Tenmars TM-305U User Manual

Page 1
Temperature/Humidity Datalogger
Users Manual V.1.0
Page 2
TABLE OF CONTENTS
FEATURES & SPECIFICATIONS ..................................................................................... 4
START UP AND GENERAL OPERATION ........................................................................ 7
LCD Display................................................................................................................................................................... 8
Sampling Rate............................................................................................................................................................... 9
Recording and Datalogging .................................................................................................................................... 10
Calibration ................................................................................................................................................................... 11
33% Calibration........................................................................................................................................................ 11
75% Calibration........................................................................................................................................................ 13
25C Calibration......................................................................................................................................................... 15
PC SOFTWARE INSTALLATION.................................................................................... 16
USING PC SOFTWARE WITH METER DATALOGGER ................................................ 18
Connect Meter Datalogger to PC ........................................................................................................................... 19
Setup Screen............................................................................................................................................................... 22
Download Records .................................................................................................................................................... 23
Graphic & Chart Analysis ........................................................................................................................................ 26
Data in Listing Format .............................................................................................................................................. 28
Data Export.................................................................................................................................................................. 30
Page 3
Model Number
Temperature/Humidity Datalogger
Page 4
FEATURES & SPECIFICATIONS
Features
Measure humidity and temperature.
Datalogging- 50,000 records.
Fast USB download .
Easy- to-Use PC software for parameter settings and data analysis.
Internal Humidity Temperature alarm.
Page 5
Specifications
General
Operation Condition -40-185℉(-40-85℃)
1-99% RH, non-condensing
Relative Humidity Sensor
Type Capacitor
Measurement Range 1%-99%
Accuracy ±3.0%RH(20~80%)
±5.0%RH(<20%,>80%)
Resolution 0.1%
Temperature Sensor
Type Thermistor
Measurement Range -40-185℉(-40-85℃)
Accuracy ±0.6℃(±1.2℉)from -20-50℃(-4 to 122℉)
±1.0(±1.8) all other ranges.
Resolution 0.1
Power Supply
Battery One 3.6V ½ AA lithium battery
Battery Operation 1 Year
External Power Supply 5 VDC with 0.5A.
Page 6
Applications
Building HVAC monitoring.
Indoor air quality survey.
Locate the presence of combustion fumes from vehicles and appliances.
LCD Display
Meter LCD
1. Temperature Display 4. USB Communication
2. Humidity Display 5. Battery Level Indicator
3. Recording
Page 7
WHAT’S IN THE BOX
START UP AND GENERAL OPERATION
Press the Start button for 2 seconds to turn on Meter Datalogger
The unit is ready to use.
1.Main Unit 2. USB Cable 3. Installation Disk
4. 3.6V ½ AA lithium battery
Page 8
LCD Display
Press Sel to jump to the next LCD display.
Main Screens (1-3)
1. 2. 3.
Main Screen
Temperature(C)
Main Screen
Temperature.(F)
% of Memory Left
Total 24,000 records
Main Screens
Page 9
Sampling Rate
(5sec / 10sec / 1min / 5min / 10min / 30min / 1 hour/ 2 hour)
Can only be set from PC setup screen.