Testo 6381 Operating Manual

testo 6381 · di f f erential pressu re t ransmitter P2A sof t ware · Parameteriz ing , adj usting and analyz ing
software
Instruction manual
2
1 Safety and the environment
Pos: 1 /TD/Überschriften/MUF/Sicherheit und Umwelt @ 3\ mod_1234793958627_79. doc @ 26223 @ 1
1 Safety and the environment
Pos: 2 /TD/Sicherheit und Umwelt/Sicherheit gewährle isten/ MUF 63xx/ Elektrische G efahr en vermeiden @ 3\m od_1234794609299_79.doc @ 26280 @ 5
Avoiding electrical hazards
> Never use the instrument and connected probes to measure on
or near live parts!
> Damaged mains cables must only be replaced by authorized
personnel.
> Only have the transmitter wired and connected by authorized
personnel with the voltage disconnected.
> You must always comply with the regulations applicable in your
Pos: 3 /TD/Sicherheit und Umwelt/Sicherheit gewährle isten/ MUF 63xx/ Personen- und Sachschäden verm eiden @ 3\mod_1234794744768_79. doc @ 26299 @ 5
country for opening and repairing electrical equipment.
Avoiding personal injury and damage to equipment
> Installation, setting and calibration work must only be carried
out by qualified and authorized personnel!
> Only open the instrument when this is expressly described in
the instruction manual for installation, maintenance or repair purposes.
> Observe the permissible storage, transport and operating
Pos: 4 /TD/Sicherheit und Umwelt/Sicherheit gewährle isten/ Nicht mit Lösungsmit teln lagern @ 0\mod_1175692375179_79. doc @ 583 @
temperature.
> Do not store the product together with solvents. Do not use any
Pos: 5 /TD/Sicherheit und Umwelt/Sicherheit gewährle isten/ MUF 63xx/ Bei Wartung M UF nicht zur Regelung verwenden @ 3\mod_1234794852377_ 79. doc @ 26318 @
desiccants.
> Do not use the instrument for control purposes at the same time
Pos: 6 /TD/Sicherheit und Umwelt/Sicherheit gewährle isten/ Produkt best immungsgem äß verwenden @ 0\mod_1173781261848_79.doc @ 386 @
as operating or servicing the transmitter.
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not
Pos: 7 /TD/Sicherheit und Umwelt/Sicherheit gewährle isten/ Nur beschr iebene Wart ungsarbeiten dur chf ühren @ 0\mod_1175692705195_79. doc @ 601 @
use any force.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the prescribed steps exactly. Use only original spare parts from
Pos: 8 /TD/Sicherheit und Umwelt/Sicherheit gewährle isten/ MUF 63xx/ Fachpersonal @ 3\ mod_1234794940409_79. doc @ 26337 @
Testo.
Any additional work must only be carried out by authorized personnel. Otherwise testo will not accept any responsibility for the proper functioning of the instrument after repair and for the validity of certifications.
3
2 About this document
Pos: 9 /TD/Überschriften/MUF/Umwelt schütz en @ 3\mod_1234858757571_79.doc @ 26363 @ 5
Pos: 10 /TD/Sicherheit und Umwelt/Umwelt schüt zen/ Produkt ent sorgen @ 0\mod_1173780307072_79.doc @ 357 @
Protecting the environment
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
Pos: 11 /TD/Übers chrift en/MUF/Zu diesem Dokument @ 3\mod_1234793991331_79. doc @ 26242 @ 1
2 About this doc ument
Pos: 12 /TD/Sicherheit und Umwelt/Zu diesem Dokument / Verwendung (Standar d) @ 0\mod_1173775068554_79. doc @ 337 @ 5
regulations) or return the product to Testo for disposal.
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay particular attention to the safety instructions and warning advice in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
Pos: 13 /TD/Sicherheit und Umwelt/Zu diesem Dokument / Symbole und Schreibkonventionen/ Warnhin weis WARNUNG @ 2\m od_1207646966234_79.doc @ 14398 @
Pos: 14 /TD/Sicherheit und Umwelt/Zu diesem Dokument / Symbole und Schreibkonventionen/ Warnhin weis VORSICHT @ 2\ mod_1207651416515_ 79. doc @ 14416 @
Pos: 15 /TD/Sicherheit und Umwelt/Zu diesem Dokument / Symbole und Schreibkonv. Sof tware [ Standard] @ 0\mod_1190203332543_79. doc @ 4883 @ 5
product.
WARNING
Indicates potential serious injuries
CAUTION indicates potential minor injuries
Symbols and writing standards Representa-
Explanation
tion
Note: Basic or further information.
1. ...
2. ...
Action: more steps, the sequence must be followed.
> ... Action: a step or an optional step.
- ... Result of an action.
Menu Elements of the program interface. [OK] Buttons of the program interface.
... | ... Functions/paths within a menu. “...” Example entries
Pos: 16 /TD/--- Seitenwechsel --- @ 0\mod_1173774430601_0. doc @ 283 @
4
3 Contents
Pos: 17 /TD/Übers chrift en/MUF/I nhalt @ 3\mod_1234794019831_79.doc @ 26261 @ 1
3 Contents
1 Safety and the environment....................................................................3
2 About this document ...............................................................................4
3 Contents....................................................................................................5
4 Transmitter................................................................................................8
4.1. Specifications...................................................................................8
4.1.1. Functions and use...........................................................................................8
4.1.2. Scope of delivery............................................................................................8
4.1.3. Accessories....................................................................................................9
4.1.4. Technical data................................................................................................9
4.1.5. Dimensions................................................................................................... 13
4.2. Product description ........................................................................13
4.2.1. At a glance....................................................................................................13
4.2.2. Display and keypad.......................................................................................15
4.2.3. Service interface........................................................................................... 15
4.2.4. Relay board (option) ..................................................................................... 15
4.2.5. Analog outputs..............................................................................................15
4.2.6. Parameters................................................................................................... 16
4.2.7. Scaling 16
4.2.8. Alarm handling..............................................................................................17
4.3. Commissioning...............................................................................18
4.3.1. Inserting Ethernet module (order no. 0554 6656).......................................... 18
4.3.2. Assembling the instrument............................................................................ 20
4.3.2.1. Wall mounting .................................................................................. 20
4.3.3. Connecting the instrument............................................................................ 21
4.3.3.1. Overview of terminals.......................................................................23
4.3.3.2. Connecting voltage supply and analog outputs.................................24
4.3.3.3. Connecting the relay outputs............................................................25
4.3.3.4. Plug-in connection option................................................................. 28
4.3.3.5. Creating the PE/earthing terminal..................................................... 29
4.3.3.6. Setting the Ethernet module.............................................................30
4.3.3.7. Closing the instrument......................................................................32
4.3.4. Ethernet communication ............................................................................... 33
4.3.4.1. Types of operation ........................................................................... 33
4.3.4.2. Mains connection.............................................................................33
4.3.4.3. LED status displays..........................................................................34
4.3.4.4. testo 6381 as Saveris subscriber...................................................... 34
4.3.4.5. Integration into customer's Ethernet system..................................... 35
4.3.5. Adjusting the instrument................................................................................45
4.3.5.1. Analog output adjustment.................................................................45
4.3.5.2. n-point adjustment............................................................................ 46
4.4. Operation .......................................................................................47
5
3 Contents
4.4.1. Relationship between user menu and mini-DIN socket is active .................... 47
4.4.2. Key cover......................................................................................................47
4.4.3. Password protection ..................................................................................... 48
4.4.4. Structure of user menu..................................................................................49
4.4.5. Overview of the testo 6381 user menu..........................................................50
4.4.6. The individual main menus............................................................................53
4.4.6.1. Editing main menu of channel 1....................................................... 53
4.4.6.2. Editing Main Menu Alarm ................................................................. 53
4.4.6.3. Editing Main Menu Settings..............................................................55
4.4.6.4. Editing Main Menu Analysis ............................................................. 58
4.4.6.5. Editing Message main menu............................................................60
4.4.6.6. Calling up Main Menu Ident..............................................................61
4.4.6.7. Editing Main Menu Adjust.................................................................61
4.4.6.8. Editing Reset main menu ................................................................. 63
4.5. Status, warning and error messages............................................. 63
4.5.1. Status messages..........................................................................................64
4.5.2. Warning messages....................................................................................... 65
4.5.3. Transmitter error messages.......................................................................... 66
4.5.4. Handling alarm messages.............................................................................66
4.5.5. Namur fault conditions.................................................................................. 67
4.6. Maintenance and cleaning............................................................. 68
4.6.1. Maintaining the instrument............................................................................ 68
4.6.2. Cleaning the instrument................................................................................ 68
5 Parameterizing, adjusting and analyzing software (P2A software).. 69
5.1. Specifications ................................................................................ 69
5.1.1. Functions and use.........................................................................................69
5.1.2. System requirements....................................................................................70
5.1.3. Scope of delivery..........................................................................................70
5.2. First steps...................................................................................... 71
5.2.1. Installing the software/driver..........................................................................71
5.2.1.1. Installing P2A software..................................................................... 71
5.2.1.2. Installing USB driver......................................................................... 71
5.2.1.3. P2A software upgrade......................................................................71
5.2.2. Starting the software..................................................................................... 71
5.2.2.1. Starting the program ........................................................................ 71
5.2.2.2. Establishing a connection with the instrument..................................71
5.2.2.3. Activating the connection with the instrument................................... 72
5.3. Using the software......................................................................... 72
5.3.1. User interface...............................................................................................72
5.3.2. Editing instrument/parameter file...................................................................75
5.3.2.1. Changing instrument/parameter file..................................................75
5.3.2.2. Saving parameters........................................................................... 84
5.3.2.3. Opening the parameter file...............................................................85
5.3.2.4. Copying and pasting parameters......................................................85
5.3.2.5. Deleting instrument/parameter file....................................................86
5.3.2.6. Creating a new instrument file.......................................................... 86
6
3 Contents
5.3.3. Analyzing/testing the transmitter ................................................................... 86
5.3.3.1. Analyzing/testing the instrument.......................................................86
5.3.3.2. Carrying out factory reset.................................................................87
5.3.3.3. Testing analog output.......................................................................87
5.3.3.4. Testing switch output relays 1 to 4................................................... 89
5.3.3.5. Displaying min./max. values............................................................. 90
5.3.4. Adjusting the transmitter ............................................................................... 91
5.3.4.1. n-point adjustment............................................................................ 91
5.3.4.2. Adjusting the analog output..............................................................92
5.3.5. Transmitter history........................................................................................ 93
6 Tips and assistance...............................................................................97
6.1. Questions and answers..................................................................97
6.2. Accessories and spare parts..........................................................97
6.2.1. Ordering options for 6381 transmitter (0555 6381)........................................ 98
Pos: 18 /TD/--- Seitenwechsel --- @ 0\mod_1173774430601_0. doc @ 283 @
7
4 Transmitter
Pos: 19 /TD/Übers chrift en/MUF/1 Messumformer @ 3\mod_1234258401060_79.doc @ 23894 @ 1
4 Transmitter
Pos: 20 /TD/Übers chrift en/MUF/1. 1/ 2.1/3.1 Leistungsbeschreibung @ 3\ mod_1234258595211_ 79. doc @ 23951 @ 2
4.1. Specifications
Pos: 21 /TD/Leist ungsbeschreibung/ Verwendung/M UF63xx/ MUF 6388 @ 4\ mod_1252422611955_79. doc @ 48769 @ 3
4.1.1. Functions and use
The testo 6381 transmitter is suitable for the following applications with Ethernet networking, amongst others:
• Test benches
• Clean rooms
• Filling processes
• Painting systems
• Drying processes
• Monitoring flow velocities or volumetric flow rates in air conditioning systems
In addition to the signal transmission of the readings to a control unit via analog outputs, the measurement data can simultaneously be recorded, documented and visualized via Ethernet. Furthermore, it is possible to issue an alarm for those responsible
Pos: 22 /TD/Leist ungsbeschreibung/ Lieferum fang/ MUF 63xx/ MUF 63xx Ether net @ 3\mod_1234447308472_79.doc @ 25117 @ 3
for the process, if necessary.
4.1.2. Scope of delivery
The scope of delivery of the testo 6381 transmitter includes the following:
• Key cover
• Rear panel bracket
• Ethernet module
• Instruction manual
• Calibration report
• CD-ROM with operating instructions (PDF), configuration files for Ethernet module and P2A update (this can only be used in conjunction with the P2A software, which has to be ordered
Pos: 23 /TD/Leist ungsbeschreibung/ Lieferum fang/ MUF 63xx/ Zubehör Übersicht 635x @ 3\mod_1236084601953_79.doc @ 27292 @ 3
8
separately).
4.1.3. Accessories
The following accessories are available for the testo 6381 transmitter, amongst others:
Mains unit
• P2A software (parameterizing, adjusting and analyzing software)
• Assembly accessories
Information about accessories and their order numbers can be found in Accessories and spare parts or on the
Pos: 24 /TD/Leist ungsbeschreibung/ Technische Dat en/M UF 63xx/M UF 6388 Ethernet @ 4\m od_1249895102003_79.doc @ 47371 @ 3555555555555555555 55
website at www.testo.com.
4.1.4. Technical data
Parameters
• Differential pressure
4 Transmitter
Differential pressure accuracy
The specifications are only valid if the positive pressure is applied at the positive pressure connection.
• 0.5 % of measuring range, additional 0.3 Pa intrinsic error 1
• T
K slope drift
= 0.03 % of measuring range per degree Kelvin of
deviation from nominal temperature 22 °C
• T
K zeroing drift
1
Measuring uncertainty in accordance with GUM: ±0.8 % of measuring range
= 0 %, as zeroing with solenoid valve2
final value ±0.3 Pa.
GUM (Guide to the Expression of Uncertainty in Measurement): ISO guideline for determining the measuring uncertainty in order to render global measurement results comparable.
The following uncertainties are used during the inquiry:
• Hysteresis
• Linearity
• Reproducibility
• Adjustment area/factory calibration
2
Minor mixtures of the media may occur at the positive and negative pressure
• Test location
side due to the automatic zeroing cycle.
