Tenmars ST-130, ST-130S User Manual

Page 1
Noise Dose Meter
ST-130/ST-130S
User’s Manual
Page 2
Page 3
Contents:
1.SAFETY PRECAUTIONS ......................................................................................................... 3
1.1. Preliminary Description ..................................................................................................... 3
1.2. Note................................................................................................................................... 3
2.PREPARATION FOR USE ....................................................................................................... 4
2.1. Initial.................................................................................................................................. 4
2.2. Calibration ......................................................................................................................... 4
2.3. Storage.............................................................................................................................. 4
3.INSTRUMENT INSTRUCTIONS............................................................................................... 5
3.1. Instrument Description ...................................................................................................... 5
3.1.1.Controls Description ................................................................................................................... 5
3.1.2.Display Description..................................................................................................................... 6
3.1.3.ST-130S Microphone.................................................................................................................. 7
3.1.4.Input interface............................................................................................................................. 7
3.2. Noise Does Measurement Procedures ............................................................................. 8
3.3. Sound Level Measurement Procedures ............................................................................ 8
3.4. Auto Data Record.............................................................................................................. 9
3.5. Single Data Record ......................................................................................................... 10
3.6. Viewing Logged Reading ................................................................................................ 10
3.7. Set Mode......................................................................................................................... 10
3.7.1.Test Mode Set ......................................................................................................................... 11
3.7.2.Auto Power Off Set................................................................................................................... 11
3.7.3.Sampling Time & Auto Record Set........................................................................................... 11
3.7.4.Real Time Clock Set................................................................................................................. 12
3.7.5.94dB Offset Adjust Set ............................................................................................................. 12
3.7.6.Noise Standard Set .................................................................................................................. 12
3.7.7.SLM Function Set .................................................................................................................... 13
4.SOFTWARE............................................................................................................................ 14
4.1. System Requirements ..................................................................................................... 14
4.2. Software Description ....................................................................................................... 14
4.3. Tools Description............................................................................................................. 15
4.4. Open File......................................................................................................................... 15
4.4.1.Zoom Function.......................................................................................................................... 15
4.4.2.Drag function ............................................................................................................................ 15
4.5. Option.............................................................................................................................. 16
4.5.1.PC Setting................................................................................................................................. 16
4.5.2.Meter Setting ............................................................................................................................ 16
4.6. Print Noise Dose Report.................................................................................................. 17
4.7. Sound level chart & Noise dose chart (LN %) ................................................................. 17
4.8. Enable PC data logger .................................................................................................... 18
4.9. Dowload Record data...................................................................................................... 18
4.10.Erase Record data .......................................................................................................... 18
5.EXPLAINATION ...................................................................................................................... 19
5.1. 1. Measurement Parameters:.......................................................................................... 19
5.2. A, C, Z Weighting Instruction:.......................................................................................... 20
6.MAINTENANCE ...................................................................................................................... 21
6.1. General Information......................................................................................................... 21
6.2. Battery Replacement....................................................................................................... 21
6.3. Cleaning .......................................................................................................................... 21
6.4. End of Life ....................................................................................................................... 21
7.TECHNICAL SPECIFICATIONS............................................................................................. 22
7.1. Feature............................................................................................................................ 22
7.2. Environment .................................................................................................................... 23
7.2.1.Environmental Conditions......................................................................................................... 23
7.2.2.EMC.......................................................................................................................................... 23
7.3. Accessories.....................................................................................................................23
Page 4
7.3.1.Standard Accessories............................................................................................................... 23
8.SERVICE ................................................................................................................................ 24
8.1. Warranty Conditions........................................................................................................ 24
8.2. Service ............................................................................................................................ 24
EN - 2
Page 5
1. SAFETY PRECAUTIONS
When taking measurements: Avoid doing measurements in humid or wet places - make sure that humidity is
within the limits indicated in section “environmental conditions”.
Avoid doing measurements in presence of explosive gas, combustible gas,
steam or excessive dust.
The following symbols are used:
Caution: refer to the user’s manual. An incorrect use may damage the
tester or its components
The instrument conforms to the CE standard
1.1. Preliminary Description
The ST-130/ST-130s is designed to test noise exposure in accordance with OSHA,MSHA,DOS,ACGIH, and ISO standards.
