Read the following precautions carefully to avoid any personal injuries, or damage to the instrument or products connected to it. Use the instrument only as specified.
Use only the power cord supplied for the instrument.
Ground the instrument.
The instrument is grounded through the ground conductor of the power cord. To avoid electric
shock, always connect to grounded outlets. Make sure the instrument is grounded correctly before
connecting its input or output terminals.
Connect the signal wire correctly.
To avoid damage, observe input polarity and maximum voltage/current ratings at all times.
Observe all terminal ratings and signs on the instrument to avoid fire or electric shock.
Before connecting to the instrument, read the manual to understand the input/output ratings.
Do not operate with suspected failures.
If you suspect that the instrument is damaged, contact the Teledyne LeCroy service department
immediately.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep the surface of the instrument clean and dry.
Avoid touching exposed circuits or wires.
Do not touch exposed contacts or components when the power is on.
Do not operate without covers.
Do not operate the instrument with covers or panels removed.
Use only the fuse specified for the instrument.
Use proper overvoltage protection.
Use anti-static protection.
Operate in an anti-static protected area. Ground measurement cable conductors before connecting
to the instrument to discharge any static electricity before connecting the cables to the instrument.
Observe ventilation requirements.
Ensure good ventilation. Check the vent and fan regularly to prevent overheating.
i
T3DAQ1-16 User Manual
Safety Terms and Symbols
The following terms may appear on the instrument:
DANGER: Direct injury or hazard may occur.
WARNING: Potential injury or hazard may occur.
CAUTION: Potential damage to instrument/property may occur.
CAT I1: IEC Measurement Category I, applicable for making measurements on ‘other’ circuits that
are not directly connected to mains. See p. v.
CAT II: IEC Measurement Category II, applicable for making measurements on circuits connected
directly to utilization points (socket outlets and similar points) of the low voltage mains installation.
See p. v.
1) CAT I as defined in IEC/EN 61010-031:2008.
Note that Measurement Category I was removed in IEC/EN 61010-031:2015 and replaced by ‘O’, indicating ”other cir-
cuits that are not directly connected to mains.”
Saftey Symbols
The following symbols may appear on the instrument:
ii
T3DAQ1-16 User Manual
Operating Environment
Temperature: 0 °C to 40 °C
Humidity: 5% to 90% relative humidity (non-condensing) up to +30°C. Upper limit derates to 50%
relative humidity (non-condensing) at +40°C.
Altitude: ≤ 2000 m
Use indoors only.
Pollution Degree 2. Use in an operating environment where normally only dry, non-conductive
pollution occurs. Temporary conductivity caused by condensation should be expected.
AC Power
Input Voltage & Frequency:
100-120 V at 50/60 Hz
200-240 V at 50/60 Hz
Manual AC selection with a slide switch.
Power Consumption: 20 W maximum
Mains Supply Connector: CAT II per IEC/EN 61010-1:2010, instrument intended to be supplied
from the building wiring at utilization points (socket outlets and similar).
Fuse Type
Current Input Terminal: 250 VAC F 10 A, 3 AG
AC Mains: 250 VAC F 300 mA, 5x20 mm
iii
T3DAQ1-16 User Manual
Input Terminal Protection Limitation
Protection limitation is defined for the input terminal.
1. Main Input (HI and LO) Terminals
HI and LO terminals are used for Voltage, Resistance, Capacitance, Continuity, Frequency,
Diode and Temperature measurement. Two protection limitations are defined:
• HI-LO protection limitation: 1000 VDC or 750 VAC. This is the maximum measurable
voltage. The limitation can be expressed as 1000 Vpk.
• LO ground protection limitation: LO terminal can “float” 500 Vpk relative to the ground
safely. The maximum protection limitation of HI terminal relative to the ground is 1000
Vpk. Therefore, the sum of the “float” voltage and the measured voltage can’t exceed
1000 Vpk.
2. Sampling (HISense and LOSense) Terminals
HISense and LOSense terminals are used for 4-wire Resistance measurement. Two protection limitations are defined:
3. Current Input (I) Terminal
The I terminal is used for current measurement. The maximum current which can go through
the I terminal is limited to 10 A by the fuse on the back panel.
NOTE: Voltage on the current input terminal corresponds to voltage on the LO terminal. To
keep good protection, only use a fuse of the specified type and value to replace this fuse
iv
T3DAQ1-16 User Manual
IEC Measurement Category II Overvoltage Protection
To avoid the danger of electric shock, the Data Acquisition Unit provides overvoltage protection for
line-voltage mains connections that meet both of the following conditions:
1. The HI and LO input terminals are connected to the mains under Measurement Category II
conditions described in the warning below.
