.g. with the enthalpy.
Temperature °C and °F
Page 17
4 Transmitter
17
Pos: 35 /TD/Produktbeschreibung/Übersi cht/MUF 63xx/Skalierung @ 3\mod_123477 5406989_79.doc @ 25783 @ 3
4.2.8. Scaling
There are three types of min./max. values: 1 The measuring range: The maximum sensor performance is in
this range. Values outside of the measuring range are displayed via messages, for example. Measuring range, see table (below).
2 Standard scaling: The output signals are assigned to this
measuring range as standard:
during delivery if no entries are made in the order code ◦ after exchanging the un it, the measuring range recorded in
the instrument is applied as standard.
The transmitter even retains its scaling with the voltage disconnected.
Measuring range, see table (below). 3 The maximum settings for the manual scaling
The maximum limits can be calculated as follows: X = difference between MIN. and MAX. value of the
standard scaling (Max. value of standard) + (50 % of X) (Min. value of standard) - (50 % of X) It is thus possible to scale beyond the measuring rang e, e.g.
for the adjustment of the scaling limits to standard values of
a PLC. With the alarm definition, however, the physical measuring
range limits are decisive.
Pos: 36 /TD/Produktbeschreibung/Übersi cht/MUF 63xx/Tabelle Skalierung M UF Pan el Druck @ 4\mod_1251818135359_79.doc @ 48019 @
Measuring range/standard scaling
Maximum scaling
0 to 50 Pa -5 to 15 Pa 0 to 50 Pa -25 to 75 Pa 0 to 100 Pa -50 to 150 Pa 0 to 500 Pa -250 to 750 Pa 0 to 10 hPa -5 to 15 hPa
-10 to 10 Pa -20 to 20 Pa
-50 to 50 Pa -100 to 100 Pa
-100 to 100 Pa -200 to 200 Pa
Page 18
4 Transmitter
18
Measuring range/standard scaling
Maximum scaling
0 to 50 Pa -5 to 15 Pa
-500 to 500 Pa -1000 to 1000
-10 to 10 hPa -20 to 20 hPa
Pos: 37 /TD/Produktbeschreibung/Übersi cht/MUF 63xx/Tabelle Skalierung Fühl er Panel @ 5\mod_1257519388289_79.doc @ 52767 @
Physical measuring range at 1013 hPa
Standard scaling of transmitter measuring range
Parameter Unit Probes
MIN MAX MIN MAX °C 6611 -20 +70 -20 +70 °F 6611 -4 +158 -4 +158 °C
6613, 6614, 6617
-40 +180 -40 +180
°F
6613, 6614, 6617
-40 +356 -40 +356
°C 6615 -40 +120 -40 +120
Temperature
°F 6615 -40 +248 -40 +248 °Ctd 6611 -20 +70 -80 +100 °Ftd 6611 -4 +158 -112 +212 °Ctd
6613, 6614, 6617
-20 +100 -80 +100
°Ftd
6613, 6614, 6617
-4 +212 -112 +212
°Ctd 6615 -60 +30 -80 +100
Dewpoint
°F
td
6615 -76 +86 -112 +212
g/m3
all probes
0 600 0 2000
Absolute humidity
gr/ft
3
all probes
0 250 0 800
relative humidity % RH
all probes
0 100 0 100
Page 19
4 Transmitter
19
Physical measuring range at 1013 hPa
Standard scaling of transmitter measuring range
WMO relative humidity
% RH 0 100 0 100
°Ctm -20 +100 -20 +100
Mixture dewpoint (H
2O2
)
°F
tm
-4 +212 -4 +212
g/kg
all probes
0 13300 0 9500
Degree of humidity
gr/lb
all probes
0 93000 0 66500
kJ/kg -40 99999 -40 8000 Enthalpy BTU/lb -18 43000 -18 3500 °Ctw -40 100 -40 180
Psychrometer temperature
°Ftw -58 210 -40 356 ppm
(vol) H
2
O
0 99999 0 99999 Water content
% vol 0 100 0 100 hPa 0 1000 0 7000
Water vapour partial pressure
inchH2O 0 400 0 2800
Pos: 38 /TD/Produktbeschreibung/Übersi cht/MUF 63xx/Alarmbehandlung @ 3\mod _12 34776787635_79.doc @ 25821 @ 3
4.2.9. Alarm handling
For upper and lower alarm limits, individual alarms as well as collective alarms can be specified. If the collective alarm function is activated, an alarm is triggered as soon as the alarm limit of an alarm is exceeded, if this alarm is assigned to the collective alarm.
The testo 6383 monitors limit values with the help of relays. If a reading is outside the limit values, a relay to be specified by the user is switched.
If the reading reverts to more than a specified hysteresis below or above the limit value, the alarm is cancelled.
In addition, information about the occurrence of error/status messages can be provided by means of a collective alarm relay, see Status, warning and error messages, page 53
If multiple alarm messages are activated at the same time, the last alarm is shown. If the alarm is cancelled again, the previous messages are no longer shown.
Page 20
4 Transmitter
20
Pos: 39 /TD/Überschriften/MUF/1.3/2. 3 I nbetriebnahme @ 3\mod_1234258805768_7 9.doc @ 24027 @ 2
4.3. Commissioning
Pos: 40 /TD/Erste Schritte/MUF 63xx/W andm ont age Panel @ 4\mod_1251818568448_79.doc @ 48051 @ 3
4.3.1. Mounting preparations
Page 21
4 Transmitter
21
1. Create a wall opening (approx. 120 mm x 220 mm) at the mounting location.
2. Hold 6383 in assembly position and mark the drill holes.
3. Drill holes suitable for the screws to be used.
4. Connect 6383.
Pos: 41 /TD/Erste Schritte/MUF 63xx/Gerä t anschließen Panel @ 4\mod_125206550656 6_79. doc @ 48245 @ 3
4.3.2. Connecting the instrument
WARNING
Electrical voltage Danger of injury!
> De-energize the mains connection before connecting the
transmitter.
Only have the transmitter wired and connected by authorized personnel with the voltage disconnected.
Page 22
4 Transmitter
22
Pos: 42 /TD/Erste Schritte/MUF 63xx/Ansc hlussübersicht 6383 @ 4\mod_125170832 7447 _79.doc @ 47697 @ 4
4.3.2.1. Overview of terminals
1 Terminal strip for voltage
supply and analog outputs
3 Relay cover (option)
2 Relay terminal strip (option),
below the relay cover
4 Insulating trough for relay
board (option), below the relay cover
The following description of the terminals refer to this overview and its numbering.
