Teledyne PPE 6 kV User Manual

HIGH VOLTAGE
PPE 6 kV
GENERAL CHARACTERISTICS
Operating temperature: 0 °C to 50 °C Cable length: 2 m Weight of probe + accessories: 70 gr Coupling: 1 MDC/AC Warranty: One year Overvoltage category: CAT I Pollution degree: 1
CLEANING
The exterior of the probe only should be cleaned with a soft cloth moistened with either water or isopropyl alcohol.
USING AND MAINTENANCE
The probe user should note that the protection provided by the equipment may be impaired, if the equipment is used in a manner not specified by the manufacturer. To guarantee accurate performance characteristics, machanical shocks should be avoided, as well as damage to the cable through excessive bending. Handle with care, especially when the extra thin spring-tip is fitted. All maintenance should be referred to qualified personnel.
SALES & SERVICE LECROY
Corporate Headquarters
U.S.A.: Chestnut Ridge Phone (1) 914 578 6020 Fax (1) 914 578 5985
European Headquarters
Germany: Heidelberg Phone (49) 6221 827 00 Fax (49) 6221 834 655
S
AFETY INFORMATION
The probe is designed to be safe under the following conditions:
Indoor use Altitude to 2000 m Temperature range from 0 °C to 50 °C Maximum relative humidity 80 % for tem­peratures up to 31 °C, decreasing linearly to 40 % relative humidity at 50 °C
WARNING
Do not disassemble or disconnect probes while they are connected to a voltage source. Use only grounded oscilloscopes.
Attention: Refer to this manual before using the probe.
Danger. High Voltage.
10
100
1000
10000
0,1 1 10 100 1000
Frequency [MHz]
Voltage [V]
ELECTRICAL CHARACTERISTICS
System attenuation: 1000:1 System input resistance: 50 M Input capacitance: < 6 pF Compensation range: 10 pF to 50 pF System BW (-3 dB): 400 MHz typical Max. non-destructive input voltage: 6 kV
See voltage derating curve
Impulse
DC + peak AC
LF PROBE ADJUSTMENT (1000:1)
Connect the probe to a 1 kHz square wave signal. Adjust compensation trimmer in the BNC-box for optimum square wave response.
Undershoot
Correct
Overshoot
HF PROBE ADJUSTMENT
The probe can be adjusted for high frequency if necessary. Remove boot and isolation from BNC-box. Connect the probe to 1 MHz square wave signal. Turn the PCB in the BNC-box by 180 ° and adjust this trimmer for optimum square wave response. Then replace all parts again in their previous position.
Undershoot
Correct
Overshoot
STANDARD ACCESSORIES SUPPLIED
WITH THE PROBE
Accessories Quantities
1. Trimming tool 1
2. Sprung hook 1
3. IC Insulating tip (red) 1
4. Insulating tip (red) 1
5. Ground lead (22 cm) 1
6. BNC Adapter 1
7. Straight tip (0.8 mm) 1
8. Spring contact tip (0.8 mm) 1
9. Crocodile clip 1
10. Flexible 4 mm security adapter 1
11. Security ground lead 1
The following accessories are rated up to:
Sprung hook 1000 V BNC Adapter 100 V Crocodile clip 1000 V Flexible 4 mm security adapter 1000 V
OPTIONAL PROBE KIT PK103
Same as standard accessories supplied with probe, except for: Straight tip (5 pieces) Spring contact tip (5 pieces)
11
2
1
5
3
4
7
6
8
9
10