2. The maximum line voltage of the mains does not exceed: 600 VAC for the instrument.
WARNING:
IEC Measurement Category II includes electrical devices connected to mains at an outlet on a
branch circuit, such as most small appliances, test equipment, and other devices that plug into a
branch outlet or socket.
The DAQ is capable of making measurements with the HI and LO inputs connected to mains in
such devices (≤ 600 VAC) or to the branch outlet itself.
However, the HI and LO terminals of the DAQ can’t be connected to mains in permanently installed
electrical devices such as the main circuit-breaker panels, sub-panel disconnected boxes and permanently wired motors. Such devices and circuits are prone to exceed the protection limits of the
DAQ.
Limits for Measurements on Other Circuits Not Directly Connected to Mains
Max. rated input voltage: 1000V
Transient overvoltage: 4000V
WARNING:
Voltages above 600 VAC can only be measured in circuits that are isolated from mains. However,
there may be transient over voltage in circuits that are isolated from mains. The DAQ is able to
withstand occasional transient overvoltage up to 4000 Vpk. Please don’t use this instrument to
measure circuits where transient overvoltage may exceed this level.
rms
pk
v
T3DAQ1-16 User Manual
Daily Maintenance and Cleaning
Maintenance
Protect the liquid crystal display from direct sunlight when storing or using the instrument.
NOTE: To avoid damage to the instrument or test leads, please don’t place them in mist, liquid or
solvent.
Cleaning
Regularly clean the instrument and test leads.
• Wipe the external dust off the instrument and test leads using a soft rag. Be careful not to
scratch the display screen when cleaning. Do not allow any liquid to enter the instrument.
• Use a damp soft rag to clean the instrument after removing the power plug. Use 75% isopropyl alcohol or water solvent to get a more thorough cleaning.
NOTE: To prevent the surface of the instrument or test leads from damage, please don’t use any
corrosive or chemical cleaning reagents. Please make sure the instrument is fully dry before reconnecting the power to avoid short circuits or personal injury.
vi
T3DAQ1-16 User Manual
Introdution to T3DAQ1-16
Teledyne Test Tools T3DAQ1-16 is a 16 channel Data Acquisition System incorporating the latest
4.3 inch (10.92 cm) dual-display technology which can be configured to show Data Histograms,
Data fluctuation Trends, Bar Graph, Statistics or the traditional Number mode, all in an easy to use
interface. The T3DAQ features 12 multi-purpose + 4 current channels and supports various measurement functions. It provides a convenient and versatile solution for test applications that require
multiple measurement points or signals and is an ideal tool for R& D burn-in and production testing.
A great feature of the Teledyne Test Tools T3DAQ is its ability to make highly accurate True RMS
AC Voltage and Current measurements, meaning no loss of accuracy even when measuring complex voltage and current waveforms.
The T3DAQ provides various measurement modes to satisfy a wide range of application requirements. 12 Multi-purpose channels can be configured individually for various scan requirements.
The Channel configuration function allows user to configure different parameters such as measurement type, range, scan mode, number of scan cycles, and duration of scan. Similarly, the
scan control menu provides flexibility to schedule, start, and stop the scan. Scanning data can be
viewed live during a scan and can also be saved for future use. An internal 1 GB flash memory is
available for data logging and to save configuration files. The T3DAQ also features a USB Host
interface to use a USB flash drive to collect data without being connected to a PC.
The T3DAQ is equipped with LAN and USBTMC interfaces for remote control of the device. The
device can be controlled remotely using EasyDMM software. The graphical interface of the software makes it easy to configure measurements, setup and execute scans or troubleshoot the
design. Live measurement data can be viewed in numeric digits or graphically by trend chart, bar
graph and histogram. The data can be manually or automatically exported as CSV files for further
analysis.
Main features:
• 4.3” TFT-LCD Display
• Dual display, Chinese and English Menu
• Built-in front panel accessible help system
• File management (support for U-disc and local storage)
• Real 6-5 digit (2,200,000 count) readings resolution
• True-RMS AC Voltage and AC Current measuring
• 1 GB flash memory for mass storage configuration files and data files
• Built-in cold terminal compensation for thermocouple
• Standard interface: USB Device, USB Host, LAN
• USB & LAN remote interfaces support common SCPI command set
Teledyne Test Tools is a brand and trademark of Teledyne LeCroy, Inc. Other product or brand
names are trademarks or requested trademarks of their respective holders. Specifications, prices,
availability and delivery subject to change without notice.
ix
Chapter 1
Quick Start
General Inspection
Please check the instrument according to the following steps.
1. Inspect the shipping container.
Keep the shipping container and packaging material until the contents of the shipment have
been completely checked and the instrument has passed both electrical and mechanical
tests. It is always good practice to save the shipping container and packaging for use when
returning the power supply to Teledyne LeCroy for service or calibration.
