TELCARE TCTM0001 User Manual

Page 1
TELCARE™ Blood Glucose Monitoring System
TELCARE™ Blood Glucose
Monitoring System
User Manual
Page 2
TELCARE™ Blood Glucose Monitoring System
e TELCARE™ Blood Glucose Monitoring system is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the ngertips, forearm, or palm. It is intended for lay use by persons with diabetes to aid in diabetes management. It is indicated for use at home (over the counter [OTC]) and should be used only by a single patient and should not be shared. Test results are plasma­calibrated for easy comparison to clinical laboratory methods. Testing is done outside the body (in vitro diagnostic use). e TELCARE™ Blood Glucose Monitoring System consists of the Telcare BGM™ Blood Glucose Meter, TELCARE™ Blood Glucose Test Strips, and TELCARE™ Glucose Control Solutions. e TELCARE™ Blood Glucose Monitoring system is not indicated for the diagnosis or screening of diabetes or for neonatal use. Palm and forearm testing should be done only during steady-state times when glucose is not changing rapidly. e Telcare BGM™ Blood Glucose Meter uses cellular data transmission to send test results to TELCARE™’s remote database, Telserve, and to receive messages from Telserve. e TELCARE™ Blood Glucose Monitoring System is not intended to provide automated treatment guidance or decisions, nor is it to be used as a substitute for professional healthcare judgment.
Intended Use
(i)
Page 3
TELCARE™ Blood Glucose Monitoring System
e TELCARE™ Blood Glucose Monitoring System and lancing device is for single patient use. Do not share it with anyone, including family members! Do not use on multiple patients!
Because all parts of the kit may come in contact with your blood, all parts are considered biohazardous. Kit parts can potentially transmit infectious diseases, even aer cleaning and disinfection.
Users should wash hands thoroughly with soap and water and dry thoroughly aer handling the meter, test strips, and any lancing device.
For Cleaning and Disinfecting procedures to prevent the spread of diseases such as Hepatitis B, see Cleaning and Disinfecting Your Meter (pg. 31).
For Technical Support contact TELCARE™ at (888) 666-4530. Hours of Operation: Monday - Friday, 9AM - 5PM EST
References
“FDA Public Health Notication: Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Blood­borne Pathogens: Initial Communication” (2010)
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm224025.htm
“CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne Pathogens” (2010)
http://www.cdc.gov/injectionsafety/Fingerstick-DevicesBGM.html
“Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline–ird Edition. Clinical and Laboratory Standards Institute (CLSI), document M29-A3, (ISBN 1-56238-567-4).
Important Safety Instructions
(ii)
Page 4
TELCARE™ Blood Glucose Monitoring System
(iii)
Your Telcare Blood Glucose Meter
The Telcare BGM
Device 2 Telcare Test Strips 3 The Display Screen 4 Setting Up Your Telcare BGM 5 Settings Options 6
Control Solution Modes
Step by Step Guide to Control Testing 8
Testing Your Blood Sugar
Preparation and Safety 14 Prepare Your Lancing Device 17 Understanding Your Blood Glucose Test Results 28
Viewing Your Meter’s Memory 29
Maintenance of Your Telcare BGM and Lancing Device
Cleaning and Disinfecting Your Meter 31 Cleaning and Disinfecting Your Lancing Device 33 Charging Your Telcare Blood Glucose Meter 35
Storing Your Meter and Test Strips 36
Troubleshooting 37
Terms of Warranty 40
Specifications 41
About Telcare 42
Contact Telcare 43
Table of Contents
Page 5
TELCARE™ Blood Glucose Monitoring System
1
Unboxing the Telcare BGM
for the first time
Before we begin, take a moment to go over the contents of your Telcare BGM kit. Your kit should include:
n Telcare BGM
Blood Glucose Meter (battery included)
n AC Charger
n Instruction Manual n Telcare Test Strips (10) n Telcare Glucose Control Solution n Lancing Device n Lancets
Make sure you charge your glucose meter for at least 12 hours before first use. To charge your glucose meter, use the AC Charger included in your kit.
You should use the Telcare BGM for:
n Testing blood glucose.
n Testing only with fresh capillary whole blood samples.
You should NOT use the Telcare BGM for:
n The diagnosis of diabetes.
n Testing blood glucose on newborns.
n Testing of arterial blood.
n Testing of venous blood.
Your Telcare Blood Glucose Meter
Page 6
TELCARE™ Blood Glucose Monitoring System
2
The Telcare BGM™ Device
2
6
5
7
1
1 Power Button
2 UP / DOWN Keys
3 Enter Button
4 Display Screen
5 Insert Test Strip
(white side up)
6 Charging Port
(do not attempt to
insert strip here)
7 Wireless Signal
Indicator
2
3
4
Page 7
TELCARE™ Blood Glucose Monitoring System
3
Telcare Test Strips
Blood Sample End
Contact End
(put into device)
IMPORTANT NOTES ON TEST STRIPS:
n Because of the autocoding feature of this meter, the strip insertion will be
snug. This is to be expected and users find it easy after a few attempts.
n Check the expiration date printed on the test strip bottle and its package.
