Latest generation testo air velocity & IAQ probes with
Bluetooth® handle
Instruction manual
www..com
information@itm.com1.800.561.8187
Page 2
1 Safety and disposal
About this document
Security
Batteries
• The instruction manual is an integral part of the instrument.
• Please read this instruction manual through carefully and familiarize yourself
with the product before putting it to use.
• Pay particular attention to the safety instructions and warning advice in
order to prevent injury and damage to the product.
• Keep this documentation to hand so that you can refer to it when
necessary.
• Always use the complete original instruction manual.
• Hand this instruction manual on to any subsequent users of the product.
• Only use the product properly, for its intended purpose, and within the
parameters specified in the technical data. Do not apply any force.
• Dangers may also arise from the systems being measured or the measuring
environment: always comply with the locally valid safety regulations when
carrying out measurements.
• Do not carry out any contact measurements on uninsulated, live parts.
• Do not store the product together with solvents. Do not use any desiccants.
• Only carry out maintenance and repair work on this instrument which is
described in the documentation. Follow the prescribed steps exactly when
doing the work. Use only original spare parts from Testo.
• Temperature information given on probes/sensors relates only to the
measuring range of the sensor technology. Do not expose handles and feed
lines to temperatures in excess of 50°C (122°F), unless they are expressly
authorized for use at higher temperatures.
• Do not operate the instrument if there are signs of damage on the housing
or supply lines.
1 Safety and disposal
• Improper use of batteries may cause destruction of the batteries, injuries
due to current surges, fire or the escape of chemicals.
• Only use the batteries supplied in accordance with the instructions in the
instruction manual.
• Do not short, take apart or modify batteries.
www..com
information@itm.com1.800.561.8187
3
Page 3
2 System description
Disposal
2 System description
• Do not expose the batteries to heavy impacts, water, fire or temperatures
above +140 °C or below -20 °C.
• Do not store unused batteries in the proximity of metal objects.
• Do not use any leaky or damaged batteries.
• In the event of contact with battery acid: rinse affected areas thoroughly
with water, and if necessary consult a doctor.
• Dispose of faulty rechargeable batteries and spent batteries in accordance
with the valid legal specifications.
• At the end of its useful life, deliver the product to the separate collection
point for electric and electronic devices (observe local regulations) or returnthe product to Testo for disposal.
You have acquired a probe with Bluetooth®. This probe comprises a handle, a
probe head and, if applicable, additional probe-specific accessories.
The handle can be connected to all the system’s probe heads and
attachments. The graphic below shows what options the system offers.
You will find detailed information about how each of the probes works
along with the testo 440 measuring instrument in the corresponding
chapter of the instruction manual for the measuring instrument.
4
www..com
information@itm.com1.800.561.8187
Page 4
System overview
1
2
3
4
5
6
(order no. 0554 2160)
2 System description
Bluetooth® handle for
connecting testo 440 probe
heads
(order no. 0554 1111)
High-precision
temperature/humidity probe
head
(order no. 0636 9770)
CO probe head
(order no. 0632 1270)
CO2 probe head, including
humidity and temperature sensor
(order no. 0632 1550)
Temperature/humidity probe
head (order no. 0636 9730)
Handle adapter for connecting
testo 440 flow probes
Extendable telescope for
testo 440 flow probes, including
16 mm vane probe head,
including temperature sensor
The handle with Bluetooth® 0554 1111 is used for the wireless connection of
probe heads to the testo 440 measuring instrument.
Observe the information on the measuring process in the detailed
instrument instruction manual.
High-precision 100 mm vane
probe head, including
temperature sensor (order no.
0635 9370)
Hot wire probe head, including
temperature and humidity sensor
Element Element
Quick-release connector for
LED status display
6
www..com
information@itm.com1.800.561.8187
Page 6
Remove battery safety strips
1
Pull safety strips out of the
battery compartment.
3.2 Hot wire probe with Bluetooth®,
including temperature and humidity
sensor (0635 1571)
Use
Structure
1
2
3
4
5
probe heads and adapter
6
7
8
9
CAUTION
Damage to sensor possible!
In conjunction with the testo 440, the hot wire probe is suitable for flow and
humidity measurements in ventilation ducts.
3 Description of the instrument
Probe head
Telescope
Quick-release connector for
Button
Battery compartment (on rear)
Probe head connection
Handle adapter
LED status display
Bluetooth® handle
- Do not touch sensor!
- Put the protective sleeve on after the measurement.
