Testec TT-HVP 40 User Manual

-3
0
Atten (dB)
Frequency
DC 300 Hz
sa
Safety Precautions
This high voltage probe must only be used by personnel who are trained, experienced, or otherwise qualified to recognize hazardous situations and who are trained in the safety precautions that are necessary to avoid possible injury when using such a device.
Do not work alone when working with high voltage circuits
For your own safety, inspect the probes for cracks and frayed or broken leads before each use. If
defects are noted, DO NOT use the probe.
Hands, shoes, floor and work bench must be dry. Avoid making measurements under humid, damp or other environmental conditions that might effect the safety of the measurement situation.
It is advisable to turn the high voltage source off before connecting or disconnecting the probe.
The probe body should be kept clean and free of any conductive contamination. Refer to the
section on cleaning.
Operation
Connect the plugs to the volts [Hi] and com [Lo] terminals of your voltmeter
Select the desired voltage function and range; do not use autoranging
Whenever possible, turn the high voltage source off before making any connections
Connect the divider probe common lead (alligator clip) to a good earth ground or reliable chassis ground
Do not attempt to take measurements from sources where the chassis or return lead is not
grounded
The ground lead is critical to the safe operation of the probe. Failure to make this connection when making high voltage measurements may result in personal injury or damage to the probe or volt­meter. This connection must always be made BEFORE the probe tip comes in contact with the high voltage and must not be removed until the probe tip has been removed from the high voltage source.
Do not connect the ground clip lead to the high voltage source for any reason.
Before turning the high voltage on, make sure that no part of your body is in contact with the device under test.
Measure the voltage remembering that the voltage being measured is 1000 times greater than the voltmeter reading.
Turn the high voltage off.
Disconnect the probe tip from the high voltage source BEFORE removing the ground clip lead.
Cleaning
Clean only the exterior probe body and cables. Use a soft cotton cloth lightly moistened with a mild solution of detergent and water. Do not allow any portion of the probe to be submerged at any time.
Dry the probe thoroughly before attempting to make voltage measurements.
Do not subject the probe to solvents or solvent fumes as these can cause deterioration of the probe body and cables.
TESTEC Elektronik GmbH Phone : +49-(0)69-943335-0 Bornheimer Landstr. 32-34 Fax : +49-(0)69-94333555 60316 Frankfurt Internet: www.testec.de
Warning
Bedienungsanleitung TESTEC Hochspannungs-Tastkopf TT-HVP 40
Dieser Hochspannungs-Tastkopf wurde so entwickelt, dass der Benutzer bei ordnungsgemäßer Benutzung vor einem versehentlichen Kontakt mit der Hochspannung geschützt wird. Vor Gebrauch des Tastkopf ist diese Bedienungsanleitung zu lesen und ihr Inhalt zur Kenntnis zu nehmen. Eine unsachgemäße Vorgehensweise oder die inkorrekte Analyse einer Messsituation kann zu schweren Unfällen führen.
Allgemeine Informationen
Der Testec Hochspannungs-Tastkopf TT-HVP 40 ist ein Zubehörteil für den Einsatz bei analogen oder digitalen Multimetern mit einem Eingangswiderstand von 10 M (± 1 %). Das Testec-Modell TT-HVP 40 ist ein 1000:1-Teiler, der die Messkapazität eines Voltmeters auf 40 kV DC/28 kV RMS AC erweitert.
Spezifikationen
Maximale Eingangsspannung CAT I Temperatur-Koeffizient
Gleichspannung: 40 kV Weniger als 200ppm/°C Wechselspannung: (0 bis 300 Hz) 28 kV RMS
Genauigkeit Genauigkeit des Teilungsverhältnisses
Gleichspannung: ± 2 % (1 kV bis 20 kV) 1000:1 ± 1 % bei Abschluss 10 MΩ ± 3 % (20 kV bis 40 kV) 2000:1 ± 6 % bei Abschluss 1 M
Andere Charakteristika
Eingangswiderstand Betriebstemperatur
1000 M 0 °C bis +50 °C
Kabellänge Lagerungstemperatur
1 Meter -20 °C bis +70 °C
Höhe Feuchtigkeit
Bis 4.6000 Meter (15.000 Fuß) Bis 80 % relative Feuchte bei +40 °C
Genauigkeit
Wechselspannung: 5 % bei 60 Hz
Der Wechselspannungs-Ausgang fällt mit der Frequenz auf –3dB bei 300 Hz ab. Die Grafik rechts zeigt die typische Genauigkeit bei Wechselspannung.
Die tatsächliche Genauigkeit bei Wechsel­spannung variiert mit der Menge der abgegriffenen Eingangskapazität.
Warnung
®
®
Bedienungsanleitung TESTEC Hochspannungs-Tastkopf TT-HVP 40
TESTEC Elektronik GmbH
Westerbachstr. 58 D-60489 Frankfurt
Phone: +49-(0)69-943335-0 Fax: +49-(0)69-94333555 Internet: www.testec.de
User’s Guide TESTEC High Voltage Probe TT-HVP 40
-3
0
Atten (dB)
Frequency
DC 300 Hz
sa
User’s Guide TESTEC High Voltage Probe TT-HVP 40
This high voltage probe is designed to prevent accidental shock to the operator when properly used. This operating note must be read and understood prior to using the probe. Improper procedures or incorrect analysis of the measurement situation can result in serious shock.