Fast and easy on-site surveys help determine noise reduction requirements.
The meter can also be used in SLM(sound level meter)mode. The SLM mode has a datalogging feature that can record up to 1000K readings which can be downloand to PC for analysis. The built-in USB interface to connect PC.
Applications: Evaluation of environmental noise, Measurements of noise at
workplaces, Assessment of product noise.
1.2. Note
Does not observe the warning and/or operation instruction, it’s
possible to damage the instrument either its components or the
Do not operate the instrument at temperature and humidity environment
beyond to reference conditions of chapter 7.2.1.
Keep the microphone dry and avoid severe vibration. Wind blowing across the microphone would bring additional extraneous
noise. Once using the instrument in the presence of wind, it must mount the windscreen to prevent the undesirable signals.
CAUTION
operator
EN - 3
Page 6
2. PREPARATION FOR USE
2.1. Initial
The instrument has been checked mechanically and electrically prior to shipment. Take care to ensure the instrument reaches you undamaged.
However, it is wise to carry out a rapid check in order to detect any possible damage that may cause during transport. If it’s damage, claims to the dealer immediately.
Check the packaging content according to packing list reported in 7.3.1 chapter .In case of discrepancies, contact the dealer immediately.
In the event of re-shipment of the instrument please follow the instructions reported in chapter 7.3.1.
2.2. Calibration
The instrument complies with the technical specifications contained in this manual and such compliance is guaranteed for 1 year. The instrument is maybe need recalibration after one year.
2.3. Storage
After a period of storage in extreme environmental conditions exceeding the limits mentioned in paragraph 7.2.1 let the instrument return to normal measuring conditions before using it.
EN - 4
Page 7
3. INSTRUMENT INSTRUCTIONS
3.1. Instrument Description
3.1.1. Controls Description
1. Microphone
2. Microphone Cable
3. Microphone Connector(Male)
4. Microphone Connector(Female)
5. LCD Display
6. record data / Reading record data button
7. Power ON/OFF Button
8. UP Button
9. Right Button
10. Enter Button
11. Down Button
12. Left Button
13. SET Button
14. Stop Button
15. Start /Pause Button
16. Belt clip
17. Battery Cover
18. USB Interface
EN - 5
Page 8
3.1.2. Display Description
Pause
Battery Status
Stop
Start
SPL High dB Range (140~70)
Auto Power Off Enable
115dB
indicator(SPL)
140 dB
indicator(Peak) USB Interface
dB Display
SPL Middle dB Range (110~50)
SPL Low dB Range (90~30) Noise Dose %
A Weighting
C Weighting
Z Weighting
dB test data > Hi level
Solid: Auto record standby: Flashing
Testing mode
Sound Level Meter mode Sound Noise dB
Fast Weighting
Slow Weighting
Impulse Weighting
dB test data < Low level Start test time
recording Visit record data
SET mode turn on
Stop test time
Test duration
EN - 6
Page 9
3.1.3. ST-130S Microphone
Diameter : 1/2 inch Polarization voltage : 0V Dynamic range : 25dBA ~140dB Sensitivity : -32±3dB (250Hz 0dB=1V/Pa) Free field frequency response : ±2dB(25Hz12.5kHz)
Frequency (KHz) Deviation of pressure
0.25 0.0 1 -0.1 2 -0.5 3 -0.6 4 -0.9 5 -1.2 6 -1.7 7 -2.2 8 -2.8 9 -3.3
10 -4.1
12.5 -6.0
3.1.4. Input interface
The front is PLT84RFR, the signal input receptacle.
ST-130
Pin 1 GND Pin 2 Power Pin 3 NC Pin 4 NC
ST-130S
Pin 1 GND Pin 2 Power(+) Pin 3 N.C Pin 4 Power(-)
EN - 7
Page 10
3.2. Noise Does Measurement Procedures
Press turn on meter
LCD do not display
Press
turn on noise dose testing. ,press again pause
symbol.
testing.
Press
select testing information
SPL→Dose%LPMAXLPMINPKMAXLEQSELLEP8TWA8
LVAGLN%
Press
to observe testing date information. Press change
date display information.
Time(hh:mm:ss)→date(YY-DD-MM)
Press
stop testing and store all test data
3.3. Sound Level Measurement Procedures
Press turn on meter
EN - 8
Page 11
LCD show symbol
Press
button to select test function.
