Teledyne QPHY-PCIe3-Tx-Rx User Manual

QPHY-PCIE3-TxRx Operator’s Manual
Revision B – March, 2014
Relating to the following release versions: Software Version Rev. 7.3.x.x Script
PCI_Express_Test_Spec_Electrical_Layer_3_0_rev_0_9
700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 teledynelecroy.com
© 2014 Teledyne LeCroy, Inc. All rights reserved. Teledyne LeCroy and other product or brand names are trademarks or requested trademarks of their
respective holders. Information in this publication supersedes all earlier versions. Specifications are subject to change without notice.
923662 Rev B March, 2014
QPHY-PCIE3-TxRx Operator’s Manual
Table of Contents
Introduction ............................................................................................................................... 1
About QualiPHY ............................................................................................................................................ 1
About QPHY-PCIE3-TxRx ............................................................................................................................. 1
Required Equipment ..................................................................................................................................... 1
Required Host Computer System ................................................................................................................. 2
Installation and Setup ............................................................................................................... 3
Install Base Application ................................................................................................................................. 3
Install SigTest.exe ......................................................................................................................................... 3
Activate Components .................................................................................................................................... 3
Set Up Dual Monitor Display ......................................................................................................................... 4
Set Up Remote Control ................................................................................................................................. 4
Using QualiPHY ......................................................................................................................... 5
Accessing the Software ................................................................................................................................. 5
General Setup ............................................................................................................................................... 6
Connection tab ....................................................................................................................................... 6
Session Info tab ..................................................................................................................................... 6
Report tab .............................................................................................................................................. 6
Advanced tab ......................................................................................................................................... 6
About tab ................................................................................................................................................ 6
QualiPHY Test Process ................................................................................................................................. 7
Set Up Test Session .............................................................................................................................. 7
Run Tests ............................................................................................................................................... 7
Run Reports ........................................................................................................................................... 8
Customizing QualiPHY .................................................................................................................................. 9
Copy Configuration ................................................................................................................................ 9
Select Tests ........................................................................................................................................... 9
Edit Variables ....................................................................................................................................... 10
Edit Test Limits .................................................................................................................................... 11
X-Replay Mode ........................................................................................................................................... 12
QPHY-PCIE3-TxRx Testing ..................................................................................................... 13
Test Preparation .......................................................................................................................................... 13
Required Test Modes .......................................................................................................................... 13
Physical Setup ..................................................................................................................................... 13
QPHY-PCIE3-TxRx Test Configurations ..................................................................................................... 13
Demo Add-in Card Tx .......................................................................................................................... 13
Add-in Card Rx .................................................................................................................................... 14
Add-in Card Tx ..................................................................................................................................... 16
System Rx ............................................................................................................................................ 18
System Tx ............................................................................................................................................ 19
QPHY-PCIE3-TxRx Variables ..................................................................................................................... 20
Specific to “TX CEM test” Group ......................................................................................................... 21
Specific to “Rx Calibration and test” Group ......................................................................................... 21
Specific to “Rx CEM Test” Group ........................................................................................................ 21
QPHY-PCIE3-TxRx Limit Sets .................................................................................................................... 22
Manual Deskewing Procedures.............................................................................................. 23
Cable Deskewing Using the Fast Edge Output........................................................................................... 23
Cable Deskewing Without Using the Fast Edge Output ............................................................................. 26
Error Messages ....................................................................................................................... 27
923662 Rev B i
Table of Figures
Figure 1 - QualiPHY framework dialog and Standard selection menu. ................................................. 5
Figure 2 - The Test Report Summary and Details pages. ........................................................................ 8
Figure 3 - Variable Setup tab. ................................................................................................................... 10
Figure 4 - Limits Manager dialog. ............................................................................................................ 11
Figure 5 – X-Replay Mode window. ......................................................................................................... 12
Figure 6 – Step 1 setup for Preset, Amplitude, RJ and SJ Calibration. ............................................... 14
Figure 7 – Step 1 setup for Differential Mode Noise, Eye height and Eye Width Calibration. ........... 15
Figure 8 – Step 2 setup for actual receiver tolerance test. ................................................................... 15
Figure 9 – Add-In Card Tx connection diagram showing oscilloscope (green) and Rx Ln0 (blue)
physical connections. ....................................................................................................................... 16
Figure 10 – Add-In Card Tx connection diagram. .................................................................................. 17
Figure 11 – Example of Signal Test dialog and Signal Test results. .................................................... 18
Figure 12 - The Skew parameter right side dialog, Skew clock 2 tab, showing default setup. ......... 27
About This Manual
This manual assumes that you are familiar with using an oscilloscopein particular the Teledyne LeCroy oscilloscope that will be used with QualiPHYand that you have purchased the QPHY­PCIE3-TxRx software option. Some of the images in this manual may show QualiPHY products other than QPHY-PCIE3-TxRx, or were captured using different model oscilloscopes, as they are meant to illustrate general concepts only. Rest assured that while the user interface may look different from yours, the functionality is identical.
