TensCare perfect BEAUTY Instructions For Use Manual

Page 1
0
INSTRUCTIONS FOR USE
READ CAREFULLY BEFORE USE
Page 2
1
QUICKSTART GUIDE
Place the electrode pads around
the area to be treated
Press and hold the OFF button
to switch the device off
ON
Button P
Channel
2 UP
Channel
2 DOWN
Channel
1 UP
Channel
1 DOWN
OFF
Select the programme by
pressing the button P
1 2 3 4 5
6
Page 3
2
Page 4
3
Dear Customer,
Thank you for choosing perfect BEAUTY. TensCare stands for high­quality, thoroughly tested products for the applications in the areas of gentle electrotherapy, muscle toning, continence management and pain relief during labour.
Please read these instructions for use carefully and keep them for later use, be sure to make them accessible to other users and observe the information they contain.
Best regards, Your TensCare Team
Page 5
4
Contents
1. INTRODUCTION ................................................................................................. 6
2. INTENDED USE .................................................................................................. 6
3. PERFECT BEAUTY FEATURES ......................................................................... 6
4. HOW ‘EMS’ WORKS ........................................................................................... 7
5. CONTRAINDICATIONS, WARNINGS & CAUTIONS .......................................... 8
6. INFORMATION ABOUT THE PROGRAMME SETTINGS ................................ 10
7. PROGRAMMES ................................................................................................ 12
7.1. PRESET PROGRAMMES .......................................................................... 12
7.2. PROGRAMMES SETTINGS ....................................................................... 13
7.3. USING THE PROGRAMMES ..................................................................... 15
8. ELECTRODE PADS .......................................................................................... 17
8.1. PAD POSITIONING .................................................................................... 17
8.2. PAD ADVICE .............................................................................................. 18
8.3. ELECTRODE PLACEMENT ....................................................................... 19
9. CONTENT ......................................................................................................... 22
10. UNIT INFORMATION ..................................................................................... 23
10.1. CONTROLS & DISPLAY ......................................................................... 23
10.2. OPERATING INSTRUCTIONS ................................................................ 24
11. SETTING UP AND USING THE PERFECT BEAUTY .................................... 26
11.1. INSTALLATION OF BATTERIES ............................................................. 26
11.2. CONNECTING LEAD WIRES .................................................................. 27
11.3. PREPARING FOR SESSION .................................................................. 27
11.4. TRAINING SESSION ............................................................................... 27
11.5. AFTER YOUR TRAINING SESSION ....................................................... 28
12. CLEANING & STORAGE ............................................................................... 28
13. EMC ............................................................................................................... 28
14. DISPOSAL OF WASTE ELECTRICAL AND ELECTRONIC PRODUCTS
(WEEE) .................................................................................................................... 29
15. ACCESSORIES ............................................................................................. 29
16. WARRANTY ................................................................................................... 30
17. TROUBLESHOOTING ................................................................................... 31
18. GENERAL SPECIFICATION .......................................................................... 32
Page 6
5
SYMBOLS USED
Attention! Please follow the instructions in the user’s instructions for use.
TYPE BF EQUIPMENT: Equipment providing a degree of protection against electric shock, with isolated applied part. Indicates that this device has conductive contact with the end user.
This symbol on the unit means “Refer to instructions for use”.
Temperature Limitation: indicates the temperature limits to which the medical device can be safely exposed.
Lot Number: indicates the manufacturer’s batch code so that the batch or lot can
be identified.
Humidity Limitation: indicates the humidity limits to which the medical device can be safely exposed.
Serial Number: indicates the manufacturer’s serial number so that a specific
medical device can be identified.
Do not dispose in household waste.
Catalogue Number: indicates the manufacturer’s catalogue number so that the
device can be identified.
Atmospheric Pressure: indicates the atmospheric limits to which the medical device can be safely exposed.
Date of Manufacture: indicates the date which the medical device was manufactured. This is included within the serial number found on the device
(usually in the battery compartment), either as “E/Year/Number” (YY/123456) or
“E/Month/Year/Number” (MM/YY/123456).
This medical device is indicated for home use.
This medical device is not water resistant and should be protected from liquids. The first number 2: Protected against access to hazardous parts with a finger,
and the jointed test finger of 12 mm ø, 80 mm length, shall have adequate clearance from hazardous parts, and protected against solid foreign objects of 12.5 mm ø and greater.
The second number 2: Protected against vertically falling water drops when
enclosure is tilted up to 15˚. Vertically falling drops shall have no harmful effects when the enclosure is tilted at any angle up to 15˚ on either side of the vertical.
LOT
S/N
REF
IP22
Page 7
6
1. INTRODUCTION
Device Description & Principles of Design
The perfect BEAUTY is a powered muscle stimulator used for body toning and shaping, as well as for the reduction of wrinkles and cellulite.
It sends a small and gentle stimulation (similar to your natural nerve impulses) to the muscles being treated via two or four auto-adhesive electrode pads, in order to help you to firm and tone your body, and to maintain your youthful appearance.
We have developed bespoke programmes dedicated to each specific area: face, abs, buttocks & hips, thighs, bust, arms and calves.
2. INTENDED USE
Perfect BEAUTY is a medical
device designed to be used in the home healthcare environment to improve muscle tone and bulk, and reduce signs of ageing. It is suitable for use by all who can control the device and understand the instructions.
Do not use the device for any purpose other than this intended use.
Warning: Not suitable for use in children without medical supervision.
3. PERFECT BEAUTY FEATURES
Dual Channel
Two independent channels with four electrode pads to treat two areas at the same time.
Comfortable Stimulation
Gentle stimulation with 90 small steps of intensity (about 1 mA per step).
8 Preset Programmes
Including facial toning and firming; as well as programmes for body toning and shaping: abs, buttocks & hips, thighs, bust, arms and calves.
Memory
Features 3 functions: programme retention (automatically starts in the last programme used), number of uses and time of usage.
Open Circuit Detection
Automatically resets the strength to zero if the electrode pads are not correctly applied.
