This manual contains proprietary information, which is protected by
copyright. All rights are reserved. No part of this manual may be
photocopied, reproduced or translated to another language without
prior written consent of the Good Will company.
The information in this manual was correct at the time of printing.
However, Good Will continues to improve products and reserves
the right to change specifications, equipment, and maintenance
procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
Declaration of Conformity ................................. 54
INDEX .............................................................................. 55
3
GOM-802 User Manual
W
A
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that
you must follow when operating the GOM-802 or when
keeping it in storage. Read the following before any
operation to insure your safety and to keep the
GOM-802 in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the GOM-802.
arning: Identifies conditions or practices that could
WARNING
CAUTION
result in injury or loss of life.
Caution: Identifies conditions or practices that could
result in damage to the GOM-802 or to other properties.
DANGER High Voltage
ttention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
4
Do not dispose electronic equipment as unsorted
municipal waste. Please use a separate collection facility
or contact the supplier from which this instrument was
purchased.
SAFETY INSTRUCTIONS
Safety Guidelines
General Guideline
CAUTION
Do not place any heavy objects on the GOM-802.
Avoid severe impact or rough handling that leads to
damaging the GOM-802.
Do not discharge static electricity to the GOM-802.
Use only mating connectors, not bare wires, for the
terminals.
Do not disassemble the GOM-802 unless you are
qualified as service personnel.
(Note) EN 61010-1:2001 specifies the measurement categories and
their requirements as follows. The GOM-802 falls under category
I.
Measurement category IV is for measurement performed at the source
of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Measurement category I is for measurements performed on circuits
not directly connected to Mains.
Power Supply
WARNING
Fuse
WARNING
AC Input voltage: 100V/ 120V/ 220V/230 V AC,
50/60Hz, 27VA, 22W
The power supply voltage should not fluctuate more
than 10%.
Connect the protective grounding conductor of the AC
power cord to an earth ground, to avoid electrical
shock.
Fuse type:
Line Voltage Rating Fuse
100V 90-110V T0.3A 250V
120V 108-132V T0.3A 250V
220V 198-242V T0.25A 250V
230V 216-250V T0.25A 250V
Make sure the correct type of fuse is installed before
power up.
To avoid fire, only replace the fuse with the specified
type and rating.
Disconnect the power cord before fuse replacement.
Make sure the cause of a fuse blowout is fixed before
fuse replacement.
5
GOM-802 User Manual
w
Cleaning the
GOM-802
Operation
Environment
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid into the
GOM-802.
Do not use chemicals or cleaners containing harsh
material such as benzene, toluene, xylene, and acetone.
Location: Indoor, no direct sunlight, dust free, almost
(Note) EN 61010-1:2001 specifies the pollution degrees and their
requirements as follows. The GOM-802 falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric
strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive pollution
occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution occurs.
Occasionally, however, a temporary conductivity caused by
condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry,
non-conductive pollution occurs which becomes conductive due to
condensation which is expected. In such conditions, equipment is
normally protected against exposure to direct sunlight, precipitation,
and full wind pressure, but neither temperature nor humidity is
controlled.
Storage
Environment
Disposal
Location: Indoor
Temperature: −10°C to 70°C
Do not dispose this instrument as unsorted municipal
aste. Please use a separate collection facility or contact
the supplier from which this instrument was purchased.
Please make sure discarded electrical waste is properly
recycled to reduce environmental impact.
6
SAFETY INSTRUCTIONS
Power cord for the United Kingdom
When using the GOM-802 in the United Kingdom, make sure the power cord
meets the following safety instructions.
NOTE: This lead / appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
Green/ Yellow: Earth
Blue: Neutral
Brown: Live (Phase)
As the colours of the wires in main leads may not correspond with the
coloured marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth
terminal marked with either the letter E, the earth symbol
or coloured
Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is
marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked
with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact
the supplier.
This cable/appliance should be protected by a suitably rated and approved
HBC mains fuse: refer to the rating information on the equipment and/or
2
user instructions for details. As a guide, a cable of 0.75mm
should be
protected by a 3A or 5A fuse. Larger conductors would normally require 13A
types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a live
socket is extremely hazardous. If a cable or plug is deemed hazardous, turn
off the mains power and remove the cable, any fuses and fuse assemblies. All
hazardous wiring must be immediately destroyed and replaced in accordance
to the above standard.
7
GOM-802 User Manual
GETTING STARTED
This chapter describes the GOM-802 in a nutshell,
including its main features as well as its front and rear
panels. After going through the panel overview, follow
the Power-up sequence before attempting to use the
instrument.
Please note the information in this manual was correct at
the time of printing. However as GW Instek continues
to improve its products, changes can occur at any time
without notice. Please see the GW Instek website for the
latest information and content.
GOM-802 is a high precision programmable DC
Milli-ohm meter suitable for low resistance
measurements of switches, relays, connectors, PCB
tracks and a variety of other devices. With the
easy-to-use features, superior performance, and
automatic test interfaces, the GOM-802 is a dependable
instrument for resistance measurements.
