Teledyne PP007, PP007-WR-1, PP007-WR-2, PP007-WS-1, PP007-WS-2 Operating Manual

PP007-WR-1
PP007-WR-2
PP007-WS-1
PP007-WS-2
Operator’s Manual
© 2017 Teledyne LeCroy, Inc. All rights reserved.
Unauthorized duplication of Teledyne LeCroy documentation materials is strictly prohibited. Customers are permitted to duplicate and distribute Teledyne LeCroy documentation for internal educational purposes.
WaveSurfer, WaveRunner, and Teledyne LeCroy are registered trademarks of Teledyne LeCroy, Inc. Windows is a registered trademark of Microsoft Corporation. Other product or brand names are trademarks or requested trademarks of their respective holders. Information in this publication supersedes all earlier versions. Specifications are subject to change without notice.
Warranty
Teledyne Lecroy warrants this oscilloscope accessory for normal use and operation within specification for a period of one year from the date of shipment. Spare parts, replacement parts and repairs are warranted for 90 days.
In exercising its warranty, Teledyne Lecroy, at its option, will either repair or replace any assembly returned within its warranty period to the Customer Service Department or an authorized service center. However, this will be done only if the product is determined by Teledyne Lecroy’s examination to be defective due to workmanship or materials, and the defect is not caused by misuse, neglect, accident, abnormal conditions of operation, or damage resulting from attempted repair or modifications by a non-authorized service facility.
The customer will be responsible for the transportation and insurance charges for the return of products to the service facility. Teledyne Lecroy will return all products under warranty with transportation charges prepaid.
This warranty replaces all other warranties, expressed or implied, including but not limited to any implied warranty of merchantability, fitness or adequacy for any particular purposes or use. Teledyne Lecroy shall not be liable for any special, incidental, or consequential damages, whether in contract or otherwise.
928597-00 Rev A July, 2017
Operator’s Manual
1
Safety Instructions
Follow generally accepted safety procedures in addition to the precautions specified in this section. The overall safety of any system incorporating this
accessory is the responsibility of the assembler of the system.
Symbols
These terms and symbols may appear on the probe body or in this manual to alert you to important safety considerations.
High Voltage WARNING, risk of electric shock or burn.
CAUTION of potential damage to equipment, or WARNING of
potential injury. Attend to the accompanying information to protect against personal injury or damage. Do not proceed until conditions
are fully understood and met.
DOUBLE INSULATION
PROTECTIVE (EARTH) TERMINAL
Precautions
To avoid personal injury or damage to property, review and comply with the following safety precautions. Use product only as specified.
Connect only to grounded instruments. Use only with compatible Teledyne LeCroy oscilloscopes that have their BNC input connected to an earth ground. Do not connect the probe reference lead to any point which is at a potential other than earth ground.
Connect and disconnect properly. Connect probe to the oscilloscope before connecting the probe to the test circuit. Disconnect the probe input and reference lead from the test circuit before disconnecting the probe from the oscilloscope. To avoid injury or death due to electric shock, do not connect or disconnect probes or probe accessories while they are connected to a voltage source.
PP007 Passive Probe
2
Do not overload. To avoid electric shock or fire, do not apply any potential to the probe leads that exceeds the maximum rating of the probe.
Comply with voltage derating curve. When measuring higher frequency signals, comply with the Voltage vs. Frequency Derating Curve.
Observe all terminal ratings. To avoid electric shock or fire, observe all markings on the oscilloscope before connecting. Consult the respective oscilloscope product manual for further ratings information.
Do not remove probe casing. Removing the probe’s case or touching exposed connections may result in electric shock.
Use only within operational environment listed. Use indoors only. Do not use in wet or explosive atmospheres. Keep product surfaces clean and dry.
Use only accessories compatible with the probe.
Handle with care. The probe tip is extremely sharp and may puncture skin or
cause other bodily injury if not handled properly.
Keep fingers behind the finger guard of probe body and accessories.
Do not operate with suspected failures. Before each use, inspect the probe
and accessories for any potential damage such as tears or other defects in the probe body, cable jacket, accessories, etc. If any part is damaged, cease operation immediately and sequester the probe from inadvertent use.
Operator’s Manual
3
Introduction
The PP007 is a miniature high impedance passive probe. Its high input resistance and low capacitance make it ideal from general purpose probing of signals with frequency content from DC through several hundred MHz. The PP007 has a large selection of connection accessories, both supplied standard with the probe and available from Teledyne Lecroy as optional accessories.
The PP007 is available in two models (-WR for use with WaveRunner and –WS for use with WaveSurfer), differing only in the design of the compensation network.
All accessories are interchangeable between all models.
Keep fingers behind the probe’s finger guard during use.
Rated for indoor use only. Do not operate where conductive pollutants may be present.
Maximum input voltage derates when frequency response is approximately 30 kHz or higher. See voltage derating curve p.4.
Specifications
Electrical Characteristics
Attenuation ÷10 ±1%
Input Resistance 10 M ±1%
Input Capacitance 9.5 pF
Input Impedance (see plot p.5)
Compensation Range 10 – 20 pF
Bandwidth 500 MHz (-3 dB)
Electrical Ratings*
Max. Input Voltage Circuits not directly connected to Mains, 400 Vrms,
1250 V transient overvoltage
CAT II, 300 Vrms
Pollution Degree 2
* See Certifications for an explanation of rating categories.