Detachable Belt Clip
Enables the user to clip the unit onto a belt.
Large LCD Screen
Clearly shows the operation of the unit and the programme and intensity being used.
Unique electrode pads
Face: Simple application to stimulate all
areas of the face in one treatment session.
Breast*: Large surface area for comfortable stimulation.
*Sold separately
Page 8
7
Note: The electrode pads need
to be applied directly to the skin
before the intensity increase buttons will operate. If the Perfect BEAUTY detects the unit is not properly connected to you (for example if the electrode pads or leads come loose) then the unit switches the intensity to zero. This safety feature prevents any unpleasant changes in output.
4. HOW ‘EMS’ WORKS
E.M.S. stands for Electrical Muscle Stimulation and has successfully been used in medical rehabilitation and training in competitive sports. EMS produces intensive and effective muscular contraction.
EMS is a complement to other physical therapy and should always be combined with active mobility, strength, coordination and functional training.
In training, the technology for electrotherapy is used for all kinds of muscular exercise - warm-up, strength, speed, power, resistance, endurance, recovery and also for rehabilitation. EMS is well known and works as an excellent complement to regular training.
Advantages of EMS
Use of EMS will help to tone and shape the body by contracting the selected muscle group. It is particularly convenient for areas that are famous for losing their shape and hard to exercise such as the upper arms and the face.
Successful athletes all over the world have discovered the advantages of electrical muscle stimulation, such as an increase of the local circulation and the size of the muscle fibres. EMS also helps to raise the intake of oxygen as well as improving the metabolic exchange and the consumption of energy of the user.
How EMS Works
Muscular activity is produced by the central and peripheral nervous systems transmitting electrical stimuli to the muscles of our body. EMS uses external electrical impulses that work through the skin to stimulate the nerves supplying a specific muscle group.
This current then passes into the nerve fibres controlling that part of the muscle stimulating it to contract. So, electrical stimulation (EMS) artificially activates a muscle for you enabling to exercise the muscles and, as with any kind of exercise if performed regularly, build strength and tone.
Different frequencies have different effects; low frequencies (1-10 Hz) coupled with long impulse times, for example, have a purifying and relaxing effect through individual contractions, whereby the circulation in the treated muscle is simultaneously improved and removal of metabolic end products is supported (lymphatic drainage). The oxygen supply to the muscle is improved.
In contrast, by means of a rapid succession of contractions (fibrillation), medium frequencies (20-50 Hz) can put a high level of strain on the muscle, thus promoting the muscular structure.
Page 9
8
5. CONTRAINDICATIONS,
WARNINGS & CAUTIONS
In this manual:
Warning is used when failure to
follow the instructions may
result in serious injury or death.
A Caution is used when failure
to follow the instructions may result in a minor or moderate injury, or damage to the device or other property.
Notes are used to provide
clarification or
recommendation.
CONTRAINDICATIONS:
Do NOT use if you have a
pacemaker (or if you have a heart rhythm problem) or with any electronic medical devices. Using
this unit with electronic medical devices may cause erroneous operation of the device. Stimulation in the direct vicinity of an implanted device may affect some models. Stimulation on the front of the neck can affect your heart rate. Very strong stimulation across the chest may cause an extra heartbeat.
Do NOT use during the first three
months of pregnancy. It is not known
whether TENS and EMS may affect foetal development.
Do NOT use on the abdomen in the
later stages of pregnancy. Stop
using immediately if you experience unexpected contractions.
WARNINGS:
Do NOT use when driving,
operating machinery, or similar actions needing muscular control. Loose electrode pads,
damaged leads, or sudden changes in contact may cause brief involuntary muscle movements.
Do NOT use to mask or relieve
undiagnosed pain. This may
delay diagnosis of a progressive condition.
Do NOT use if you have, in the
area being treated: active or suspected cancer or undiagnosed pain with a history of cancer. Stimulation
directly through a confirmed or suspected malignancy should be avoided as it may stimulate growth and promote spread of cancer cells.
CAUTIONS:
Caution should be used if you
have suspected or diagnosed epilepsy as electrical stimulation may affect seizure threshold.
Caution should be observed
when using the device at the same time as being connected to monitoring equipment with body worn electrode pads. It may interfere with the signals being monitored.
Caution should be used
following recent surgical procedures. Stimulation may disrupt the healing process.
Caution Simultaneous
connection to high frequency surgical equipment may result in burns and damage to the stimulator.
Page 10
9
Caution Strong
electromagnetic fields (electro-
surgery/ microwave cookers/ mobile phones) may affect the correct operation of this unit. If it appears to behave unusually, move it away from these devices.
Caution Do not permit use by
children unable to understand the instructions or persons with
cognitive disabilities, i.e.; Alzheimer’s
disease or dementia.
Caution Keep away from
children under 5 years of age.
Long cord - risk of strangulation in infants.
Caution should be observed
when using the Perfect BEAUTY at high strength settings. Prolonged use at high settings may cause muscle injury or tissue inflammation. The Perfect BEAUTY is capable of delivering outputs in excess of 10 mA (RMS) or 10 V averaged over any period of 1 sec.
DO NOT PLACE ELECTRODE PADS:
On skin, which does not have normal
sensation. If the skin is numb too
great a strength may be used, which could result in skin inflammation.
On broken skin. The electrode pads
could encourage infection.
On the front of the neck. This could
cause the airway to close, giving breathing problems. May cause sudden drop in blood pressure (vasovagal response).
Over the eyes. May affect eyesight
or cause headaches.
Across the front of the head. Effect
on patients who have had strokes or seizures is not known. May affect your sense of balance. The effects of
stimulation on the brain are unknown.
ELECTRODE PADS CAUTION:
Caution: Do not ignore any
allergic reaction to the
electrode pads: If a skin irritation
develops, stop using TENS, as this type of electrodes may not be suitable for you. Alternative electrode pads specially made for sensitive skin are available (see E-696-SS).