7
www..com
information@itm.com1.800.561.8187
Page 7
3 Description of the instrument
Connect hot wire probe to the handle
1
Turn quick-release connector,
2
Insert handle adapter into the handle
3
Turn quick-release connector back
4
Insert telescope into the handle
5
Push telescope in slightly and turn to
6
Push probe head onto the telescope
7
Low flow velocities may result in greater measurement uncertainties
during temperature and humidity measurement. The probe should be
switched on outside a duct under the following conditions:
Ambient temperature: 20°C
Flow: approx. 0 m/s.
viewed from the handle end, through
90° anticlockwise until it locks in
place.
through 90° until it locks in place.
(pay attention to coding).
adapter (pay attention to the guide
groove and guide pin).
lock in place.
(pay attention to the guide groove
and guide pin).
Push probe head in slightly and turn
to lock in place.
The hot wire probe is ready to use.
8
www..com
information@itm.com1.800.561.8187
Page 8
3 Description of the instrument
3.3 Vane probe (Ø 16 mm) with Bluetooth®,
including temperature sensor (0635
9571)
Application
Structure
1
Probe head
2
Probe head connection
3
Telescope
4
Handle adapter
5
probe heads and adapter
6
7
Button
8
Bluetooth® handle
9
Battery compartment (on rear)
CAUTION
Damage to sensor possible!
Connect the vane probe (Ø 16 mm) to the handle
1
Turn quick-release connector,
2
Insert handle adapter into the handle
3
Turn quick-release connector back
In conjunction with the testo 440, the vane probe (Ø 16 mm) is suitable for flow
measurements in ventilation ducts.
Quick-release connector for
- Do not touch sensor!
- Put the protective sleeve on after the measurement.
viewed from the handle end, through
90° anticlockwise until it locks in
place.
(pay attention to coding).
through 90° until it locks in place.
LED status display
www..com
9
information@itm.com1.800.561.8187
Page 9
3 Description of the instrument
4
Insert telescope into the handle
5
Push telescope in slightly and turn to
6
7
Push probe head in slightly and turn
3.4 High-precision vane probe (Ø 100 mm)
with Bluetooth®, including temperature
sensor (0635 9371)
Application
Structure
1
Probe head
2
Handle adapter
3
4
5
6
Bluetooth® handle
7
Battery compartment (on rear)
adapter (pay attention to the guide
groove and guide pin).
lock in place.
Push probe head onto the telescope
(pay attention to the guide groove
and guide pin).
to lock in place.
The vane probe (Ø 16 mm) is ready to use.
In conjunction with the testo 440, the high-precision vane probe (Ø 100 mm) is
used for flow measurement, e.g. at ventilation outlets.
Quick-release connector for
probe heads and adapter
Button
10
LED status display
www..com
information@itm.com1.800.561.8187
Page 10
CAUTION
Damage to sensor possible!
- Do not touch sensor!
Connect high-precision vane probe (Ø 100 mm) to the handle
1
Turn quick-release connector, viewed
2
Insert handle adapter into the handle
3
through 90° until it locks in place.
4
Insert probe head into the handle
5
to lock in place.
The testovent bracket included in the package is intended for the
testovent 417 measurement funnel.
from the handle end, through 90°
anticlockwise until it locks in place.
(pay attention to coding).
Turn quick-release connector back
adapter (pay attention to the guide
groove and guide pin).
Push probe head in slightly and turn
3 Description of the instrument
The high-precision vane probe is ready to use.
If necessary, the telescope (order no. 0554 0960) and the 90° angle
(order no. 0554 0991) can be connected, e.g. for measurements at
ceiling outlets.
11
www..com
information@itm.com1.800.561.8187
Page 11
3 Description of the instrument
3.5 Vane probe (Ø 100 mm) with Bluetooth®,
including temperature sensor (0635
9431)
Application
Structure
1
Probe head
2
Handle adapter
3
probe heads and adapter
4
5
Button
6
Bluetooth® handle
7
Battery compartment (on rear)
CAUTION
Damage to sensor possible!
Connect vane probe (Ø 100 mm) to the handle
1
place.
2
In conjunction with the testo 440, the vane probe (Ø 100 mm) is used for flow
measurement, e.g. at ventilation outlets.
Quick-release connector for
- Do not touch sensor!
The testovent bracket included in the package is intended for the
testovent 417 measurement funnel.
Turn quick-release connector,
viewed from the handle end, through
90° anticlockwise until it locks in
Insert handle adapter into the handle
(pay attention to coding).
LED status display
12
www..com
information@itm.com1.800.561.8187
Page 12
3 Description of the instrument
3
through 90° until it locks in place.
