Teledyne Isco Syringe Pump Technical Bulletin

D-Series and HL-Series Syringe Pump Controller Versions
How to Distinguish Between Versions
Overview
The D-Series and HL/HLf-Series Syringe Pump Controller designs have undergone several updates. While old or new controllers work with ISCO pumps, some operating and maintenance procedures for the current controller are different from those for the legacy controller. To ensure that you are following the correct procedures, you must know which controller you are using.
Where procedural differences o
Pump Technical Bulletins available on our web site
distinguish the two types as the “legacy” and “current” controller versions.
Legacy Pump Controller
ccur, the Syringe
Syringe Pump Technical Bulletin
Sept 2020, TB28
Pump systems using the legacy controller require a different user manual than pump systems using the current controller.
This bulletin explains how to tell the two controllers apart, and should always be referenced before servicing or operation if the controller version is in question.
Differences in Appearance
Systems using the updated pump controller are iden­tified by the lighted display and the blue Dispense (DISP) button in the fourth row on the control panel.
D-Series Controllers
Lighted display
DISP button
Close-Up view of legacy display
Current Pump Controller
Close-Up view of current display
Syringe Pump Technical Bulletin Sept 2020, TB28
Lighted display
Legacy Pump HL Controller
HL-Series Controllers
Close-Up view of legacy display
Current Pump HL Controller
DISP button
Date of Sale
Be sure to use the above indicators to identify
your controller.
Any D-Series controller that was purchased
before November 2011 is a legacy controller.
Any HL-Series controller that was purchased
before October 2013 is a legacy controller.
Latest Firmware Features
For systems using the current controller, firmware versions 1.09 (D-Series) or 2.00.0003 (HL-Series) and later give your pump syste additional capabilities.
Quad Pump Systems
With this firmware and later, the current controller is capable of operating four connected pumps independently or in pairs of two, connected with air valve packages.
Refer to Technical Bulletin TB26 D-Series Quad
Dual Pump Systems for Continuous Flow and
m
Close-Up view of current display
Independent Operation for complete information
about setup and operation.
Modbus RTU Communication
With this firmware and later, the current controller is capable of Modbus communication using RTU protocol over an RS-485 connection.
Refer to Technical Bulletin TB27 Modbus RTU
with Teledyne ISCO Syringe Pumps for complete
information about Modbus RTU communication.
Finding Your Software Version
To check your controller’s software version, press
,
MORE ( ), and STATUS ( ). The
software version will appear in the upper right corner of the run screen.
Contact the Tel
edyne ISCO service department at 800-775-2965 for assistance in obtaining the correct firmware for your pump application.
Teledyne Isco
P.O. Box 82531, Lincoln, Nebraska, 68501 USA
Toll-free: (800) 775-2965 • Phone: (402) 464-0231 • Fax: (402) 465-3001
E-mail: IscoService@teledyne.com
Teledyne Isco is continually improving its products and reserves the right to change product specifications, replacement parts, schematics, and instructions without notice.