Easy to Use
Features
Performance
The GOM-802 includes an easy to use comparator
function (HI-LO-GO) that is able to easily set upper and
lower limits for pass/fail testing. The alarm buzzer can be
used with the comparator function. The flexible handler
interface can be used to monitor the status of the
pass/fail testing.
The relative feature enables the GOM-802 to easily
compensate for any stray resistance. Up to 20 different
sets of HI-LO-GO settings can be stored to satisfy a
number of different testing conditions. The GOM-802 is
also able to recall the last test setting that was used every
time it is turned on.
The GOM-802 has nine selectable measurement ranges
from 30mΩ to 3MΩ, a constant current source of 1uA
to 1A, an accuracy of 0.05%, a 1uΩ resolution and
performs measurements using four wire Kelvin
connections for accurate, consistent measurements.
The ability to choose between high measurement
accuracy with 7 samples/sec (full scale at 30000) or high
speed measurements with 30 samples/sec (full scale at
3000) allows the GOM-802 the flexibility to fulfill a
number of different measurement roles.
9
GOM-802 User Manual
Temperature
Compensation
Automatic
Tes tin g
Applications
Temperature Compensation (Optional):
The optional temperature probe (PT-100) can be used to
extrapolate the resistance of a DUT at a certain
temperature. When the temperature coefficient and the
required temperature (of the resistance measurement) is
keyed in under TC mode, the GOM-802 will display the
extrapolated measurement.
For automatic testing The GOM-802 has a handler
interface designed for automatic testing. The handler
interface outputs the status of PASS, FAIL, HI, LO,
READY and EOT signals and inputs a trigger control
signal. An RS-232 and GPIB option is also available for
computer control applications.
Production testing for contact resistance of switches,
relays, connectors, cables and printed circuit boards
and other low resistance devices.
Component testing of resistors, motors, fuses and
heating elements.
Incoming inspection and quality assurance testing.
Conductivity evaluation for product design.
10
GETTING STARTED
Key Features
30,000 counts.
Measurement Range: 30mΩ~3MΩ.
0.05% accuracy.
Hi/Lo comparator and limit percentage setting with
20 memory sets.
REL, Actual and % value measurements.
Manual or Auto-ranging.
Continuous or Triggered measurement modes.
Temperature compensation and measurement.
Four-wire measurement method.
Auto-recall last setting on power-up.
Diode test.
Alarm setting for PASS/FAIL test result.
Sampling rate: 7 or 30 sampling/sec.
Standard interface: Scan/Handler, optional interfaces:
RS-232 + GPIB.
11
GOM-802 User Manual
Front Panel Overview
Power Switch
Primary Value
Display
Normal Display
Comparator
Indicators
Turns On or Off the main
power. For details about the power up
sequence, see page 18.
Shows the primary measurement results.
Shows the Normal (nominal) value setting.
The comparator indicators indicate the
status of comparison judgments.
HI: measurement exceeds the HI limit
LO: measurement exceeds the LO limit
GO: measurement is within the HI and
LO limits.
12
GETTING STARTED
Function Control
Indicators
Measurement
Ter min als
Current Source
Ter min als
Negative
Ter min al
Indicates when a function key is active.
Sense HI and Sense LO terminals.
Current Source Terminals, Source HI
and Source LO.
Negative Terminal. This terminal has
the same potential as earth, but cannot
be substituted for it.
Function Keys
The COMP
TEMP
key activates the
comparator function. The secondary
function turns on the temperature
compensation function.
The Cursor keys are used to edit
parameters and navigate the menu tree.
The Up/Down arrow keys are used to
change the values and range of the
different parameters.
The secondary function recalls
previously saved settings.
Temperature
compensation
Recall settings
The AUTO/MAN
range between automatic and manual
SPEED
key toggles the
and toggles between 7 samples per
second and 30 samples per second.
13
GOM-802 User Manual
Sample speed
The REL%
a zero adjustment to the test leads or a
VA LU E
key is used to perform
DUT. Using the shift key will display
the measured values as a percentage of
the normal value.
Percentage
The LOCAL
milliohm meter between local and
GPIB/RS232
values
key will switch the
remote mode.
The secondary function will set the I/O
interface to GPIB or RS-232.
GPIB/
RS232 mode
Sets the High limit % values. The
secondary function activates the
Handler function.
HANDLER
SHIFT
Sets the normal (Nominal) value for the
comparator function.
HIGH
Handler mode
Scan function
Sets the lower % limit for the
comparator function and turns the
buzzer on for either a PASS or FAIL
judgment or turns the buzzer off.
Buzzer
ON/OFF
14
GETTING STARTED
Lower Limit
Display (LO)
Upper Limit
Display (HI)
Turns the external trigger on to allow
TRIG
the MANUAL TRIG
key to be used
as a manual trigger. The secondary
function will set the trigger to internal.
