Teledyne CP030A, CP031, CP030, CP031A Operator's Manual

Operator’s Manual
CP030/CP030A CP031/CP031A Current Probes
CP03x/CP03xA Current Probes
Operator’s Manual
September, 2016
CP030/CP030A and CP031/CP031A Current Probes Operator’s Manual
© 2016 Teledyne LeCroy, Inc. All rights reserved.
Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal sales and distribution purposes is strictly prohibited. However, clients are encouraged to duplicate and distribute Teledyne LeCroy documentation for internal educational purposes.
Teledyne LeCroy is a registered trademark of Teledyne LeCroy, Inc. Windows is a registered trademark of Microsoft Corporation. Other product or brand names are trademarks or requested trademarks of their respective holders. Information in this publication supersedes all earlier versions. Specifications are subject to change without notice.
926878-00 Rev A September, 2016
Operator’s Manual
i
Contents
Safety Instructions ................................................................................................................................ 1
Symbols .............................................................................................................................................. 1
Precautions ........................................................................................................................................ 1
Introduction ............................................................................................................................................ 3
Specifications ........................................................................................................................................ 4
Warranted Characteristics ................................................................................................................ 4
Nominal Characteristics .................................................................................................................... 4
Typical Characteristics ...................................................................................................................... 5
Environmental Characteristics ......................................................................................................... 5
Physical Characteristics ................................................................................................................... 5
Safety Ratings .................................................................................................................................... 6
CP030/CP030A Derating Curve - Maximum Input Current vs. Frequency*................................... 7
CP030/CP030A Insertion Impedance vs. Frequency* .................................................................... 7
CP031/CP031A Derating Curve - Maximum Input Current vs. Frequency .................................... 8
CP031/CP031A Insertion Impedance vs. Frequency ...................................................................... 8
Operation ................................................................................................................................................ 9
Precautions ........................................................................................................................................ 9
Connecting to the Test Instrument ................................................................................................10
Connecting to the Test Circuit ........................................................................................................10
Operating with an Oscilloscope ......................................................................................................11
Performance Verification ....................................................................................................................13
Test Equipment Required ................................................................................................................14
Preliminary Procedure .....................................................................................................................14
Functional Check .............................................................................................................................15
Check LF Accuracy ..........................................................................................................................15
Performance Verification Test Record ...........................................................................................16
CP03x/CP03xA Current Probes
ii
CP03x Test Record ..............................................................................................................................17
Care and Maintenance .........................................................................................................................18
Cleaning ............................................................................................................................................18
Calibration Interval ..........................................................................................................................18
Service Strategy ...............................................................................................................................18
Troubleshooting ...............................................................................................................................18
Returning a Product for Service ..........................................................................................................19
Teledyne LeCroy Service Centers .......................................................................................................20
Warranty ...............................................................................................................................................21
Certifications ........................................................................................................................................22
EMC Compliance ..............................................................................................................................22
Safety Compliance ...........................................................................................................................23
Environmental Compliance .............................................................................................................24
Operator’s Manual
1
Safety Instructions
Follow these instructions to keep the probe operating in a correct and safe condition. Observe generally accepted safety procedures in addition to the precautions specified here. The overall safety of any system incorporating this accessory is the responsibility of the assembler of the system.
Symbols
These symbols appear on the probe body or in documentation to alert you to important safety considerations.
CAUTION, possibility of electric shock.
CAUTION of damage to probe or instrument, or WARNING of hazard to health.
Attend to the accompanying information to protect against personal injury or damage. Do not proceed until conditions are fully understood and met.
Do not apply around or remove from UNINSULATED HAZARDOUS LIVE
conductors which may render electric shock, electric burn, or arc flash.
Equipment protected by double insulation or reinforced insulation.
CAUTION, hot surface.
Precautions
Comply with the following safety precautions to avoid personal injury or damage to your equipment:
Use only as specified. The probe is intended to be used only with compatible Teledyne LeCroy instruments. Using the probe and/or the equipment it is connected to in a manner other than specified may impair the protection mechanisms.
Do not use the probe for measurements on Mains circuits. The probe should only be applied around or removed from insulated limited energy circuit conductors that are not directly connected to the mains.
CP03x/CP03xA Current Probes
2
Do not overload; observe all ratings. To avoid electric shock or fire, do not connect the
current probe to any wire that carries voltages or currents that exceed the ratings of the probe.
Connect and disconnect properly. Connect the probe to the measurement instrument before connecting to the circuit/conductor being measured. Avoid damaging the cable through excessive bending.
Never install or remove the probe on bare conductors which are energized. The transformer core and shield are grounded but not insulated and may contact the conductor when the locking lever is open.