SPL→LeqSELPeakMAX
Press
testing , Press again pause testing
If test data bigger than HI test range at the same time, LCD is showing
If test data is smaller than Low test range at the same time, LCD is
showing
Leq integral time setting is along with the same sampling time.
When the sampling time is set to zero, the integration time until the user
exits
Press
stop testing
CAUTION
Wind blowing across the microphone would bring additional extraneous noise. Once using the instrument in the presence of wind with speed higher than 10m/s, it must mount the windscreen to prevent the undesirable signals. Keep the microphone dry and avoid severe vibration.
3.4. Auto Data Record
Press enabled Auto record function.
LCD The bottom left of LCD display ”Write” ,
memory.
EN - 9
symbol will be flash.
this mean the data will be writing to
Page 12
The bottom left of LCD display “FULL”, this mean the data will be full. Auto record function can not use menu record.
3.5. Single Data Record
Press each time to store the display reading and symbol flash. The bottom left of LCD display ”Write” ,
this mean the data will be writing to
memory.
3.6. Viewing Logged Reading
Press button more than 1 sec into the viewing logged reading mode.
Press
Press
Test modelStart timeTest duration time Total pause timeTest end time
Press
or to scroll through the readings.
select dose record information. (Nosie Dose Meter mode)
to change data or date, Press change Time.
(hh:mm:ssYY-MM-DD)
Press
more than 1 sec again to exit viewing logged reading mode.
3.7. Set Mode
Press into set mode , can set 7 functions in set mode
Press
exit set mode
Test Mode→Auto Power OffSampling Time & Auto RecordReal
Time Clock94dB Offset AdjustNoise standardSLM Fuunction
Press
again,save set and into next set mode
EN - 10
Page 13
3.7.1. Test Mode Set
Press or , change test mode. (NDM SLM)
NDM: Noise Dose Meter SLM:Sound Level Meter
3.7.2. Auto Power Off Set
Press or , enable or disable Auto Power Off function
3.7.3. Sampling Time & Auto Record Set
Press or , select auto records set or sampling time.
Press
or enable or disable Auto record, adjust sampling time.
Minimum sampling time: 1 second; Maximum sampling time: 23 hours 59
minutes 59 seconds
EN - 11
Page 14
3.7.4. Real Time Clock Set
Press or , select option to adjust.
Press
or , adjust time digit.
3.7.5. 94dB Offset Adjust Set
Press , auto run 94dB offset adjust.
Press
Press
or , change frequency weighted
or , adjust offset
3.7.6. Noise Standard Set
Press or ,select NDM testing law
OSHA→MSHSDODACGIHISO85ISO90USER
EN - 12
Page 15
3.7.7. SLM Function Set
Press or , change test function
Peak C Peak Z
Press
or ,select test function
Sound level range→Time weightedFrequency weightedPeak frequency
EN - 13
Page 16
4. SOFTWARE
4.1. System Requirements
Operating: Windows® XP/Windows Vista/Windows 7/Windows 8  Storage:200 MB of available hard disk space.  Processor: Intel Pentium® 4  Memory:1 GB RAM (XP), 1.5 GB (Windows Vista/Windows 7/Windows 8)  Other:
4.2. Software Description
PL2303 Windows Drive ;Microsoft .NET Framework 3.5 Service Pack 1
EN - 14
Page 17
4.3. Tools Description
Open file
Print noise dose report (*.ndr) Noise dose chart (LN%)
Stop to log sound level
Options
Sound level chart
Start to log sound level (*.csv)
Download sound level logs;
Doenload noise does report Erase measured data that stored in meter
Automatic detect port which
connect with meter
4.4. Open File
Sound Level Log List Tool
SPLLEQSELPeakMax
FastSlowImpluse
ACZ
Sound Level Logs Graph
Calculate LEQ & SEL Save file as…. Graph
Zoom Drag
Print Print Preview
Print Setup
4.4.1. Zoom Function
Click the left mouse button to zoom in Click the left mouse button to original size Hold the left mouse button to moving selection range Keyboard '+', '-' can zoom in or zoom out
4.4.2. Drag function
Hold the left mouse
button can moving view range
Hold CTRL+ left mouse button can be change view range and zoom
EN - 15
Page 18
4.5. Option
4.5.1. PC Setting
4.5.2. Meter Setting
PC Setting
Meter Setting
PC data logger sample time.
