Tenmars TM-185, TM-185D User Manual

Page 1
Temperature/Humidity Monitor
TM-185 TM-185D
Users Manual
HB2TM1850000
Page 2
TM-185/TM-185D
1
Table of Contents
1 PREFACE ......................................................... 2
2 CHARACTERISTICS ........................................ 2
3 GENERAL SPECIFICATIONS .......................... 3
4 ELECTRICAL SPECIFICATION ........................ 4
5 INSTRUMENT DESCRIPTION .......................... 5
6 LED DESCRIPTION .......................................... 8
7 CLOCK AND DATE SETUP .............................. 9
8 ALARM RANGE SETUP ................................. 10
9 BATTERY REPLACEMENT ............................ 12
10 EXTERNAL DC POWER ................................ 13
11 SAFETY PRECAUTION .................................. 13
12 SOFTWARE INSTALLATION (TM-185D) ........ 14
13 MAINTENANCE .............................................. 17
14 END OF LIFE .................................................. 18
15 AUTO STORES DATA .................................... 18
16 SINGLE DATA MEMORY ................................ 19
17 VIEWING LOGGED DATA .............................. 19
Page 3
TM-185/TM-185D
2
LED display.
It shows both the value of temperature
and humidity.
It give the selection of °C or °F.
Real time data.
Data Hold function.
Alarm function.
1 PREFACE
Thank you for your patronage, please do read the operation instructions in details before you use this meter, so that operate correctly, make a machine give play to the best function.
2 CHARACTERISTICS
Page 4
TM-185/TM-185D
3
Display: Double rows LED, humidity MAX
reading 999,and temperature MAX reading
1999.
Temperature sensor: Diode.
Humidity Sensor: Capacitor Sensor.
Sampling: 1 times/second.
Power: AC to DC Adapter. (9V/1A).
Backup Power:9V battery NEDA 1604IEC
6F22 or JIS 006P(Only data logger use).
Resolution:0.1%RH0.1°C0.1°F.
Size:260x178x47mmLxWxH. Weight: 1000g. Operating Temperature and Humidity: -20°C ~
+60°C, <95% RH. (Non-condensing).
Storage Temperature and Humidity: -10 to
60°C, <70% RH. (Non-condensing).
Consumption Current:700mA. Standard Accessories:User’s manual, 9V battery
AC to DC Adapter, Carrying case,LCD Wall Mount.
TM-185D Accessories:
Datalogging Box. (Datalogging capacity with
Memory Size30,000 data sets). MINI USB 4P(MALE) to USB A Type cable. Install CD disk of TM-185D .
3 General Specifications
Page 5
TM-185/TM-185D
4
Measurement Range:
Humidity: 5%~95%. Temprature:-20.0°C60.0°C
/-4.0°F140.0°F.
Temperature Accuracy:±1.0°C/±1.8°F (-5.0°C
~40.0°C /23°F~104°F)other ±2.0°C /3.6°F.
Humidity Accuracy:
±2.5%RH(at 25°C,10%RH~90%RH). ±5.0%RH(at 25°C,<10%RH,>90%RH).
Note1: If you require more accurate we can do it
according to your requirement.
Note2: When the AC adaptor is off will start the
backup power only to provide Data Logger use, the LED wont be lightened.
Note3: Datalogging operation instruction at
Install CD disk of TM-185D.
4 Electrical Specification
Page 6
TM-185/TM-185D
5
5 Instrument description
Page 7
6
TM-185/TM-185D
Page 8
TM-185/TM-185D
7
1. Temperature & Humidity sensor
2. Time Set Button
3. Alarm Set Button
4. °C /°F Set Button/Down Button
5. Enter Button
6. Hold Button/UP Button
7. LED Display
8. Data Logger Box (optional)
9. Data Logger Box interface (-)
10. Power Switch
11. Terminal Blocks (Always off)
12. Audio Output
13. Lo Battery LED Display
14. External power DC 9V
15. ARM Hold
16. Battery Cover.
17. Record Button
18. Memory Button
19. USB interface
20. Data Logger Box interface (+)
Page 9
8
6 LED description
1. Temperature reading value
2. Humidity reading value
3. Time reading value
4. Time unit
5. Data logger symbol
6. Temperature unit
7. Alarm symbol
8. HI symbol
9. LO symbol
10. Humidity unit
TM-185/TM-185D
Page 10
TM-185/TM-185D
9
7 Clock and Date setup
Push button into time clock and date
setup.
Push button to select option to adjust Push or button to change the
digit.
Push button to store the setup then
exit the mode, If not push button then exit clock and data setup.
This meter clock to introduce 12 hour time, if
you want change 24 hours, push .
Page 11
TM-185/TM-185D
10
8 Alarm Range Setup
Push button into alarm setup LED will
display ALARM ONIf you just reading alarm setup please push button again LED will display ALARM OFF.
