8.4. Accessories and spare parts ......................................................... 21
9 EC declaration of conformity ................................................................ 22
3
2 Safety and the environment
Representation
Explanation
Warning advice, risk level according to the
signal word:
Warning! Serious physical injury may occur.
Caution! Slight physical injury or damage to
the equipment may occur.
> Implement the specified precautionary
measures.
Menu
Elements of the instrument, the instrument
display or the program interface.
[OK]
Control keys of the instrument or buttons of
the program interface.
2 Safety and the environment
2.1. About this document
Use
> Please read this documentation carefully and familiarize
yourself with the product before putting it to use. Pay particular
attention to the safety instructions and warning advice in order
to prevent injuries and damage to the product.
Symbols and writing standards
2.2.Ensure safety
> Do not operate the instrument if there are signs of damage at
the housing, mains unit or hoses.
> Do not store the product together with solvents. Do not use any
desiccants.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation.
> Dangers may also arise from the refrigeration systems being
measured or the measuring environment: Note the safety
regulations valid in your area when performing the
measurements.
4
> If the manifold falls or another comparable mechanical load
occurs, the pipe sections of the refrigerant hoses may break.
The valve stem shutoff may also be damaged, whereby further
damage to the interior of the manifold may occur that cannot be
identified from the outside. The refrigerant hoses must therefore
be replaced with new, undamaged refrigerant hoses every time
the manifold falls or following any other comparable mechanical
load. Send the manifold to Testo Customer Service for a
technical check for your own safety.
> To prevent damage from ESD (electro-static discharge) or
transient voltage spikes make sure that your refrigeration
system is properly grounded, as otherwise the manifold might
get damaged.
2.3. Protecting the environment
> Dispose of spent batteries in accordance with the valid legal
specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
regulations) or return the product to Testo for disposal.
> Refrigerant gases can harm the environment. Please note the
applicable environmental regulations.
3 Product description
3 Product description
3.1. Use
Testo 549 and testo 550 are digital manifolds for maintenance and
service work on refrigeration systems and heat pumps. They are
intended for use by qualified personnel only.
The functions of the testo 549 and testo 550 are designed to
replace analog manifolds, thermometers and pressure/temperature
charts. Pressures and temperatures can be applied, adapted,
tested and monitored.
Testo 549 and testo 550 are compatible with most non-corrosive
refrigerants, water and glycol. Testo 549 and testo 550 are not
compatible with refrigerants containing ammonia.
The instruments must not be used in explosive environments!
5
3 Product description
Icon
Significance
Battery status
Bluetooth®
//
Measuring mode
3.2.Overview
Display and control elements
1 Mini-DIN probe socket for NTC temperature probe, with socket
cover
2 Foldable hanging hook (on rear)
3 Display. Instrument status icons:
Switching the instrument on/off
Switch the display illumination on/off.
When not in use for long period: Remove batteries/rechargeable
batteries.
5 Control keys:
6 Sight glass for refrigerant flow
7 2 x valve stem shutoff
8 3 x hose holders for refrigerant hoses
9 3 x connections 1/4" SAE, brass
Left/right: Low pressure/high pressure, for refrigerant hoses
with quick connect fitting
Center: for charge / discharge of refrigerant
10 Mini-USB connection for firmware update, inside the battery
compartment.
4First steps
Inserting batteries / rechargeable batteries
1. Fold out the hanging hook and open the battery compartment
by squeezing the clip lock.
2. Insert batteries (included in delivery) or rechargeable batteries
(4x 1.5 V, type AA/Mignon/LR6) in the battery compartment.
Observe the polarity!
3. Close the battery compartment.
- After inserting the batteries, the instrument switches on
automatically and goes into the settings menu.
7
4 First steps
Key functions
Description
[▲] or [▼]
Change parameters and select units
[Set]
Confirm units / parameters
Selectable
parameters
Description
°F, °C
Temperature unit
Psi, bar, kPa, MPa
Unit of pressure
Psig or Pabs, Prel
Switch between absolute and relative
pressure display
Pamb
Set current absolute pressure in inches of HG
or bar
/ /
Select measuring mode: heat pump / cooling
/ or Auto
AUTO OFF
Activate or deactivate Automatic power-off.
Instrument switches off after 30 minutes if no
temperature probe is connected and there is
no pressure apart from ambient pressure
T
fac
Activate or deactivate surface temperature
compensation factor, icon is shown on the
display if the function is deactivated
Units / Parameter selection
1. Press [Set] to confirm or change unit parameter settings
2. Press [▲] or [▼] to change the units / parameters.
The settings will be accepted once the last selection has been
made.
Settings will be applied following the final selection.
Operating valve stem shutoffs
The digital manifold acts like a conventional two-way manifold with
regard to the refrigerant path: The passages are opened by
opening the valves. The adjacent pressure is measured with valves
closed as well as with them open.
> Open valve: Turn valve positioner counterclockwise.
> Close valve: Turn valve positioner clockwise.
8
Loading...
+ 16 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.