9
4 Transmitter
Pressure measuring range, resolution and overload of differential pressure
Pressure measuring range depending on version ordered
0 to 50 Pa 0.1 Pa 20,000 Pa 0 to 50 Pa 0.1 Pa 20,000 Pa 0 to 100 Pa 0.1 Pa 20,000 Pa 0 to 500 Pa 0.1 Pa 20,000 Pa 0 to 10 hPa 0.01 hPa 200 hPa 0 to 50 hPa 0.01 hPa 750 hPa 0 to 100 hPa 0.1 hPa 750 hPa 0 to 500 hPa 0.1 hPa 2500 hPa 0 to 1000 hPa 1 hPa 2500 hPa
-10 to 10 Pa 0.1 Pa 20,000 Pa
-50 to 50 Pa 0.1 Pa 20,000 Pa
-100 to 100 Pa 0.1 Pa 20,000 Pa
-500 to 500 Pa 0.1 Pa 20,000 Pa
-10 to 10 hPa 0.01 hPa 200 hPa
-50 to 50 hPa 0.01 hPa 750 hPa
-100 to 100 hPa 0.1 hPa 750 hPa
-500 to 500 hPa 0.1 hPa 2500 hPa
-1000 to 1000 hPa 1 hPa 2500 hPa
Upon delivery and following a factory reset the readings are shown in the display in the unit that was ordered via the KMAT option Fxx, see Ordering options for 6381 transmitter (0555 6381), page 98.
Resolution Overload
10
Meas. cycle
• 1/sec
Interface
• Mini-DIN for P2A software (parameterizing and adjusting
software) and handheld testo 400/650
4 Transmitter
Voltage supply
• 4-wire (separate signal and supply lines): 20 to 30 V AC/DC, 300 mA power consumption
Maximum load
• 4-wire: 500 Ω (power output)
Maximal load
• 4-wire: 10 k (voltage output)
Analog output
• 0 to 1 V ± 1.5 mV (4-wire) or
• 0 to 5 V ± 7.5 mV (4-wire) or
• 0 to 10 V ± 15 mV (4-wire) or
• 0 to 20 mA ± 0.03 mA (4-wire) or
• 4 to 20 mA ± 0.03 mA (4-wire)
Resolution of analog output
• 12 bit
Relay
• 4 relays, 250 V AC/DC, 3 A (optional)
11
4 Transmitter
Display
• 2-line LCD with plain text line (optional)
Operating temperature
• -5 to 50 °C/23 to 122 °F
Storage temperature
• -20 to 60 °C/-4 to +140 °F
Process temperature
• -20 to 65 °C/-4 to 149 °F
Housing, weight
• Metal: 1.960 kg
• Ethernet module: 0.610 kg
Protection class
• IP 65 only if the transmitter is wired properly (closed cable
entries), Ethernet connector and Harting PushPull connector are inserted and/or sealing plugs are inserted.
12
Directives, standards and tests
• EC Directive: 2004/108/EC
• DIN 14644-4
• EN 61000-6-2 interference immunity
• EN 61000-6-3 interference emission
• EN 61326-1+A1+A2
Ethernet module
• Interface:
1 x mini-DIN 1 x RJ45 (Ethernet 10 BaseT/100 BaseTX)
• LED:
2 x green
Warranty
• Duration: 2 years
• Warranty conditions: see website www.testo.com/warranty
Pos: 25 /TD/Leist ungsbeschreibung/ Technische Dat en/M UF 63xx/M UF 63xx Abmessungen @ 3\m od_1234450671494_79.doc @ 25194 @ 3
4.1.5. Dimensions
4 Transmitter
Dimensions in mm a b
with M20 cable couplings 144 147 With NPT cable coupling 144 144 With M plug-in connection 143
Pos: 26 /TD/Übers chrift en/MUF/1. 2/ 2.2 Produkt beschreibung @ 3\ mod_1234258723551_79. doc @ 24008 @ 2
4.2. Product description
Pos: 27 /TD/Pr oduktbeschr eibung/Übersicht /M UF 63xx/Auf einen Blick MUF 6388 Ethernet @ 4\m od_1252479453527_79.doc @ 48803 @ 3
4.2.1. At a glance
1 Keys (only with optional display) 2 Service flap screw connection
(self-locking, 2 pcs.) 3 Display (optional) 4 Service flap 5 Negative pressure connection 6 Positive pressure connection,
marked with a red washer 7 Lower part of housing 8 M 16 x 1.5 screw connection*,
e.g. analog outputs 9 M 16 x 1.5 screw connection*,
e.g. voltage supply 10 Earthing/PE connection
13
4 Transmitter
11 M 20 x 1.5 screw connection*,
e.g. R3 and R4 relays
12 M 20 x 1.5 screw connection*,
e.g. R1 and R2 relays 13 Adjusted probe plug 14 Upper part of housing * Alternatively, NPT cable
couplings or M plug-in
connections are available A Ethernet module 1 DIP switch 2 Ethernet port 3 LED: LAN connection status 4 LED: Supply
5 Housing screws 6 Hole for fastening to rear panel
bracket (M3 x 6 screw) 7 Plastic bracket for assembly on
rear panel
14
Pos: 28 /TD/Pr oduktbeschr eibung/Übersicht /M UF 63xx/Display und Tastatur @ 3\mod_1234773965059_79.doc @ 25650 @ 3
4.2.2. Display and keypad
The display option allows operation of the testo 6381 transmitter via the display and four keys.
The LCD display consists of two 7-segment lines for displaying readings and units and of an information line (for status messages, for example).
The brightness and contrast of the display and the background lighting (permanent or off) can be changed via the user menu or the
Pos: 29 /TD/Übers chrift en/MUF/1. 2. x Serviceschnitts telle @ 3\m od_1237306891654_79.doc @ 29795 @ 3
4.2.3. Service interface
Pos: 30 /TD/Pr oduktbeschr eibung/Übersicht /M UF 63xx/Serviceschnit tst elle 635x @ 3\mod_1237217388492_79. doc @ 29723 @
P2A software.
The parameterizing socket (mini-DIN) is located behind the service
Pos: 31 /TD/Pr oduktbeschr eibung/Übersicht /M UF 63xx/Relaisplatine (O pt ion) @ 3\mod_1234774184843_79.doc @ 25688 @ 3
flap as an interface to the P2A software.
4.2.4. Relay board (option)
This has a floating switch capacity of 250 V AC/3 A. The switching limits and hysteresis as well as the function as relay for the collective alarm can be set via the display or the P2A software.
Further features include:
• Function of changeover contacts (NC/NO contacts) freely selectable
Pos: 32 /TD/Pr oduktbeschr eibung/Übersicht /M UF 63xx/Analogausgänge 635x @ 3\m od_1234774510463 _79.doc @ 25726 @ 3
• 12 terminals for a total of 4 relays.
If no relays are available, settings for monitoring limit values or alarms can still be controlled via the display. The alarm status will be shown on the display.
Only have the transmitter wired and connected by authorized personnel with the voltage disconnected.
4 Transmitter
4.2.5. Analog outputs
As analog outputs, the testo 6381 has either
• 1 current output of 0 to 20 mA (4-wire)/4 to 20 mA (4-wire) or
Pos: 33 /TD/Pr oduktbeschr eibung/Übersicht /M UF 63xx/Mes sgrößen 635x @ 3\mod_1234775082160_79. doc @ 25764 @ 3
15
• 1 voltage output of 0 to 1 V/0 to 5 V/0 to 10 V (4-wire)
4 Transmitter
4.2.6. Parameters
The following parameters are displayed:
• Differential pressure in Pa, hPa, kPa, mbar, bar, mmH
O, inch HG, kg/cm2, PSI
H
2
3
in m/s, ft/min
• Pos: 34 /TD/Pr oduktbeschr eibung/Übersicht /M UF 63xx/Skalierung @ 3\m od_1234775406989_79.doc @ 25783 @ 3
• Flow
• Volumetric flow rate
4.2.7. Scaling
There are three types of min./max. values: 1 The measuring range: The maximum sensor performance is in
this range. Values outside of the measuring range are displayed via messages, for example. Measuring range, see table (below).
2 Standard scaling: The output signals are assigned to this
measuring range as standard:
during delivery if no entries are made in the order code after exchanging the unit, the measuring range recorded in
the instrument is applied as standard.
The transmitter even retains its scaling with the voltage disconnected.
Measuring range, see table (below). 3 The maximum settings for the manual scaling
The maximum limits can be calculated as follows: X = difference between MIN. and MAX. value of the
standard scaling (Max. value of standard) + (50 % of X) (Min. value of standard) - (50 % of X) It is thus possible to scale beyond the measuring range, e.g.
for the adjustment of the scaling limits to standard values of
a PLC.
4
in m3/h, l/min, Nm3/h, Nl/min
O, inch
2
3
To prevent fluctuating flow rate values at the zero point (depressurized), the flow rate values are only calculated as of differential pressures > 0.2 Pa or > 0.1 % of the respective measuring range (whichever is the greater). With smaller differential pressures, the flow rate value remains at 0.00 m/s.
4
Calculated 16
Pos: 35 /TD/Pr oduktbeschr eibung/Übersicht /M UF 63xx/Tabelle Skalierung MUF 638x @ 3\ mod_1236343950039_79. doc @ 27502 @
Measuring range/standard scaling
0 to 50 Pa -5 to 15 Pa 0 to 50 Pa -25 to 75 Pa 0 to 100 Pa -50 to 150 Pa 0 to 500 Pa -250 to 750 Pa 0 to 10 hPa -5 to 15 hPa 0 to 50 hPa -25 to 75 hPa 0 to 100 hPa -50 to 150 hPa 0 to 500 hPa -250 to 750 hPa 0 to 1000 hPa 500 to 1500 hPa
-10 to 10 Pa -20 to 20 Pa
-50 to 50 Pa -100 to 100 Pa
-100 to 100 Pa -200 to 200 Pa
-500 to 500 Pa -1000 to 1000 Pa
-10 to 10 hPa -20 to 20 hPa
-50 to 50 hPa -100 to 100 hPa
-100 to 100 hPa -200 to 200 hPa
-500 to 500 hPa -1000 to 1000 hPa
-1000 to 1000 hPa -2000 to 2000 hPa
Pos: 36 /TD/Pr oduktbeschr eibung/Übersicht /M UF 63xx/Alarmbehandlung @ 3\ mod_1234776787635_7 9. doc @ 25821 @ 3
4 Transmitter
With the alarm definition, however, the physical measuring range limits are decisive.
Maximum scaling
4.2.8. Alarm handling
For upper and lower alarm limits, individual alarms as well as collective alarms can be specified. If the collective alarm function is activated, an alarm is triggered as soon as the alarm limit of an alarm is exceeded, if this alarm is assigned to the collective alarm.
The testo 6381 monitors limit values with the help of relays. If a reading is outside the limit values, a relay to be specified by the user is switched.
If the reading reverts to more than a specified hysteresis below or above the limit value, the alarm is cancelled.
17
4 Transmitter
In addition, information about the occurrence of error/status messages can be provided by means of a collective alarm relay, see Status, warning and error messages, page 63
If multiple alarm messages are activated at the same time, the last alarm is shown. If the alarm is cancelled again, the
Pos: 37 /TD/Übers chrift en/MUF/1. 3/ 2.3 Inbetriebnahme @ 3\mod_1234258805768_ 79.doc @ 24027 @ 2
4.3. Commissioning
Pos: 38 /TD/Erste Schritte/MUF 63xx/Ethernet -spezifisch/Ether netm odul einsetzen 635x @ 3\mod_1238061282310_7 9. doc @ 30183 @ 3
previous messages are no longer shown.
4.3.1. Inserting Ethernet module (order no. 0554 6656)
The Ethernet module can be ordered retroactively as an accessory. It can easily be installed in the testo 6381 transmitter.
18
1. Loosen screw connection (1) of service flap and open the flap.
2. Loosen and remove housing screws (2).
3. Remove upper part of housing (3) and place on a clean surface.
4 Transmitter
4. Place Ethernet module (A) on lower part of instrument (4).
First set the desire operating mode via the DIP switch (see Setting the Ethernet module, page 30) before fixing the instrument in place.
5. Set on upper part of instrument (5) and fix in place using the housing screws (6) provided in the accessories.
19
4 Transmitter
Pos: 39 /TD/Erste Schritte/MUF 63xx/Wandmontage 635x _6321 @ 3\mod_1236065568006_79.doc @ 27063 @ 3455
4.3.2. Assembling the instrument
4.3.2.1. Wall mounting
Attaching rear panel bracket
1. Remove locking screw (see item (4) of drawing below) and
detach rear panel bracket from plastic bracket (see item (2) of drawing below).
2. Hold rear panel bracket in assembly position and mark the three
drill holes.
3. Drill three holes (Ø 5 mm) and insert dowels where necessary.
4. Screw on rear panel bracket.
Remember that the clamping brackets (1) must face the wall.
Fastening instrument to rear panel bracket
20
1. Slide plastic bracket (2) on the back of instrument onto rear
panel bracket until it engages (see arrows).
2. Insert screw (4) through hole (3) and screw into rear panel
Pos: 40 /TD/Erste Schritte/MUF 63xx/Gerät anschließen Ethernet @ 3\mod_1234779843987_79.doc @ 25897 @ 35
bracket.
4.3.3. Connecting the instrument
4 Transmitter
Opening the instrument
1. Loosen screw connection (1) of service flap and open the flap.
21
4 Transmitter
22
2. Loosen and remove housing screws (2).
The Ethernet module (A) is already detached from the upper and lower parts of the housing by removing the housing screws (2).
3. Remove upper part of housing (3) and place on a clean surface.
4. Remove Ethernet module (A) from lower part of housing (4) and
also place on a clean surface.
WARNING
El ec t r i c a l vo l t a ge Danger of injury!
> De-energize the mains connection before connecting the
transmitter.
Only have the transmitter wired and connected by
Pos: 41 /TD/Erste Schritte/MUF 63xx/Anschlussübersicht 635x @ 3\ mod_1234780293877_79. doc @ 25916 @ 4
4.3.3.1. Overview of terminals
authorized personnel with the voltage disconnected.