Caution: Do not use this
device with leads or electrode pads other than those recommended by the manufacturer. Performance
may vary from specification. Electrodes with smaller surface area may cause tissue irritation.
Note: Electrode pads supplied have no measurable latex content.
TO KEEP YOUR DEVICE IN GOOD WORKING ORDER, OBSERVE THE FOLLOWING ADDITIONAL CAUTIONS:
Caution: Do not immerse your
device in water or place it close to excessive heat such as a fireplace or radiant heater or sources of high humidity such as a nebulizer or kettle as this may cause it to cease to operate correctly.
Caution: Keep the device
away from sunlight, as long­term exposure to sunlight may affect the rubber causing it to become less elastic and crack.
Caution: Keep the device
away from lint and dust, as
Page 11
10
long-term exposure to lint or dust may affect the sockets or cause the battery connector to develop a bad contact.
Caution: Temperature & Relative Humidity of storage:
-20°C to +40°C, 8% to 70% R.H. Temperature & Relative Humidity of transportation: -20°C to +40°C, 8% to 70% R.H.
Caution: Do not attempt to open or modify the TENS unit.
This may affect the safe operation of the unit and will invalidate the warranty.
There are no known side effects to use and long-term
use is not harmful. However, the device works your muscles and unaccustomed amounts of exercise may produce muscle ache. If this happens, simply reduce the strength of the machine or allow your muscles to rest until the aching disappears.
6. INFORMATION ABOUT
THE PROGRAMME SETTINGS
Each programme has its own combination of Frequency and Pulse Width settings which allow for different muscle groups to be exercised and toned through the electrode pads.
Frequency (measured in Hz -
pulses per second)
Low frequencies (1-10 Hz) have a purifying and relaxing effect through individual contractions.
Medium frequencies (20-50 Hz) can put a high level of strain on the muscle, thus promoting the muscular structure.
Pulse Width (measured in μs -
millionths of a second)
The perfect BEAUTY unit has pulse widths of 80 to 600 μs. Generally speaking, the higher the pulse width, the more "aggressive" the stimulation feels, if the pulse width is set high enough, it will usually elicit a muscle contraction, which is required for an effective toning of the muscle group selected.
Work/Rest:
All the programmes are intermittent. Once you have reached a comfortable
level, 5 seconds after you stop pressing the button, the intermittent Work/Rest phase will start.
For each programme, the unit will go through a stimulation phase (Work) for 0-10 seconds, followed by a phase without stimulation (Rest) for a similar time.
This is to allow your muscles to recover from exercise and avoid fatigue.
During the Work period (W), the unit will generate contractions, try to contract the muscle voluntarily at the time to get the best possible results.
This cycle Work/Rest will continue for the duration of the programme selected.
W and R are displayed on the screen.
Page 12
11
Each programme is a balanced mix of toning, firming, and shaping specifically designed for the designated muscle group.
All the programmes except FACE 1 consist of seven phases, with a combination of settings optimised for each part of the body selected from the following.
Capillarization (3 Hz to 8 Hz) Treatment at these frequencies significantly increases arterial flow in the area treated. This is useful for recovering after intense work and improves local microcirculation.
Tissue Elasticity (5 Hz to 15 Hz) Stimulates the surface muscle fibres. These frequencies facilitate the removal of substances accumulated on the surface and improve the dynamic appearance of the skin.
Micro lifting (12 Hz to 20 Hz) Toning facial muscles to improve both the appearance and the dynamism of facial muscles.
Modelling (12 Hz to 30 Hz) Due to a combination of capillarizing and toning impulses, this frequency range helps mobilise fat in areas where it tends to accumulate.
Firming up (20 Hz to 30 Hz) Firming up muscles working mainly on slow-twitch fibres.
Lipolysis (30 Hz to 40 Hz) This frequency range evokes a significant increase in the temperature of the subcutaneous tissue. This results in an increased metabolism, and thus facilitates lipolysis.
Toning (40 Hz to 50 Hz) Toning up working mainly on fast-twitch fibres.
Definition and bulk (60 Hz to 100 Hz)
Defining muscles in the arms and bust working mainly on explosive fibres. Grows muscle size.
For all programmes except FACE 1, follow the guidelines below to set the intensity level.
Phase 1 - Preparation: Turn up the stimulation gradually, until you feel an intermittent tingling.
Phases 2 to 6 - Activation: Increase the intensity until you feel a very smooth tensing under the pads, there should be little visible movement.
Use a low intensity for the first session to help you to understand how the machine works).
During the programme and over the next few days the intensity should be increased gradually to produce marked, but not painful muscle contractions.
Phase 7 - Cool down: Turn down the intensity gradually until you can only feel a mild tingling pulse.
For the programme FACE 1, which has just one phase, increase the intensity to the point where you can just see the muscle moving.
Note: The stimulation should remain comfortable and relaxing at all times.
Page 13
12
7. PROGRAMMES
7.1. PRESET PROGRAMMES
Note: These programmes must be used in combination with a healthy diet
and regular active exercises.
Programme
Muscle group
Description
Special Feature
1
FACE 1
Facial Quick Tone
Helps tone, firm, lift and tighten the soft tissues of the face, counteracting the effects of ageing.
Intensity limited to 40 mA.
Use with face electrode pads (see E-CM3050).
2
FACE 2
Facial Deep Tone
In the quest for a youthful appearance, facial muscles and skin may also be stimulated to improve circulation, help activate fibroblasts which produce collagen and elastin and can also tone flaccid and sagging face muscles.
Intensity limited to 40 mA.
Use with face electrode pads (see E-CM3050).
3
ABS
Abdominal
Suitable for people who wish to tone and shape their body to obtain a thinner silhouette and to enhance general fitness levels.
4
HIP
Buttocks & Hips
5
THIGH
Thighs
Circulation boosting to improve the appearance of cellulite.
6
BUST
Bust/Pectorals
For women who have sagging muscles and a build-up of fat around the bra strap area.
Can be used with breast electrode pads* (see E- CMR60130) for improved comfort.