Turns the
internal trigger
ON/OFF
Enter key.
The SHIFT key is used to access the
secondary functions.
Displays the lower limit as as a
percentage.
Displays the upper limit as a percentage.
Interface
Indicator
The interface indicator shows the status
mode of the GOM-802. Remote mode
(RMT), Handler mode (HAND) or
Scan mode.
15
GOM-802 User Manual
A
A
A
Rear Panel Overview
AC Input
Fuse Socket
RS-232C port
GPIB port
ccepts the power cord. AC 100/120V
or 220/230V; 50/60Hz.
For the power on sequence, see page
18.
Holds the main fuse:
100/120V: T0.3A 250V;
220/230V: T0.25A 250V;
For the fuse replacement details, see
page 47.
ccepts an RS-232C cable for remote
control; DB-9 male connector.
For remote control details, see page 43.
ccepts a GPIB cable for remote
control (page 42).
Handler / Scan
port
Temperature
Sensor
16
The Handler/Scan port is used to
output pass/fail/high/low comparison
results.
The temperature sensor input for the
optional PT-100 temperature probe.
Set Up
Tilt Stand
GETTING STARTED
Steps
1. Pull handle base away from the casing.
2. Turn handle into any of the preset positions.
Stand Position Carry Position
17
Power Up
GOM-802 User Manual
Steps
CAUTION
1. Ensure that the correct voltage is lined up with the
arrow on the fuse cover on the rear panel.
100V 120V 220V 230V
100
120
230
220
120
220
100
230
220
230
120
100
230
100
220
120
2.Connect the power cord to the AC Voltage input.
0
2
2
230
120
1
0
0
Ensure the ground connector of the power cord is
connected to a safety ground. This will affect the
measurement accuracy.
3.Press the main power switch on the front panel.
R
E
W
O
P
4. The display will light up and show the last setting
used before the last shut down.
18
GETTING STARTED
Example:
Resistance
measurement
mode
4 Wire Kelvin Connection
Background
The GOM-802 uses 4 wire Kelvin connections for
accurate measurements.
Diagram
-
Description Source HI The Source HI carries the measuring
current source. It is connected to the +
side of the DUT.
Source LO The Source LO terminal accepts the
signal return current. Connects to the –
side of the DUT.
+
Sense HI Monitors the positive (+) potential.
Sense LO Monitors the negative (-) potential.
GND If the component has a large metallic
area not connected to either terminal,
connect it to ground to reduce noise.
19
GOM-802 User Manual
A
v
Zeroing (Relative Function)
Background
1. Short the
cables
1. Set the
Reference value
The Relative function is used to perform a zero
adjustment on the test leads.
fter the Relative value is pre-set, each measurement that
is displayed is equal to the actual value – relative preset
alue.
If the measured value minus the relative value is a
negative, a minus sign is shown on the MSB (most
significant bit). At low measurement speeds the minus
sign will alternate with the displayed measurement.
Short the test
cables together
using a short thick
copper wire if
necessary.
Press the REL
%VALUE
key.
2. Relative mode
display appears
REL Indicates the Relative function is active
The Auto-range function will be disabled when using
Buzzer Buzzer Function ..................................................... 30
Percentage Display as Percentage ............................................ 31
Relative Relative Function .................................................... 32
Temperature Temperature Measurement .................................... 32
Relative Function .................................................... 33
Temperature
Compensation
Temperature Compensation ................................... 34
Select the Resistance Range ................................... 35
Relative Function .................................................... 35
21
GOM-802 User Manual
A
Measurement Overview
Background
Measurement type
The Measurement chapter refers to the measurements
listed below. For measurements using the Handler or
Scan interfaces, see page 36.
TEMP
COMP
SHIFT+ COMP
TEMP
Comparison function
Temperature sensor
function
AUTO/MAN
SPEED
uto-range Manual
range
SHIFT+ AUTO/MAN
SPEED
Measurement speed
function
%/VALUE
REL
SHIFT+ REL
%/VALUE
Relative function selector
Toggle Relative % or Value
HIGH
SHIFT+ HIGH
NORMAL
SHIFT+ NORMAL
LOW
SHIFT+ LOW
MANUAL TRIG
SHIFT+ MANUAL TRIG
HANDLER
SCAN
HANDLER
SCAN
INT
High comparison limit
Handler function
Normal (nominal) value
Scan mode
Low comparison limit
Buzzer on/off
Trigger settings
INT
Internal trigger setting
22
MEASUREMENT
T
W
Resistance Measurement
1. Select the
Resistance
function.
Press SHIFT + COMP
access the main function mode.
Press the Up and Down arrow keys
to select resistance measurement
mode (shown above).
There are three different modes:
(resistance), (temperature)
and
(temperature compensation)
mode.
EMP
to
2. Resistance
mode display
appears.