4 Transmitter
1 Lower part of housing 6 Terminal board 2 Relay board (option) 7 M 16 x 1.5 screw connection* 3 Relay terminals 8 Eyelet for measuring point
panel
4 Insulating trough for relay
9 M 20 x 1.5 screw connection*
board
5 Terminal strip for voltage
supply and analog outputs
* Alternatively, NPT cable
coupling or M plug-in connection
23
4 Transmitter
The following description of the terminals refer to this
Pos: 42 /TD/Erste Schritte/MUF 63xx/Spannungsversorgung/ Analogausgänge ansc hließen 6387/6388 @ 4\ mod_12506759569 13_79. doc @ 47503 @ 45
4.3.3.2. Connecting voltage supply and analog outputs
overview and its numbering.
Terminal strip for voltage supply and analog outputs, item (5) of overview of terminals
Channels 2 and 3 shown on the circuit board cannot be used with this instrument.
1. Feed cable with voltage supply and analog signal lines through
opened M 16 x 1.5 screw connection (item (8) in the overview of terminals).
2. Strip the cable ends, clamp wire end ferrules on and screw
down onto voltage terminals.
3. Close M 16 x 1.5 screw connection (item (8) in the overview of
terminals).
Wiring diagram for 4-wire system (0 to 20 mA/4 to 20 mA/0 to 1 V/0 to 5 V/0 to 10 V)
1-channel 0 to 20 mA/4 to 20 mA
max. load per 500 0 to 1 V/0 to 5
V/0 to 10 V
Channels 2 and 3 shown on the circuit board cannot be used with this instrument.
24
Requirement for the connecting cable of the supply:
• Insulated with cross-section of at least 0.25 mm².
• The supply line must be secured against exceeding 8 A.
• An OFF switch must be installed in an easily accessible position close by and be marked as such.
1. Feed connection cable of the channel through opened
M 16 x 1.5 screw connection (item (8) in the overview of terminals).
2. Strip the cable ends, clamp wire end ferrules on and screw to channel terminals as shown in diagram.
3. Close M 16 x 1.5 screw connection (item (8) in the overview of
Pos: 43 /TD/Erste Schritte/MUF 63xx/Relaisausgänge anschließen @ 3\ mod_1234786460813_ 79. doc @ 25992 @ 455555
4.3.3.3. Connecting the relay outputs
terminals).
Only have the transmitter wired and connected by authorized personnel with the voltage disconnected.
There is the option of twelve terminals for a total of four relays. The designations NC/C/NO (normally closed contact/root or pin/normally open contact) are etched on the surface of the board.
4 Transmitter
Relay terminal strip, item (3) of overview of terminals
Using PG screw connection
1. Feed connection cables for the relays through opened M 20 x 1.5 screw connection (item (10) of overview of terminals).
2. Strip cable ends and clamp on wire end ferrules.
3. Connect relays according to chosen function (NC/NO) (see diagrams below; relay 1 is shown as an example of a connection).
Using plug-in connections (optional)
Only insert or disconnect the plug-in connection when the voltage is disconnected.
4. Clean the connector of the probe line and the coupling of any foreign matter.
25
4 Transmitter
Do not disconnect the connector of the probe line from the instrument for extended periods to protect against contamination.
Connection note
• For the connection, a double-insulated mains cable (sheathed cable) with a cross-section of at least
1.5 mm² must be used.
• Cable connection (2) may not be routed in a loop within the tray (1).
• It is recommended that you always tie 3 cores to one another using a cable tie (3).
• The insulation of the cable must be fed at least 5 mm (4) into the tray.
26
Use of relay as NC contact (NC = normally closed)
1 Alarm/status light
(example of installation)
2 250 V AC/DC, 3 A
4 Transmitter
The busy light (alarm/status light) is permanently on until the relay opens or the circuit is interrupted. This circuit can therefore be used to monitor the functionality of the alarm circuit, as a cable break, for instance, is indicated by the busy light going off.
Use of relay as NO contact (NO = normally open)
1 Alarm/status light
(example of installation)
2 250 V AC/DC, 3 A
27
4 Transmitter
The busy light (alarm/status light) only comes on when the relay is switched (closed). Monitoring the functionality of the alarm circuit is therefore not possible with this switching operation.
5. Close M 20 x 1.5 screw connection (item (9) in overview of
Pos: 44 /TD/Erste Schritte/MUF 63xx/Option Stec kerver bindung 635x @ 3\m od_1237387714896_79.doc @ 30053 @ 45
terminals).
4.3.3.4. Plug-in connection option
As an option, the PG screw connections of the signal and supply lines can be replaced with plug-in connections that are installed at the housing, see item 1 and 2. The relay cabling occurs via standard cable entries and PG screw connections, see item 3 and
4.
28
Plug-in connections for power supply and channels
M12 plug-in connection (5-pin) socket (item 1) View of the plug-in connections in the installed state from outside
PIN Assignment
1 V 24­2 V 24+ 3 + Ch1 4 - Ch1 5 PE
Pos: 45 /TD/Erste Schritte/MUF 63xx/PE-/Erdungsanschluss her stellen 638x @ 3\m od_1234788670248_79.doc @ 26049 @ 455
4.3.3.5. Creating the PE/earthing terminal
As the testo 6381 has a metal housing, we recommend that the instrument be earthed. This can be done using the earthing terminal within the instrument (1) or the earthing terminal outside of the instrument (2).
Only use the external earthing terminal in dry interiors.
4 Transmitter
Using the earthing terminal within the instrument
1. Guide shielded cable (5) through the cable coupling (x) and fit cable lug (8). Fix this to the side of the instrument (6) using M 5 screw (3), washer (4) and snap ring (7) on the internal earthing terminal (1).
2. Place the other cable end on an appropriate functional earth, e.g. an earthing bar.
Using an earthing terminal outside of the instrument
1. Use PE line (yellow-green) (5) with cable lug (8). Fix this using M 5 screw (3), washer (4) and snap ring (7) on the external earthing terminal (2).
2. Place the other cable end on an appropriate functional earth, e.g. an earthing bar.
29
4 Transmitter
Pos: 46 /TD/Erste Schritte/MUF 63xx/Ethernet -spezifisch/Ether netm odul einstellen @ 3\mod_1234875066055_7 9. doc @ 26443 @ 4
4.3.3.6. Setting the Ethernet module
Using a DIP switch the Ethernet module can be switched between two main functions:
• Saveris subscriber function (DIP switch no. 1 = on, DIP switch no. 2 = on), testo 6381 is used as a testo Saveris component.
• XML server function (DIP switch no. 1 = off, DIP switch no. 2 = on), integration of the testo 6381 into the customer's Ethernet system.
In the condition on delivery, the switch position is set to on (Saveris mode) with the testo 6381 with Ethernet module.
The testo 6381 must be separated from the Ethernet network.
1. Open transmitter (see Opening the instrument, page 21).
2. Set DIP switch no. 1 (1) at Ethernet module. In the picture: DIP switch no. 1 off, DIP switch no. 2 on
3. Set Ethernet module on lower part of housing (see arrow).
5
DIP switch no. 2 non-functional
30
5
4 Transmitter
4. For the configuration of the Ethernet module, see following chapter.
5. If you do not wish to perform a configuration, close the transmitter.
31
4 Transmitter
Pos: 47 /TD/Erste Schritte/MUF 63xx/Ethernet -spezifisch/Gerät schließen Ethernet @ 3\ mod_1235986061555_79. doc @ 26965 @ 4
4.3.3.7. Closing the instrument
32
1. Place Ethernet module (A) on lower part of instrument (1). 2 Set on upper part of instrument (2) and fix in place using
housing screws (3).
4 Transmitter
3. Close the service flap and tighten screws (4).
4 Connect Ethernet network cable to the transmitter via the
Pos: 48 /TD/Erste Schritte/MUF 63xx/Ethernet -spezifisch/Komm unikation Ethernet @ 3\mod_1234877476523_79. doc @ 26502 @ 3444455
Ethernet jack.
4.3.4. Ethernet communication
4.3.4.1. Types of operation
In general, the Ethernet module performs two functions:
• a corresponding testo 6381 becomes a Saveris subscriber
• a corresponding testo 6381 can be integrated into any Ethernet systems
4.3.4.2. Mains connection
33
> Connect the network cable to the transmitter.
The network cable must not be connected directly to the telephone network (ISDN).
In order to ensure the IP protection class, an RJ 2.5 Harting PushPull connector must be used.
4 Transmitter
4.3.4.3. LED status displays
Characteristic LED 1 LED 2
Colour green green Status display for Voltage supply LAN connection Status: off No voltage supply No LAN connection Status: lights up Voltage supply present LAN connection
Status: flashes Data transfer
4.3.4.4. testo 6381 as Saveris subscriber
Setting the IP address
Upper part of housing must be removed.
Network cable must be connected to the Saveris base and to
the Ethernet module.
Service plug must be inserted.
The IP address of the transmitter can either be set via the P2A software (transmitter accessories) or via the configuration menu of testo Saveris
TM
(testo SaverisTM Ethernet Wizard).
present
34
1. Set the IP address with the P2A software (see Using the software, page 72) or with the testo Saveris
TM
Ethernet Wizard
(see instruction manual).
2. Disconnect service plug.
3. Disconnect the Ethernet module from the voltage supply.
4. Close instrument (see Closing the instrument, page 32).
Using Saveris
> Start Saveris software (see instruction manual Measurement
data monitoring with testo Saveris)
Ensure that the Saveris software is not already open, for example in multi-user operation under Windows
®
Vista. If multiple clients are installed in a network, make sure that no simultaneous changes are made to the system configuration by the clients during simultaneous operation of the clients.
4 Transmitter
1. [Start] | All Programs | Testo |
• Select Saveris Professional Client (full version). The entry is
available if Saveris Professional Client is installed
• Select Saveris Viewer (limited functionality). The entry is
available if Saveris Viewer is installed
- The Testo Saveris software program window is opened with the Select project dialogue.
If the software will not start, check whether the testo
tdassvcs service is started in the service management of
the operating system and restart it, if needed.
2. Select the
Only active projects option if the data from a running
project should be opened
All projects option if the data from a completed project
should be opened.
3. Select the project that is to be opened in the tree structure.
4. Confirm with [OK].
- The Testo Saveris software program window is shown with
Pos: 49 /TD/Erste Schritte/MUF 63xx/Ethernet -spezifisch/I nt egr ation Ethernet -Syst em 635x @ 3\mod_1234881594495_79.doc @ 26521 @ 4555555555
the selected data record in the foreground.
4.3.4.5. Integration into customer's Ethernet system
Precondition
The user should be familiar with the structure of an XML document. Furthermore, enough knowledge should be available that this type of document can be downloaded and decoded via an Internet connection with the aid of a programming/script language.
35
4 Transmitter
Interface
The communication takes place according to the client/server principle, whereby the Ethernet module takes on the role of the server:
The XML interface will be approached via a URL. The URL consists of the IP address of the Ethernet module, the path for the XML document and, depending on the URL, with a clearly defined parameter. If a parameter is missing or an incorrect value is transferred, the Ethernet module sends back a simple error message.
Example: IP address = 254.169.100.100 URL for serialnumber.xml: http://254.169.100.100/data/getserialnumber If a parameter is required, a "?" is always in the first place after the
URL, then the parameter is transferred to the common query string form (name=value).
Example: IP address = 254.169.100.100 URL for / identification.xml: http://254.169.100.100/config/getidentification?param=0 (compare
table xml codes) After calling up the compound URL, if everything was correct, an
XML document is returned. With faulty parameters a simple HTML answer comes that points out the reason that an XML answer could not be given.
If an XML document is to be loaded on the server, access is provided by means of a POST request
Example of an upload via the program wget: IP address = 254.169.100.100 URL for / usersettings.xml:
Path details of the folder in which the wget program is located
H:/wget/wget-complete-stable /wget --post-file= C:/usersettings.xml 254.169.100.100/config/
Comm and P ath de ta ils o f the
folder in which the usersettings XML file is located
IP address of the transmitter and required URL (see table on the nex t page)
setusersettings
The Ethernet module supports reading out of
• Readings
36
4 Transmitter
• Instrument type (testo 6381)
• Firmware date and version (testo 6381)
• Status and status messages (testo 6381)
• Alarm messages (testo 6381)
• Service hour counter (testo 6381)
as well as the reading and writing of the:
• Adjustment data (testo 6381)
• Configuration data of analog outputs (testo 6381)
• Configuration data of relays (testo 6381)
• User settings (testo 6381)
XML codes (download)
The table of XML codes is available for download at www.testo.de/transmitter as well as on the product CD
URL Description Parameter Answer (see
Appendix)
/data/getserialnumber Read off serial number of
/data/getidentification Read off type of
/data/getversion Read off firmware ve r sio n
/data/getfirmwaredate Re a d off firmware date of
/data/getonlinevalue Read off onl i n e values of
/data/getviewchannels Read off vie w channel s of
/data/getstatus Read off stat us of
/data/getlaststatusmessa ge
/c onfig/gethourscount Re ad off se rvice hour
/config/getusersettings Read off settings of
/config /getc a libration Re a d off a dj u stment data
connecte d trans m i tte r
connecte d trans m i tte r
of co nnected trans m itter
connecte d trans m i tte r
connecte d trans m i tte r
connecte d trans m i tte r
connecte d trans m i tte r Read off last status
me ss ages of co nnected transmitter
counter o f transm i tte r
transmitter
of transmitter
param =0 (for tra nsmitte r)
param=0 (adjustment set 1) param=1 (adjustment set 2)
param=2 (adjustment set 3)
serialnumber.xml
identification.xml
version.xml
firmwaredate.xml
onlinevalue.xml
viewchannels.xml
status.xml
laststatusmessage.xml
hourscount.xml
usersettings.xml
calibration.xml
37
4 Transmitter
URL Description Parameter Answer (see
Appendix)
/config/getreldefinition Read off relay info rmation
of transmitter
/config/getoptions Read off options of
transmitter
/c onfig/getcoll ectiveal arm Read off alarm me ssages
of transmitter
param =0 (rel ay 1) param =1 (rel ay 2) param =2 (rel ay 3)
param =3 (rel ay 4)
reldefinition.xml
options.xml
collectivealarm.xml
Upload XML documents
URL Description Parameter Post Answer (see
Appendix)
/config/setusersettings /config/setcalibration S et adj u stm ent da ta of
/config/setreldefinition Set relay information
/config/setoptions Set options of
/action/setresettm Not yet impleme n te d
Perform settings of transmitter
transmitter
of transmitter
transmitter
param=0 (adjustment set 1) param=1 (adjustment set 2)
param=2 (adjustment set 3)
param =0 (rel ay 1) param =1 (rel ay 2) param =2 (rel ay 3)
param =3 (rel ay 4)
usersettings.xml usersettings.xml calibration.xml calibration.xml
reldefinition.xml reldefinition.xml
options.xml options.xml
resettm.xml
Description of the XML elements
General elements
XML tag Description Type
measurement_value Parent element. Contains the chil d
elements value, unit, resolution value Reading Numeric a l , decim al num ber unit Unit ASCII number_values Qty. Numeri cal, whole number
38
4 Transmitter
Elements in calibration.xml
XML tag Description Type
calibration_data Base element. Contains the child
el em ents unit, attenuatio n, cal_reserved , cal_offset, cal_scale.