*Sold separately.
7
ARM
Arms
Our upper arms often have little shape, usually caused by a lack of movement and muscular work. This programme is designed to firm and bulk muscles.
Circulation boosting to improve skin tone and appearance.
8
CALF
Calves
It is usually a lack of trained muscles in the legs which disturbs us most. Targeted muscle training for the calves is very simple to perform.
Page 14
13
7.2. PROGRAMMES SETTINGS
PROGRAMME 1: Facial Quick Tone
Phase
1 2 3 4 5 6 7
Total
Time (min)
20
20
Stimulation (s)
0
Pause (s)
1 Frequency (Hz)
70 Pulse width (µs)
100
Ramp (s)
1/1
PROGRAMME 2: Facial Deep Tone
Phase
1 2 3 4 5 6 7
Total
Time (min)
2 3 2 2 2 2 2
15
Stimulation (s)
2 4 3 6 3 4 1.2
Pause (s)
2.5 4 3.5 4 3.5 2 .3 Frequency (Hz)
100
10
50 4 40
4
100
Pulse width (µs)
140
200
140
220
140
220
80
Ramp (s)
.5
.5
.5
.5
.5
.5
.2
PROGRAMME 3: Abdominal
Phase
1 2 3 4 5 6 7
Total
Time (min)
5 5 10 8 7 8 2
45
Stimulation (s)
2 4 8
10 6 4
1.2 Pause (s)
2.5 4 6 1 4 4 .5
Frequency (Hz)
90
50
10 4 30
65
100
Pulse width (µs)
200
300
500
600
360
300
80 Ramp (s)
.5
.5
.9
1.3
.6
.5
.2
PROGRAMME 4: Buttocks & Hips
Phase
1 2 3 4 5 6 7
Total
Time (min)
5 5 5 5 5 3 2
30
Stimulation (s)
3 6 8 4 3 5 1.2 Pause (s)
2 5 4 4 4 5 .5 Frequency (Hz)
100
65
10
50
75
45
100
Pulse width (µs)
200
300
500
300
260
260
80
Ramp (s)
.5
.7
.9
.5
.5
.5
.2
Page 15
14
PROGRAMME 5: Thighs
Phase
1 2 3 4 5 6 7
Total
Time (min)
5 5 5 5 5 3 2
30
Stimulation (s)
5 3 1
10 6 4
1.2 Pause (s)
4 4 .7 1 4 4 .5 Frequency (Hz)
10
65
100
4
45 8 100
Pulse width (µs)
400
300
260
500
360
500
80 Ramp (s)
.9
.5
.3
.9
.6
.5
.2
PROGRAMME 6: Bust/Pectorals
Phase
1 2 3 4 5 6 7
Total
Time (min)
5 5 5 5 5 3 2
30
Stimulation (s)
3 4 5 1 4 5 2
Pause (s)
3
2.5 3 .3
4.5 6 .5 Frequency (Hz)
90
60
10 4 45
65
90 Pulse width (µs)
150
150
220
300
200
200
80 Ramp (s)
.3
.5
1.3
.9
.6
.5
.2
PROGRAMME 7: Arms
Phase
1 2 3 4 5 6 7
Total
Time (min)
5 5 5 5 5 3 2
30
Stimulation (s)
3 4 8 4 6 4 1.2 Pause (s)
2.5 3 3 8 4 4 .5 Frequency (Hz)
70
60
10
50
10
65
100
Pulse width (µs)
200
200
300
220
300
220
80 Ramp (s)
.3
.5
.9
.9
.6
.5
.2
PROGRAMME 8: Calves
Phase
1 2 3 4 5 6 7
Total
Time (min)
5 5 5 5 5 3 2
30
Stimulation (s)
1 5 4 6 8 4 1.2 Pause (s)
1 4 3.5 4 4 4 .5 Frequency (Hz)
100
40
50
10 4 25
100
Pulse width (µs)
200
220
200
300
300
220
80 Ramp (s)
.3
.5
.5
.5
.9
.5
.2
Page 16
15
7.3. USING THE PROGRAMMES
FACIAL PROGRAMMES:
The FACE 1 and FACE 2 programmes are a non-invasive option for helping lift and/or tighten the soft tissues of the face, improving facial muscle tone and reducing the signs of ageing.
Recommended use: One treatment session daily, 5 days/week, for 12 weeks.
You may notice some positive effects after just one session, but normally it will take about 6 weeks before you notice a difference.
Take care when adjusting the intensity as facial muscles are particularly sensitive; intensity should be increased gradually, starting with a very low level of stimulation (just above perception) and increasing with care until you reach a good level of stimulation, represented by good muscle activation.
On the face, it is not necessary to reach levels of intensity capable of causing discomfort! Great and significant results are obtained through consistency and patience.
Warning: Use these programmes only with the face electrode pads included in the
kit (see E-CM3050).
Warning: a minimum distance of 3 cm. must be kept between
the electrode pad and the eyeball.
The TensCare face electrode pads (see E-CM3050) are designed to reduce any risk to a minimum. Some secondary effects may subsist but they should be temporary.
Note: You might notice mild skin redness at the site of the
electrode pads after the treatment session. It should fade within 10 to 20 minutes.
Note: Very rarely temporary after-effects, such as mild eyelid
flutter, have been reported after facial stimulation. This has always resolved within a few days.
ABS, HIP, THIGH PROGRAMMES:
These programmes help improving body toning and body shaping by generating passive muscle contractions in the selected area.
It may improve your general fitness levels as well as achieving a thinner silhouette by building up the muscles.
The THIGH programme contains specific settings that help increase the flow of blood circulation, and modify the metabolism of the lipocytes. It allows to stimulate the subcutaneous deposits of fat and it assists reducing or eliminating the “Orange Peel effect (cellulite) of the skin surface.
Page 17
16
Recommended use:
Generally, 3 sessions per week, over a period of 4 to 6 weeks, gives good initial results depending on the body shaping, toning, sports training or rehabilitation and recovery needs.