Pos: 43 /TD/Erste Schritte/MUF 63xx/Spannu ngsversorgung/Analogausgänge an sc hließen 6383/6385 @ 4\mod_1254489681844 _79.doc @ 51173 @ 45
Page 23
4 Transmitter
23
4.3.2.2. Connecting voltage supply and analog outputs
Terminal strip for voltage supply and analog outputs, item (1) of overview of terminals
Wiring diagram for 4-wire system (0 to 20 mA/4 to 20 mA/0 to 1 V/0 to 5 V/0 to 10 V)
1 1 or 3 channels,
0 to 20 mA/4 to 20 mA max. load per 500 0 to 1 V/0 to 5
V /
0 to 10 V
2 Functional earth
Requirement for the connecting cable of the supply:
Shielded and insulated with cross-section of at least
0.25 sq. mm, maximum 1.5 sq. mm without wire end sleeves.
T he supply line must be secured against exceeding
0.5 A.
An OFF switch must be installed in an easily accessible position close by and be marked as such.
1. Disconnect terminal strip for voltage supply and analog outputs.
2. Strip the cable ends, clamp wire end ferrules on and screw down with voltage terminals/channel terminals.
3. Tie together each of the two adjacent cores using a cable tie.
4. Attach terminal strip for voltage supply and analog outputs.
Page 24
4 Transmitter
24
Pos: 44 /TD/Erste Schritte/MUF 63xx/Re lai sa usgänge anschließen Panel @ 4\mod_12518 14425917_79.doc @ 47891 @ 4555
4.3.2.3. Connecting the relay outputs
Only have the transmitter wired and connected by authorized personnel with the voltage disconnected.
Relay terminal strip, item (2) of overview of terminals
There is the option of twelve terminals for a total of four relays.
1. Remove relay cover.
2. Disconnect relay terminal strip.
3. Strip cable ends and clamp on wire end ferrules.
4. Connect relays according to chosen function (NC/NO) (see
diagrams below; relay 1 is shown as an example of a connection).
Connection note
For the connection, a double-insulated mains cable (sheathed cable) with a cross-section of at least 1.5 mm² must be used.
Cable connection (2) may not be routed in a loop within the relay trough (1).
Page 25
4 Transmitter
25
It is recommended that you always tie 2 adjacent cores to one another using a cable tie (3).
T he insulation of the cable must be fed at least 5 mm (4) into the relay tray up to the elevated part.
Use of relay as NC contact (NC = normally closed)
1 Alarm/status light
(example of installation)
2 250 V AC/DC, 3 A
The busy light (alarm/status light) is permanently on until the relay opens or the circuit is interrupted. This circuit can therefore be used to monitor the functionality of the alarm circuit, as a cable break, for instance, is indicated by the busy light going off.
Page 26
4 Transmitter
26
Use of relay as NO contact (NO = normally open)
1 Alarm/status light
(example of installation)
2 250 V AC/DC, 3 A
The busy light (alarm/status light) only comes on when the relay is switched (closed). Monitoring the functionality of the alarm circuit is therefore not possible with this switching operation.
5. Insert relay terminal strip into socket.
6. Set on relay cover.
Page 27
4 Transmitter
27
Pos: 45 /TD/Erste Schritte/MUF 63xx/Gerä t schließen 6383 @ 4\mod_1252501761161_79.doc @ 48997 @ 4
4.3.2.4. Closing the instrument
1. Connect probe.
Version with int egrated humidity probe:
1. Insert probe into probe socket.
2. Guide probe through opening of front plate.
3. Tighten screw at opening.
Version with external h umid ity probe: > Insert probe into probe socket.
Page 28
4 Transmitter
28
4. Pull sealing frame (1) over the edge of the 6383.
5. Screw on 6383. Initially, only tighten the screws slightly. When all screws are in position, also align 6383 and tighten the screws.
6. Seal 6383 using a measure appropriate for the installation point (e.g. with a silicone strip).
Pos: 46 /TD/Erste Schritte/MUF 63xx/Gerä t abgleichen 6383/6385 @ 4\mod_1252502215678 _79.doc @ 49029 @ 4
Page 29
4 Transmitter
29
4.3.2.5. Adjusting the instrument
The testo adjusting concept allows the entire signal chain from the sensor signal (probe) and the digital signal (within the transmitter) through to the analog signal (transmitter output signal) to be adjusted (see diagram).
1-point adjustment
Adjustment via
• testo 400/650 handheld instrument with adjustment adapter
• P2A software
• User menu
Analog adjustment n-point adjustment
Adjustment using a precise multimeter and transmission of analog reference value in
• User menu or
• P2A software
Adjustment using a precise pressure sensor and transmission of analog reference value in
• User menu or
• P2A software
The 1-point adjustment is suitable for adjusting the sensor signal ­digital signal chain.
The testo 6383 transmitter has digital probes whose adjustment information is stored in the probes' internal memory. The 1-point adjustment can thus be carried out on another testo 6383 (e.g. in the calibration laboratory).
Page 30
4 Transmitter
30
Pos: 47 /TD/Erste Schritte/MUF 63xx/Übersi cht Abgleichtasten und Prüfkontak te 6383/6385 @ 4\mod_1252502914330_79.doc @ 49061 @ 4
4.3.2.6. Overview: Adjustment keys and test contacts
1 Contact ch. 1
2. Contact ch. 2
3. Contact ch. 3
4. Service interface
Pos: 48 /TD/Erste Schritte/MUF 63xx/1-Pun kt-Abgleich (Offset) @ 3\mod_1234790066 779_79.doc @ 26106 @ 45
4.3.2.7. 1-point adjustment (offset - humidity/temperature)
In the 1-point adjustment, the reading at the working point is raised to the reference value so that there is no longer any deviation in the working point. The reference condition can be measured using a precise handheld instrument (e.g. testo 400/650 with precision humidity probe) or be created in an air conditioning cabinet.
1 Deviation 2 Working point 3 Working range
The advantage of the 1-point adjustment is the good measuring result in the working range. But the further away the measurement is from the working point, the greater the deviation can become. The 1-point adjustment should therefore only be used for a
Page 31
4 Transmitter
31
relatively narrow measuring range (working range), e.g. clean room applications, air conditioning applications for storage and similar.
The 1-point adjustment can be performed
via the user menu (see Editin g Main Menu Adjust, page 50) or
via the P2A software (see volume 2, 1-point adjustment, page
122) or
directly by means of a T esto handhel d instrument (testo 400/650) (see description of how to proceed below).
Please note that the 1-point adjustment is generally performed on the basis of the % RH and °C/°F parameters.
Adjusting testo 6383 using testo handheld instrument
The service flap is open, a testo 400/650 handheld instrument
with a precision humidity probe is ready.
1. Connect testo handheld instrument 400/650 (1) with connected humidity reference probe (3) (order no. reference set 0699 3656/20) to the service interface (5) of the testo 6383 via the adjustment adapter (2) (connected to probe socket 1 of the handheld instrument).