First push button into alarm range
setupLED display show LO ALARM ON °C or °F ,and set LOW alarm Temperature, Push or button to change the digit.
Second push button into alarm range
setup,LED display show HI ALARM ON °C or
°F”,and set Hi alarm Temperature, Push or button to change the digit.
Third push button into alarm rang
setup, LED display slow LO ALARM ON %RH”, and set Hi alarm Humidity, Push or button to change the digit.
Fourth push button into alarm range
setup, LED display show HI ALARM ON %RH”, and set Hi alarm Humidity, Push or button to change the digit.
If you want to store this setup push the
button, If not push button then exit alarm setup.
Page 12
TM-185/TM-185D
11
If test Temperature or Humidity exceed set
range, buzzer were bleep, at this time socket output sin wave signal at 2000Hz,terminal blocks were short.
Page 13
TM-185/TM-185D
12
Turn off the instrument.
Open the battery cover and remove the battery.
Replace with four-9V NEDA 1604, IEC 6F22 or
JIS 006P size battery.
Install the battery cover.
WARNING
If the LO Battery LED were turn on, please replace the battery immediately
9 Battery replacement
Page 14
TM-185/TM-185D
13
10 External DC Power
External AC to DC adapter: Voltage
9VDC(8~14VDCMax)
Socketpin Positive, Ground Casing External Diameter 6.3mm internal Diameter 2.0 mm
11 Safety Precaution
For cleaning the instrument use a soft dry cloth.
Never use a wet cloth, solvents or water, etc.
Datalogging capacity with Memory Size500 data sets
Operation Altitude: Up to 2000M.
Operating Environment: Indoors use. This instrument
has been designed for being used in an environment of pollution degree 2.
Page 15
TM-185/TM-185D
14
12 Software installation (TM-185D)
Please put the CD in the PC that will be connected to this meter.
Please select the USB driver that will be installed, such as E:\TM-185D\PL-2303 Driver Installer.exe (windows 2000 SP4/windows XP SP2), click twice on the left key of the mouse to install the USB driver.
Page 16
TM-185/TM-185D
15
Select the SETUP.EXE i.e., E:\TM-185D\SETUP.EXE and installs the desktop icon
Tack out the CD from PC after completed the installation of the desktop icon.
Page 17
TM-185/TM-185D
16
Use the USB cable to connect the meter and computer according to the drawing.
Select the desktop icon (TM-185) and click twice on left key of the mouse to run the procedure.
Page 18
TM-185/TM-185D
17
13 Maintenance
Do not use the meter in an environment with severe
change; do not store the unit in an environment with high temperature, high humidity, and high vibration
Take battery off if the meter has not been used for a
long period of time
The Diode temperature probe is used to measure
temperature and the capacitive humidity sensor is used to measure take humidity
The thermocouple and humidity sensor will start aging
under the influence of oxidation, reduction, corrosion, pollution, vaporization, diffusion or other metallurgy. The aging process will affect its precision seriously.
Cleaning and inspection of temperature probe:
The smoke, coal, dust, grease attached on the
protective tube of temperature probe will slow down the heat conduction of the thermocouple and cause measuring error. Therefore, it should be cleaned periodically. The metal coating of thin thermocouple should be replaced properly upon the occurrence of corrosion.
Cleaning and inspection of humidity sensor:
The smoke and dust attached on the humidity sensor
will slow down the function of humidity sensor and cause measuring error. Therefore, it should be cleaned periodically. Blow off the dust with mild compressed air instead of water or alcohol. The aluminum plate inside the humidity sensor should be replaced upon the occurrence of corrosion.
Page 19
TM-185/TM-185D
18
Caution: this symbol indicates that equipment and its accessories shall be subject to a separate collection and correct disposal
14 End of life
15 Auto Stores Data
Push button into time clock and
date setup.
Push button again into auto stores
time clock setup.
This time LED display show auto stores
data timeyou can push to change time unit (Max auto stores data time :23hour 59min 59sec)
Push or button to change
the digit.
If you want to store this setup push
button, If not push button then exit auto stores data time setup.
If stores data will be writhing into memory
LED display show Record on
Page 20
TM-185/TM-185D
19
16 Single data memory
Push Record button (in TM-185D) each
time to store the display reading and memory location in memory.
17 Viewing logged Data
Push Memory button (in TM-185D) into
view logged data modepush Memory button again exit view logged data mode
In view logged data mode push
button reading stores time push button to change the temp unitPush
or button to change view logged data
Page 21
TENMARS ELECTRONICS CO., LTD 6F, 586, RUI GUANG ROAD, NEIHU,
TAIPEI 114, TAIWAN.
E-mail: service@tenmars.com
http://www.tenmars.com