Auto Indicates automatic ranging
3. Connect the
test lead and
measure
Press ENTER to confirm
mΩ
Ω
kΩ
Milliohms
Ohms
Kilo ohms
4-wire resistance:
Use the SOURCE HI and
the SOURCE LO port for
measurement, and the
SENSE HI, and SENSE LO
port for sensing.
hen switching from a high range to a low range, please
Note
allow a moment for the circuits to settle before
measuring.
23
GOM-802 User Manual
Select the Resistance Range
Background
The resistance range can be used with normal resistance
measurement as well as the temperature compensation
function.
Auto Range
Manual
Selection List
Press the Up and Down arrow keys
to manually select the resistance
range.
Press the AUTO/MAN key to use
automatic ranging.
Range Resolution(7 meas./s)
30mΩ1uΩ
300mΩ 10uΩ
3Ω 100uΩ
30Ω 1mΩ
300Ω 10mΩ
3kΩ 100mΩ
30kΩ 1Ω
300kΩ 10Ω
3MΩ 100Ω
Note
For detailed specifications, please see the specifications
on page 51.
24
MEASUREMENT
Select Measurement Rate
Background
1. Select Rate
LO rate
The resistance measurement speed has 2 ranges: low and
high. Low speed is the most accurate with 7
measurements/second and a full scale of 30,000 counts.
High speed has 30 measurements/second with a full scale
3,000 counts.
The rate function is applicable to the resistance,
temperature and temperature compensation
measurements.
Press the SHIFT +
AUTO/MAN
SPEED
keys to switch
between the LO and HI rates.
The low rate is shown with a full scale of 5 digits
in the resistance measurement mode.
HI rate The high rate is shown with a full scale of 4 digits.
For detailed specifications, please see the specifications
Note
on page 49.
Using the Trigger Function
Background
1. Select Manual
Trigger
The GOM-802 can use internal or manual triggering for
the resistance, temperature and temperature
compensation measurement modes.
External triggering is only supported with the Handler
and Scan modes.
By default the GOM-802 is in internal triggering mode.
Press MANUAL TRIG
to manual triggering mode.
INT
to switch
25
GOM-802 User Manual
A
The EXT indicator will be lit when the manual
trigger is active.
2. Manually
Triggering
Measurements
The EXT indicator will be lit when the manual
Press the MANUAL TRIG
each time you want to start a single
measurement.
trigger is active.
INT
key
Internal Triggering
Press SHIFT + MANUAL TRIG
to return the triggering mode back
INT
to internal mode.
The EXT indicator will become cleared from the
display.
Compare Function
Background
26
The compare function compares a measured value to a
“Normal” value that has an upper (HI) and lower (LO)
limit. The upper and lower limit is set as a percentage of
the Normal value.
measured value that falls within the upper and lower
limits is considered a GO (pass), a value that falls below
the lower limits is considered LO, and a value that falls
over the upper limit is a HI.
MEASUREMENT
T
1. Select the
compare function
The GO, HI or LO indicators will light up for each
measured value that is compared to the upper and lower
limits.
2. Normal value
setting
Press COMP
compare mode, as shown above.
Press NORMAL
NORMAL value setting.
Use the Left and Right arrow keys to
select a digit. The selected digit will
EMP
to access the
SCAN
to set
flash.
Use the Up and Down arrow keys to
edit the value of the selected digit.
Range 33000~00000
Note
3. HI limit
setting
Press ENTER to confirm the setting.
After setting the Normal value, the HI and LO limits will
be changed to reflect the new Normal value setting.
Press HIGH
percentage limit.
SCAN
to set the HI
Use the Left and Right arrow keys to
select a digit. The selected digit will
flash.
27
GOM-802 User Manual
L
Range 000~999
4. LO limit
setting
Use the Up and Down arrow keys to
edit the value of the selected digit.
Press ENTER to confirm the
setting.
Press LOW
percentage limit.
to set the LO
Use the Left and Right arrow keys to
select a digit. The selected digit will
flash.
Use the Up and Down arrow keys to
edit the value of the selected digit.
Range 000~999
Press ENTER to confirm the
setting.
Save Compare Settings
Background
1. Select Save
mode
The save function saves the Normal, High and Low limit
settings in memory. If these values are not saved, the
settings will be lost after exiting Compare mode.
Up to 20 Normal, High, Low settings can be saved on
the GOM-802.
Press SHIFT +
RECAL
to access
the RECALL/SAVE mode.
Press the Up and Down arrow key
to show “SAVE” in the main display,
as shown above.
28
MEASUREMENT
A
v
L
2. Select memory
location
Press enter to confirm.
Use the Left and Right arrow keys to
select a digit. The selected digit will
flash.
Use the Up and Down arrow keys to
edit the value of the selected digit.
Range 00~19
Press ENTER to confirm the
setting.
fter saving the settings, the meter will return to the
compare mode function.
Pressing SHIFT before pressing ENTER will exit the Save
Note
mode.
Recall Compare Settings
Background
1. Select Recall
mode
The Recall function retrieves the Normal, High and Low
limit settings from one of 20 memory locations. If these
alues are not saved, the settings will be lost after exiting
Compare mode.
