Tenmars TM-305U User Manual

Page 1
Page 1
Temperature/Humidity Datalogger
Users Manual V.1.0
Page 2
Page 2
TABLE OF CONTENTS
FEATURES & SPECIFICATIONS ..................................................................................... 4
START UP AND GENERAL OPERATION ........................................................................ 7
LCD Display................................................................................................................................................................... 8
Sampling Rate............................................................................................................................................................... 9
Recording and Datalogging .................................................................................................................................... 10
Calibration ................................................................................................................................................................... 11
33% Calibration........................................................................................................................................................ 11
75% Calibration........................................................................................................................................................ 13
25C Calibration......................................................................................................................................................... 15
PC SOFTWARE INSTALLATION.................................................................................... 16
USING PC SOFTWARE WITH METER DATALOGGER ................................................ 18
Connect Meter Datalogger to PC ........................................................................................................................... 19
Setup Screen............................................................................................................................................................... 22
Download Records .................................................................................................................................................... 23
Graphic & Chart Analysis ........................................................................................................................................ 26
Data in Listing Format .............................................................................................................................................. 28
Data Export.................................................................................................................................................................. 30
Page 3
Page 3
Model Number
Temperature/Humidity Datalogger
Page 4
Page 4
FEATURES & SPECIFICATIONS
Features
Measure humidity and temperature.
Datalogging- 50,000 records.
Fast USB download .
Easy- to-Use PC software for parameter settings and data analysis.
Internal Humidity Temperature alarm.
Page 5
Page 5
Specifications
General
Operation Condition -40-185℉(-40-85℃)
1-99% RH, non-condensing
Relative Humidity Sensor
Type Capacitor
Measurement Range 1%-99%
Accuracy ±3.0%RH(20~80%)
±5.0%RH(<20%,>80%)
Resolution 0.1%
Temperature Sensor
Type Thermistor
Measurement Range -40-185℉(-40-85℃)
Accuracy ±0.6℃(±1.2℉)from -20-50℃(-4 to 122℉)
±1.0(±1.8) all other ranges.
Resolution 0.1
Power Supply
Battery One 3.6V ½ AA lithium battery
Battery Operation 1 Year
External Power Supply 5 VDC with 0.5A.
Page 6
Page 6
Applications
Building HVAC monitoring.
Indoor air quality survey.
Locate the presence of combustion fumes from vehicles and appliances.
LCD Display
Meter LCD
1. Temperature Display 4. USB Communication
2. Humidity Display 5. Battery Level Indicator
3. Recording
Page 7
Page 7
WHAT’S IN THE BOX
START UP AND GENERAL OPERATION
Press the Start button for 2 seconds to turn on Meter Datalogger
The unit is ready to use.
1.Main Unit 2. USB Cable 3. Installation Disk
4. 3.6V ½ AA lithium battery
Page 8
Page 8
LCD Display
Press Sel to jump to the next LCD display.
Main Screens (1-3)
1. 2. 3.
Main Screen
Temperature(C)
Main Screen
Temperature.(F)
% of Memory Left
Total 24,000 records
Main Screens
Page 9
Page 9
Sampling Rate
(5sec / 10sec / 1min / 5min / 10min / 30min / 1 hour/ 2 hour)
Can only be set from PC setup screen.
Page 10
Page 10
Recording and Datalogging
Press the Start button to start data recording.
The LCD will display REC when Meter is in the datalogging mode.
Press the Start button again to stop recording.
Page 11
Page 11
Calibration
33% Calibration
Press Start + SEL to enter 33% humidity calibration mode.
Press Start to start 33% calibration.
Page 12
Page 12
During the calibration, the LCD will display an antenna sign.
Meter Datalogger will start 33% Rh sampling 45 minutes after the Start button is
pressed.
The COM icon will show up when calibration is completed.
Press SEL to select the next calibration page.
Page 13
Page 13
75% Calibration
Press and SEL button then press the Start button at the same time to enter the calibration mode.
Press Sel one more time to switch to the 75% Rh calibration page.
Press Start to start 75% calibration.
Page 14
Page 14
Meter Datalogger will start 75% Rh sampling 45 minutes after the Start button is
pressed. During the calibration, the LCD will display an antenna icon.
The COM icon will show up when calibration is completed.
Press SEL to select the next calibration page
Page 15
Page 15
25C Calibration
Press Sel one more time to jump to the 25C calibration page.
Press Start to start 25C calibration.
During the calibration, the LCD will display an antenna icon.
When the calibration is completed, the COM will show up.
Page 16
Page 16
PC SOFTWARE INSTALLATION
Installation Requirements:
Minimum system requirements for the software:
-Pentium III - 500 MHz processor or equivalent with 128 megabytes of RAM.
Optimal Performance:
-Pentium 4 - 2.8 gigahertz processor or equivalent with 256 megabytes or RAM.
Operating System Requirements:
-Windows 2000
-Windows XP
-Window 7
-Window 8
Page 17
Page 17
Insert the CD into the disk drive.
Click Setup to start installation.
Page 18
Page 18
Follow the steps on the Installation Shield Wizard and click Finish to complete the installation.
USING PC SOFTWARE WITH Meter Datalogger
Start Window PC software.
Click on All programs, Click on “
Meter Logger”
Click on Logger to start the datalogger program.
Page 19
Page 19
Connect Meter Datalogger to PC
No Connection Screen
Page 20
Page 20
Connect Meter Datalogger to the PC with a USB cable.
Page 21
Page 21
Press Start for 2 seconds to turn on Meter Datalogger. Meter Datalogger is now
connected to the PC.
Connected Screen
If the datalogger is properly connected, these three values will show up.
1. Temperature Reading.
2. Humidity Reading.
3. Battery Status.
4. Connection Status- “Connected”.
Page 22
Page 22
Setup Screen
Click “Setup” to enter the Setup screen.
Page 23
Page 23
The values in the white area can be changed by the user.
The value in the grey area can not be changed by the user.
Click “Save Settings” to save the values to the datalogger.
Download Records
Page 24
Page 24
Click Download to download the stored data.
Page 25
Page 25
Click “Download” to start the download.
Page 26
Page 26
Download Completed
Graphic & Chart Analysis
Page 27
Page 27
Downloaded data will first be displayed in graph format.
Each data groups are listed according to the date of the sampling. Lowest group
number represents earliest sampling date.
Page 28
Page 28
Data in Listing Format
To display data in Listing format:
Click File
Click Chart Preview
Page 29
Page 29
Data Listing Format
Page 30
Page 30
Data Export
Select export data type: Text or Excel CSV.
Click Save to save the file in .CSV/Txt format.
File Saved.
Loading...