5.1 Power
Place the calibrator on a flat surface with at least 6 inches of free space around the instrument.
Overhead clearance is required. Do not place under a cabinet or structure. Insert equilizing block
into the well with help of tong, after inserting remove the tong from equalizing block. Plug the
power cord into mains outlet of the proper voltage, frequency Typically this will be (230 VAC ±10,
50/60 Hz). Turn the “Mains” (Heater) switch located on front panel. The controller dispaly should
illuminate after 3-4 sec . After a brief self test the controller should begin normal operation. If the
unit fails to operate please check the power connection.
5.2 Setting the Temperature
Press “UP” or “DOWN” buttons to change the change the temperature set-point value. When the
set-point temperature is changed the controller will switch the CALsys 1200 ON or OFF to raise or
lower the temperature. The displayed temperature will gradually change until it reaches the setpoint
temperature. The temperature bath requires 15 to 20 minutes to reach the set-point depending on
the span. Another 10 to 15 minutes is required to stabilize the temperature bath within ± 0.5ºC of
the set-point.
5.3 Operating Instruction:-
1. Insert the inconel block into well of furnace.
2. Connect the ‘CALsys 1200’ to a suitable power supply & Switch ON the Dry bath source and set
the desire temperature value in PID by using UP & Down key.
3. Insert the reference (Master) temperature sensor and UUC (Unit under calibration) into the
thermowell (inconel block).
4. PV (Present value) display in controller will gradually rise until it reaches the set point
temperature. The controller takes some times to reach the set-point depending on the span.
Furnace is stable when PV is equal to SV (Set Value).
5. Master sensor take some times to reach the setpoint temperature and stable at temperature
near about controller set temperature.
6. When temperature of the master and UUC (Unit under calibration) are stable record the
readings of master sensor.
7. Compare the UUC reading with the master’s reading & find out the error by comparison method.
8. Reset the controller and / or repeat the calibration for another calibration point or for another
sensor.
9. When the calibration is complete, reset the controller to 0°C & wait until the unit has cooled to
below 100°C, before moving the ‘CALsys 1200 to new location the ‘CALsys 1200’ must be cooled
below 100°C before it can be put back into its carrying case.