BUST PROGRAMME:
Electrical Stimulation can build, strengthen, and firm the pectoral muscles giving bulk and lift.
Exercise and massage therapy to enhance the size and shape of your breasts
Natural method to enhance your all­natural body
Recommended use:
Generally, three sessions per week, over a period of 4 to 6 weeks, gives good initial results depending on the body shaping, toning, sports training or rehabilitation and recovery needs.
Note: For improved comfort, use this programme with the
breast electrode pads* (see E- CMR60130). *Sold separately.
ARM PROGRAMME:
The upper arms often get saggy and flabby especially in women with the triceps (back of the upper arm). It is
commonly referred to as “Bingo Wings” or “Bat Wings”. It is tricky to tone this
muscle group and some women find this area ungraceful or even embarrassing.
Perfect BEAUTY helps to tone, firm and lift the biceps as it allows you to generate contractions in this specific muscle group.
Recommended use: Generally, 3 sessions per week, over a
period of 4 to 6 weeks, gives good initial results depending on the body shaping, toning, sports training or rehabilitation and recovery needs.
CALF PROGRAMME:
Most people are aiming for thin but muscly legs with a harmonious balance between the size of the thighs and the size of the calves. Often, the calves are either too skinny with only little muscle mass or too big with a long muscle mass especially around the ankle.
The aim of this programme is to tone,
shape and lift the calves’ muscles to
achieve slender legs with improved definition of the muscles as well as a slimming effect on the ankles.
Recommended use: Generally, 3 sessions per week, over a
period of 4 to 6 weeks, gives good initial results depending on the body shaping, toning, sports training or rehabilitation and recovery needs.
GENERAL ADVICE:
Ideally, use perfect BEAUTY once every two days for each muscle with a day of rest in between. It is possible to work on pairs of muscles, for example
Page 18
17
thighs and abdominal muscles, treating one set one day and the other the next day.
Warning: Working on too
many muscles at the same
time is not recommended.
Once good results have been obtained the sessions may be reduced to once or twice a week until the user feels that optimum body shaping has been achieved.
8. ELECTRODE PADS
8.1. PAD POSITIONING
The electrode pads must always be used in pairs (two electrode pads on each channel), so that the signal can flow in a circuit.
Note: The face electrode pad
acts as a pair of standard electrode pads. Connect both pigtail leads to the same channel on the unit.
Electrode placement for EMS is very important for obtaining the best results.
Place two electrode pads over the bulk of the muscle, with one electrode over
the muscle’s motor point. The motor
point is the area on the skin that is
located closest to the motor nerve’s
entry into the muscle – about 1/3 of the way down the muscle from the spine. Here, it is easiest to trigger a contraction by electrical stimulation.
Experiment by moving the electrode pads across the skin until you locate the point over the muscle that gives the cleanest contraction. In the examples in section 8.3., the electrode pads are marked +ve (red) and -ve (black) to match the red and black leads. The positive (red connector) electrode pad should be over the motor point.
Large muscle groups may require stimulation with two channels, that is, four electrodes simultaneously.
When exercising smaller muscles, take care to adjust the intensity level slowly
Page 19
18
as the motor nerves may be more sensitive and using smaller electrode pads increases the current density.
You may feel more sensation in one pad than the other. This is normal – it depends on where the electrode pads are in relation to your nerves.
Use the square electrode pads (see E- CM5050) for all areas except the face where specific electrode pads are required (see E-CM3050).
For the bust area, it is recommended to
use the optional breast electrode pads for improved comfort (see E-CMR60130).
Always check that the unit is OFF before attaching or removing electrode pads.
Perfect BEAUTY resets strength to zero if the pad or lead is disconnected from your body. This is to prevent sudden changes in sensation when the pad is re­connected.
8.2. PAD ADVICE
The electrode pads supplied are reusable but for single patient use.
In order to obtain the best conductivity through the electrode pads always ensure that they are in good condition and tacky.
Before use make sure your skin is clean and dry.
Peel the electrode pads from their protective plastic shield by holding and lifting one corner of the pad and pulling. Do not pull on the pigtail wire of the pad.
After use always replace the electrode pads on the plastic liner and replace them in the re-sealable plastic bag.
If the electrode pads dry out, then it is best to buy a replacement pack. In an emergency, it may be possible to restore some of the tackiness of the pad by adding a tiny drop of water on each pad and spreading around. If too much water is added, the electrode pad will become too soft. If that happens then it is suggested in order to try and re-establish some adhesiveness, to place them sticky side up in a refrigerator for a few hours.
In very hot weather the gel on the electrode pads may become soft. In such cases place them, still on their plastic liners and in their pouch into a fridge until they return to their normal condition.
The electrode pads provided are latex-free.
Replace the electrode pads when they lose their stickiness.
Poor connection may cause discomfort and skin irritation.
Storage life of an unopened pack of electrode pads is 2 years. This may be affected by very high temperatures or very low humidity.
Page 20
19
8.3. ELECTRODE PLACEMENT
The electrode pads placement is specific to the programme used. Please refer to the placement guidelines below.
Note: The + electrode pad needs to be plugged in to the red lead wire connector, the – refers to the one for the black lead wire connector.
FACE 1 and FACE 2 – Pads placement
The face electrodes stimulate all the muscles on the face simultaneously.
