Testo 184 H1 Operating Manual

testo 184 · Data logger
Instruction manual
2
1 Safety and the environment
1.1. About this document
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay particular attention to the safety instructions and warning advice in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
1.2. Ensure safety
product.
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not use any force.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the prescribed steps exactly. Use only original spare parts from Testo.
2 Specifications
1.3. Protecting the environment
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
2 Specifications
2.1. Use
regulations) or return the product to Testo for disposal.
The testo 184 data loggers are used to save and read out individual readings and measurement series. They have been specifically designed for monitoring the transportation of products subject to cold chain requirements.
Temperature and humidity readings are saved throughout the entire duration of the measurement program.
2 Specifications
Acceleration readings are monitored throughout the duration of the measurement program and saved when the set limit value is exceeded.
Data logger programming and measurement report output are implemented via PDF files, no software needs to be installed.
Product versions T1 and T2 are single-use data loggers with a time-limited service life.
2.2. Technical data
testo 184 T1, T2, T3, T4
Feature Values
Display T1, T4: no
T2, T3: yes
Probe type T1, T2, T3: internal NTC temperature sensor
T4: internal PT1000 temperature sensor
Measuremen
1 internal
t channels
Measure-
Temperature [°C, °F] ment parameters [unit]
Measuring range
T1, T2, T3: -35 to 70 °C
T4: -80 to 70 °C
Accuracy T1, T2, T3: ±0.5 K
T4: ±0.8 K (-80 to -35.1 °C), ±0.5 K (-35.0 to 70 °C)
Resolution 0.1 °C
Operating temperature
Storage temperature
T1, T2, T3: -35 to 70 °C
T4: -80 to 70 °C
T1, T2, T3: -35 to 70 °C
T4: -80 to 70 °C
Battery type T1: internal, non-replaceable
T2: internal, non-replaceable
T3: CR2450, replaceable
T4: TLH-2450, replaceable
4
2 Specifications
Feature Values
Battery life (reusable
T3: 500 days (15-minute measuring cycle, 25 °C) T4: 100 days (15-minute measuring cycle, -80 °C)
data loggers)
Operating time (single­use data loggers)
Protection
T1: 90 days from the first time the program is started (5-minute measuring cycle, -35 °C)
T2: 150 days from the first time the program is started (5-minute measuring cycle, -35 °C)
IP67
class
Measuring
1 minute to 24 hours
interval
Memory
T1: 16,000 readings T2, T3, T4: 40,000 readings
Dimensions
T1: 33 x 9 x 74 mm T2, T3, T4: 44 x 12 x 97 mm
Weight T1: 25 g
T2, T3, T4: 45 g
Directives, standards, certificates
Warranty/ guarantee
2014/30/EC, EN 12830, HACCP-certified, temperature calibration certificate traceable according to ISO 17025
T1, T2: 12-month warranty from the production date, production date: see circled date code (MMDD) on the identification plate.
T3, T4: 24-month warranty, warranty terms: see website www.testo.com/warranty
testo 184 H1, G1
Feature Values
Display yes
Probe type H1: internal digital humidity sensor
G1: internal 3-axis accelerometer
Measuremen t channels
Measure­ment parameters
H1: 2 internal G1: 5 internal
H1: temperature [°C, °F], relative humidity [%] G1: temperature [°C, °F], relative humidity [%],
acceleration [g, m/s²]
[unit]
2 Specifications
Feature Values
Measuring range
-20 to 70 °C
0 to 100 % (not for condensing atmospheres)
1
G1: 0 to 27 g
Accuracy ±0.5 K (0.0 to 70 °C), ±0.8 K (-20 to -0.1 °C)
±1.8 % RH + 3 % of the reading (at 25 °C, 5 to
80 %), ±0.03 % RH / K (at 0 to 60 °C)
G1: ±1,1,1 m/s² + 5 % of the reading
Resolution 0.1 °C
0.1 % RH
G1: 0.1 g
Operating
-20 to 70 °C
temperature
Storage
-55 to 70 °C
temperature
Battery type CR2450, replaceable
Battery life (reusable
H1: 500 days (15-minute measuring cycle, 25 °C)
G1: 120 days (15-minute measuring cycle, 25 °C) data loggers)
Protection
IP 30 class
Measuring interval
1 minute to 24 hours (temperature and relative
humidity)
1 second (acceleration)
Scan
1600Hz (acceleration) frequency
Memory 64,000 readings (temperature and relative humidity)
G1: 1,000 readings (acceleration)
Dimensions 44 x 12 x 97 mm
Weight 45 g
Directives,
2014/30/EC, HACCP-certified standards, certificates
1
For continuous application in high humidity (> 80 % RH at 30 °C for > 12 h, > 60 % RH at > 30 °C for > 12 h), please contact us via www.testo.com/service­contact