Up to 20 Normal, High, Low settings can be saved on
the GOM-802.
Press SHIFT +
the RECALL/SAVE mode.
Press the Up and Down arrow key
to show “CALL” in the main display,
as shown above.
RECAL
to access
Press ENTER to confirm.
29
GOM-802 User Manual
A
2. Select memory
location
Use the Left and Right arrow keys to
select a digit. The selected digit will
flash.
Use the Up and Down arrow keys to
edit the value of the selected digit.
Range 00~19
Press ENTER to confirm the
setting.
fter saving the settings, the meter will return to the
compare mode function.
Pressing SHIFT before pressing ENTER will exit the Recall
Note
mode.
Buzzer Function
Background
1. Select Buzzer
Function
2. Buzzer
function display
appears
NON Buzzer function off
bF Buzz on Fail (No Go)
bP Buzz on Pass (Go)
The buzzer function can make the buzzer sound when
the compare test has passed (GO) or failed (NOGO).
The buzzer settings are only applicable in Compare
mode.
Press the SHIFT key and then press
LOWkey.
Press the Up and Down arrow key
to choose the buzzer mode.
30
Press ENTER to confirm the
selection.
MEASUREMENT
Pressing SHIFT before pressing ENTER will exit the
Note
Buzzer function settings.
Display as Percentage
Background
1. Select Buzzer
Function
2. Percentage
display appears
3. Return to
normal display
Note
In the compare function, the measured value can be
displayed as a percentage of the Normal value rather than
the actual value.
For example, if the Normal value is 20.000, and the
actual value is 10.000, then as a percentage it would be
displayed as 50.00%.
Press the SHIFT key and then press
the REL
%VALUE
.
SHIFT
Press the SHIFT key and then press
the REL
The Auto-range function will be disabled when using the
percentage display.
%VALUE
.
31
GOM-802 User Manual
A
T
Relative Function
Background
The Relative function is used to perform a zero
adjustment on the test leads or the resistance of the
DUT.
fter the Reference value is pre-set, each measurement
that is displayed is equal to the actual value – relative
preset value.
If the measured value minus the relative value is a
negative, a minus sign is shown on the MSB (most
significant bit). At low measurement speeds the minus
sign will alternate with the displayed measurement.
1. Set the
Reference value
2. Relative mode
display appears
REL Indicates the Relative function is active
Press the REL
%VALUE
key.
The Auto-range function will be disabled when using
Note
compare mode with the relative function.
Temperature Measurement
Background
1. Select the
Temperature
function.
The temperature measurement function uses the optional
PT-100 temperature probe. The measured temperature is
displayed on the NORMAL display. For more
information on the optional PT-100 sensor, see the
appendix on page 48.
There is only one range in when using the temperature
function.
Press SHIFT + COMP
EMP
to
access the main function mode.
32
MEASUREMENT
Press the Up and Down arrow keys
to select the ˚C measurement mode
(shown above).
There are three different modes:
(resistance), (temperature)
and
(temperature compensation)
mode.
2. Temperature
mode display
appears.
Press ENTER to confirm
The temperature is displayed on the NORMAL display.
3. Temperature
mode
connection.
The temperature sensor uses the rear panel TC port for
input.
Relative Function
Background
The Relative function can also be used with the
temperature function. The temperature displayed is the
temperature of the sensor minus the temperature of the
sensor when the REL
%VALUE
key was pressed.
1. Set the
Reference value
2. Relative mode
display appears
REL Indicates the Relative function is active
Press the REL
%VALUE
key.
33
GOM-802 User Manual
T
Temperature Compensation
Background
1. Select the
Temperature
compensation
mode.
If the resistance of a DUT at a particular temperature is
needed the compensation function can be used. This
function can simulate the resistance of a DUT at a
desired temperature. If the ambient temperature and the
temperature coefficient of the DUT are known, it is
possible to determine the resistance of a DUT at any
temperature.
Press SHIFT + COMP
access the main function mode.
EMP
to
Press the Up and Down arrow keys
to select the TC measurement mode
(shown above).
There are three different modes:
(resistance), (temperature)
and
(temperature compensation)
mode.
2. Temperature
compensation
mode appears.
Press ENTER to confirm.
Edit the desired temperature and the
temperature coefficient of the DUT.
Use the Left and Right arrow keys to
select a digit and use the Up and
Down arrow keys to edit the digit.
Desired
Temperature
Temperature
Coefficient
Press ENTER to confirm the
0~99.9
0000~9999
selection and start measurement.
34
MEASUREMENT
3. Temperature
compensation
connection.
Sensor Connection:
DUT connection
4 Wire:
Select the Resistance Range
Background
The resistance range can be selected when using the
temperature compensation function.
Auto Range
Press the Up and Down arrow keys
to manually select the resistance
range
Manual
Press the AUTO/MAN key to use
automatic ranging.
Relative Function
Background
1. Set the
Reference value
2. Relative mode
display appears
The Relative function can also be used with the
temperature compensation function.
