Tenmars TM-4001, TM-4002, TM-4003 User Manual

Hot-Wire
Air Velocity Meter
User’s Manual
TM-4001 / TM-4002
/ TM-4003
TENMARS ELECTRONICS CO., LTD
HB2TM4001002
CONTENTS
1.
PREFACE ............................................................. 2
2.
GENERAL DESCRIPTION .................................. 2
3.
SPECIFICATIONS (23℃±5℃) ............................. 3
4.
INSTRUMENT DESCRIPTION ............................ 6
5.
LCD DESCRIPTION ............................................ 7
6.
POWER ON/OFF ............................................... 10
7.
Auto power off .................................................... 10
8.
BACKLIGHT ....................................................... 10
9.
CHANGE AIR VELOCITY UNIT ......................... 10
10. CHANGE UNIT.................................................... 11
11. CHANGE FUNCTION ......................................... 11
12. TIME CONSTANT SETTING MODE .................. 11
13. AIR FLOW SETTING MODE ............................. 12
14. DATA HOLD........................................................ 12
15. SINGLE DATA RECORD ................................... 12
16. VIEWING OR CLEARING DATA RECORDS .... 13
17. MAX/MIN/AVG .................................................... 13
18. PRECAUTIONS ................................................. 14
19. SOFTWARE INSTALLATION(TM-4003) ............ 15
20. COMPUTER GRADE ......................................... 15
21. EXTERNAL DC POWER ................................... 16
22. SAFETY AND MAINTENANCE STANDARDS .. 16
23. BATTERY REPLACEMENT ............................... 17
24. END OF LIFE ..................................................... 17
TM-4001/4002/4003
2
1. PREFACE
Thank you for purchasing out product. Please, read the operating instructions in details before you use this Hot-Wire Velocity meter, so you will operate the meter correctly. This meter can be operated in industry and home.
WARNING
Non-compliance warning and/or operating instructions may damage the instrument or its parts or injuring the operator.
2. GENERAL DESCRIPTION
Fast response probe. Air flow volume.(Use platinum-resistors sensor) Telescoping probe. Instant/ AVG/ 2/3V MAX flow measurement. Max/AVG/Min & data hold function. Low batter indication “ ”. LCD triple display with LED backlight. Manual data memory and read function
(99 records). (TM-4001/TM-4002)
Humidity measurement. (TM-4002/TM-4003) Calculate dew point temperature, wet bulb
temperature. (TM-4002/TM-4003)
Absolute pressure measurement. (TM-4003) Auto power off with disable function. USB PC serial interface. (TM-4003) Data logging capacity(8000 records).(TM-4003) Sampling interval : Max 24 hours; Min 1S.
(TM-4003)
TM-4001/4002/4003
3
3. SPECIFICATIONS (23±5)
Air velocity measurement: Unit Range Resolution Accuracy m/s 0.01~25.00 0.01
±(3%of reading +1.6%FS)
Km/hr(kph) 0.02~90.00 0.1 mph 0.12~55.9 0.1 Knots(kts) 0.11~48.6 0.01 Ft/min(fpm) 0~4921.0 0.1 Beaufort 1~10 1 -
Air flow(volume) calculate: Unit Range Resolution Accuracy CMM 0~9999 0.1 ­CFM 0~9999 0.1 -
Temperature Measurement: Unit Range Resolution Accuracy
-20~50 0.1
±1
-4~122 0.1
±1.8
Humidity(TM-4002TM-4003): Unit Range Resolution Accuracy
%RH 20~80 0.1 ±3.5%RH %RH <20,>80 0.1 ±5.0%RH
TM-4001/4002/4003
4
Absolute pressure(TM-4003): Unit Range Resolution Accuracy hPa 350~1100 0.1 ±2hPa mmHg 263~825 0.1 ±1.5mmHg inHg 10.3~32 0.1 ±0.1
Meter Dimension: 156x73x35mm(Lx W x H). Probe Dimension:
Wire length: 160cm Probe length: 128cm Probe diameter of tip: 1.6cm Probe diameter of base: 3.2cm
Meter weight: 250g.(without batteries) Probe weight: 240g. Power supply: UM-4(AAA)1.5V*6PCS or
AC100~240V DC9V/0.5A(9mm).
Battery life: 10 hours. Operating temperature & humidity:
-20°C to +50°C, <80 %RH.
Storage temperature & humidity :
-10°C to +50°C, <70 %RH.
Data output: USB PC serial interface. (TM-4003) Display: 60mm(W) x 47mm(L) monochrome
LCD and backlight.
TM-4001/4002/4003
5
Standard accessories:
UM-4(AAA)1.5V battery×6 PCS,
User’s manual, USB cable.(TM-4003), AC100~240V DC9V/0.5A.(9mm)×1(TM-4003),
Install CD disk (TM-4003).
EMC: This test was designed in accordance with
EMC standards in force and its compatibility has been tested in accordance with EN61326-1 (2006).
WARNING
To ensure that the accuracy of the meter, we recommend for calibration once a year.