unit Se e general elem ents attenuation Damping (0 - 15) Numeri cal, decimal num ber
cal_offset Offset Numeri cal, whole number cal_scal e Parent e l ement. Contains the child
elemen ts cal_m in sca l e , cal_ m a xscale cal_minscale Scaling value Numerical, decimal number cal_maxscale Scaling value Numerical, decimal number
Elements in collectivealarm.xml
XML tag Description Type
colalarmtable Base element. Contains the child
alarm_numbers Number of alarm messages Nume ri cal, whole number alarm Pare nt ele me nt. Contains the chil d
alarm_event Type of ala rm ASCII alarm_state
el em ents alarm _num be rs, alarm
el em ents alarm _event, al arm_sta te
Status of the alarm
0 = alarm i nactive
1 = alarm active
Numerical, whol e num ber
Elements in firmwaredate.xml
XML tag Description Type
firmware_date Base element. Contains the child
el em ents ye ar, month, day year Year Numeri cal, whole number mo nth Month Numerical, whole number day Day Numerical, whole number
Elements in hourscount.xml
XML tag Description Type
hourcount Base element. Contains the child
el em ent hours hours S ervice hour co unter in h Num eric al , whol e num ber
39
4 Transmitter
Elements in identification.xml
XML tag Description Type
ident Base element. Contains the child
element device_id
device_id Transmitter type Numeri cal, whole number
Elements in laststatusmessage.xml
XML tag Description Type
mufmsg Base element. Contains the child
el em ents msg, sn, hours
msg Status m essage ASCII sn Se ri al numbe r ASCII, 8 characters hours S ervic e hour counter in h Numeri cal, whole number
Elements in onlinevalue.xml
XML tag Description Type
online_values Base element. Contains the child
el em ents number_val ues, measurement_value
number_values Se e general elem ents measurement_value See general elements
Elements in options.xml
XML tag Description Type
options Base element. Contains the child
ele m en ts devi ce_ options, production_options
device_options Se e de v i ce_options description Numeri cal, whole number production_opti ons S ee prod uction_optio ns descriptio n Numerical , whol e num ber
Elements in reldefinition.xml
XML tag Description Type
relay_data Base element. Contains the child
el em ents relay_channel, rel ay_number, rel ay_status, sw_point_character, sw_p oint_val ue, hys teresis_val ue
relay_channel M easurement channel l i nked to the
relay
rel ay_numb er Relay number (0 to 4) Num eric al , whol e num ber
Numeri cal, whole number
40
4 Transmitter
XML tag Description Type
relay_status
sw_point_charact
sw_p oint_value Switch p oint Numeri cal, decimal num ber hys teresis_value Hysteresis Num eric al , decim al num ber
Status of relay
0 = off
1 = o n
Switch point: Switc h point:
0 = low-li mit mo ni toring
1 = high-lim i t monito ring
Numerical, whol e num ber
Numerical, whol e num ber
Elements in serialnumber.xml
XML tag Description Type
serialnumber Base element. Contains the child
el em ent number number Se ri al number ASCII, 8 characte rs
Elements in status.xml
XML tag Description Type
mufstatus Base element. Contains the child
statemsg
staterel
statecounter Counter Numeric al , whol e num ber
ele m en ts st atems g, st a terel ,
statecounter
Status m essage
See statemsg description
Status relay
See staterel description
Numeri cal, whole number
Numerical, whol e num ber
Elements in usersettings.xml
XML tag Description Type
usersettings Base element. Contains the child
elements (pressure)*6,
(abs_pressure_pa_process)*,
(temperature_c_process)*,
(abs_pressure_pa_norm)*,
(temperature_c_norm)*,
(abs_pressure_pa)*, (area)*,
(correction_factor)*, (pi tot_factor)*,
setting_display, backlight, contrast,
lang uage, disp_msg pressure Absolute pressure Num eric al , decim al numbe r
* Chi ld elements are opt io nal
41
4 Transmitter
XML tag Description Type
setting_display
backlight contrast language
disp_msg
abs_pressure_pa_process, temperature_c_process abs_pressure_pa_norm humidity_norm temperature_c_norm area
correction_factor p ito t_ fa ctor
Auto OFF background lighting 0 -> b ackground lighting auto off
1 -> b ackground lighting stay s on Brightnes s of background lighting
0 to 9 (0 = off, 9 = max. ) Displ a y contrast 0 to 9 (0 = m i n., 9 = max.) Language 0 ->German 1 ->Engli sh 2 ->French 3 ->Spanis h 4 ->Ital i an 5 ->J apane se 6 ->Swedish Status m essages displ ay 0 = off 1 = o n Pr ess u re pr oc ess data, absolute pressure in Pa Pr ess u re pr oc ess data, pro cess temperature i n °C Pre ssure standard da ta, absol ute pressure in Pa Pre ssure s tandard data, process humidity in % RH Pre ssure s tandard data, process temperature i n °C Cross-section of the duct in mm²
Correction factor Pi tot tube facto r
Numerical, whol e num ber
Numerical, whol e num ber Numerical, whol e num ber Numerical, whol e num ber
Numerical, whol e num ber
Numeri cal, decimal num ber Numeri cal, decimal num ber Numeri cal, decimal num ber Numeri cal, decimal num ber Numeri cal, decimal num ber Numeri cal, decimal num ber
Numeri cal, decimal num ber Numeri cal, decimal num ber
Elements in versions.xml
XML tag Description Type
firmware_version Base element. Contains the child
element version
version Firm ware v ersio n: ASCII, 6 characte rs
Elements in viewchannels.xml
XML tag Description Type
view_channels Base element. Contains the child
el em ents number_val ues, view_channel
number_values Se e general elem ents
42
4 Transmitter
XML tag Description Type
view_channel Pare nt ele me nt. Contains the chil d
elements channel_info,
me asureme nt_v al ue, me as_status channel_info Pare nt ele me nt. Contains the chil d
elements connector_info, channel_type measurement_value See general elements me as_status Parent ele me nt. Contains the child
el em ents mi n, max, mean connector_info Channel (transmitter) ASCII channel_type Parameter details ASCII min. M i nimum readi ng Numeri cal, decimal num ber max. Max i m um read i ng Nume ri cal, decimal num ber me an Mean v al ue Numeri cal, decimal num ber
production_options description
Content of production_options is a double word type number (32 bit). The individual hardware options are bit-coded here.
xxxx xxxx xxxx xxxx xxxx xxx1 2222 3334
1 0=2-wire
1=4-wire 2 free 3 0=4 to 20 mA
1=0 to 20 mA
2=0 to 1 V
3=0 to 5 V
4=0 to 10 V 4 0=1 analog output
device_options description
Content of device_options is a double word type number (32 bit). The individual instrument options are bit-coded here.
xxxx xxxx xxxx xxxx xxxx xxxx 1222 2345
1 free 2 free 3 free
43
4 Transmitter
xxxx xxxx xxxx xxxx xxxx xxxx 1222 2345
4 0=no relay present
1=relay present
5 0=no display
1=display present
statemsg description
Content of statemsg is a double word type number (32 bit). The individual status messages are bit-coded here.
If statemsg = 0 there are no new messages. If statemsg != 0
xxxx xxxx xxxx xxxx xxxx xxxx 1234 5678
1 free
2 free
3 free
4 free
5 free
6 1=transmitter information
7 1=transmitter warning
8 1=transmitter error
44
staterel description
Content of staterel is a double word type number (32 bit). The individual relay states are bit-coded here.
xxxx xxxx xxxx xxxx xxxx xxxx xxxx x123
1 0=relay 4 is not set
1=relay 4 is set
2 0=relay 3 is not set
1=relay 3 is set
3 0=relay 2 is not set
1=relay 2 is set
4 0=relay 1 is not set
1=relay 1 is set
More information can be found in the download area at www.testo.com.
Pos: 50 /TD/Übers chrift en/MUF/1. 3. 3 Gerät abgleichen @ 3\ mod_1236081099337_79. doc @ 27173 @ 3
4.3.5. Adjusting the instrument
Pos: 51 /TD/Erste Schritte/MUF 63xx/Analogausgangs-Abgleic h 635x @ 3\m od_1237373284088_79.doc @ 29825 @ 45
4.3.5.1. Analog output adjustment
The purpose of adjusting the analog outputs is to adjust the signal chain from the digital signal (within the transmitter) to the analog outputs. The signal type that was appointed for the transmitter is adjusted respectively for each channel (e.g. 4 to 20 mA or 0 to 1 V, etc.).
4 Transmitter
1 Status LED
2. Contact ch. 1+
3. Contact ch. 1-
4. Service interface
Adjusting analog output 1
With testo 6381 with current output: Load of max. 500Ω is
connected to channel 1 (see Plug-in connections for power supply and channels, page 28)
A precise multimeter (minimum requirement: resolution 6.5
digits, at least 5-times more accurate than the 6381) is available.
If only a simple multimeter is available, the analog output must not be adjusted.
The service flap is open.
45
4 Transmitter
1. Connect the inputs of the multimeter with the contacts (2) and (3) for channel 1.
2. Transfer the reference analog value measured with the multimeter into the P2A software (see Adjusting the analog output, page 92) or enter it via the user menu (see Performing analog adjustment, page 61).
3. Disconnect connections between the multimeter and the
Pos: 52 /TD/Erste Schritte/MUF 63xx/n-Punkt- Abgleich _635x @ 3\mod_1236352382511_79. doc @ 27522 @ 4
contacts of the testo 6381 and close the service flap.
4.3.5.2. n-point adjustment
With an n-point adjustment, the parameters at the 3-6 measurement points are adjusted to the reference value. The reference conditions are obtained by using a precise pressure sensor that should be 5-times more accurate than the transmitter.
1 Positive pressure connection
2. Negative pressure connection
3. Pressure sensor
46
The number of measuring points is set to 3 by the factory and can only be changed using the P2A software (see n- point adjustment, page 91).
The n-point adjustment must always be carried out to its full extent and in good time at all selected measurement points.
A precise pressure sensor (5-times more accurate than the
transmitter, e.g. DPC precision pressure sensor from testo industrial services) is available.
1. Connect the positive output of the pressure sensor (3) to the positive pressure connection of the transmitter (1) and the negative output of the pressure sensor (3) to the negative pressure connection of the transmitter (2).
2. Transfer the reference pressure value created with the pressure sensor into the P2A software (see n-point adjustment, page 91) or enter it via the user menu (see Performing analog adjustment, page 61).
3. Repeat step 2 for all of the measuring points.
4 Transmitter
4. Disconnect connections between the pressure sensor and the
Pos: 53 /TD/Übers chrift en/MUF/1. 4 Bedienung @ 3\mod_1234443160034_79. doc @ 25001 @ 2
4.4. Operation
Pos: 54 /TD/Pr odukt verwenden/MUF 63xx/ Zusamm enhang Bedienmenü – Mini DIN Buchse aktiv @ 3\m od_1234454016014_79.doc @ 25213 @ 3
pressure connections of the testo 6381.
4.4.1. Relationship between user menu and mini-DIN socket is active
The testo 6381 can be parameterized using either the user menu or the P2A software (see Parameterizing, adjusting and analyzing software (P2A software) page 69).
The testo 6381 transmitter can only be operated via the display and keypad if the display option is available.
If the testo 6381 is connected to the P2A software, the user menu is blocked for the duration of the communication. The message Service plug is shown in the display of the testo
6381. As soon as the P2A software is disconnected, the
Pos: 55 /TD/Pr odukt verwenden/MUF 63xx/ Tastenblende_Ether net @ 3\ mod_1237385298320_79. doc @ 30011 @ 35
user menu is accessible again.
4.4.2. Key cover
To prevent unauthorized operation of the keys, the standard key frame can be replaced with a key cover.
If the key cover has been assembled, the service flap must be opened for operation.
47
4 Transmitter
Pos: 56 /TD/Pr odukt verwenden/MUF 63xx/ Passworts chutz _635x @ 3\mod_1237383967006_79.doc @ 29969 @ 3
Attaching the key cover
The service flap is opened, see Opening the instrument, page
21.
1. Unscrew screws (3) and remove key frame (2).
2. Insert key cover (1) into service flap and tighten screws (3).
3. Close and screw down the service flap.
4.4.3. Password protection
The user menu can be protected with a four-digit numerical code (see Editing Main Menu Settings, page 55) so that access to the user menu is denied to unauthorized persons not familiar with this numerical code.
If the password protection is not to be used, the numerical code
Pos: 57 /TD/Pr odukt verwenden/MUF 63xx/ Aufbau des Bedienmenüs 635x @ 3\ mod_1236162940119_79. doc @ 27333 @ 3
48
"0000" must be entered. This is also the status upon delivery.