Interface Function Check ....................................... 44
36
HANDLER/SCAN/ INTERFACE
Handler Overview
Background
The Handler Interface is used to help bin components
based on the Go-NoGo comparator function test. There
are 6 TTL outputs and one TTL input. The Handler
interface can only be used with the resistance
measurement and compare measurement modes.
Interface and pin
assignment
Pin Assignment START
9 Pin D-SUB
(Female)
Start the trigger for a single
measurement. Negative edge trigger.
READY
High when the measurement has
finished. The instrument is ready for
the next trigger..
EOT
High when the AD conversion has
completed. The DUT is ready to be
changed.
LOW
High when the compare result is
deemed LO.
HIGH
High when the compare result is
deemed HI.
FAIL
High when the compare result is either
HI or LO (fail).
PASS
High when the compare result is GO
(pass).
The output current from all the terminals and the +5V
Note
terminal cannot exceed 60mA.
37
GOM-802 User Manual
A
T
Scan Overview
Background
The Scan function is used to automatically bin groups of
up to 100 components. The scan function, unlike the
handler interface must first be activated using the panel
keys.
There are a total of 6 outputs as well as a GND and
power (+5V) pin.
Interface and pin
assignment
Pin Assignment Relay
Pass
9 Pin D-SHELL
(Female)
Controls the relay output.
Pass signal. Indicates a GO compare
result is GO (pass).
Low
Low signal. Indicates a LO compare
result.
High
High signal. Indicates a HI compare
result.
Clock
The clock signal will pulse high when
each group of output signals (Ready,
Pass, Low, High) are ready. There are up
to 100 groups of output signals
STRB
fter all (100) output groups are
ready, the STRB signal will pulse high.
Scan Setup
Background
1. Select the
compare function
The scan function is accessed via the compare mode.
Press COMP
EMP
to access the
compare mode, as shown above.
page 26
38
HANDLER/SCAN/ INTERFACE
T
2. Select the
resistance function.
3. Resistance
mode and
Compare display
appears.
4. Resistance
Range
Press SHIFT + COMP
EMP
to
access the main function mode.
Press the Up and Down arrow keys
to select the resistance measurement
mode (shown above).
There are three different modes:
(resistance), (temperature)
and
(temperature compensation)
mode.
Press ENTER to confirm.
Press the Up and Down arrow keys
to manually select the resistance
range.
5. Select the
Scan mode.
Press SHIFT + NORMAL
SCAN
to
access the scan mode.
Press the Up and Down or Left and
right arrow keys to select the
number of counts.
Count range 0~100
6. Select the
Delay time.
Press ENTER to confirm.
Press the Up and Down or Left and
Right arrow keys to set the sweep
delay time. Each count unit has a
delay time of 16.2 ms.
39
GOM-802 User Manual
Count units 30~30,000
7. The Ready
Display appears
Note
8. Start the scan.
Press ENTER to confirm.
After the Ready indicator is shown it is not
possible to make any changes to the compare
function or range.
Press the MANUAL TRIG
start the sweep measurements.
INT
key to
The results will be displayed on
screen and output through the scan
port until finished.
9. View Results
10. Exit Scan
mode
Press the Up and Down arrow keys
to cycle through each result. The
measurement number will be
displayed on the Normal display and
the measurement on the main
screen.
Press SHIFT + NORMAL
SCAN
at
any time to exit the scan mode.
40
HANDLER/SCAN/ INTERFACE
A
Scan Output
Background
The timing diagrams for the scan output under different
conditions are shown below.
Ready message displayed…
Scan channel 1. Delay time has
elapsed.
fter the MANUAL TRIG
INT
key is
pressed…
Relay
Pass
Low
High
Clock
STRB
1299100
Scan channel n, delay time has elapsed.
Scan Channel 100. Delay time has
Scan output signal timing.
elapsed.
41
GOM-802 User Manual
Configure Interface
Overview
The RS-232 and GPIB interfaces are factory installed
options for remote control of the GOM-802. The
remote control interfaces allow the GOM-802 to be
programmed for automatic testing.
For more information on the remote control
programming, please see the GOM-802 programming
manual.
Interface
GPIB
RS-232
Configure GPIB Interface
Background
1. Select the
GPIB address.
The GPIB interface is SCPI-1994, IEEE488.1 and
IEEE488.2 compliant.
24 pin female GPIB port
DB-9 male port
Press SHIFT + LOCAL
GPIB/RS232
to
access the GPIB/RS232 settings.
Press the Left or Right arrow key to
enter the GPIB address settings.
Press the Up and Down arrow keys
to set the GPIB address.
GPIB Address 0~31
Press ENTER to confirm. The
display returns to the main display.
