Teledyne ISCO TN11 Instruction Manual

Adjustable Rf Solid Load Cartridge Cap for Redi
Sep
Product Information and Instructions
®
Rf Cartridges
Chromatography Technical Note
TN11, June 2020
Overview
Three cap sizes are available (refer to Table 1) to fit 5, 25, and 65 gram RediSep Rf empty or prepacked solid load cartridges for use with the CombiFlash® system.
Designed for Convenience
The design of the solid load sample cartridge cap is simple, yet effective.
Easily adjustable – A lever on the cap provides
for quick vertical movement of the metal plunger rod, allowing for adjustment to any bed height.
No dead volume – Eliminating dead volume
improves gradient accuracy because there is no space for solvent mixing.
Secure lock – The sleeve’s internal locking
mechanism holds the plunger and cartridge securely in place, without the use of clumsy external springs and locking mechanisms.
Low maintenance – The plunger uses a virgin
PTFE seal, eliminating costly, leak-prone O-rings.
No leaks – Internal stops within the cap keep the
cartridge from rotating. The sleeve tightly holds the cartridge and prevents leaks, even under high operating pressures.
High operating pressure – The cap, when used
according to these instructions, is rated to 200 psi.
Preparing an Empty Cartridge
1. Adsorb the sample onto loose silica or other suitable media.
2. Pour the sample and media mixture into the car­tridge (Figure 1).
3. Place a frit into the cartridge.
4. Push the frit down on top of the sample and media mixture.
5. If you did not remove the solvent before loading the mixture into the cartridge, continue with Dry- ing the Cartridge.
Preparing Cartridges
Teledyne Isco’s Rf adjustable cartridge cap can be used with any type of sample, and are effective with low solubility samples. The adjustable caps can be used with a variety of RediSep Rf cartridges. To meet your specific needs, RediSep Rf disposable cartridges are available empty, prepacked with silica gel, or prepacked with C18 reversed phase media. After selecting a car­tridge, prepare the solid sample.
Preparing a Prepacked Cartridge
1. Dissolve the sample and introduce it onto the prepacked cartridge (Figure 1).
2. Remove the solvent by drying the cartridge.
Figure 1: Preparing Cartridges
prepacked cartridges (left) or pour adsorbed sample into empty cartridges (right).
Pipette
sample onto
Drying the Cartridge
Solvent should be removed from the prepared car­tridge. You can do so using a vacuum dryer or the air pump in the CombiFlash system. Instructions to use the system’s air pump are included in the next sections, Loading Cartridges.
Chromatography Technical Note #11 69-5233-423
Loading 5 and 25 gram Cartridges
To load the 5 or 25 gram cartridge on the system:
1. The solid load cartridge cap must match the size of your selected cartridge. If it does not match, install the correct cap now.
2. Press the lever on the side of the cap and fully extend the plunger (Figure 2).
3. Slide the plunger into the cartridge until it reaches the top frit (Figure 3).
4. Press the lever and push the cartridge into the cap. Align the cartridge so that it fits fully into the recess in the cap (Figure 4).
2
1
CAUTION
Operating the cartridges at pressures greater than 50 psi may cause leaks.
1
Figure 2: Press the lever and extend the plunger
2
Figure 4: Align and push cartridge into the cap
Figure 3: Insert the plunger to the bed