Do Not use expired test strips.
n Use each test strip immediately after removing it from the bottle.
n Do not use wet or damaged test strips.
n Keep away from direct sunlight and heat. Store the test strip bottle in a dry,
cool place.
n Record the “date opened” on the bottle label. Discard the bottle and any
remaining test strips 3 months after the date of opening.
n Do not operate the meter outside the temperature range of 50˚F to 104˚F.
Page 8
TELCARE™ Blood Glucose Monitoring System
The Display Screen
4
From the Display screen, you can see the following things:
n Wireless Signal Strength n Current Time n Battery Level n Last Test Reading n Time of Last Reading
In this Display screen, use the UP/DOWN buttons (located on the left of your meter) to access the meter’s options:
n Test Mode n L1 Control Solution Mode (Low) n L2 Control Solution Mode (High) n Settings Menu
The Display Screen
1 Wireless Signal Strength
2 Current Time
3 Battery Level
4 Last Test Reading
5 Time of Last Reading
3 2
1
4
5
Page 9
TELCARE™ Blood Glucose Monitoring System
Setting Up Your Telcare BGM
5
Your Telcare BGM is preset with your information, including your time zone. During each cellular connection, Telserve will validate the time displayed is accurate. Telserve will automatically update your BGM clock, if appropriate. Also, if your physical location changes, Telserve will recognize your time zone change and update your meter.
If you would like to view your meter’s preset settings, navigate to the Settings screen.
Getting to the Settings Screen
1 Power (PWR) your Telcare BGM device on.
2 Use the UP/DOWN buttons to reach the “Settings” screen.
IMPORTANT: Slight delays occur when meter powers on. If delay is greater than 20 seconds, do not use meter. Contact technical support at (888) 666-4530.
Page 10
TELCARE™ Blood Glucose Monitoring System
6
Settings Options
Alarms
Create up to 3 alarms to help remind yourself when to test next.
Date and Time
This page allows you to view the Date and Time settings of your device.
Volume
This page allows you to select the volume of your meter. Use the UP/DOWN keys to select the appropriate volume. There are six possible volumes you may select from.
Mobile Network
This is an advanced feature. This page allows you to view the details of the mobile network your device uses for communication to the MyTelcare.com Web Portal. We recommend you do not change the factory settings.
Auto-Sync
Disabling Auto-sync will prevent the meter from communicating with the MyTelcare.com Web Portal (Not Recommended). Readings will still be stored in the meter.
Page 11
TELCARE™ Blood Glucose Monitoring System
7
Memory Recall
Your last 300 glucose and control test readings are stored within the meter. Please see “Viewing Your Meter’s Memory” (page 29) for further details. To send a reading to MyTelcare.com, highlight it and push the Enter button. If a reading has already been stored on MyTelcare.com and you attempt to send it, an error will appear on screen.
Airplane Mode
Turning on Airplane Mode disables cellular network access. You will not be able to transmit your readings while Airplane Mode is turned on. Your meter’s clock will continue to work properly while in this mode.
If a time zone change occurs, your meter will not automatically update to the new time until Airplane Mode has been disabled. The meter will reset the time from the new network.
About
This screen provides a listing of current software and related information about your meter. This is helpful if you need technical support.
Page 12
TELCARE™ Blood Glucose Monitoring System
Control Solution Modes
8
Control solution testing is to make sure your Telcare BGM™ and Test Strips are working properly. You should perform control solution testing when:
n Using the meter for the first time. n Using a new bottle of Telcare Test Strips. n You left the test strip bottle cap open. n You dropped the meter. n You suspect the meter and test strips are not working properly. n The test results do not reflect how you feel. n Practicing the testing. n The test strips are exposed to extreme environmental conditions.
Purchase additional control solution by calling Telcare at (888) 666-4530 (9AM - 5PM EST, M-F).
Step by Step Guide to Control Testing
Power on the Telcare BGM device with the
PWR button on the right side of the BGM.
The device will automatically start in
“Normal Test Mode,” ready to check blood
glucose. You will see the “Insert test strip”
screen as shown.
1 2
Page 13
TELCARE™ Blood Glucose Monitoring System
9
3
Use the UP / DOWN keys on the left side of
the BGM to access the main menu. Switch
modes from “Normal Test Mode” to
L1 Control Solution Mode.
To test Level 1 (L1) control solution, use the
UP/DOWN buttons to highlight L1 mode.
Press the middle (ENTER) button to select
L1 mode.
Upon successful completion of Level 1
(L1), test Level 2 (L2) control solution.
Use the UP / DOWN keys on the left side of
the BGM to access the main menu. Switch
modes from “Normal Test Mode” to
L2 Control Solution Mode.
To test Level 2 (L2) control solution, use the
UP/DOWN buttons to highlight L2 mode.
Press the middle (ENTER) button to select
L2 mode.
