Testec TT-SI 9101 User guide

Page 1
INSTRUCTION MANUAL
TT-SI 9101
100MHz Active Differential Probe
Page 2
These probe is in compliance with IEC-61010-031 CAT III, Pollution Degree 2
Safety Symbols
1. Safety Terms and Symbols
Terms appear in this manual:
___________________________________________________
WARNING. Warning statements identify conditions or practice that
could result in injury or loss life.
___________________________________________________
CAUTION. Caution statements identify conditions or practice that
could result in damage to this product or other property.
Connect it to safety earth ground using the wire recommended in the user’s manual.
High voltage danger
The symbol on an instrument indicates that the user should refer to the operating instructions located in the manual.
2. General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to this probe or any products that connected to it.
Observe Maximum Working Voltage
To avoid any injury, do not use the probe under the condition that the voltage between either input lead or earth is above 1000Vrms CAT III. This voltage rating applies to both settings 1/10 & 1/100.
Page 3
Must be Grounded
This probe is grounded with the shell of BNC connector and an auxiliary grounding terminal, through the grounding conductor of the power cord of the measurement instrument. Before making connections to the input leads of this probe, ensure that the output BNC connector is attached to the BNC connector of the measurement instrument and the auxiliary grounding terminal is connected to a proper ground, while the measurement instrument is properly grounded.
Use Fused Test Prods if Necessary
If this probe is intended to use for measurements in circuits of MEAUSREMENT CATEGORY III, it should incorporate with fused test prods.
Do Not Operate Without Covers
To avoid electric shock or fire hazard, do not operate this probe with covers removed.
Do Not Operate in Wet/Damp Conditions
To avoid electric shock, do not operate this probe in wet of damp conditions.
Do Not Operate in Explosive Atmosphere
To avoid injury or fire hazard, do not operate this probe in an explosive atmosphere.
Avoid Exposed Circuit
To avoid injury, remove jewelry such as rings, watches, and other metallic objects. Do not touch exposed connections and components when power is present.
Use Proper Power Source
To ensure this probe function well, use four AA cells or 6VDC/60mA or regulated 9VDC/40mA mains adaptor or power lead. Do not operate this probe from a power source that applies more than the voltage specified.
Do Not Operated With Suspected Failures
If you suspect there is damage to this probe, have it inspected by qualified service personnel.
3. Description
By enabling conventional oscilloscopes to display and measure in-circuit waveforms that are referenced to high common mode voltages. The differential probe extends the measurement capability of oscilloscopes in electronic power converters, inverters, motor speed controls, switch mode power supplies, and many applications.
Page 4
4. Installation
a. Simply plug-in the BNC output connector to the vertical input of a general purposed oscilloscope or other measurement instrument, and connects the auxiliary grounding terminal to a proper ground. The measurement instrument must have a ground referenced. b. Connect an appropriate power source to this probe and or enter the batteries, then turn it on. c. Select the proper attenuation ratio. When measuring signals below 70V, switch the attenuation ratio to 1/10 in order to get higher resolution and less noise ratio. Otherwise, set the attenuation ratio to 1/100.
WARNING. To protect against electric shock, use only the
accessories supplied with this probe.
d. Using the appropriate probe accessories, connect the inputs to the circuits under
measurement.
CAUTION. This probe is to carry out differential measurement between two points on the circuit under measurement. This probe is not for electrically insulating the circuit under measurement and the measuring instrument.
5. Appearance
The differential probe looks as follows.
a. Output Cable The BNC output connector and an auxiliary grounding terminal
are connected to the oscilloscope.
b. Input Leads The input leads of the differential probe connect to sprung hooks
that come with the probe.
c. Sprung Hooks The sprung hooks are connected safely to test points in circuits under measurement.
