Teledyne QPHY-HDMI User Manual

QPHY-HDMI
Operator’s Manual
Revision A – December, 2013
Relating to the following release versions:
Software Version Rev. 7.3.x.x Script HDMI.IRT Style Sheet Rev. 1.2
700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 teledynelecroy.com
© 2013 Teledyne LeCroy, Inc. All rights reserved. Teledyne LeCroy and other product or brand names are trademarks or requested trademarks of their
respective holders. Information in this publication supersedes all earlier versions. Specifications are subject to change without notice.
922540 Rev A December, 2013
QPHY-HDMI Operator’s Manual
Table of Contents
Introduction ............................................................................................................................... 1
About QualiPHY ............................................................................................................................................ 1
About QPHY-HDMI ....................................................................................................................................... 1
Required Hardware ....................................................................................................................................... 1
Test Fixtures .................................................................................................................................................. 2
Required Host Com puter S yst em ................................................................................................................. 2
Installation and Setup ............................................................................................................... 3
Install Base Application ................................................................................................................................. 3
Activate Components .................................................................................................................................... 3
Set Up Dual Monitor Display ......................................................................................................................... 3
Set Up Remote Control ................................................................................................................................. 4
Configure Oscilloscope for Remote Control .......................................................................................... 4
Add Connection to QualiPHY ................................................................................................................ 4
Select Connection .................................................................................................................................. 4
Using QualiPHY ......................................................................................................................... 5
Accessing the Software ................................................................................................................................. 5
General Setup ............................................................................................................................................... 6
Connection tab ....................................................................................................................................... 6
Session Info tab ..................................................................................................................................... 6
Report tab .............................................................................................................................................. 6
Advanced tab ......................................................................................................................................... 6
About tab ................................................................................................................................................ 6
QualiPHY Test Process ................................................................................................................................. 7
Set Up Test Session .............................................................................................................................. 7
Run Tests ............................................................................................................................................... 8
Run Reports ........................................................................................................................................... 9
Customizing QualiPHY ................................................................................................................................ 10
Copy Configuration .............................................................................................................................. 10
Select Tests ......................................................................................................................................... 10
Edit Variables ....................................................................................................................................... 11
Edit Test Limits .................................................................................................................................... 12
X-Replay Mode ........................................................................................................................................... 13
QPHY-HDMI Testing ................................................................................................................ 14
Test Preparation .......................................................................................................................................... 14
Required Test Modes .......................................................................................................................... 14
Physical Setup ..................................................................................................................................... 14
Deskew ................................................................................................................................................ 14
QPHY-HDMI Test Configurations ................................................................................................................ 15
Clock and Data0 Basic Tests ............................................................................................................... 15
Clock and Data0 Tests at All Standard Frequencies ........................................................................... 15
Data0 & Data1 & Data2 Inter-Pair Skew (7-6) ..................................................................................... 15
Clock and Data0 Tests at Custom Frequency (340 MHz) ................................................................... 15
Demo of All Tests................................................................................................................................. 15
QPHY-HDMI Test Descriptions ................................................................................................................... 16
Test7-2: TMDS-VL ............................................................................................................................... 16
Test7-4: TMDS TRise, TFall ................................................................................................................ 17
Test7-5: TMDS Over/Undershoot ........................................................................................................ 17
Test7-6: TMDS Inter-P air S kew ........................................................................................................... 18
Test7-7: TMDS Intra-P air S kew ........................................................................................................... 19
Test7-8: TMDS Clock Duty Cycle ........................................................................................................ 20
922540 Rev A i
Test7-9: TMDS Clock Jitter .................................................................................................................. 21
Test7-10: TMDS Data Eye Diagram .................................................................................................... 22
QPHY-HDMI Variables ................................................................................................................................ 23
Specific to “Test7-2: TMDS-VL” Group ................................................................................................ 24
Specific to “Test7-4: TMDS TRise, TFall” Group ................................................................................. 24
Specific to “Test7-6: TMDS Inter-Pair Skew” Group ............................................................................ 24
Specific to “Test7-7: TMDS Intra-Pair Skew” Group ............................................................................ 25
Specific to “Test7-8: TMDS Clock Duty Cycle” Group ......................................................................... 25
Specific to “Test7-9: TMDS-Clock Jitter” Group .................................................................................. 25
Specific to “Test7-10: TMDS-Data Eye Diagram” Group ..................................................................... 26
QPHY-HDMI Limit Sets ............................................................................................................................... 26
Manual Deskewing Procedures.............................................................................................. 27
Cable Deskewing Using the Fast Edge Output........................................................................................... 27
Cable Deskewing Using the DUT Signal .................................................................................................... 30
Error Messages ....................................................................................................................... 32
Table of Figures
Figure 1 - QualiPHY framework dialog and Standard selection menu. ................................................. 5
Figure 2 - The Test Report Summary Table and Detail page. ................................................................. 9
Figure 3 – X-Replay Mode window. ......................................................................................................... 13
Figure 4 - The Skew parameter right side dialog, Skew clock 2 tab, showing default setup. ........... 31
About This Manual
This manual assumes that you are familiar with using an oscilloscopein particular the Teledyne LeCroy oscilloscope that will be used with QualiPHYand that you have purchased the QPHY­HDMI software option. Some of the images in this manual may show QualiPHY products other than QPHY-HDMI, or were captured using different model oscilloscopes, as they are meant to illustrate general concepts only. Rest assured that while the user interface may look different from yours, the functionality is identical.
ii 922540 Rev A
QPHY-HDMI Operator’s Manual