2. Expose the humidity probe (4) of the testo 6383 and the reference probe (3) to the same reference conditions (e.g. in the humidity generator) and allow climatic conditions to equalize.
3. Switch on the testo 400/650. The two-part display of the handheld instrument will show the values of the transmitter on the left, and the values of the reference probe on the right. The humidity and temperature values are adjusted to the reference probe using the Probe > Probe Adjustment menu item on the
Page 32
4 Transmitter
32
testo 400/650. The 1-point adjustment is performed for both the humidity and the temperature.
4. Disconnect the adjustment adapter (2) from the service
interface (5).
5. Close the service flap.
Pos: 49 /TD/Erste Schritte/MUF 63xx/Hin weis 6614, 6615 @ 3\mod_1234791706885_79.doc @ 26163 @
Adjustment with the standard testo adjustment salt pots is not suitable for the testo 6614 (heated for high-humidity applications) and testo 6615 (trace humidity) probes.
The reference conditions should be generated in a humidity generator to adjust these probes. In addition, these probes can also be adjusted at a third adjustment point by Testo Service.
testo 6614: third adjustment point at 90 % RH
testo 6615: third adjustment point at -40 °Ctd/-40 °Ftd
Pos: 50 /TD/Erste Schritte/MUF 63xx/Analo gausgangs-Abgleich 6383/6385 @ 4\mod_125 2498044594_79.doc @ 48837 @ 45
4.3.2.8. Analog output adjustment
The purpose of adjusting the analog outputs is to adjust the signa l chain from the digital signal (within the transmitter) to the analog outputs. The signal type that was appointed for the transmitter is adjusted respectively for each channel (e.g. 4 to 20 mA or 0 to 1 V, etc.).
1 Contact ch. 1
2. Contact ch. 2
3. Contact ch. 3
4. Service interface
Page 33
4 Transmitter
33
Adjusting analog outputs 1, 2 and 3 (optional)
With testo 6383 with current output: Load of max. 500 Ω is
connected to channel that is to be adjusted (see Connecting voltage supply and analog outputs, page 23)
A precise multi meter (minimum requirement: resolution 6.5
digits, at least 5-times more accurate than the 6383) is available.
If only a simple multimeter is available, the analog outputs must not be adjusted.
The service cover is open.
1. Connect the inputs of the multimeter with the contacts (1) and (2) for channel 1 or with contacts (3) and (4) for channel 2, (5) and (6) for channel 3.
2. Transfer the reference analog value measured with the multimeter into the P2A software (see volume 2, Adjusting the analog output, page 125) or enter it via the user menu (see Editing Main Menu Adjust, page 50).
3. Disconnect connections between the multimeter and the contacts of the testo 6383 and close the service cover.
Pos: 51 /TD/Erste Schritte/MUF 63xx/n-Pun kt-Abgleich _638x @ 3\mod_1237384429887_ 79.doc @ 29990 @ 4
4.3.2.9. n-point adjustment (pressure)
With an n-point adjustment, the parameters at the 3-6 measurement points are adjusted to the reference value. T he reference conditions are obtained by using a precise pressure sensor that should be 5-times more accurate than the transmitter.
1 Positive pressure connection
2. Negative pressure connection
3. Pressure sensor
The number of measuring points is set to 3 by the factory and can only be changed using the P2A software (see n- point adjustment page 124)
Page 34
4 Transmitter
34
The n-point adjustment must always be carried out to its full extent and in good time at all selected measurement points.
A precise pr essure sensor (5-times more accurate than the
transmitter, e.g. DPC precision pressure sensor from testo industrial services) is available.
1. Connect the positive output of the pressure sensor (3) to the
positive pressure connection of the transmitter (1) and the negative output of the pressure sensor (3) to the negative pressure connection of the transmitter (2).
2. Transfer the reference pressure value created with the pressure
sensor into the P2A software (see volume 2, n-point adjustment, page 124) or enter it via the user menu (see Editing Main Menu Adjust, page 50).
3. Repeat step 2 for all of the measuring points.
4. Disconnect connections between the pressure sensor and the
pressure connections of the testo 6383.
Pos: 52 /TD/Erste Schritte/MUF 63xx/Hochf euchteabgleich beim testo 6614 @ 3\mod_123 5661029461_79.doc @ 26922 @ 4
4.3.2.10. High-humidity adjustment for testo 6614
With the testo 6614, the rear of the Testo humidity sensor is heated, creating a microclimate around the sensor (within the filter) that is constantly 5 K warmer than the actual process conditions.
Page 35
4 Transmitter
35
As can be seen in the Mollier diagram, this reduces the relative humidity at the sensor from around 100 % RH to a lower value, e.g. 73 % RH. In this range, the reaction time of the sensor is noticeably shorter than in the condensation range and the risk of the sensor corroding is also reduced. Using the separate temperature probe, the testo 6383 transmitter compensates the microclimate conditions and displays the process readings.
The reference conditions (11.3 % RH and 75.3 % RH) for the 2-point adjustment of the testo 6614 should be generated in a humidity generator, as humid ity adjustment sets cannot be used due to the heat generated.
The adjustment can also be carried out at a third adjustment point (90 % RH) by Testo Service so that optimum accuracy is also achieved in the high humidity ranges.
Pos: 53 /TD/Erste Schritte/MUF 63xx/Selb stabgleich des Restfeuchtefühlers tes to 6615 @ 3\mod_1235661074070_79.doc @ 26942 @ 4
4.3.2.11. Self adjustment of testo 6615 trace humidity probe
Conventional trace humidity probes show a steep rise in measuring uncertainty at low humidities. In the testo 6615 trace humidity probe, these measuring uncertainties are corrected by means of an automatic self-adjustment process. This means that extremely accurate measuring results are also attained to -60 °Ctd.
To this end, a temperature sensor is fitted on the back of the testo 6615 which is used as a heater. A humidity and temperature value pair is taken in both the unheated and heated state. The deviation of the probe obtained from these pairs of values is automatically corrected.
Page 36
4 Transmitter
36
The heating time and storing of cycles can be edited in the P2A software; for example, they can be deactivated by setting the two parameters to "0".
Deactivating th e adjustment function of the testo 6615 will reduce measuring accuracy and should therefore be restricted to the shortest possible length of time.
During the heating phase, the relay and analog outputs, the display value and output value are "frozen", see diagram above. Self-adjustment active is shown in the display until it has finished. The factory setting for self­adjustment time (incl. heating time, calculation time, cooling time) requires 30 minutes daily.
In the factory setting, a third adjustment point (-40 °Ctd) is approached for the testo 6615 in addition to the 2-point adjustment. This special adjustment can be performed again by your Testo Service team if necessary.