General Information
The Testec High Voltage Probe TT HVP 40 is an accessory to be used with analog or digital multimeters having an input resistance of 10 M Ω (± 1 %). The Testec Model TT-HVP 40 is a 1000:1 divider which extends a voltmeter’s measurement capability to 40 kV dc/ 28 kV rms ac.
Specifications
Maximum Input Voltage CAT I Temperature Coefficient
DC Voltage: 40kV Less than 200ppm/°C AC Voltage: (0 to 300 Hz) 28 kV RMS
Accuracy Division Ratio Accuracy
DC volts: ± 2 % (1 kV to 20 kV) 1000:1 ± 1 % when terminated in 10 MΩ ± 3 % (20 kV to 40 kV) 2000:1 ± 6 % when terminated in 1 M
Other Characteristics
Input Resistance Operating Temperature
1000 M 0 °C to +50 °C
Cable Length: Storage Temperature:
1 meter -20 °C to +70 °C
Altitude: Humidity
Up to 4.6000 meters (15.000 ft) Up to 80 % relative humidity at +40 °C
Accuracy
AC Volts: 5 % at 60 Hz
AC output derates with frequency to -3db at 300 Hz. The graph at the right shows typical AC accuracy.
Actual AC accuracy varies with the amount of shunted input capacitance.
Warning
Sicherheitsvorkehrungen
Dieser Hochspannungs-Tastkopf darf nur von Personen verwendet werden, die über
entsprechende Ausbildung und/oder Erfahrung verfügen oder die auf Grund einer anderweitigen Qualifikation in der Lage sind, gefährliche Situationen zu erkennen. Auch muss eine Unterweisung in den Sicherheitsvorkehrungen erfolgt sein, die befolgt werden müssen, um Unfälle bei der Verwendung eines solchen Gerätes zu vermeiden.
Arbeiten Sie bei Umgang mit Hochspannungs-Schaltkreisen niemals allein.
Untersuchen Sie zu Ihrer eigenen Sicherheit den Tastkopf vor jedem Gebrauch auf Risse sowie Defekte der Kabelisolation. Werden Defekte festgestellt, darf der Messkopf NICHT verwendet werden.
Hände, Schuhe, Boden und Arbeitsplatz müssen trocken und sauber sein. Vermeiden Sie Messungen in Feuchtigkeit oder anderen Umgebungsbedingungen, die die Sicherheit beeinträchtigen können.
Vor dem Anschließen oder Abklemmen des Tastkopfes ist die Hochspannungsquelle abzuschalten.
Der Körper des Tastköpfes ist sauber und frei von leitenden Verunreinigungen zu halten. Siehe auch „Reinigung“.
Verwendung
Verbinden Sie die Stecker mit den Spannungs- und Masse-Anschlüssen Ihres Voltmeters.
Wählen Sie die gewünschte Spannung sowie den gewünschten Bereich aus; verwenden Sie nicht die automatische Bereichswahl.
Schalten Sie die Hochspannungsquelle ab, wann immer dies möglich ist, bevor Sie Verbindungen herstellen.
Verbinden Sie die Masse-Leitung des Tastkopfes (Krokodilklemme) mit einer guten Erdung oder
einer verlässlichen Gehäuseerde.
Versuchen Sie nicht, Messungen an Quellen durchzuführen, deren Gehäuse oder Rückleitung nicht geerdet sind.
Die Erdleitung ist von entscheidender Bedeutung für eine sichere Verwendung des Tastkopfes.
Wird bei Hochspannungs-Messungen versäumt, diesen Anschluss vorzunehmen, so kann dies zu Personenschäden oder zu Schäden am Tastkopf oder Voltmeter führen. Diese Verbindung ist immer herzustellen, BEVOR der Kontakt zwischen dem Tastkopf und der Hochspannungsquelle hergestellt wird. Sie darf erst getrennt werden, wenn der Tastkopf von der Hochspannungsquelle getrennt wurde.
Verbinden Sie unter keinen Umständen die Erdleitung mit der Hochspannungsquelle.
Stellen Sie vor dem Einschalten der Hochspannungsquelle sicher, dass kein Teil Ihres Körpers zu dieser Kontakt hat.
Nehmen Sie die Messung vor und bedenken Sie dabei, dass die gemessene Spannung 1000 mal größer ist als die Anzeige des Voltmeters.
Schalten Sie die Hochspannung ab.
Entfernen Sie den Tastkopf von der Hochspannungsquelle, BEVOR Sie die Erdleitung abtrennen.
Reinigung
Reinigen Sie nur das Äußere von Tastkopf und Kabel. Verwenden Sie dafür ein weiches Baumwolltuch, sowie eine milde Lösung von Reinigungsmittel in Wasser. Tragen Sie dafür Sorge, dass keinesfalls ein Teil des Messkopfes in Wasser eingetaucht wird.
Trocknen Sie den Messkopf gründlich ab, bevor Sie damit beginnen, Spannungsmessungen vorzunehmen.
Setzen Sie den Messkopf niemals Lösungsmitteln oder Lösungsmitteldämpfen aus, denn diese können eine Schädigung des Gerätekörpers sowie der Kabel verursachen.
TESTEC Elektronik GmbH Tel.: +49-(0)69-943335-0 Bornheimer Landstr. 32-34 Fax: +49-(0)69-94333555 D-60316 Frankfurt Internet: www.testec.de
Warnung
®
®
TESTEC Elektronik GmbH
Westerbachstr. 58 D-60489 Frankfurt
Tel.: +49-(0)69-943335-0 Fax: +49-(0)69-94333555 Internet: www.testec.de