TES TES-2700 Instruction Manual

AUTO - RANGING
DIGITAL MULTIMETER
TES-2700
INSTRUCTION MANUAL
TES ELECTRICAL ELECTRONIC CORP.
1. SAFETY INFORMATION
Read the following safety information carefully before attempting to operate or servi ce the met er.
To av oid damages to the instrument do not apply the signals which exceed the maximum limit s shown in the tech nical specifications tables.
Never measure current whi le the test leads are inserted into the input jacks.
Do not use th e mete r or test l eads if they look damaged. Use extreme cau tion when working around bare conductors or bus bar s.
Accidental contact with the conductor could result in electric shock.
Use the meter onl y as spe cified in thi s manual ; otherwise, the protect ion prov ided by the meter may be impaired.
Read the operating instructions before use and follow all safety information.
Caution when working with voltages above 60V DC or 30 V AC RMS. Such voltages pose a shock hazard.
Before taking resistance measurements or testing acoustic continuity, disconnect circuit from main power supply and all loads from circuit.
Safety symbols:
Caution refer to this manual before using the meter.
Dangerous voltages.
Meter is protected throughout by double insulation or reinforced insulation.
When servicing, use only specified replacement parts.
Comply with EN-61010-1
1
2. FEATURES
3-1/2 (3200 COUNT S) Di git al LCD with Bar-graph
Auto- Ranging on Volt, Ohm , Current measurement
Auto-power off
Provides data HOLD mode
Ohms , Dio de / Audible Contin uity and Ca paci tance input Overl oad protect ion to 50 0Vrms
20A / 380V hig h ene rgy fuse protection o n 20A A C/DC
Flip - up stand
3. SPECIFICATIONS
3-1 General Information
Environment conditions :  Installation Categories II
Pollution Degree 2
Altitude up to 2000 meters
Indoor use o nly Relatively humidity 80% max. Operation Ambien t 0〜40℃
mai ntenance & Cle aring: Repairs or servicing not covered in this
manual should only be performed by qual ifi ed personal.
Periodically wipe the case with a dry
cloth and detergent.
Do not use abrasives or solvents on this
instruments.
Operating Principle :
Dual slope integration
Numerical Display 3 1/2 digit liquid crystal display (LCD) maximum
reading 3260 and Bargraph indication.
2
Display Symbol Defini tion
Decimal point
AC
Alternating current or voltage
DC
Direct current or voltage
V
Voltage
mV
Millivolt ( 1×10-3 volt )
A
Ampere (amps) Current
mA
Milliampere ( 1×10-3 amp )
Ω
Ohm. Resistance
KΩ
Kilohm ( 1×10
3
ohm ). Resistance
MΩ
Megohm (1×10
6
ohm ). Resistance
Diode
Continuity Beeper Low Battery Negative polarity
MANU
Manual Range
Date Hold
Range Selection All ranges are select ed by sin gle Range Switch
operati on.
Over Range Indication LCD will show a “OL”.
Low Battery Indication The
is displayed when the battery
voltage drops below the pr oper opera ting voltage.
Sampling Rate :12.5 times/sec of Bar-graph indication, 2 times
/sec of digital display
Power Source 2 pcs of “AAA” 1.5V Battery
3
Battery Life 500h rs approx.
Polar ity Automatic polarity “
“ indicates negative
input.
Auto Power Off Time Approx. 10 mi nutes
Operating Temperature :0℃ to 40℃ ( 32℉ to 104℉ ) and Humid ity be low 80 % RH
Storage Temperature :-10℃ to 60 ( 14 to 140 ) and Humidity below 70% RH
Dimensions 180 (L)
×
82 (W) × 38(H) mm
Weight Approx. 365
g
Accessories Test leads, Spa re fuse (0.5A/250V), Battery
and Instruction manual.
3-2 Electrical Specifications
Accuracies are ± (...% of reading + ...dig its) at 23℃± 5℃,below 80% RH.
DC Voltage : (
Autoran ging & Manu al range )
Range Resolution Accuracy
Input
Impedance
Overload
Protection
320m V 0.1mV
> 100MΩ
3.2 V 1mV
11MΩ
32V 10mV
320V 100mV
0.5 %+2
600V 1V 0.8 %+2
10MΩ
AC/DC 750V
4
AC Voltage
: 40Hz〜400Hz ( Autorange & Manual range )
Range Resolution Accuracy
Input
Impedance
Overload Protection
3.2 V 100uV
11MΩ
32V 1mV
320V 10mV
600V 1V
1.2 %+4 10MΩ
AC/DC 750V
DC Current :
(uA mA Autorangin g & Manual Range )
Range Resolution Accuracy
Burden Voltage
Overload Protection
320u A 0.1uA 0.2 V max
3200 uA 1uA 1.8V max
32mA 10uA 1.2V max
320m A 100uA
1%+2
1.8 V max
0.5 A/250 V fast bl ow fuse & Diode
20A 10mA 1.5 %+2 0.6 V max
20A/380V fast blow fuse 10A Continuity 20A for 30 sec. max.
AC Current :
40Hz〜400Hz ( uA mA Autorange & Manual range )
Range Resolution Accuracy
Burden Voltage
Overload Protection
320u A 0.1uA 0.2 V max
3200 uA 1uA 1.8V max
32mA 10uA 1.2V max
320m A 100uA
1.5 %+5
1.8 V max
0.5 A/250 V fast bl ow fuse & Diode
20A 10mA 2.0 %+5 0.6 V max
20A/380V fast blow fuse 10A Continuity 20A for 30 sec. max.
5