Introduction

About QualiPHY

QualiPHY is highly automated compliance test software meant to help you develop and validate the PHY (physical-electrical) layer of a device, in accordance with the official documents published by the applicable standards organizations and special interest groups (SIGs). You can additionally set custom variables and limits to test compliance to internal standards.
QualiPHY is composed of a “framework” application that enables the configuration and control of separate tests for each standard through a common user interface. Features include:
Multiple Data Source Capability: Connect to your X-Stream oscilloscope via LAN or other interfaces.
User-Defined Test Limits: Tighten limits to ensure devices are well within the passing region, even if subsequently measured with different equipment.
Flexible Test Results Reporting that includes XML Test Record Generation. Understand a device performance distribution, or obtain process related information from the devices under test.

About QPHY-HDMI

QPHY-HDMI is an automated test package performing all the normative, real-time oscilloscope tests for sources in accordance with version 1.4 of the High Definition Multimedia Interface Compliance Test Specification (HDMI).
The software can be run on any Teledyne LeCroy oscilloscope with at least 8 GHz bandwidth (for testing clock rates above 148.5 MHz). The oscilloscope must also have the QualiPHY software installed and the QPHY-HDMI software option activated.

Required Hardware

Real-time Teledyne LeCroy oscilloscope with all required software installed.
Note: QPHY-HDMI tests are designed to run on WavePro/SDA/DDA 7Zi/ZI-A, WaveMaster/SDA/DDA Zi/Zi-A, LabMaster 9Zi/Zi-A Series oscilloscopes with minimum 8GHz bandwidth. The HDMI test specification requires 8 GHz bandwidth for testing clock rates above 148.5 MHz.
SDAII or SDAIII oscilloscope software option. (SDAII / SDIII is standard on SDA model oscilloscopes)
8 SMA Male to SMA Male cables
4 TF-HDMI-3.3V test adapters
HDMI test fixtur e
For the most complete and up-to-date list, see the HDMI SIG website: http://www.hdmi.org
922540 Rev A 1

Test Fixtures

Teledyne LeCroy recommends using the Wilder Technologies HDMI-TPA-P (plug test adapter) used in conjunction with HDMI-TPA-E (EDI D board with EEPROM). This equipment is available from Wilder Technologies individually or as part of the HDMI-TPA-PRRCE Test Adapter Kit. The HDMI-TPA-PRRCE test adapter kit provides a very high bandwidth, low noise evaluation vehicle that enables high-performance characterization of both plug and receptacle HDMI™ devices with bandwidth greatly exceeding the 340 MHz (10.2 Gbps) of the latest v1.4b HDMI specification.

