Tel-Tru Check-Set Specifications

Page 1
TEL-TRU® Manufacturing Co.
CHECK-SET® THERMOMETER CALIBRATORS
An accurate, safe, convenient, and cost­effective thermometer calibration system
www.teltru.co mwww.teltru.co m
Page 2
Verify Thermometer Accuracy at Your Critical Control Point!
The Tel-Tru Check-Set Series of Thermometer Calibrators provides accurate, unambiguous, and cost-effective calibration and accuracy verification for a wide range of thermometers and temperature measurement instruments.
The Check-Set series is the perfect temperature reference device for:
Plant quality assurance
Process control
Food safety
Instrument calibration and service
HACCP compliance (Hazard Analysis Critical
Control Points)
The Check-Set is easy to use, reliable, accurate and compatible with a wide range of:
Dial and digital thermometers
Filled system bulbs
Electronic temperature sensors
Typical users of the Tel-Tru Check-Set Calibrators are food safety professionals responsible for operating, inspecting or certifying:
Food preparation and food service facilities
Industrial canning, meat, dairy and beverage
processing plants
Users include:
- HACCP compliance officers
- QA managers
- In-plant maintenance technicians
- Field calibration contractors
- Local, state, and federal agency compliance officers
FEATURES
The series includes:
- Check-Set I with one calibration set point
- Check-Set II with two calibration set points
- Check-Set IV with four calibration set points
High accuracy
- Certificate of Calibration traceable to NIST
- Built-in warning circuitry to alert user when unit exceeds set-point by 1°F
Fast calibration verifications
- Simple to use
- Easy operator training
Serial number assigned to each unit
Portable - optional padded shoulder strap
available (except CS1-HOT models)
Rugged enclosure and electronics for reliable
service
Quick Start Guide
Compatible with thermometers and temperature probes of all sizes!
2
www.teltru.com
CALIBRATION SERVICE
An annual factory recertification service is
available.
Completion of recertification service extends the
warranty one additional year, for up to nine years.
Page 3
Check-Set Accuracy ±0.2ºF (0.1°C)
GENERAL SPECIFICATIONS
Accuracy ±0.2°F (0.1°C)
Input Power 115 volts, 50/60 Hz, 1.5 amps
Recommended Ambient Operating temperature range:
- 65-85°F (18-29°C)
Type 304 stainless steel enclosure
Weight:
- All models: 9 lbs. (4 kg)
Dimensions:
- Check-Set I HOT: 11 x 65⁄8 x 31⁄8" (279 x 168 x 79mm)
- All other models: 83⁄4 x 87⁄8 x 41⁄4" (222 x 225 x 108mm)
US patent #6,170,983
IMPORTANT CALIBRATION TEMPERATURES
The Check-Set series of thermometer calibrators verifies the calibration of your temperature sensing devices at multiple temperatures. Their accuracy also allows you to validate your temperature measuring devices at temperatures specific to your application. For example, in the food and dairy industry, important temperatures or “set-points” include:
40°F (5°C): Recommended cold food holding
temperature
140°F: Recommended hot food holding
temperature
150°F: Clean-in-Place return lines
160°F: E.Coli safety temperature
161°F: HTST flash pasteurization - milk
175°F: HTST flash pasteurization - ice cream,
eggnog, frozen desserts
212°F: Boiling point of water
Dial Adjustment Tool makes calibrations easy! (sold separately)
“READY TO USE” TIMES
Users often keep the Check-Set units on all day so they are ready for use “on demand.” Typical “ready to use” times from ambient:
Model Calibration
Temperature
Check-Set I HOT
Check-Set I COLD
Check-Set II 40°F 5 min. ±0.03°F
Check-Set IV 40°F 6 min. ±0.03°F
140°F 10 min. ±0.06°F
160°F 14 min. ±0.06°F
212°F 20 min. ±0.06°F
40°F 5 min. ±0.03°F
160°F 7 min. ±0.03°F
5°C 5 min. ±0.02°C
90°C 12 min. ±0.02°C
150°F 6 min. ±0.03°F
161°F 7 min. ±0.03°F
175°F 9 min. ±0.03°F
Ready to
Use Time
Stability
Eliminate Error-Prone and Cumbersome Calibration Methods
The Ice Water Method
Difficult
A “pain”
Time consuming
Error-prone
The Boiling Water Method
Dangerous
Time consuming
Difficult
Error-prone -- boiling
point of water varies with altitude
www.teltru.com
3
Page 4
Markets and Applications Served
Commercial Food Service
Chain Restaurants
Grocery Chains
Corporations
Airlines
Hotels
Restaurants
Theme Parks
Sports Arenas
Institutional Food Service
Military Bases
Public School Districts
Colleges and Universities
Hospitals
Government Medical Centers
Dept. of Corrections
Food Processing, Storage, and Distribution
Red Meats
Poultry
Seafood
Sausage and Hot Dogs
Baked Goods
Dairy Products
Grains
Tomatoes
Pickles
Corn Syrup
Produce
Wholesale Food Distribution
Training and Education
Chef Training/Culinary Arts School
HACCP and Food Safety Training
Environmental Safety and Health Professionals
City Health Dept.
