Teledyne PP018 User Manual

PP017/PP018 Passive Probe
Instruction Manual
Manufactured under an ISO 9000 Registered Quality Management System. Visit www.teledynelecroy.com to view the certificate.
This electronic product is subject to disposal and recycling regulations that vary by country and region. Many countries prohibit the disposal of waste electronic equipment in standard waste receptacles. For more information about proper disposal and recycling of your Teledyne LeCroy product, visit www. teledynelecroy.com/recycle.
Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 www.teledynelecroy.com
PP017/PP018 Passive Probe Instruction Manual © 2012 Teledyne LeCroy Corporation. All rights reserved.
Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal sales and distribution purposes is strictly prohibited. However, clients are encouraged to distribute and duplicate Teledyne LeCroy documentation for their own internal educational purposes.
Teledyne LeCroy, HDO 4000 and HDO 6000 are registered trademarks of Teledyne LeCroy Corporation. Other product or brand names are trademarks or requested trademarks of their respective holders. Information in this publication supersedes all earlier versions. Specifications are subject to change without notice.
920876-00 Rev A
Instruction Manual
920876-00 Rev A 1
Warranty
NOTE: THE WARRANTY BELOW REPLACES ALL OTHER WARRANTIES, EXPRESSED
OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY, FITNESS, OR ADEQUACY FOR ANY PARTICULAR PURPOSE OR USE. TELEDYNE LECROY SHALL NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER IN CONTRACT OR OTHERWISE. THE CUSTOMER IS RESPONSIBLE FOR THE TRANSPORTATION AND INSURANCE CHARGES FOR THE RETURN OF PRODUCTS TO THE SERVICE FACILITY. TELEDYNE LECROY WILL RETURN ALL PRODUCTS UNDER WARRANTY WITH TRANSPORT PREPAID.
Teledyne LeCroy warrants this oscilloscope accessory for normal use and operation within specification for a period of one year from the date of shipment. Spare parts, replacement parts and repairs are warranted for 90 days.
In exercising its warranty, Teledyne LeCroy, at its option, will either repair or replace any assembly returned within the warranty period to the Customer Service Department of an authorized service center. However, this will be done only if the product is determined by Teledyne LeCroy’s examination to be defective because of workmanship or materials, and the defect is not caused by misuse, neglect, accident, abnormal conditions of operation; or, damaged by attempted repair or modifications by a non-authorized service facility.
The customer will be responsible for the transportation and insurance charges for the return of products to the service facility. Teledyne LeCroy will return all products under warranty with transportation charges prepaid.
This warranty replaces all other warranties, expressed or implied, including, but not limited to, any implied warranty of merchantability, fitness, or adequacy for any particular purposes or use. Teledyne LeCroy shall not be liable for any special, incidental, or consequential damages, whether in contract or otherwise.
PP017/PP018 Passive Probe
2 920876-00 Rev A
WARNING
Warnings identify conditions or practices that could result in injury or loss of life.
DANGER
High Voltage
ATTENTION
Refer to Manual
DOUBLE INSULATION PROTECTIVE (EARTH) TERMINAL
WARNING
To avoid electric shock, keep fingers behind the probe’s finger
guard during use.
Terms and Symbols
The following terms appear in this manual:
The following symbols may appear on the product:
Introduction
The Teledyne LeCroy PP017 and PP018 are passive high impedance voltage oscilloscope probes designed and calibrated for use with
general purpose oscilloscopes having an input impedance of 1 MΩ
shunted by 16 pF. However, they may be compensated for use with instruments having an input capacitance between 10 to 30 pF. They are compatible with oscilloscopes that automatically detect probe attenuation and adjust their readout accordingly.
Instruction Manual
920876-00 Rev A 3
Model
Attn.
Input
Imped.
B/W
Rise
Time
Length
Compensation
Range
R (MΩ)
C (pF)
(MHz)
(ns)
(m)
(pF)
PP017
10x
10
12
250
1.4
1.3
10 - 30
PP018
10x
10
10
500
0.7
1.3
6 - 22
WARNING
To reduce risk of fire or electric shock, do not exceed the voltage or category ratings of the probe.
