8.4. Accessories and spare parts ......................................................... 23
9 EC declaration of conformity ................................................................ 24
3
Page 4
2 Safety and the environment
2 Safety and the environment
2.1. About this document
Use
> Please read this documentation carefully and familiarize
yourself with the product before putting it to use. Pay particular
attention to the safety instructions and warning advice in order
to prevent injuries and damage to the product.
Symbols and writing standards
Representation Explanation
Warning advice, risk level according to the
signal word:
Warning! Serious physical injury may occur.
Caution! Slight physical injury or damage to
the equipment may occur.
> Implement the specified precautionary
Menu
Elements of the instrument, the instrument
display or the program interface.
[OK]
Control keys of the instrument or buttons of
the program interface.
2.2. Ensure safety
> Do not operate the instrument if there are signs of damage at
the housing, mains unit or hoses.
> Do not store the product together with solvents. Do not use any
desiccants.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation.
> Dangers may also arise from the refrigeration systems being
measured or the measuring environment: Note the safety
regulations valid in your area when performing the
measurements.
> If the manifold falls or another comparable mechanical load
occurs, the pipe sections of the refrigerant hoses may break.
The valve stem shutoff may also be damaged, whereby further
measures.
4
Page 5
damage to the interior of the manifold may occur that cannot be
identified from the outside. The refrigerant hoses must therefore
be replaced with new, undamaged refrigerant hoses every time
the manifold falls or following any other comparable mechanical
load. Send the manifold to Testo Customer Service for a
technical check for your own safety.
> To prevent damage from ESD (electro-static discharge) or
transient voltage spikes make sure that your refrigeration
system is properly grounded, as otherwise the manifold might
2.3. Protecting the environment
get damaged.
> Dispose of spent batteries in accordance with the valid legal
specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
regulations) or return the product to Testo for disposal.
> Refrigerant gases can harm the environment. Please note the
3 Product description
applicable environmental regulations.
3 Product description
3.1. Use
Testo 557 is a digital manifold for maintenance and service work on
refrigeration systems and heat pumps. It is intended for use by
qualified personnel only.
The function of the testo 557 is designed to replace analog
manifolds, thermometers and pressure/temperature charts.
Pressures and temperatures can be applied, adapted, tested and
monitored.
Testo 557 is compatible with most non-corrosive refrigerants, water
and glycol. Testo 557 is not compatible with refrigerants containing
ammonia.
The instruments must not be used in explosive environments!
5
Page 6
3 Product description
1
2
2 3 4
5
6 7 8
8
8
8
9
9
10
10
11
12
13
3.2.Overview
Display and control elements
1 Front connection for external vacuum probe
2 Mini-DIN probe socket for NTC temperature probe, with socket
cover
3 Foldable hanging hook (on rear)
4 Display. Instrument status icons:
Icon Meaning
//
Battery status
Bluetooth®
Select measuring mode
6
Page 7
3 Product description
5 Battery compartment. The rechargeable batteries cannot be
charged inside the instrument!
6 Control keys:
Ke y Function
[Set]
[R, ►, ■]
[Mode]
[Min/Max/Mean]
[▲]
[P=0]
Esc
[▼]
[/]
Set units
Select refrigerant / Start-Stop leak test
Switch between measuring modes
Display min, max, mean values
Up-key: Scroll through menu.
Pressure zeroing
Switches to the measurement/home view.
Down-key: Scroll through menu.
Switching the instrument on/off
Switching display ilumination on/off.
7 Sight glass for refrigerant flow.
8 4 x valve stem shutoff
9 4 x hose holders for refrigerant hoses
10 2 x Connection 1/4" SAE, brass.
High pressure, for refrigerant hoses with quick release screw
fitting, passage for valve actuator lockable.
11 Connection 3/8" SAE, brass, for vacuum pump
12 Connection 1/4" SAE, brass, for e.g. refrigerant cylinders
13 Mini-USB connection for firmware update, inside the battery
compartment.