Page 14
TELCARE™ Blood Glucose Monitoring System
10
You can select any Mode you want by
pushing the ENTER button.
Take out a Telcare Test Strip from the test
strip bottle. Close the bottle immediately.
IMPORTANT:
- Insert the strip into the strip port at a
slight downward angle for it to properly
seat in the meter.
- The word “Telcare” should be facing up.
- The sample tip should be facing up.
- Continue to push the strip in until Apply
Blood appears.
Sample Tip
4 5
6
Page 15
TELCARE™ Blood Glucose Monitoring System
11
Squeeze a drop of control solution onto a
clean, dry, non-absorbing surface. Use Level
1 solution for L1 mode. Use Level 2 solution
for L2 mode. Do not apply control solution
to the test strip directly from the bottle.
Immediately replace the bottle cap on the
control solution bottle.
Gently touch the tip of the test strip to the
drop of control solution. You should see the
blood sample end of the test strip fill.
The screen will start to count down. After
6 seconds, the control test result will appear
on the meter’s display screen.
7
8
9
Page 16
TELCARE™ Blood Glucose Monitoring System
12
Compare the reading on the screen to
the “L1 Control Solution” or “L2 Control
Solution” range printed on the test strip
bottle or package. If the reading on the
meter falls within the appropriate range,
select the YES tag. If not, select the NO
tag. If the reading falls outside the level
printed on the bottle or package that
you used, please see Control Solution
Troubleshooting on page 13.
10
IMPORTANT TIPS:
1. Only use Telcare Glucose Control Solution with the Telcare Test Strips. Other
brands of control solution or test strips may produce inaccurate results.
2. Always check the expiration date. DO NOT use expired control solution.
3. DO NOT FREEZE. Store the control solution at room temperature.
4. Never touch the tip of the control solution bottle to the test strip.
It will damage all remaining solution in the bottle.
Control Solution Range
Page 17
TELCARE™ Blood Glucose Monitoring System
13
Control Solution Troubleshooting
If your control solution testing is out of range (too high or too low), it may be caused by the following:
n Wrong brand of control solution.
n Control solution not at room temperature.
n Expired or contaminated control solution.
n Damaged test strip.
n Touched the tip of solution bottle to test strip.
n Meter malfunction.
n Wrong ambient temperature (should be between 50˚F and 104˚F).
n Check the expiration date and open bottle date on the control solution
and on the test strips. Repeat testing using a new test strip. If the result is still out of range, use a new bottle of control solution and retest.
If the problem persists, turn your meter off and then call TELCARE™ at (888) 666-4530. Do not perform any blood glucose tests until you resolve this issue.
Page 18
TELCARE™ Blood Glucose Monitoring System
Testing Your Blood Sugar
14
n Set up your Telcare BGM
Blood Glucose Meter properly. Run a control test
before first use. (See Step by Step Guide to Control Testing on page 8.)
n Wash your hands and testing site thoroughly with warm
water and soap. Dry your hands well.
n Test in a place where the temperature range is between
50˚F and 104˚F. (See Troubleshooting on page 37 for more information.)
You will need:
n Your Telcare BGM
Blood Glucose Meter
n A new Telcare Test Strip
n A lancing device with a clean, unused lancet (or safety lancets)
IMPORTANT:
Inaccurate results may occur in severely hypotensive individuals or patients
in shock. Inaccurate low results may occur for individuals experiencing a
hyperglycemic hyperosmolar state, with or without ketosis. Critically ill
patients should not be tested with blood glucose meters.
Before you begin, make sure to:
Page 19
TELCARE™ Blood Glucose Monitoring System
Important Safety Information Before Testing
15
n Please follow cleaning and disinfecting instructions for the meter
(page 31) and lancing device (page 33) to prevent the risk of blood-borne pathogen transmission!
n Follow any cleaning/disinfection instructions that may come with
the lancing device you use.
n The meter and lancing device are intended for a single user and
should not be shared.
n Make sure you wash and dry hands thoroughly with soap and
water before and after handling the meter, lancing device, or test strips.
Page 20
TELCARE™ Blood Glucose Monitoring System
16
Step by Step Blood Sugar Testing
Power on the Telcare BGM™ device with
the PWR button on right side of meter.
The device will automatically start in
“Normal Test Mode.” When ready to
check blood, “Insert test strip” will be
displayed.
Take out a Telcare Test Strip from the
bottle, and close the bottle immediately.
(Some meters require the user to enter a
code from the vial. Telcare’s strips do not
require a code.)
Firmly insert the test strip into your Telcare
BGM device until it is fully seated.
1
2
3
Page 21
TELCARE™ Blood Glucose Monitoring System
Inserting a Lancet
Wash your hands and puncture site with soap and warm water. Rinse and dry thoroughly.
Twist off the Lancing Device Cover
Insert a new Lance firmly into the Lancet Carrier
Hold the Lancet Needle Cover and gently twist off of the Lancet
The Adjustable Tip offers 5 levels of skin penetration. To select the desired depth, twist the Adjustable Tip in either direction until the number line up with the Arrow.