Specifications
Attenuation Ratio ................................... 10x (10:1) 10:1 ±0.5%
Max. Input Voltage ................................. 600 V CAT I or 300 V CAT II (DC + peak AC)
................................................................... Derated with Frequency (see Derating Curve)
Pollution Degree .................................... Pollution Degree 2 (as defined in EN 61010-031)
Max. Operating Temp ........................... 0° to +50 C°
Humidity .................................................. 85% RH or less (at 35° C)
Definitions:
Measurement Category I (CAT I) is for measurements performed on circuits not directly connected to mains.
Measurement Category II (CAT II) is for measurements performed on circuits directly connected to the low voltage installation. Examples are on household appliances, portable tools, and similar equipment.
Pollution Degree 2 refers to an operation environment where normally only dry non-conductive pollution occurs. Temporary conductivity caused by condensation must be expected.
PP017/PP018 Passive Probe
4 920876-00 Rev A
EC Declaration of Conformity
Compliance was demonstrated to the specification as listed in the Official Journal of the European Communities: Low Voltage Directive 2006/95/EC.
EN 61010-031/A1:2008 Safety requirements for electrical equipment for measurement, control and laboratory use – Part 031: Safety requirements for handheld probe assemblies for electrical measurement and test.
General Safety Information
To avoid personal injury and to prevent fire or damage to the probe or any product connected to it, review and comply with the following safety precautions.
Using the probe and/or the test instrument it is connected to in a manner not specified by the manufacturer may impair the protection mechanisms.
Connect to properly grounded instruments. Use only with test instruments
having their BNC input connected to earth ground. Do not connect the probe ground terminal to any point which is at a potential other than earth ground.
Do not disconnect the probe from instrument during measurement. Connect the probe to the measurement instrument before connecting the probe to the test circuit.
Do not apply to the input any potential that exceeds the maximum rating of the probe. Use only the accessories provided with the probe.
Comply with voltage derating curve. When measuring higher frequency
signals, be sure to comply with the Voltage vs. Frequency Derating Curve.
Do not remove probe casing. Removal of the probe’s casing may expose you to electric shock.
Do not use if any part is damaged. All maintenance should be referred to qualified service personnel only.
Instruction Manual
920876-00 Rev A 5
Do not use in wet or explosive atmospheres.
Avoid physical injury. The probe tip is extremely sharp. Use care when
handling to prevent injury, including accidental skin puncture.
For indoor use only.
Cleaning
Use the following guidelines:
Clean only the exterior of the probe, cables, and accessories. Use a
soft cotton cloth moistened with a mild detergent and water solution. Do not allow any portion of the probe to be submerged at any time.
Dry the probe and accessory thoroughly before attempting to
make any voltage measurement.
Do not subject the probe to solvents or solvent fumes as these can
cause deterioration of the probe body, cables, and accessories.
PP017/PP018 Passive Probe
6 920876-00 Rev A
Probe Compensation
Proper compensation of the probe is required to assure amplitude accuracy of the waveform being measured by matching the probe to the oscilloscope’s input capacitance. Compensation should be adjusted whenever the probe is connected to or transferred between oscilloscopes.
Low Frequency (LF)
1. Apply a 1 kHz square wave to the probe or connect to the
oscilloscope’s calibrator output.
2. Adjust the left trimmer (LF) located on the PP017probe body:
or the trimmer located on the PP018 BNC Box:
until you achieve a flat-topped square wave per the center illustration in the figure below.
Instruction Manual
920876-00 Rev A 7
High Frequency (HF)
High frequency should rarely require adjustment. However, if adjustment is required:
1. Apply a 1 MHz square wave to the probe (< 0.7 ns rise-time).
2. Adjust the right trimmer (HF) located on the PP017probe body:
or both trimmers on the PP018 BNC Box (HF2 or HF1):
until you achieve a flat-topped square wave per the center illustration in the figure below.
PP017/PP018 Passive Probe
8 920876-00 Rev A
PP017 Input Impedance Profile
PP018 Input Impedance Profile
Instruction Manual
920876-00 Rev A 9
Part Name
Toxic or Hazardous Substances and Elements
Lead
(Pb)
Mercury
(Hg)
Cadmium
(Cd)
Hexavalent
Chromium
(Cr6+)
Polybrominated
Biphenyls
(PBB)
Polybrominated
Diphenyl Ethers
(PBDE)
PCBAs X O X X X X
Mechanical Hardware
O O X O O
O
Sheet Metal
O O X O O
O
Plastic Parts
O O O O X X Protective Case
O O O O X
X
Cable Assemblies
X O X O X X Probe Tips
X O X O X
X
O: Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement specified in SJ/T11363-2006.