The consigner or carrier will be responsible for damage to the instrument resulting from shipping. Teledyne LeCroy will not provide free maintenance or replacement in this instance.
2. Inspect the instrument.
If the instrument is found to be damaged, defective or fails in electrical or mechanical tests,
please contact the Teledyne LeCroy service department immediately.
3. Check the accessories.
Please check that you have received the accessories on the packing list:
1 DAQ
1 Test Lead Set
1 USB Cable
If the accessories are incomplete or damaged, please contact Teledyne LeCroy immediately.
1
T3DAQ1-16 User Manual
Dimensions
2
T3DAQ1-16 User Manual
Handle Adjustment
To adjust the handle position of the DAQ, grip the handle by the two sides and pull outward. Then,
rotate the handle to the appropriate position.
3
T3DAQ1-16 User Manual
Front Panel
A USB Host
Users can store the current state or measurement data into a USB storage device. Users
can also read the state files or updated firmware from a USB storage device.
B Power Key
Turn the instrument on or off.
C LCD Display
The instrument provides a 4.3 inch high resolution color TFT-LCD display screen with 480*272
pixels that displays the function menus, measurement parameter settings, system status, and
prompt messages.
D Menu Operation Keys
Press any softkey to activate the corresponding menu.
4
T3DAQ1-16 User Manual
E Measurement and Assistant Function Keys
5
T3DAQ1-16 User Manual
F Range and Direction Keys
G Signal Input Terminals
The measured signal (device) will be connected into the DAQ through these terminals. Different measurement objects have different connection methods.
For details, please refer to “Measurement Connections”.
6
T3DAQ1-16 User Manual
Rear Panel
A Power Socket
The DAQ accepts two types of AC supplies. Please use the power cord provided
in the accessories to connect the DAQ to the AC power through this socket.
Note: The correct voltage scale must be first selected (through the Voltage Selector) before power connection.
B Power Fuse
The DAQ is already installed with a power fuse before leaving factory. To change the fuse
please:
• Turn off the DAQ and remove the power cord.
• Press down the block tongue using a straight screwdriver (in the direction of the dotted
arrow in the figure below) and pull out of the fuse seat.
• Select a proper voltage scale.
• Replace a specified fuse.
• Reinstall the fuse seat into the slot.
7
T3DAQ1-16 User Manual
C AC Voltage Selector
Select the correct voltage scale (110 V or 220 V) for the AC supply used.
D 16-channel Data Acquisition Module
E USB Device
Connect the PC through this interface. SCPI commands or PC software can be used to
control the DAQ remotely.
F LAN
Through this interface, the DAQ can be connected to the network for remote control.
G VMC Output
The DAQ outputs a low-true pulse from the [VM Comp] connector after every measurement.
H Ext Trigger
Trigger the DAQ by connecting a trigger pulse through the [Ext Trig] connector. Note the
external trigger source must be selected.
I Current Input Fuse
The DAQ is already installed with a current Input fuse to provide 10 A maximum input protection before leaving factory. To replace a new one, please:
(a) Turn off the DAQ and remove the power cord.
(b) Turn the fuse seat counterclockwise as shown in the figure using a straight screw driver
and then pull out the fuse seat
(c) Place a new 10 A specified fuse.
(d) Reinstall the fuse seat into the slot.
J Instrument Kensington Lock Point
A Kensington lock (not supplied) can be used to lock the DAQ to a fixed place if necessary.
8
T3DAQ1-16 User Manual
Starting the DAQ
Before connecting the instrument to a power source please adjust the AC voltage
selector on the rear panel of your DAQ according to your local power supply voltage.
Then connect the power cord as shown in the following figure.
Press the Power key on the front panel to turn on the DAQ. If the DAQ does not start normally,
then try the following:
1. Make sure the power cord is in good connection and connected to the DAQ and the wall
socket.
2. Ensure that the wall socket has power and is turned on.
3. Try to restart the DAQ, if it fails, check the power fuse and replace with a new one if necessary.
4. If the problem still remains, please contact the Teledyne LeCroy service department for help.
9
T3DAQ1-16 User Manual
Operating Instructions
NOTE: The Scanner Card is not designed to be ”Hot Swappable”.
Please make sure the instrument is turned off before installation or removal of the Scanner Card.
Hot swapping the card could cause damage to the instrument and is not covered under warranty.
10
T3DAQ1-16 User Manual
Scanner Card Connections
An example of 2-wired connection application is shown below.
First 12 channels can be used to measure DCV/DCI ACV/ACI/CAP/FREQ/DIODE/CONT/TEMP
and the last 4 channels are dedicated to current measurement.
4-Wire Resistance measurement example is shown below.
CH1 to CH6 can be used for measurements and CH7 to CH12 can be used as sense channels. To
minimize voltage errors, the remote sense connections (CH7,CH8 etc..) should be made as close
to the device-under-test (DUT) as possible.
