Teledyne ECO-02-0227 User Manual

INSTRUCTION MANUAL
311XL SERIES
0-2 PPM Full Scale
Trace Oxygen Analyzer
D ANGER
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM. HAZARDOUS VOLT A GES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MA Y PERSIST FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONL Y A UTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SER VICING. BEFORE CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/ MANAGER.
Teledyne Analytical Instruments
P/N M70255
08/14/02
ECO # 02-0227
INSTRUCTION MANUAL
FOR
TRACE OXYGEN ANALYZER
MODEL 311XL
TELEDYNE ANALYTICAL INSTRUMENTS
16830 CHESTNUT STREET
CITY OF INDUSTRY, CA 91749
TELEPHONE: (888) 789-8168 (626) 934-1500 (626) 961-9221
FAX: (626) 961-2538 (626) 934-1651
Web: www.teledyne-ai.com
WARNING
The sensor(s) used in this instrument uses electrolytes which contain substances that are extremely harmful if touched, swallowed, or inhaled. Avoid contact with ANY fluid or powder in or around the unit. What may appear to be plain water could contain one of these toxic substances.
SKIN CONTACT: Flush contact area with cold water for several minutes. Seek medical attention.
EYE CONTACT: Flush with cold water for at least 15 minutes. Seek immediate medical attention.
IF SWALLOWED: Drink milk or milk of magnesia. Seek immediate medical attention.
IF INHALED: Move immediately to an area of fresh air, or assist victim to fresh air. If victim is having difficulty breathing, use artificial respiration or administer oxygen. Seek immediate medical attention.
TABLE OF CONTENTS
311XL SERIES
1. INTRODUCTION 1
1.1 Description 1
1.2 Method of Analysis 1
1.3 Outstanding Features 1
1.3.1 Micro-Fuel Cell 1
1.3.2 Reliable Calibration 2
1.3.3 Integral Power Supply 2
1.3.4 Accuracy and Response 3
1.3.5 Compact Packaging 3
2. SUPPORTING EQUIPMENT AND SERVICES 4
2.1 Sampling Equipment 4
2.2 Power Service 4
3. OPERATION 5
3.1 Introduction 5
3.2 Calibration 6
3.2.1 Standard (Span) Calibration Procedure 6
3.3 Positive Pressure Sampling 6
3.4 Atmospheric Pressure Sampling 7
4. MAINTENANCE 8
4.1 Battery Power Supply Services 8
4.2 Routine Maintenance 8
4.3 Cell Replacement 8
4.4 Cell Warranty 19
5. TRANSDUCTION AND TEMPERATURE COMPENSATION 10
6. LEAK TESTING 10
7. SPECIFICATION DATA 12
SPARE PARTS LIST 13
DRAWING LIST 14
APPENDIX - Effect of CO2 on B-2CXL Cell Life 15
311XL SERIES
TRACE OXYGEN ANALYZER
1. INTRODUCTION
1.1 Description: The Teledyne Analytical Instruments (TAI) Model 311XL and 311TCXL are portable, intrinsically safe trace oxygen analyzer which can be operated without an external power source. These analyzers will be referred to as Model 311XL or 311XL Series analyzers in this manual.
The instrument provides for trace oxygen analysis in decade steps ranging from 0-2 to 0-10,000 ppm (full scale) plus a special calibration range that encompasses the known oxygen concentration of atmospheric air (209,000 ppm).
Sample oxygen is read from an extremely accurate integral meter whose range of measurement is determined by the position of the range selector switch. The linear 5 inch scale (mirror equipped to eliminate parallax) provides excellent resolution and accuracy.
Sample gas is introduced and vented via a pair of quick disconnect fittings that feature integral shutoff valves which automatically close when the mating male fitting is withdrawn. The fittings are an integral part of the measuring cell manifold so that internal sample passage volume is at an absolute minimum. Sample flow control of 0.5 to 5 liters/min. must be accomplished with accessory equipment.
1.2 Method of Analysis. The sample oxygen is measured by a unique electrochemical transducer which functions as a fuel cell. Oxygen diffusing into the cell reacts electro-chemically to produce an electrical current that is proportional to the oxygen concentration in the gas phase immediately adjacent to the transducer’s sensing surface. The linear, but minute, signal produced by the transducer from trace oxygen is amplified by a two stage amplifier. The dual stages of amplification provide enough gain to drive meter and thermistor controlled network utilized to compensate for the positive temperature co-efficient of the transducer.
1.3 Outstanding Features the Model 311XL:
1.3.1 Micro-Fuel Cell ducer with no electrolyte to change or electrodes to clean. When the cell reaches the end of its useful life, it is merely thrown away and replaced, as one would replace a worn out battery in a flashlight.
. The following unique features are incorporated into
: The Micro-Fuel Cell* is a sealed electrochemical trans-
-1-
1.3.2 Reliable Calibration. Although this instrument can be calibrated with air, the use of a calibration gas gives better results on the lower PPM ranges. If the instrument is exposed to high concentrations of oxygen (percent levels), it will take a significant amount of time to purge the oxygen from the sensor and analyzer for use in the low PPM level measurements (24 hours or longer).
1.3.3 Integral Power Supply. The differential power +/- 3.6 volts D.C.) of the instrument amplifier is furnished by two internally mounted 750 milliampere hour nickel cadmium batteries. Fully charged, these batteries will provide enough power to operate the instrument continuously for a period of about 30 days. Furthermore, an overnight charge on a three weeks duty cycle should keep the original batteries supplied usable for many years.
An integral charging circuit and a detectable power cord are provided so that the batteries may be recharged from any 50 or 60 cycle, 100 to 115 volt, convenience outlet.
The instrument is designed to either sample or have its batteries recharged. Both operations cannot be carried out simultaneously. TAI has deliberately interlocked the circuitry so that both operations cannot be carried on at the same time.
Only when the selector switch is placed in the OFF position will the neon lamp on the back plate of the Model 311XL light up to indicate power to the battery charging circuit.
* U.S. Pat. Nos. 3,767,552 and 3,668,101
-2-
Loading...
+ 12 hidden pages