Page 1
User Manual
T3AFG5 and T3AFG10 Series Function/Arbitrary
Waveform Generator
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
Declaration
Copyright © Teledyne Test Tools. All rights reserved.
Contents in this Manual are not allowed to be copied, extracted and
translated without permission from Teledyne Test Tools.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
I
Page 3
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage
to this product or any products connected to it. To avoid potential hazards,
use this product only as specified.
Only qualified personnel should perform service procedures.
To Avoid Fire or Personal Injury
l Use proper power line. Only the special power line of the products
approved by the state should be used.
l Ground the instrument. This generator is grounded through the
protective terra conductor of the power cord. To avoid electric
shock, the grounding conductor must be connected to the earth
ground. Make sure that the instrument is properly grounded before
connecting the input or output terminals.
l Observe all the ratings of the terminal. To avoid fire or shock,
observe all the ratings and symbols that marked on the instrument.
Read the user guide carefully before making connections to the
instrument.
l Do not operate without Covers. Do not operate the product with
covers or panels removed.
l Avoid circuit or wire exposed. Do not touch the exposed
connections or components when the power is on.
l Do not operate with suspected failures. If you suspect there is
damage with this product, you should have it inspected by
qualified service personnel authorized by Teledyne Test Tools
before further operations.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
I I
Page 4
l Provide proper ventilation.
l Do not operate in wet/damp conditions.
l Do not operate in an explosive atmosphere.
l Keep the product’s surfaces clean and dry.
l Not position the equipment so that it is difficult to operate the
disconnecting device (detachable plug)
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
I I I
Page 5
If the equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
This product has been tested to the requirements of
CAN/CSA-C22.2 No. 61010-1, second edition, including Amendment 1,
or a later version of the same standard incorporating the same level of
testing requirements.
Not to use the product for measurements within other measurement
categories, such as CAT II, CAT III, CAT IV.
Not to use the equipment for measurements on mains circuits, not to
use the equipment for measurements on voltage exceed the voltage
range describe in the manual.
Only probe assemblies which meet the manufacturer’s specifications
shall be used.
The Responsible body or operator should refer to the instruction
manual to preserve the protection afford by the equipment. If the
equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
Any parts of the device and its accessories are not allowed to be
changed or replaced, other than authorized by the manufacturer of his
agent.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
I V
Page 6
Safety Terms and Symbols
This electronic product is subject to disposal and recycling regulations
that vary by country and region. Many countries prohibit the disposal
of waste electronic equipment in standard waste receptacles. For more
information about proper disposal and recycling of your Teledyne
LeCroy product,
Terms and Symbols
The following symbols appear on the product or in its documentation:
WARNING
WARNING or CAUTION Condition or practice could result in injury or
loss of life, or damage to equipment.
DOUBLE INSULATION
PROTECTIVE (EARTH) TERMINAL
Rated for indoor use only. Do not operate where conductive pollutants
may be present.
High Voltage. Risk of electric shock or burn.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
V
Page 7
Introduction of T3AFG5-10 Series
The manual covers the following two types of T3AFG5-10 Series
Function/Arbitrary Waveform Generators: T3AFG5 and T3AFG10. It
does not cover the T3AFG40, T3AFG80 or T3AFG120. They have
their own manual.
T3AFG5-10 Series Function/Arbitrary Waveform Generators adopt the direct
digital synthesis (DDS) technology, which can provide stable, high-precision,
pure and low distortion signals. Its combination of excellent system features,
easiness in usage and versatile functions makes this generator a perfect
solution for your job now and in the future.
T3AFG5-10 Series Function/Arbitrary Waveform Generator has a clear and
simple front-panel. The user-friendly panel layout and instructions, versatile
terminals, direct graph interface, built-in instructions and help system have
greatly simplified the operation process, with the help of which, users do not
have to spend a great deal of time learning and familiarizing the operation of
the generator before they can use it proficiently. The built-in AM, DSB-AM,FM,
PM, ASK, FSK and PWM modulation functions generate modulated
waveforms at ease, without the help of a separate modulating source. USB
I/O is a standard accessory, while GPIB is optional. Remote instructions meet
the SCPI specification requirements.
From the characteristics and specifications given below, you will understand
how T3AFG5-10 can satisfy your requirements.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
VI
Page 8
l DDS technology provides precise, stable and low distortional
output signal.
l 3.5’TFT color LCD display.
l 125MSa/s sampling rate, 14-bit resolution.
l Frequency characteristics:
Sine: 1μHz to 5 MHz / 10 MHz
Square: 1μHz to 5 MHz / 10 MHz
Ramp: 1μHz to 300kHz
Pulse: 500μHz to 5MHz
White Noise: 5 MHz / 10MHz bandwidth (-3dB)
Arbitrary: 1μHz to 5MHz
l 5 standard waveforms: Sine, Square, Ramp, Pulse, Noise
l Self-defined arbitrary waveform
l Multiple modulation function: AM, FM, PM, ASK, FSK, PWM,
DSB-AM, Sweep and Burst.
l Multiple interfaces: USB host & USB device(USBTMC)
l Support USB storage device. Software updating could also be
performed using USB devices.
l Up to 16k sample points of internal waveform depth, which can
rebuild or simulate any complex waveform.
l 2 languages (English and Chinese) user interface and built-in help
system.
Note:
All the specifications described in this guide are according to T3AFG5-10. If
you need to know the particular specifications about the other types, please
see the data sheet.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
VI I
Page 9
Catalogue
General Safety Summary................................................................................. II
Introduction to the T3AFG5-10 Series .......................................................... VI
1. Getting Started.......................................................................................... 1
1.1. General Inspection ........................................................................... 2
1.2. Handle Adjustment ...........................................................................3
1.3. The Front/Rear Panel.......................................................................4
1.4. To Set a Waveform........................................................................... 7
1.5. To Set Modulate/Sweep/Burst ........................................................ 12
1.6. To Set Output ................................................................................. 14
1.7. To Use Digital Input ........................................................................ 15
1.8. To Use Store/Utility/Help Function ..................................................16
2. Operating Your Generator........................................................................ 17
2.1. To Set Sine Signals ........................................................................18
2.2. To Set Square Signals.................................................................... 23
2.3. To Set Ramp Signals...................................................................... 26
2.4. To Set Pulse Signals ...................................................................... 29
2.5. To Set Noise Signals ......................................................................33
2.6. To Set Arbitrary Signals ..................................................................35
2.7. To Generate the Modulated Waveform........................................... 42
2.8. To Generate Sweep........................................................................51
2.9. To Generate Burst .......................................................................... 53
2.10. To Store and Recall ..................................................................... 57
2.11. To Set the Utility Function............................................................ 63
2.12. Test/Cal .......................................................................................74
2.13. Edition Information ...................................................................... 79
2.14. Updating Firmware......................................................................81
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
VI I I
Page 10
2.15. How to Use the Built-in Help System .......................................... 82
3. Application and Examples ....................................................................... 83
3.1. Example 1:Generate a Sine Wave .................................................84
3.2. Example 2:Generate a Square Wave............................................. 86
3.3. Example 3:Generate a Ramp Wave............................................... 88
3.4. Example 4:Generate a Pulse Wave ............................................... 90
3.5. Example 5:Generate a Noise Wave ............................................... 92
3.6. Example 6:Generate an Arbitrary Wave ......................................... 93
3.7. Example 7:Generate a Sweep Linear Wave................................... 95
3.8. Example 8:Generate a Burst Wave ................................................ 97
3.9. Example 9:Generate an AM Wave ................................................. 99
3.10. Example 10:Generate a FM Wave ............................................101
3.11. Example 11:Generate a PM Wave ............................................ 103
3.12. Example 12:Generate a FSK Wave .......................................... 105
3.13. Example 13:Generate an ASK Wave ........................................107
3.14. Example 14: Generate a PWM Wave........................................ 109
3.15. Example 15: Generate a DSB-AM Wave .................................. 111
4. Troubleshooting ..................................................................................... 113
General Inspecting ................................................................................ 113
Troubleshooting .................................................................................... 114
5. Service and Support .............................................................................. 115
Maintain summary ................................................................................ 115
Contact Teledyne Test Tools .................................................................. 116
6. Appendix................................................................................................ 117
Appendix A: Accessories ...................................................................... 117
Appendix B: Daily Maintain and Cleaning............................................. 118
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
I X
Page 11
1. Getting Started
This chapter covers the following topics:
n General Inspection
n Handle Adjustment
n The Front/Rear Panel
n To Set a Waveform
n To Set Modulate/Sweep/Burst
n To Set Output
n To Use Digital Input
n To Use Store/Utility/Help Function
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1
Page 12
1.1. General Inspection
When you get a new T3AFG5-10 Series Function/Arbitrary Waveform Generator,
you are suggested to take the following steps to inspect the instrument.