4
Insert probe head into the handle
5
Push probe head in slightly and turn
3.6 High-precision temperature/humidity
probe with Bluetooth® (0636 9771)
Application
Structure
1
Probe head
2
Quick-release connector
3
LED status display
4
Button
5
Bluetooth® handle
6
Battery compartment (on rear)
CAUTION
Damage to sensor possible!
Turn quick-release connector back
adapter (pay attention to the guide
groove and guide pin).
to lock in place.
The vane probe is ready to use.
If necessary, the telescope (order no. 0554 0960) and the 90° angle
(order no. 0554 0991) can be connected, e.g. for measurements at
ceiling outlets.
In conjunction with the testo 440, the high-precision temperature/humidity
probe is used to measure humidity and temperature.
- Do not touch sensor!
13
www..com
information@itm.com1.800.561.8187
Page 13
3 Description of the instrument
Connect high-precision temperature/humidity probe to the handle
1
place.
2
Insert probe head into the handle
3
Turn quick-release connector back
3.7 Temperature/humidity probe with
Bluetooth® (0636 9731)
Turn quick-release connector,
viewed from the handle end, through
90° anticlockwise until it locks in
(pay attention to coding).
through 90° until it locks in place.
The high-precision temperature/humidity probe is ready to use.
In conjunction with the testo 440, the temperature/humidity probe is used to
measure humidity and temperature.
Probe head
LED status display
Quick-release connector for
probe heads and adapter
Button
14
- Do not touch sensor!
www..com
information@itm.com1.800.561.8187
Page 14
3 Description of the instrument
Connect temperature/humidity probe to the handle
1
Turn quick-release connector,
2
Insert probe head into the handle
3
Turn quick-release connector back
3.8 CO2 probe with Bluetooth®, including
temperature and humidity sensor (0632
1551)
Application
Structure
1
2
3
4
5
Bluetooth® handle
6
Battery compartment (on rear)
viewed from the handle end, through
90° anticlockwise until it locks in
place.
(pay attention to coding).
through 90° until it locks in place.
The temperature/humidity probe is ready to use.
In conjunction with the testo 440, the CO2 probe is used to determine CO2 and
to measure humidity and temperature.
Probe head
LED status display
Quick-release connector for
probe heads and adapter
Button
www..com
15
information@itm.com1.800.561.8187
Page 15
3 Description of the instrument
CAUTION
•
•
•
•
•
•
s to acclimatize to
•
Connect CO2 probe to the handle
1
Turn quick-release connector,
2
Insert probe head into the handle
3
The probe contains sensitive visual components. Please handle the probe
with care.
Strong vibrations alter the factory calibration. Checking of readings in fresh
air 350 to 450 ppm CO2 (urban air up to 700 ppm CO2).
Avoid dew on the probe, otherwise long-term stability will be impaired. If
If the ambient temperature changes (change of measuring location, e.g.
After the instrument is switched on, there is a sensor warm-up phase of
The CO2 concentration in the sensor requires approx. 60
Keep the probe as far away from your body as possible. This avoids any
there is dew on the probe, this can lead to higher CO2 readings.
indoors to outdoors or vice versa), the probe requires an acclimatization
phase of a few minutes.
approx. 30 s.
the environment. Gently waving the probe reduces the acclimatization time.
influences due to the CO2 content of the air you breathe.
viewed from the handle end, through
90° anticlockwise until it locks in
place.
16
(pay attention to coding).
Turn quick-release connector back
through 90° until it locks in place.
The CO2 probe is ready to use.
www..com
information@itm.com1.800.561.8187
Page 16
3 Description of the instrument
3.9 CO probe with Bluetooth® (0632 1271)
Application
Structure
1
2
probe heads and adapter
3
LED status display
4
Button
5
Bluetooth® handle
6
Battery compartment (on rear)
DANGER
Risk of death!
Carbon monoxide is a colourless, odourless and tasteless gas. High
concentrations can cause death.
Connect CO probe to the handle
1
Turn quick-release connector,
2
Insert probe head into the handle
3
Turn quick-release connector back
In conjunction with the testo 440, the CO probe is used to determine the CO
concentration in the environment.
Probe head
- Do not use the CO probe as a monitoring instrument for personal safety.
The CO probe is ready to use.
Quick-release connector for
viewed from the handle end, through
90° anticlockwise until it locks in
place.
(pay attention to coding).
through 90° until it locks in place.
www..com
17
information@itm.com1.800.561.8187
Page 17
4 Commissioning
4 Commissioning
Switching on
1
- Press button on the handle.
Switching off
1
-
LED status display
not connected.
Bluetooth is connected.
The handle is switched on.
Press and hold the button on the handle for three seconds.
The handle is switched off.