7
Page 8
4 First steps
4First steps
Inserting batteries/rechargeable batteries
1. Fold out the hanging hook and open the battery compartment
by squeezing the clip lock.
2. Insert batteries (included in delivery) or rechargeable batteries
(4x 1.5 V, type AA/Mignon/LR6) in the battery compartment.
Observe the polarity!
3. Close the battery compartment.
- After inserting the batteries, the instrument switches on
automatically and goes into the settings menu.
When not in use for long period: Remove batteries /
rechargeable batteries.
Units / Parameter selection
1. Press [Set] to confirm or change unit parameter settings
2. Press [▲]
or [▼] to change the units / parameters.
- The settings will be accepted once the last selection has been
made.
Key Functions Description
[▲] or [▼]
[Set]
Selectable
parameters
°C, °F
bar, kPa, MPa, psi
Pabs, Prel or psig
micron, inHg, Pa,
hPa, mTorr,Torr,
inH2O, mbar
/ /
Change parameters and select units
Confirm units / parameters
Description
Temperature unit
Unit for pressure.
Switch between absolute and relative
pressure display
Unit for vacuum pressure.
Select measuring mode: heat pump / cooling
/ or Auto
8
Page 9
4 First steps
WARNING
Over tightening the valve stem shutoffs may cause:
Selectable
Description
parameters
AUTO OFF
Activate or deactivate Automatic power-off.
Instrument switches off after 30 minutes if no
temperature probe is connected and there is
no pressure apart from ambient pressure
T
fac
Activate or deactivate surface temperature
compensation factor, icon is shown on the
display if the function is deactivated
- Settings will be applied following the final selection.
Operating the valve stem shutoffs
The digital manifold acts like a conventional two-way manifold with
regard to the refrigerant path: The passages are opened by
opening the valves. The adjacent pressure is measured with valves
closed as well as with them open.
> Open valve: Turn valve positioner counterclockwise.
> Close valve: Turn valve positioner clockwise.
• Damage to the PTFE seal.
• Mechanical deformation of the valve piston leading to the
PTFE seal falling out.
• Damage to the thread of the threaded spindle and the valve
screw.
• valve knobs to brake.
Lightly tighten the valve knob. Do not use tools to tighten the valve
stem shutoffs.
9
Page 10
5 Using the manifold
WARNING
5 Using the manifold
5.1. Preparing for measurement
5.1.1. Switching the instrument on
> Press [/].
Zero the pressure sensors before every measurement.
✓All connections must be at ambient pressure.
> Press [P=0] key for 3 seconds to execute zeroing.
Connecting the refrigerant hoses
Before each measurement check whether the refrigerant
hoses are in flawless condition.
✓ Make sure the valve stem shutoffs are closed.
1. Connect the refrigerant hoses to the low-pressure side (blue)
and high-pressure side (red).
2. Connect the refrigerant hoses to the AC/R system.
Dropping this instrument or any other comparable mechanical
shock can damage the refrigerant pipes and hoses. The valve
actuators may also suffer damage, which in turn could result in
further damage inside the instrument and may not be detectable
from outside.
> For your own safety you should return the manifold to the
Testo Service Department for technical inspection.
> You should therefore always replace the refrigerant hoses with
new ones after an accidental drop has occurred or after any
visible wear and tear.
10
Page 11
5 Using the manifold
Choosing the refrigerant
1. Press [R, ►, ■].
- It activates the refrigerant menu and the currently selected
refrigerant flashes.
2. Setting the refrigerant:
Key functions Description
[▲] or [▼]
[R, ►, ■]
Selectng another refrigerant
Confirm the selection and exit the refrigerant
menu
Available refrigerants
Representation Description
R...
Refrigerant number of refrigerant acc. to
ISO 817
---
no refrigerant selected.
Example: Setting refrigerant R401B
1. Press [R, ►, ■]to activate refrigerant menu.
2. Press [▲] or [▼] several times, until R401B flashes.
3. Press [R, ►, ■]to confirm the setting.
Exiting the refrigerant selection
> Press [R, ►, ■] or automatically after 30 s, if no other key has
been pressed.