Replace the Lancing Device Cover
3
6
17
Prepare Your Lancing Device
4
Twist off the Lancing Device Cover
Insert a new Lance firmly into the Lancet Carrier
3
A Twist off the lancing device cover.
Insert a new lancet firmly into the
lancet carrier.
B Hold the lancet needle cover and
gently twist it off of the lancet.
Continued on next page
IMPORTANT SAFETY PROCEDURES:
Use a new lancet every time you test to avoid contamination. It is important
to discard the used lancet carefully after each use to avoid unintended lancet
stick injuries. Used lancets may be considered biohazardous waste in your
area. Be sure to follow your local regulations for proper disposal.
If alcohol wipes are used to clean your finger or palm, make sure your skin is
dry before the blood sample is obtained.
Lancing devices should NEVER be used by more than one person.
A
B
Page 22
TELCARE™ Blood Glucose Monitoring System
18
C Replace the lancing device cover.
The adjustable tip offers 5 levels of skin
penetration. To select the desired depth,
twist the adjustable tip in either direction
until the number lines up with the arrow.
To select the best depth:
1-2 for soft or thin skin
3 for average skin
4-5 for thick or callused skin
D Hold the lancing device cover in one
hand. Using the other hand, slowly
pull the sliding barrel away from the
lancing device cover. You will hear a click,
indicating that the lancet carrier is locked
into position. Release the sliding barrel to
return it to its original position.
NOTE:
• For ngertip testing, use the blue or grey cap included on the lancing device.
• For alternate site testing on your palm, use the clear cap.
Twist off the Lancing Device Cover
Insert a new Lance firmly into the Lancet Carrier
The Adjustable Tip offers 5 levels of skin penetration. To select the desired depth, twist the Adjustable Tip in either direction until the number line up with the Arrow.
Replace the Lancing Device Cover
3
6
Hold the Lancet Needle Cover and gently twist off of the Lancet
The Adjustable Tip offers 5 levels of skin penetration. To select the desired depth, twist the Adjustable Tip in either direction until the number line up with the Arrow.
Replace the Lancing Device Cover
To select the best depth: 1-2 for soft or think skin 3 for average skin 4-5 for thick or calloused skin
6
Place the Lancing Device aside and wait a few seconds for a blood drop to form. To help form a blood drop, lower the hand to waist level and gently massage the finger from palm to fingertip.
Position the end of the Adjust- able Comfort Tip against the side-tip of the finger. Press the Trigger Button, and then lift the Lancing Device away from the finger after the puncture is complete.
Hold the Lancing Device Cover in one hand. Using the other hand, slowly pull the Sliding Barrel away from the Lancing Device Cover. You will hear a click, indicating that the Lancet Carrier is locked into position. Release the Sliding Barrel to return it to its original position.
C
D
Page 23
TELCARE™ Blood Glucose Monitoring System
19
5
Fingertip Testing
Position the end of the adjustable
comfort tip against the side of the
fingertip. Press the trigger button,
and then lift the lancing device away
from the finger after the puncture is
complete.
TIPS FOR FINGERTIP TESTING
• Gently massage your hand and nger toward the puncture site to form
a drop of blood. Do not “milk” or squeeze around the puncture site.
• Lance the side of your ngertip to avoid soreness. To avoid calluses,
choose a different lancing site each time.
Note:
DO NOT re-use the Lancet
The Adjustable Tip offers 5 levels of skin penetration. To select the desired depth, twist the Adjustable Tip in either direction until the number line up with the Arrow.
Replace the Lancing Device Cover
To select the best depth: 1-2 for soft or think skin 3 for average skin 4-5 for thick or calloused skin
6
Place the Lancing Device aside and wait a few seconds for a blood drop to form. To help form a blood drop, lower the hand to waist level and gently massage the finger from palm to fingertip.
Position the end of the Adjust- able Comfort Tip against the side-tip of the finger. Press the Trigger Button, and then lift the Lancing Device away from the finger after the puncture is complete.
Note:
DO NOT re-use the Lancet
The Adjustable Tip offers 5 levels of skin penetration. To select the desired depth, twist the Adjustable Tip in either direction until the number line up with the Arrow.
To select the best depth: 1-2 for soft or think skin 3 for average skin 4-5 for thick or calloused skin
6
Place the Lancing Device aside and wait a few seconds for a blood drop to form. To help form a blood drop, lower the hand to waist level and gently massage the finger from palm to fingertip.
Page 24
TELCARE™ Blood Glucose Monitoring System
20
Testing Using an Alternate Test Site
Your Telcare BGM™ is capable of testing
your blood glucose from your palm,
usually referred to as an alternate site
test, or AST.
Alternate site testing can be less painful
than fingertip testing. However, because
of the physiological differences between
your fingertip and palm, AST results may
be significantly different.
You should consult with your doctor
before using alternate site testing.