LED status display Meaning
Flashing red Low battery status
Flashing yellow Measurement probe is switched on and is
Flashing green Measurement probe is switched on and
searching for a Bluetooth connection, but is
18
www..com
information@itm.com1.800.561.8187
Page 18
5 Maintenance
5 Maintenance
Replacing the batteries
1
Open the battery compartment:
2
Take spent batteries out of the
3
Put battery cover on and push lock
5.1 Maintaining the probes
5.1.1 Cleaning the instrument
push lock down and remove cover.
battery compartment and replace
with new batteries (4 x AA type).
attention to polarity.
back up.
Handle is ready to use again.
Do not use any aggressive cleaning agents or solvents; instead use
mild household cleaning agents or soapy water.
Always keep the connections clean and free from grease and other
Clean the instrument and the connections with a damp cloth and dry them off.
deposits.
Pay
www..com
19
information@itm.com1.800.561.8187
Page 19
6 Questions and answers
5.1.2 Calibration
use service software.
6 Questions and answers
the batteries.
The probes are supplied with a factory calibration protocol as
standard.
Recalibration of the probes once every 12 months is recommended in
many applications.
This can be carried out by Testo Industrial Services (TIS) or other
certified service providers with the aid of easy-to-
Question Possible cause / solution
LED status display flashes
red
The probe switches itself off Remaining battery capacity is too low. Change
Batteries are almost empty. Change the
batteries.
20
www..com
information@itm.com1.800.561.8187
Page 20
7 Technical data
• Adjustment conditions for flow probes:
Please do not use the humidity probes in condensing atmospheres.
Bluetooth® handle for connecting testo 440 probe heads (0554 1111)
Technical data
Storage temperature
-10°C to +50°C
Operating temperature
0 to +50°C
Hot wire probe with Bluetooth®, including temperature and humidity sensor
(0635 1571)
700 to 1100 hPa
±3 hPa
• Note for flow probes:
• Note for humidity probes:
Feature Value
7 Technical data
Adjustment in free jet Ø 350 mm, reference pressure 1013 hPa,
based on testo reference Laser Doppler Anemometer (LDA).
Low flow velocities may result in greater measurement
uncertainties during humidity and temperature measurement!
For continuous use in high-humidity ranges
> 80% RH at ≤ 30°C for > 12 h
> 60% RH at > 30°C for > 12 h
please get in touch with Testo Service or contact us via the Testo
website.
Dimensions 170 mm x 50 mm x 40 mm
Warranty Duration: 2 years
Feature Value
Measuring range 0 to +50 m/s
Accuracy
(at 22°C, ±1 digit)
-20 to +70°C
5 to 95% RH
±(0.03 m/s + 4% of m.v.) (0 to 20 m/s)
±(0.5 m/s + 5% of m.v.) (20.01 to 30 m/s)
±0.8°C (-20 to 0°C)
±0.5°C (0 to +70°C)
21
www..com
information@itm.com1.800.561.8187
Page 21
7 Technical data
- Long-term stability: ±1% RH / year
0.1 hPa
Temperature coefficient
type (k=1) 0.06% RH/K
Storage temperature
-20 to +60°C
Battery life
21.5 h
Probe head length
230 mm
Ø probe head on the sensor
9 mm
800 mm
tests
Vane probe (Ø 16 mm) with Bluetooth®, including temperature sensor
(0635 9571)
-10 to +70°C
±1.8°C
0.1°C
Storage temperature
-20 to +60°C
Feature Value
Accuracy
(at 25°C, ±1 digit)
Resolution 0.01 m/s
Operating temperature -5 to +50°C
Bluetooth® range 20 m (free field)
±3.0% RH (10 to 35% RH)
±2.0% RH (35 to 65% RH)
±3.0% RH (65 to 90% RH)
±5% RH (remaining measuring range)
Additional uncertainty:
- Hysteresis: ±1.0% RH
0.1°C
0.1% RH
Dimensions of probe head
with telescope
Directives, standards and
Feature Value
Measuring range 0.6 to 50 m/s
Accuracy
(at 22°C, ±1 digit)
Resolution 0.1 m/s
Operating temperature -5 to +50°C
22
Extension length: 1000 mm
Probe length from handle end to probe head
end with retracted telescope:
EU directive: 2014/30/EU
±(0.2 m/s + 1% of m.v.) (0.6 to 40 m/s)
±(0.2 m/s + 2% of m.v.) (40.1 to 50 m/s)
Handle adapter for connecting testo 440 flow probes
0554 2160
bag)
including 90° angle
440 flow probes
Combi-case for testo 440 and multiple probes
0516 4401
Description Order no.