Required Host Computer System

Usually, the oscilloscope is the host computer for the QualiPHY software, and all models that meet the acquisition requirements will also meet the host system requirements. However, if you wish to run the QualiPHY software from a remote computer, these minimum requirements apply:
Operating System:
o Windows 2000 Professional with service pack 4 o Windows XP Professional with service pack 2 o Windows VISTA with service pack 1 o Windows 7 Professional
1 GHz or faster processor
1 GB (32-bit) or 2 GB (64-bit) of RAM
Ethernet (LAN) network capability
Hard Drive:
o At least 100 MB free to install the framework application o Up to 1GB per standard installed to store the log database (each database grows
from a few MB to a maximum of 1 GB)
See Set Up Remote Control
for configuration instructions.
2 922540 Rev A
QPHY-HDMI Operator’s Manual
Installation and Setup
QualiPHY is a Windows-based application that can be configured with one or more serial data compliance components. Each compliance component is purchased as a software option.

Install Base Application

Download the latest version of the QualiPHY software from: teledynelecroy.com/support/softwaredownload under Oscilloscope Downloads > Software Utilities. If the oscilloscope is not connected to the Internet, copy the installer onto a USB memory stick,
then transfer it to the oscilloscope desktop or a folder on a D:\ drive to execute it. Run QualiPHYInstaller.exe and follow the installer prompts. Choose all the components you
plan to activate. If you omit any components now, you will need to update the installation to activate them later.
By default, the oscilloscope appears as local host when QualiPHY is executed on the oscilloscope. Follow the steps under Add Connection to QualiPHY is 127.0.0.1.
to check that the IP address

Activate Components

The serial data compliance components are factory installed as part of the main application in your oscilloscope and are individually activated through the use of an alphanumeric code uniquely matched to the oscilloscope’s serial number. This option key code is what is delivered when purchasing a software option.
To activate an option on the oscilloscope:
1. From the menu bar, choose Utilities > Utilities Setup.
2. On the Options tab, click Add Key.
3. Use the Virtual Keyboard to Enter Option Key, then click OK.
If activation is successful, the key code now appears in the list of Installed Option Keys.
4. Restart the oscilloscope application by choosing File > Exit, then double-clicking the
Start DSO icon on the desktop.

Set Up Dual Monitor Display

Teledyne LeCroy recommends running QualiPHY on an oscilloscope equipped with Dual Monitor Display capability. This allows the waveform and measurements to be shown on the oscilloscope LCD display, while the QualiPHY application and test results are displayed on a second monitor.
See the oscilloscope Operator’s Manual or Getting Started Manual for instructions on setting up dual monitor display.
922540 Rev A 3

Set Up Remote Contr ol

QualiPHY software can be executed from a remote host computer, controlling the oscilloscope through a LAN Connection. To set up remote c ontrol:
The oscilloscope must be connected to a LAN and assigned an IP address (fixed or dynamic).
The host computer must be on the same LAN as the oscilloscope.
Note: LXI, GPIB, LSIB, and USBTMC remote control is available for some model instruments. Follow the same procedure, selecting the appropriate connection type on the oscilloscope and in QualiPHY.

Configure Oscilloscope for Remot e Cont rol

1. From the menu bar, choose Utilities  Utilities Setup...
2. Open the Remote tab and set Remote Control to TCP/IP.
3. Verify that the oscilloscope shows an IP address.

Add Connection to QualiPHY

1. On the host PC, download and run QualiPHYInstaller.exe.
2. Start QualiPHY and click the General Setup button.
3. On the Connection tab, click Scope Selector.
4. Click Add and choose the connection type. Enter the oscilloscope IP address from Step 3 above. Click OK.
5. When the oscilloscope is properly detected, it appears on the Scope Selector dialog. Select the connection, and click OK.
QualiPHY is now ready to control the oscilloscope.