4.4.4. Structure of user menu
At the main menu level, the user menu comprises the following:
• Main Menu Channel 1
• Main Menu Alarm
• Main Menu Settings
• Main Menu Analysis
• Main Menu Messages
• Main Menu Ident
• Main Menu Adjust
• Main Menu Reset
4 Transmitter
1 Channel 1 display 2 No display 3 Display for messages
Four keys enable the user to navigate/scroll through the menus and enter/amend values and settings:
Key Function/description
SET • In Measuring Mode: changes to
parameterization
• In Parameterizing Mode: confirms a selection or setting
ESC • Leaves a menu (without modifying any
settings)
X • Selecting: scrolls through menus
(downwards) or selectable alternatives
• Editing: changes to next digit (to the right)
49
4 Transmitter
Key Function/description
S • Selecting: scrolls through menus (upwards)
or selectable alternatives
• Editing: increases the value of the current
digit by 1
Pos: 58 /TD/Pr odukt verwenden/MUF 63xx/ Übersicht Bedienmenü @ 3\ mod_1234510821302_79. doc @ 25303 @ 3
4.4.5. Overview of the testo 6381 user menu
50
4 Transmitter
51
4 Transmitter
52
Pos: 59 /TD/Pr odukt verwenden/MUF 63xx/ Hauptmenü Kanäle bearbeit en @ 3\m od_1234511119158_79.doc @ 25322 @ 34
4.4.6. The individual main menus
4.4.6.1. Editing main menu of channel 1
An overview is given in Overview of the testo 6381 user menu, page 50).
You can perform basic settings for channel 1.
1. In the Measuring Mode press SET, select Main Menu
Channel 1 with X or S and confirm selection with SET.
One of the following parameters can now be selected using X or S, after which the selection must be confirmed with SET:
• Channel 1 Unit
The parameter for this channel is selected. Edit/select parameter with X or S, confirm with SET or abort
entry with ESC.
• Scale minimum for channel 1:
The lower scale limit is edited; Unit as selected above. Editing the value: Scroll one digit to the right using X and
increase value of digit by 1 using S. Confirm with SET or abort entry with ESC.
• Scale maximum for channel 1
The upper scale limit is edited; Unit as selected above. Editing the value: Scroll one digit to the right using X and
increase value of digit by 1 using S. Confirm with SET or abort entry with ESC.
• Signal delay ("Attenuation") for channel 1
The analog signal can be delayed ("Attenuation"); a time constant is selected for this (1 = no delay; 15 = longest delay)
Edit/select parameter with X or S, confirm with SET or abort entry with ESC.
2. Continue to the main menu with X or S or return to Measuring
Pos: 60 /TD/Pr odukt verwenden/MUF 63xx/ Hauptmenü Alarm bearbeiten 635x @ 4\mod_1245756301576_79.doc @ 45273 @ 455
Mode with ESC.
4 Transmitter
4.4.6.2. Editing Main Menu Alarm
With the alarm, the relays, available as options, are programmed. In addition, the alarm statuses are shown on the display (top right) (even without relays).
53
4 Transmitter
You can choose whether the alarm is to be used to monitor limit values or as a collective alarm. If an alarm is to be used to monitor limit values, you can choose between monitoring the minimum or maximum value and set a limit value and hysteresis for each alarm.
In addition, every alarm can be linked to a clearly visible visual alarm (display background lighting flashes).
An alarm delay between 0 and 240 seconds can still be assigned to every alarm used for limit value monitoring so that both the corresponding relay effect and the visual alarm are delayed. If the alarm status goes out within the set alarm delay time, neither the visual alarm nor a relay connection is triggered. .
With an alarm status present, the visual alarm and all relay outputs can be reset by means of acknowledgement. The triggering of a new alarm cannot be enabled until after the alarm status goes out.
1. In the Measuring Mode press SET, select Main Menu Alarm with X or S and confirm selection with SET.
- Four alarms can be parameterized.
2. Select Alarm x with X or S and confirm selection with SET.
54
Using alarm to monitor limit values
NO contact
Monitoring minimum Monitoring maximum
Hysteresis
Limit value Limit value
Hysteresis
NC contact
Monitoring minimum Monitoring maximum
Hysteresis
Limit value Limit value
Hysteresis
3. Select Channel x (e.g. "Channel 1") with X or S and confirm selection with SET.
4. Select Max control or Min control with X or S (see graphic).
5. Press SET and edit Limit value as well as Hysteresis: Scroll one digit to the right using X and increase value of digit by 1 using S. Confirm with SET or abort entry with ESC.
6. Select Visual alarm with X or S. Select YES or NO with X or
7. Press SET and edit Alarm delay: Scroll one digit to the right
8. Return to Channel x with ESC.
9. Return to Alarm x with ESC.
10. Change to the other relays using X or S and perform settings
Using alarm as collective alarm or not using it at all
If an alarm is assigned to the collective alarm, the relay is switched and a visual alarm can be issued via the display as soon as (at least) one of the warning or error messages of the testo 6381 transmitter becomes active.
Alarm is selected (see previous steps 1 and 2).
1. Use X or S to determine whether Alarm x should be used as a
2. If collective alarm is selected: Select Visual alarm with X or S.
3. Change to another alarm using X or S and perform settings in
4. Return to Main Menu Alarm with ESC.
5. Continue to Main Menu Settings with X or S or return to
Pos: 61 /TD/Pr odukt verwenden/MUF 63xx/ Hauptmenü Einst ellungen bearbeit en 635x @ 3\m od_1236608659829_79.doc @ 27573 @ 4555555
4 Transmitter
S. Confirm with SET or abort entry with ESC.
using X and increase value of digit by 1 using S (0 to 240 seconds possible). Confirm with SET or abort entry with ESC.
in the same way.
The messages affecting the collective alarm can only be selected in the P2A software, see Using the software, page 72
Collective alarm or not used. Confirm selection with SET.
Select YES or NO with X or S. Confirm with SET and return to
Alarm x.
the same way.
Measuring Mode with ESC.
4.4.6.3. Editing Main Menu Settings
You can edit instrument settings and other settings. > In Measuring Mode, press SET, select Main Menu Settings
using X or S and confirm selection with SET.
You can edit settings for:
• Display
• Language
• Code
• Units
55
4 Transmitter
Absolute pressure Area Temperature Standard data Process data
Editing display settings
You can set the brightness and contrast of the display.
1. Select Display Settings with X or S and confirm selection with
SET.
2. Select Backlight or Contrast with X or S and confirm selection with SET.
One of the following parameters can now be selected using X or S, after which the selection must be confirmed with SET:
Backlight The display illumination is changed. Edit/select parameter with X or S, confirm with SET or cancel
entry with ESC (the effect of the change in parameter can be seen during input).
Contrast The brightness difference between the display background and
the displayed values is changed. Edit/select parameter with X or S, confirm with SET or cancel
entry with ESC (the effect of the change in parameter can be seen during input).
Backlight 24h on Select On or Off using X or S and confirm with SET. Off: The display light switches off automatically if no button was
pressed for 10 seconds. On: The display light is activated
3. Return to Display Settings with ESC and use continue to Language.
X or S to
56
Selecting language
You can select the language for the plain text line in the display. > Press SET, select required language with X or S, confirm
selection with SET and return to Language.
4 Transmitter
Only choose a language that you can understand well.
Select unit
This setting affects the unit of the standard and process data.
1. Press SET, select Change parameters with X or S, confirm selection with SET or cancel with ESC.
2. Select Change unit with X or S, confirm selection with SET or cancel with ESC.
3. Select the required variable (absolute pressure/area/temperature) with X or S, confirm selection with
SET or cancel with ESC.
4. Select the required unit with X or S, confirm selection with SET or cancel with ESC.
5 Return to Change unit with ESC and continue to Standard
data with X or S.
Editing standard data
Setting individual values for the standard data to calculate the volumetric flow rate.
1. Select Standard data with X or S, confirm selection with SET or cancel with ESC.
2. Select the required variable (absolute pressure/temperature) with X or S, confirm selection with SET or cancel with ESC.
3. Scroll one digit to the right using
X and increase value of digit
by 1 using S. Confirm with SET or abort entry with ESC.
4. Return to Standard data with ESC and useX or S to continue to Pressure process data.
Editing pressure process data
Setting of the process data for the Pitot tube calculation.
1. Select Pressure process data with X or S and confirm selection with SET.
2. Select the required variable (absolute pressure/humidity/temperature/cross-section/Pitot tube factor/correction factor) with X or S, confirm selection with SET or cancel with ESC.
3. Scroll one digit to the right using X and increase value of digit by 1 using S. Confirm with SET or abort entry with ESC.
4. Return to Pressure process data with ESC.
57
4 Transmitter
5. Return to Change parameters with ESC and useX or S to continue to Code.
Editing code settings
You can set the access code (password).
If a code other than "0000" (factory setting) is set, the transmitter can only be operated once this code has been entered via the menu.
1. Select Code with X or S and confirm selection with SET.
2. Scroll one digit to the right using X and increase value of digit by 1 using S. Confirm with SET or abort entry with ESC.
3. Return to Code with ESC.
Pos: 62 /TD/Pr odukt verwenden/MUF 63xx/ Hauptmenü Analyse bear beiten 635x @ 3\ mod_1237277461094_79. doc @ 29753 @ 45555
4. Return to Main Menu Settings with ESC.
4.4.6.4. Editing Main Menu Analysis
You can test the functionality of analog and relay outputs. In addition, you can read off the minimum and maximum values (since the last voltage supply or reset of the min./max. values).
58
Testing functionality of analog outputs
This function affects the analog outputs directly, not only the test contacts.
1. In the Measuring Mode press SET, select Main Menu Analysis with X or S and confirm selection with SET.
- Test Analog Output is shown.
2. Press SET, scroll one digit to the right using X and increase value of digit by 1 using S. Any analog output value can be predefined, e.g. for an analog output of 4 to 20 mA, the value "6.0 mA". Confirm with SET or abort entry with ESC.
3. Accept setting by pressing SET and test with multimeter (minimum requirement: resolution 6.5 digits, at least 2-times more accurate than the 6381) as follows:
Analog output 1: Via test contacts under service flap, see diagram.
4 Transmitter
1 Channel 1
test contacts
2 Service
interface
3 Multimeter
4. Return to Test Analog Output with ESC and useX or S to continue to Test Relay Output.
Testing functionality of the pressure sensor (Test pressure
sensor)
This function is only required to calibrate the pressure sensor.
Testing functionality of relay outputs
1. Press SET and choose between Alarm 1, 2, 3, 4 with X or S.
2. Press SET. The relay can now be tested. You can choose between OFF
and ON using X or S. If ON is chosen, the NO contact is closed, the NC contact opened. If OFF is chosen, the NC contact is closed, the NO contact opened.
3. To test, route a measuring cable from the relay terminals (see Connecting the relay outputs, page 25) out of the transmitter to a multimeter (resistance measurement) or continuity tester.
4. Return to Test Relay Output with SET (starts relay test) or
ESC (exits the menu without relay test).
Reading off min./max. values of channels
To reset the max./min. values, see Editing Reset main menu, page 63
1. Read off the min./max. values of the channel in succession with X or S and return to the Main Menu Analysis with ESC.
2. Continue to Main Menu Message with X or S or return to Measuring Mode with ESC.
59
4 Transmitter
Pos: 63 /TD/Pr odukt verwenden/MUF 63xx/ Hauptmenü M eldungen bearbeiten @ 3\ mod_1234526541417_79. doc @ 25398 @ 4
4.4.6.5. Editing Message main menu
Messages can be confirmed/acknowledged, the last messages can be called up and the display of the messages can be switched on or off.
Using the P2A software (see Parameterizing, adjusting and analyzing software (P2A software), page 69) you can predefine whether messages are to be shown in the display.
1. In the Measuring Mode press SET, select Main Menu Message
with X or S and confirm selection with SET.
2. Confirm Confirm message using SET.
3. Select Last messages with X or S and confirm with SET.
4. Scroll between the warning and error messages recorded so far
using X or S and return to Last messages using ESC.
5. Select Information with X or S and confirm with SET.
6. Scroll between the status messages recorded so far using X or
S and return to Information using ESC.
7. Continue to Display of message with X or S.
ON: Measurements are shown on the display in Measuring
Mode.
OFF: No messages shown on display.
8. Select ON or OFF with X or S and confirm selection with SET.
9. Return to Main Menu Message with ESC.
1 Operating hours at the
time of message
2 Message code (see
Status, warning and error
messages, page 63). 3 Message text 4 Message number
(example: "4/7" refers to
the fourth of seven
messages) 5 Number of messages
present (example: "4/7"
refers to the fourth of
seven messages)
60
10. Continue to Main Menu Ident with X or S or return to Measuring Mode with ESC.
An overview of the messages can be found in Status,
Pos: 64 /TD/Pr odukt verwenden/MUF 63xx/ Hauptmenü I dent abfr agen 635x @ 3\mod_1238158282021_79.doc @ 30233 @ 4
4.4.6.6. Calling up Main Menu Ident
warning and error messages, page 63
4 Transmitter
1 Instrument type 2 Build number 3 Firmware
version
4 Serial number
1. In the Measuring Mode press SET, select Main Menu Ident with X or S and confirm selection with SET.
- The type, firmware version, build number and serial number of the transmitter are displayed.
This information is required when servicing.
2. Return to Main Menu Ident with ESC.
3. Continue to Main Menu Adjust with X or S or return to
Pos: 65 /TD/Pr odukt verwenden/MUF 63xx/ Hauptmenü Abgleich bearbeiten 635x @ 3\m od_1236081643663_ 79.doc @ 27232 @ 455
Measuring Mode with ESC.
4.4.6.7. Editing Main Menu Adjust
Also see Analog output adjustment, page 45, for instructions on
61
how to perform the analog adjustment.
Performing analog adjustment
1. In the Measuring Mode press SET, select main menu Adjust with X or S and confirm selection with SET.
Each channel is adjusted at three points in the analog range (at 10 %; 50 %; 90°% of the analog scale).
2. Select Analog Adj. Ch. 1 with X or S and confirm with SET.
3. Use X or S to select Adj. Point 1.
4 Transmitter
Pos: 66 /TD/Pr odukt verwenden/MUF 63xx/ Hauptmenü Reset bearbeiten 635x @ 3\mod_1238480830340_79.doc @ 30373 @ 4
4. Press SET. Read off multimeter display and enter this value in
the user menu. Do this by scrolling one digit to the right using X and increasing the value of digit by 1 using S. Confirm with
SET or abort entry with ESC.
5. Use X or S to select Adj. Point 2.
6. Press SET. Read off multimeter display and enter this value in
the user menu. Do this by scrolling one digit to the right using X and increasing the value of digit by 1 using S. Confirm with
SET or abort entry with ESC.