5.1.2.Connecting the temperature sensor
Surface temperature sensor
At least one NTC temperature probe must be connected to
measure the pipe temperature, for automatic calculation of
superheating and subcooling.
Deactivating the surface compensation factor for insertion and
air temperature sensor
A surface compensation factor has been set in the measuring
instrument to improve the measuring accuracy of surface
temperature readings.
If the manifold is used in combination with insertion or air
temperature probe (optional), this factor must be deactivated:
to Off.
fac
fac
is displayed.
1. Press [Set] repeatedly until T
2. Press [▲] or [▼] to set T
11
Page 12
5 Using the manifold
3. Press [Set] to continue through the settings menu until the
measurement/home view is displayed.
is shown on the display if T
- T
fac
is disabled.
fac
5.1.3.Connecting the vacuum probe
> Open the front cover of the connector and connect up the
5.1.4. Switching Bluetooth® on and off
vacuum probe.
In order to to establish a connection via Bluetooth, on an Android
or iOS device, the Testo App Refrigeration must be already
installed.
You can get the App for iOS instruments in the App Store or for
Android instruments in the Play Store.
Information about compatibility can be found in the relevant app
store.
1. To turn on the Bluetooth press [▲] and [▼] simultaneously and
hold down for 3 seconds.
- Once the Bluetooth icon is shown on the display, Bluetooth is
switched on.
Display Description
flashes
is permanently
There is no Bluetooth connection, or a
potential connection is being searched for.
There is a Bluetooth connection.
displayed
is not displayed
Bluetooth is disabled.
12
2. To turn off the Bluetooth press [▲] and [▼] simultaneously and
hold down for 3 seconds.
- Once the Bluetooth icon is no longer shown on the display,
Bluetooth is switched off.
Page 13
5.1.5. Measuring mode
pressure. This automatic change occurs
WARNING
Display Mode Function
Refrigeration
system
Heat pump Normal function of the digital manifold
Automatic
mode
5 Using the manifold
Normal function of the digital manifold
If the automatic mode is activated, the
testo 557 digital manifold automatically
changes the display of the high and low
when the pressure on the low-pressure
side is 1 bar (15 psi) higher than the
pressure on the high-pressure side.
During the change, Load (2 s) is shown
in the display.
This mode is especially suited to air
conditioning systems which cool and
heat (heat pumps).
5.2.Performing the measurement
Risk of injury caused by refrigerant that is at high pressure,
hot, cold, or poisonous!
> Wear safety goggles and protective gloves..
> Before pressurizing the measuring instrument: Always fasten
the measuring instrument at the hanging hook in order to
prevent it from falling (risk of breakage)
> Check if the refrigerant hoses are intact and connected
correctly before each measurement. Do not use a tool to
connect the hoses. Only tighten the hoses by hand (max.
torque 5.0 Nm/3.7 ft*lb).
> Do not exceed the permissible measuring range (0 to 870 psi /
0 to 60 bar). Pay particular attention with systems with
refrigerant R744, as these are often operated with higher
pressures.
13
Page 14
5 Using the manifold
Refrigerant evaporation
(psi/bar)
Refrigerant condensation
(psi/bar)
Measured temperature T1/t
(psi/bar)
Measured temperature T2/
(psi/bar)
Superheating SH/Δt
(°F/°C)
(psi/bar)
Subcooling SC/Δt
(°F/°C)
(psi/bar)
Measuring
1. Connect and apply pressure to the manifold.
2. See readings.
Note: With refrigerants that have a temperature glide, “Zeotropes”
the evaporation temperature Ev/to and condensation temperature
Co/tc are displayed after evaporation and condensation are
complete.
Zeotropes (refrigerants blends mix together) can separate from
each other, unlike azeotropes which mix together to become one.
Zeoptropes often blend refrigerants with different boiling points
(saturation temps), where one will change from liquid to vapor
before the other as they go through the evaporator. The glide is
the difference between the lowest boiling point and the highest
boiling point. If they are 3 degrees apart, for example, the blend
has a 3 degree glide..