ii 923662 Rev B
QPHY-PCIE3-TxRx Operator’s Manual
Introduction
About QualiPHY
QualiPHY is highly automated compliance test software meant to help you develop and validate the PHY (physical-electrical) layer of a device, in accordance with the official documents published by the applicable standards organizations and special interest groups (SIGs). You can additionally set custom variables and limits to test compliance to internal standards.
QualiPHY is composed of a “framework” application that enables the configuration and control of separate tests for each standard through a common user interface. Features include:
Multiple Data Source Capability: Connect to your X-Stream oscilloscope via LAN or
other interfaces.
User-Defined Test Limits: Tighten limits to ensure devices are well within the passing
region, even if subsequently measured with different equipment.
Flexible Test Results Reporting that includes XML Test Record Generation.
Understand a device performance distribution, or obtain process related information from the devices under test.
About QPHY-PCIE3-TxRx
QPHY-PCIE3-TxRx is an automated test package performing all the Tx and Rx receiver tests in accordance with PCI Express Card Electromechanical Specification, Rev. 3.0.
The software can be run on any Teledyne LeCroy real-time oscilloscope with at least 13 GHz bandwidth and a sample rate of at least 40 GS/s.
Required Equipment
Real-time Teledyne LeCroy oscilloscope with the QualiPHY software, an activated
QPHY-PCIE3-TxRx component, and SigTest.exe installed
PeRT3 Phoenix (for Rx tests) DC blocks 1 pair standard SMA-SMP cable and 3 pair short SMA-SMP cables 1 pair standard SMA-SMA cable and 2 pair high-quality SMA-SMA cables PCIE 3.0 test fixtures , CBB3 and CLB3 SMP extractor tool Torque wrench Power supply
For the most complete and up-to-date list, see the PCI SIG website: www.pcisig.com/specifications/pciexpress/compliance
923662 Rev B 1
Required Host Computer System
Usually, the oscilloscope is the host computer for the QualiPHY software, and all models that meet the acquisition requirements will also meet the host system requirements. However, if you wish to run the QualiPHY software from a remote computer, these minimum requirements apply:
Operating System:
o Windows 2000 Professional with service pack 4 o Windows XP Professional with service pack 2 o Windows VISTA with service pack 1 o Windows 7 Professional
1 GHz or faster processor 1 GB (32-bit) or 2 GB (64-bit) of RAM Ethernet (LAN)network capability Hard Drive:
o At least 100 MB free to install the framework application o Up to 1GB per standard installed to store the log database (each database grows
from a few MB to a maximum of 1 GB)
See Set Up Remote Control for configuration instructions.
2 923662 Rev B
QPHY-PCIE3-TxRx Operator’s Manual
Installation and Setup
QualiPHY is a Windows-based application that can be configured with one or more serial data compliance components. Each compliance component is purchased as a software option.
Install Base Application
Download the latest version of the QualiPHY software from: teledynelecroy.com/support/softwaredownload under Oscilloscope Downloads > Software Utilities If the oscilloscope is not connected to the Internet, copy the installer onto a USB memory stick
then transfer it to the oscilloscope desktop or a folder on a D:\ drive to execute it. Run QualiPHYInstaller.exe and follow the installer prompts. Choose all the components you
plan to activate. If you omit any components now, you will need to update the installation to activate them later.
By default, the oscilloscope appears as local host when QualiPHY is executed on the oscilloscope. Follow the steps under Add Connection to QualiPHY to check that the IP address is 127.0.0.1.
Install SigTest.exe
SigTest.exe ver. 3.2.0, which is the current version required by the PCI SIG, must be installed on the oscilloscope for the PCIE3-TxRx script to execute properly. It is available for download from the PCI SIG website.
Activate Components
The serial data compliance components are factory installed as part of the main application in your oscilloscope and are individually activated through the use of an alphanumeric code uniquely matched to the oscilloscope’s serial number. This option key code is what is delivered when purchasing a software option.
To activate a component on the oscilloscope:
1. From the menu bar, choose Utilities > Utilities Setup.
2. On the Options tab, click Add Key.
3. Use the Virtual Keyboard to Enter Option Key, then click OK.
If activation is successful, the key code now appears in the list of Installed Option Keys.
4. Restart the oscilloscope application by choosing File > Exit, then double-clicking the
Start DSO icon on the desktop.