7. Use X or S to select Adj. Point 3.
8. Press SET. Read off multimeter display and enter this value in
the user menu. Do this by scrolling one digit to the right using S and increasing the value of digit by 1 using X. Confirm with
SET or abort entry with ESC.
9. Return to main menu Adjust with ESC.
10. Continue to main menu Reset with X or S or return to Measuring Mode with ESC.
Performing the pressure adjustment
Also see n-point adjustment, page 46, for instructions on how to perform the pressure adjustment.
1. In the Measuring Mode press SET, select main menu Adjust with X or S and confirm selection with SET
.
The pressure adjustment can be performed at 3 to 6 adjustment points.
2. Select Adjust Pressure with X or S and confirm with SET.
3. Use X or S to select Adj. Point 1.
4. Press SET. Read off pressure sensor display and enter this value in the user menu. Do this by scrolling one digit to the right using X and increasing the value of digit by 1 using S. Confirm with SET or abort entry with ESC.
5. Repeat steps 1 to 4 for adjustment points 2 to 6.
6. Return to main menu Adjust with ESC.
7. Continue to main menu Reset with X or S or return to Measuring Mode with ESC.
62
4.4.6.8. Editing Reset main menu
You can reset the factory settings for the following:
• Instrument
Pos: 67 /TD/Übers chrift en/MUF/1. 5 Status-, Warn- und Fehler meldungen @ 3\ mod_1234442925442_79. doc @ 24963 @ 2
4.5. Status, warning and error messages
Pos: 68 /TD/Pr odukt verwenden/MUF 63xx/ Stat us-War nmeldungen/Status - , Warn- und Fehlermeldungen 635x @ 3\mod_1237458634551_7 9. doc @ 30125 @
Pos: 69 /TD/Pr odukt verwenden/MUF 63xx/ Stat us-War nmeldungen/Status meldungen 635x @ 3\m od_1235651943832_79.doc @ 26823 @ 3
Min./max. values
Resetting to the factory settings means resetting to the order specification, i.e. the specific condition at the time of supply to the customer.
1. In the Measuring Mode press SET, select Main Menu Reset with X or S and confirm selection with SET.
- Reset Device to factory settings is displayed.
2. Select the setting to be reset using X or S and confirm selection with SET.
- Reset Completed is displayed.
3. Return to the reset setting with ESC or SET and return to Main
Menu Reset with ESC.
4. Continue to the Main Menu Channel 1 with X or S or return to Measuring Mode with ESC.
To achieve optimum operational reliability (machine availability), the transmitter shows the following via the user menu or the P2A software
• Status messages
• Warning messages
• Error messages
in each case for the testo 6381 transmitter. All messages are stored in the transmitter with an operating hours
stamp. Use the user menu (see Editing Message main menu, page
60) or the P2A software (see Transmitter history, page 93) to view
the message history. In the transmitter, the last 60 status messages and the last 120
error and warning messages are stored in a ring memory. There is no limit in the P2A software.
4 Transmitter
63
4 Transmitter
4.5.1. Status messages
Status messages show the current operating status of the testo
6381.
Message Display Description
02506 Sensor
initialization
01D19 Service plug The Mini-DIN socket is connected
00300 New limit value The limit value has been changed
00301 Scaling changed The scaling has been changed 00117 Adjustment DeltaP An n-point adjustment was
02104 Analog adjustment An analog adjustment has been
00530 Change solenoid
valve
00500 Transmitter reset The transmitter was reset to the
0052F Reset Min/Max Resets all saved Min/Max values
00503 Reset device to
fact
00307 User Setting
Change
Message appears while the transmitter is starting up. If the message disappears, the transmitter is ready for operation.
to: the USB adapter for P2A software, the adjustment adapter or the service plug (is not recorded/no number)
or shifted
performed.
made The solenoid valve should be
changed
factory settings and is restarted.
for all channels A factory reset of the transmitter
was performed User Setting Change: General
settings were changed for the transmitter.
Pos: 70 /TD/Pr odukt verwenden/MUF 63xx/ Stat us-War nmeldungen/Warnm eldungen 635x @ 3\m od_1236081573428_79.doc @ 27192 @ 3
64
4 Transmitter
4.5.2. Warning messages
Warning messages show an early warning or a current malfunction
Message Display Cause Remedying of fault
00E00 T ambient high The ambient
00E01 T ambient low The ambient
00809 Pressure too
00E02/00E 05
0081C Alarm 1 Depending on the
0081D Alarm 2 Depending on the
0081E Alarm 3 Depending on the
0081F Alarm 4 Depending on the
Pos: 71 /TD/Pr odukt verwenden/MUF 63xx/ Stat us-War nmeldungen/Fehlermeldung en Messum for mer 635x/6781 @ 3\mod_1236081613819_79. doc @ 27212 @ 3
which may negatively impact measuring.
temperature exceeds the permissible temperature for the transmitter
temperature is below the permissible temperature for the transmitter
The process pressure
high
exceeds the pressure intended for the transmitter
Supply voltage low
The supply voltage is below the minimum voltage required for the transmitter
parameterization of the relay
parameterization of the relay
parameterization of the relay
parameterization of the relay
Take necessary measures to lower ambient temperature, e.g. through venting or cooling
Take necessary measures to raise ambient temperature, e.g. through heating
Remove the transmitter from the process and take any necessary measures to lower the pressure
Take measures to ensure sufficient voltage supply
Depending on the parameterization of the relay
Depending on the parameterization of the relay
Depending on the parameterization of the relay
Depending on the parameterization of the relay
65
4 Transmitter
4.5.3. Transmitter error messages
Message Display Cause Remedying of fault
01505 Watchdog error Due to a processor
01115 Low adjustment
01116 High adjustment
Pos: 72 /TD/Pr odukt verwenden/MUF 63xx/ Stat us-War nmeldungen/Behandlung von Alarmmeldungen 635x @ 3\mod_1236081852932 _79.doc @ 27252 @ 3
4.5.4. Handling alarm messages
Error messages show a current malfunction.
If the problem occurs error, the transmitter performs an automatic
frequently, contact
Testo Service. restart.
temperature
temperature
Shown on the display
7
The ambient temperature is too low during the pressure adjustment.
The ambient temperature is too high during the pressure adjustment.
Can be used for collective alarm8
Take necessary
measures to raise
ambient temperature,
e.g. through heating.
Take necessary
measures to lower
ambient temperature,
e.g. through venting.
Additional message end9
New limit value x Scaling changed x Adjustment DeltaP x Pressure too high x x Alarm 1 x Alarm 2 x
7
If multiple messages/alarms are activated at the same time, only the last message/alarm is shown. If this is cancelled, the other messages that are still active are no longer displayed.
8
The message can be assigned the collective alarm function, which means that the collective alarm is activated as soon as at least one of the messages assigned to it is activated. The collective alarm can be assigned to each of the 4 optional relays. The collective alarm is then always the same, as it can only be defined once.
9
The message is shown upon the occurrence of the event causing the message as well as when closing. Two entries appear in the history in the P2A software: Message text_start and Message text_end.
66
4 Transmitter
Shown on the display
7
Can be used for collective alarm8
Alarm 3 x Alarm 4 x Transmitter reset x Analog adjustment x T ambient high x x T ambient low x x Supply voltage low x x Watchdog error x
Perform the Confirm message function (acknowledgement of the alarm via the control keys on the transmitter):
• The message/alarm is no longer shown on the display and the optical alarm goes out, where applicable. If multiple messages/alarms are active at the same time, all are reset simultaneously.
• If at least one message is assigned to the collective alarm, the collective alarm is reset. If the collective alarm is set on a relay,
Pos: 73 /TD/Pr odukt verwenden/MUF 63xx/ Stat us-War nmeldungen/Namur Fehlerbedingu ngen 635x @ 3\ mod_1236939722916_79. doc @ 29433 @ 3
the relay is also reset, meaning switched to its neutral position.
4.5.5. Namur fault conditions
If the faults named in the following table occur, the analog outputs output special values that enable a general fault warning in the higher-level control system. The values correspond to the "Namur"
Status message in the display
0 to
Watchdog error
Value below min. scale
Value above max. scale
industry standard.
Class Display
Analog output value in the display
20 mA
Error Previous
21 mA 3.8 mA 1.1 V 5.5 V 11 V value stops
Underrange Reading 0 mA 3.8 mA 0 V 0 V 0 V
Overrange Reading 20.5 mA 20.5 mA 1.1 V 5.5 V 11 V
4 to 20 mA
Additional message end9
1 V 5 V 10 V
67
4 Transmitter
Status message in the display
Pressure too
Class Display
Analog output value in the display
Overrange ooooo · 20.5 mA 20.5 mA 1.1 V 5.5 V 11 V
high
Pos: 74 /TD/Übers chrift en/MUF/1. 6 Wartung und Reinigung @ 3\mod_1234443039129_79.doc @ 24982 @ 2
4.6. Maintenance and cleaning
Pos: 75 /TD/Pr odukt instand halten/M UF 63xx/G erät warten/r einigen 635x @ 4\mod_1245756987352_79. doc @ 45305 @ 33
4.6.1. Maintaining the instrument
We recommend that the adjustment and settings of the transmitter be checked at regular intervals using the
• User menu (Operation, page 47) or
• P2A software (Parameterizing, adjusting and analyzing software (P2A software), page 69)
Convenient "remote monitoring" of the transmitter can be implemented, for example by using a relay as a collective alarm (see Editing Main Menu Alarm, page 53) whose messages are forwarded to a local alarm transmitter (horn, light) or PLC.
4.6.2. Cleaning the instrument
• Only clean the instrument carefully with a moist cloth.
• Do not use aggressive cleaning agents.
Pos: 76 /TD/Übers chrift en/MUF/3 Parametrier-, Abgleich und Analysesof tware ( P2A-Soft ware) @ 3\mod_1234258523713_79. doc @ 23932 @ 1
• Do not use any solvents.
68
5 Parameterizing, adjusting and analyzing software (P2A software)
5 Parameterizing, adjusting and analyzing
Pos: 77 /TD/Übers chrift en/MUF/1. 1/ 2.1/3.1 Leistungsbeschr eibung @ 3\mod_1234258595211_ 79. doc @ 23951 @ 2
5.1. Specifications
Pos: 78 /TD/Leist ungsbeschreibung/ Verwendung/M UF63xx/ MUF 63xx P2A @ 3\m od_1234258967326_79.doc @ 24065 @ 355
software (P2A software)
The P2A software is used for the parameterizing, adjustment and analysis of testo transmitters. The following applies:
• Generally, all newer testo transmitters (as of 2007) are supported.
• Included with every testo transmitter that is bought new is a CD that contains a free upgrade of the software, which includes the device drivers for all transmitters that can be attached at this time.
• This upgrade can be downloaded at any time via the testo homepage "www.testo.com/Download/P2A".
The software must only be bought one time, even for owners of several testo transmitters.
5.1.1. Functions and use
In the P2A software, two different file types are used: The instrument and the parameter file.
Instrument file
The parameters of a particular transmitter are stored in its so-called instrument file. Using this file, the parameters can be edited and the instrument can be tested and adjusted.
Instrument files also contain the respective histories in addition to the parameter data, i.e. "log books" are kept for the previous parameterizations, adjustments and messages (see Transmitter history page 93).
Instrument files are ".cfm" format files.
Parameter file
Parameter files are not tied to a specific individual transmitter and contain only parameter data/no history data.
If you use various instruments of the same type, you can create parameter files once (e.g. by saving the appropriate instrument file as the parameter file) and transmit these onto the other instruments.
69
5 Parameterizing, adjusting and analyzing software (P2A software)
Parameter files are ".cfp" format files.
Pos: 79 /TD/Leist ungsbeschreibung/ System vorausset zungen/ MUF 63xx @ 3\mod_1234260654399_79. doc @ 24084 @ 3555
5.1.2. System requirements
Operating system
• Windows® 2000 SP4
• Windows® XP Home/Professional
• Windows® Vista
Computer
• Pentium processor of at least 400 MHz or equivalent
• 128 MB RAM
• Graphics resolution of at least 1024 x 768
• Unused hard drive capacity of at least 15 MB
• CD-ROM drive
• USB interface
• At least Internet Explorer 5.0.
Software
The P2A software must be purchased and installed separately from the transmitter. If it is a new software version, the transmitter is already supported completely. Older P2A software versions can be updated via the P2A software upgrade (cf. product CD included
Pos: 80 /TD/Leist ungsbeschreibung/ Lieferum fang/ MUF 63xx/ MUF 63xx P2A @ 3\mod_1234260991646_79.doc @ 24103 @ 3
with the transmitter).
5.1.3. Scope of delivery
Included in the scope of delivery are:
• P2A software
70
• USB driver
When working with the parameterizing, adjusting and analyzing software (P2A software), previous knowledge of Windows in this instruction manual relates to Windows
®
operating systems is assumed. The description
®
XP.
5 Parameterizing, adjusting and analyzing software (P2A software)
Pos: 81 /TD/Übers chrift en/MUF/3. 2 Erste Schritte @ 3\mod_1234258633304_79.doc @ 23970 @ 2
5.2. First steps
Pos: 82 /TD/Erste Schritte/MUF 63xx/P2A/Software/ Treiber installieren @ 3\ mod_1234261192065_79. doc @ 24123 @ 3444
5.2.1. Installing the software/driver
Administrator rights are required to install programs and drivers under Windows
®
2000 SP4, XP and Vista.
5.2.1.1. Installing P2A software
1. Insert CD with P2A software.
If the installation program does not start automatically:
> Open Windows Explorer and start the file Setup.exe on the
product CD.
2. Follow the directions of the installation wizard.
5.2.1.2. Installing USB driver
Before installing the USB driver, please read the separate documentation that is enclosed with the USB driver CD.
5.2.1.3. P2A software upgrade
1. Insert product CD (supplied with the transmitter).
®
Explorer and start the file P2A upgrade.exe
Pos: 83 /TD/Erste Schritte/MUF 63xx/ P2A/ Sof t war e st arten @ 3\mod_1234261605199_79. doc @ 24142 @ 3444
2. Open Windows on the product CD.
3. Follow the directions of the installation wizard.
5.2.2. Starting the software
5.2.2.1. Starting the program
> Select: [Start] > All Programs > Testo > P2A Software.
