1 Safety and disposal

About this document
Security
Batteries
The instruction manual is an integral part of the instrument.
Please read this instruction manual through carefully and familiarize yourself
with the product before putting it to use.
Pay particular attention to the safety instructions and warning advice in
order to prevent injury and damage to the product.
Keep this documentation to hand so that you can refer to it when
necessary.
Always use the complete original instruction manual.
Hand this instruction manual on to any subsequent users of the product.
Only use the product properly, for its intended purpose, and within the
parameters specified in the technical data. Do not apply any force.
Dangers may also arise from the systems being measured or the measuring
environment: always comply with the locally valid safety regulations when carrying out measurements.
• Do not carry out any contact measurements on uninsulated, live parts.
• Do not store the product together with solvents. Do not use any desiccants.
Only carry out maintenance and repair work on this instrument which is
described in the documentation. Follow the prescribed steps exactly when doing the work. Use only original spare parts from Testo.
Temperature information given on probes/sensors relates only to the
measuring range of the sensor technology. Do not expose handles and feed lines to temperatures in excess of 50°C (122°F), unless they are expressly authorized for use at higher temperatures.
Do not operate the instrument if there are signs of damage on the housing
or supply lines.
1 Safety and disposal
Improper use of batteries may cause destruction of the batteries, injuries
due to current surges, fire or the escape of chemicals.
Only use the batteries supplied in accordance with the instructions in the
instruction manual.
Do not short, take apart or modify batteries.
www. .com
information@itm.com1.800.561.8187
3

2 System description

Disposal
2 System description
Do not expose the batteries to heavy impacts, water, fire or temperatures
above +140 °C or below -20 °C.
Do not store unused batteries in the proximity of metal objects.
• Do not use any leaky or damaged batteries.
In the event of contact with battery acid: rinse affected areas thoroughly
with water, and if necessary consult a doctor.
Dispose of faulty rechargeable batteries and spent batteries in accordance
with the valid legal specifications.
At the end of its useful life, deliver the product to the separate collection
point for electric and electronic devices (observe local regulations) or return the product to Testo for disposal.
You have acquired a probe with Bluetooth®. This probe comprises a handle, a probe head and, if applicable, additional probe-specific accessories. The handle can be connected to all the system’s probe heads and attachments. The graphic below shows what options the system offers.
You will find detailed information about how each of the probes works along with the testo 440 measuring instrument in the corresponding chapter of the instruction manual for the measuring instrument.
4
www. .com
information@itm.com1.800.561.8187
System overview
1
2
3
4
5
6
(order no. 0554 2160)
2 System description
Bluetooth® handle for connecting testo 440 probe heads (order no. 0554 1111) High-precision temperature/humidity probe head (order no. 0636 9770) CO probe head (order no. 0632 1270)
CO2 probe head, including humidity and temperature sensor (order no. 0632 1550)
Temperature/humidity probe head (order no. 0636 9730)
Handle adapter for connecting testo 440 flow probes
5
www. .com
information@itm.com1.800.561.8187

3 Description of the instrument

7
8
9
90° angle (order no. 0554 0960)
10
(order no. 0635 1570)
11
(order no. 0635 9570)
3 Description of the instrument

3.1 Bluetooth® handle for connecting testo 440 probe heads (0554 1111)

Application
Structure
1
probe heads and adapter
2 3
Button
4
Battery compartment (on rear)
100 mm vane probe head (order no. 0635 9430)
Extendable telescope for testo 440 flow probes, including
16 mm vane probe head, including temperature sensor
The handle with Bluetooth® 0554 1111 is used for the wireless connection of probe heads to the testo 440 measuring instrument.
Observe the information on the measuring process in the detailed instrument instruction manual.
High-precision 100 mm vane probe head, including temperature sensor (order no. 0635 9370)
Hot wire probe head, including temperature and humidity sensor
Element Element
Quick-release connector for
LED status display
6
www. .com
information@itm.com1.800.561.8187
Remove battery safety strips
1
Pull safety strips out of the
battery compartment.

3.2 Hot wire probe with Bluetooth®, including temperature and humidity sensor (0635 1571)

Use
Structure
1
2
3
4
5
probe heads and adapter
6 7
8
9
CAUTION
Damage to sensor possible!
In conjunction with the testo 440, the hot wire probe is suitable for flow and humidity measurements in ventilation ducts.
3 Description of the instrument
Probe head Telescope Quick-release connector for
Button Battery compartment (on rear)
Probe head connection Handle adapter LED status display
Bluetooth® handle
- Do not touch sensor!
- Put the protective sleeve on after the measurement.
7
www. .com
information@itm.com1.800.561.8187
3 Description of the instrument
Connect hot wire probe to the handle
1
Turn quick-release connector,
2
Insert handle adapter into the handle
3
Turn quick-release connector back
4
Insert telescope into the handle
5
Push telescope in slightly and turn to
6
Push probe head onto the telescope
7
Low flow velocities may result in greater measurement uncertainties during temperature and humidity measurement. The probe should be switched on outside a duct under the following conditions: Ambient temperature: 20°C Flow: approx. 0 m/s.
viewed from the handle end, through 90° anticlockwise until it locks in place.
through 90° until it locks in place.
(pay attention to coding).
adapter (pay attention to the guide groove and guide pin).
lock in place.
(pay attention to the guide groove and guide pin).
Push probe head in slightly and turn to lock in place.
The hot wire probe is ready to use.
8
www. .com
information@itm.com1.800.561.8187
3 Description of the instrument
3.3 Vane probe (Ø 16 mm) with Bluetooth®, including temperature sensor (0635
9571)
Application
Structure
1
Probe head
2
Probe head connection
3
Telescope
4
Handle adapter
5
probe heads and adapter
6 7
Button
8
Bluetooth® handle
9
Battery compartment (on rear)
CAUTION
Damage to sensor possible!
Connect the vane probe (Ø 16 mm) to the handle
1
Turn quick-release connector,
2
Insert handle adapter into the handle
3
Turn quick-release connector back
In conjunction with the testo 440, the vane probe (Ø 16 mm) is suitable for flow measurements in ventilation ducts.
Quick-release connector for
- Do not touch sensor!
- Put the protective sleeve on after the measurement.
viewed from the handle end, through 90° anticlockwise until it locks in place.
(pay attention to coding).
through 90° until it locks in place.
LED status display
www. .com
9
information@itm.com1.800.561.8187
3 Description of the instrument
4
Insert telescope into the handle
5
Push telescope in slightly and turn to
6 7
Push probe head in slightly and turn

3.4 High-precision vane probe (Ø 100 mm) with Bluetooth®, including temperature sensor (0635 9371)

Application
Structure
1
Probe head
2
Handle adapter
3
4 5
6
Bluetooth® handle
7
Battery compartment (on rear)
adapter (pay attention to the guide groove and guide pin).
lock in place.
Push probe head onto the telescope (pay attention to the guide groove and guide pin).
to lock in place.
The vane probe (Ø 16 mm) is ready to use.
In conjunction with the testo 440, the high-precision vane probe (Ø 100 mm) is used for flow measurement, e.g. at ventilation outlets.
Quick-release connector for probe heads and adapter
Button
10
LED status display
www. .com
information@itm.com1.800.561.8187
Loading...
+ 21 hidden pages