Test Equipment Depot BST-CM273 Instruction Manual

1000A
CAT.
600V
AC V
OFF
AC A
DC V
DC A
COM
V
AC
000
V
MAX
600V
HOLD
MAX
CAT.
AC/DC CLAMP METER
99 Washington Street Melrose, MA 02176
Phone 781-665-1400
Visit us at www.TestEquipmentDepot.com
Toll Free 1-800-517-8431
BST-CM273
AC/DC CLAMP METER
INSTRUCTION MANUAL
INDEX
1. INTRODUCTION................................
Page
1
2. SAFETY NOTES................................
3. FEATURES.........................................
4. SPECIFICATIONS..............................
5. INSTRUMENT LAYOUT.....................
6. MEASUREMENT................................
7. MAINTENANCE.................................
2
3
4-6
7-8
9-11
12
1. INTRODUCTION
NOTE
This meter has been designed and tested according to CE Safety Requirements for Electronic Measuring Apparatus, EN61010-1, EN 61010-2-32 and other safety standards. Follow all warnings to ensure safe operation.
WARNING
READ "SAFETY NOTES" (NEXT PAGE) BEFORE USING THE METER.
-1-
2. SAFETY NOTES
Read the following safety information carefully before attempting to operate or service the meter.
● Use the meter only as specied in this manual,
otherwise the protection provided by the meter may be
impaired.
● Always keep hands behind the meter barrier.
● Use extreme caution when clamping around uninstalled
conductors or bus bars.
● Never clamp around any conductor carrying a voltage
above 600V R.M.S.
● Rated environmental conditions:
1. Indoor use.
2. Installation Category III.
3. Pollution degree II.
4. Altitude up to 2000 Meter.
5. Relative Humidity 80% Max.
6. Ambient Temperature 0~40°C.
● Observe the international Electrical Symbols listed
below:
Meter protected throughout by double insulation or reinforced insulation.
Warning ! Risk of electric shock. Caution ! Refer to this manual before using the
meter. Alternating current.
Earth (ground) terminal.
-2-
3. FEATURES
● Full Autorange for all functions.
● Data Hold function freezes the reading.
● Max Hold function holds the absolutely maximum of
readings.
● LCD backlight function.
● Flashlight function.
● Low battery indication.
● Safety design throughout with no exposed metal parts,
shielded banana plugs and recessed input terminals.
● The Ohm function is ideal for checking continuity of
relays, transformers and motor coils.
● Diode check function.
-3-
4. SPECIFICATIONS
Ranges:
AC Voltage: 400/600 V AC Current: 400/1000 A DC Voltage: 400/600 V DC Current: 400/1000 A
Resistance: 400/2000 Ω
Diode : 400/2000 mV at 1mA
ACV
Range Resolution Accuracy 400V 0.1V 600V 1V
ACA
Range Resolution Accuracy 400A 0.1A 1000A 1A
*Frequency Response 40Hz~500Hz
DCV
Range Resolution Accuracy 400V 0.1V 600V 1V
±(1.0%rdg+3dgt)
±(1.5%rdg+3dgt)
±(0.75%rdg+3dgt)
-4-
DCA
Range Resolution Accuracy 400A 0.1A 1000A 1A
Resistance
Range Resolution Accuracy
400Ω 0.1Ω 2000Ω
*Buzzer sounds below 38.0Ω
Overload protection
ACV 750V rms. DCV 1000V. Diode & Ohm 600V rms.
● Conductor Size
Approx. 40mm Max.
● Operating Principle: Dual slope integration.
● Over Rage Indication: "O.L" indicated.
● Low Battery Indication: " " sign appears on the
display.
● Response Time:
Approx.1 second.
±(1.5%rdg+3dgt)
±(1.0%rdg+3dgt)
-5-
● Sample Rate:
Approx.2 times per second.
● Temperature & Humidity for Guaranteed
0°C~40°C at 80% max. relative humidity.