County Dept. of Environmental Health
State Dept. of Health
Federal Agencies
Military Food Safety Inspectors
Industrial/OEMs
Research and Development Lab
Industrial Laboratory
Service and Calibration of:
– Refrigeration Equipment – Industrial Ovens – Smokers for Meat/Poultry
Manufacturer of Food and
Beverage Testing Equipment
FOOD SAFETY/
HACCP
Background
HACCP (Hazard Analysis Critical Control Point) is a systematic
preventative approach to food safety that involves the definition of specific measurements to reduce risk. HACCP techniques focus on the prevention of hazards and not upon final inspection. HACCP principles are widely applied to FDA and USDA regulated industries and a growing list of international industries, government agencies, and organizations.
Requirements
Satisfactory monitoring of CCPs requires accurate and precise measurements of parameter values (e.g. temperature, pressure, pH, etc.) to ensure that defined Critical Limits are met. Equipment used to make those measurements must be properly calibrated.
The HACCP plan must describe instrument calibration procedures and frequency, responsible individuals and required documentation records for instruments used measuring those parameter values.
Check-Set Series Delivers
Check-Set Calibrators are designed for Food Safety Professionals as an easy-to-use tool to verify
the calibration of temperature measurement instruments in compliance with HACCP requirements.
Check-Set users can check thermometers and other temperature sensing instruments at temperatures equal to or close to the critical control points of their processes.
4
www.teltru.com
Page 5
Markets and Applications Served
DAIRY
Background
The ”Pasteurized Milk Ordinance” (“PMO”) regulates the production,
transportation, processing, handling, sampling, examination, labeling, and sale of all “Grade A” milk and milk products, as well as the inspection of dairy farms, milk plants, receiving stations, transfer stations, milk tank truck cleaning facilities, milk tank trucks and bulk milk hauler/samplers.
Requirements
The PMO defines the performance requirements and testing methods for test, indicating, and recording thermometers, and other temperature instruments. The PMO requires that temperature instruments be “compared to a certified temperature source” upon installation and every 3 months thereafter, and after equipment is altered, serviced, repaired or replaced.
Check-Set Series Delivers
The Tel-Tru Check-Set IV Calibrator is optimized for
compliance with the temperature instrument calibration and verification requirements of the dairy industry
cheese processing plants.
Typical users include in-plant staff responsible for QA, food safety, HACCP compliance and instrument maintenance. Other users include field calibration contractors and regulatory agency personnel who inspect and certify these plants.
Enjoy less plant down time due to faster and easier certification of temperature instruments
No ice water, no ice buckets
No temperature controller circulators
Easy to carry throughout the plant
Probe inserts for most probe sizes
including milk, ice cream, and
FOOD PROCESSING/
RETORTS
Background
Thermal processing of food in various types of pressure retorts is regulated by
the Food and Drug Administration through the establishment of rules for equipment, and temperature indicating, recording and controlling devices.
A major food processing industry trade association also recommends voluntary guidelines for food processors to test and document the temperature distribution within their retorts.
Requirements
FDA Rule 21 CFR 113.40 specifies that each retort must be equipped with at least one temperature­indicating device, a temperature recording device, and a temperature controller. These instruments must be checked with a reference device that is directly traceable to National Institute of Standards and Testing (NIST) “to ensure accuracy during processing.”
FDA Rule 21 CFR 113.100 defines the required record keeping for reference devices including: identification of the device, a Certificate of Calibration, date and results of the test, and date of the next test.
The Institute for Thermal Processing (IFTPS) recommends voluntary guidelines for operators of retorts in food processing plants to test the thermal efficiency of their retort equipment:
All temperature instruments conform to applicable
Federal Regulations and are tested against an accurate and traceable reference device.
A calibration process is recommended before and
after testing to assure that the temperature indicators and sensors on the retort and on an external monitoring system are consistent and accurate.
Check-Set Series Delivers
Tel-Tru Check-Set Calibrators are accurate and
certified temperature reference devices that meet the requirements of the FDA and the recommendations of the IFTPS
accuracy of temperature instruments in food processing plants. Tel-Tru Check-Set features include a unique serial number, a certificate of accuracy traceable to NIST, and available annual recertification and maintenance service.
for verifying the
www.teltru.com
5
Page 6
Do you need to verify hot temperatures only?