1. Inspect the shipping container for damage.
If there are damages in the packing or foam, keep them until the whole
machine and the accessories pass the electric and mechanical testing.
2. Check the accessories.
Accessories supplied with the instrument are listed in chapter 6
‘Appendix A: Accessories’.
If the contents are incomplete or damaged, please notify your sales
representative.
3. Inspect the instrument.
In case any mechanical damage or defect, or if the instrument does not
operate properly or pass performance tests, notify your sales
representative. If the shipping container is damaged, or the cushioning
materials show signs of stress, notify the carrier as well as your sales
office. Keep the shipping materials for the carrier’s inspection. Offices
will arrange for repair or replacement at their option without waiting for
claim settlement.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2
Page 13
1.2. Handle Adjustment
To adjust the handle position of T3AFG5-10 Function/Arbitrary Waveform
Generator, please grip the handle by the sides and pull it outward.
Then, make the handle rotate to the desired position.
Figure 1- 1 Viewing Position and Carrying Position
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3
Page 14
1.3. The Front/Rear Panel
When you get a new T3AFG5-10 Series Function/Arbitrary Waveform Generator,
first you need to understand how to operate the front/rear panel correctly.
This chapter will make a brief introduction and description for the operation
and functions of the front/rear panel.
The T3AFG5-10 Series Function/Arbitrary Waveform Generator has a clear
and simple front panel. See Figure 1- 2 and Figure 1- 3. The front panel has a
knob and functional keys. The 5 blue grey buttons on the right side of the screen
are menu buttons (named F1 to F5 from top to bottom) with the help of which,
you can enter different functions menu or have direct specific applications. The
signal input and output interfaces are set at the front and rear panels which can
help generating multiple arbitrary waveforms. The various interfaces can meet
the need of the multiple interface communications.
Figure 1- 2 T3AFG5-10 Series
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4
Page 15
Back
Waveform
keys
Number
keys
Power
LCD
Display
Figure 1- 3 Front Panel of T3AFG5-10 Series
Menu
Function
Operation keys
Output
Control
Direction
keys
Power Socket
Figure 1- 4 Rear Panel of T3AFG5-10 Series
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5
Page 16
Waveform
Operation
Display area
Menu:
Parameter area
Figure 1- 5 Display Interface (Sine Wave is the default display signal)
Character definitions in this User Manual:
The signs for buttons in this manual are the same as the panel buttons.
Please note that, the signs for the functional buttons on the operation panel
are represented by squared words, such as Sine , which represents the
transparent functional key with Sine on it on the front panel, while the menu
buttons are represented by brighten words such as Freq, which means the
frequency option in the Sine menu.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6
Page 17
1.4. To Set a Waveform
On the operation panel, there is a set of buttons with waveform icon. See
Figure 1- 6. The exercise below will help you familiarize with the waveform
selection settings.
Figure 1- 6 Waveform Selection Buttons
1. Press Sine button and the waveform window will display sine waveform.
T3AFG5-10 Series Generator can generate sine signal with a frequency
from 1μHz to 30MHz. By setting frequency/period, amplitude/high level,
offset/low level, sine signal with different parameters can be generated.
Figure 1- 7 Sine Signal Display Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7
Page 18
As is shown in Figure 1- 7, the default signal parameters are: 1kHz frequency,
4.0Vpp amplitude and 0Vdc offset.
2. Press Square button, and the waveform window displays square
waveform. T3AFG5-10 Series Generator can generate square signal
with a frequency from 1μHz to 5MHz or 10MHz and variable duty
cycle.
Figure 1- 8 Square Signal Display Interface
As is shown in Figure 1- 8, the default signal parameters are: 1kHz frequency,
4.0Vpp amplitude, 0Vdc offset and 50% duty cycle.
3. Press Ramp button, and the waveform window displays ramp waveform.
T3AFG5-10 Series Generator can generate ramp signal with a frequency of
from 1μHz to 300kHz and variable symmetry.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8
Page 19
Figure 1- 9 Ramp Signal Display Interface
As is shown in Figure 1- 9, the default signal parameters are: 1kHz frequency,
4.0Vpp amplitude, 0Vdc offset and 50% symmetry.
4. Press Pulse button, and the waveform window displays pulse waveform.
T3AFG5-10 Series Generator can generate pulse signal with a frequency
from 500μHz to 5 MHz and variable pulse width and delay.
Figure 1- 10 Pulse Signal Display Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
9
Page 20
As is shown in Figure 1-10, the default signal parameters are: 1kHz
frequency, 4.0Vpp amplitude, 0Vdc offset, 200μs pulse width.
5. Press Noise button, and the waveform window displays noise waveform.
T3AFG5-10 Series Generator can generate noise signal with a band width
up to 5MHz or 10MHz depending on the model.
Figure 1-11 Noise Signal Display Interface
As is shown in Figure 1-11, the default signal parameters are: 128mV Stdev
and 0V Mean.
6. Press Arb button, and the waveform window displays arbitrary waveform.
T3AFG5-10 Series Generator can generate repeatable arbitrary waveform
signals with at most 16K points and 5MHz frequency.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1 0
Page 21
Figure 1-12 Arbitrary Waveform Signal Display Interface
As is shown in Figure 1-12, the default signal parameters are: 1kHz
frequency, 4.0Vpp amplitude and 0mVdc offset.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1 1
Page 22
1.5. To Set Modulate/Sweep/Burst
As shown in Figure 1-13, there are three buttons on the front panel, which
are used for modulation, sweep and burst settings. The instructions below will
help you familiarize yourself with the setting of these functions.
Figure 1- 13 Modulate/Sweep/Burst Button
1. Press Mod button, the modulated waveforms will be generated. The
modulated waveform can be changed by modifying the parameters
such as type, internal/external modulation, depth, frequency, waveform,
etc. T3AFG5-10 Series can modulate waveform using AM, FM, PM,
ASK, FSK, PWM and DSB-AM. Sine, square, ramp and arbitrary
waveforms can be modulated (pulse, noise and DC can not be modulated).
Figure 1-14 Modulated Waveform Display Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1 2
Page 23
2. Press Sweep button, sine, square, ramp or arbitrary waveform can be
swept (pulse, noise and DC can not be swept).
In the sweep mode, T3AFG5-10 Series generate signal with variable
frequencies.
Figure 1-15 Sweep Waveform Display Interface
3. Press Burst button, burst for sine, square, ramp, pulse or arbitrary
waveform can be generated.
Figure 1-16 Burst Waveform Display Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1 3
Page 24
Term Explanation
Burst: Output waveforms with set cycle times.
Burst can last for certain times of waveform cycle (N-Cycle Burst) or be
controlled by external gated signals (Gated Burst). Burst applies to all
kinds of waveforms, but noise can only be used in gated burst. Generally it
is called burst function within every signal generator.
1.6. To Set Output
As is shown in Figure 1-17, there are two buttons on the right side of the
operating panel, which are used for output/trigger control. The instructions
below will help you familiarize with these functions.
Figure 1-17 Output Buttons
Press Output button, to activate or deactivate the output signal.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1 4
Page 25
1.7. To Use Digital Input
As is shown in Figure 1-18, there are three sets of buttons on the operating
panel, which are direction buttons, the knob and the keypad. The instruction
below will help you familiarize yourself with the digital input function.
Figure 1-18 Front Panel Digital Input
1. The up and down keys are used to shift parameters and the left and right
keys are used to shift digits.