ABS – Pad placement 1 OR ABS – Pad placement 2
Page 21
20
HIP – Pads placement
THIGH – Pad placement 1 OR THIGH – Pad placement 2
BUST – Pad placement 1
Page 22
21
BUST – Pad placement 2 with the optional breast electrode pads (E-CMR60130)
ARM – Pad placement 1 OR ARM – Pad placement 2
CALF – Pad placement
Page 23
22
9. CONTENT
The pack contains:
1 x Perfect BEAUTY multi-purpose muscle stimulator and body toner
2 x Lead wires (L-CPT)
4 x 50x50 mm electrode pads (E-
CM5050)
4 x 30x50 mm face electrode pads (E-CM3050)
2 x AA 1.5V alkaline batteries
1 x Detachable belt clip
1 x Storage pouch
1 x Instructions for use
Page 24
23
10. UNIT INFORMATION
10.1. CONTROLS & DISPLAY
Programme selected
Channel 2 strength
Time remaining
Channel 2 strength UP
Programme selector
Channel 2 strength DOWN
Phase Skip
Channel 1 strength DOWN
Channel 1 strength UP
Low Battery
Keypad Lock
Channel 1 strength
ON/OFF
Output active indicators
Programme phase
Work / Rest
Page 25
24
10.2. OPERATING INSTRUCTIONS
ON/OFF To turn the unit ON, press the
ON/OFF button on the top of the unit and hold for 3 to 5 seconds until the display shows.
To turn the unit OFF, press the ON/OFF button on the top of the unit and hold for 3 to 5 seconds until the display stops
The unit will turn OFF automatically: When the Timer reaches zero, If it is left at zero strength for more than
5 minutes. The unit will start with the programme
you used last. At first use, or after changing batteries,
the display shows that the unit is automatically set in programme FACE 1 at zero strength.
Note: Always check unit is OFF before applying or removing electrode pads.
The backlight will turn off 10 seconds after the last button press.
On the front of the unit there are six buttons:
STRENGTH CONTROLS The buttons marked CH1+, CH1-,
CH2+, CH2- are the strength controls.
The left-hand buttons change strength in the left-hand lead, CH1, and the right-hand buttons change the right-hand lead, CH2.
To increase strength in steps of 1 mA, press and release the top part of the channel buttons, CH1+ and CH2+.
The strength levels are shown on the LCD.
The strength control buttons will not operate until the unit is properly connected to you. If the Perfect BEAUTY detects a disconnection in either channel it will automatically return the strength to zero in that channel and flash ‘00’.
The unit has 90 levels of strength except for the facial programmes (FACE 1 and FACE 2), which only go up to 40 mA (40 steps).
If you hold down the strength controls for 3 to 5 seconds, the strength will start increasing continuously.
You may feel nothing over the first few presses. Continue pressing until the sensation is strong but comfortable. At each phase of the programme the intensity automatically adjusts to give a comfortable stimulation, however further increases during use may be necessary if your body becomes used to the sensation. To decrease the strength, press and release the lower part of the buttons marked CH1- and CH2-.
The yellow LED on the output socket indicates that there is an active output which can be high enough to be harmful if not used correctly. The display will remain on for 5 seconds after the plug is removed.
PROGRAMME CONTROL
The button marked P is the programme control. The
P
CH1+
CH1-
CH2+
CH2-
Page 26
25
perfect BEAUTY has eight different,
specially developed programmes. When the unit is first switched on it automatically enters the programme FACE 1.
Each time you press and release the P button, the programme changes and the letter is shown on the LCD.
Each time you change the programme, the strength level reverts back to zero. This is a safety feature to alleviate any sudden feeling of a surge, as each programme gives a different sensation.
PHASE SKIP
All the programmes except FACE 1 have seven phases.
The bar column in the middle of the display increases as each phase starts:
Phase 1 Phase 2 Phase 3
And so on up to the 7th phase (7 bars).
This button can be used to skip a
phase. Press and hold this button for 3 to 5 seconds to skip and go to the next phase of the selected treatment session.
LOW BATTERY
An empty battery symbol will
show when you need to change the batteries. The unit will shut down about 2 minutes after this.
KEYPAD LOCK
If you do not press any keys
for 30 seconds, the keypad will lock so that only the keys marked CH1- or CH2- will work. This is to avoid
accidental changes in setting. To unlock, press either CH1- or CH2- keys.
OPEN CIRCUIT CUTOUT
If the perfect BEAUTY is not correctly connected to your body, the strength in the channel which is not properly connected will automatically reset to zero. This is to prevent sudden changes if a broken connection is re-made.
MEMORY
The perfect BEAUTY has a Memory with three functions:
1) Programme Retention. When you switch the unit on, it will automatically start in the programme which was being used when it was switched off.
2) Usage. Press the phase skip button and CH2- together and hold down
for 3 to 5 seconds. The display will show the number of times used and
duration in hours used.
Press the same buttons again to return to normal controls.
3) Memory Reset. To reset memory to
zero, hold down the phase skip button
and CH1- buttons together for 3 to
5 seconds.
Number of uses
Time used in hours
Page 27
26
11. SETTING UP AND USING THE PERFECT BEAUTY
11.1. INSTALLATION OF
BATTERIES
1) Remove belt clip by sliding down.
2) Remove battery cover by pulling on
tag.
3) Insert batteries.
Ensure that the batteries are inserted
the right way as shown in battery compartment and that the ribbon is behind them.
4) Replace battery cover and belt clip.
When the batteries are running
low, a low battery indicator will show on the screen and it is important to change the batteries as soon as possible.
Rechargeable batteries
The unit will work with rechargeable batteries, but the display may appear dim.
Storage Remove batteries from your perfect
BEAUTY if the unit is unlikely to be
used for a long period. Some types of batteries may leak corrosive fluid.
Battery Life
Batteries should last about 48 hours at full power.
Unused batteries have a nominal shelf life of 3 years, but will usually last longer than this.
Battery Warnings
Do NOT pierce, open,
disassemble, or use in a humid
and/or corrosive environment.
Do NOT expose to temperatures over 60°C(140F).
Do NOT put, store or
leave near sources of heat, in direct strong sunlight, in a high temperature location, in a pressurized container or in a microwave oven.
Do NOT immerse in water, or
get wet.
Do NOT short-circuit.
Do NOT connect the device
unless the battery cover is in
place.
If battery leakage occurs and comes in contact with the skin or eyes, wash thoroughly with lots of water.
Keep batteries out of the reach of small children.
Page 28
27
Caution NEVER attempt to
recharge an alkaline battery. Risk of explosion.