Pos: 54 /TD/Überschriften/MUF/1.4 Bedi enung @ 3\mod_1234443160034_79.doc @ 25001 @ 2
4.4. Operation
Pos: 55 /TD/Produkt verwenden/MUF 63xx/Zusa mmenhang Bedienmenü – Mini DIN Buchse akti v @ 3\mod_1234454016014_79.doc @ 25213 @ 3
4.4.1. Relationship between user menu and mini-DIN socket is active
The testo 6383 can be parameterized using either the user menu or the P2A software (see volume 2, Parameterizing, adjusting and analyzing software (P2A software) page 101).
The testo 6383 transmitter can only be operated via the display and keypad if the display option is available.
If the testo 6383 is connected to the P2A software, the user menu is blocked for the duration of the communication. The message Service plug is shown in the display of the testo
6383. As soon as the P2A software is disconnected, the user menu is accessible again.
Pos: 56 /TD/Produkt verwenden/MUF 63xx/Pass wortschutz _638x @ 3\mod_1234455136144 _79.doc @ 25251 @ 3
4.4.2. Password protection
The user menu can be protected with a four-digit numerical code (see Editing Main Menu Settings, page 44) so that access to the user menu is denied to unauthorized persons not familiar with this numerical code.
If the password protection is not to be used, the numerical code "0000" must be entered. This is also the status upon delivery.
Page 37
4 Transmitter
37
Pos: 57 /TD/Produkt verwenden/MUF 63xx/Auf bau des Bedienmenüs 638x @ 3\mod_1234455236596 _79.doc @ 25270 @ 3
4.4.3. Structure of user menu
At the main menu level, the user menu comprises the following:
Main Menu Channel 1
Main menu of channel 2 (if this option is available)
Main menu of channel 3 (if this option is available)
Main Menu Alarm
Main Menu Settings
Main Menu Analysis
Main Menu Messages
Main Menu Ident
Main Menu Adjust
Main Menu Reset 1 Channel 1 display
2 Channel 2 display 3 Channel 3 display or for
messages
Four keys enable the user to navigate/scroll through the menus and enter/amend values and settings:
Key Function/description
SET
In Measuring Mode: changes to parameterization
In Parameterizing Mode: confirms a selection or setting
ESC
Leaves a menu (without modifying any settings)
X Selecting: scrolls through menus
(downwards) or selectable alternatives
Editing: chang es to ne xt digit (to the right)
Page 38
4 Transmitter
38
Key Function/description
S Selecting: scrolls through menus (upwards)
or selectable alternatives
Editing: increa ses t he value of the current
digit by 1
Pos: 58 /TD/Produkt verwenden/MUF 63xx/Über sicht Bedienmenü @ 3\mod_1234510821302 _79.doc @ 25303 @ 3
4.4.4. Overview of the testo 6383 user menu
Page 39
4 Transmitter
39
Page 40
4 Transmitter
40
Page 41
4 Transmitter
41
Pos: 59 /TD/Produkt verwenden/MUF 63xx/Haup tmenü Kanäle bearbeiten @ 3\mod_12345111 1915 8_79.doc @ 25322 @ 34
4.4.5. The individual main menus
4.4.5.1. Editing main menu of channel 1
An overview is given in Overview of the testo 6383 user menu, page 38).
You can perform basic settings for channel 1.
1. In the Measuring Mode press SET, select Main Menu
Channel 1 with X or S and confirm selection with SET.
One of the following parameters can now be selected using X or S, after which the selection must be confirmed with SET:
Channel 1 Unit The parameter for this channel is selected. Edit/select parameter with X or S, confirm with SET or abort
entry with ESC.
Scale minimum for channel 1: The lower scale limit is edited; Unit as selected above. Editing the value: Scroll one digit to the right using X and
increase value of digit by 1 using S. Confirm with SET or abort entry with ESC.
Scale maximum for channel 1 The upper scale limit is edited; Unit as selected above. Editing the value: Scroll one digit to the right using X and
increase value of digit by 1 using S. Confirm with SET or abort entry with ESC.
Signal delay ("Attenuation") for channel 1 The analog signal can be delayed ("Attenuation"); a time
constant is selected for this (1 = no delay; 15 = longest delay) Edit/select parameter with X or S, confirm with SET or abort
entry with ESC.
2. Continue to the main menu with X or S or return to Measuring Mode with ESC.
Pos: 60 /TD/Produkt verwenden/MUF 63xx/Haup tmenü Kanal 2/3 638x @ 3\mod_1236085000820_79. doc @ 27312 @ 44
4.4.5.2. Editing Main Menu Channel 2 (if this option is available)
See channel 1.
Page 42
4 Transmitter
42
4.4.5.3. Editing Main Menu Channel 3 (if this option is available)
See channel 1.
Pos: 61 /TD/Produkt verwenden/MUF 63xx/Haup tmenü Alarm bearbeiten 638x @ 3\mod_123451 8541 385_79.doc @ 25341 @ 455
4.4.5.4. Editing Main Menu Alarm
With the alarm, the relays, available as options, are programmed. In addition, the alarm statuses are shown on the display (top right) (even without relays).
You can choose whether the alarm is to be used to monitor limit values or as a collective alarm. If an alarm is to be used to monitor limit values, you can choose between monitoring the minimum or maximum value and set a limit value and hysteresis for each alarm.
In addition, every alarm can be linked to a clearly visible visual alarm (display background lighting flashes).
An alarm delay between 0 and 240 seconds can still be assigned to every alarm used for limit value monitoring so that both the corresponding relay effect and the visual alarm are delayed. If the alarm status goes out within the set alarm delay time, neither the visual alarm nor a relay connection is triggered. .
With an alarm status present, the visual alarm and all relay outputs can be reset by means of acknowledgement. The triggering of a new alarm cannot be enabled until after the alarm status goes out.
1. In the Measuring Mode press SET, select Main Menu Alarm
with X or S and confirm select ion with SET.
- F our alarms can be parameterized.
2. Select Alarm x with X or S and confirm sele ction with SET.
Using alarm to monitor limit values
NO contact
Monitoring minimum Monitoring maximum
Hysteresis
Hysteresis
Limit value Limit value
Page 43
4 Transmitter
43
NC contact
Monitoring minimum Monitoring maximum
Hysteresis
Hysteresis
Limit value Limit value
3. Select Channel x (e.g. "Channel 1") with X or S and confirm selection with SET.
4. Select Max control or Min control with X or S (see graphic).
5. Press SET and edit Limit value as well as Hysteresis: Scroll one digit to the right using X and increase value of digit by 1 using S. Confirm with SET or abort entry with ESC.
6. Select Visual alarm with X or S. Select YES or NO wit h X or S. Confirm with SET or abort entry with ESC.
7. Press SET and edit Alarm delay: Scroll one digit to the right using X and increase value of digit by 1 using S (0 to 240 seconds possible). Confirm with SET or abort entry with ESC.