2. Keypad is used to directly set the parameters value.
3. Knob is used to change a signal digit value whose range is 0~9.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1 5
Page 26
1.8. To Use Store/Utility/Help Function
As is shown in Figure 1-19, there are three buttons on the operating panel,
which are used to call the store/recall, utility and help function. The instruction
below will help you familiarize yourself with these functions.
Figure 1- 19 Store/Recall Utility and Help Button
1. The Store/Recall button is used to store waveform data and configure
information.
2. The Utility button is used to set the auxiliary system function, change
the output configure parameters, interface setting, system setting
information or perform the instrument self-test and read the calibration
information, etc.
3. The Help button is used to read the help information.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1 6
Page 27
2. Operating Your Generator
Up to now you have got a brief understanding about T3AFG5-10 series
with the front/rear panel, every function control area and keys. You
should also know how to set your Function/Arbitrary Waveform
Generator for your usage. If you are not familiar with these operations,
you are suggested to read chapter one ‘Getting Started’ again.
This chapter covers the following topics:
n Setting Sine Signal
n Setting Square Signal
n Setting Ramp Signal
n Setting Pulse Signal
n Setting Noise Signal
n Setting Arb Signal
n Output Modulated Signal
n Output Sweep Signal
n Output Burst Signal
n Store/Recall
n Utility Setting
n Help System
You are suggested to read this chapter carefully so as to understand
T3AFG5-10 Series Generator’s versatile waveform setting functions and
more operation methods.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1 7
Page 28
2.1. To Set Sine Signals
Press Sine button to call the sine operation. The sine waveform parameters
are set by using the sine operation menu.
The parameters of sine waveforms are: frequency/period, amplitude/high
level, offset/low level and phase. Different sine signals are generated by
setting these parameters. As is shown in Figure 2- 1, in the soft key menu,
select Freq. Cursor is located in the frequency parameter area in the
parameter display window, and users can set the frequency value here.
Figure 2- 1 Sine Parameter Display Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1 8
Page 29
Figure 2- 2 Table2- 1 Menu Explanations of Sine Waveform
Function
menu
Set the signal frequency or period;
Freq/
The current parameter will be switched at
Period
a second press.
Set the signal amplitude or high level; The
Ampl/
current parameter will be switched at a
HLevel
second press.
Set the signal offset or low level;
Offset/
The current parameter will be switched at
LLevel
a second press.
Phase Set the phase of the signal;
To Set the Output Frequency/Period
Explanations
1. Press Sine →Freq, to set the frequency parameter.
The frequency shown on the screen when the instrument is powered is
the default value or the set value beforehand. When setting the function,
if the current value is valid for the new waveform, it will be used
sequentially. If you want to set the period for the waveform, press
Freq/Period button again, to switch to the period parameter (the current
operation is displayed in inverse color).
2. Input the desired frequency.
Use the keypad to input the parameter value directly, and press the
corresponding button to select the parameter unit. Or you can use the
direction button to select the digit you want to edit, and then use the knob
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1 9
Page 30
to change its value.
Figure 2-3 Setting the Frequency
Instruction :
When using the keypad to enter the digit, you can use the left direction button
to move the cursor backward and delete or change the value of the previous
digit.
To Set the Output Amplitude
1. Press Sine →Ampl, to set the amplitude.
The amplitude shown on the screen when the instrument is powered is
the default value or the set value beforehand. When changing the
function, if the current value is valid for the new waveform, it will be used
sequentially. If you want to set the waveform by high level or low level,
press the Ampl/HLevel or Offset/LLevel button again, to switch into the
high level or low level parameter (the current operation is displayed in
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2 0
Page 31
inverse color).
2. Input the desired Amplitude
Use the keypad or the knob to input the desired value, choose the unit,
and press the corresponding button.
Figure 2-4 Setting the Amplitude
To Set the Output Offset
1. Press Sine →Offset, to set the offset.
The offset shown on the screen when the instrument is powered is the
default value or the set value beforehand. When changing the function, if
the current value is valid for the new waveform, it will be used
sequentially.
2. Input the desired Offset
Use the keypad or the knob to input the desired value, choose the unit,
and press the corresponding button.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2 1
Page 32
Figure 2-5 Setting the Offset
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2 2
Page 33
2.2. To Set Square Signals
Press Square button to call the Square operation. The square waveform
parameters are set by using the Square operation menu.
The parameters of Square waveforms are: frequency/period, amplitude/high
level, offset/low level, phase and duty. As is shown in Figure 2- 6, select Duty.
Cursor is located in the duty parameter area in the parameter display window,
and users can set the duty value here.
Figure 2- 6 Square Parameter Display Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2 3
Page 34
Figure 2- 7 Table2- 2 Menu Explanations of Square Waveform
Function
Menu
Freq/
Period
Ampl/
HLevel
Offset/
LLevel
Phase Set the phase of the signal;
Duty Set the duty cycle for square waveform.
Settings Explanation
Set the signal frequency or period;
The current parameter will be switched at a
second press.
Set the signal amplitude or high level;
The current parameter will be switched at a
second press.
Set the signal offset or low level;
The current parameter will be switched at a
second press.
Term Explanation:
Duty Cycle: The percentage that the high level takes up the whole period.
Please Note : for the Frequency Duty Cycle Value
Below 10MHz: 20% to 80%
To Set the Duty Cycle
1 Press Square →Duty, to set the duty cycle.
The duty cycle shown on the screen when the instrument is powered is
the default value or the set value beforehand. When changing the
function, if the current value is valid for the new waveform, it will be used
sequentially.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2 4
Page 35
2 Input the desired Duty Cycle
Use the keypad or the knob to input the desired value, choose the unit,
and press the corresponding button. The generator will change the
waveform immediately.
Figure 2- 8 Setting the Duty Cycle
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2 5
Page 36
2.3. To Set Ramp Signals
Press Ramp button to call the ramp operation. The ramp waveform
parameters are set by using the ramp operation menu.
The parameters for ramp waveforms are: frequency/period, amplitude/high
level offset/low level, phase and symmetry. As is shown in Figure 2-9, in the
soft key menu, select Symmetry. Cursor is located in the symmetry parameter
area in the parameter display window, and users can set the symmetry value
here.
Figure 2- 9 Ramp Parameter Display Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2 6
Page 37
Figure 2-10 Table2-3 Menu Explanations of Ramp Waveform
Function
Menu
Freq/
Period
Ampl/
HLevel
Offset/
LLevel
Phase Set the phase of the signal;
Symmetry Set the symmetry for ramp waveform.
Settings Explanation
Set the signal frequency or period;
The current parameter will be switched at
a second press.
Set the signal amplitude or high level; The
current parameter will be switched at a
second press.
Set the signal offset or low level;
The current parameter will be switched at
a second press.
Term Explanation:
Symmetry: The percentage that the rising period takes up the whole Period.
Input Range: 0~100%.
To Set the Symmetry
1. Press Ramp →Symmetry, to set the symmetry.
The symmetry shown on the screen when the instrument is powered is
the default value or the set value beforehand. When changing the
function, if the current value is valid for the new waveform, it will be used
sequentially.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2 7
Page 38
2. Input the desired Symmetry
Use the keypad or the knob to input the desired value, choose the unit,
and press the corresponding button. The generator will change the
waveform immediately.
Figure 2-11 Setting the Symmetry
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2 8
Page 39
2.4. To Set Pulse Signals
Press Pulse button to call the pulse operation. The pulse waveform
parameters are set by using the pulse operation menu.
The parameters for pulse waveforms are: frequency/period, amplitude/high
level, offset/low level, pulse width/Duty and Rise/Fall. As is shown in Figure
2- 12, in the soft key menu, select PulWidth. Cursor is located in the pulse
width parameter area in the parameter display window, and users can set the
pulse width value here.
Figure 2- 12 Pulse Parameter Display Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2 9
Page 40
Figure 2- 13 Table 2- 4 Menu Explanations of Pulse Waveform
Function
Menu
Freq/
Period
Ampl/
HLevel
Offset/
LLevel
PulWidth
/Duty
Rise /
Fall
Explanation
Set the signal frequency or period;
The current parameter will be switched at a
second press.