Click the left mouse button to change the color box Graph (Y-axis) sound level range
Serial number(0~9;A~Z)16 word
Alarm setting
dB Offset setting MAX:+3.9dB MIN:-3.9dB
Setting meter real time clock
Enable/disable auto Record. Setting sample time
Enable/disable auto play noise dose function. Meter must be turn on and setting in NDM(noise does meter) mode
Select noise dose standard
Select noise dose play time. Default standard(8hour), Other 5;10;15;30min
12 48101224 hour
Setting information import or export ,file
EN - 16
format(*.ncg)
Page 19
4.6. Print Noise Dose Report
Select noise dose logger report (*.ndr)
Click
or and key in user information , click “ok” ,the output report
4.7. Sound level chart & Noise dose chart (LN %)
Sound level char. X-axis is the time coordinate.Y-axis is the sound coordinate.  Noise does chart. X-axis is the LN%. Y-axis is the sound coordinate.
EN - 17
Page 20
4.8. Enable PC data logger
Click enable data logger.
Create a new log file (*. csv) in the hard disk.  Click
stop data logger.
4.9. Dowload Record data
Click
Download 100K record data. it takes about 15 minutes time-consuming, please be
patiently.
download sound level logs or noise dose report on the meter memory.
4.10. Erase Record data
Click
delete all records of the meter.
EN - 18
Page 21
5. EXPLAINATION
5.1. 1. MEASUREMENT PARAMETERS:
Test Function Screen parameter Explaination
SPL LAFp Sound pressure level (SPL)
SPL LASp Sound pressure level (SPL)
SPL LCFp Sound pressure level (SPL)
SPL LCSp Sound pressure level (SPL)
SPL LZFp Sound pressure level (SPL)
SPL LZSp Sound pressure level (SPL)
Equivalent continuous level for the
Leq LAFq
Leq LCFq
duration of the measurement for A weighting
Equivalent continuous level for the duration of the measurement for C weighting
Equivalent continuous level for the
Leq LZFq
duration of the measurement for Z weighting
Frequency weighted sound exposure
SEL LAE
level for the duration of the measuremen for A weighting
Frequency weighted sound exposure
SEL LCE
level for the duration of the measuremen for C weighting
Frequency weighted sound exposure
SEL LZE
level for the duration of the measuremen for A weighting
Peak Lcpeak Instantaneous C peak level
EN - 19
Page 22
5.2. A, C, Z WEIGHTING INSTRUCTION:
1. : The A weighting curve is based on 40 Phon Fletcher-Munson Equal Loudness Contour, Noise assessment in human, suggest to use the A weighting.
2. The C weighting in essentially is approximate smooth. With labor safety concern, suggest using the C weighting.
3. The Z weighting for the electric instrument interior not the linear signal which processes after the filter, suits in wants to output AC or the DC signal does other research to use.
The Z weighting is a linear signal which is not processed through the filter.
It’s suitable to output AC or DC signal for research.
4. Sound Level Meter Class Description:
Class 0: use in the laboratory reference standard.  Class 1: laboratory or field use.  Class 2: laboratory or field use.  Class 3: general field use.
EN - 20
Page 23
6. MAINTENANCE
6.1. GENERAL INFORMATION
This is a precision instrument. To guarantee its performance be sure to use it or keep it stored on suitable environmental conditions. Do not expose it to high temperatures or humidity or direct sunlight. Be sure to turn the power off after use. If you expect not to use the instrument for a long period remove batteries to avoid leakages of battery liquid which could damage the its inner components.
6.2. BATTERY REPLACEMENT
The low battery “ ” indication is displayed; the battery should be replaced.
Turn off the instrument.  Remove the battery cover.  Remove all the batteries from the battery holder.  Insert four new batteries of the same type respecting the polarity signs.  Install the battery cover.  Please follow the local laws and regulations to process the waste battery.
6.3. CLEANING
To clean the instruments use a soft dry cloth. Never use a wet cloth, solvents or wate.
6.4. END OF LIFE
Caution: this symbol indicates that equipment and its accessories shall be
subject to a separate collection and correct disposal.
