TES TES-1362 Instruction Manual

RS
Printing Humidity/Temperature Meter
1362
INSTRUCTION MANUAL
TABLE OF CONTENTS
TITLE PAGE
IV. GENERAL PRECAUTIONS AND PREPARATIONS
FOR MEASUREMENT ................................................8
VI. SETTING MODE.........................................................10
VII. INTERVAL PRINTING SET-UP ...............................14
VIII. BATTERY REPLACEMENT.....................................15
IX. RECALIBRATION PROCEDURE............................15
I. SAFETY INFORMATION
Read the following information carefully before attempting
to operate or service the meter. Use the meter only as specified in this manual, otherwise
the protection provided by the meter may be impaired.
Environmental conditions
c Altitude up to 2000 metres d Indoor use only e Relatively humidity; 95% max. f Operating temperature; 0 to 50 (32 to 122℉)
Maintenance & clearing
c Repairs or servicing not covered in this manual should
only be performed by qualified personnel.
d Periodically wipe the case with a dry cloth. Do not use
abrasives or solvents.
Compliance symbols
Complies with EMC Directive 89/336/EEC
1
II. SPECIFICATIONS
2-1 General information
Display: Triple 4-digit LCD.
-
Over range indication : Low battery indication: symbol appears on the LCD
Sampling rate: approximately 1 per second Power requirements: Qty 6, 1.5V size AAA alkaline batteries,
Battery life: Approx. 90 hours. (interval printing = 60 minutes,
excluding beeper).
100 hours ( without printing & beeper). Printing interval:6 sec. to 59 min. 59 sec. Operating temperature and humidity:
0℃ to 50℃ (32℉ to 122℉) below 95% RH Storage temperature and humidity:
-10℃ to 60℃ (14℉ to 140℉) below 70% RH
Dimensions: Meter : 193 (L) × 74 (W) × 37(H) mm
Probe : 160 (L) × 13 (D) mm
Weight: Approx. 430g with batteries & paper roll. Accessories suppllied: Carrying case, instruction manual,
Printer:Thermal type with 16 characters per line using
38mm width thermal paper.
-
appears on the LCD.
when the battery voltage falls below the operating voltage.
or AC adaptor; 9Volts @ 500mA minimum.
Cable length : 1m approx.
batteries, 2 rolls of thermal printer paper, type K thermocouple probe.
2
2-2 Electrical specifications
Humidity:
Range: 10%RH to 95%RH Resolution: 0.1%RH Accuracy: ±3%RH from 30%RH to 95%RH
±5%RH from 10%RH to 30%RH
Temperature T1 (Integral sensor):
Range: -20℃ to +60℃ (-4℉ to +140℉) Resolution: 0.1℃ / 0.1 Accuracy: ±0.8℃, ±1.5
Dew-point:
Range: -44.0℃ to 58.5℃ (-47.2℉ to 137.3℉) Accuracy: 1.0℃/1.8
Temperature (T2) (External thermocouple sensor):
Range: -200℃ to +1333℃ (-328℉ to +2431℉) Resolution: 0.1℃ from -200℃ to 999
0.1℉ from -328℉ to 999.9℉ (1℃ or 1 outside this range).
Accuracy: -50℃ to -200℃: ±(0.2%rdg + 1.8℃)
-50℃ to 500℃: ±(0.1%rdg + 0.8℃) 500℃ to1333℃: ±(0.2%rdg + 1.8℃)
-58℉ to -328℉: ±(0.2%rdg + 3℉)
-58℉ to 1000℉: ±(0.1%rdg + 1.6℉) 1000℉ to 2431℉: ±(0.2%rdg + 3℉)
Temperature coefficient:
0.1 times the applicable accuracy specification per from 0℃ to 18℃ and 28℃ to 50℃ (32℉ to 64℉ and 82℉ to 122℉)
3
III. LOCATION OF PARTS
Fig-1
1. LCD: Measured values, units, symbols, and decimal points are displayed on the screen.
2.
: Power on/off button
3. ℃
4. PRINT
Button for selecting temperature units ℃ or ℉.
c Press this button to start printing the current data. The data format is as follows:
LINE 1 20:12:01 02 - 21 LINE 2 58.9 % RH LINE 3 td 15.7 LINE 4 T1 24.3 LINE 5 T2 24.8