The display illumination will flash if:
• The critical pressure of the refrigerant is within 15 psi (1 bar)
of the highest pressure (and temperature) where the
refrigerant can still condense.
• The maximum. permissible pressure of 870 psi (60 bar) is
exceeded.
Key functions
> Press [▲]
Possible display combinations:
or[▼] to change the reading in the display.
temperature Ev/to (°F/°C)
Evaporation pressure
(°F/°C)
Evaporation pressure
Evaporation pressure
14
temperature Co/tc (°F/°C)
Condensation pressure
oh
tcu
(°F/°C)
Condensation pressure
oh
cu
Condensation pressure
Page 15
5 Using the manifold
Changeable
- Evaporation
Temperature (Ev)
- Measured
Temperature (T1)
- Superheating
(SH)
Changeable
- Condensation
Temperature (Co)
- Measured
Temperature (T2)
- Subcooling (SC)
Evaporation
pressure
Condensation
pressure
With both NTC temperature probes connected, Δt is also shown.
> Press [Mean/Min/Max]: to display min. / max. readings and
mean values.
Leak test / pressure drop test
Systems can be tested for tightness with the temperaturecompensated leak test. The system pressure and the ambient
temperature are measured over a defined period of time, typically
with an inert gas such as Nitrogen. A temperature probe can be
connected that measures the ambient temperature. : Optional air
temperature probe, part. no. 0613 1712) is recommended.
Measurement data of the temperature-compensated differential
pressure and temperature, from start to the end of the test, is
displayed. It is possible to perform a leak test without connecting a
temperature probe.
1. Press [Mode] ΔP is displayed.
2. Start the leak test: Press [R, ►, ■]. ΔP is now flashing and
hh:mm timer is on.
3. End the leak test: Press [R, ►, ■]. ΔP stops flashing and
hh:mm timer stops.
- Result is displayed. Note: Leak test time duration and ΔP value
4. Confirm the message: Press [Mode] to exit leak mode.
15
Page 16
5 Using the manifold
Duration
of test
Differential
pressure start
compared to
actual
Measured
pressure
beginning
of test
Actual
measured
pressure
Vacuum measurement
✓ The vacuum probe is plugged into the front connection of the
manifold and connected to the system.
1. Press [Mode] the vacuum mode is displayed.
If ambient pressure is applied to the vacuum probe, then oooo is
shown on the display.
2. Start the vacuum pump.
- Once the measuring range 0 to 20,000 microns is reached, the
current vacuum value is shown on the instrument display. The
instrument also displays the current ambient temperature, the
water evaporation temperature, which corresponds to the
vacuum reading, and the delta between these two
temperatures.
3. To leave vacuum mode, remove the vacuum probe from the
testo 557 or switch to the standard measurement view using the
[Mode] button.
16
Page 17
6 Technical data
6.1.1. Bluetooth Modul
The Bluetooth® option may only be operated in countries in
which it is type approved.
Feature Values
Bluetooth Range >20 m (free field)
Bluetooth type
LSD Science & Technology Co., Ltd
L Series BLE Module (08 Mai 2013) based
on TI CC254X chip
Qualified Design ID B016552
Bluetooth radio
Class 3
class
Bluetooth company 10274
European Union
Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary,
Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands,
Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden,
United Kingdom.
EFTA countries
Island, Suisse, Norway, Lichtenstein.
Other countries
USA, Turkey, Hong Kong, Canada
Information from the FCC (Federal Communications Commission)
This device complies with part 15 of the FCC Rules. Its
commissioning is subject to the two following conditions: (1) This
instrument must not cause any harmful interference and (2) this
instrument must be able to cope with interference, even if this has
undesirable effects on operation.
Changes
The FCC demands that the user be informed that any changes or
modifications to the instrument that are not explicitly approved by
Testo AG may void the user's right to use this instrument.