Set the signal amplitude or high level;
The current parameter will be switched at a
second press.
Set the signal offset or low level;
The current parameter will be switched at a
second press.
Set the signal pulse width or duty;
The current parameter will be switched at a
second press.
Setting the rising edge for pulse waveform.
Setting the falling edge for pulse waveform
Term Explanation:
Pulse Width:
Positive Pulse Width: the time span between thresholds of 50% of the rising
edge amplitude to the next 50% of the falling edge amplitude;
Negative Pulse Width: the time span between thresholds of 50% of the falling
edge amplitude to the next 50% of the rising edge amplitude.
To Set the Pulse Width
1. Press Pulse →PulWidth, to set the pulse width.
The pulse width shown on the screen when the instrument is powered is
the default value or the set value beforehand. When changing the
function, if the current value is valid for the new waveform, it will be used
sequentially.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3 0
Page 41
2. Input the desired Pulse Width
Use the keypad or the knob to input the desired value, choose the unit,
and press the corresponding button. The Generator will change the
waveform immediately.
Figure 2- 14 Setting the Pulse Width
To Set the Rising Edge
1. Press Pulse →Rise, to set the Rise edge.
The rising edge shown on the screen when the instrument is powered is
the default value or the set value beforehand. When changing the
function, if the current value is valid for the new waveform, it will be used
sequentially.
2. Input the desired rising edge
Use the keypad or the knob to input the desired value, choose the unit,
and press the corresponding button. The generator will change the
waveform immediately.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3 1
Page 42
Figure 2-15 Setting the Rise edge
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3 2
Page 43
2.5. To Set Noise Signals
Press Noise button to call the Gaussian White noise operation. The noise
waveform parameters are set by using the noise operation menu.
The parameters for noise waveforms are: Stdev and mean. As is shown in
Figure 2- 16, in the soft key menu, select Stdev, Cursor is located in the Stdev
parameter area in the parameter display window, and users can set the Stdev
value here. Noise is non-regulated signal which has no frequency or period.
Figure 2-16 Noise Parameter Display Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3 3
Page 44
Figure 2- 17 Table 2-5 Menu Explanations of Noise Waveform
Function
Menu
Stdev Set the signal standard deviation
Mean Set the signal mean
Settings Explanation
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3 4
Page 45
2.6. To Set Arbitrary Signals
Press Arb button to call the Arb operation. The Arb waveform parameters are
set by using the Arb operation menu.
The Arb signal consists of two types: the system built-in waveform and the
user-definable waveform. The parameters for Arb waveforms are:
frequency/period, amplitude/high level, offset/ low level and phase.
Figure 2-18 Arb Parameter Display Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3 5
Page 46
Figure 2-19 Table 2-6 Menu Explanations of Arb Waveform (Page 1/2)
Function
Menu
Freq/
Period
Ampl/
HLevel
Offset/
LLevel
Phase Set the phase of the signal;
Settings Explanation
Set the signal frequency or period;
The current parameter will be switched at
a second press.
Set the signal amplitude or high level;
The current parameter will be switched at
a second press.
Set the signal offset or low level;
The current parameter will be switched at
a second press.
Figure 2-20 Table 2- 7 Menu Explanations of Arb Waveform (Page 2/2)
Function
Menu
Load Wform
Settings Explanation
Select the built-in arbitrary signal as
output.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3 6
Page 47
To select the built-in Arbitrary Waveform
There are forty-eight built-in Arbitrary Waveforms and user-definable Arbitrary
Waveforms inside the Generator. To select one of them, follow the
instructions below:
Press Arb →Load Wform, to enter the interface below.
Figure 2- 21 Table 2- 8 Menu Explanations of Built-in Arbitrary Waveform
Function
Menu
Built-In
Stored
Wforms
Cancel
Settings Explanation
Select one of the 46 types built-in arbitrary
waveforms.
Select one of arbitrary waveforms stored in
the non-volatile memory.
Cancel the current operation, and return to
the upper menu (the followings are the same
and will not be explained).
1. To Select the Built-in Waveform
Press Arb →Load Wform->Built-In, and enter the following interface.
As is shown in Figure 2-22, there are five kinds of arbitrary waveform.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3 7
Page 48
Figure 2-22 Table 2-9 Menu Explanations of Built-In Arbitrary Waveform
Function
Menu
Common Select common waveform.
Math Select math waveform.
Project Select project waveform.
Winfun/
Triangle
Select Validate the built-in waveform.
Figure 2-23 Common Built-In Arbitrary Waveform interface
Settings Explanation
Select windows function
/triangle waveform.
Table 2-10 Menu Explanations of Common Built-In Arbitrary Waveform
Function
Menu
StairUp Select the built-in stair up waveform.
StairDn Select the built-in stair down waveform.
StairUD
PPulse Select the built-in positive pulse waveform.
NPulse Select the built-in negative pulse waveform.
Trapezia Select the built-in trapezoid waveform
UpRamp Select the built-in up ramp waveform.
DnRamp Select the built-in down ramp waveform.
Settings Explanation
Select the built-in stair up and down
waveform.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3 8
Page 49
Figure 2-24 Math Built-In Arbitrary Waveform Interface
Table 2-11 Menu Explanations of Math Built-in Arbitrary Waveform
Function
Menu
ExpFall Select the built-in exponential fall waveform.
ExpRise Select he built-in exponential rise waveform.
LogFall Select the built-in logarithmic fall waveform.
LogRise Select the built-in logarithmic rise waveform.
Sqrt Select the built-in square root waveform.
Root3 Select the built-in Root3 waveform.
X^2 Select the built-in X^2 waveform.
X^3 Select the built-in X^3 waveform.
Sinc
Gaussian Select the built-in gaussian waveform.
Dlorentz Select the built-in D-lorentz waveform.
Haversin Select the built-in haversine waveform.
Lorentz Select the built-in lorentz Waveform.
Gauspuls
Gmonpuls
Tripuls
Settings Explanation
Select the built-in sinc waveform; Sinc sin(x)/x.
Select the built-in gaussian-modulated sinusoidal
pulse waveform.
Select the built-in Gaussian monopulse waveform.
Select the built-in triangle pulse waveform.
Figure 2-25 Project Built-In Arbitrary Waveform interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3 9
Page 50
Table 2-12 Menu Explanations of Project Built-in Arbitrary Waveform
Function
Menu
Cardiac
Quake
Chirp
TwoTone Select the built-in two tone signal waveform.
SNR
Figure 2-26 Winfun/Triangle Built-In Arbitrary Waveform interface
Settings Explanation
Select the built-in electrocardiogram (ECG) signal
waveform.
Select the built-in loma prieta earthquake
waveform.
Select the built-in swept-frequency cosine
waveform.
Select the built-in sin wave with white noise
waveform.
Table 2- 13 Menu Explanations of Winfun/Triangle Built-in Arbitrary Waveform
Function
Menu
Hamming Select the built-in hamming window waveform.
Hanning Select the built-in hanning window waveform.
Kaiser Select the built-in kaiser window waveform.
Blackman
Gaussian Select the built-in gaussian window waveform.
Triangle Select the built-in triangle window waveform.
Hairs Select the built-in hairs window waveform.
Bartlett Select the built-in bartlett window waveform.
Tan Select the built-in tangent waveform. Cot
Select the built-in cotangent waveform.
Sec Select the built-in secant waveform. Csc
Select the built-in cosecant waveform
Asin Select the built-in inverse sine waveform.
Acos Select the built-in inverse cosine waveform.
Settings Explanation
Select the built-in blackman windows
waveform.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4 0
Page 51
Atan Select the built-in tangent waveform.
Acot
Select the built-in inverse cotangent waveform.
2. To Select the Stored Waveform
Press Arb →Load Wform->Stored Wforms, and enter the following interface.
As is shown in Figure 2-27, use the direction keys or knob to choose the
corresponding arbitrary waveform and press Select.