Connect one end of the GPIB cable
to the computer and the other end
to the GPIB port on the GOM-802.
or
42
HANDLER/SCAN/ INTERFACE
GPIB Constrains
Maximum 15 devices altogether, 20m cable length,
2m between each device
Unique address assigned to each device
At least 2/3 of the devices turned on
No loop or parallel connection
Configure RS-232 Interface
Background
Settings Baud rate 1200, 2400, 4800, 9600
Parity None
Data bits 8
The GOM-802 uses an RS-232C connection for remote
control. When connecting to a PC ensure the correct
baud rate, parity, data bits, stop bit and data control
settings are used.
Stop bit 1
Data flow
None
control
1. Select the
RS-232 baud rate
Press SHIFT + LOCAL
GPIB/RS232
to
access the GPIB/RS232 settings.
Press the Up and Down arrow keys
to set the baud rate.
Baud rate 1200, 2400, 4800, 9600
Press ENTER to confirm. The
display goes back to the main
display.
Connect the RS-232C cable to rear
panel RS232 port.
RS-232C pin
assignment
Pin 2: RxD
Pin 3: TxD
9 8 7 6
Pin 5: GND
Pin 1, 4, 6 ~ 9: No Connection
5 4 3 2 1
43
GOM-802 User Manual
PC – GOM
RS-232C
Connection
The RS232 connection uses a Null-modem connection,
in which transmit (TxD) and receive (RxD) lines are
cross-linked.
Interface Function Check
Operation
Run this query from the terminal.
*idn?
This should return the Manufacturer, Model number, and
Firmware version.
GW. Inc, GOM-802 , FW1.00
Function Check
Errors
If you do not receive a proper response from the Meter,
please check if the power is on, the GPIB address/RS232
baud rate is correct, and all cable connections are active.
44
FAQ
FAQ
● What are the different measurement speeds?
● The GOM-802 performance does not match the specifications.
What are the different measurement speeds?
There are two measurement speeds for both resistance and temperature
measurement. At 30,000 counts the measurement speed is 7 readings/s, at
3,000 counts the measurement speed is 30 readings/s.
The GOM-802 performance does not match the specifications.
Make sure the device is powered on for at least 30 minutes, within
+18°C~+28°C with a humidity not exceeding 80%. This is necessary to
stabilize the unit to match the specifications.
If there is still a problem, please contact your local dealer or GWInstek at
General .................................................................. 53
46
APPENDIX
Fuse Replacement
Replace the AC source fuse
Steps
1.Remove the power cord.
0
2
2
230
120
1
0
0
2.Remove the fuse socket using a flat screwdriver.
0
2
2
230
120
1
0
0
3.Replace the fuse in the holder.
0
2
2
230
120
1
0
0
47
GOM-802 User Manual
˚K
˚
4. Ensure the correct line voltage is lined up with the
arrow on the fuse holder. Insert the fuse socket.
0
Rating
2
2
230
120
1
0
0
100V/120V TT0.3A 250V
0
2
2
230
120
1
0
0
220V/230V T0.25A 250V
Temperature Measurement
Reference Temperature Table
Overview
Background
The International Temperature Scale (ITS) is based on the
following table. The table has 17fixed calibration points as
of 1990.
Temperature
Element Type
(H2) Hydrogen Triple point 13.8033 -259.3467
(Ne) Neon Triple point 24.5561
(O2) Oxygen Triple point 54.3584
(Ar) Argon Triple point 83.8058 -189.3442
(Hg) Mercury Triple point 234.325 -38.8344
(H2O) Water Triple point 273.16 +0.01
(Ga) Gallium Melting point 302.9146 29.7646
(In) Indium Freezing point 429.7485 156.5985
(Sn) Tin Freezing point 505.078 231.928
(Zn) Zinc Freezing point 692.677 419.527
(Al) Aluminum Freezing point 933.473 660.323
(Ag) Silver Freezing point 1234.93 961.78
(Au) Gold Freezing point 1337.33 1064.18
C
48.5939
18.7916
48
APPENDIX
y
Hig
g
y
W
RTD Sensors
Overview
Resistive Thermal Devices (RTDs) are commonly used as
temperature sensors. RTDs change resistance linearly over
a specific range of temperature. The table below shows
some of the inherent features of RTDs compared to
thermocouples.
Feature
Accurac
Resolution
Speed of response
Self-heatin
Long term stabilit
Output characteristics Approx. 0.4ohm/˚C, near
Optional Platinum Sensor
Introduction
The optional platinum sensor is a PT-100 sensor.
The PT-100 sensor meets the German DIN43760:
1968 3 wire measurement specification.
Description
her accuracy
0.1~1.0˚C, higher resolution
Slower
Yes
Good
linear
These sensors are one of the most common
temperature sensors used in industry. These sensors
have a nominal resistance of 100
Ω at 0˚C.
The relationship between temperature and resistance
for the PT-100 sensor can be described with the
Gallendarvan Dusen equation shown below:
RRTD=R0[1+AT+BT2+CT3(T-100)]
here: RRTD is the calculated resistance of the
RTD.R0 is the known RTD resistance at 0˚C.