8. Return to Channel x with ESC.
9. Return to Alarm x with ESC.
10. Change to the other relays using X or S and perform settings in the same
wa
y.
Using alarm as collective alarm or not using it at all
If an alarm is assigned to the collective alarm, the relay is switched and a visual alarm can be issued via the display as soon as (at least) one of the warning or error messages of the testo 6383 transmitter (or the connected testo 6610 probe) becomes active.
The messages affecting the collective alarm can only be selected in the P2A software, see volume 2, Using the software, page 104
Alarm is selected (see previous steps 1 and 2).
1. Use X or S to determine whether Alarm x should be used as a
Collective alarm or not used. Confirm selection with SET.
2. If collective alarm is selected: Select Visual alarm with X or S. Select YES or NO with X or S. Confirm with SET and return to
Alarm x.
Page 44
4 Transmitter
44
3. Change to another alarm using X or S and perform settings in
the same way.
4. Return to Main Menu Alarm with ESC.
5. Continue to Main Menu Settings with X or S or return to
Measuring Mode with ESC.
Pos: 62 /TD/Produkt verwenden/MUF 63xx/Haup tmenü Einstellungen bearbeiten 6383/6 385 @ 4\mod_1252503503580_79.doc @ 49125 @ 4555 55
4.4.5.5. Editing Main Menu Settings
You can edit instrument settings and other settings. > In Measuring Mode, press SET, select Main Menu Settings
using X or S and confirm selection with SET.
You can edit settings for:
• Display
• Language
• Code
• Units
Absolute pressure Area Temperature
Humidity process data
H2O2 percentage by weight ◦ Humidity process pressure
Editing display settings
You can set the brightness and contrast of the display.
1. Select Display Settings with X or S and confirm selection wit h
SET.
2. Select Backlight or Contrast with X or S and confirm
selection with SET.
One of the following parameters can now be selected using X or S, after which the selection must be confirmed with SET:
Backlight
The display illumination is changed. Edit/select parameter with X or S, confirm with SET or cancel
entry with ESC (the effect of the change in parameter can be seen during input).
Contrast
Page 45
4 Transmitter
45
The brightness difference between the display background and the displayed values is changed.
Edit/select parameter with X or S, confirm with SET or cancel entry with ESC (the effect of the change in parameter can be seen during input).
Backlight 24h o n Select On or Off using X or S and confirm with SET. Off: The display light switches off automatically if no button was
pressed for 10 seconds. On: The display light is activated
3. Return to Display Settings with ESC and useX or S to continue to Language.
Selecting language
You can select the language for the plain text line in the display. > Press SET, select required language with X or S, confirm
selection with SET and return to Language.
Only choose a language that you can understand well.
Editing code settings
You can set the access code (password).
If a code other than "0000" (factory setting) is set, the transmitter can only be operated once this code has been entered via the menu.
1. Select Code with X or S and confirm selection with SET.
2. Scroll one digit to the right using X and increase value of digit by 1 using S. Confirm with SET or abort entry with ESC.
3. Return to Code with ESC and useX or S to continue to
Change parameters.
Select unit (Change unit)
This setting affects the displayed unit of all readings.
1. Press SET, select Change parameters with X or S, confirm selection with SET or cancel with ESC.
2. Select Change unit with X or S, confirm selection with SET or cancel with ESC.
Page 46
4 Transmitter
46
3. Select the required variable (absolute
pressure/area/temperature) with X or S, confirm selection with
SET or cancel with ESC.
4. Select the required unit with X or S, confirm selection with SET
or cancel with ESC.
5 Return to Change unit with ESC.
6. Return to Change parameters with ESC.
7. Return to Main Menu Settings with ESC.
Editing Humidity process data
This menu is only used to parameterize humidity measurem ents in H
2O2
atmospheres (e.g. in sterilization processes) and determines
whether °Ctm or °Ftm is the output parameter.
1. Select Humidity process data with X or S and confirm
selection with SET.
2. Select H202 weight prop. with X or S, confirm selection with
SET or cancel with ESC.
The selection H
2O2
water or H2O2 vapour in the submenu
describes whether H2O2 is generated by means of evaporation or is actively vaporized in the process.
3. Select H
2O2
water or H2O2 vapour with X or S and confirm
with SET.
4. Edit proportion by weight of the H
2O2
in % (% H2O2 proportion by weight in the liquid initial solution): Scroll one digit to the right using X and increase value of digit by 1 using S. Confirm with
SET or abort entry with ESC.
5. Return to H202 weight prop. with ESC and useX or S to continue to Humidity process pressure.
6. Select Humidity process pressure with X or S and confirm selection with SET.
7. Edit humidity process pressure: Scroll one digit to the right using X and increase value of digit by 1 using S. Confirm with
SET or abort entry with ESC.
8. Return to Change parameters with ESC.
9. Return to Main Menu Settings with ESC.
Pos: 63 /TD/Produkt verwenden/MUF 63xx/Haup tmenü Analyse bearbeiten 6383/6385 @ 4\mod_1 252507887718_79.doc @ 49449 @ 45555
Page 47
4 Transmitter
47
4.4.5.6. Editing Main Menu Analysis
You can test the functionality of analog and rela y outputs. In addition, you can read off the minimum and maximum values (since the last voltage supply or reset of the min./max. values).
Testing functionality of analog outputs
This function affects the analog outputs directly, not only the test contacts.
1. In the Measuring Mode press SET, select main menu Analysis with X or S and confirm select ion with SET.
- Test Analog Output is shown.
2. Press SET and choose between Analog output 1, 2, 3 with X or S.
3. Press SET, scroll one digit to the right using X and increase value of digit by 1 using S. Any analog output value can be predefined, e.g. for an analog output of 4 to 20 mA, the value "6.0 mA". Confirm with SET or abort entry with ESC.
4. Accept setting by pressing SET and test with multimeter (minimum requirement: resolution 6.5 digits, at least 2-times more accurate than the 6383) as follows:
Analog output 1, 2 or 3: Via test contacts under service cover, see diagram.
1 Positive test
contact channel 1
2 Negative test
contact channel 1
3 Multimeter
4. Return to Test Analog Output with ESC and useX or S to continue to Test Relay Output.
Page 48
4 Transmitter
48
Testing functionality of the pressure sensor (Test pressure
sensor)
This function is only required to calibrate the pressure sensor.
Testing functionality of relay outputs
1. Press SET and choose between Alarm 1, 2, 3, 4 with X or S.
2. Press SET.
The relay can now be tested. You can choose between OFF and ON using X or S. If ON is chosen, the NO contact is closed, the NC contact opened. If OFF is chosen, the NC contact is closed, the NO contact opened.
