This book was written as a general guide. The author and publisher have neither
liability not can they be responsible to any person or entity for any
misunderstanding, misuse, or misapplication that would cause loss or damage of
any kind, including the loss of rights, material, or personal injury, or alleged to be
caused directly or indirectly by the information contained in this publication.
The author and publisher do not assume and expressly disclaim any obligations
to obtain and include any additional information. The reader is expressly warned
to consider and adopt all safety precautions that might be indicated by activities
herein, and to avoid all potential hazards. By following instructions contained
herein, the reader willingly assumes all risks in connection with such instructions.
WARNING
Information contained is only for use by formally trained competent technicians
practicing within the HVAC/R community. The manufacturers’ installation,
operation, and service information should always be consulted, and should be
considered the first and best reference for installing, commissioning and
servicing equipment. The author and publisher assume no liability for
typographical errors or omissions of information in this guide.
CAUTION
EPA-Approved Section 608 certification is legally required to service
building air conditioning and refrigeration systems containing CFC and
HCFC (Class 1 and 2 refrigerants). This includes the connection of analog
refrigerant pressure gauges or digital refrigeration system analyzers to any
Note: When installing the button cell, hold the
meter with the carbineer toward you, (upside
down). This will prevent the battery from
accidentally falling into the wireless card slot.
1) Insert button cell (included, 1 CR2032) in the
battery compartment (+) side up. This is a
buffer battery to hold memory when batteries
are changed.
2) If equipped, insert wireless radio module
ribbed side up, contacts up, label down until
module engages.
3) Insert 4 type AA alkaline batteries, or 4
rechargeable into the battery compartment,
Battery compartment located under
carbineer, Remove battery
compartment door, insert batteries
and wireless card (if equipped) as
shown.
observe polarity. Do not mix battery types
4) Close the battery compartment. Push until a
click is felt.
Note: When not using for a long period of time,
upload the memory to a PC and remove the AA
batteries. Do not remove the button cell.
Switching the Instrument on:
The first time a new instrument is powered or when the batteries are replaced,
the unit requires a general configuration including the language, battery type and
date and time. After these parameters are set, they will not need to be adjusted
until the batteries expire.
First time power up or after batteries are replaced.
•Press the power button
•Initialization phase:
•All segments are displayed for 2 seconds.
•Instrument model, firmware version and serial number are displayed for 2
seconds.
1) Set the language using the arrow keys and confirm the entry with the OK
function key (Soft Key) English users select “English US”
i. Set battery type is displayed
2) Set the battery type. Using the arrow keys select the battery type and
confirm using the OK soft key.
i. Set date/time displayed
3) Set the time and date. Set the blinking number with the arrow keys.
Change to the next number using the left or right soft keys. Confirm entry
with the OK soft key
i. The configuration menu is opened.
Normal Power Up
•Press the power button
•Initialization phase:
•All segments are displayed for 2 seconds.
•Instrument model, firmware version and serial number are displayed for 2
seconds.
•The measurement view is opened
Initial Unit configuration
The 556/560 requires initial configuration of the units of measure. Once
configured, these parameters are stored in memory and will not need to be
changed again unless the button cell is removed for an extended period of time.
Changing the main batteries will not affect these settings.
Setting Relative or Absolute Pressures and Nulling (zeroing the sensor)
Use this menu to configure for relative or absolute pressure measurement.
Relative pressure refers to standard atmospheric pressure, 0 psig at sea level;
absolute pressure is 14.7 psia at sea level. After the pressure type has been
selected, remove any hoses if installed and null the pressure sensor
From the measurement menu
1) Press the Menu key
2) Menu: Pressure Nulling (Zeroing) relative/absolute is displayed
3) Using the arrow keys scroll up to select relative or absolute pressure
a. US users will select relative pressure (Prel)
4) Confirm entry with the OK soft Key
5) Press Menu
6) Menu: Pressure Nulling (Zeroing) relative/absolute is displayed
a. The p=0 is displayed
7) Confirm entry with the OK soft Key
a. The pressure sensors have now been nulled or zeroed.
b.
Note it is not possible to null the sensors with over 14.7 psi (rel) on the sensors.
Configuring the Units of Measure
The Units configuration menu allows the user to select the units of measure
desired. Selectable units are for pressure, temperature, altitude, mass, and in the
560 vacuum. Units can be selected in metric or U.S. Once configured, the
selections are stored and will not normally need to be reconfigured.
Loading...
+ 10 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.