6 Technical data
17
Page 18
6 Technical data
5000 micron (from 10000 to 20.000 micron)
6.1.2.General technical data
Feature Values
Pressure: psi / kPa / MPa / bar
Parameters
Sensors
Measuring cycle 0.5 s
Interfaces
Measuring
ranges
Overload 940 psi, 65 bar, 6500 kPa, 6.5 MPa
Resolution
Temperature: °F / °C / K
Vacuum: micron / inHg / inH
mTorr /Torr / Pa
2 Pressure: sensors, 2 Temperature (NTC
Thermistors)
1. Fold out the hook, loosen the clip and remove the cover of the
battery compartment.
2. Remove discharged batteries/rechargeable batteries and insert
new batteries/rechargeable batteries (4x 1.5 V, type AA,
Mignon, LR6) in the battery compartment. Observe the polarity!
3. Set on and close cover of the battery compartment (clip must
engage).
4. Switch the instrument on.
20
Page 21
7 Maintaining the product
> Remove the vacuum probe from the testo 557!
Changing the valve or valve stem shutoff
Replacement of valve stem shutoffs or valves must be performed
by a qualified technician. Warranty coverage with respect to valve
stem shut-offs and valves will NOT apply to miss-use or damage
and is only applicable if failure is attributable to defective parts and
workmanship.
Send the measuring instrument to Testo Customer Service.
Cleaning the vacuum probe
Contaminants such as oil may impair the accuracy of the vacuum
sensor.
CAUTION
Carrying out cleaning with the probe connected may result in
damage to the probe!
CAUTION
Damage to the sensor due to sharp objects!
> Do not insert any sharp objects into the probe!
1. Remove the vacuum probe from the testo 557.
2. Put a few drops of rubbing alcohol into the sensor opening.
3. Seal the opening by placing your finger on it and shake the
vacuum probe briefly.
4. Remove all the alcohol from the probe.
5. Repeat this process at least twice.
6. Leave the probe to dry for at least 1 hour. To dry the sensor
faster, you can connect the probe directly to a vacuum pump
and draw vacuum.
21
Page 22
8 Tips and assistance
8 Tips and assistance
8.1. Questions and answers
Question Possible causes/solution
flashes
The instrument switches
off automatically.
Batteries are almost empty.
> Change batteries.
Residual capacity of the batteries is too
low.
> Change batteries.
uuuu lights up instead of
the parameter display
The permissible measuring range has
been undershot.
> Keep to the permitted measuring
range.
oooo lights up instead of
the parameter display
The permissible measuring range has
been exceeded.
> Keep to the permitted measuring
range.
8.2.Measurement parameters
Name Description
bar, °C psi, °F
Δtoh SH Superheating, evaporation pressure
Δtcu SC Subcooling, condensation pressure
to Ev Refrigerant evaporation temperature
tc Co Refrigerant condensation temperature
toh T1 Measured temperature, evaporation
tcu T2 Measured temperature, condensation
22
Page 23
8.3.Error reports
Question Possible causes/solution
----is lit up instead of
measurement parameter
display
Display EEP FAIL
Sensor or cable defective
> Please contact your dealer or
Testo Customer Service
Eeprom defective
> Please contact your dealer or
Testo Customer Service
Display BT ERR
No BT module connected or BT
module defective.
> Please contact your dealer or Testo
Customer Service
Display ERR 1-5
Vac-Probe defect
> Please contact your dealer or
Testo Customer Service
8.4.Accessories and spare parts
Description Article no.
Clamp probe for temperature measurement
at pipes (1,5m)
Clamp probe for temperature measurement
at pipes (5m)
Pipe wrap probe with Velcro tape for pipe
diameters of up to max. 75 mm, Tmax.
+75 °C, NTC
Watertight NTC surface probe 0613 1912
Precise, robust NTC air probe 0613 1712
External vacuum probe 0638 1557
For a complete list of all accessories and spare parts, please refer
to the product catalogues and brochures or look up our website at:
www.testo.com
If you have any questions, please contact your dealer or Testo
Customer Service. The contact details can be found on the back of
this document or on the Internet at www.testo.com/service-
contact.
Click onto your country’s flag for local support.