3. To test, route a measuring cable from the relay terminals (see
Connecting the relay outputs, page 24) out of the transmitter to a multimeter (resistance measurement) or continuity tester.
4. Return to Test Relay Output with SET (starts relay test) or
ESC (exits the menu without relay test).
Reading off min./max. values of channels
To reset the max./min. values, see Editing Reset main menu, page 52
1. Read off the min./max. values of the three channels in
succession with X or S and return to the Main Menu Analysis with ESC.
2. Continue to Main Menu Message with X or S or return to
Measuring Mode with ESC.
Pos: 64 /TD/Produkt verwenden/MUF 63xx/Haup tmenü Meldungen bearbeiten @ 3\mod_123452 6541417_79.doc @ 25398 @ 4
Page 49
4 Transmitter
49
4.4.5.7. Editing Message main menu
Messages can be confirmed/acknowledged, the last messages can be called up and the display of the messages can be switched on or off.
1 Operating hours at the
time of message
2 Message code (see
Status, warning and error
messages, page 53). 3 Message text 4 Message number
(example: "4/7" refers to
the fourth of seven
messages) 5 Number of messages
present (example: "4/7"
refers to the fourth of
seven messages)
Using the P2A software (see volume 2, Parameterizing, adjusting and analyzing software (P2A software), page 101 ) you can predefine whether messages are to be shown in the display.
1. In the Measuring Mode press SET, select Main Menu Message with X or S and confirm select ion with SET.
2. Confirm Confirm message using SET.
3. Select Last messages with X or S and confirm with SET.
4. Scroll between the warning and error messages recorded so far using X or S and return to Last messages using ESC.
5. Select Information with X or S and confirm with SET.
6. Scroll between the status messages recorded so far using X or S and return to Information using ESC.
7. Continue to Display of message with X or S.
ON: Measurements are shown on the display in Measuring
Mode.
OFF: No messages shown on display.
8. Select ON or OFF with X or S and confirm selection with SET.
9. Return to Main Menu Message with ESC.
Page 50
4 Transmitter
50
10. Continue to Main Menu Ident with X or S or return to
Measuring Mode with ESC.
An overview of the messages can be found in Status, warning and error messages, page 53
Pos: 65 /TD/Produkt verwenden/MUF 63xx/Haup tmenü Ident abfragen 638x @ 3\mod_12345269 9773 1_79.doc @ 25417 @ 4
4.4.5.8. Calling up Main Menu Ident
1 Instrument or
probe type 2 Build number 3 Firmware
version 4 Serial number
The serial numbers of the transmitter and probe can be read off.
1. In the Measuring Mode press SET, select Main Menu Ident with X or S and confirm select ion with SET.
- The type, firmware version, build number and serial number of the transmitter are displayed.
This information is required when servicing.
2. Return to Main Menu Ident with ESC or read off the type, firmware release, build number and serial number of the probe with X or S and then return to Main Menu Ident with X or S.
3. Continue to Main Menu Adjust with X or S or return to Measuring Mode with ESC.
Pos: 66 /TD/Produkt verwenden/MUF 63xx/Haup tmenü Abgleich bearbeiten 6383/6385 @ 4\ mod _12 52503280594_79.doc @ 49093 @ 4555
4.4.5.9. Editing Main Menu Adjust
A reference value can be entered for both relative humidity (RH) and temperature (°C/°F) for the 1-point adjustment. Please refer to the description in 1-point adjustment (offset ­humidity/temperature), page 30
Reference values for pressure can be entered for the n-point adjustment. Please refer to the description in n-point adjustment (pressure), page 33
Page 51
4 Transmitter
51
In addition, the analog outputs can be adjusted. Also see Analog output adjustment, page 32 for instructions on how to do this.
Enter reference value for 1-point adjustment
Please also refer to 1-point adjustment (offset ­humidity/temperature), page 30
1. In the Measuring Mode press SET, select main menu Adjust with X or S and confirm select ion with SET.
- Reference value % RH is displayed.
2. Press SET, edit value: Scroll one digit to the right using X and increase value of digit by 1 using S. Confirm with SET or abort entry with ESC.
3. Continue to Reference value temp with X or S.
4. Press SET, Reference value °C or Reference value °F is displayed.
5. Press SET, edit value: Scroll one digit to the right using X and increase value of digit by 1 using S. Confirm with SET or abort entry with ESC.
6. Continue to Analog Adj. Ch. 1 with X or S.
7. Continue with the adjustment of the analog outputs (see belo w, step 2) or press ESC to return to main menu Adjust.
8. Continue to main menu Reset with X or S or return to Measuring Mode with ESC.
Performing analog adjustment
Please refer to Analog output adjustment, page 32.
1. In the Measuring Mode press SET, select main menu Adjust with X or S and confirm select ion with SET.
Each channel is adjusted at three points in the analog range (at 10 %; 50 %; 90°% of the analog scale).
2. Select Analog Adj. Ch. 1 with X or S and confirm with SET.
3. Use X or S to select Adj. Point 1.
4. Press SET. Read off multimeter display (e.g. 5601 mA) and enter this value in the user menu. Do this by scrolling one digit to the right using X and increasing the value of digit by 1 using S. Confirm with SET or abort entry with ESC.
5. Use X or S to select Adj. Point 2.
6. Press SET. Read off multimeter display (e.g. 12,001 mA) and enter this value in the user menu. Do this by scrolling one digit
Page 52
4 Transmitter
52
to the right using X and increasing the value of digit by 1 using S. Confirm with SET or abort entry with ESC.
7. Use X or S to select Adj. Point 3.
8. Press SET. Read off multimeter display (e.g. 18,401 mA) and
enter this value in the user menu. Do this by scrolling one digit to the right using S and increasing the value of digit by 1 using X. Confirm with SET or abort entry with ESC.
9. With X or S continue to Analog Adj. Ch. 2or 3 (repeat steps 3
to 8).
10. Return to main menu Adjust with ESC.
11. Continue to main menu Reset with X or S or return to
Measuring Mode with ESC.
Performing the pressure adjustment
Please refer to n-point adjustment (pressure), page 33.
1. In the Measuring Mode press SET, select main menu Adjust
with X or S and confirm select ion with SET.
The pressure adjustment can be performed at 3 to 6 adjustment points.
2. Select Adjust Pressure with X or S and confirm with SET.
3. Use X or S to select Adj. Point 1.
4. Press SET. Read off pressure sensor display (e.g. 30.1 Pa) and
enter this value in the user menu. Do this by scrolling one digit to the right using X and increasing the value of digit by 1 using S. Confirm with SET or abort entry with ESC.
5. Repeat steps 1 to 4 for adjustment points 2 to 6.
6. Return to main menu Adjust with ESC.
7. Continue to main menu Reset with X or S or return to
Measuring Mode with ESC.