Hot wire probe head, including temperature and
humidity sensor
16 mm vane probe head, including temperature sensor 0635 9570
High-precision 100 mm vane probe head, including
temperature sensor
100 mm vane probe head 0635 9430
0635 1570
0635 9370
CO2 probe head, including temperature and humidity
Bluetooth® handle for connecting testo 440 probe
heads
Measuring stand for comfort level measurements with
standard-compliant positioning of probes (including
Extendable telescope for testo 440 flow probes,
Telescope extension (length 0.40 to 0.85 m) for testo
90° angle for connecting the testo 440 vane probe (Ø
100 mm)
0632 1550
0554 1111
0554 1590
0554 0960
0554 0990
0554 0991
28
www..com
information@itm.com1.800.561.8187
Page 28
9 Authorisations and certification
Product
0554 1111
Mat.-No.:
0554 1111
Date
06.12.2017
IC Warnings
Turkey
Authorized
Japan Information
Please note the following country-specific information for the product
authorization.
The use of the wireless module is subject to the regulations and
stipulations of the respective country of use and the module may only
be used in countries for which a national certification has been
granted. The user and every owner has the obligation to adhere to
these regulations and prerequisites for use and acknowledges that the
re-sale, export, import, etc., in particular in, to or from countries
without wireless authorization, is their responsibility.
Country Comments
Australia
9 Authorisations and certification
E 1561
Canada Product IC ID: 6127B-05541111
Europa + EFTA
Japan
EU countries:
Belgium (BE), Bulgaria (BG), Denmark (DK), Germany (DE), Estonia (EE),
Finland (FI), France (FR), Greece (GR), Ireland (IE), Italy (IT), Latvia (LV),
Lithuania (LT), Luxembourg (LU), M alta (MT), Netherlands (NL), Austria (AT),
Poland (PL), Portugal (PT), Romania (RO), Sweden (SE), Slovakia (SK),
Slovenia (SI), Spain (ES), Czech Republic (CZ), Hungary (HU), United
Kingdom (GB), Republic of Cyprus (CY).
This device complies with Industry Canada licence-exempt RSS standard(s).
Operation is subject to the following two conditions:
(1) this device may not cause interference, and
(2) this device must accept any interference, including interference that may cause undesired
operation of the device.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio
exempts de licence.
L'exploitation est autorisée aux deux conditions suivantes :
(1) l'appareil ne doit pas produire de brouillage, et
(2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage
est susceptible d'en compromettre le fonctionnement.
Caution: Radio Frequency Radiation Exposure
This equipment complies with IC radiation exposure limits set forth for an uncontrolled environment
and meets the IC radio frequency (RF) Exposure Guidelines. This equipment should be installed and
operated keeping the radiator at least 20 cm or more away from person´s body in normal use position.
Co-Location:
This transmitter must not be co-located or operated in conjunction with any other antenna or
transmitter.
Attention : exposition au rayonnement de radiofréquences
Cet équipement est conforme aux limites d'exposition aux radiofréquences IC fixées pour un
environnement non contrôlé et aux Lignes directrices relatives à l'exposition aux radiofréquences (RF).
Cet équipement devrait être installé et utilisé à une distance d'au moins 20 cm d'un radiateur ou à une
distance plus grande du corps humain en position normale d'utilisation.
Co-location
Ce transmetteur ne peut pas être installé en colocation ou être utilisé avec une autre antenne ou
transmetteur, quel qu'en soit le type.
Bluetooth range < 20 m (free field)
Bluetooth type L Series BLE module (08 May 2013) based on
Output power 0 dBm
Information from the FCC (Federal Communications Commission)
30
www..com
information@itm.com1.800.561.8187
Page 30
FCC warning statement
Caution
Warning
Caution: Radio Frequency Radiation Exposure
Japan Information
9 Authorisations and certification
Shielded cables should be used for a composite interface. This is to ensure continued protection
against radio frequency interference.
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference in a residential installation. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instructions, may
cause harmful interference to radio communications. However, there is no guarantee that interference
will not occur in a particular installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
• Consult the dealer or an experienced radio/TV technician for help.
Changes or modifications not expressly approved by the party responsible for compliance could void
the user's authority to operate the equipment. Shielded interface cable must be used in order to
comply with the emission limits.
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received,
including interference that may cause undesired operation.
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment
and meets the FCC radio frequency (RF) Exposure Guidelines. This equipment should be installed and
operated keeping the radiator at least 20 cm or more away from person´s body in normal use position.
当該機器には電波法に基づく、技術基準適合証明等を受けた特定無線設備を装着している。
www..com
31
information@itm.com1.800.561.8187
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.