● Storage Temperature & Humidity:
-10°C~50°C at 80% max. relative humidity.
● Dimensions
255(L) x 80(W) x 35(D)mm.
● Weight
Approx. 420g (battery included).
● Power Source
9V (6F22) x 1.
● Accessories:
Test leads. Soft pouch. Instruction manual. 9V (6F22) x 1.
-6-
5. INSTRUMENT LAYOUT
(1)
(2)
1000A
CAT.Ⅲ 600V
OFF
AC V
(3)
(4)
DC V
AC A
DC A
A ZERO
AC/DC CLAMP METER
MAX
(8)
(7)
HOLD
(9)
(5)
(6)
AC
COM
000
MAX
600V
-7-
V
V
CAT.
(1) Transformer Jaws
Pick up the conductor within the jaws center.
(2) Barrier
Provide a protective distance from hands to conductor.
(3) Jaw Trigger
Press to open the jaws.
(4) DCA zero adjustment shaft
(5) LCD Display
3¾ digit LCD with the maximum reading of 3999.
(6) Input Terminals
(7) Function Switch
(8) Max Button
Hold the absolute maximum of readings with " " indicated.
(9) Data Hold Button
Freeze the reading for all ranges with " "
HOLD
indicated.
MAX
-8-
6. MEASUREMENT
Before proceeding with measurement, read the safety notes. (1) AC Voltage measurement
● Insert the BLACK test lead to COM and the RED
one to the other terminal.
● Set the function switch to ACV.
● Connect the test leads to the object to be
measured.when measuring AC voltage, the polarity of the leads can be ignored.
● Read the display.
(2) DC Voltage measurement
● Insert the BLACK test lead to COM and the RED
one to the other terminal.
● Set the function switch to DCV.
● Connect the test leads to the object to be measured.
● Read the display.
※ Reversing the polarity of the test leads displays a negative value.
-9-
+
-
(3) AC Current measurement
● Set the function switch to ACA.
● Press the jaw trigger to open the transfomer jaw
and clamp onto one conductor only.
● Read the display.
LOADLOAD
1000A
MAX
CAT.Ⅲ 600V
OFF
AC V
1000A
MAX
CAT.Ⅲ 600V
OFF
AC V
DC V
AC A
DC A
A ZERO
HOLD
AC/DC CLAMP METER
V
AC
000
V
MAX
COM
600V
CAT.
O X
DC V
AC A
DC A
A ZERO
HOLD
AC/DC CLAMP METER
V
AC
000
V
MAX
COM
600V
CAT.
(4) DC Current measurement
● Set the function switch to DCA.
※ If the inital reading of DCA is not zero, adjust
the "DCA zero adjustment shaft" until the display shows zero.
● Press the jaw trigger to open the transfomer jaw
and clamp onto one conductor only.
● Read the display.
-10-
LOADLOAD
(5) Resistance & Diode measurement CAUTION! Make sure that there is no voltage in the circuit or
components being measured.
For resistance test:
● Insert the BLACK test lead to COM and the RED
one to the other terminal.
● Set the function switch to Ω.
● Connect the test leads to the object to be measured.
● Read the display.
For Diode test:
● Connect the test leads to the diode for testing.
● Read the forward Voltage of diode directly from the
display. ※ Reversing the polarity of the test leads displays
O.L.
-11-
7. MAINTENANCE
Battery Replacement:
When the low battery warning sign appears, replace with a new battery as follows: (1) Disconnect the test leads from the clamp meter and
turn off the power.
(2) Unscrew the battery cover and replace with a new
Battery.
Cleaning and Storage:
WARNING
To avoid electrical shock or damage to the meter, do not get water inside the case.
Periodically wipe the case with a damp cloth and detergent. Do not use abrasives or solvents.
If the meter is not to be used for a long time over 60 days, please remove the battery for storage.
-12-
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com