Be careful not to damage the insulation surface when making measurements. Before clamping to the conductor being measured, check that the insulation on the conductor is undamaged, and take care not to damage the insulation when clamping the conductor. Any damage to the insulation could cause an electric shock.
Use only indoors and within the operational environment listed. Do not use in wet or explosive atmospheres.
Do not remove the probe's casing. Touching exposed connections may result in electric shock.
Keep product surfaces clean and dry.
Comply with the maximum input current vs. frequency derating when measuring current that includes a high frequency component. Using the probe at high frequencies or in strong magnetic fields may cause the device to become abnormally hot, resulting in fire, equipment damage or burns.
Do not operate with suspected failures. Before each use, inspect the probe and accessories for any damage such as tears or other defects in the probe body, cable jacket, accessories, etc. If any part is damaged, cease operation immediately and sequester the probe from inadvertent use.
NOTE: Depending on the amplitude and frequency of the current being measured, the sensor head may emit a resonant sound. This sound may also occur during demagnetizing operation, but it does not represent a malfunction (device failure).
Operator’s Manual
3
Introduction
The CP030, CP030A, CP031, and CP031A current probes are designed for easy, highly accurate current measurements. The compact probes offer wide bandwidth with over-current protection. The probes utilize a combination of Hall-effect and transformer technology which enables measurements to be made on DC, AC and impulse currents.
The CP030 and CP030A have a 50 MHz bandwidth and are designed to measure continuous currents up to 30 Amp. The CP030-3M probe has a 10 MHz bandwidth and is designed to measure continuous currents up to 30 Amp.
NOTE: All CP030A probes, regardless of cable length, will indicate default probe bandwidth of 50 MHz in the Attributes section of the Probe dialog.
The CP031 and CP031A have a 100 MHz bandwidth and are designed to measure continuous currents up to 30 Amp.
The CP03x probes are compatible with any Teledyne LeCroy oscilloscope with a ProBus interface running Windows XP Professional
®
or later.
The CP03xA probes are compatible with most Teledyne LeCroy oscilloscopes with a ProBus interface running X-Stream™ firmware version 7.8.x.x or later. See table below.
Compatible
Compatible with Upgrade
Not Compatible
LabMaster 9MZi-A
DDA/SDA/WaveMaster 8 Zi/Zi-A/Zi-B DDA/SDA/WavePro 7 Zi/Zi-A HRO/WaveRunner 6 Zi WaveRunner Xi/Xi-A/Mxi/Mxi-A HDO8000, HDO6000, HDO4000 MDA800 WaveSurfer 10 WaveSurfer MXs/MXs-A/MXs-B
WaveSurfer Xs-A
WavePro 7000/7000A*
WaveSurfer Xs†
LabMaster 10 Zi/Zi-A
LabMaster 9 SZi-A SDA13000, SDA18000 SDA9000, SDA11000 WaveExpert WaveRunner 6000/6000A WaveSurfer 400 WaveSurfer 3000 All other legacy instruments
*Compatible if running Windows XP Professional and X-Stream 7.8.x.x or later. †Must be upgraded to Windows XP Professional, not Windows XP Embedded. Contact your regional service center regarding upgrades.
With the ProBus interface, the probe becomes an integral part of the measuring instrument. The bandwidth limit, Auto Zero and Degauss functions are all controlled from the instrument’s touch screen user interface. The interface provides power to the probe, so no external power supply is needed.
CP03x/CP03xA Current Probes
4
Specifications
For the most current specifications, see the probe product pages at teledynelecroy.com. Specifications are subject to change without notice.
Warranted Characteristics
CP030/CP030-3M
CP030A
CP031
CP031A
Amplitude Accuracy
±1.0 % of reading to 30 Arms
(45 to 65 Hz)
±2.0% of reading to 50 Apeak
±1.0 % of reading to 30 Arms
(45 to 65 Hz)
±2.0% of reading to 50 Apeak
Bandwidth (probe only)
DC to 50 MHz/10 MHz
DC to 100 MHz
Nominal Characteristics
CP030
CP030A
CP031
CP031A
Max Continuous Input Current at 0.1 V/A
30 Arms
30 Arms Max Peak Current
50 Apeak non-continuous
50 Apeak non-continuous
Intended Output Load
1 MΩ
1 MΩ
Measured Conductors
Insulated Conductors only
Insulated conductors only
Sensitivity *
10 mA/div to 15 A/div
1 mA/div to 15 A/div
10 mA/div to 15 A/div
1 mA/div to 15 A/div
Offset Range (1x Attenuation)
N/A
±5 A
N/A
±5 A
Max Continuous Input
Current at 1 V/A
(100mA/div or less)
N/A
5 A
N/A
5 A
* Values are based on oscilloscopes with 1 mV/div sensitivity. Numbers will be higher on instruments with lower sensitivity.