Important Information on Alternate Site Testing
Page 25
TELCARE™ Blood Glucose Monitoring System
21
Testing on Your Palm
Prepare the lancing device with a new lancet and clear cap. Cock the lancing
device so that it is ready to use. Select a site on your palm that does not have
veins, moles, excessive hair, or bone. Vigorously rub the side of your palm with
your other hand for at least 10 seconds. You should feel it get quite warm.
Press the opening of the clear cap against a fleshy area of the palm. Press the
lancing device up and down in a slow pumping motion for several seconds to
help the flow of blood.
Keep steady pressure on the area and press the release button. Do not be afraid
to apply pressure to the area with the lancing device to assist the flow of blood.
Continue to press the cap against the area for a few seconds to allow blood to
come to the surface. You can see the blood drop through the clear cap. Do not
smear the blood when removing the lancing device. If the blood drop is too
small, the meter will show “Error: Not Enough Blood.” Insert a new strip and
obtain a new blood drop.
If the results from alternate site testing differ from how you feel, use fingertip testing instead.
user tip
Telcare users have found that a 26 gauge lancet or thicker (a number lower
than 26) is needed in order to obtain an adequate blood drop for palm testing.
Ultra-fine lancets that are adequate for finger testing will often fail to yield
enough blood for palm testing.
Page 26
TELCARE™ Blood Glucose Monitoring System
22
IMPORTANT INFORMATION FOR LANCING DEVICE
A Use the TELCARE
Lancing Device included in your TELCARE™ Blood
Glucose Monitoring System to obtain a blood sample. The minimum
blood sample volume is 0.8 microliters.
B Purchase an additional TELCARE™ Lancing Device by calling
TELCARE™ at (888) 666-4530 (9AM - 5PM EST, M-F).
C If using another type of lancing device, please refer to its user
manual.
IMPORTANT SAFETY PROCEDURES:
Use a new lancet every time you test to avoid contamination.
It is important to discard the used lancet carefully after each use to avoid
unintended lancet stick injuries. Used lancets may be considered biohazardous
waste in your area. Be sure to follow your local regulations for proper disposal.
Lancing devices should NEVER be used by more than one person.
Please see your lancing device’s user manual for further instructions.
Page 27
TELCARE™ Blood Glucose Monitoring System
23
Important Information on Alternate Site Testing
We recommend that Alternate Site Testing SHOULD NOT BE USED WHEN:
n You have hypoglycemic unawareness (not able to tell if you have low
blood sugar).
n You are within 2 hours of eating a meal, exercising, or taking medication.
n You will be operating machinery or driving a car.
n You are sick.
n You think your blood glucose is low.
n Your Alternate Site Testing results do not match the way you feel.
n You are testing for hyperglycemia.
n Your routine blood glucose results are fluctuating often.
n Alternate Site Testing should not be used for patients who are dehydrated,
hypotensive, in shock, critically ill, or in a hyperosmolar state.
n Alternate Site Testing measurements should never be used to calibrate a
continuous glucose monitor (CGM).
n Alternate Site Testing measurements should never be used for insulin dose
calculations.
Please see your lancing device’s user manual for further instructions.
Page 28
TELCARE™ Blood Glucose Monitoring System
24
Applying Blood to the Test Strip
Gently touch the test strip to the drop of
blood. The test strip will absorb the blood.
Keep the test strip in the blood drop until
you see the blood sample end of the test
strip fill.
Tagging Your Data
The screen will start to count down. After
6 seconds, your glucose level will appear
on the display screen. You will then have
time to select a Reading Tag (e.g., Before
Breakfast) from the list displayed. Use the
DOWN button to scroll to the correct tag.
Press the ENTER button to select that tag.
Your test result will be automatically stored
into the meter memory. If a reading tag
is not selected in the allotted time, it will
automatically be tagged as “No Selection.”
Remove the Test Strip
In order to send results to the Telcare server,
remove the test strip from the meter when
you see the screen shown here.
6
7
8
Page 29
TELCARE™ Blood Glucose Monitoring System
25
IMPORTANT
If you see “High” or “Low” displayed, your blood glucose level may be above
600 mg/dL or below 20 mg/dL. Wash your hands. Dry your hands completely.
Test again using fingertip testing, DO NOT test on your palm or forearm. If you
still receive the same message, call your caregiver immediately.
Sending Your Results
As soon as you remove the test strip, your
BGM™ will send the results to Telserve,
Telcare’s secure server. Results can be sent to
your caregiver’s electronic medical records
and to other locations. Other locations could
be your cell phone or computer, depending
on the setup instructions you gave Telcare.
Telserve data can be viewed at MyTelcare.com.
TWO BARS OR LESS OF WIRELESS SIGNAL
STRENGTH MAY RESULT IN UNRELIABLE
TRANSMISSION AND RECEPTION OF DATA. THE
STRENGTH INDICATOR IS LOCATED AT THE TOP
LEFT OF THE METER’S SCREEN.
9
Page 30
TELCARE™ Blood Glucose Monitoring System
26
OPTIONAL: After successful data
transmissions to Telserve, you may receive a
multiple choice question to answer. This may
be a question from your caregiver or relate
to reordering supplies. Use the UP/DOWN
buttons to select the appropriate answer.