Page 5
6. Available Power Sources
a. 4 x AA batteries b. Mains adaptor (6VDC/60mA or regulated 9VDC/40mA), c. Lemo® Power Cord, for oscilloscopes with power output - Lemo® connector. d. Probus® Power Cord, for oscilloscopes with power output - Probus
®
connector.
e. USB Power Cord, for oscilloscopes which offer USB connector.
7. Accessories supplied
Type Order-No. Description TT-SI G1 15103
Grabber red and black TT-SI NT 15100 Mains Adapter TT-SI HC 15160
Hardcase
8. Optional Accessories
Type Order-No. Description TT-SI PROBUS 15150 TT-SI LEMO 15151
Power Lead with PROBUS-Connector
Power Lead with LEMO-Connector TT-SI USB 15152 TT-SI EPL1 15140 TT-SI EPL2 15141 TT-SI G2 15104
Power Lead with USB-Connector
1 to 3 Power Splitter
1 to 4 Power Splitter
Grabber red and black
Lemo® and Probus® are the registered trademarks
Page 6
Length of BNC Cable
put Leads
9. Specifications
TT-SI 9101
Bandwidth DC to 100MHz (-3dB) Attenuation Ratio 1:10 / 1:100 Accuracy ±2% Rise Time 3,5ns Input Impedance Input Voltage
- Differential Range Input Voltage
- Common Mode Range Input Voltage
- Absolute Max. (Differential or Common)
Measurement Category CAT III Output Voltage
- Swing Output Voltage
- Offset (typical) Output Voltage
- Noise (typical) Source Impedance
(typical) CMRR (typical) -85dB @50Hz, -55dB @1MHz Ambient Operating
Temperature Ambient Storage
Temperature Ambient Operating Humidity Ambient Storage Humidity Power Requirements
- Standard Power Requirements
- Optional
4MΩ // 7pF each side to ground 1:10 ±70V (DC+peak AC) or 70Vrms
1:100 ±700V (DC+peak AC) or 700Vrms 1:10 and 1:100
±700V (DC+peak AC) or 500Vrms
1:10 ±1000V (DC+peak AC) or 1000Vrms 1:100 ±1400V (DC+peak AC) or 1000Vrms
±7V (into 50kΩ load)
<±5mV
0,9mVrms
50 (for using 1M input system oscilloscope)
-10°C to 40°C
-30°C to 70°C 25% to 85% RH 25% to 85% RH 4 x AA Cells Power lead or Mains Adapter (6VDC/60mA or regulated VDC/40mA)
95cm Length of In Weight 500g Dimensions (LxWxH) 202mm x 83mm x 38mm
30cm
a. The supplied voltage must be less than 12V and greater than 4.4V, otherwise the probe
could be damaged or can’t be operated properly.
b. Polarity is “+” inside and “–” outside. For wrong polarity, built-in circuit protects the
probe, no danger or damage will occur.
c. When the voltage of the cells become too low, the power indicator on the will flicker.
Page 7
10. Derating Curve
The derating curve of the absolute maximum input voltage in common mode is shown as follows.
11. Offset Adjustment
If the offset voltage is too large, short the input leads and turn the adjust variable resistor (DC voltage adjustment) which you find in the hole of the panel by using a flat-head screwdriver until the offset voltage is lowest.
12. Inspection Procedure
a. Connect the BNC output connector to the vertical input of a general purposed
oscilloscope.
b. Install four AA cells or connect an appropriate mains adaptor or power lead to the correct
line voltage. c. Set the oscilloscope input coupling to DC and the 1V/div. Center the trace on the display. d. Connect the inputs of the probe to power lines. e. Set the range of the probe to 1/100. f. Then, a 50Hz/60Hz sine-wave of proper amplitude will be displayed on the screen of the
oscilloscope and this means the probe is working properly.
13. Cleaning
Use a soft cloth to clean the dirt. Prevent damage to probe. a. Avoid immersing the probe. b. Avoid using abrasive cleaners. c. Avoid using chemicals contains benzene or similar solvents.
Page 8