Select Connection

Multiple oscilloscopes may be accessible to a single remote host. In that case, go to General Setup and use the Scope Selector at the start of the QPHY session to choose the correct connection.
QualiPHY tests the oscilloscope connection when starting a test. The system warns you if there is a connection problem.
4 922540 Rev A
QPHY-HDMI Operator’s Manual
Using QualiPHY
This section provides an overview of the QualiPHY user interface and general procedures. For detailed information about the QPHY-HDMI software option, see QPHY-HDMI Testing

Accessing the Softw ar e

Once QualiPHY is installed and activated, it can be accessed from the oscilloscope menu bar by choosing Analysis > QualiPHY, or by double-clicking the QualiPHY desktop icon on a remote computer.
The QualiPHY framework dialog illustrates the overall software flow, from general set up through running individual compliance tests. Work from left to right, making all desired settings on each sub-dialog.
.
Figure 1 - QualiPHY framework dialog and Standard selection me nu.
The sub-dialogs are organized into tabs each containing configuration controls related to that part of the process. These are described in more detail in the following sections.
If Pause on Failure is checked, QualiPHY prompts to retry a measure whenever a test fails. Report Generator launches the manual report generator dialog. The Exit button at the bottom of the framework dialog closes the QualiPHY application.
922540 Rev A 5

General Setup

The first sub-dialog contains general system settings. These remain in effect for each session, regardless of Standard, until changed.

Connection ta b

Shows IP Address of the test oscilloscope (local host 127.0.0.1 if QualiPHY is run from the oscilloscope). The Scope Selector allows you to choose the oscilloscope used for testing when several are connected to the QualiPHY installation. See Set Up Remote Control

Session Info tab

Optional information about the test session that may be added to reports, such as: Operator Name, Device Under Test (DUT), Temperature (in °C) of the test location, and any additional Comments. There is also an option to Append Results or Replace Results when continuing a
previous session. To optimize report generation, enter at least a DUT name at the beginning of each session.

Report tab

Settings related to automatic report generation. Choose:
Reporting behavior of:
for details.
o “Ask to generate a report after tests,” where you’ll be prompted to create a new
file for each set of test results.
o “Never generate a report after tests,” where you’ll need to manually execute the
Report Generator to create a report.
o “Always generate a report after tests,” to autogenerate a report of the latest test
results.
Default report output type of XML, HTML, or PDF.
A generic Output file name, including the full path to the report output folder.
Optionally, check Allow style sheet selection in Report Generator to enable the use of a custom .css when generating reports. The path to the .css is entered on the Report Generator dialog.
Report Generator launches the Report Generator dialog, which contains the same settings as the Report tab, only applied to individual reports.

Advanced tab

This tab launches the X-Replay Mode dialog. See X-Replay Mode.

About tab

Information about your QualiPHY installation.
6 922540 Rev A
QPHY-HDMI Operator’s Manual

QualiPHY Test Process

Once general system settings are in place, these are the steps for running test sessions.

Set Up Test Session

1. Connect the oscilloscope to the DUT. See QPHY-HDMI Testing Physical Setup.
2. Access the QualiPHY software to display the framework dialog.
3. If running QualiPHY remotely, click General Setup and open the Scope Selector to select the correct oscilloscope connection.
4. If you have more than one component activated, click Standard and select the desired standard to test against. Otherwise, your one activated component will appear as the default selection.
Note: Although all the QualiPHY components appear on this dialog, only those selected when installing QualiPHY are enabled for selection.
5. Click the Configuration button and select the test configuration to run. These pre­loaded configurations are set up to run all the tests required for compliance and provide a quick, easy way to begin compliance testing. See QPHY-HDMI Test Configurations a description of your configurations.
You can also create custom configurations for internal compliance tests by copying and modifying the pre-loaded configurations. See Customizing QualiPHY
6. Close the Edit/View Configuration dialog to return to the framework dialog.
for de tails.
for
922540 Rev A 7

Run Tests

1. On the framework dialog, click Start to begin testing. When tests are in progress, this button changes to Stop. Click it at any time to stop the
test in process. You’ll be able to resume from the point of termination or from the beginning of the test.
2. Follow the pop-up window prompts. QualiPHY guides you step-by-step through each of the tests described in the standard specification, including diagrams of the connection to the DUT for each required test mode.
3. When all tests are successfully completed, both progress bars on the framework dialog are completely green and the message “All tests completed successfully” appears. If problems are encountered, you’ll be offered options to:
Retry the test from the latest established point defined in the script
Ignore and Continue with the next test
Abort Session
8 922540 Rev A
Loading...
+ 25 hidden pages