TES TES-2730 Instruction Manual

DIGITAL MULTIMETER
TES-2730
TES ELECTRICAL ELECTRONIC CORP.
CONTENTS
TITLE PAGE
SAFETY INFORMATION ................................................................ 1
I. SPECIFICATIONS ........................................................................ 2
1-1 General Specifications ......................................................... 2
1-2 Electrical Specifications ....................................................... 4
II. NAME OF PARTS AND POSITIONS .......................................... 8
III. PRECAUTIONS AND PREPARATIONS FOR
MEASUREMENT ........................................................................10
IV. MEASUREMENTS ..................................................................... 11
4-1 DC Voltage Measurement ................................................... 11
4-2 AC Voltage Measurement ................................................... 12
4-3 DC Current Measurement ................................................... 12
4-4 AC Current Measurement .................................................... 13
4-5 Resistance Measurement .................................................... 13
4-6 Capacitance Measurement .................................................. 14
4-7 Frequency Measurement ..................................................... 14
4-8 Temperature Measurement ................................................. 14
4-9 Diode Tests .......................................................................... 15
4-10 Continuity Measurement .................................................... 15
V. BATTERY & FUSE REPLACEMENT ......................................... 16
5-1 Battery Check-up & Replacement ....................................... 16
5-2 FUSE Replacement .............................................................16
VI. OPTIONAL ACCESSORY ......................................................... 19
1
SAFETY INFORMATION
Read the following safety information carefully before attempting to operate or service the meter.
To avoid damages to the instrument do not apply the signals which exceed the maximum limits shown in the technical specifications tables.
Never measure current while the test leads are inserted into the input jacks.
Do not use the meter or test leads if they look damaged. Use extreme caution when working around bare conductors or bus bars.
Accidental contact with the conductor could result in electric shock.
Use the meter only as specified in this manual ; otherwise, the protection provided by the meter may be impaired.
Read the operating instructions before use and follow all safety information.
Caution when working with voltages above 60V DC or 30 V AC RMS. Such voltages pose a shock hazard.
Before taking resistance measurements or testing acoustic continuity, disconnect circuit from main power supply and all loads from circuit.
Safety symbols:
Caution refer to this manual before using the meter.
Dangerous voltages.
Meter is protected throughout by double insulation or reinforced insulation.
When servicing, use only specified replacement parts.
Comply with EN-61010-1
2
1. SPECIFICATIONS
1-1 General Information
Environment conditions
Installation Categories II
Pollution Degree 2
Altitude up to 2000 meters
Indoor use only
Relatively humidity 80% max
Operation Ambient 0〜40℃
maintenance & Clearing :
Repairs or servicing not covered in this manual should only be performed by qualified personal.
Periodically wipe the case with a dry cloth
and detergent. Do not use abrasives or solvents on this
instruments.
Operating Principle Dual slope integration.
Numerical Display
3 1/2 digit liquid crystal display (LCD)
7mm height. maximum reading 1999 .
Unit and Sign Display :
Fig-1
3
DC Direct current or voltage
AC Alternating current or voltage
Negative polarity
Decimal point
V Volts mV Millivolts ( 1×10
-3
volts ) A Ampere ( amps ). Current uA Microamperes ( 1×10
-6
amps )
uF Microfarads ( 1×10
-6
Farads ), Capacitance.
nF Nanofarads ( 1×10
-9
Farads ), Capacitance.
pF Picofarads ( 1×10
-12
Farads ), Capacitance. Ω Ohms. Resistance KΩ Kilohms ( 1×10
3
ohms ). Resistance
MΩ Megohm ( 1×10
6
ohms ). Resistance
MHz Megahertz ( 1×10
6
ohms ). Frequency
KHz Kilohertz ( 1×10
3
cycles / sec ). Frequency
Celsius temperature scale
Fahrenheit temperature scale
Low Battery
Data Hold
Continuity Beeper
Diode
Range Selection
All ranges are measured by single Range switch operation.
Over Range Indication
OL
“ appears on the display.
Low Battery Indication
The
is displayed when the battery
voltage drops below the operating voltage.
Sampling Rate
2.5 time per second except in Freq. Mode. 1 time per second in Freq. Mode.
Power Requirements
9-Volt battery (NEDA or JIS 006p IEC6f22)
Battery Life ( typical )
Approx. 60hours. (Alkaline Battery)
Operating Temperature
:0℃
to 40℃ ( 32℉ to 104℉ )
4
and Humidity RH below 80%
Storage Temperature
-10℃ to 60℃ ( 14℉ to 140℉ )
and Humidity RH below 70% Dimensions
180 (L) × 82 (W) × 38(H) mm
7.09”(L)×3.2”(W)×1.5”(H)
Weight
365g
Accessories
Test leads (pair), Operating Instruction,
Spare fuse ( 0.5A / 250V ), Software, Battery, RS-232C Cable, 9pin to 25pin Gender Changer.
Optional Accessories
K type thermocouple.
1-2 Electrical Specifications
Accuracies are ±(...% of reading + ...digits) at 23℃±5℃,below 80% RH.
DC Voltage
Range Resolution Accuracy
Input
Impedance
Overload
Protection
200m
V
0.1mV 600Vrms
2V 1mV
20V 10mV
200V 100mV
0.5%+1
600V 1V 0.5%+2
10MΩ
750Vrms
AC Voltage ( 50Hz to 500Hz )
Range Resolution Accuracy
Input
Impedance
Overload
Protection
200mV 0.1mV 600Vrms
2V 1mV
20V 10mV
1%+5
10MΩ
750Vrms
5
200V 100mV
600V 1V 1.2%+5
DC Current
Range
Resolutio
n
Accuracy
Burden
Voltage
Overload Protection
200uA 0.1uA
2mA 1uA
20mA 10uA
200mA 100uA
1%+1
0.35V
0.5A / 250V Fast
Blow Fuse & Diode
20A 10mA 1.2%+3 0.8V
20A / 380V Fast Blow Fuse 20A for 30sec. max. 10A continuity.
AC Current ( 50Hz to 500Hz )
Range
Resolutio
n
Accuracy
Burden
Voltage
Overload Protection
200uA 0.1uA
2mA 1uA
20mA 10uA
200mA 100uA
1.2%+3 0.35V
0.5A / 250V Fast
Blow Fuse & Diode
20A 10mA 1.5%+5 0.8V
20A / 380V Fast Blow Fuse 20A for 30sec. max. 10A continuity.