Figure 2-27 Stored Wform Display Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4 1
Page 52
2.7. To Generate the Modulated Waveform
Use the Mod button to generate modulated waveform. T3AFG5-10 Series
can generate AM, FM, ASK, FSK, PM, PWM and DSB-AM modulated
waveforms. Modulating parameters vary with the types of the modulation.
In AM, users can set the depth, modulating frequency, modulating waveform
and carrier waveform; In FM, users can set the modulating frequency,
frequency deviation, modulating waveform and carrier waveform; In ASK,
users can set the Key Freq and carrier waveform; In FSK, users can set
the key frequency, Hop Freq and carrier waveform; In PM, users can set
the phase deviation, modulating frequency, modulating waveform and carrier
waveform. We will cover how to set these parameters in details according
to the modulation types.
Modulating
Frequency
Carrier
Frequency
Figure 2- 28 Display Interface of Modulated Waveform Parameter
Carrier
Amplitude
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4 2
Page 53
AM
The modulated waveform consists of two parts: the carrier waveform and the
modulating waveform. In AM, the amplitude of the carrier waveform varies
with the instantaneous voltage of the modulating waveform. The parameters
for the AM are in Figure 2-29
Press Mod →Type→ AM, to enter the following menu.
Figure 2-29 Table 2-14 Menu Explanations of the AM Parameters
Function
Menu
AM Freq
AM Depth Set the amplitude range.
Type AM Amplitude modulation.
Shape
Settings Explanation
Set the modulating waveform
frequency. Frequency range:
2mHz~20kHz (internal source only).
Sine
Square
Choose the modulating waveform.
Triangle
To change the carrier waveform
UpRamp
parameter, press Sine, Square, Ramp,
DnRamp
Arb
Noise
Arb
Term Explanation
Modulation Depth
The amplitude range (also called ‘Percentage Modulation’). Modulation depth
varies from 0% to 120%.
l In the 0% modulation, the output amplitude is the half of the set value.
l In the 100% modulation, the output amplitude is the same as the
set value.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4 3
Page 54
FM
The modulated waveform consists of two parts: the carrier waveform and the
modulating waveform. In FM, the frequency of the carrier waveform varies
with the instantaneous voltage of the modulating waveform. The parameters
for the FM are as shown in Figure 2-30.
Figure 2-30 Setting Interface of FM Waveform Parameter
Press Mod →Type→ FM, to enter the following menu.
Figure 2-31 Table 2-15 Menu Explanations of the FM Parameters
Function
Menu
FM Freq
FM Dev Set the maximum frequency deviation
Type FM Frequency modulation
Shape
Settings Explanation
Set the modulating waveform frequency.
Frequency range 2mHz~20kHz (internal
source).
Sine
Square
Triangle
UpRamp
DnRamp
Noise
Arb
Choose the modulating waveform. To
change the carrier waveform parameter,
press Sine, Square, Ramp, Arb
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4 4
Page 55
Term Explanation
Frequency Deviation
l The deviation should be equal to or less than the carrier waveform
frequency.
l The sum of the deviation and the carrier frequency should be equal to
or less than maximum frequency of the selected function.
ASK (Amplitude Shift Keying)
ASK is a form of modulation that represents digital data as variations in the
amplitude of a carrier wave. The amplitude of an analog carrier signal varies
in accordance with the bit stream (modulating signal), keeping frequency and
phase constant. The parameters for the ASK are as shown in Figure 2- 32
Figure 2-32 Setting Interface of ASK Waveform Parameter
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4 5
Page 56
Press Mod →Type→ ASK, to enter the following menu.
Figure 2-33 Table 2-16 Menu Explanations of the ASK Parameters
Function
Menu
Key Freq
Type ASK Amplitude shift keying modulation.
Settings Explanation
Set the frequency at which the output
amplitude shifts between the carrier
amplitude and zero (internal modulation
only): 2mHz~50kHz.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4 6
Page 57
FSK (Frequency Shift Keying)
The FSK Modulation is a modulation method, the output frequency of which
switches between the two pre-set frequencies (carrier waveform frequency
and the hop frequency). The frequency at which the output frequency
switches is called the key frequency.
Figure 2- 34 Setting Interface of FSK Waveform Parameter
Press Mod →Type→ FSK, to enter the following interface
Figure 2- 35 Table 2- 17 Menu Explanations of the FSK Parameters
Function
Menu
Key Freq
Type FSK Frequency shift keying modulation.
Hop Freq Set the hop frequency.
Settings Explanation
Set the frequency at which the output
frequency shifts between the carrier
frequency and the hop frequency (internal
modulation only):
2mHz~50kHz.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4 7
Page 58
PM (Phase Modulation)
The modulated waveform consists of two parts: the carrier waveform and the
modulating waveform. In PM, the phase of the carrier waveform varies with
the instantaneous voltage level of the modulating waveform. The parameters
for the PM are as shown in Figure 2- 36.
Figure 2- 36 Setting Interface of PM Waveform Parameter
Press Mod →Type →PM, enter the following interface.
Figure 2-37 Table 2-8 Menu Explanations of the PM Parameters
Function
Menu
PM Freq
Phase
Dev
Type PM Phase modulation
Shape
Settings Explanation
Set the modulating waveform frequency.
Frequency range: 2mHz~20kHz
Range from 0° ~ 360°.
Sine
Square
Triangle
UpRamp
DnRamp
Noise
Arb
Choose the modulating waveform.
To change the carrier waveform parameter,
press Sine, Square, Ramp, Arb
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4 8
Page 59
PWM (Pulse Width Modulation)
The modulated waveform consists of two parts: the carrier waveform and the
modulating waveform, the carrier waveform is only a pulse. In PWM, the pulse
width of pulse varies with the instantaneous voltage of the modulating
waveform. The parameters for the PMW are as shown in Figure 2-38.
Figure 2-38 Setting Interface of PWM Waveform Parameter
Press Mod→ Pulse →PWM, to enter the following menu.
Figure 2-39 Table 2-19 Menu Explanations of the PWM Parameters
Function
Menu
PWM Freq
Width Dev
Duty Dev
Type PWM Amplitude modulation.
Shape
Settings Explanation
Set the modulating waveform
frequency. 2mHz~20kHz
Set the width or Duty range.
Sine
Square
Triangle
Choose the modulating waveform.
UpRamp
The carrier waveform is pulse.
DnRamp
Noise
Arb
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
4 9
Page 60
DSB-AM
Press Mod→ Type →DSB-AM. The parameters for the DSB-AM are as
shown in Figure 2- 40.
Figure 2- 40 Setting Interface of DSB-AM Waveform Parameter
Figure 2- 41 Table 2- 20 Menu Explanations of the DSB-AM Parameters
Function
Menu
DSB Freq
Type DSB-AM Amplitude modulation.
Shape
Settings Explanation
Set the modulating waveform frequency.
Frequency range: 2mHz~20kHz
Sine
Square
Triangle
UpRamp
DnRamp
Noise
Arb
Choose the modulating waveform.
To change the carrier waveform parameter,
press Sine, Square, Ramp, Arb
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5 0
Page 61
2.8. To Generate a Sweep
In the frequency sweep mode, the function generator ‘steps’ from the start
frequency to the stop frequency at the sweep time you specify. Sweep can be
generated by sine, square, ramp or arbitrary waveforms (pulse, noise and DC
are not allowed).
Figure 2- 42 Setting Interface of Sweep Waveform Parameter
Press Sweep button to enter the following menu. Set the waveform
parameters by using the operation menu.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5 1
Page 62
Figure 2-43 Table 2-21 Menu Explanations of Waveform Sweep (Page 1/2)
Function
Menu
Swp Time
Stop Freq
Freq Span
Start Freq
Mid Freq
Source
Sweep Frequency Setting
Settings Explanation
Set the time span of the sweep in which
the frequency changes from the start
frequency to stop frequency.
Set the stop frequency of the sweep;
Set the frequency span of the sweep.
Set the start frequency of the sweep;
Set the center frequency of the sweep.
Internal Choose internal source.
External Choose external source.
Manual Set the start and stop time by hand.
Use start freq and stop freq or center freq and freq span to set the range of
the frequency. Press the button again to switch between each setting.