Pos: 67 /TD/Produkt verwenden/MUF 63xx/Haup tmenü Reset bearbeiten 638x @ 3\mod_123452 9155 827_79.doc @ 25455 @ 4
4.4.5.10. Editing Reset main menu
You can reset the factory settings for the following:
• Instrument
• Sensor/probe
• Min./max. values
Resetting to the factory settings means resetting to the order specification, i.e. the specific condition at the time of supply to the customer.
Page 53
4 Transmitter
53
1. In the Measuring Mode press SET, select Main Menu Reset with X or S and confirm select ion with SET.
- Reset Device to factory settings is displayed.
2. Select the setting to be reset using X or S and confirm selection with SET.
- Reset Completed is displayed.
3. Return to the reset setting with ESC or SET and return to Main
Menu Reset with ESC.
4. Continue to Main Menu Channel 1 with X or S or return to Measuring Mode with ESC.
Pos: 68 /TD/Überschriften/MUF/1.5 Sta tus-, Warn- und Fehlermeldungen @ 3\mod_123444 2925442_79.doc @ 24963 @ 2
4.5. Status, warning and error messages
Pos: 69 /TD/Produkt verwenden/MUF 63xx/Sta tus-Warnmeldungen/Status-, W arn- und Fehlermeldungen 638x @ 3\mod_1234441087981 _79.doc @ 24844 @
To achieve optimum operational reliability (machine ava ilability), the transmitter shows the following via the user menu or the P2A software
• Status messages
• Warning messages
• Error messages
The status and warning messages for the respective testo 6610 probes connected to the transmitter can be evaluated via the P2A software.
All messages are stored in the transmitter with an operating hours stamp. Use the user menu (see Editing Message main menu, page
49) or the P2A software (see volume 2, Transmitter history, page
126) to view the message history.
In the transmitter, the last 60 status messages and the last 120 error and warning messages are stored in a ring memory. There is no limit in the P2A software.
Pos: 70 /TD/Produkt verwenden/MUF 63xx/Sta tus-Warnmeldungen/Statusmeldung en 6383/6385 @ 4\mod_1252506138420_79.doc @ 49257 @ 3
Page 54
4 Transmitter
54
4.5.1. Status messages
Status messages show the current operating status of the testo
6383.
Message Display Description
02506
Sensor initialization
Message appears while the transmitter is starting up. If the message disappears, the transmitter is ready for operation.
01D19 Service plug
The Mini-DIN socket is connected to: the USB adapter for P2A software, the adjustment adapter
or the service plug (is not recorded/no number)
00300 New limit value
The limit value has been changed
or shifted 00301 Scaling changed The scaling has been changed 00500 Transmitter reset
The transmitter was reset to the
factory settings and is restarted. 0052F Reset Min/Max
Resets all saved Min/Max values
for all channels 02518 Probe reset
Probe reset: The probe performs a
reset 00503
Reset device to fact
A factory reset of the transmitter
was performed 02503
Reset probe to fact
A factory reset of the probe was
performed 00530
Change solenoid valve
The solenoid valve should be
changed 00307
User Setting Change
User Setting Change: General
settings were changed for the
transmitter. 00117 Adjustment DeltaP
An n-point adjustment was
performed. 02104 Analog adjustment
An analog adjustment has been
made 02101 1-point adjustment A 1-point adjustment is performed.
Page 55
4 Transmitter
55
Message Display Description
02105
Self-adjustment active
For testo 6615 probe only: The probe performs an automatic self­adjustment
Pos: 71 /TD/Produkt verwenden/MUF 63xx/Sta tus-Warnmeldungen/Warnmeldungen 6 383/ 6385 @ 4\mod_1252506208031_79.doc @ 49289 @ 3
4.5.2. Warning messages
Warning messages show an early warning or a current malfunction which may negatively impact measuring.
Message Display Cause Remedying of fault
00809
Pressure too high**
The process pressure exceeds the pressure intended for the transmitter
Remove the transmitter from the process and take any necessary measures to lower the pressure
00E00 T ambient high**
The ambient temperature exceeds the permissible temperature for the transmitter
Take necessary measures to lower ambient temperature, e.g. through venting or cooling
00E01 T ambient low**
The ambient temperature is below the permissible temperature for the transmitter
Take necessary measures to raise ambient temperature, e.g. through heating
00E04/00E 05
Supply voltage low**
The supply voltage is below the minimum voltage required for the transmitter
Take measures to ensure sufficient voltage supply
02822 T process high**
The process temperature exceeds the temperature designated for the probe
Remove the probe from the process and take any necessary measures to lower the process temperature
02821 T process low**
The process temperature is below the temperature designated for the probe
Take necessary measures to raise process temperature
0081C Alarm 1**
Depending on the parameterization of the relays
Depending on the parameterization of the relays
Page 56
4 Transmitter
56
Message Display Cause Remedying of fault
0081D Alarm 2**
Depending on the parameterization of the relays
Depending on the parameterization of the relays
0081E Alarm 3**
Depending on the parameterization of the relays
Depending on the parameterization of the relays
0081F Alarm 4**
Depending on the parameterization of the relays
Depending on the parameterization of the relays
02806 Condensation*
100 % RH has been reached, condensation developing
Take measures to reduce process humidity
02807
Values less than 0 % RH**
The adjustment or sensor is faulty
Check adjustment (via P2A adjustment history, perform 2-point adjustment where necessary). If the problem persists, contact Testo Service
02809
Sensor early warning*
For testo 6617 probe only: The cover electrode of the sensor is damaged; this may soon cause the sensor to break
Carry out visual inspection. If the mirror­like surface of the sensor is dirty or damaged, contact Testo Service
* Early warning ** Current malfunction
Pos: 72 /TD/Produkt verwenden/MUF 63xx/Sta tus-Warnmeldungen/Fehlermeldung en Messumformer 638x @ 3\mod_123453585107 6_79. doc @ 25512 @ 3
4.5.3. Transmitter error messages
Error messages show a current malfunction.
Message Display Cause Remedying of fault
03401 No probe signal
The probe communication is interrupted
Ensure that the probe connector is fully engaged in the transmitter.
If communication still cannot be established, contact Testo Service
Page 57
4 Transmitter
57
Message Display Cause Remedying of fault
03508 Wrong probe
The connected probe is not compatible with the present transmitter
Use a compatible probe Note: Probe 661x for
transmitter 638x
01505 Watchdog error
Due to a processor error, the transmitter performs an automatic restart
If the problem occurs frequently, contact Testo Service
0300A
% RH short­circuit
Short-circuit in humidity sensor
Contact Testo Service
0300B
% RH sensor broken
The humidity sensor is damaged (sensor broken)
Contact Testo Service
0300C T short-circuit
Short-circuit in temperature sensor
Contact Testo Service
0300D T sensor broken
The temperature sensor is damaged (sensor broken)
Contact Testo Service
03105
Self-adjustment error
For testo 6615 probe only: The automatic self-adjustment was faulty
Contact Testo Service
03106 Adjustment error
The adjustment of the probe was faulty
Contact Testo Service
01115
Low adjustment temperature
The ambient temperature is too low during the pressure adjustment
Take necessary measures to raise ambient temperature, e.g. through heating
01116
High adjustment temperature
The ambient temperature is too high during the pressure adjustment
Take necessary measures to lower ambient temperature, e.g. through venting.