Press the ENTER button to transmit.
CLINICAL RESPONSE SCREEN: After
data transmission, you will see a graph
that compares your reading to your set
target. To change your set target, go to
MyTelcare.com or call Customer Support.
You may also see helpful guidance from your
caregiver regarding this reading to manage
your diabetes.
Access your readings at www.MyTelcare.com. Use the login credentials that you can obtain by calling (888) 666-4530. For additional questions or login support, turn your meter off and contact Telcare by phone or e-mail
support@MyTelcare.com.
Page 31
TELCARE™ Blood Glucose Monitoring System
27
If Your Data Transmission Is Not Successful
If the device cannot communicate with Telserve, the following message will appear:
“No data connection, Data will be stored”
Your test results will be stored in your device. You can send the reading on your own by going into Memory Recall within Settings. See page 7.
IF DATA TRANSFER FAILS, YOU MAY USE THIS PRODUCT LIKE ANY OTHER GLUCOSE METER. FOLLOW YOUR HEALTHCARE PROFESSIONAL’S INSTRUCTIONS ON TREATMENT.
Page 32
TELCARE™ Blood Glucose Monitoring System
References
1. American Diabetes Association. Standards of Medical Care in Diabetes—2008. Diabetes Care. 2008;31:S12–S54.
28
Understanding Your Blood Glucose Test Results
Blood glucose value will vary depending on food intake, medication, health, stress, and exercise. The ideal ranges for people without diabetes are as follows
1
:
n 70 to 100 mg/dL before meals, and n 70 to 140 mg/dL after meals
What to do if you get a high or low reading AND you feel ill.
If the Telcare BGM™ displays results that are “High” or “Low,” or you get a result that is more than 250 mg/dL or below 50 mg/dL AND you feel ill:
n Treat your diabetes according to the instructions from your doctor.
If the Telcare BGM displays results that are “High” or “Low,” or you get a result that is more than 250 mg/dL or below 50 mg/dL AND you DO NOT feel ill:
n Test your Telcare BGM with a control solution, refer to page 8, ”Step by
Step Guide to Control Testing.”
n Wash your hands, dry thoroughly and test again using your fingertip.
If you still get a high or low reading, contact your doctor.
Page 33
TELCARE™ Blood Glucose Monitoring System
Viewing Your Meter’s Memory
29
Your Telcare BGM™ Blood Glucose Meter stores up to 300 test and control results. Each result is stored with date and time.
Power on the TELCARE BGM device by
using the PWR button.
Use the UP/DOWN keys to access
“Settings Mode.”
Select “Memory Recall” from the
“Settings Mode” screen.
Use the UP/DOWN keys to scroll through
the stored readings.
To exit the Memory Recall menu, you
must scroll to the bottom of your reading
history and select “BACK.”
(Shortcut: Press the UP button when your
most recent reading is highlighted. Doing
so will display the Back option.)
1 2
3
4
Page 34
TELCARE™ Blood Glucose Monitoring System
Cleaning & Disinfecting
30
What is the difference between cleaning and disinfecting?
Cleaning and disinfecting are not the same thing. Cleaning simply removes blood
and dirt. Disinfecting kills germs and provides an extra level of safety. In the following
procedure, the first disinfectant towel is used for cleaning, and the second towel
provides for disinfection.
You should disinfect areas where there are both high concentrations of germs
and a possibility that they will be spread to others. The Caltech DISPATCH Cleaner/
Disinfectant with Bleach Premoistened Towels™ have ingredients that clean and
disinfect your meter, lancing device, and endcaps. While surfaces may look clean,
many infectious germs may be lurking around. Some germs can live on surfaces for
hours and even for days! The disinfection process shown in the pages that follow is
effective in destroying dangerous Hepatitis B viruses.
If the meter or lancing device is being operated by a second person who is
providing testing assistance to the user, the meter should be decontaminated prior to
use by the second person. See your lancing device’s user manual for instructions on
decontaminating your lancing device.
To acquire canisters of Caltech DISPATCH towels, please contact: Medex Supply - www.medexsupply.com - (888) 433-2300 Sears - www.sears.com - (800) 349-4358 Berktree - www.berktree.com - (866) 505-5333 Optics Planet - www.opticsplanet.com - (800) 504-5897 Amazon.com - www.amazon.com A full list of authorized dealers of Caltech DISPATCH towels is at www.MyTelcare.com.
Page 35
TELCARE™ Blood Glucose Monitoring System
31
Clean and disinfect your meter at least 1 time per week using Caltech DISPATCH Cleaner/Disinfectant with Bleach Premoistened Towels
using the
following method:
1. Dispense the DISPATCH towel from its package or
container.
2. Clean the outside of the meter to remove any blood or
dirt.
3. Discard used towel in a sealed container where it will not
be touched by others.
4. Dispense a second DISPATCH towel from its package or
container.
6. Discard used towel in a sealed container where it will not
be touched by others.