Figure 2-44 Table 2-22 Menu Explanations of Waveform Sweep (Page 2/2)
Function
Menu
Linear/
Log
Direct
Settings Explanation
Set the sweep with linear spacing;
Set the sweep with logarithmic spacing.
Sweep upward;
Sweep downward.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5 2
Page 63
2.9. To Generate a Burst
Burst function can generate versatile waveforms in burst, which can last
specific number of waveform cycles (N-Cycle burst), or when external gated
signals (gated burst) is applied, any waveform could be used, but noise can
only be used in Gated Burst.
Press Burst button to enter the following interface. Set the waveform
parameters by using the operation menu.
Figure 2- 45 Setting Interface of Burst Waveform Parameter
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5 3
Page 64
Set the N-Cycle Burst
Press Burst → N Cycle, to enter the following interface.
Figure 2-46 Table 2-23 Menu Explanations of the N-Cycle Parameters(Page 1/2)
Function
Menu
Period Set the burst Period.
Start
Phase
NCycle
Gated
Source
Settings Explanation
Set the start phase of the burst.
Use the N-Cycle mode.
Use the Gated mode.
Internal Choose internal source.
External Choose external source.
Choose external source, set the start time
Manual
by hand.
Burst Period
Set the time span between an N-Cycle burst and the next. If necessary the
period will increase to allow the specific number of cycles in a burst. Burst
Period>Carrier Period × Burst Number
Start Phase
Define the start point in a waveform. The phase varies from 0° to 360°, and
the default setting is 0°. For an Arbitrary Waveform, 0° is the first waveform
point.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5 4
Page 65
N-Cycle/Gated
N-Cycle has specific number of waveform cycles, and every burst is activated
by a trigger event. Gated burst use external source to control burst when
it’s activated.
Figure 2-47 Table 2-24 Menu Explanations of the N-Cycle Parameters (Page2/2)
Function
Menu
Cycles/
Infinite
Delay Set the delay time before the burst starts.
Settings Explanation
Set the number of the bursts in a N-Cycle.
Set the number of the bursts in a N-Cycle to
be infinite.
Cycles
Set the number of waveform cycle in an N-Cycle (1 to 50,000 or Infinite).
If you choose Infinite, then a continuous waveform will be generated which
will not stop until a trigger event happens.
l If required, Burst Period will increase to cater to the specific number
of cycles.
l For an infinite-cycle burst, external or manual trigger is needed to
activate the burst.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5 5
Page 66
Delay
Set the time delay between the trigger input and the start of the N-Cycle burst.
The max delay is 240ns.
Set the Gated Burst
Press Burst →Gated, to enter the following interface.
Figure 2- 48 Table 2- 25 Menu Explanations of the Gated Burst Parameters
Function
Menu
NCycle
Gated
Polarity
Settings Explanation
Set the NCycle mode;
Set the gated mode.
Positive
Set the polarity for the gated signal.
Negative
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5 6
Page 67
2.10. To Store and Recall
Press Store/Recall button to enter the following interface. You can save or
recall the state documentation inside the generator. The state file on the U
Disk can also be recalled or deleted. File names can only be in English. User
can only recall or delete the data documentation saved via CSV from the
Oscilloscope.
Figure 2- 49 Save and Read Interface
Figure 2-50 Table 2-26 Menu Explanations of Save and Recall (Page 1/2)
Function
Menu
File Type
Browser
Save
Recall
Delete Delete the selected file.
Settings Explanation
State
Data
Directory
File
The setting of the generator;
Arbitrary waveform file;
Shift between the directory and file.
Save the waveform to the appointed
place.
Recall the waveform or setting
information in the specific position of the
memory.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5 7
Page 68
About the browser
The directory selection shift is done by the direction keys. In the directory
mode, pressing the right key will open the lower directory while the left key
will fold the directory. Up and down key are used to shift between the
directories;
To Save the Instrument State
Users are allowed to store the instrument state in any of the 10 non-volatile
memories. The state storage will ‘memorize’ the selected function (including
the arbitrary waveform, frequency, amplitude, DC offset, duty cycle, symmetry,
and other modulation parameter used.)
To save the instrument state, the procedures are given as followed:
1. Choose the file type to store
Press Store/Recall →Type→ State, and choose state as the storage
type.
2. Choose the location of the file.
There are ten positions in the Local C:, choose any one of them by
rotating the knob.
3. Name the file and save it
Press Save button, enter the desired name. Press Save to finish.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5 8
Page 69
To Use USB Storage
As is shown in Figure 2-51, the storage location is divided into: The internal
storage Local(C: ) and the U Disk storage USB Device (A: ). At the left side of
the front panel, there is a USB interface. When USB storage is connected,
the storage menu will show ‘USB Device (A:)’. Otherwise, the default location
is the internal location Local(C: ).
Figure 2-51 USB Storage Interface
1. Install the USB Device
Insert the USB Device into the USB interface on the front panel, and the
screen will show ‘USB flash device plug in’, and storage menu will show
‘USB Device (A :)’
2. Choose the USB Device
Press Browser->Directory, move the cursor with the up or down direction
key to select ‘USB Device (A: )’. Press the right key to open the lower
directory, use the up and down direction key to choose the file ‘T3AFG5-10’.
Use the right key to open the lower directory, and up and down key to
select the file ‘Workspace’. Input the file name and save.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5 9
Page 70
3. Remove the USB Device
Remove the USB Device from the interface. The system will inform you
‘USB flash device plug out’, and the ‘USB Device (A: )’ in the storage
menu will disappear.
Note: USB Device can only be use U Disk; portable hard disk is not
supported.
To Save a File
Press Store/Recall →Store, to enter the following interface. Enter the desired
file name in the ‘File Name’ frame. In the middle of the figure below is the
input keypad, used to edit the file name. Use the up and down direction keys
and knob to select the desired character; use the left and right direction keys
to edit the input file name.
Figure 2-52 File Storage Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6 0
Page 71
Figure 2-53 Table 2-27 Menu Explanation of File Storage
Function
Menu
Input
Type
Select Select the current character.
Delete Delete the current character.
Save Store the file with the current name
Settings Explanation
En English input.
1. English Input
The English input interface is as shown in Figure 2-54, to save a file named
‘NEWFILE’, follow the steps below:
Figure 2-54 English Input Interface
(1) Press InType->En, to enter the English interface.
(2) Input the file name’ NEWFILE’.
Use the Knob to adjust the cursor’s horizontal position and the up and
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6 1
Page 72
down key to adjust the vertical position. Select the Character ‘N’ and
press Select. Repeat this until you have inputted ‘NEWFILE’.
(3) Edit the File Name
When you have entered a wrong character, move the cursor to the
wrong character to be deleted and press Delete to remove it. Re-enter
the file name.
(4) Press Save, to finish and save the file.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6 2
Page 73
2.11. To Set the Utility Function
With the Utility Function, you can set the parameters of the generator such as:
DC On/Off, Sync On/Off, Output Parameter, Interface Parameter, System
Setting and Testing Parameter. The DC On/Off offers the options of DC output
or Arbitrary Waveform Output. Sync On/Off offers the option to choose the
Sync Signal or not. Output Setting provides the parameter setting for
Load/HighZ and Normal/Inverse. The System Setting provides the setting for
Language, Display, Beep, Screen Guard, Format, Power System Configure
and default setting; Test provides the self-testing and calibration function.
Press Utility button, to enter the Utility Menu. Its functions are listed below in
Figure 2-55
Figure 2-55 Table 2-28 Menu Explanations of Utility System Setting (Page1/2)
Function
Menu
DC
Output
Setup
Sync Set the sync output
Settings Explanation
On
Off
Set the output waveform to be DC. Set
the output waveform to be arbitrary.
Set the output parameter.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6 3
Page 74
Figure 2-56 Table 2-29 Menu Explanations of Utility System Setting (Page2/2)
Function
Menu
System Set the system configuration.
Test/Cal Test and calibrate the instrument.
EditInfo Information of the system.
Update Update function.