Caution Do not mix old, new or
different types of batteries as this may lead to battery leakage or low battery indication.
Disposal: Always dispose of batteries responsibly according to local government guidelines. Do not throw batteries onto a fire. Risk of explosion.
11.2. CONNECTING LEAD
WIRES
Insert the lead wire plug into the base of the unit.
Attaching the electrode pads to the lead wire.
Push the pin ends firmly into the pigtail ends of the electrode pads.
The lead wires may be damaged by rough handling, and should be treated with care.
Lead wire colour coding.
The ends of the lead wires are coloured red or black. The positive (red connector) electrode pad should be the closest to the spine (see 8.3.).
11.3. PREPARING FOR SESSION
1) Before use make sure your skin is
clean and dry. You can use conductivity barrier wipes for effective electrode pads contact such as TensCare Contact Skin Wipes (see E-CW50).
If you are using the FACE programmes, clean the skin carefully before applying electrode pads, removing all makeup, oils and creams. Residue from these can prevent the electrode pads from adhering properly. Use an Aloe Vera based cream after treatment to soothe any skin irritation.
2) Place the electrode pads on the skin,
see section 8.
3) If you are using your perfect
BEAUTY while moving, you can attach the unit to your belt using the belt clip or hang it around your neck using a neck cord (not included). For safety, use thread or wool that will snap easily if the unit snags on something.
4) You can safely lean on the electrode
pads. The sensation might change.
Warning: Ensure the Perfect BEAUTY is switched OFF
before applying the electrode
pads on the skin.
11.4. TRAINING SESSION
1) Press and hold the ON button on the
control unit for 3 to 5 seconds to switch the control unit on.
When switched on for the first time, the perfect BEAUTY will automatically select the programme FACE 1. Next time, it will
Page 29
28
automatically select the programme you were using the last time it was switched off.
2) You can select from the 8 preset programmes. Details in section 7 will help you identify the best programme to suit you.
3) Identify the correct pads positioning for the selected area and apply the electrode pads as described in section 8.3.
4) Once the electrode pads are correctly positioned, adjust the intensity as described in 7.3. with the buttons CH1-, CH1+, CH2- and CH2+. The LCD display shows the strength of intensity used for each channel.
Note: If the sensation becomes uncomfortable, reduce the intensity.
11.5. AFTER YOUR TRAINING
SESSION
When the timer reaches zero, your session is complete and the unit turns off.
1) Check that the control unit is OFF. If it is not, hold down the OFF button to switch off then remove the electrode pads from your skin by holding the pad itself and gently pulling. Be careful not to pull the wires as this may damage them.
2) Replace the electrode pads to their protective plastic shield and return them to the re-sealable plastic bag.
Note: When removing the electrode pads, DO NOT PULL ON LEAD WIRES.
12. CLEANING & STORAGE
Clean the case and lead wires at least once a week by wiping with a damp cloth and a solution of mild soap and water. Wipe dry.
Do not immerse your perfect
BEAUTY unit in water.
Do not use any other cleaning
solution than soap and water.
Storage life
Storage life of an unopened pack of
self-adhesive electrode pads is 2 years. This may be affected by very high temperatures or very low humidity.
The unit has no fixed shelf life.
13. EMC
Wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies can affect this equipment and should be kept at least a distance d = 3,3 m away from the equipment.
(Note. As indicated in Table 6 of IEC 60601-1-2:2007 for ME EQUIPMENT, a typical cell phone with a maximum output power of 2 W yields d = 3,3 m at an IMMUNITY LEVEL of 3 V/m).
Note: For hospital use, full EMC advice tables are available on request.
Page 30
29
14. DISPOSAL OF WASTE ELECTRICAL AND ELECTRONIC PRODUCTS (WEEE)
One of the provisions of the European Directive 2002/96/CE is that anything electrical or electronic should not be treated as domestic waste and simply thrown away. To remind you of this Directive all affected products are now being marked with a crossed-out wheelie bin symbol, as depicted below.
To comply with the Directive, you can return your old electro-therapy unit to us for disposal. Simply print a postage­paid PACKETPOST RETURNS label from our website www.tenscare.co.uk, attach this to an envelope or padded bag with the unit enclosed, and post it back to us. Upon receipt, we will process your old device for components recovery and recycling to help conserve
the world’s resources and minimise
adverse effects on the environment.
15. ACCESSORIES
Expected Service Life
The machine will often last for more
than 5 years, but is warrantied for 2 years. Accessories (lead wires, electrode pads, and batteries) are not covered by the warranty.
Lead life depends greatly on use.
Always handle the leads with care. We recommend to replace the lead
wires regularly (about every 6 months).
Electrode pads should last 12 to 20 applications, depending on skin condition and humidity.
AA alkaline batteries should last about 48 hours of continuous use.
Replacement electrode pads, new batteries and lead wires are available from your supplier or distributor (see back cover for contact details), by mail order from TensCare, by telephone using a credit or debit card, or through the TensCare website.
The following replacement parts may be ordered from TensCare at www.tenscare.co.uk or +44(0) 1372
723434.
E-CM3050 Face electrode pads
Pack of 4 pairs
E-CMR60130 Breast electrode pads
Pack of 4 (2 pairs)
E-CM5050 Square electrode pads
Pack of 4
E-CM50100 Large electrode pads
Pack of 4
E-696-SS Square hypoallergenic
electrode pads
Pack of 4 L-CPT Lead wire B-AA 1.5V AA batteries X- BC-PT Replacement belt clip X- BL-PTT Replacement battery
cover E-CW50 Pack of 50 wipes
Page 31
30
16. WARRANTY
This warranty refers to the unit only. It does not cover, electrode pads, battery, or the lead wires.
PRODUCT WARRANTY INFORMATION
This product is warranted to be free from manufacturing defects for 2 years from date of purchase.
This warranty is void if the product is modified or altered, is subject to misuse or abuse; damaged in transit; lack of responsible care; is dropped; if incorrect battery has been fitted; if the unit has been immersed in water; if damage occurs by reason of failure to follow the written instruction booklet enclosed; or if product repairs are carried out without authority from TensCare Ltd.
