Teledyne PPE 20 kV User Manual

HIGH VOLTAGE
PPE 20 kV
GENERAL CHARACTERISTICS
Operating temperature: 0 °C to 50 °C Cable length: 3 m Weight of probe + accessories: 495 gr Coupling: 1 MDC/AC Warranty: One year Overvoltage category: CAT I Pollution degree: 1
CLEANING
The exterior of the probe only should be cleaned with a soft cloth moistened with either water or isopropyl alcohol.
USING AND MAINTENANCE
The probe user should note that the protection provided by the equipment may be impaired, if the equipment is used in a manner not specified by the manufacturer. To guarantee accurate performance characteristics, machanical shocks should be avoided, as well as damage to the cable through excessive bending. Handle with care, especially when the extra thin spring-tip is fitted. All maintenance should be referred to qualified personnel.
SALES & SERVICE LECROY
Corporate Headquarters
U.S.A.: Chestnut Ridge Phone (1) 914 578 6020 Fax (1) 914 578 5985
European Headquarters
Germany: Heidelberg Phone (49) 6221 827 00 Fax (49) 6221 834 655
SAFETY
INFORMATION
The probe is designed to be safe under the following conditions:
Indoor use Altitude to 2000 m Temperature range from 0 °C to 50 °C Maximum relative humidity 80 % for tem­peratures up to 31 °C, decreasing linearly to 40 % relative humidity at 50 °C
WARNING
Do not disassemble or disconnect probes while they are connected to a voltage source. Use only grounded oscilloscopes.
Attention: Refer to this manual before using the probe.
Danger. High Voltage.
ELECTRICAL CHARACTERISTICS
System attenuation: 1000:1 System input resistance: 100 M Input capacitance: < 3 pF Compensation range: 10 pF to 65 pF System BW (-3 dB): 100 MHz typical Input voltage:
1. Max. voltage: 20 kV DC incl. peak AC
2. AC: 14 kV AC rms 50/60 Hz Sinus
3. HF-Sinus-AC: See derating chart
4. Pulse peak: 40 kV (see peak pulse derating chart
LF PROBE ADJUSTMENT (1000:1)
Connect the probe to a 1 kHz square wave signal. Adjust compensation trimmer in the BNC-box for optimum square wave response.
Undershoot
Correct
Overshoot
HF PROBE ADJUSTMENT
If necessary the probe can be adjusted for high frequency. Connect the probe to a 1 MHz square wave signal and adjust the trimme rin order to receive an optimum square wave response.
Undershoot
Correct
Overshoot
STANDARD ACCESSORIES SUPPLIED
WITH THE PROBE
Accessories Quantities
1. Trimming tool 1
2. Ground lead with a crocodile clip 1
3. Hook 1
HIGH VOLTAGE PROBE KIT PK104
Accessories Quantities
1. Trimming tool 1
2. Ground lead with a crocodile clip 1
3. Hook 1
4. BNC Adapter 1
Derating PPE 20 kV
1000
10000
100000
0,01 0,1 1 10 100
Frequency [MHz]
Maximum Impuls Derating - PPE20kV
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
10 100 1000
Duration [msec]
10% 25% 50% 100%
Duty Cycles