03000
Heat function defective
For testo 6614 probe only: Heat function defective
Contact Testo Service
Page 58
4 Transmitter
58
Pos: 73 /TD/Produkt verwenden/MUF 63xx/Sta tus-Warnmeldungen/Behandlung von Alar mmeldungen 6383/6385 @ 4\mod_12525062849 72_79.doc @ 49321 @ 3
4.5.4. Handling alarm messages
Shown on the display
3
Can be used for collective alarm4
Additional message end5
New limit value x Scaling changed x Pressure too high x x Delta p adjustment x Alarm 1 x Alarm 2 x Alarm 3 x Alarm 4 x Transmitter reset x Reset Min/Max Transmitter refresh x User Setting Change Analog adjustment x 1-point adjustment x Probe reset x T ambient high x x T ambient low x x Supply voltage low x x T process high x x T process low x x
3
If multiple messages/alarms are activated at the same time, only the last message/alarm is shown. If this is cancelled, the other messages that are still active are no longer displayed.
4
The message can be assigned the collective alarm function, which means that the collective alarm is activated as soon as at least one of the messages assigned to it is activated. The collective alarm can be assigned to each of the 4 optional relays. The collective alarm is then always the same, as it can only be defined once.
5
The message is shown upon the occurrence of the event causing the message as well as when closing. Two entries appear in the history in the P2A software: Message text_start and Message text_end.
Page 59
4 Transmitter
59
Shown on the display3
Can be used for collective alarm4
Additional message end5
Condensation x x Values less than
0 % RH
x x
Sensor early warning x No probe signal x Watchdog error x % RH short-circuit x x % RH sensor broken x x T short-circuit x x T sensor broken x x Heat function
defective
x x
Perform the Confirm message function (acknowledgemen t of the alarm via the control keys on the transmitter):
The message/alarm is no l onger shown on the display and the optical alarm goes out, where applicable. If multiple messages/alarms are active at the same time, all are reset simultaneously.
If at least one message is assigne d to the collective alarm, the collective alarm is reset. If the collective alarm is set on a relay, the relay is also reset, meaning switched to its neutral position.
Pos: 74 /TD/Produkt verwenden/MUF 63xx/Sta tus-Warnmeldungen/Namur Fehlerbedi ngungen 638x @ 3\mod_1234536611418_79.doc @ 25550 @ 3
4.5.5. Namur fault conditions
If the faults named in the following table occur, the analog outputs output special values that enable a general fault warning in the higher-level control system. The values correspond to the "Namur" industry standard.
Status message in the display
Display value in display
Class Analog output
0 to 20 mA
4 to 20 mA
1 V 5 V 10 V
No probe signal
None Error 21 mA 21 mA 1.1 V 5.5 V 11 V
Wrong probe None Error 21 mA 21 mA 1.1 V 5.5 V 11 V
Page 60
4 Transmitter
60
Status message in the display
Display value in display
Class Analog output
Values < 0 %RH
uuuuu Underrange 0 mA 3.8 mA 0 V 0 V 0 V
Condensation ooooo Overrange 20.5 mA 20.5 mA 1.1 V 5.5 V 11 V % RH short-
circuit
----- Error 21 mA 21 mA 1.1 V 5.5 V 11 V
% RH sensor broken
----- Error 21 mA 21 mA 1.1 V 5.5 V 11 V
T short-circuit ----- Error 21 mA 21 mA 1.1 V 5.5 V 11 V T sensor
broken
----- Error 21 mA 21 mA 1.1 V 5.5 V 11 V
T process low uuuuu Underran ge 0 mA 3.8 mA 0 V 0 V 0 V T process
high
ooooo Overrange 20.5 mA 20.5 mA 1.1 V 5.5 V 11 V
Probe disconnected
None Error 21 mA 3.8 mA 1.1 V 5.5 V 11 V
Watchdog error
Previous value stops
Error 21 mA 3.8 mA 1.1 V 5.5 V 11 V
Value below min. scale
Reading Underrange 0 mA 3.8 mA 0 V 0 V 0 V
Value above max. scale
Reading Overrange 20.5 mA 20.5 mA 1.1 V 5.5 V 11 V
Pressure too high
ooooo Overrange 20.5 mA 20.5 mA 1.1 V 5.5 V 11 V
Heat function defective
----- Error 21 mA 21 mA 1.1 V 5.5 V 11 V
Pos: 75 /TD/Überschriften/MUF/1.6 W ar tung und Reinigung @ 3\mod_1234443039129_79. doc @ 24982 @ 2
Page 61
4 Transmitter
61
4.6. Maintenance and cleaning
Pos: 76 /TD/Produkt instand halten/MUF 63xx/ Gerät warten/reinigen 6383/6385 @ 4\mod_1 255077977417_79.doc @ 51703 @ 3355
4.6.1. Maintaining the instrument
We recommend that the adjustment and settings of the transmitter be checked at regular intervals using the
User menu (Operati on, page 36) or
P2A software (volume 2, Parameterizing, adjusting and analyzing software (P2A software), page 101)
Convenient "remote monitoring" of the transmitter can be implemented, for example by using a relay as a collective alarm (see Using alarm as collective alarm or not using it at all, page 43) whose messages are forwarded to a local alarm transmitter (horn, light) or PLC.
4.6.2. Cleaning the instrument
Integrated humidity probe
The humidity sensor of the integrated humidity probe in particular must be protected.
The sensors of the integrated humidity probe must not make contact with cleaning agents, otherwise the functionality can no longer be guarantee d.
1. Put supplied plastic cap on the integrated humidity probe.
2. Wipe surface of the instrument with mild cleaning agents and a soft cloth.
3. Remove plastic cap.
Instrument surface
Integrated hum idit y probe is protected by the plastic cap.
Clean the surface of the instrument with mild cleaning agent s (e.g. washing-up liquid) and water. Only use gentle and standard cleaners.
Do not use any aggressive or chafing cleaning agents such as thinner, liquid scouring agent, wax, benzine, concentrated acids and lyes.
Carefully rub down the instrument surface with a moist and soft cloth.
Page 62
4 Transmitter
62
Page 63
4 Transmitter
63
Page 64
0970 6383 en V01.0 V01.40-1 en-GB Vol1
Loading...