Settings Explanation
To Set the DC Output
Press Utility →DC→DC On, to enter the following interface. Please note
that there is a ‘DC On’ sign at the middle left of the screen.
Figure 2- 57 DC Setting Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6 4
Page 75
DC Offset
Set the DC voltage level.
To Shift into the Arbitrary Waveform Output
1. Press Utility →DC→DC Off, to close DC output and return to arbitrary
waveform output.
2. Press any functional button, and the waveform output setting turns into
the arbitrary waveform output. The DC output is turned off automatically.
To Set IO
Press Utility →IO Setup, to set the IO interface. The equipment supports
RAW protocol and TMC protocol, user may setup corresponding protocol by
IO Setup.
To Set Output Parameter
Press Utility →Output Setup, to enter the following interface.
Figure 2-58 Table 2-30 Menu Explanations of Output Setting (Page 1/2)
Function
Menu
Load
HighZ
Normal
Invert
Settings Explanation
Set the load connected to the Output
Connector;
Set the load connected to the Output
Connector to be HighZ.
Normal output;
Inverse output.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6 5
Page 76
1. To Set the Output Load
For the [Output] connector on the front panel, the generator has a built-in
50Ω series impendence. If the actual load does not match the set one,
the displayed amplitude and offset will be incorrect. This function is used
to match the displayed voltage with the expected one.
Steps for setting the load: Press Utility →Output Setup→ Load, to enter
the following interface. Please note that the load parameter shown on the
right bottom is the default setting when the power is on or the pre-set
load value. If the current value is valid for the output, then current value
will be used.
Figure 2-59 Set the Output Load
Instruction
T3AFG5-10 Series has a fixed 50Ω Series Impendence. No matter what value
the set parameter is, if the real load is different from the set one, the displayed
voltage will not equal the real voltage.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6 6
Page 77
2. To Set the Invert Waveform
Press Utility →Output Setup→ Invert, to set the Inverse Waveform Output.
When the waveform is in inverse mode, offset is disabled.
3. To Set the Sync Output
The generator provides Sync output through the [Sync] connector on the rear
panel. All standard output functions (except DC and Noise) have a
corresponding Sync signal. For some applications, they can be disabled if
users do not want to use it,
l In the default setting, the Sync signal should be connected to the
[Sync] connector (activated). When the Sync Signal is disabled, the
output voltage of the [Sync] connector is level low.
l In the Inverse Mode, the Waveform that corresponds to the Sync
Signal does not invert.
l The Sync Signal is a Pulse Signal with fixed positive pulse width,
which is more than 50ns.
l For non-modulated waveform, the Sync Signal reference is the
carrier.
l For internal modulating AM, FM and PM, the Sync signal reference
is the modulated signal (not the carrier signal).
l For ASK and FSK, the Sync Signal Reference is the keying
Frequency.
l For a Sweep, when the sweep starts, the Sync Signal becomes TTL
Level High. The Sync frequency equals the specific Sweep time.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6 7
Page 78
l For the Burst, when the burst starts, the Sync Signal is Level High.
l For the External Gated Burst, the Sync Signal follows the External
Gated Signal.
To Set the System
Press Utility → System, to enter the following interface.
Figure 2-59 Table 2-31 Menu Explanations of System Setup (Page 1/2)
Function
Menu
Number
format
Language Set the display language.
Power On
Set to
Default
Figure 2-60 Table 2-32 Menu Explanations of System Setup (Page 2/2)
Function
Menu
Beep
ScrnSvr
Settings Explanation
Set the number format.
All the settings return to default when
Default
Last
Settings Explanation
On
Off
1min
5min
15min
30min
1hour
2hour
5hour
Off Deactivate the screen saver program.
powered;
All the settings return to the last one
when powered.
Set all the settings to default.
Open beep;
Close beep.
Activate the screen saver program.
Screen saver will be on if no action is
taken within the time that you have
selected. Press any button the resume.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6 8
Page 79
Key points:
Power On
Choose the configuration setting when the machine is powered on.
Two choices are available: the default setting and the latest. Once selected,
the setting will be used when the instrument is first powered on.
Beep
Activate or deactivate the sound when an error occurs from the front panel or
the remote interface. Activate or deactivate any sound made by the button or
knob on the front panel. The current setting is stored in the non-volatile
memory.
1. Set the Format
Press Utility →System→ Number Format, to enter the following interface.
Figure 2-61 Set the number Format
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6 9
Page 80
Figure 2-62 Table 2-33 Menu Explanations of Setting the Number Format
Function
Menu
Point
Separator
Settings Explanation
Using dot to represent point;
Using comma to represent point.
On
Off
Space
Enable the Separator;
Close the Separator;
Use Space to separate.
According to the different choices of the point and the separator, the format
can have various forms.
(1) as point, press Separator->On, the example is as followed:
Figure 2- 63 Set Format
(2) as point, press ->Separator->On, the example is as followed:
Figure 2- 64 Set Format
(3) as point, press Separator->Off, the example is as followed:
Figure 2-65 Set Format
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7 0
Page 81
(4) as point, press Separator->Off, the example is as followed:
Figure 2-66 Set Format
(5) as point, press Separator->Space, the example is as followed:
Figure 2-67 Set Format
(6) as point, press Separator->Space, the example is as followed:
Figure 2-68 Set Format
2. Language Setup
The T3AFG5-10 Series Generator offers two languages (English and Simplified
Chinese) for user to choose.
To Select Language, press Utility and then Language to select the language.
The Procedure is as followed:
Press Utility →System→ Language, to change the language.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7 1
Page 82
3. To Return to Default Setting
Press Utility →System→ Set to Default, to set the system to the default
setting. The default settings of the system are as followed:
Table 2- 34 Factory Default Setting
Output Default
Function Sine Wave
Frequency 1kHz
Amplitude/Offset 4Vpp/0Vdc
Phase 0°
Terminals High Z
Modulation Default
Carrier 1kHz Sine Wave
Modulating 100Hz Sine Wave
AM Depth
FM Deviation 500Hz
Key Freq 100Hz
Key Freq 100Hz
FSK Hop Frequency 1MHz
Phase Deviation 180°
Sweep Default
Start/Stop Frequency 100Hz/1.9kHz
Sweep Time 1S Trig
Out Off
Mode Linear
Direction ↑
Burst Default
Period 10ms
Phase 0°
Count 1Cycle
Trig Off
Trigger Default
Source Internal
100
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7 2
Page 83
2.12. Test/Cal
Press Utility →Test/Cal, to enter the following menu.
Figure 2-69 Test/Cal function Menu
Figure 2-70 Table 2-35 Menu Explanations of Test Setting
Function
Menu
SelfTest Perform system self-test.
SelfCal Do self calibration
Settings Explain
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7 4
Page 84
SelfTest
Press Utility →Test/Cal →SelfTest, to enter the following menu.
Figure 2-71 Table 2-36 Menu Explanations of Self Test
Function
Menu
Scr Test Run screen test program.
Key Test Run keyboard test program.
LED Test Run LED test program.
Settings Explain
1. Scr Test
Select Scr Test to enter the screen test interface. The words ‘Press ‘7’ Key
to continue, Press ‘8’ Key to exit’ is displayed. You could press the ‘7’ for test.
Figure 2- 72 Screen Test Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7 5
Page 85
2. Key Test
Select ‘keyboard Test’ to enter the keyboard test interface, the on-screen
rectangle shapes represent the front panel keys. The shapes with two
arrows beside them represent the front panel knobs. Test all keys and knobs
and you should also verify that all the backlit buttons illuminate correctly.
Note:
l When you operate, the screen would display the white (color LCD).
l The tested button or knobs corresponding area would display green
(color LCD).
l At the bottom of the screen display ‘Press ‘8’ Key Three Times to
exit’ information is displayed to indicate that pressing ‘8’ three times
Is required to exit the test.
Figure 2-73 Key Test Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7 6
Page 86
3. LED Test
Select ‘LED Test’ to enter the below interface. The on-screen rectangle
shapes represent the front panel keys; the shapes with two arrows beside
them represent the front panel knobs. The instructions ‘Press ‘7’ Key to
continue, ‘Press ‘8’ Key to exit’ is displayed, You could press the ‘7’ button
continuously for testing, when buttons are lit the corresponding area on the
screen would display green (color LCD).