- The program window is opened (see User interface page 72).
5.2.2.2. Establishing a connection with the instrument
Multiple instruments can be attached, however only one connection is active at all times.
USB driver is installed (see Installing USB driver page 71).
1. Start the P2A software.
2. Connect adapter (supplied with the P2A software) to the service interface of the instrument (see Service interface page 15).
71
5 Parameterizing, adjusting and analyzing software (P2A software)
3. Connect instrument/adapter to the PC via the USB interface.
- The instrument file of the attached instrument is shown in the file list.
5.2.2.3. Activating the connection with the instrument
> Click on the desired instrument file.
- The selected file is marked in colour and the connection with the instrument is activated.
If a connection with the instrument is established when the program is started, the corresponding instrument file is marked
Pos: 84 /TD/Übers chrift en/MUF/3. 3 Software verwenden @ 3\mod_1234258679599_ 79.doc @ 23989 @ 2
5.3. Using the softwar e
Pos: 85 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Soft ware verwenden @ 3\ mod_1234262654547_79. doc @ 24162 @ 3
5.3.1. User interface
automatically.
72
1 Menu bar:
Menu Command Explanation
File Open Shows the Windows dialogue for
searching and opening files.
5 Parameterizing, adjusting and analyzing software (P2A software)
Menu Command Explanation
Save as Saves the parameters of an
instrument or parameter file under a new name.
Edit Copy Copies the parameters of the
marked instrument or parameter file in the cache.
Paste Pastes the parameters from the
cache in the marked instrument or parameter file.
View Toolbar
Status bar
? Check instrument
connections
Activates/deactivates the toolbar or status bar.
Checks the connections to a connected instrument without the instrument having to be activated.
Service A text file with the most important
information on the computer and the software is opened via
Display service data.
Information Shows the version number of the
P2A software.
2 Toolbar: Shows the Windows-compliant icons for editing. 3 File:
Icon File Explanation
Symbol shows a transmitter
Instrument file
Instrument file Connection to the instrument has
been established.
<Type> <Serial number>.cfm
File name should not be changed.
Symbol shows a transmitter with a red minus sign in the upper left corner
Instrument file
Instrument file Connection to the instrument has
not been established.
73
5 Parameterizing, adjusting and analyzing software (P2A software)
Icon File Explanation
Symbol shows a transmitter with a white P for parameter file in the upper left corner
Parameter file
<Type> <Serial number> <Date> <Time>.cfp
File name can be changed. The name can be selected freely,
but it is recommended that you retain the reference to the instrument.
Parameter files are always marked red; the parameter values they contain are only forwarded to the instrument after being transmitted to the instrument file.
4 Function buttons: Dialogues on editing and testing the
instrument are opened by means of the buttons.
[Change parameterization] see Changing instrument/parameter
file page 75
[Test/analyze transmitter] see Chapter Analyzing/testing the
transmitter page 86
[Adjusting the transmitter] see Chapter Adjusting the transmitter
page 91
[Transmitter history] see Transmitter history page 93
5 File information:
Status Shown in the window
An instrument file is selected
A parameter file is selected
Type, serial number, firmware version of the instrument.
Type, serial number and firmware version of instrument for which the parameter file was created.
Connection status Green = connection is active
Red = connection is inactive
6 Status bar: Shows the current status when editing via the menu
bar.
74
5 Parameterizing, adjusting and analyzing software (P2A software)
Pos: 86 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Ger äte-/Par am et er datei bearbeiten @ 3\mod_1234358080444_7 9. doc @ 24303 @ 34
5.3.2. Editing instrument/parameter file
5.3.2.1. Changing instrument/parameter file
The desired instrument/parameter file is marked.
1. Click on [Change parameterization].
- The Properties of <Instrument type> <Serial number> dialogue is opened with the Change parameterization register.
If the parameters were transmitted from other parameter files into the instrument file, a message is shown with which you can transmit the new parameters to the connected instrument using
[Yes].
> If the parameters should not be transmitted, click on [No].
Pos: 87 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Einheit / Analogausgan g @ 3\m od_1234366217517_79.doc @ 24592 @
2. Change or enter parameters in the corresponding fields.
Field Explanation
Unit/ Analog output
All analog outputs are parameterized in this mask.
75
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
Unit/analog output (graphic)
Unit: 0 to 1 V/5 V/10 V or 0 to 20 mA/4 to 20 mA.
Vertical: Current version of the analog output (cannot be changed).
Horizontal: Min./max. scale end points of selected unit.
The curve changes in accordance with the entered value of scale minimum and
maximu m. Scale minimum/ maximu m
The endpoints of the scaling can be selected
up to the stored scale minimum and
maximum. In the process, scaling can take
place beyond the measuring range in order
to adjust the analog output to the customer
system, see Scaling, page 16 Unit Selection of the physical unit.
When changing the unit, standard values are
set for scale minimum and maximum.
Caution!
When changing the phys. unit, the relay limit
values are set to the assigned default values. Signal delay
(graphic)
Curve changes according to the set signal
delay. Signal delay Time interval in stages 1 – 15:
1 = no delay
15 = longest delay.
The signal delay is added to the reaction time
of the sensor. The signal delay shows
averaging over the time interval of the
selected stage in seconds:
Example
Stage 10 = average of the readings from the
last 10 seconds.
Pos: 88 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Gr enzwert e Relais @ 3\mod_1234366741293_79.doc @ 24611 @
76
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
Limit values, relay 1 to 4/alarm values, alarm 1 to 4
In this mask, the relays or display alarms are parameterized.
Relay x/alarm x Four relays or alarm values are
available (optional).
Not used Relay is not used.
Hysteresis image and input options are hidden.
Collective alarm A relay can be used as a collective
alarm detector in the event that selected messages appear. Selection of the messages (OR linkage) by selecting the checkbox.
Min control If switched to ON (NO contact) or
OFF (NC contact) under the limit value; in the event of a subsequent exceeding of Limit
value plus Hysteresis, it is
switched to OFF (NO contact) or ON (NC contact).
77
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
Max control If switched to ON (NO contact) or
OFF (NC contact) above the limit value; in the event of a subsequent undershooting of
Limit value minus Hysteresis, it
is switched to OFF (NO contact) or ON (NC contact).
78
The graphic display in the centre of the screen refers to the relay wiring as a NO contact (ON).
Hysteresis To avoid switching cycles. Channel Selection of the channel that is to
be monitored.
Limit value Values for the limits of the unit
selected in Unit/analog output; 4 decimal places.
When changing the phys. unit, the relay limit values are set to the default values.
Visual alarm display Checkbox is activated: Upon the
appearance of the selected alarm, the background lighting of the display flashes.
Field Explanation
Alarm delay The desired alarm delay for the
Pos: 89 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Prozessdat en Druck+Norm daten 635x/ 6381 @ 3\mod_1235994444633_79. doc @ 26986 @
Field Explanation
Basic settings Setting the pressure process data for Pitot
5 Parameterizing, adjusting and analyzing software (P2A software)
alarms of the min/max control and the visual alarm is entered in the input field (0 to 240 seconds possible).
The alarm delay has no effect on the collective alarms.
tube measurement and standard data for volumetric flow rate measurement.
Absolute pressure Absolute pressure existing in the process.
The entered absolute pressure value is included in the Pitot tube calculation.
Temperature Temperature existing in the process. The
entered temperature value is included in the Pitot tube calculation.
Humidity Humidity existing in the process. The entered
humidity value is included in the Pitot tube calculation.
79
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
Pitot tube factor The Pitot tube correction factor depends on
the shape of the Pitot tube (see operating
instructions of the Pitot tube).
The entered correction factor has a direct
proportional effect on the volumetric flow
rate. Cross-section of the
duct Correction factor The correction factor enables an adjustment
The entered value is included in the Pitot
tube calculation.
to the flow profile in the duct.
The entered correction factor is included in
the Pitot tube calculation.
80
Absolute pressure The entered value and the selected unit are
included in the calculation of the standard
volumetric flow rate. Temperature The entered value and the selected unit are
included in the calculation of the standard
volumetric flow rate.
[Reset] Button for resetting the standard data to the
factory settings.
Pos: 90 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Ether net @ 3\mod_1234432069559_79. doc @ 24732 @
5 Parameterizing, adjusting and analyzing software (P2A software)
Button Explanation
Ethernet Networking the transmitters via Ethernet.
For a multitude of applications, measurement data can be simultaneously recorded, documented and visualized.
Start wizard... Address allocation of the testo 6381 with
Ethernet module:
IP address Transmitter IP address
81
5 Parameterizing, adjusting and analyzing software (P2A software)
Button Explanation
Prior to the automatic allocation of the IP address, the network cable must be connected to the transmitter (see Mains connection, page
33). If the instrument is used as a Saveris
subscriber:
• Saveris base must be functional.
• Saveris base must be connected to the
IP address of the device
network.
Address allocation of the testo 6381 with
Ethernet module:
Manual • Define the IP address of the transmitter
• Enter netmask
• Enter gateway
Automatic Automatic allocation of the IP address
82
Only when using the transmitter Ethernet modules in Saveris mode
Manual • Take the IP address from the InfoBase
menu window of the Saveris base
• Enter IP address
# Pos: 91 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Display @ 3\mod_1234424330 866_79. doc @ 24673 @
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
Display Setting the display functions (if a display is
available on the transmitter).
Continuous display
Display lighting is permanently switched on.
lighting Display lighting
when button is
When a particular button on the instrument is
pressed, the display lights up for 10 seconds. pressed (10 seconds)
Display brightness (0 to 9)
Setting the brightness between 0 and 9:
0 = dark
9 = light Display contrast
(0 to 9)
Setting the contrast between 0 and 9:
0 = low contrast
9 = high contrast Display of the error
messages
Selection of whether the error messages
should be shown in the display. Display language Selection of the language. New password The password consists of four numbers,
each of which must be between 1 and 9.
If the password protection is not to be used,
the numerical code "0000" must be entered.
83
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
[Adopt new
Button for confirming the new password.
password] Valid password Display of the current password.
Pos: 92 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Nullungszyklus @ 3\mod_123599595293 6_79. doc @ 27005 @
Field Explanation
Zeroing cycle Setting the interval of the automatic zeroing
with solenoid valve.
The accuracy specifications are only valid for the zeroing cycle of 15 sec set by the factory.
Zeroing cycle setting Selection of the time interval after which the
automatic zeroing cycle is to occur. Recommendation: 15 sec
Pos: 93 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Param eter speichern @ 3\mod_1234358425697_79. doc @ 24322 @ 4
5.3.2.2. Saving parameters
Parameters can be saved in new parameter files.
1. Mark instrument/parameter file.
2. Click on File > Save as in the menu bar.
3. Select storage location and enter the file name.
4. Click on [Save].
- The new parameter file is shown in the file list.
84
5 Parameterizing, adjusting and analyzing software (P2A software)
Only the parameters are saved from an instrument file; the history data are not adopted.
The original name (Instrument type, Serial number) is suggested with the current date/time as standard, e.g. "testo 6381 01234578 061120 1403.cfp".
For a standard installation, the files are saved under "C:\Documents and Settings\All Users\Shared Documents\P2A Software". The path can differ depending
Pos: 94 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Param eter datei öffnen @ 3\mod_1234358634164_79. doc @ 24341 @ 4
on the version of the operating system.
5.3.2.3. Opening the parameter file
All parameter files stored in the standard directory path are automatically displayed in the file list when the software is started.
You can also open parameter files that are stored in other directories.
1. Click on File > Open in the menu bar.
2. Select the storage location and click on the requisite file.
3. Click on [Open].
- The selected file is opened. This can be changed and saved
Pos: 95 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Param eter kopieren/einfügen @ 3\m od_1234358826054_79.doc @ 24360 @ 4
(see Editing instrument/parameter file page 75).
5.3.2.4. Copying and pasting parameters
The parameters of a parameter file can be transmitted to an instrument file or another parameter file from the same instrument type.
1. Select file from which parameters are to be copied.
2. Click on Edit > Copy in the menu bar.
3. Select the file which is to be modified.
4. Click on Edit > Paste in the menu bar.
- The parameters are transmitted to the file.
You can also use the common keyboard shortcuts for copying (CTRL+C) and pasting (CTRL+V).
Parameters can also be transmitted using drag & drop, where you drag the icon of the parameter file onto the icon of the target instrument file.
5. Connect and select corresponding instrument.
6. Click on [Change parameterization].
7. Confirm confirmation request.
85
5 Parameterizing, adjusting and analyzing software (P2A software)
Pos: 96 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Ger äte-/Par am et er datei löschen @ 3\m od_1234359025194_79.doc @ 24379 @ 4
- Parameter data are transferred to the instrument.
5.3.2.5. Deleting instrument/parameter file
Instrument/parameter files can be deleted from the file list.
1. Click on the file that is to be deleted with the right mouse button.
2. Select the command Delete in the context menu.
Pos: 97 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Ger ätedat ei erzeugen @ 3\mod_1237206865717_79. doc @ 29703 @ 4
- The instrument or parameter file is deleted from the list.
5.3.2.6. Creating a new instrument file
It is possible to create an instrument file without restarting the P2A software.
Transmitter must be connected.
1. Click on File > New connection in the menu bar.
Pos: 98 /TD/Übers chrift en/MUF/x.x.x Messumformer analysieren / testen @ 3\mod_1237380874787_79. doc @ 29928 @ 3
5.3.3. Analyzing/testing the transmitter
Pos: 99 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Mess umf ormer analysieren/t est en 635x @ 3\mod_1234359444253_79.doc @ 24398 @ 4
- Connection to the transmitter is established.
In this section, you can test the outputs of the connected instrument, read off the limit values and reset the parameters to the factory settings.
The function is only available for instrument files.
5.3.3.1. Analyzing/testing the instrument
86
The required instrument file is marked.
1. Click on [Test/analyze transmitter].
- The Properties of <Instrument type> <Serial number> dialogue is opened with the Test/analyze transmitter register.
2. Perform action:
Action Explanation
Carrying out factory reset
Testing analog output
Reset the unit, limit value and hysteresis parameters to factory settings (see below).
Test channel 1 (see Testing functionality of analog outputs, page 58).