7. Wait at least one minute and make sure device is
completely air dried before next blood test.
5. Disinfect the meter with the second towel by wiping all
outside surfaces.
Cleaning and Disinfecting Your Meter
12
6
9 3
Page 36
TELCARE™ Blood Glucose Monitoring System
32
n Do not allow cleaning solution to run in or around the strip port. Doing so
may cause a malfunction.
n Do not let the meter come in contact with water.
n Do not use any glass or household cleaners on the meter.
n IMPORTANT: If the meter is being operated by a second person who
is providing testing assistance to the user, the meter should ALWAYS be decontaminated prior to use by the second person.
n Stop using the meter and contact customer service for a replacement meter
device immediately if you notice any of the following signs of deterioration:
- The meter does not turn on.
- The meter’s screen has been cracked or has become clouded.
- Any buttons have fallen off or stopped working.
For technical support, contact Telcare at (888) 666-4530.
The Telcare BGM™ Blood Glucose Meter has been validated for 1,825 cleaning and disinfection cycles. Please obtain a new meter after cleaning and disinfecting the meter 1,825 times or once the lifetime of the meter (5 years) has been reached, whichever comes first.
Page 37
TELCARE™ Blood Glucose Monitoring System
Cleaning and disinfecting of your lancing device should be done with the Caltech
DISPATCH Cleaner/Disinfectant with Bleach Pre-moistened Towels when it appears
dirty or contaminated with blood. The lancing device should be cleaned and
disinfected daily. The Telcare Lancing Device has been validated for 1,825 cleaning
and disinfection cycles, which corresponds to daily cleaning over a five year lifetime.
Please obtain a new lancing device after five years of daily cleaning or if signs of
deterioration are seen. Please follow this cleaning method:
1. Dispense the DISPATCH towel from its
package or container.
2. Clean the outside of the lancing
device and the inside and outside of the
cap to remove any blood or dirt.
3. Discard used towel in a sealed
container where it will not be touched
by others.
33
Cleaning and Disinfecting Your Lancing Device
Continued on page 34
4. Dispense a second DISPATCH towel
from its package or container.
Page 38
TELCARE™ Blood Glucose Monitoring System
34
5. Wipe all surfaces again, just as you
did in step 2, in order to disinfect
them.
6. Discard the second towel in a
sealed container where it will not be
touched by others.
7. Wait at least one minute. Then
rinse briefly with water and allow to
dry before next use.
IMPORTANT: If the lancing device is being operated by a second person who is providing testing assistance to the user, the lancing device and endcap should ALWAYS be decontaminated prior to use by the second person.
Stop using the lancing device and contact customer service immediately for a replacement if you notice any of the following signs of deterioration:
-Any buttons have fallen off or stopped working.
-The device is cracked.
-The device is discolored
-The device does not yield an adequate blood drop.
Page 39
TELCARE™ Blood Glucose Monitoring System
Charging Your Change to Telcare BGM™ Blood Glucose Meter
35
Your Change to Telcare BGM™ Blood Glucose Meter uses a rechargeable battery. It can be charged and discharged hundreds of times, but eventually the battery will lose its charge. If you notice significant reduction in the duration of your meter’s operating time, your battery may need replacing. Please contact Telcare technical support at (888) 666-4530, or e-mail
support@MyTelcare.com.
If left unused, a fully charged battery will lose its charge over time.
Always charge indoors.
To Charge Your Device
n Plug the power cord into an
electrical outlet.
n Plug the other end of the cord
into the BGM’s Charging Port.
Note:
If your device does not turn on when you switch the power button on, your
battery may be weak. Charge the battery. If after 5 hours of charging the
device does not turn on, contact Telcare technical support at (888) 666-4530.
Important: Do Not Test Blood Glucose While Charging the Device or for 30 Minutes After Charging Is Complete.
Page 40
TELCARE™ Blood Glucose Monitoring System
Storing Your Telcare BGM™ and Test Strips
36
n Your meter is a delicate electronic device that should be handled with care.
Severe shock, such as dropping the meter, could damage it.
n The meter and test strips only work in the temperature range of 50˚F to 104˚F.
n Store the test strips between 39°F and 86°F. Do not freeze. See the test
strip instructions for more details.
n Don’t leave the meter in a place that is extremely hot or extremely cold
(e.g., anywhere near a heat source or in an extremely hot or cold car).
n Do not expose the meter or test strips to high levels of humidity, such as in
the bathroom or kitchen.
n Immediately close the bottle cap tightly after removing a test strip.
n Do not attempt to open the meter. Opening the meter will void the
warranty and may cause it to operate improperly. If you have technical problems or questions, turn your meter off and contact Telcare at (888) 666-4530. Support is available Monday through Friday, 9AM - 5PM EST.
Page 41
TELCARE™ Blood Glucose Monitoring System
Troubleshooting
37
Error Codes
In the event your device experiences an error, you will see the following screen along with an error message.
1. Error Message: Meter Failure
What it means: As with any computer device, sometimes the meter may need to
be restarted.