Figure 2-74 Led Test Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7 7
Page 87
SelfCal
Press Utility →1/2→Test/Cal →SelfCal, to enter SelfCal, as is shown in
Figure 2-5
SelfCal: If the environment where the generator is being used changes,
then it is recommended to do a self calibration.
Figure 2-75 SelfCal Interface
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7 8
Page 88
2.13. Edition Information
Press the EditInfo option button of the Utility Menu to view the generator’s
hardware and software configuration.
Figure 2-76 Edit Info Interface
Edition Information introduction
Boot-strap No:
The times of boot-strap
Software version:
Software version of current equipment
Hardware version:
02-00-00-21-25 represents ordinally: PCB version, BOM version, Daughter
card version, FPGA version, CPLD version.
Model:
Contains information of brand of product, series, bandwidth.
For example: T3AFG5 represents Teledyne Test Tools T3AFG5-10 series
function/arbitrary waveform generator with 5MHz bandwidth.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7 9
Page 89
Serial No:
The serial number information includes a date code.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8 0
Page 90
2.14. Updating Firmware
Using USB flash drive to update firmware
The software of the generator can be updated directly via USB flash drive.
This process takes about two minutes. Follow the next steps:
1. Insert USB flash drive with firmware procedure to USB host interface on
the front panel of the generator.
2. Press the Utility button to enter the ‘Utility Menu’.
3. Press ‘1/3’ option button to enter the third page of ‘Utility Menu’.
4. Press the ‘Update’ option button.
5. Press ‘Browser’ option button to select ‘Directory’, then select the ‘USB
Device (A:)’ through direction key.
6. Press ‘Browser’ option button to select ‘File’, then select the
‘XXXX.ADS’ file using the direction keys.
7. Press the ‘Recall’ option button to update.
8. After finishing the update, restart the generator.
Note : Don’t disconnect the power whilst the generator is being updated.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8 1
Page 91
2.15. How to Use the Built-in Help System
You can get help for every button on the front panel by using the built-in help
system. Or you can get help about the operation of the front panel
buttons with the help list.
Press Help to enter the following interface.
Figure 2-77 Help Menu
Figure 2-78 Table 2-37 Help Menu Explanations
Function
Menu
Choice Select to read the information.
Settings Explanation
Cursor upward to select.
Cursor downward to select.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8 2
Page 92
3. Application and Examples
To help the user master how to use the Function/ Arbitrary Waveform
Generator more efficiently, we will describe some examples in detail. All the
examples below use the default setting of the instrument with some
Explanation on setup.
This chapter includes the following topics:
l Example 1: Generate a Sine Wave
l Example 2: Generate a Square Wave
l Example 3: Generate a Ramp Wave
l Example 4: Generate a Pulse Wave
l Example 5: Generate a Noise Wave
l Example 6: Generate an Arbitrary Wave
l Example 7: Generate a Sweep Wave
l Example 8: Generate a Burst Wave
l Example 9: Generate an AM Wave
l Example 10:Generate a FM Wave
l Example 11:Generate a PM Wave
l Example 12:Generate a FSK Wave
l Example 13:Generate an ASK Wave
l Example 14:Generate a PWM Wave
l Example 15:Generate a DSB-AM Wave
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8 3
Page 93
3.1. Example 1:Generate a Sine Wave
Generate a sine wave with 50kHz frequency, 5Vpp amplitude and 1Vdc
offset.
Ø Steps:
l Set the frequency.
1. Press Sine →Freq and choose frequency which will display in white color.
2. Input ‘50’ from the keyboard and choose the unit ‘kHz’. The frequency is
set to be 50kHz.
l Set the amplitude.
1. Press Ampl to choose Ampl which will display in white color.
2. Input ‘5’ from the keyboard and choose the unit ‘Vpp’. The amplitude is
set to be 5Vpp.
l Set the Offset.
1. Press Offset to choose Offset which will display in white color
2. Input ‘1’ from the keyboard and choose the unit ‘Vdc’. The offset is set to
be 1Vdc.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8 4
Page 94
When the frequency, amplitude and offset are set, the wave generated is
shown in Figure 3- 1;
Figure 3-1 Sine Waveform
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8 5
Page 95
3.2. Example 2:Generate a Square Wave
Generate a square wave with 5kHz frequency, 2Vpp amplitude, 0Vdc offset
and 30% duty cycle.
Ø Steps:
l Set the frequency.
1. Press Square →Freq and choose Frequency which will display in white
color.
2. Input ‘5’ from the keyboard and choose the unit ‘kHz’. The frequency is
set to be 5kHz.
l Set the amplitude.
1. Press Ampl to choose Ampl which will display in white color.
2. Input ‘2’ from the keyboard and choose the unit ‘Vpp’. The amplitude is
set to be 2Vpp.
l Set the offset.
1. Press Offset to choose Offset which will display in white color
2. Input ‘0’ from the keyboard and choose the unit ‘Vdc’. The Offset is set to
be 0Vdc.
l Set the duty
1. Press Duty to choose Duty which will display in white color
2. Input ‘30’ from the keyboard and choose the unit ‘%’. The duty is set to be
30%.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8 6
Page 96
When the frequency, amplitude, offset and duty cycle are set, the wave
generated is shown in Figure 3-2.
Figure 3- 2 Square Waveform
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8 7
Page 97
3.3. Example 3:Generate a Ramp Wave
Generate a ramp wave with 10μs period, 100mVpp amplitude, 20mVdc offset,
45°phase and 30% symmetry.
Ø Steps:
l Set the period.
1. Press Ramp →Freq and choose Period which will display in white color.
2. Input ‘10’ from the keyboard and choose the unit ‘μs’. The period is set to
be 10μs.
l Set the amplitude.
1. Press Ampl to choose Ampl which will display in white color.
2. Input ‘100’ from the keyboard and choose the unit ‘mVpp’. The amplitude
is set to be 100mVpp.
l Set the offset.
1. Press Offset to choose Offset which will display in white color
2. Input ‘20’ from the keyboard and choose the unit ‘mVdc’. The offset is set
to be 20mVdc.
l Set the phase
1. Press Phase to choose Phase which will display in white color
2. Input ‘45’ from the keyboard and choose the unit ‘ °’. The phase is set to
be 45°.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8 8
Page 98
l Set the symmetry
1. Press Symmetry to choose Symmetry which will display in white color.
2. Input ‘30’ from the keyboard and choose the unit ‘30%’. The symmetry is
set to be 30%.
When the period, amplitude, offset, phase and symmetry are set, the wave
generated is shown in Figure 3-3:
Figure 3-3 Ramp Waveform
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8 9
Page 99
3.4. Example 4:Generate a Pulse Wave
Generate a pulse wave with 5kHz frequency, 5V high level, -1V low level,
40μs pulse width and 20ns delay.
Ø Steps:
l Set the frequency.
1. Press Pulse → Freq and choose Freq, which will display in white color.
2. Input ‘5’ from the keyboard and choose the unit ‘kHz’. The frequency is
set to be 5 kHz.
l Set the high level
1. Press Ampl and choose the HLevel which will display in white color.
2. Input ‘5’ from the keyboard and choose the unit ‘V’. The high level is set
to be 5V.
l Set the low level
1. Press Offset and choose the LLevel which will display in white color.
2. Input ‘-1’ from the keyboard and choose the unit ‘V’. The low level is set to
be -1V.
l Set the pulse width
1. Press PulWidth and choose PulWidth which will display in white color.
2. Input ‘40’ from the keyboard and choose the unit ‘μs’. The pulse width is
set to be 40μs.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
9 0
Page 100
l Set the Rising Edge
1. Press Rising Edge and choose Rising Edge which will display in white.
2. Input ‘20’ from the keyboard and choose the unit ‘ns’. The delay is set to
be 20ns.
When the frequency, high level, low level, pulse width and delay are set, the
wave generated is shown in Figure 3-4
Figure 3-4 Pulse Waveform
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
9 1