Drawing A
(CS1-HOT)
What temperature(s) do you
want to verify?
4.0" Insulated Zone 4.0" Te mp. Control Zone
PROBE SENSING AREA
What are the stem diameters of the
instruments you are checking?
STEm-LENGTH COmPATIbILITY
For best results, the sensing area of the probe must be inserted completely into the Temperature Control Zone. The minimum immersion depth is dependent on the thermometer type. Check the
manufacturer’s user manual of your thermometer for the correct
immersion length.
CS1 w/ single hole configuration—Code 100
Frequently used in food preparation or food processing plant environments to verify accuracy of digital or dial thermometers at a single hot test point. Built-in wrench is for easy calibration adjustment of spot check or meat cooking dial thermometers.
CS1 w/ multi-hole configuration—Code 400
Used in food preparation, food processing and industrial environments to verify accuracy of several types of temperature measurement instruments at a single hot test point. Built-in wrench is for easy calibration adjustment of spot check or meat cooking dial thermometers.
OPTIONS AND ACCESSORIES AVAILABLE FOR ALL UNITS ...........................................................................................Part #
Dial Adjustment Tool for 1" and 2" thermometers (set of 3) ..............................................................................................96100055
Cleaning Brushes for models that fit probe diameter up to .150" (set of 3)......................................................................96100054
Cleaning Brushes for multi-hole models (set of 2) ............................................................................................................96100145
Carrying Strap (for all except CS1-HOT) ..........................................................................................................................96100224
FACTORY RECALIBRATION/RECERTIFICATION SERVICE
Tel-Tru Manufacturing Co. offers annual maintenance, inspection, cleaning, and accuracy recertication of each Check-Set thermometer calibrator. Recertication automatically extends the product warranty one additional year,
for up to nine years. Contact Tel-Tru Customer Service for details.
Visit our website at www.teltru.com/calibrator.asp for more information on the Check-Set series of thermometer calibrators and accessories.
6
www.teltru.com
Page 7
Do you need to verify cold and/or hot temperatures?
Drawing A
(CS1-HOT)
4.0" Insulated Zone 4.0" Te mp. Control Zone
PROBE SENSING AREA
Drawing B
(All other models)
What temperature(s) do you
want to verify?
3.0" Insulated 3.0" Te mp. Zone Control Zone
PROBE SENSING AREA
What are the stem diameters of the
instruments you are checking?
STEm-LENGTH COmPATIbILITY
For best results, the sensing area of the probe must be inserted completely into the Temperature Control Zone. The minimum immersion
depth is dependent on the thermometer type. Check the manufacturer’s
user manual of your thermometer for the correct immersion length.
CS1-F42-310 shown with removable well P/N 96100250
Designed for use with refrigeration.
ADDITIONAL WELL INSERTS FOR CHECK-SET I COLD, II AND IV .................................................... PART #
Check-Set Well, 3-Hole (.125", .150", .187" Probes w/recal) .................................................................... 96100249
Check-Set Well, 3-Hole (.187", .250", .375" Probes) ................................................................................. 96100250
Check-Set Well, 1-Hole (.375" Probe)........................................................................................................ 96100252
Check-Set Well, 1-Hole (.500" Probe)........................................................................................................ 96100253
Check-Set Well, 1-Hole (.750" Probe)........................................................................................................ 96100254
Note: Wells with other hole diameters available upon request
Most
popular
unit
CS2-F50-320 shown with removable well P/N 96100249 - multi-hole configuration and built in wrench
Hot-Cold dual set point unit for general HAACP verification of many types of temperature instruments in food service, food processing, and meat packing environments.
Built-in wrench is for easy calibration adjustment of spot check or meat cooking dial thermometers.
www.teltru.com
CS4-F80-310 shown with removable well P/N 96100250
Designed with 4 critical set points for use in dairy products plants.
7
Page 8
ADDITIONAL PRODUCTS
AVAILABLE FROM TEL-TRU
Food / Dairy / beverage
Spot Check / Laboratory Testing Thermometers
Sanitary Bimetal Thermometers
Sanitary RTDs and Transmitters
Non-Contact Thermometers
Digi-Tel
Sanitary Pressure Gauges
Sanitary Pressure Transmitters
Industrial
Digi-Tel Thermometers
Bimetal Thermometers
Glow Dial Thermometers
Industrial RTDs and Transmitters
Pressure Gauges
Pressure Transmitters
Thermowells
Sanitary Thermometers
Gas Actuated Thermometers
Vapor Tension Thermometers
Surface Thermometers
Non-Contact Thermometers
ISO
9001:2015
CERT IFIED
MADE IN USA
CS616a
© 2018