Action to take: Restart meter and contact Telcare support if the problem
continues. Turn your meter off before calling Telcare at (888) 666-4530.
2. Error Message: Used Strip Inserted
What it means: Your test strip may be used or damaged.
Action to take: Repeat the test with a new test strip. Remember, add the sample
only when you see the prompt on the display screen. If you believe you
received a damaged test strip, return it to Telcare for a free replacement.
Page 42
TELCARE™ Blood Glucose Monitoring System
3. Error Message: Not Enough Blood
What it means: There is not enough blood or too much blood or control solution
on the test strip to give a reliable reading.
Action to take: Remove the test strip and repeat the test with a new test strip.
4. Error Message: Temp. Too Low / Temp. Too High
What it means: The meter is too hot or too cold.
Action to take: Move the meter into an area that is between 50˚F and 104˚F.
IMPORTANT: After receiving this error, allow the meter to be at room temperature for at
least 30 minutes before testing blood or control solution.
38
If your meter displays a “Low Battery” icon, charge the battery.
Other Messages
High / Low Displayed
If you see either High or Low displayed on
your device, your blood sugar level may be
out of range (either in excess of 600 mg/dL
[High] or lower than 20 mg/dL [Low]).
Page 43
TELCARE™ Blood Glucose Monitoring System
39
Communication Errors
If you are having trouble uploading your results to Telserve, you may want to try the following:
n If testing indoors, move closer to a window for better cellular reception.
n Test outdoors.
n Check to see if there is cellular coverage in your area. Two bars or less of
wireless signal strength may result in unreliable transmission and reception of data.
n Restart meter and send reading within Memory Recall (see page 7).
If the above actions do not resolve your problem, turn your meter off and then contact Telcare technical support at (888) 666-4530, or e-mail
support@MyTelcare.com.
5. Error Message: No data connection, Data will be stored
What it means: The meter cannot find a cellular connection.
Action to take: See “Communication Errors” below.
Page 44
TELCARE™ Blood Glucose Monitoring System
Terms of Warranty
40
Telcare warrants the Telcare BGM™ Blood Glucose Meter to be free of defects in workmanship and materials under normal use for a period of one (1) year from the date of purchase.
The liability of Telcare is limited to repair or replacement. In no event shall Telcare be liable for any collateral or consequential damages or loss.
Instruments subjected to misuse, abuse, neglect, exposure to water, unauthorized repair, or modification are excluded from this warranty.
This guarantee specifically excludes expendables and consumables.
Please direct any questions or claims to the authorized dealer from whom you purchased your system, or to:
Telcare, Inc. Bethesda, MD 20814 USA Phone: (888) 666-4530
This warranty applies only to the original purchaser of the system.
Page 45
TELCARE™ Blood Glucose Monitoring System
Specifications
41
Test Strips: Telcare Test Strips
Minimum Sample: 0.8 microliters
Test Range: 20~600 mg/dL
Calibration: Plasma
Hematocrit Range: 30~55%
Display Type: Color LCD screen
Memory: 300 blood test results with date and time
Result Averaging: 7, 14, and 30 days averaging displayed on Telserve Data System
(not on TELCARE BGM™)
Dimension: 100L x 60W x 15H (mm)
Weight: 115g
Battery: Rechargeable, Lithium Ion 1100 mA hour battery, permanently installed
Battery Life: 3 years
Automatic Sleep Mode: After 2.0 minutes of nonuse
Operating Temperature: 50°F~104°F
Relative Humidity: Less than 85%
Storage Condition: Meter at 39°F~122°F, Test Strips at 39°F~86°F
Page 46
TELCARE™ Blood Glucose Monitoring System
About Telcare
42
TELCARE™ is committed to creating the very best medical device experience for
patients and doctors. Connecting powerful hardware and innovative software via
a “no wires” approach, Telcare’s offerings aim to provide convenient and effective
disease management tools.
The TELCARE™ Blood Glucose Monitoring System won the People’s Choice Award at
CTIA Wireless 2009, Best in Show at MobiTechFest 2009, was an E-tech Finalist at
CTIA Wireless 2010, and won the First Prize 2011 CTIA E-Tech Award for Enterprise
Healthcare Solution and the First Prize 2011 CTIA E-Tech Award for Mobile Healthcare
Solution.
Page 47
TELCARE™ Blood Glucose Monitoring System
43
Contact Telcare
Telcare, Inc.
Bethesda, MD 20814 USA
Phone: (888) 666-4530
Hours: M - F, 9AM - 5PM EST E-mail: support@MyTelcare.com Website: www.MyTelcare.com
Federal Communication Commission Interference Statement
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 18 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
n This device complies with Part 18 of the FCC Rules. Operation is subject to the following
two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
n FCC Caution: Any changes or modifications not expressly approved by the party
responsible for compliance could void the user’s authority to operate this equipment.
FCC Radiation Exposure Statement
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with minimum distance 20cm between the radiator & your body.
This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.
Please contact your healthcare professional at all other times.