T is the temperature in˚ C
A=alpha [I+(delta/100)]
B=-I(alpha)(delta)(Ie-4)
C=-I(alpha)(beta)(Ie-8)
The Alpha (A), Beta (B), Delta (D) values for the
49
GOM-802 User Manual
A
PT-100 sensor are listed below:
Type
PT-100 ITS90 0.003850 0.10863 1.49990
Standard Alpha
Beta
Delta
Ω @ 0˚C
100Ω
Temperature
Calculation Example
Example—Calculating the resistance of a PT-100
RTD at 100˚C (T). The following R
(Ωat 0˚C),
0
alpha, beta, and delta values are used for the
PT-100 RTD:
*When the instrument is set to 30mΩ or 300mΩ ranges, the resistance value
will be cha
due to the different temperature between internal and external parts of the
instrument. Therefore, please must wait 1 minute in order to obtain an
accurate value after the test leads have been connected or disconnected.
* When Ke
please wa
** In high speed mode, first ensure that the instrument has been grounded
through the power cord.
nged while connecting or disconnecting the test lead to the panel
lvin clips are used to resume
it for a short time to stabilize the measurement.
testing after a long period of time,
51
GOM-802 User Manual
Measurement Four-terminal method.
Auto-ranging Provided.
Over input range “OL” indication
Maximum Applied
voltage
30m~3Ω range: 30VpDC
Other range: 100VpDC
Comparator 20 sets of comparator status can be selected.
Buzzer mode
Temperature
correction range
Reference temperature
range
Thermal coefficient
range
Temperature range Accuracy of temperature compensation for 3930
0˚C~39.9˚C
40˚C~100˚C
*The temperature coefficient for the other settings must be calculated
individually according to different conditions.
*If the temperature coefficient or the difference between the environmental
temperature and the required temperature exceeds normal operation, after
calculating the compensation, the variation to the reading value will be
significant.
0.0˚C~39.9˚C
0˚C~99.9˚C
±9999 ppm
ppm/Cu wire, speed: 7 readings/second.*
0.3%+resistance measurement accuracy.
0.6%+resistance measurement accuracy.
52
APPENDIX
Interface
Handler interface* Signal: START TTL input
Signal: LOW, HIGH, FAIL, PASS, EOT, READY total 6 TTL
outputs.
(This function is valid only under the resistance
measurement mode and the compare mode is enabled.)
Scan* Signal: READY, PASS, LOW, HIGH, CLOCK, STRB total 6
TTL outputs.
(This function is valid only under the resistance
measurement mode and the compare mode is enabled.)
RS-232+ GPIB (option) IEEE488.1-1987, IEEE488.2-1992 and SCPI-1994
* The Scan and Handler interface use the same
connector
Environmental
Operation
Environment
Storage temperature -10˚C to 70˚C.
Indoor use, altitude up to 2000m.
Ambient Temperature 0˚C to 40˚C.
Relative Humidity 80% (Maximum).
Installation category II
Pollution Degree 2
General
Power source
Accessories Test Lead × 1, User manual × 1,
Dimension 251(W)×91(H)×291(D) mm
Weigh Approx. 3 kg
AC 100V/120V/220V/230V±10%, 50/60Hz, 27VA, 22W.
Programmable manual × 1 (option),
Temperature sensor (option) × 1
53
GOM-802 User Manual
Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
(1) No.7-1, Jhongsing Rd., Tucheng Dist., New Taipei City, Taiwan
(2) No. 69, Lu San Road, Suzhou City (Xin Qu), Jiangsu Sheng, China
declare, that the below mentioned product
Type of Product: DC Milliohm Meter
Model Number: GOM-802
is herewith confirmed to comply with the requirements set out in the Council
Directive on the Approximation of the Law of Member States relating to
Electromagnetic Compatibility (89/336/EEC, 92/31/EEC, 93/68/EEC).
For the evaluation regarding the Electromagnetic Compatibility and Low
Voltage Equipment Directive, the following standards were applied:
◎ EMC
EN 61326: Electrical equipment for measurement, control and laboratory
use –– EMC requirements (1997+A1: 1998+A2: 2001)
Conducted and Radiated Emission
EN 55011: 1998 class A
Current Harmonic
IEC 61000-3-2: 2000
Voltage Fluctuation
IEC 61000-3-3: 1995
-------------------------
-------------------------
-------------------------
-------------------------
◎ Safety
Electrostatic Discharge
IEC 61000-4-2: 1995+A1: 1998
Radiated Immunity
IEC 61000-4-3: 1996+A1: 1998
Electrical Fast Transients
IEC 61000-4-4: 1995
Surge Immunity
IEC 61000-4-5: 1995
Conducted Susceptibility
IEC 61000-4-6: 1996
Power Frequency Magnetic Field
IEC 61000-4-8: 1993
Voltage Dips/ Interrupts
IEC 61000-4-11: 1994
Low Voltage Equipment Directive 73/23/EEC & amended by 93/68/EEC
IEC / EN 61010-1: 2001