X: Indicates that this toxic or hazardous substance contained in at least one of the homogenous materials used for this part is above the limit requirement specified in SJ/T11363-2006.
Voltage vs. Frequency Derating Curve
China RoHS Compliance
EFUP (Environmental Friendly Use Period): 30 Years. Use Conditions: refer to the environmental conditions stated in the Specifications section of this Manual.
PP017/PP018 Passive Probe
10 920876-00 Rev A
920876-00 Rev A 11
Replaceable Parts
Item
Description
Quantity
1
Sprung Hook, 5 mm, Black
1
2
GND Lead w/Alligator Clip
1
3
BNC Adapter, 5 mm
1
4
Replacement Tip, Black
1
5
Deluxe Trimmer Tool
1
6
IC Tip Insulator, 5 mm, Black
1
7
Tip Insulator, 5 mm, Black
1
8
Identifier Rings
1
9
Probe Tip GND
1
Instruction Manual
Standard Accessory Replacement Kit part number is PKIT4-5MM-101.
PP017/PP018 Passive Probe
12 920876-00 Rev A
Teledyne LeCroy Offices World Wide
United States and Canada - World Wide Corporate Office Teledyne LeCroy Corporation 700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Ph: 800-553-2769 (Sales 1, Applications 2, Service 3) or 845-425-2000 Fax: 845-578-5985 contact.corp@teledynelecroy.com www.teledynelecroy.com
US Protocol Solutions Group Teledyne LeCroy Corporation 3385 Scott Boulevard Santa Clara, CA, 95054 Sales and Service: Ph: 800-909-7211/408-727-6600 Fax: 408-727-0800 contact.corp@teledynelecroy.com www.teledynelecroy.com Support: Ph: 800-909-7112/408-653-1260 psgsupport@teledynelecroy.com
Europe Teledyne LeCroy Europe 4, Rue Moïse Marcinhes Case postale 341 1217 Meyrin 1 Geneva, Switzerland Ph: +41 22 719 2228/2323/2277 Fax: +41 22 719 2233 contact.sa@teledynelecroy.com, applications.indirect@teledynelecroy.com www.teledynelecroy.com/europe Protocol Analyzers: Ph: +44 12 765 0397 1
Contact Teledyne LeCroy
Contact your local Teledyne LeCroy office for sales and support.
Instruction Manual
920876-00 Rev A 13
China Teledyne LeCroy Beijing Rm. 2001 Unit A, Horizon Plaza No. 6, Zhichun Rd., Haidian Dist. Beijing 100088, China Ph: ++86 10 8280 0318/0319/0320 Fax: +86 10 8280 0316 www.teledynelecroy.com.cn Service: Rm. 2002 Ph: ++86 10 8280 0245
Taiwan LeColn Technology Co Ltd. Far East Century Park, C3, 9F No. 2, Chien-8th Road, Chung-Ho Dist., New Taipei City, Taiwan Ph: ++ 886 2 8226 1366 Fax: ++ 886 2 8226 1368 sales_twn@teledynelecroy.com.tw www.teledynelecroy.com.tw
Japan Teledyne LeCroy Japan Hobunsya Funchu Bldg, 3F 3-11-5, Midori-cho, Fuchu-Shi Tokyo, 183-0006 Japan Ph: ++ 81 4 2402 9400 Fax: ++ 81 4 2402 9586 www.teledynelecroy.com/japan
Korea Teledyne LeCroy Korea 10th fl. Ildong Bldg. 968-5 Daechi-dong, Gangnam-gu Seoul 135-280, Korea Ph: ++82 2 3452 0400 Fax: ++82 2 3452 0490 www.teledynelecroy.com.kr
Singapore (Oscilloscopes) Teledyne LeCroy Singapore Pte Ltd. Blk 750C Chai Chee Road #02-08 Technopark@ChaiChee Singapore 469003 Ph: ++ 65 64424880 Fax: ++ 65 64427811
Singapore (Protocol Analyzers) Genetron Singapore Pte Ltd. 37 Kallang Pudding Road, #08-08 Tong Lee Building Block B Singapore 349315 Ph: ++ 65 9760-4682
PP017/PP018 Passive Probe
14 920876-00 Rev A
Loading...