Testo Instruments 0635 1047 User manual

Page 1
Thermal anemometer probe
0635.1047
Application Information en
General Information
Please also refer to the documentation included with the mea­suring instrument.
Intended Use
The thermal anemometer probe is designed to measure and inspect laboratory fume cupboards in conjunction with the testo 400 measuring instrument.
Technical data
Feature Values
Application temperature 0 to +50 °C/32 to 122 °F Sensor Thermal Measurement range 0 to 5.00 m/s Accuracy (at 22 °C) ± (0.02 m/s + 5 % of reading) Time constant < 0.5 s Adjustment In free jet Ø 350 mm, reference pressure 1013 hPa,
referred to testo reference DKD laboratory velocity. Absolute pressure Compensation in testo 400, testo 454 Standard Corresponds to EN 14175-3, -4 (Draft)
Operation
Setting up probe
1 Connect probe to measuring instru-
ment.
2 Push down probe protection cap until it can go no further, both openings must be uncovered.
3 Pull out probe to the required length.
Measuring
Align the probe head such that the
flow direction corresponds to the arrow on the probe head.
In the case of deviations from the exact alignment, the probe shows the correct reading even if the angle is turned by 20°.
Finish measurement
Push up probe protection cap as far
as it will go so that the measurement sensor is protected.
Find Quality Products Online at: [email protected]
www.GlobalTestSupply.com