Test switch outputs Manually switch relays 1 to 4 to test for
proper function (see Testing functionality of relay outputs, page 59).
5 Parameterizing, adjusting and analyzing software (P2A software)
Action Explanation
Display min./max. values
Overview of the minimum and maximum values measured since the last reset of the transmitter (see Displaying min./max. values, page 90).
Pos: 100 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Wer ksreset durchführen @ 3\mod_1237379864072_79.doc @ 29887 @ 4
3. Click on [OK] or [Cancel] to close the dialogue.
5.3.3.2. Carrying out factory reset
The required instrument file is marked.
1. Click on [Test/analyze transmitter].
- The Properties of <Instrument type> <Serial number> dialogue is opened with the Test/analyze transmitter register.
2. Mark transmitter test.
- Current operating hours are shown.
3. Confirm control query to perform the reset.
- The values are reset to the customer-specific factory settings.
Pos: 101 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Analogausgang t est en 635x @ 3\mod_1237382683201_79.doc @ 29948 @ 4
4. Click on [OK] or [Cancel] to close the dialogue.
5.3.3.3. Testing analog output
The required instrument file is marked.
1. Click on [Test/analyze transmitter].
- The Properties of <Instrument type> <Serial number> dialogue is opened with the Test/analyze transmitter register.
2. Mark channel and test values.
Field/button Explanation
Transmitter test Check the analog outputs (see Testing
functionality of analog outputs, page
58).
87
5 Parameterizing, adjusting and analyzing software (P2A software)
Field/button Explanation
Current reading Readings are updated every second. Unit Unit according to the type of analog
output.
Default value Freely definable output value for the
respective type of analog output (V or mA), 1 decimal place.
[Activate] The entered default value is forwarded
to the corresponding analog output and to the test contacts by clicking.
A warning informs that the value is being transmitted to the connected instrument in the event of existing cabling.
Now check the analog output using a precise multimeter.
[Deactivate] Finish entering the electrical variables
at the analog output. The analog output returns to the
current reading again.
3. Click on [OK] or [Cancel] to close the dialogue.
- The analog output and the relay return to Measuring Mode again.
88
5 Parameterizing, adjusting and analyzing software (P2A software)
Pos: 102 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Schaltausgang Re lais testen 635x @ 3\m od_1237374457598_79.doc @ 29866 @ 4
5.3.3.4. Testing switch output relays 1 to 4
The required instrument file is marked.
1. Click on [Test/analyze transmitter].
- The Properties of <Instrument type> <Serial number> dialogue is opened with the Test/analyze transmitter register.
2. Mark Relay tests and test the values.
Field/button Explanation
Test the relay function (see Testing
functionality of relay outputs, page 59).
[Activate relay n] Close contact.
A warning informs that the value is being transmitted to a connected PLC, external display, etc. in the event of existing cabling.
[Deactivate relay n]
Open contact. A warning informs that the value is being
transmitted to a connected PLC, external display, etc. in the event of existing cabling.
3. Click on [OK] or [Cancel] to close the dialogue.
- The analog output returns to Measuring Mode again.
89
5 Parameterizing, adjusting and analyzing software (P2A software)
Pos: 103 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Min-/ Max -Werte anzeigen 635x @ 3\mod_1242135985269_79. doc @ 32803 @ 4
5.3.3.5. Displaying min./max. values
The transmitter saves the minimum or maximum value for each channel (measured since the last voltage supply or since the last manual reset).
The required instrument file is marked.
1. Click on [Test/analyze transmitter].
- The Properties of <Instrument type> <Serial number> dialogue is opened with the Test/analyze transmitter register.
2. Mark Min./max. values.
Field/button Explanation
Min./max. values View the min./max. values of each channel.
Only the values within the measuring range are shown.
90
Channel Channel 1 min./max. Value Min. or max. value, 1 decimal place. Unit Unit selected in Unit/analog output.
3. Reset Min./max. values.
4. Click on [R eset mi n . /max. va l ues] .
5. Confirm control query to perform the reset.
- The values are reset to the factory settings.
5 Parameterizing, adjusting and analyzing software (P2A software)
Pos: 104 /TD/Übers chrift en/MUF/x.x.x Messumformer abgleichen @ 4\mod_1244530152797_79. doc @ 44885 @ 3
5.3.4. Adjusting the transmitter
Pos: 105 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Mess umf ormer abgleichen 635x @ 3\m od_1235996515230_79.doc @ 27024 @
6. Click on [OK] or [Cancel] to close the dialogue.
This function is used to adjust an attached instrument. The following adjustments may be carried out using the software:
• Analog adjustment (entry via assistant/wizard)
• n-point adjustment (entry via assistant/wizard)
Pos: 106 /TD/Pr odukt verwenden/MUF 63xx/ P2A/n-Punkt -Abgleich _635x @ 3\m od_123745734571 4_79.doc @ 30083 @ 4
Also see Editing Main Menu Adjust, page 61
5.3.4.1. n-point adjustment
1. Connect precision pressure sensor (see n-point adjustment, page 46).
2. Mark the instrument file of the connected instrument.
3. Click on [Adjusting the transmitter].
- The Properties of <Instrument type> <Serial number> dialogue is opened with the Adjusting the transmitter register.
4. Mark n-point adjustment.
5. Click on [Start wizard …] and follow the instructions of the wizard.
- The adjustment is performed when the wizard is closed.
Field Explanation
91
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
How much pressure is actually applied
Required field: Entry of the value read off at the pressure sensor.
The n-point adjustment must always be carried out to its full extent and in good time at all selected adjustment points.
The number of adjustment points (3 to 6) is stored in the user menu of the transmitter and can only be changed via
Pos: 107 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Analogausgang abgle ichen _635x @ 3\ mod_1234363369967 _79. doc @ 24531 @ 4
the P2A software.
5.3.4.2. Adjusting the analog output
1. Connect precision multimeter (see Analog output adjustment, page 45).
2. Mark the instrument file of the connected instrument.
3. Click on [Adjusting the transmitter].
- The Properties of <Instrument type> <Serial number> dialogue is opened with the Adjusting the transmitter register.
4. Click on [Start wizard …] and follow the instructions of the wizard.
- The adjustment is performed when the wizard is closed.
Field Explanation
92
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
Default value The analog output value from the last
performed adjustment is given at the output. Value of the factory adjustment:
• Lower adjustment point: approx. 10 % of the max. value
• Centre adjustment point: approx. 50 % of the max. value
• Upper adjustment point: approx. 90 % of the max. value
Measured analog value
Pos: 108 /TD/Übers chrift en/MUF/2. 3. 5 Messumformer-Hist orie @ 3\ mod_1237373806926_79. doc @ 29846 @ 3
5.3.5. Transmitter history
Pos: 109 /TD/Pr odukt verwenden/MUF 63xx/ P2A/Mess umf ormer-Histor ie 635x @ 3\mod_1236082095561_79.doc @ 27272 @
Required field: Entry of the value read off at the multimeter.
Parameterizations, adjustment processes and messages that have occurred are registered in the transmitter with an operating hours stamp.
In the history overviews (explained later in more detail), past processes and events can be made visible.
For parameter changes or adjustments that are performed directly at the instrument (via the user menu), "Transmitter" is entered in the User field and only the operating hour is entered in the Date/time field instead of operating hour/date/time.
For entries that are performed using the P2A software, the name of the user logged into Windows appears in the User field, while the date/time and operating hour are shown in the Date/time field.
1. Mark the instrument file of the connected instrument.
2. Click on the [Transmitter history] button.
- The Properties of <Instrument type> <Serial number> dialogue is opened with the Transmitter history register.
3. Click on the required entry in the list to change the display.
93
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
Operating hours / date/time
Operating hour/time stamp at which the change at the instrument was performed.
User Name with which the user is logged into the
operating system. "Transmitter" entry if the change was
performed at the instrument.
Comments Type of parameter change, e.g. "Unit of
channel 1 from Pa to bar".
94
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
Selection: Analog adjustments.
Operating hours / date/time
Operating hour/time stamp at which the change at the instrument was performed.
User Name with which the user is logged into the
operating system. "Transmitter" entry if the change was
performed at the instrument.
Unit Unit during the adjustment.
Channel Analog adjustment: Channel 1 to n.
Specification Analog adjustment: Actual value.
Pressure specification
n-point adjustment: Reference value set at the pressure sensor.
Multimeter Analog adjustment: Reference value set at
the multimeter.
Offset Analog adjustment: Deviation at time of
adjustment.
95
5 Parameterizing, adjusting and analyzing software (P2A software)
Field Explanation
The table is shown only for error and status messages that were generated in the transmitter and were transferred and saved there via the connection to the P2A software.
Operating hour Operating hour at which the message
appeared in the instrument.
Message e.g. "Watchdog error". Due to a processor
error, the transmitter performs an automatic restart.
Type of message e.g. early warning, status message.
> To print out the history data, click on [Print].
The printing job is automatically sent to the default printer for the operating system.
Pos: 110 /TD/Übers chrift en/8. Tipps und Hilf e @ 0 \m od_1173789887985_79. doc @ 406 @ 1
4. Click on [OK] or [Cancel] to close the dialogue.
The printout can be edited using [Set up printer …].
96
6 Tips and assistance
Pos: 111 /TD/Übers chrift en/8.1 Fragen und Ant worten @ 0\mod_1177402017078_79.doc @ 1093 @ 2
6.1. Questions and answers
Pos: 112 /TD/Tipps und Hilf e/F r agen und Antworten/ MUF 635x @ 3\mod_1236607627223_79. doc @ 27543 @
Question Possible causes/solution
Connection to instrument
Check connection cable/plug contacts
cannot be established A message appears on the
display Malfunction (with and
without display)
See Status, warning and error messages, page 63
Analysis using the P2A software, see Analyzing/testing the transmitter, page 86
When does a stable
After approx. 20 seconds
current reading appear?
If we could not answer your question, please contact your dealer or Testo Customer Service. For contact details see the rear side of
Pos: 113 /TD/Übers chrift en/8.2 Zubehör und Er satzteile @ 0\mod_1177402058734_79. doc @ 1102 @ 2
6.2. Accessories and spare parts
Pos: 114 /TD/Tipps und Hilf e/Z ubehör und Ersat zteile/M UF63xx/Zubehör Er satzt eile MUF 635x @ 3\m od_1235640016819_79.doc @ 26783 @
this document or the web page www.testo.com/service-contact
Description Article no. Ethernet
Ethernet module 0554 6656 Ethernet connector 0554 6653 Interface and software P2A software (parameterizing, adjusting,
analyzing) incl. USB adapter Plug-in connections Set of M12 plug-in connections (connector
and socket) for power and signal lines Hoses Silicone hose ID 4 transparent 0086 0001, sold by
TYGON hose ID 4.8 transparent 0086 0031, sold by
6 Tips and assistance
0554 6020
0554 6682
the metre
the metre
97
6 Tips and assistance
Description Article no. Pitot tubes
Pitot tube, length 350 mm, stainless steel 0635 2145 Pitot tube, length 1000 mm, stainless steel 0635 2345 Supply Mains unit (desktop, wall-mounted) 0554 1748 Mains unit (top-hat rail mounting) 0554 1749 External display
Process display for testo 54-2 AC.... 5400 7553
Process display for testo 54-7 AC.... 5400 7555
Calibration Standard ISO calibration certificate,
transmitter only Standard DKD calibration certificate,
transmitter only
For a complete list of all accessories and spare parts, please refer to the product catalogues and brochures or look up our website at:
Pos: 115 /TD/Tipps und Hilf e/Z ubehör und Ersat zteile/M UF63xx/Best elloptionen M UF 6380 @ 3\mod_1234277777760_79. doc @ 24261 @ 3
www.testo.com
0520 1000
0520 1200
6.2.1. Ordering options for 6381 transmitter (0555 6381)
Order code Characteristic Axx Measuring range
A01 0 to 50 Pa A02 0 to 50 Pa A03 0 to 100 Pa A04 0 to 500 Pa A05 0 to 10 hPa A07 0 to 50 hPa A08 0 to 100 hPa A09 0 to 500 hPa A10 0 to 1000 hPa A21 -10 to 10 Pa A22 -50 to 50 Pa
98
6 Tips and assistance
Order code Characteristic
A23 -100 to 100 Pa A24 -500 to 500 Pa A25 -10 to 10 hPa A27 -50 to 50 hPa A28 -100 to 100 hPa A29 -500 to 500 hPa A30 -1000 to 1000 hPa
Bxx Analog output/supply
B02 0 to 1 V (4-wire, 24 V AC/DC) B03 0 to 5 V (4-wire, 24 V AC/DC) B04 0 to 10 V (4-wire, 24 V AC/DC) B05 0 to 20 mA (4-wire, 24 V AC/DC) B06 4 to 20 mA (4-wire, 24 V AC/DC) Cxx Display C00 Without display C02 With display/English C03 With display/German C04 With display/French C05 With display/Spanish C06 With display/Italian C07 With display/Japanese C08 With display/Swedish Dxx Cable entry D01 M16 cable entry (relay: M20) D02 Cable entry NPT 1/2" D03 Cable contacting via M plug-in
connection for signal and supply
Exx Ethernet E00 Without Ethernet module
99
6 Tips and assistance
Order code Characteristic
E01 With Ethernet module
Fxx Differential pressure unit
F01 Pa/Min/Max F02 hPa/Min/Max F03 kPa/Min/Max F04 mbar/Min/Max F05 bar/Min/Max F06 mmH2O /Min/Max F07 inchH2O /Min/Max F08 inch HG/Min/Max F09 kg/cm2 /Min/Max F10 PSI/Min/Max F11 m/s /Min/Max F12 ft/min /Min/Max F13 m3/h /Min/Max F14 l/min /Min/Max F15 Nm3/min /Min/Max F16 Nl/min /Min/Max
Gxx Optional analog output for humidity probe connection testo 6610/units
G00 Without connection option for testo
G01 % RH/Min/Max G02 °C/Min/Max G03 °F/Min/Max G04 °Ctd/Min/Max G05 °Ftd/Min/Max G06 g/kg /Min/Max
10
6610 humidity probe
10
Scaling 10 to 100 % of measuring range is possible, but at least 10 Pa
100
Loading...