We will repair, or at our option replace free of charge, any parts necessary to correct material or workmanship, or replace the entire unit and return to you during the period of the warranty. Otherwise, we will quote for any repair which will be carried out on acceptance of our quotation. The benefits conferred by this warranty are in addition to all other rights and remedies in respect of the product, which the consumer has under the Consumer Protection Act
1987.
Our goods come with guarantees that cannot be excluded under the UK consumer Law. You are entitled to have
the goods repaired or replaced if the goods fail to be of acceptable quality.
Before you send your unit for service
Before sending in your unit for service, please take a few minutes to do the following:
Read your manual and make sure you follow all the instructions.
Returning your unit for service
Should repair be needed within the warranty period, enclose the tear off section of the warranty card (see page
33) and your proof of purchase receipt. Please ensure all relevant details are completed before sending your unit in for service. Please ensure your contact details are still current and include a brief description of the problem you are experiencing together with your purchase receipt.
For hygiene reasons, please do not include used electrode pads. Send only the unit and lead wires.
Please return the unit and warranty card (see page 33) at your cost to:
TensCare Ltd
PainAway House, 9 Blenheim Road, Longmead Business Park, Epsom, Surrey KT19 9BE, UK
Should you require any further information please do not hesitate to contact us by calling our number:
+44 (0) 1372 723 434.
Page 32
31
17. TROUBLESHOOTING
If your perfect BEAUTY is not working properly, please check the following:
Problem
Possible causes
Solution
No display
Flat batteries.
Replace batteries.
Batteries inserted incorrectly.
Remove plastic wrap. Check + / - .
Damaged springs in battery compartment.
Contact supplier.
Low battery display
Low batteries.
Replace batteries.
No sensation
Intensity is not strong enough.
Increase strength. Most users will feel something at a setting
below 20 mA.
Output will not increase above zero
Open circuit cut-out operating.
Lead not connected to body or faulty/damaged.
Ensure machine is attached to your body correctly.
Try using the second lead wire. Purchase replacement if necessary.
(Lead wires can break at the bend where they leave the machine giving no, or intermittent, output).
Sudden change in sensation
If you disconnect and re­connect a few minutes later, the signal will feel quite a lot stronger.
Always return strength to zero after
disconnecting the leads or electrode pads.
If the above review has failed to resolve your problem, or to report unexpected operation or events, call TensCare or your local supplier or distributor (address on back cover) for advice.
Contact TensCare customer service on +44 (0) 1372 723 434. Our staff are trained to assist you with most issues you may have experienced, without the need to send your product in for service.
Page 33
32
18. GENERAL SPECIFICATION
Waveform
Asymmetrical Bi-Phasic
Amplitude FACE over 500 Ohm load Other programmes over 1
kOhm load
40 mA zero to peak +ve in 40 steps. 90 mA zero to peak +ve in 90 steps +/- 10% Modified constant charge control
Output plug
Fully shielded: touch proof
Channels
Dual channel
Batteries
2 x AA alkaline (two AA batteries)
Weight
75 gms without batteries
Dimensions
120 x 60 x 20 mm
Safety Classification
Internal power source. Designed for continuous use. No special moisture protection.
Environmental specifications:
Operating:
Storage:
Temperature range: 5 to 40C Humidity: 15 to 93% RH non-condensing Atmospheric pressure: 70 kPa to 106 kPa Temperature range: -25 to +70C Humidity: Up to 93% RH non-condensing
TYPE BF EQUIPMENT
Equipment providing a degree of protection against electric shock, with isolated applied part.
This symbol on the unit means “Refer to Instructions for
use”.
IP22
The unit is not water resistant, and should be protected from liquids.
Complies with EU WEEE regulations
Note: The electrical specifications are nominal and subject to variation from
the listed values due to normal production tolerances of at least 5%.
Page 34
33
PLEASE RETAIN THIS WARRANTY CARD. RETURN THIS PORTION ONLY WHEN YOU RETURN YOUR PRODUCT FOR
REPAIR UNDER WARRANTY.
NAME: ADDRESS:
POSTCODE: DAYTIME TELEPHONE: E-MAIL:
MODEL: DATE OF PURCHASE:
ATTACH PROOF OF PURCHASE
DO NOT SEND IN ELECTRODE PADS
RETAILERS NAME: RETAILERS ADDRESS:
RETAILERS POSTCODE: BRIEF DESCRIPTION OF PROBLEM YOU ARE EXPERIENCING:
WARRANTY IS VOID UNLESS THE ABOVE INFORMATION IS COMPLETED AND CORRECT.
Page 35
34
TensCare aim to give you the best possible product and service. We listen to your suggestions and are constantly trying to improve our products. We also want to learn about the way our products are used, and the benefits they give. If you have anything you would like to share with us, please contact:
www.tenscare.co.uk
Follow us:
https://www.facebook.com/pages/TensCare-Ltd
https://twitter.com/TensCareLtd
https://www.linkedin.com/company/tenscare-limited
https://plus.google.com/+TenscareLtdEpsom
https://uk.pinterest.com/TensCareLtd/
https://tenscareltd.wordpress.com/
https://www.youtube.com/channel/UCzpik9dmLlJ3j0aHOpQ-0sg
EC Declaration of Conformity
TensCare Ltd hereby declare that an examination of the production quality assurance system has been carried out following the requirements of the UK national legislation according to Annex V of the Directive 93/42/EEC on medical devices. We certify that the production quality system conforms with the relevant provisions of the aforementioned legislation, and the result entitles the organization to use the CE 0088 marking on this product.
Page 36
35
Distributed by:
Manufactured by:
TensCare Ltd, 9 Blenheim Road, Epsom, Surrey KT19 9BE, UK Tel: +44(0) 1372 723434 www.tenscare.co.uk
Pub No.: I-PBEAUTY-UK Rev 1.0 08/17