923662 Rev B 3
Set Up Dual Monitor Display
Teledyne LeCroy recommends running QualiPHY on an oscilloscope equipped with Dual Monitor Display capability. This allows the waveform and measurements to be shown on the oscilloscope LCD display while the QualiPHY application and test results are displayed on a second monitor.
See the oscilloscope Operator’s Manual or Getting Started Manual for instructions on setting up
dual monitor display.
Set Up Remote Control
QualiPHY software can be executed from a remote host computer, controlling the oscilloscope through a LAN Connection. To set up remote control:
The oscilloscope must be connected to a LAN and assigned an IP address (fixed or
dynamic).
The host computer must be on the same LAN as the oscilloscope.
Note: LXI, GPIB, LSIB, and USBTMC remote control is available for some model instruments. Follow the same procedure, selecting the appropriate connection type on the oscilloscope and in QualiPHY.
Configure Oscilloscope for Remote Control
1. From the menu bar, choose Utilities Utilities Setup...
2. Open the Remote tab and set Remote Control to TCP/IP.
3. Verify that the oscilloscope shows an IP address.
Add Connection to QualiPHY
1. On the host PC, download and run QualiPHYInstaller.exe.
2. Start QualiPHY and click the General Setup button.
3. On the Connection tab, click Scope Selector.
4. Click Add and choose the connection type. Enter the oscilloscope IP address from Step 3 above. Click OK.
5. When the oscilloscope is properly detected, it appears on the Scope Selector dialog. Select the connection, and click OK.
QualiPHY is now ready to control the oscilloscope.
Select Connection
Multiple oscilloscopes may be accessible to a single remote host. In that case, go to General Setup and use the Scope Selector at the start of the session to choose the correct connection.
QualiPHY tests the oscilloscope connection when starting a test. The system warns you if there is a connection problem.
4 923662 Rev B
QPHY-PCIE3-TxRx Operator’s Manual
Using QualiPHY
This section provides an overview of the QualiPHY user interface and general procedures, using examples from the QPHY-BroadR-Reach option. For detailed information about the QPHY-PCIE3-TxRx software option, see QPHY-PCIE3-TxRx Testing.
Accessing the Software
Once QualiPHY is installed and activated, it can be accessed from the oscilloscope menu bar by choosing Analysis > QualiPHY, or by double-clicking the QualiPHY desktop icon on a remote computer.
The QualiPHY framework dialog illustrates the overall software flow, from general set up through running individual compliance tests. Work from left to right, making all desired settings on each sub-dialog.
Figure 1 - QualiPHY framework dialog and Standard selection menu.
The sub-dialogs are organized into tabs each containing configuration controls related to that part of the process. These are described in more detail in the following sections.
If Pause on Failure is checked, QauliPHY prompts to retry a measure whenever a test fails. Report Generator launches the manual report generator dialog. The Exit button at the bottom of the framework dialog closes the QualiPHY application.
923662 Rev B 5
General Setup
The first sub-dialog contains general system settings. These remain in effect for each session, regardless of Standard, until changed.
Connection tab
Shows IP Address of the oscilloscope (local host 127.0.0.1 if QualiPHY is run from the oscilloscope). The Scope Selector allows you to choose the oscilloscope used for testing when several are connected to the QualiPHY installation. See Set Up Remote Control for details.
Session Info tab
Optional information about the test session that may be added to reports, such as: Operator Name, Device Under Test (DUT) name, Temperature (in C) of the test location, and any
additional Comments. There is also an option to Append Results or Replace Results when continuing a previous session.
To optimize report generation, enter at least a DUT name at the beginning of each session.
Report tab
Settings related to report generation. Choose:
Reporting behavior of:
o “Ask to generate a report after tests,” where you’ll be prompted to create a new
file for each set of test results.
o “Never generate a report after tests,” where you’ll need to manually execute the
Report Generator to create a report.
o Always generate a report after tests,” to autogenerate a report of the latest test
results.
Default report output format of XML, HTML, or PDF.
Note: Unlike some QualiPHY components, PCIE-TxRx does not overwrite a generic file
when autogenerating reports, but creates a new report file each session named “LeCroyReport_<DUT name>” (or if DUT is empty, “LeCroyReport_DUT”). You can omit entering an Output file name.
Optionally, check Allow style sheet selection in Report Generator to enable the use of a custom .xslt when generating reports (XML and HTML output only). The path to the .xslt is entered on the Report Generator dialog.
Report Generator launches the Report Generator dialog, which contains the same settings as the Report tab, only applied to individual reports.
Advanced tab
This tab launches the X-Replay Mode dialog. See X-Replay Mode.
About tab
Information about your QualiPHY installation.