11
T3DAQ1-16 User Manual
Front Panel Operations
Scanner mode functions can be accessed by pressing ”Shift” and ”Temp” buttons.
The following table gives a short description about different functions available.
Function
MenuSettingsDescription
ModeScan/StepSet the operation mode
Sets the duration between
each scan loop (Scan mode)
0ms-
Time
CyclesAuto/Man
Channel
Setup
StartOn/OffStart or stop scan operation
ExitExit the scanner function
1. Operation mode setup
•Scan: Measures all specified channel ranges (Channel MIN-MAX) for each trigger event. Time
settings are applied between each scan for the whole channel range.
•Step: Measures a single channel in the specified range (Channel MIN-MAX) at each trigger event.
Time settings are applied for each channel.
2. Time setup
Use the direction keys to set the duration between each scan loop (Scan mode) or between each
scanned channel (Step mode)
999.999s
or between each scanned
channel (Step mode)
Sets the number of scan operations
Sets the scanned channel
range,measurement function, and measurement parameters
3. Cycles
•Auto: The instrument will scan specified channel circularly after the scan operation start and you
should stop the operation manually.
12
T3DAQ1-16 User Manual
•Manual: Sets the number of scan operations by direction keys. The range of the setting is from 1
to 999. After starting the scan operation the instrument will not stop scanning until reaches cycle
number.
4. Channel setup
Press the [Channel Setup] to enter the setup interface.
The table below gives a brief description of Function menu.
Function
MenuDescription
Open/close the channel and set the meaChannel Config
surement function, measurement parame-
ters of specified channel.
LowSet the low value of scanned channel range.
HighSet the high value of scanned channel range.
Press the [Channel Setup] to enter the channel configuration interface and set the channel switch, function,
range and speed.
The range setting is applicable for the following functions: DC/AC Voltage (DCV/ACV), 2/4 Wire Resistance
(2W/4W), Capacitance (CAP), Frequency (FRQ).
Ranges available for various functions is given below.
13
T3DAQ1-16 User Manual
Function
MenuDescription
DCV/ACV/
FRQAuto, 200mV, 2V, 20V, 200V
DCI/ACI2A (fixed)
Auto, 200Ω, 2kΩ, 20kΩ, 200kΩ, 2MΩ, 10MΩ,
2W/4W
The scanner function provides two measurement speeds: Fast (50 reading/s) and Slow (5 reading/s). (Fast:
1PLC, Slow: 10 PLC)
The speed setting is applicable for the following functions: DC/AC Voltage (DCV/ACV), 2/4 Wire Resistance
(2W/4W)
100MΩ
Operating instructions:
• Move the cursor to choose the wanted parameter by direction keys and the background color of
cursor’s position turns to gray.
• Select the current item by pressing “OK” key and the background color of the selected item turns
to green.
• Set the parameter by up and down direction keys.
• Press “OK” key again to store the setting of the selected item of which the background turns back
to gray. Move the cursor and repeat the prior steps to set the next parameter.
• Press [Done] to save the current settings and return to the higher level menu.
5. Channel range setting
Select [High] or [Low] and then input numerical value by direction keys. Note: The upper limit value
should be always bigger than the lower limit value.
6. Start scan operation
Set the [Start] to on to start the scan operation
The upper part of the interface displays the function, range, and result of the current channel. The
table below records the measurement result of each channel.
7. Enter trend chart and statistics mode (optional operation)
Press ”Shift” and ”Math” to open trend chart and statistics function.
14
T3DAQ1-16 User Manual
Set the [Inquire Channel] by direction keys and the interface displays the minimum, average, maximum, span, standard deviation, samples and trend chart of measurement results of the setting
channel during the scan operation. Press [Done] to return to higher level menu.
8. Stop scan operation
Set the [Start] to off to stop the scan operation. If the [Cycles] is set to manual, then the instrument
will stop scan when it reaches the cycle number
9. Store measurement data (optional operation)
Press ”Shift” and ”Dual”, then Select [Store/Recall] → [Store Settings] to enter the storage function
interface.
15
T3DAQ1-16 User Manual
User Interface
16
T3DAQ1-16 User Manual
Measurement Connections
The DAQ is designed with many measurement functions. After selecting the desired measurement function, please connect the signal (device) under test to the DAQ according to the method
below. Do not switch the measurement function when measuring as it may cause damage to the
DAQ. For example, when the test leads are connected to the related current terminals, AC voltage
measurement should not be used.
17
T3DAQ1-16 User Manual
18
T3DAQ1-16 User Manual
19
T3DAQ1-16 User Manual
Connecting to USB and LAN Ports
The DAQ has LAN and USB I/O ports. Connect to the ports as in the diagrams below:
20
Loading...
+ 81 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.