EN - 21
Page 24
7. TECHNICAL SPECIFICATIONS
7.1. FEATURE
Environmental conditions: temperature 23 ± 5, relative humidity < 80%.
Display Single LCD MAX reading 999999
Display Refresh Rate 1 Time/sec
Standards IEC 61252-1993
IEC 61672-1-2003
ANSI S1,25-1992
ANSI S1,4-1983
ANSI S1,43-1997
Microphone(ST-130S) 1/2” pre-polarized condenser microphone build in
preamplifier: 1V/Pa@250HZ, frequency range: 20
Hz~12.5 kHz, Thermal noise: <25 dB(A)
Microphone(ST-130) 1/2 inch Electret condenser microphone
Measurement Items(NDM)
Does%,Lxyp,Lxmax,Lxmin,Lxeq,SEL(LAE),PeakMAX,L
AVG,TWA,LEP,LN%
Measurement Items(SLM) Lxyp,Lxmax,Lxmin,Lxeq,SEL(LAE),PeakMAX
Display Range
Primary RMS Range @1KHZ
Maximum Peak C
30dB to 90dB (L)
50dB to 110dB (M)
70dB to 140dB (H)
41dB to 86dB (L)
55dB to 106dB (M)
75dB to 125dB (H)
90~143 dB Weighting Sound Level Measurement
Dynamic Range 60 dB
Accuracy ±1.4dB@94dB /1KHZ
Internal memory
MAX Datalogger data10000(NDM);999999(SLM)
Time Weighting Fast, Slow, Impulse
Frequency Weighting A/C/Z
Frequency Range 20Hz~8KHz
Starting Time
10 Second
Battery Life(ST-130) 24 hours ( 9V×1 battery Alkaline )
Battery Life(ST-130S) 20 hours ( 9V×1 battery Alkaline )
Dimensions 107(L) x 65(W) x 33(H) mm
EN - 22
Page 25
7.2. ENVIRONMENT
7.2.1. Environmental Conditions
For inside use, max height: 2000m Reference temperature: 23° ± 5
Operation temperature: 5 ~ 40 Operation humidity: <80% RH
Storage temperature -10 ~ 60 Storage humidity <70%
7.2.2. EMC
This instrument was designed in accordance with EMC Standards in force and its compatibility has been tested in accordance with EN61326-2 (2006).
7.3. ACCESSORIES
7.3.1. Standard Accessories
Noise Dose Meter. User’s manual. Carrying case. 9 V battery NEDA 1604 IEC 6F22 or JIS 006P.
Microphone diameter windscreen.  Software.  MINI USB Cable (Mini B type).
EN - 23
Page 26
8. SERVICE
8.1. WARRANTY CONDITIONS
This instrument is guaranteed for one year against material or production defects, in accordance with our general sales conditions. During the warranty period the manufacturer reserves the right to decide either to repair or replace the product. Should you need for any reason to return back the instrument for repair or replacement take prior agreements with the local distributor from whom you bought it. Do not forget to enclose a report describing the reasons for returning (detected fault). Use only original packaging. Any damage occurred in transit due to non­original packaging will be charged anyhow to the customer.
The warranty doesn’t apply to: Accessories and batteries (not covered by warranty) Repairs made necessary by improper use (including adaptation to particular applications not foreseen in the instructions manual) or improper combination with incompatible accessories or equipment. Repairs made necessary by improper shipping material causing damages in transit. Repairs made necessary by previous attempts for repair carried out by non-skilled or unauthorized personnel. Instruments for whatever reason modified by the customer himself without explicit authorization of our Technical Dept.
The contents of this manual may not be reproduced in any form whatsoever without the manufacturer’s authorization.
Our products are patented. The logotypes are registered. We reserve the right to modify characteristics and prices as part of technological developments which might require them.
8.2. SERVICE
Shouldn’t the instrument work properly, before contacting your distributor make sure that batteries are correctly installed and working, check the test leads and replace them if necessary.
EN - 24
Page 27
TENMARS ELECTRONICS CO., LTD
6F, 586, RUI GUANG ROAD, NEIHU,
TAIPEI 114, TAIWAN.
E-mail: service@tenmars.com
http://www.tenmars.com
Loading...